Metrology And Measurements Laboratory

CALIBRATION OF VERNIER CALIPER


Experiment No: 1
Date:
Aim:
To calibrate the given vernier caliper using slip gauge and
draw the calibration graph and error graph.
Apparatus required:
1. Slipgauge set
Make
:
Grade
:
Material:
Size
:
S.No.
Range (mm) In steps of No.
(mm)
pieces

of Total no. of
pieces

2. Vernier caliper
Make
:
Range
:
Least count:
Principle:
The principle of vernier is that when two scales or divisions
slightly different in size are used, the difference between them
can be utilized to enhance the accuracy of measurement.

United Institute of Technology

Department of Mechanical Engineering

Metrology And Measurements Laboratory

1. Outside jaws: used to measure external diameter or width of an object


2. Inside jaws: used to measure internal diameter of an object
3. Depth probe: used to measure depths of an object or a hole
4. Main scale: scale marked every mm
5. Main scale: scale marked in inches and fractions
6. Vernier scale gives interpolated measurements to 0.1 mm or better

United Institute of Technology

Department of Mechanical Engineering

Metrology And Measurements Laboratory

7. Vernier scale gives interpolated measurements in fractions of an inch


8. Retainer: used to block movable part to allow the easy transferring of a measurement

Least count:
Value of 50 V.S.D
Value of 1 V.S.D
Value of 1 M.S.D
Least count
=

=
49mm
=
0.98mm
=
1mm
=
1 M.S.D- 1 V.S.D
0.02mm

Procedure:
1. The slip gauge, vernier caliper are cleaned using

acetone

and lint free cloth.


2. The zero error of the given vernier is checked (if there is
any error, zero correction must be accounted while
measuring)
3. The least count of the vernier caliper is calculated.
4. Slip gauges are held between the vernier caliper jaws.
5. Main scale reading (MSR) and the vernier scale reading
(VSR) are noted. The value measured is MSR+ (VSR*LC)
6. The above procedure is repeated for various standard
sizes.
7. The error is found out by comparing the standard value
(Slip gauge value)and vernier caliper reading.
8. The calibration graph is drawn with the standard reading
on X-axis and indicated reading on Y-axis
9. Error graph is drawn by taking indicated reading along Xaxis and the error value along Y-axis.
Observations:
Zero error
=
Zero correction =
United Institute of Technology

Department of Mechanical Engineering

Metrology And Measurements Laboratory

LC of the Vernier caliper =

S.N
o

Standar Vernier caliper reading


d value (mm)
(mm)
(2)
(1)
Trial
1

Trial 2 Trial
3

Sample calculation:
Standard reading in mm(S)
Measured reading in mm (I)
Error in mm (I S )

Error
(mm)
(2)

(1)

Trial 4

Trial 5

Mean

=
=
=

Result:
The given vernier caliper is calibrated and the following graphs
are drawn.
1. Calibration graph
2. Error graph
United Institute of Technology

Department of Mechanical Engineering

Metrology And Measurements Laboratory

Inference:

United Institute of Technology

Department of Mechanical Engineering