Escolar Documentos
Profissional Documentos
Cultura Documentos
Chromatography
What is HPLC ?
H: High
P : Performance (Pressure)
L : Liquid
C : Chromatography
Figure 1
Stationary Phases
Normal-phase (NP) chromatography generally uses
the same types of mobile phase as for the adsorption
mode. The difference, however, is the nature of the
stationary phase. In NP chromatography, the packing is
silica gel that has been bonded with a polar phase. The
usual polar phases widely available from many
manufacturers include cyano, amino, nitro, and diol
phases.
Mobile Phases
A variety of solvent can be used for adsorption
chromatography, ranging non-polar to polar.
Adsorption chromatography functions very well
for the separation of non-polar and moderately
polar compounds using moderately polar
eluents. It does not function well with high
polarity eluents.
example
In a normal-phase partition column, a solute
was found to have a retention time of 29.1 min, and
an unretained sample had a retention time of 1.05
min when the mobile phase was 50% by volume
chloroform and 50% n-hexane.
Calculate (i) k for the solute and (ii) a solvent
composition that would bring k down to a value of
about 10.
Reversed-phase
chromatography
Neutral Compounds
Stationary Phases
Historically, silica gel was the most common material
used in the early development of column liquid
chromatography (LC).However, silica is a polar material
that contains hydroxyl groups (silanols) that are both
acidic and strongly hydrogen-bonding in character.
38
Mobile Phase
The mobile phase generally consist of mixtures of water or
aqueous buffer solutions with various water-miscible
solvents.
The strength of binary mixtures is not a well defined
function of %B but depends on analyte and stationary
phase properties. Nevertheless, it is possible to give
numerical data which allow to obtain a good
approximation of mobile phase composition when it is
necessary to try more than one solvent in order to find the
best selectivity.
Example
In a reversed-phase column, a solute was found to have
a retention time of 31.3 min, and an unretained species
required 0.48 min for elution when the mobile phase
was 30% (by volume) methanol and 70% water. Calculate (a) k and (b) a water-methanol composition that
should bring k to a value of about 5.
Solution
Problem
If a separation with 70% of methanol gives
adequate retention but poor selectivity, which
other solvent mixtures could be tried?
IONIC SAMPLES:
REVERSED-PHASE &
ION-PAIR
CHROMATOGRAPHY
55
Ionizable Compounds
analysis by RP
57
58
59
61
62
64
65
Choice of Buffers
Whenever acids or bases are separated, it is necessary
to buffer the mobile phase in order to maintain a
constant pH and reproducible retention during the
separation.
In selecting a buffer for RPC separation, several buffer
properties may prove relevant:
69
Factors to be considered
pKa and buffer capacity
solubility
UV absorbance (when UV detection is used)
volatility (when mass-spectrometric or
evaporative light-scattering detection is used)
ion-pairing properties
stability and compatibility with the equipment
Imad Abu Reid
70
71
72
73
ION-PAIR CHROMATOGRAPHY
(IPC)
74
75
76
77
78
Basis of Retention
Two possible retention processes or
mechanisms exist for separation by IPC.As an
example, we will use the ion-pairing of an ionized
acidic solute A by a tetraalkylammonium IPC
reagent R+. The ion-pairing of a protonated basic
Solute B+ by an alkylsulfonate IPC reagent R can
be described similarly.
Imad Abu Reid
79
80
81
82
83
84
85
86
87
DETECTION
90
92
93
2. Sample-Specific Detectors
Some characteristic of a sample is unique to that sample,
or at least is not common to all analytes, and the samplespecific detector responds to that characteristic. The UV
detector is the most commonly used sample-specific
detector. It responds to compounds that absorb UV light
at a specific wavelength. Other common sample-specific
detectors rely on the ability of an analyte to fluoresce
(fluorescence), conduct electricity (conductivity), or react
under specific conditions (electrochemical).
Imad Abu Reid
94
3. Hyphenated Techniques
Hyphenated techniques refer to the coupling of an
independent analytical instrument to the HPLC system
to provide detection, and often are abbreviated with a
hyphenas LC-(plus the technique). The most common
hyphenated technique is LC-MS, where a mass
spectrometer is coupled with an HPLC system. Other
less widely used techniques are LC-IR or LC-FTIR and LCNMR.
Imad Abu Reid
95
I. UV Detector
UV-Vis detectors are the most commonly used detectors in
HPLC. Solutes which absorb UV or visible radiation (typically
190 - 600 nm) can be detected. The degree of absorption is a
function of the molar absorptivity of the sample molecule, the
path length of the detector flow cell and the solute
concentration. The solute concentration is directly
proportional to the absorbance allowing quantification. UVVis detectors can routinely achieve detection of only a few
nanograms. They have a large linear dynamic range and are
very robust.
96
97
98
99
diode-array UV detector
100
101
102
103
104
105
III. ELECTROCHEMICAL
(AMPEROMETRIC) DETECTORS
Many compounds that can be oxidized or reduced in the
presence of an electric potential can be detected at very low
concentrations by selective electrochemical (EC) measurements.
By this approach the current between polarizable and reference
electrodes is measured as a function of applied voltage. Because
a constant voltage normally is imposed between the electrodes,
and only the current varies as a result of solute reaction, EC
detectors are more accurately termed amperometric devices. EC
detectors can be made sensitive to a relatively wide variety of
compound types.
Imad Abu Reid
106
107
108
109
110
111
112
113