Instruction for REX console


Parameter setting

This information has been drafted to give our engineer basic structure on
parameters of REX control console.

All of them will be set in factory, so you

do not need to adjust them as long as the factory permit adjustment.

That

means this information is not for your maintenance service.


However, please keep this information in case you should adjust parameter
under the factorys permission.

In any case, please contact factory for any

problem first to get exact solution.


LISTEM Factory, R&D center

How to enter into setting mode


After Power on the system, press any key to enter

If you want to set


system Parameter?
-> Press any

into Parameter Setting mode.Console.

Top menu
(1) REX_mode : Setting for each REX Series

Please select Mode


>REX_ Mode Set para
Tube,Vsrc
Spare
Check_key
Calibrat.

model
(2) Tube,Vsrc : Setting for Tube model and
Line power voltage
(3) Check_Key : Console KEY Checking
mode (Key input Error check mode)

(4) Set para : Setting mode for several options in each REX model
(5) Calibrat : Calibration Mode
NOTE - ESC: Return to previous mode
SET: Selection and Saving

REX_Mode------------------------------------------------------------------------------(1)

Please select Mode


>REX_ Mode Set para
Tube,Vsrc
Spare
Check_key Calibrat.

Select REX_Mode of Right LCD


to select Model which you are
currently using.

(2)

And then press SET button to


enter into REX_Mode.

> REX-325R REX-325RF


REX-525R REX-525RF
REX-550R REX-550RF

(1)

Please select the model name which you are


currently using.

You can see the model

name on the rear side of control console.


(2)

To move to other model name, you should

press following buttons.

Up
Down

(3)

After you select the model name, press SET button. Then, Model name will be saved.

(4)

Press ESC to escape from REX_Mode.

Tube, Vsrc mode----------------------------------------------------------------------Tube Selection

Please select Mode


REX_ Mode Set para
>Tube,Vsrc
Spare
Check_key Calibrat.

Select Tube Type !


> Set Tube1 Type(R)?
Set Tube2 Type(F)?
Source Voltage?

(1)

Select Tube,Vsrc

(2)

Press SET button to enter into


this mode.

(1)

Select the tube type which you want.

(2)

R system Only Tube 1 should be


set.

(3)

RF system Tube 1 and 2 and


should be set.

(4)

To set tube model in R system,

press SET Tube 1 Type(R)? and Press SET button to enter into this mode.
(1)

E7204
>E7239
E7242
E7252

E7813
E7255

LTN-25(Listem) = E7239
(Toshiba)

(2)

LTN-50(Listem) = E7252
(Toshiba)

(3)

In case of 10kW system,

the tube should be E7204.


(4)

After select tube unit, press SET to save it.

(5)

In case of RF system, perform same manner to select tube 2.

Source voltage selection

Select Tube Type !


Set Tube1 Type(R)?
Set Tube2 Type(F)?
>Source Voltage?

220VAC
> 380/400VAC

(1) Select Source Voltage? to select your


line power.
(2) Press SET button to enter into this
mode.

(1) Select your source power.


(2) Press SET button to save it.

Check_Key mode----------------------------------------------------------------------(1)

Please select Mode


REX_ Mode
Set para
Tube,Vsrc
Spare
>Check_key Calibrat.

Select Check Key and press


SET button to enter into this
mode.

(2)

You can check whether there is


problem in console key
operation. When you press the
key, you may see the name of

the key in Right LCD.


(3)

For example, when you operate RkV UP button, you may see the name of the key in
Right LCD.

Please Push Button


Return ; <Esc> key
RkV UP

If there is no problem, press ESC to escape from the mode.

Set Para mode-------------------------------------------------------------------Select Set Para menu and press SET

Please select Mode

button to enter into this mode.

REX_ Mode >Set para


Tube,Vsrc
Spare
Check_key Calibrat.

>1) Rotor Type

2)Rotor REF.V
3)Rotor Delay
4)Rotor Check

55
30
1

To move to other sub-menu, use the following keys.


UP
DOWN

SET : Enter into the menu and saving.


ESC : Move to previous menu.
Rotor Type

>0:Nomal Speed
1:high Speed Rotor

(1)

Select the Rotor Type.

(2)

Standard(Normal) : 0
(REX-125,325,525)

(3)

High speed Rotor : 1


(REX-550,650,850)

(4)

Press SET to save it and ESC


to escape from the mode.

Rotor REF.V
(1)

Set rotor Ref Volt


Scale(255=5[V])
Rotor REF V :

To adjust Rotor output level voltage, use


following keys. (Factory setting : 55 /
Default : 60)

55

Up
Down

(2)

Press SET(Save) and ESC(Escape)

Rotor Delay
(1) READY time adjustment of

Set Rotor Delay


Scale(10=1Sec)
Rotor REF. V :

Radiography.(Factory setting : 40 /
Default : 30)

40

(2) To adjust Ready time, use following


keys
Up

NOTE : If the ready time value is too low, mA can be measured lower than
setting value as it has not been pre-heated enough.

Down

Rotor Check

0:No Check Rotor


>1:Check Rotor RPM

(1)

Rotor Error check setting

(2)

Use the side Cursor to select the


mode.

(3)

Factory setting : 1(Check Rotor


RPM)

Check kVp
(1)

kVp exposure output Error


checking

>0:No Check kVp


1:Check kVp

(2)

Use the side cursor to select the


mode.

(3)

Factory setting : 0 (Currently no


use.)

Check Filament
(1)

Filament low voltage and over


voltage checking

0:No Check Filament


>1:Check Filament

(2)

Use the side cursor to select the


mode.

(3)

Factory setting : 1 (Check kVp)

(1)

AEC ion chamber setting in each

AEC Sensor

Select AEC Sensor


>Bucky0;
AEC Not
Bucky1;
AEC Not
Bucky2;
AEC Not

bucky unit.
(2)

Use the side cursor to select the


mode.

EX) If ion chamber should be used in Bucky 1,

Move to Bucky1 and press SET.

Select AEC Sensor


Bucky0;
AEC Not
>Bucky1;
AEC Not
Bucky2;
AEC Not

Use the side curse and move to 1; Ion

0; AEC Not

3;PM Tube

Chamber and press SET.

> 1; Ion Chamber


2; Photo D

Select AEC Sensor


Bucky0;
AEC Not
>Bucky1;
Ion Chamber
Bucky2;
AEC Not

(1)

Display will be changed as left picture.

(2)

As like above manner, you can select


AEC mode for each bucky unit.

(3)

This mode will be set at factory before


delivery.

Set Film Speed


(1)

Select Film Speed


0; Slow(200,220)
> 1; Medium(400,440)
2; Fast(600-800)

Film speed setting against


intensifying screen.

(2)

Use the side cursor to move to


each memu.

(3)

Factory setting : 1(Medium)

(4)

According to Film Speed,


exposure time for AEC will be
changed.

NOTE
If you select improper speed, AEC function can not be performed properly.

ABC Sensor

Select ABC Sensor


0; ABC Not
> 1; CCU
2; Photo Diode

(1)

Applied only for RF or C-arm unit


having

Automatic

Brightness

Control function.
(2)

Factory setting : 0(ABC by CCU)

ABC Ref.V

Set ABC Ref Volt


1- 255 (255 (5V))
ABC Ref. V :
160

(1)

ABC standard voltage setting

(2)

Use the following keys, and adjust the


value.
Up
Down

(3)

Factory setting : 160

ABC Filter

Set ABC Filter


1-255 (35 (100msec))
ABC Filter : 20

(1)

ABC sensing interval setting

(2)

Use the following keys, and adjust the


value.
Up
Down

(3)

Factory setting : 20

ABC Bound

Set ABC Bound


1-255 (255(5V))
ABC Bound :

(1)

ABC non-control time setting

(2)

Use the following keys, and adjust the


value.

50
Up

(3)

Down

Factory setting : 50

ABC Settle.
(1)

Set ABC Settleing


1-255 (35 (100msec))
ABC Set. :
250

ABC control start time setting in initial


fluoro.

(2)

Use the following keys, and adjust the


value.
Up

(3)

Factory setting : 250


Down

Max F_Time
(1)

Set Max. Flu. Time


1-10 Min
Set_FTime:
5(5Min)

Fluoro time setting (Max upto 10min.) in


RF system

(2)

Use the following keys, and adjust the


value.
Up

(3)

Factory setting : 5Min

(4)

30 second before setting time, the alarm will work.

Down

FkV Limit
(1)

Set FkV max(40-125)


Scale (kV)
FkV Max. V :

Maximum Radiography kVp setting in RF


system (Max upto 125kV)

(2)

120

Use the following keys, and adjust the


value.
Up

(3)

Factory setting : 120kVp

Down

16) PPS Ref

60
50
30
15

8
4
2
> No Pulsed

(1)

Setting for the number of frame in Pulsed Fluoro in RF system

(2)

Use the cursor key to select the number for frame.

(3)

Factory setting : No Pulsed (Currently no use)

17) Bucky Type

Select Bucky Type


>0:Motor Type
1:Spring Type

(1)

Bucky type setting

(2)

Use the cursor key to select the


bucky type.

(3)

Factory setting : 1 (Default : 0)

(1)

AEC setting in Spot Radiography

18) Spot Semi

of RF system.

Select Spot Semi


>0: Not Use
1: Use Spot Sem

(2)

Use the cursor key to select the


mode.

(3)

Factory setting : 0 (No use)

19) Spot Delay

Set Spot Delay


Scale(150=1sec)
Spot Delay : 50

(1)

Applied only for RF system

(2)

Spot delay time setting

(3)

Factory setting : 50

(4)

Use the following keys to adjust the value.


Up

NOTE : If the value is too low, mA can be lower than setting mA

Down

because of insufficient pre-heating time.

20) Serial_Com

Set Serial Com


> 0: Com_Single
1: com_Multi

(1)

Communication method setting

(2)

Use the side cursor to set this


mode.

(3)

Factory setting : Com Single

(1)

Applied only for Orthopedic RF

21) Ortopedia

Set Ortopedia
> 0: Nomal
1: Ortopedia
->1Tube(R, R/F)

system(LIBERO-X)
(2)

Use the side cursor to set this


mode.

(3)

Factory setting : 0.Normal (REX


system)

(4)

Factory setting :
1.Ortopedia(LIBERO-X)

22) Exhibit
(1)

Exhibit Program?
> 0: No(Nomal)
1: Yes(Exhibit)

Demo

unit

Exhibition(Console

setting
can

in
work

without generator)
(2)

Use the side cursor key to select


the mode.

(3)

Factory setting : 0

23) Cable Length

Set Cable length


1-12 Meter
Cable Length:

(1)

HV cable length setting

(2)

Use the following key to adjust the length.


Up

Down
(3)

Factory setting : 8

24) FmA Limit

Set FmA Max(2-30)


Scale 5(=1mA)
FmA Max.
:

(1)

Applied only for RF system

(2)

Max Fluoro mA setting

(3)

Use the following keys to adjust the value.

(4)

Factory setting : 20(4mA) / Default : 5

(5)

Maximum 30(6mA)

Up
Down

25) I/I Size

4.5 Inch
>6 Inch
9
Inch
12 Inch

(1)

I.I size setting

(2)

Applied only for RF

16 Inch

system
(3)

Use the side cursor to


select I.I size.

(4)

Factory setting : 6inch


or 9inch
(Default : 6inch)

26) Rewrite APR

APR Data Writing


APR Adress : xxx

(1)

Select the manu by pressing SET button

(2)

APR data will be set as initial value.

(3)

After rewriting, automatically escape from


this mode.

27) Rewrite Def Para

Def Para Writing

(1)

Select the manu by pressing SET button

(2)

All parameter setting value will return to


initial value.

(3)

NOTE : After rewriting, all parameter


should be adjusted properly again.

Calibration mode-----------------------------------------------------------------To enter into Calibration mode, please select

Please select Mode


REX_ Mode Set para
Tube,Vsrc
Spare
Check_key > Calibrat.

>RkV_RmA
FkV_FmA
Preheat
Fil_Limit

Calibrat by using side cursor key.


Press SET to enter into Calibration mode.

Density
Anatomic

Description of submenu
(1)

RkV_RmA : Radiography Calibration. In case of R system, reboot with RkV_RmA


and recalibrate Radiography factor using Tube 1 menu. In case of RF system, reboot
with RkV_RmA and recalibrate each Radiography factors using Tube 1 and 2 menu.

(2)

FkV_FmA : Fluoro Calibration. Only applied to RF system.

(3)

Preheat : Preheat voltage Calibration

(4)

Fil_Limit : No use

(5)

Density : Density Calibration of AEC. Only applied to AEC system

(6)

Anatomic : No use

Detailed information how to calibrate the system, please refer additional


instruction hereinafter.

CALIBRATION
Intorduction
This Calibration document provides information and procedures to perform all the
adjustments required to establish an optimal performance of this generator

Calibration procedures
NOTE Record all the calibration data in the Date Book
Previous to calibration, perform the following steps:
1. with the generator power OFF, connect :
Non-invasive kV Meter to measure kV
mAs Meter to the Test point on the HV Transformer to measure mA or mAs
NOTE

Test points on the HT controller Board can also be used to monitor the

kV and mA readings but should not be used to calibrate the unit.

These test points must be

checked with scope:


- mA test point is +mA or -mA and the scale factor is:
1Volt = 20mA
- kV test pint is +kV or -kV

and the scale factor is

1volt = 10kV
2. Power On Display Console Press NO(2) in the Figure 2
2.1 During the Programming process, it will show as below.
If you want to set
system Parameter?
-> Press any Key
(Twinkle Parameter)
Press the Any key(Button) in the Membrain. If Any key is selected , The Parameter display
will be as below.
Figure 3

Please select Mode


REX_ Mode
Tube,Vsrc

Rotor Type
Max FTime

Check_key > Calibrat.

Select " Calibrat. " and press SET.

And then, the Parameter display will be as below.

>RkV_RmA

Density

FkV_FmA

Anatomic

Preheat
Fil_Limit

Press the " SET " Button then it will be Returned to the Figure 3 .
Press the " ESC " Button then All of Configuration data Saved and Program will be
restarted for Recalibration automatically.
The calibration Parameter display will be automatically shown as below.
Left LCD Window(SettingValue)

Right LCD Window(Data Adjust Value)

3. Adjust the kVP , mA , Sec as it above Values

41kVP , 200mA , 0.05Sec

Time is fixed 0.05Sec automatically


4. Refer to the Calibration Data Sheet drafted by factory.

Calibration Operating Buttons


(1) kV Data Increment Button
(2) kV Data Decrement Button
(3) mA Data Increment Button
(4) mA Data Decrement Button
(5) EXP Data Increment Button
(6) EXP Data Decrement Button
5. Adjust No(1) and No(2) of Figure 3, and press Exposure SW while pressing Ready SW
in order to 41kVP in D-III or Oscilloscope.
6. Adjust No(3) and No(4) of Figure 3, and press Exposure SW while pressing Ready SW
in order to get 200mA in D-III or Oscilloscope.
7. Equalize the slope of mA pressing No(5) and No(6) of Figure 3,
( With oscilloscope,feedback wave can be obtained and the value of EXP mA should
be designated plus(+) )

8. Save the data by pressing " SET " Button.


9. According to the above mentioned instructions the edge of kV data each mA Steps.
Refer to the Table1 as below.
10. At First, make calibration each kV positions 17steps for 200mA.

For all 17steps,

please press SET button for each 17steps(in 200mA line) after adjusting value.

* Table 1 - Position of kV and mA(REX-525)


Position kV

mA step
25mA

41kV

81kV

121kV

50mA

41kV

81kV

121kV

75mA

41kV

81kV

121kV

100mA

41kV

81kV

121kV

150mA

41kV

81kV

121kV

200mA

41
kV

46
kV

51
kV

56
kV

61
kV

66
kV

71
kV

76
kV

81
kV

86
kV

91
kV

96 101 106 111 116


121kV
kV kV kV kV kV

250mA

41kV

81kV

121kV

300mA
400mA
500mA

41kV
46kV
51kV

81kV
81kV
80kV

121kV
100kV
80kV

11.After adjustment the value in 17steps(of 200mA line), perform re-calibration according to
the above position for mA value more than 200mA(250,300,400,500mA,), and when it finishes,
make calibration of the value less than 200mA(25,50,75,100,150mA).
Each mA Position and kV Positions of 3steps(ex.41-81-121kV) Data should be saved
separately by pressing SET button.

After saving all 3 steps separately(that means : after you

press SET button in 121kV), press


" SET " button with press Anatomic region " FOOT " button for auto-calibration between
41~81kV and between 81~121kVp.

< First press button >

< Second press button >

Between the Edges 41kV - 81kV - 121kV data will be automatically saved in
microprocessor.
12. When calibration for all the values data saving is over.
" Interpolating R _ Cal ----> Interpolated R _ Data " will be displayed automatically
in the Right window LCD.
13. When calibration for above procedure is over , Turn OFF the Switch

REX SERIES
TROUBLE SHOOTING

This instruction is to provide you with guide line of trouble shooting


flow on REX series.

LISTEM Corporation
1- REX T/S

DISPLAY TEST (1)

POWER SUPPLY
Connection

1) Put Demo IC into U1 on Main Console B/D.


2) Connect the power of Power supply to J5 on Main Console B/D.

* How to adjust RAD LCD DISPLAY


1) Adjust the brightness of characters in Left LCD(Small one) by
adjusting VR1 on LCD PCB. (Contrast Adjustment-Precise
adjustment required.)

RAD. LCD
DISPLAY Operation

2) Adjust the brightness of Left LCD(Small one) by adjusting VR4 on


LCD PCB. (Back Light Adjustment-Medium level adjustment
required.)
3) Adjust the brightness of characters in Right LCD(Big one) by
adjusting VR2 on LCD PCB. (Contrast Adjustment-Precise
adjustment required.)
4) Adjust the brightness of Right LCD(Big one) by adjusting VR3 on
LCD PCB. (Back Light Adjustment-Medium level adjustment
required.)

* How to adjust FLUORO LCD DISPLAY


1) Adjust the brightness of characters in Left LCD(Small one) by
adjusting VR1 on LCD PCB. (Contrast Adjustment-Precise
adjustment required.)

FLUORO. LCD
DISPLAY

2) Adjust the brightness of Right LCD(Big one) by adjusting VR2


on LCD PCB. (Contrast Adjustment-Precise adjustment
required.)
A) Check the condition of solder of LCD power line and mis-insertion
of circuit elements.
B) Replace LCD, U1,U2 and VR1,2,3,4

(A) RAD LCD DISPLAY

(B) FLUORO LCD DISPLAY

2- REX T/S

MEMBRANE TEST (2)

Switch Operation

1) Check the condition of Membrane switch by pressing each


switch on Membrane panel.
2) Check the condition of LED Lamp operation when pressing
each switch.
NO
A) Replace Membrane sheet after checking of switch condition.
B) Replace Main Console B/D.
C) Replce CN1,CN2 and FLAT CABLE

Repetitive Operation

1) Check again LCD PCB B/D.


2) Check again Main Console B/D.

<Radiography Console>

<R/F Console>

3- REX T/S

RELAY2 BOARD TEST (3)


1) Normal Operation of K10 onRELAY2 B/D.

Turn on the Power

NO

Turn off the Power


NO

A) Check if input voltage is 12VAC on J4-5,6


(K9,K10 Operation voltage)
B) Check if input voltage is 220VAC on J4-1,2.
C) Check if output voltage is 220VAC on J4-3,4
(Control Operation voltage)
D) Check mis-inserted elements.(BD1,TLP5604)
E) Replace K10 Relay.(12VAC)
1) Point of contact of K10 comes off when K9 on RELAY2 B/D
is operating.
A) Check the condition of point of contact of Membrane switchs
B) Replace K9 Relay(12VAC)
1) After operation of K8, K2 on RELAY2 B/D,
operating with K4 operation.

During Display
Transmitting
NO

K1, K2 on Main B/D is

A) Check if input voltage is +24VDC on J3-1,2.


(Relay operation voltage)
B) Check if MPOWER of J1-5 become GND.
C) Replace Main console B/D.
D) Replace CN1,CN2 & FLAT CABLE

1) When selecting BUCKY1 : operation of K4

2) When selecting BUCKY2 : operation of K5

3) When exposing X-ray : operation of K3

4) When power on : operation of K7

5) When switching from Small to Large : operation of K8


* When it is not working as the above : Output a signal from J7 on the H.T Control.
Usually it is output the approved voltage 24VDC.

When the above 1)~5) is selected on

Display Control, it is dropped to 0VDC(GND).


* Check whether U20(ULN2803), U14(82C55), U9(TIBPAL) and program IC U3(27C512) is operating
normally or not and, if necessary, replace them by new one.

4- REX T/S

RELAY1 BOARD TEST (4)

During DISPLAY
TRANSMITTTING

NO

1) Operating in order K3(Pre charge)---K1(Main)---K2(Charge) and


after then, K3 Relay comes off.
2) R-Series : K1(ON) MG1(ON) K3(ON) K2(ON) MG2(ON)
K3(OFF)
3) RF-Series : K1(ON), K3(ON) MG1(ON) K2(ON) MG2(ON)
* When K3 doesnt work : Check whether J1-5 has GND.
Check whether J8-5 on H.T Control has +24VDC and then, GND.
* When K1 doesnt work : Check whether J1-3 has GND.
Check whether J8-3 on H.T Control has +24VDC and then, GND.
* When K2 doesnt work : Check whether J1-4 has GND.
Check whether J8-4 has +24VDC and then, GND.
A) Check wiring condition of J1 on RELAY1 B/D.
B) Check Diode polarity & mis-insertion.
C) Check input voltage of J1(Check if J1-3,4,5 become GND.)
D) Check if input voltage is +24VDC on J1-1,2.
E) Check if polarity of J1-1,2 is reversed.

5- REX T/S

POWER BOARD TEST (5)

During DISPLAY
TRANSMITTTING

POWER BOARD
Voltage Adjustment

NO

1) Put GND of Tester device on CON1-1 on Power B/D.


2) Put another probe on the point you want to check.
3) Adjust VR7 so that the voltage between CON1-3 can be +5VDC.
4) Adjust VR2 so that the voltage between CON1-7 can be +12VDC.
5) Adjust VR4 so that the voltage between CON1-5 can be -12VDC.
A) Check input voltage of CON1.
(20VAC,0VAC,20VAC // 15VAC,0VAC)
B) Check the condition of F1,F2 Fuse in Power B/D.
C) Check if there is any misinserted elements.
D) Check variable point of VR1 & VR4 is located in center.
E) In case +5VDC,+12VDC,-12VDC voltage become low
after turning the power.
: Check the condition of element on Filament B/D, HT control
B/D and whether there is any short-circuits.

6- REX T/S

FILAMENT BOARD TEST (6)


Voltage measurement

1) Insert the GND of tester into J9-4.


2) Put the rest terminals of tester to J4-3(filament board)
3) When programming at display console, about +18VDCis output.
When finishing the program, about +16VDC is maintained
(According to the length of HT cable, the voltage is changed a little
by a basis on 200mA)

Wave measurement

4) Locate the GND terminal of scope on GND point in PCB and the rest
terminal on heat sink of Q1 and Q2(IRF630)
5) Adjust variable resistance R1 to 10KHZ.
6) adjust the variable resistance R4 and then save not to be shape of
spherical wave and peak

NO

A) Check if NF1 input voltage is 50 VAC


B) Check if D1 output voltage and J2-1,3 is +60VDC
C) Check if F1,F2 fuse is disconnected or not.
D) Turn the variable resistance R38 to the right(2/3 location)
E) Replace U3,4,5,(TL082)
F) In case the voltage between GND-J4 No.1 outputs +60~+70VDC
* Caution : DO NOT EXPOSURE
1> Check if Q7 is damaged, and then replace it.
2> Check if ARRAY TR PCB is defective, and then replace it.
3> Check if the PCB lag of Connector J4-1,2,3 has bad contact or
disconnection
4> Check if the HV cable contact connected to HV transformerTube-Change box

is bad (HV Cable Bushing should be

combined with HV transformer bushing tightly)

7- REX T/S

H.T CONTROL TEST (7)


1) LED1,2,3,4 is Off for a moment and On

Power On

(Indicates the state of CPU RESET)


2) When the program starts process, LED 1 is on a light.

NO
A) Check if the parts are inserted wrong.
B) Check if Q4(2N2907A),Q5(2N222A) is defective and replace it
C) Check the output signal wave in the No 15 pin of U1(80C552) is

IPM , FEED BACK


Voltage and frequency
measurement

okay and replace it.


D) Check if the power board output voltage(+5VDC,+12VDC,12VDC) is okay and the fuse is disconnected.
1) Set the frequency between GND-TP9 to 30KHZ by adjusting VR 1
(Tube 1)
2) Set the frequency between GND-TP9 to 10KHZ by

adjusting VR 2

(Tube 2- Fluoro)
3) Set the voltage between GND-TP2 to 3.0VDC by
(Tube 1)

8- REX T/S

adjusting VR 3

A) Check U38(TL082) and replace it


NO
1) Set the voltage between GND-TP17 to 1VDC(after READY

ROTOR signal Voltage


adjustment(ROTSEN)

status-LED3 on) by adjusting VR 13


2) Before READY(Rotor sensing) finish, make sure the voltage
should not be more than 5V.(approx.3V). After Rotor sensing,
the voltage will be approx.1VDC.
A) Check the voltage of TP17(after READY), and adjust VR13.

B) Check the CT1 operation of the main power board and


NO

the connector contact badness


C) Check the ROTOR input voltage (Ready 220VAC input /
after ready 60VAC input)

Filament Signal Voltage

Warning! : In case of High speed system such as

adjustment (FILSEN)

REX550,650,850, fix ROTOR input voltage to 220VAC)


D) Check if F7 fuse is disconnected
E) Check if the Rotor line W,B,G is disconnected and wrong
wiring
1) All process will be taken by CPU automatically, no need to adjust.
A) Check the U5(TL082) of filament board and replace it.
NO

B) U22-1,7 should be LOW

9- REX T/S

1) Adjust READY TIME in Rotor Delay setting mode of control

READY
Time Adjustment

console.(See Parameter setting mode section)


2) Adjust READY TIME of Spot Radiography in Spot Delay setting
mode of control console. (See Parameter setting mode section)

READY FINISH
1) When READY button lights on control console, ROTOR sensing is
finished.(Ready status-LED 3 on).

NO

A) Place a tester to TP17 to check ROTOR output voltage


B) And then, increase output voltage to 1VDC by adjusting VR13 during
operating READY button. (Counter-clockwise : Increase)

10- REX T/S

IPM DRIVE PCB, IPM GATE WAVE TEST (8)


1) Inset Drive PCB into IPM DRIVE PCB test tool

IPM DRIVE TEST

2) Connect J1,J2 connectors.


3) Put the tester on the end of GND and +15VDC and set + 15VDC by
adjusting VR1 and VR2 (In case of REX125 and 325, no need to adjust

NO

it. Different PCB applied for both models.)


4) Check if LEC is switching in turn by pressing switches.
A) When +15VDC is not be output, check D1,D2(DF06) U1,U2(LM317)
and replace it.
B) When LED is not be switching in turn, check PC1,2,3,4 and replace it.

C) Replace PC(HCPL4503) in the side that the switching is inferior

1) When Espousing, there must be feed back wave at GND and +KV,-

Repeat Operation

KV,+mA,-mA
2) If the wave is not measured, check it as follows.
A) The drive signal must be output at U38-9,10 as follows.

B) The drive signal must be output at U38-9,10 as follows.

11- REX T/S

When the A) wave is not be output at EXPOSURE


1) Check if KOUT,EXP,KSY signal is input
2) Check U38(TL594) and replace it.
NO

* When the B) wave is not be output at EXPOSURE


1) Check U37(AD201G) and replace it.
2) Check U41(ULN2004) and replace it.
3) Check IPM DRIVE PCB and replace it.
4) Check IPM and replace it.
5) Check HV transformer and replace it.

<HT control B/D>

12- REX T/S

TERMINAL FUNCTION AND VOLTAGE MEASUREMENT(9)


Function and Voltage output R-Series
1) TB2-1,2 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 1 and 2.
* When it doesnt work : follow the check instructions of Relay 2 Board Test.
2) TB2-3,4 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 3 and 4.
* When it doesnt work : follow the check instructions of Relay 2 Board Test.
3) TB2-5,6 : 24VAC voltage is output to operate the collimator.
* When it doesnt work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
4) TB2-7,8 : 30VDC LOCK voltage is output to operate Stand, Table and Bucky Stand.
* When it doesnt work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
Check out the operation of MG1. (LOCK voltage is approved with the point of contact of
MG1)
5) TB2-7,8,9,10 : 110VAC voltage is output to operate Stand and Table.
* When it doesnt work : Check out the operation of MG2. (LOCK voltage is approved with
the point of contact of MG2)
6) TB2-11,12 : It is supplied Power to operate the board of the optional photo-fluorography
and so on. (The voltage to operate photo-fluorographic relay on Interface board)
7) TB2-15 : It is Exposure Signal and for photo-fluorography.
8) TB2-16,17,18 : It is output the voltage to operate the rotor of tube.
* When it doesnt work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
9) TB2-19,20 : It is Temp Sensor and detects the overload of tube This is optional and
when connecting, it works normally. ( Usually the contact point of thermometer sensor in
the tube is connected to the rotor wire and when overloading of tube, it is treated the rotor
error.)

Function and Voltage output RF-Series


1) TB2-1,2,3 : It is output the voltage to operate the rotor of tube2.(Tube under the table
for Fluoroscopy)
* When it doesnt work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
2) TB2-4,5,6 : When selecting Tube 2, it is supplied Power to operate I.I, Power Supply,
Monitor, Camera and so on to be inflicted the voltage.
13- REX T/S

3) TB2-7,8 : It is Thermometer Sensor and detects the overload of tube This is optional
and when connecting, it works normally. ( Usually the contact point of thermometer sensor
in the tube is connected to the rotor wire and when overloading of tube, it is treated the
rotor error.)
4) TB2-9,10 : When selecting Bucky1 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 9 and 10.
* When it doesnt work : follow the check instructions of Relay 2 Board Test.
5) TB2-11,12 : When selecting Bucky2 on Display Console and pressing Ready, 110VAC
voltage is output between the terminal 11 and 12.
* When it doesnt work : follow the check instructions of Relay 2 Board Test.
6) TB2-13,14 : 24VAC voltage is output to operate the collimator.
* When it doesnt work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
7) TB2-15,16 : 30VDC LOCK voltage is output to operate Stand, Table and Bucky Stand.
* When it doesnt work : check whether F1 fuse on Main Power Board is disconnected or not
and, if necessary, replace it by new one.
Check out the operation of MG1. (LOCK voltage is approved with the point of contact of
MG1)
8) TB2-17,18,19,20 : 110VAC voltage is output to operate Stand and Table.
* When it doesnt work : Check out the operation of MG2. (110VAC voltage is approved with
the point of contact of MG2)
9) TB2-21,22 : It is supplied Power to operate the board of the optional photo-fluorography
and so on. (The voltage to operate photo-fluorographic relay on Interface board)
10) TB2-23 : It is Exposure Signal and Signal Wire for photo-fluorography and the other
options.
8) TB2-24,25,26 : It is output the voltage to operate the rotor of tube 1.(Tube on the
tubestand for Radiography)
* When it doesnt work : follow the check instructions of Rotor signal voltage adjustment on
H.T control Board Test(7).
9) TB2-27,28 : It is Thermometer Sensor and detects the overload of tube This is optional
and when connecting, it works normally. ( Usually the contact point of thermometer sensor
in the tube is connected to the rotor wire and when overloading of tube, it is treated the
rotor error.)

14- REX T/S

REX-R SERIES CALIBRATION METHOD (2)


Membrane Switch Function

()
1) POWER OFF Switch
3) Small Focus LED Indicator
5) Over Load LED Indicator
7) Kvp Increase Switch
9) mA Increase Switch
11) mAs / sec Increase Switch
13) Density Increase Switch (no use)
15) Tube1 Selection Switch
17) Bucky1 Selection Switch
19) Tube2 Selection Switch
21) Pre-exposure (Ready) Switch
23) Pre-exposure (Ready) LED Indicator

2) POWER ON Switch
4) Large Focus LED Indicator
6) Error LED Indicator
8) Kvp Decrease Switch
10) mA Decrease Switch
12) mAs/sec Decrease Switch
14) Density Decrease Switch (no use)
16) NO Bucky Selection Switch
18) Bucky2 Selection Switch
20) mAs / sec Change Switch
22) Exposure (x-ray) Switch
24) Exposure (x-ray) LED Indicator

15- REX T/S

B) Connect DIVIDER to H.T.T and Tube.


C) Connect DISPLAY and Communication cable of DIVIDER
D) Put the power to D-3.

* KV Offset : ROM DATA (no change) - - - - + , - RAM DATA (changeable)


* mA Offset : ROM DATA (no chagne) - - - - + , - RAM DATA (changeable)
* Exp.mA : mA feedback wave can be flat. ( change only plus(+) side )

16- REX T/S

kV & mA CALIBRATION
CALIBRATION SETTING
1) Power ON
2) Follow the above Section 6.3.1 procedure
Calibration mode will start.
3) Adjust kVp value to 40kVp by using kV UP/DOWN button, mA value to 200mA by using mA UP/DOWN
button and time to 0.05sec.
4) Start calibration on the basis of 200mA
5) Prepare the calibration data sheet.

CALIBRATION BY DYNALYZER lll (D-lll)

1) To calibrate by using D-lll, connect Display and Divider of D-llI as below


- Connect with Divider between HT transformer and tube.
- Connect with the correspondence cable between Display and Divider.
- Supply the power to D-lll.
* kV Offset : ROM DATA(Non-changeable) ----- +, - RAM DATA(Changeable)
* mA Offset : ROM DATA(Non-changeable) ----- +, - RAM DATA(Changeable)
* EXP mA : mA feedback wave can be flat.( possible to change to plus(+) side only)
(Factory Adjustment Only)

17- REX T/S

Fixed

(1) Kvp DATA Increase Switch


(3) mA DATA Increase Switch
(5) EXP mA Increase Switch
(7) DATA Value Save Switch

Changeable

(2) Kvp DATA Decrease Switch


(4) mA DATA Decrease Switch
(6) EXP mA Decrease Switch

Calibration with
Oscilloscope

18- REX T/S

REX-R/F SERIES Calibration Method (4)


Membrane Switch Function

Left

Right

LCD

LCD

1)POWER OFF Switch


3) Small Focus LED Indicator
5) Over Load LED Indicator
7) Kvp Increase Switch
9) mA Increase Switch
11) mAs / sec Increase Switch
13) Density Increase Switch (no use)
15) Tube1 Selection Switch
17) Bucky1 Selection Switch
19) Tube2 Selection Switch
21) Pre-exposure (Ready) Switch
23) Pre-exposure (Ready) LED Indicator
25) Fluoro Kvp Increase Switch
27) Fluoro mA Increase Switch
29) Fluoro Timer Reset Switch

2) POWER ON Switch
4) Large Focus LED Indicator
6) Error LED Indicator
8) Kvp Decrease Switch
10) mA Decrease Switch
12) mAs/sec Decrease Switch
14) Density Decrease Switch (no use)
16) NO Bucky Selection Switch
18) Bucky2 Selection Switch
20) mAs / sec Change Switch
22) Exposure (x-ray) Switch
24) Exposure (x-ray) LED Indicator
26) Fluoro Kvp Decrease Switch
28) Fluoro mA Decrease Switch
30) Image Enlargement Normal Switch(no use)
31) Image Enlargement MAG1 Switch(no use) 32) Image Enlargement MAG2 Switch(no use)
33) Automatic Brightness Control( ABC Switch(no use))

19- REX T/S

Calibration Setting Method

A) Connect DISPLAY and DIVIDER of D-3 to calibrate with D-3


like below.

B) Connect DIVIDER to H.T.T and Tube.


C) Connect DISPLAY and Communication cable of DIVIDER
D) Put the power to D-3.
* KV Offset : ROM DATA (no change) - - - - + , - RAM DATA (changeable)
* mA Offset : ROM DATA (no change)
- - - + , - RAM DATA (changeable)
* mA -COMP : mA feedback wave can be flat. ( change only plus(+) side )

20- REX T/S

(A) TUBE1 RAD CALIBRATION MODE

(B) TUBE2 RAD CALIBRATION MODE

Calibration with
an Oscilloscope

Feedback terminal

A) Put Oscilloscopes GND on PCBs GND.


B) Place Oscilloscopes terminal on +KV , -KV, and press 22 switch in the middle of pressing
21 switch of the right LCD to check KVP(10KVP per 1V)
C) Place Oscilloscopes terminal on +mA , -mA, and press 22 switch in the middle of
pressing 21 switch of the right LCD to check mA(10mA per 1V)

21- REX T/S

* TUBE2 FL CAL MODE CALIBRATION METHOD ( FLUORO MODE)

J1-9:FLU mode calibration


On / Off selection terminal

(C) TUBE2 FL(FLUORO) CALIBRATION MODE

22- REX T/S

Calibration with
an Oscilloscope

Feedback terminal

A) Put Oscilloscopes GND on PCBs GND.


B) Place Oscilloscopes terminal on +KV , -KV, and press 22 switch in the middle
of pressing 21 switch of the right LCD to check KVP(10KVP per 1V)
C) Place Oscilloscopes terminal on +mA , -mA, and press 22 switch in the middle of
pressing 21 switch of the right LCD to check mA ( 2mA per 0.1V)

23- REX T/S

Error Message & Troubleshooting


*

REX-R(F) SERIES

01

02

03

04

Checking point

ERROR

CODE

Filament value is LOW Error in


preheating
After Ready, Filament value is Low

HVC connecting point. Filament board defect, and refer TEST


A setting voltage of U22-3. Output voltage of U22-1 in H.T
Output voltage of U11-2, and replace if faulty
Filament board. U22,30,25 in H.T. Replace if faulty

Error
Filament value is HIGH Error in
preheating
After Ready, Filament value is High

HVC connecting point. Filament board defect, and refer TEST


A setting voltage of U22-6. Output voltage of U22-7 in H.T
Output voltage of U11-2, and replace if faulty
Filament board. U22,30,25 in H.T. Replace if faulty

Error
Rotor driven voltage is 220VAC,60VAC input and output voltage

05

ROTOR ERROR

CT1 and fuse 7 cutting in MAIN POWER BOARD


A setting voltage of U27-3,12. Output voltage of U27-1,14 in H.T
U27-2,13 ROTSEN Voltage input in H.T
Checking main fuse1,2 cutting. MG1,MG2 operation checking

06

CHARGE ERROR

Program down. P2 connector dislocation in CHARGE CPB


Part broken(D1,D2,U1,Resistance) in CHARGE PCB

07

OVER HEAT(TEMP SENSOR) ERROR

08

COMMUNICATION ERROR

09

KV LOW ERROR

10

KV HIGH ERROR

No relevance (TEMP SENSOR built in Tube)


Disconnection of TEMP wire on TB1 terminal
Connecting point between J1 and J2 of communication line
Defect of TIBPAL IC and program IC

Checking IPM DRIVE PCB voltage. Replace it if faulty


00

IPM FAULT ERROR

Checking the feedback wave of +KV,-KV,+mA,-mA


Checking U37,38,42 IC in H.T
Defect of high voltage PCB inside H.T.T
Checking light of LED1,2,3,4 on H.T control.

In case the program does not go on


without ERROR code.

(check CPU RESET is possible.)


Checking defect of Q4,Q5 unit in H.T control
Break of CPU(U1) unit in H.T control
Fuse1, fuse2 cutting in POWER BOARD
Unit broken or short in Filament board

24- REX T/S

CALIBRATION WHEN REPLACING THE FILAMENT BOARD WITH MAS METER(6)


* When the filament board is defective and new one is replaced, follow the below steps.
1. Each board has some differences of voltage. Accordingly, refer to the below standard
datasheet when replacing filament board by new one.
* In case of High Voltage Cable HVC-6M (Test Point : between GND and the connector J4-3)
mA step

Initial value

Ready (Preparation)

25mA

13.84VDC

29.5VDC

50mA

13.84VDC

32.0VDC

75mA

13.84VDC

33.5VDC

100mA

13.84VDC

34.5VDC

150mA

13.74VDC

39.0VDC

200mA

13.74VDC

40.7VDC

250mA

13.74VDC

41.9VDC

300mA

13.74VDC

43.0VDC

400mA

13.74VDC

44.3VDC

500mA

13.74VDC

45.5VDC

* The above values may have some differences in accordance with the length of High Voltage
Cable or the condition of calibration.
2. With reference of the above values, make measure the voltage of the old filament board
in advance and write down the values. Replace the old board by new one.
3. Test Point on PCB of the replaced board : Make place the tester between GND and J4-3
and set up the display condition to 200mA/70kV. When the ready, adjust its value to the same as
the voltage of PCB on the old filament board with the pot resistor R60.

25- REX T/S

* Follow the above steps to obtain the more precise mA/mAs values.
(mA, mAs Calibraion with mAs Meter)
1. Make preparation mA/mAs Meter.
2. Disconnect the wires and connect the terminal of mAs Meter to the below connector (A).

3. Select the display condition to 200mA/70kV/20mAs.


4. Make reset mAs meter and press the exposure switch. Check whether the established
mAs value is displayed. Otherwise, adjust its value to the same as the established
20mAs on the display with the pot resistor R60 of the filament board.
5. Make establishment each mAs(each mA X 0.1sec) as the above steps. Adjust mA Data
on the display console and make calibration as REX-R SERIES CALIBRATION(2) and
REX-R/F SERIES CALIBRATION(4)

26- REX T/S

REX WIRING

This instruction is to provide you with guide line how to connect


ctrical wire between the generator and other mechanical component.

LISTEM Corporation
1-REX Wiring

REX-R Series Brief Wiring

2-REX Wiring

REX-R Series Terminal

ROTOR1 PHASE(G)

- (16)

ROTOR1 MAIN(B)

- (17)

1> TB1 (Terminal Block 1) Terminal Explanation(Classfied by Power Source)


1) Single Phase : Composited by L1 , L2 , GND
2) Single Phase PSU : Composited by OVDC , 320VDC(Main Power) 0VAC , 220VAC(Control Power)
3) Three Phase : Composited by L1 , L2 , L3 , GND
2> TB2 (Terminal Block 2) Terminal Explanation (When you connect Stand and Table)
1) BUCKY1 : Connect to BUCKY START in the Table (KOB-1,KOB-3,KOB-60)
2) BUCKY1 : Connect to BUCKY START in the Table (KOB-1,KOB-3,KOB-60)
3) BUCKY2 : Connect to BUCKY START in the Stand (SFC-31R,CSTS-28,S-BRS)
4) BUCKY2 : Connect to BUCKY START in the stand (SFC-31R,CSTS-28,S-BRS)
5) 24VAC(CM) : Connect to Collimator CM power supply in Handle-Bar, and connect to table power supply.
6) 0 VAC(CM) : Connect to Collimator CM power supply in Handle-Bar, and connect to table power supply.
7) LOCK(DC+) : Connect to LOCK power supply(DC+) in the Table and Stand
8) LOCK(DC-) : Connect to LOCK power supply(DC-) in the Table and Stand
9) 100VAC : Only for Ceiling suspended Tube stand power(CSTS-28)
10) 0 VAC : Only for Ceiling suspended Tube stand power(CSTS-28)
11) 100VAC : Connect to the power supply of Mass chest system(Indirect Radiography) (Subsidery
power supply)
12) 0 VAC : Connect to the power supply of Mass chest system(Indirect Radiography) (Subsidery
3-REX Wiring

power supply)
13) +24VDC : Connect to the power supply of INTERFACE BOARD(J1-1) in Mass chest systems
controller.
14) 0 VDC : Connect to the power supply of INTERFACE BOARD(J1-2) in Mass chest systems
controller.
15) EXP SGN(EXP SIGNAL) : Connect to the power supply of INTERFACE BOARD(J2-2) in Mass
chest systems controller.
16) MAIN (G) : It is included in stand wiring. Connect to Rotor wiring(START) in handle bar.
17) PHASE (B) : It is included in stand wiring. Connect to Rotor wiring(RUNNING) in handle bar.
18) COM (W) : It is included in stand wiring. Connect to Rotor wiring(COM) in handle bar.
19) TEMP : This is TEMPERATURE SENSOR, connect to TH line.
20) TEMP : This is TEMPERATURE SENSOR GND, connect to TH line.
3> TB1 (Terminal Block 1) Termial connection

( A ) Line (single phase)


L3 should be added in case of three phases.

( B ) Line (PSU)

4> STAND Terminal connection of TB2(Terminal Block 2)

( C ) SFC-31R
* W : Connect to TB2-18

( D ) TB2 Handle bar line


* B : Connect to TB2-17
4-REX Wiring

* G : Connect to TB2-16

* DC(+) : Connect to TB2-7


* CM : Connect to TB2-5

* DC(-) : Connect to TB2-8 * GND : Connect to GND of the Case


* CM : Connect to TB2-6

( E ) CSTS-28

( F ) TB2 Handle Bar line

* W : Connect to TB2-18
* B : Connect to TB2-17
* 0VAC : Connect to TB2-9 * 110VAC : Connect to TB2-8
* CM : Connect to TB2-5
* CM : Connect to TB2-10

* G : Connect to TB2-16
* GND : Connect to GND of the Case

( G ) BS-20 , TB2 - BS20 Line


* DC(+) : Connect to TB2-11
* DC(-) : Connect to TB2-12
* GND : Connect to GND of the Case
* START : Connect to TB2-3
* START : Connect to TB2-4

5-REX Wiring

( H ) SFM-31
( I ) TB2 Handle Bar line
* W : Connect to TB2-18
* B : Connect to TB2-17
* G : Connect to TB2-16
* DC(+) : Connect to TB2-7
* DC(-) : Connect to TB2-8
* GND : Connect to GND of the Case
* CM : Connect to TB2-5
* CM : Connect to TB2-6

( J ) INTER FACE BOARD

( K ) DDN-100
* 0 VAC : Connect to J6-1
* X-RAY : Connect to J6-3

( L ) TB2 DDN100 Line


* 110VAC : Connect to J6-2
* X-RAY : Connect to J6-4

* Above J6 is the connector in INTER FACE BOARD


6-REX Wiring

* GND : Connect to GND of the Case

( M ) SFI-1X(E)
( N ) MS-101(E)
( O ) DDN100 Dark Hood Line
* SFI-1X:Supporing Tube+Tube arm
* MS-101:Dark Hood using 4digit numbering system normally.
* (E) : Dark Hood in case adoption Elevator(Option) higher than normal one.

( P ) KX100-2(Lens camera)
( Q ) INTERFACE BOARD Camera connecting line
* (1) : Connect to J5-1 * (2) : Connect to J5-2 * (3) : Connect to J5-3
* (4) : Connect to J5-4 * (5) : Connect to J5-3 * (6) : Connect to J5-4
* (7) : GND, Connect to GND of the Case * Above J5 is Connector in INTER FACE BOARD.

7-REX Wiring

( R ) MSE-24(Elevator)

(S) Communication cbl.

(T) Camera controller

(Btwn Camera-Camera controller)

* Using general Plug for power line


5> Connection TABLE Terminal of TB2 (Terminal Block 2)

(1)
(2)
(3)
(4)
(5)

: 0VAC
: 24VAC
: Start
: Start
: GND

(U)

KOB-1

* 0 VAC : Connect to TB2-6


* 24VAC : Connect to TB2-5
* GND : Connect to GND of the Case
* START(Bucky start) : Connect to TB2-1
* START(Bucky start)

8-REX Wiring

: Connect to TB2-2

(6) : 0VAC
(7) : 24VAC
(8) : Start
(9) : Start
(10): GND

(V)

KOB-3

* 0 VAC : Connect to TB2-6


* 24VAC : Connect to TB2-5
* GND : Connect to GND of the Case
* START(Bucky start) : Connect to TB2-1
* START(Bucky start)

( W) KOB-60

: Connect to TB2-2

( X ) TB2 KOB60 (CON2)

* Using general plug for Power line CON1


* START: Connect to TB2-1
* START: Connect to TB2-2
* GND: Connect to GND of the Case - CON2

9-REX Wiring

REX-RF SERIES Brief Wiring

10-REX Wiring

REX-RF SERIES System Terminal

ROTOR2 PHASE(G)
ROTOR2 MAIN(B)

-1
-2

1> TB1 (Terminal Block 1) Terminal (According to inpup power)


1) Single phase : Composited by L1 , L2 , GND
2) Single phase PSU : OVDC , 320VDC(Main Power) 0VAC , 220VAC(Control Power)
3) Three phase : Composited by L1 , L2 , L3 , GND
2> TB2 (Terminal Block 2) Terminal (In case of Stand, Table connection)
1) ROTOR2 MAIN (G) : Connect to Rotor wiring(START) of UNDER TUBE(FLUORO TUBE)
2) ROTOR2 PHASE (B) : Connect to Rotor wiring(RUNNING) of UNDER TUBE(FLUORO TUBE)
3) ROTOR2 COM (W) : Connect to Rotor wiring(COM) of UNDER TUBE(FLUORO TUBE)
4) TUBE2 (110VAC) : Connect to the FAN of UNDER TUBE(FLUORO TUBE)
5) TUBE1 (110VAC) : Not using
6) TUBE (0VAC) : Connect to Power supply in the FAN of UNDER TUBE(FLUORO TUBE)
7) TUBE2 TEMP : TUBE2 TEMPERATURE SENSOR, Connect to TH line.
8) TUBE2 TEMP RTN : TUBE2 TEMPERATURE SENSOR GND, Connect to TH line.
9) BUCKY1 110VAC(START) : Connect to BUCKY START of the table (DMT-80)
10) BUCKY1 0VAC(START) : Connect to BUCKY START of the table (DMT-80)
11) BUCKY2 110VAC(START) : Connect to BUCKY START of the stand (BS-20)
12) BUCKY2 0VAC(START) : Connect to BUCKY START of the stand (BS-20)
13) 24VAC(CM) : It is included in Stand wiring. Connect to Collimator CM power supply in Handle Bar.
11-REX Wiring

14) 0 VAC(CM) : It is included in Stand wiring. Connect to Collimator CM power supply in Handle Bar.
15) LOCK(DC+) : : Connect to LOCK Power supply(DC+) of the Table and the Stand.
16) LOCK(DC-) : Connect to LOCK Power supply(DC-) of the Table and the Stand.
17) 100VAC : Connect to Power supply of the Table and the Stand.
18) 0 VAC : Connect to Power supply of the Table and the Stand.
19) 100VAC : CCD Camera , Connect to POWER SUPPLY of I.I (Subsidery power)
20) 0 VAC : CCD Camera , Connect to POWER SUPPLY of I.I (Subsidery power)
21) +24VDC : Connect to the power of INTERFACE BOARD(J1-1) in Controller.
22) 0 VDC : Connect to the power of INTERFACE BOARD(J1-2) in Mass chest systems
controller.(Indirect Radiography)
23) EXP SGN(EXP SIGNAL) : Not using.
24) ROTOR1 MAIN (G) : For OVER TUBE (RAD TUBE), It is included in stand wiring.
Connect to Rotor wiring (START) in Handle bar.
25) ROTOR1 PHASE (B) : For OVER TUBE (RAD TUBE), It is included in stand wiring.
Connect to Rotor wiring (RUNNING) in Handle bar.
26) ROTOR1 COM (W) : For OVER TUBE (RAD TUBE), It is included in stand wiring.
Connect to Rotor wiring (COM) in Handle bar.
27) TUBE1 TEMP : TEMPERATURE SENSOR, Connect to TH line.
28) TUBE1 TEMP RTN : TEMPERATURE SENSOR GND, Connect to TH line.
3> TB3 (Terminal Block 3) Terminal
* Connect to the power of DMT-20 in case of 0VAC , 110VAC , 220VAC
AEC adoption B/D connection
Console
NO BUCKY
BUCKY 1
BUCKY 2

Ion chamber

adaptation board

SPOT
IONCHAMBER
TABLE
BUCKY
IONCHAMBER
WALL BUCKY
IONCHAMBER

connection

J3

connection

J2

connection

J1

12-REX Wiring

4> TB1 (Terminal Block 1) Terminal

( A ) Line(Single phase) :

( B ) Line (PSU)

L3 should be added in case of Three phases

5> STAND Terminal Connection of TB2 (Terminal Block 2)

( C ) SFC-31
( D ) TB2 Handle bar line
* W : Connect to TB2-26
* B : Connect to TB2-25
* G : Connect to TB2-24
* DC(+) : Connect to TB2-15
* DC(-) : Connect to TB2-16
* GND : Connect to GND of the case
* CM : Connect to TB2-13
* CM : Connect to TB2-14

13-REX Wiring

( E ) CSTS-28

* W : Connect to TB2-26
* 0VAC : Connect to TB2-17

( F ) TB2 Handle bar connection line

* B : Connect to TB2-25
* 110VAC : T Connect to B2-18

* G : Connect to TB2-24

* GND : Connect to GND of the Case


* CM : Connect to TB2-13
* CM : Connect to TB2-14

( G ) BS-20 , TB2 - BS20 Connection Line

* DC(+) : Connect to TB2-15


* DC(-) : Connect to TB2-16
* GND : Connect to GND of the Case
* START : Connect to TB2-11
* START : Connect to TB2-12

14-REX Wiring

6>)TABLE Terminal connection of TB2 (Terminal Block 2)

( H ) DMT-80

( I ) TB2,3 INTERFACE BOARD DMT80 Conection line

* 110VAC,0VAC,220VAC : Connect to TB3-1,2,3


* GND : Connect to GND of the Case
* BUCKY : Connect to TB2-9,10
* FOOT,SPOT Connector : Connect to J5 of INTERFACE BOARD

( J ) INTER FACE BOARD

* J1-3 : Connect to FOOT SW


* J1-5 : Connect to SPOT SW

15-REX Wiring

* J1-4 : Connect to FOOT SW


* J1-6 : Connect to SPOT SW

( K ) INTER FACE BOARD

* W : Connect to TB2-3
* 0VAC : Connect to TB2-6

* B : Connect to TB2-2
* 110VAC : Connect to TB2-4

* G : Connect to TB2-1

( L ) Image Intensifier (Integrated type)


( M ) Monitor (PM129,PM179)
* There are Seperated type and Integrated type in Image Intensifier
* I.I Size : 6inch / 9inch

* Monitor size : 12inch / 17inch

( N ) I.I POWER SUPPLY(+24VDC)

( O ) Camera POWER SUPPLY (+12VDC)

*Integrated type Image Intensifier(Including CCD camera) : POWER SUPPLY inbuilted type / Power supply exterior type.

16-REX Wiring

( P ) CCD Camera(CS8320B , +12VDC)

( Q ) Coaxial cable(btwn Monitor CCD camera)

* CS8320B CCD camera : Using for I.I and CCD Camera seperated type.(Camera POWER SUPPLY required.)

( R ) TB2 Camera POWER SUPPLY connection


( S ) POWER SUPPLY CCD camera connection
* Make sure not to change GND and +12VDC in case of Power supply connecting of CCD camera.
(Pin Number of CCD Power == * 10 : GND ,
* 11 : +12VDC)

17-REX Wiring

TABLE CONNECTION (DMT-94)


3.1.1 Brief configuration
Briefly, DMT-94 consist of Table body part, Table Tower part and Spot film device part.
3.1.2 Electrical connection.
Electrical connection for DMT-94 is as below:
3.1.2.1

Electrical connection between TABLE TOWER part and SPOT FILM DEVICE
As described below, cables from SFD should be connected to Connection unit Board of

Table parts.

For detailed information, see below diagram.

CONNECTION
UNIT BOARD

SPOT FILM DEVICE

JP5

J5
J8

JP9

J9
J11

GENERATOR
OPTION

<Electrical connection between SFD and CONNECTION UNIT BOARD of table part>

18-REX Wiring

<Detailed diagram between SFD and CONNECTION UNIT BOARD of Table part>

19-REX Wiring

3.1.2.2

Electrical Connection between TABLE part and GENERATOR

TABLE
SFD

REX

HT CONTROL
J11

J11
TABLE DRIVE
BOARD

INTERFACE
BOARD J1

J6

BUSHING
TB2
TB1

TB3

TUBE

<Electrical connection diagram between REX generator and Table part>

20-REX Wiring

SFD

OPTION
J11

RS485

6
5
4
3
2
1

6
5
4
3
2
1

J11
6
5
4
3
2
1

GND
Y
Z
B
A

6
5
4
3
2
1
CONN TRBLK 6

HT CPU BOARD
TERMINAL BLOCK
10
BUCKY IN
10 9
BUCKY IN
9 7
7 6
6 3
3 2
2 1
1

TB1
1 1
2 2
3 3
4 4
5 5
6 6
7 7
8 8
9 9
1010
1111
1212
1313
1414
1515

AC0V
AC220V
AC110V

TB2

GND
FOOT SW IN
FOOT SW IN

J1
3
3 4
4 5
5 6
6

FOOT SW
FOOT SW
SPOT SW
SPOT SW

INTERFACE BOARD

J6
BUCKY IN
BUCKY IN
FOOT SW
FOOT SW
SPOT SW
SPOT SW

GND

TB3
3
3 2
2 1
1

AC220V
AC0V
AC110V

1
2
3
4
5
6
7

1
2
3
4
5
6
7
TABLE DRIVE BOARD

TERMINAL BLOCK

.
10
9
8
7
6
5
4
3
2
1

REX

10
9
8
7
6
5
4
3
2
1
TUBE ROTOR CONNECTOR

DMT-94
TABLE
FOOT SW IN
FOOT SW IN

FOOT SW
<Detailed Electrical connection diagram between REX generator and Table part>

21-REX Wiring

J1

K1
1
2
3
4
5
6

DISCHARGE1
DISCHARGE1
DISCHARGE2
DISCHARGE2
MPOWER
GND

K2
14

14

50

50
32
5

32
5

DISCHARGE RELAY

41

POWER RELAY

41

23

23

59
68

CON6

59
68
K9

K10
14

RELAY AC/DC

D1
1N4001

RELAY AC/DC

D2
1N4001

14

50
32
5

32
5

41

41
23

23

K3

RELAY AC/DC

14

+ C1

23
59
68

R1 1.5K/0.25w

BD1
DB104G

ISO1
100uF/35v

R3
1.5K/0.25w
ISO2

C2
104j

C3
104j

R2
100/0.5W

RELAY AC/DC

D3
1N4001

FSAFTY
FOCAL

41
3

+24VDC
GND
DISCHARGE
POWER
BUCKY START
BUCKY1
BUCKY2

RELAY AC/DC

+ 4

6 -

32
5

50

1
2
3
4
5
6
7
8
9
10

59
68

59
68

J3

50

R4
100/0.5W
J4

CON10
C

220VAC INPUT
220VAC INPUT
220VAC OUTPUT
220VAC OUTPUT
AC12
AC12
POWER ON
POWER OFF
COM

J2

K4
AC100
AC100RTN
BUCKY1
BUCKY1 RTN
BUCKY2
BUCKY2 RTN

14
50
32
5
41
23
59
68

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8
9

CON9
RELAY AC/DC

D4
1N4001

K6
14
50
32
5

L1
J6
K5

INDUCTOR

14

23
59
68

1
2

50

41

CON2

32
5

RELAY AC/DC

41
23
59
68
K7

K8
14

D5
1N4001

RELAY AC/DC

32
5

41

FILOUT1
FILCOM
LARGE
COM
SMALL

41
23

23

59
68

J5

50
32
5

D7
1N4001

14

50

59
68

RELAY AC/DC

1
2
3
4
5
CON5

L2

L3

INDUCTOR

INDUCTOR

RELAY AC/DC
D8
1N4001

TITLE

MODEL REX R, R/F SERIES

B
A

DESCRIPTION

0
REVISION
5

02.11.04
98.8.20
DATE
4

MYK

DRW.BY

JCK

MY KIM

DRW

CHK.BY

SH LEE

APPR.BY

DWG

KD KIM

APPR

RELAY BOARD2

REV

AAZE100001

A
LISTEM CORPORATION

J2
MAIN MAG1(+24VDC OUT)
MAIN MAG1(GND OUT)
CHARGE MAG2(+24VDC OUT)
CHARGE MAG2(GND OUT)
PRE-CHARGE CONTACT OUT
PRE-CHARGE CONTACT OUT

K1
1
9

D2
1N4007

5
4

12

J1
+24V DC
GND
MAIN RELAY
CHARGE RELAY
PRE-CHARGE RELAY

1
2
3
4
5

1
2
3
4
5
6

CON6

8
14
13
RELAY 2
D1
1N4007

CON5

K2
1
9

D4
1N4007

5
4
12
C

8
14
13
RELAY 2
D3
1N4007

F1
FUSE
K3
1
9
5
4
12
8
14
13

RELAY 2
D5
1N4007

C
B
A

F1 C CHARGE

03.08.14

MYK

99.08.17

JCK

DATE

DRW APPR

DESCRIPTION

REVISION
5

MODELREX

R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

JC. KIM

SH LEE

KD KIM

TITLE

RELAY BOARD1

REV

DWG AAZE100016
LISTEM CORPORATION
1

:QP-0410-1(R0)

J?
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20

SW1

SW2

SW3

SW4

SW5

SW6

SW7

SW9

SW10

SW11

SW12

SW13

SW14

S17

SW18

SW19

SW20

SW21

SW22

SW23

SW25

SW26

SW27

SW28

SW29

SW30

SW31

SW8

CON20
SW32

D?
R?
680

SW33

SW34

SW35

SW36

SW37

SW38

SW39

LED
D?

SW40

R?

680
LED
SW41

SW42

SW43

SW44

SW45

SW46

SW47
SW51
J?
1
2
3
4
5
6

SW49

SW52

CON6
SW50

SW

SW

FUNCTION

SW1

KVP DEC

SW9

TUBE1

SW17

SW

FUNCTION

SW2

KVP INC

SW10

TUBE2

FIELD1

SW18

SW25

APR SEL1

SW33

APR1

SW41

100KG

SW3

mA DEC

SW11

NON BUCKY

FIELD2

SW19

SW26

APR SEL2

SW34

APR2

SW42

SW49

POWER ON

SW50

80KG

SW

FUNCTION

SW

FUNCTION

SW4

mA INC

SW12

WALL BUCKY

FIELD3

SW20

SW27

APR SEL3

SW35

APR3

SW43

40KG

SW

FUNCTION

SW5

mAS DEC

SW13

TABLE BUCKY

RESET

SW21

SW28

APR SEL4

SW36

APR4

SW44

10K

SW

FUNCTION

SW6

mAS INC

SW14

mAS/TIME

APR SET

SW22

SW29

APR SEL5

SW37
SW45

density DEC

SW8

density INC

SW15

**

SW16

**

Function1

SW23

Function2

SW24

**

SW30

APR SEL6

SW31

APR SEL7

SW32

APR SEL8

APR5

SW38

APR6

SW39

APR7

SW40

APR8

APR21

SW46

APR22

SW47

Control

SW48

**

REVISION

POWER OFF

MODEL REX R,R/F SEIRES

FUNCTION

SW7

SW

FUNCTION

SW22,23,31,32,40,45,46,47

DESCRIPTION

02.07.31

DRW. BY

99.08.20

DATE

REVISER

CHK. BY

APPR. BY

KD. KIM

TITLE
DWG

RADIO & ANATOMICAL KEYBOARD

ABBE100005
LISTEM CORPORATION

REV

D1
1N4007

3 -

R6
0.5/3W

BR1
KBPC606

TR2
2N3055

TR1
TIP31

+ 1
+

R1
200/2W

C1
3300uF/80V

C3
102

AC15V
AC220V

13
14
8
1
4

R2
680

AC15V
AC20V

POWER SUPPLY(PS1)

5
4
3
2
1

AC20V COM
AC20V
TRANSFORMER STEPUP

R7
680 1/2W

V+

T2

VC

3A 250V

12

11

D2
1N4745A

F1

COMP

VOUT

ILIM

R11
100 1/2W
R8
68 1/2W

10

D3
1N4007

VR1
500 1/2W

R10
1.2K 1/2W

F2
R3
27K

VREF

ISEN

VR2
2K

C4
220uF/35V

R9
2.7K 1/2W

U1
LM723CN

3A 250V

CON2
CON5PSM

VZ

TP3
TEST POINT

R12
10K

V-

R4
4.99K

R5
6.8K

R13
4.7K

C2
4.7uF/50V

CON1
8 8
7 7
6 6
5 5
4 4
3 3
2 2
1 1

C6
10uF/35V

VR4
100

F3

R39
1K
TP5
TEST POINT

3A 250V

C8
2200uF/80V

3 -

BR2
KBPC606

U2
LM337H
R20
120

ADJ

R19
2.2K/3W

D10
1N4007

VOUT

VIN

D6
1N4007

C7
100uF/25V

1
TP6
TEST POINT

R27
0.5/3W
TR5
2N3055

TR4
TIP31

+ 1
C11
3300uF/80V
D7

D5
1N4007

R26
200 2W

1
TP4
TEST POINT

R23
27K

12
V+

VC
VREF

ILIM

R24
4.99K

R25
6.8K

R28
680 1/2W

VOUT

COMP

V-

C9 13
102
14
R22
8
680
1
4

11

1N4745A

CON8SL

10

VR5
500 1/2W

VZ

R21
68 1/2W

D9
1N4007
VR7
2K

ISEN

R38
100 1/2W

R29
2.7K 1/2W

R30
1.2K 1/2W

C12
220uF/35V

R34
820

U3
LM723CN
R33
4.7K

C10
4.7uF/50V

F1,F2,F3

03.08.22

03.08.22

MYK

99.08.20

JCK

DATE

DRW

REVISION

DESCRIPTION
5

MYK

TITLE

POWER BOARD

REX R, R/F SERIES

MODEL
DRW.BY

CHK.BY

APPR.BY

JC KIM

SH LEE

KD KIM

DWG

LISTEM CORPORATION

APPR
3

REV

AAZE100026

:QP-0410-1(R0)

F10
F1

10A

10A
440VAC
0VAC
415VAC
20VAC

380VAC

2
1
2
3
4
5
6
7
8

6A

5A

24VAC
24VAC
20VAC
20VAC
50VAC
50VAC
110VAC
110VAC

10A 250V

D1
3

CON4

+ C1
2200uF/100V

KBPC2051
F2

CON8

0VAC

6.3A 250V

240VAC
50VAC

3A

D2
+

J9
+ C2
2200uF/100V

1A
KBPC2051
4

0VAC

+24VDC
GND
50VAC
50VAC
220VAC
220VAC
110VAC
110VAC

220VAC

K1-2

F3
4A 250V

110VAC 1A

1
2
3
4
5
6
7
8
CON8

2A
K2-1

0VAC
12VAC

F4
2.5A 250V

K2-2

1A

1A

J10

0VAC

1
2
3
4
5
6
7
8

220VAC 1A
2A

0VAC

0VAC

12VAC
12VAC
T2AC
T1AC
ACCOM

F5
2.5A 250V

J2

110VAC
40VAC

12VAC
12VACRTN
220VAC-1
220VAC-1RTN

1
2
3
4
5

F6
2.5A 250V

CON5
J11

220/380VAC
T1RTMAIN
T1RTSHF
T1RTCOM
T2RTMAIN
T2RTSHF
T2RTCOM
T1AC
T2AC
ACCOM

220/380VAC RTN

CON8

F7
10A 250V
F9
20A 250V

F8
20A 250V
CT1

9
AC INPUT
AC INPUT

J12
CON3

10

11

1
2
3

1
2
3

1
2
3
4
5
6
7
8
9
CON9

J3

1
2
3
4

24VAC

24VAC
24VAC
+30VDC
+30VDC RTN

10A 250V

T1

J8

J1

CON3

2
3

J6
1
2
3
4
5
6
7
8

J4
1
2
3
4
5
6
7
8
9

LOCK SW1
LOCK SW2
RT SW1
RT SW2
24VDC
GND
MAIN
CHARGE
DISCHARGE

D3
1N4001

D4
1N4001

K1

K2

K3

ZNR1
11

SSR1

+
SSR

C3

1
4

30uF
J7
+24V

220V RTN
110VAC
+24VDC
RTON
+24VDC
220/60 RELAY
+24 VDC
TSEL RELAY

D6
1N4001
J5
1
2
3
4
5
6
7
8
9

T1RTMAIN
T1RTSHF
ACCOM
T2RTMAIN
T2RTSHF
ACCOM
T1AC
T2AC

MPOWER
220V/60V
HIGH SPEED
TSEL RELAY
MAIN
CHARGE
DISCHARGE
RTSEN
RTSEN RTN

1
2
3
4
5
6
7
8
9

3
6

K3

K4
R1
470/2W

D5
1N4001

10

220V 5
8

60V

CON9

D7
1N4001

C
B

F10 , J3 2 3
F9

LINE

DESCRIPTION

REVISION

2003.05.21 MYK
2002.05.13 MYK
2001.04.28 MYK
99.08.20 JCK
DATE

DRW APPR

MODEL
DRW.BY

JC KIM

REX R/F SERIES


CHK.BY

SH LEE

APPR.BY

KD KIM

TITLE

DWG

MAIN POWER BOARD

AAZE100013

REV

C
LISTEM CORPORATION

RA1,2,3

RA4,5,6

RA3

RA6

2.2K

2.2K

+5V

+5V

+5V

+12V
D29
1N4007

2.2K

2.2K

RA4

D30
1N4007

2.2K

VR3
200

2.2K
CON1

19
20
21
B

22

1
2
3
4
5
6
7
8
D2
Tube2

D3
Bucky1

D4
Bucky2

D5
Bucky3

D6
Power_on

D7
Field1

D8
Field2

D9
Field3

D10
Func1

D11
Func2

D12
Reset

D13
Person1

D14
Person2

D15
Person3

D16
Person4

RA7
1KX8

CS2

RA8
1KX8

CS3
+12V
+12V
+5V

CS5

+5V

1
2
3
4
5
6
7
8
9

CS4

CS6

CON4

CS7

D0
D2
D4
D6
RS
E1

Small_Focus

Error_Lamp

D17
D18
D19
D20
D21
D22
D23
D24
Ap-position8
Ap-position1Ap-position2Ap-position3Ap-position4Ap-position5Ap-position6Ap-position7

J5
LED

Over_Load

Large-Focus
6

18

D1
Tube1

17

LCD(16)

J4
LED

Ready

1
3
5
7
9
11
13
15
17
19

2
4
6
8
10
12
14
16
18
20

D1
D3
D5
D7
R/W
E2

CON20A
D25
Shift

X-Ray

23

D26

D27

D28
set
Ap-position21 Ap-position22

10

16

VR4
500

CS1

15

C1
10uF 35V

CS0

14

10

13

D32
1N4007

12

RS
R/W
E1
D0
D1
D2
D3
D4
D5
D6
D7

ROW7

11

VOUT

VIN

vss
v0
vdd
rs
r/w
e
db0
db1
db2
db3
db4
db5
db6
db7

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16

ROW6

10

22

VR1
10K

U2
LM317T

ROW5

ROW4

7
C

CON2

LCD(20-4)

1
2
3
4
5
6
7
8
9

ROW3

ROW2

ROW1

C2
10uF 35V

ROW0

1
2
3
4
5
6
7
8

R26
22/2W

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18

RS
R/W
E2
D0
D1
D2
D3
D4
D5
D6
D7

VOUT

VIN
ADJ

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

RA1

VR2
10K

U1
LM317T
3

D31
1N4007

CON3

ADJ

R25
22/2W

RA5

RA2

+5V

+12V

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

24
25
26
+12V

27
28
29
30

CON30

C
B
A

D15,17,18,26,27,28,RA5

DESCRIPTION

0
REVISION
5

TITLE

MODEL REX R, R/F SERIES

U1,U2,D29,D30,D31,D32,C1,C2,VR3,VR4,R25,R26
backlight 22(LCD1) 10(LCD2)

06.05.02

ESJ

02.11.04
98.8.20

MYK

DRW.BY

JCK

ES JEON

DATE
4

DRW

CHK.BY

APPR.BY

HM LEE

YM SEO

DWG

APPR

LCD DISPLAY CONSOLE BOARD

REV

AAZE100020

LISTEM CORPORATION
3

RA1,2,3

RA4,5,6

RA3

RA6

2.2K

2.2K

+5V

+5V

+5V

+12V
D29
1N4007

2.2K

2.2K

RA4

D30
1N4007

2.2K

VR3
200

2.2K
CON1

19
20
21
B

22

1
2
3
4
5
6
7
8
D2
Tube2

D3
Bucky1

D4
Bucky2

D5
Bucky3

D6
Power_on

D7
Field1

D8
Field2

D9
Field3

D10
Func1

D11
Func2

D12
Reset

D13
Person1

D14
Person2

D15
Person3

D16
Person4

RA7
1KX8

CS2

RA8
1KX8

CS3
+12V
+12V
+5V

CS5

+5V

1
2
3
4
5
6
7
8
9

CS4

CS6

CON4

CS7

D0
D2
D4
D6
RS
E1

Small_Focus

Error_Lamp

D17
D18
D19
D20
D21
D22
D23
D24
Ap-position8
Ap-position1Ap-position2Ap-position3Ap-position4Ap-position5Ap-position6Ap-position7

J5
LED

Over_Load

Large-Focus
6

18

D1
Tube1

17

LCD(16)

J4
LED

Ready

1
3
5
7
9
11
13
15
17
19

2
4
6
8
10
12
14
16
18
20

D1
D3
D5
D7
R/W
E2

CON20A
D25
Shift

X-Ray

23

D26

D27

D28
set
Ap-position21 Ap-position22

10

16

VR4
500

CS1

15

C1
10uF 35V

CS0

14

10

13

D32
1N4007

12

RS
R/W
E1
D0
D1
D2
D3
D4
D5
D6
D7

ROW7

11

VOUT

VIN

vss
v0
vdd
rs
r/w
e
db0
db1
db2
db3
db4
db5
db6
db7

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16

ROW6

10

22

VR1
10K

U2
LM317T

ROW5

ROW4

7
C

CON2

LCD(20-4)

1
2
3
4
5
6
7
8
9

ROW3

ROW2

ROW1

C2
10uF 35V

ROW0

1
2
3
4
5
6
7
8

R26
22/2W

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18

RS
R/W
E2
D0
D1
D2
D3
D4
D5
D6
D7

VOUT

VIN
ADJ

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

RA1

VR2
10K

U1
LM317T
3

D31
1N4007

CON3

ADJ

R25
22/2W

RA5

RA2

+5V

+12V

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

24
25
26
+12V

27
28
29
30

CON30

C
B
A

D15,17,18,26,27,28,RA5

DESCRIPTION

0
REVISION
5

TITLE

MODEL REX R, R/F SERIES

U1,U2,D29,D30,D31,D32,C1,C2,VR3,VR4,R25,R26
backlight 22(LCD1) 10(LCD2)

06.05.02

ESJ

02.11.04
98.8.20

MYK

DRW.BY

JCK

ES JEON

DATE
4

DRW

CHK.BY

APPR.BY

HM LEE

YM SEO

DWG

APPR

LCD DISPLAY CONSOLE BOARD

REV

AAZE100020

LISTEM CORPORATION
3

5
D3
1N4007
1

GND

J1-3

R3
330

HCPL4503

VPC(GND1)

J3-3
J3-4

VI VO

2
R4
240

J4

1N4001

+ C2

C3

470uF

0.1

J3-5

D5

FPO(-FAULT1)

1
2
U1
1N4007
LM317
ADJ

J3-1

HCPL4503

D4

2
3

VP1(+15VOUTPUT1)

J3

J2

CP(-KV DR1)

J3-2

PC2
8

J1

:QP-0410-1(R0)

C7
0.1

+5VDC

J1-4

R1
6.8K

R2
330

SPR(+5V INPUT1)

8
6

C8
0.1

PC1

DR1
DR2

J1-1
J1-2

+ C1
470uF

+
AC
DF06

1
2
U2
1N4007
LM317

FUSE
J2-2

D2
+
AC
DF06

+
C4
470uF

VI VO

R6
240

D8
1N4001

VR2
5K

VNC(GND2)

+ C5

VN1(+15VOUTPUT2)

FNO(-FAULT2)

C6

J4-3
J4-4

J4-1

J4-5

0.1
HCPL4503

+ C11
10uF

FAULT

MODEL REX R,R/F SERIES

B
MYK

F1

03.08.14
99.08.20

JCK

REVISION

DESCRIPTION

DATE

DRW

CN(-KV DR2)

J4-2

PC4

470uF

J2-4

C9
0.1

HCPL4503

AC

ADJ

0VAC

R5
6.8K

D6

SNR(+5VINPUT2)

8
6

J2-1

AC

F1
115VAC

VR1
5K

PC3

T1

D1

+ C10
10uF

D7
1N4007

DRW. BY
APPR

CHK. BY

APPR. BY

MY KIM SH LEE KD KIM

TITLE IPM DRIVER BOARD


REV

DWG

AAZE100011

LISTEM CORPORATION

1k

R90

1K
C109 C29

C28

104

104

9
8
7
6
5

104

CON9

12
16
1
3
14
15

OUT
A0
A1
EN
VDD
VSS

C27

EXP_STOP
104

S1
S2
S3

S4
S5
S6
S7
S8

CH_SEL2
CH_SEL0
CH_SEL1

4
2

A2
GND

TP3
TP12

+5V
+12V
-12V
C97

C96

104

104

12

1K

AOUT

LM339

C30
104

AD7501SJ

AEC_END

R92
1M

R93
10K

U39A

R89

U40
13
11
10

1K

R102
R88

ION1
ION2
ION3
ION4
PT_INPUT

1
2
3
4
5
6
7
8
9

J12

+12V

C95

+12V

104
TL074
7
U36B
PT_AEC_RESET
ADG201A
11

+12V

3
C110
104

VR4
2

10V/1N4740

+5V

1
2
3

U36C
JM5
CON3
+5V

FSEN
100k

LM339

8
1K

C33
104

1
2
3

10

ZD10

R97

R99
10K

-kVP

104

TL084
TL084

R190
1

6
10k

U39C

FSEN_ANL

U28A

D30
U28B

104

+12V

3.3K

R104

TP16

R188

C121

10k
R189

TL074

JM4
CON3

-12V

+12V

10K

R103

60%

10K

1
TL074
U36A

RY24W-K

1K

R187 60k
R98

11

+kVP

11

4
1
8

11

13

U39D

-12V

C124
104

10

10K

105

5
3

LM339

VR5
2

U37C
C31

6
7

D19
1N4148

IN
D

1K

K1

+24V

+12V

R100

68K

-12V

9
10

R101

1K

R94

R95 22
D17
1N4148

R96 22

D18
1N4148

1N5231A

14
LM339

C122
U39B
U37B
16
15

1
2
3

JM6
CON3

TP11

U36D
14

1
ZD11
10V/1N4740

U26B

D31
D32
1N4001 1N4001

J10

1K
10uF/25V

ROTSEN
ROTSEN RTN

1
2
3

+
C123
100uF/25V

VR13
R176
1K
2
16K

ABC_IN
CCU_VSYNC

ULN2004A

C118

ABC_ANL
+ C114

TL074

10K

1K

1
2
3
4

R177

U28C

16K

10

C119
103

D33
D34
1N4001 1N4001

J18

2
9

103
TP17

IN
COM

RSEN
U26C

R191
C120
103

TL084

3
9

16k

IN
COM

U26D
4
9

U52
I1
I2
I3
I4
I5
I6
I7
I8
GND

Q1
Q2
Q3
Q4
Q5
Q6
Q7
Q8
COM

18
17
16
15
14
13
12
11
10

+12V

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15

13

OUT

CON15

U26F
6
9

IN
COM

11

OUT

ULN2004A
TP19

GND3

SW7

+12V

R179
1K

R180
1K

VR15
10K

VR16
10K

U53
1
2
3
4

1A
1C
2A
2C

1OC
1OE
2OC
2OE

8
7
6
5

J21
1
2
3
CON3

1
2
3
4
5
6
7
8

16
15
14
13
12
11
10
9

FLT1
FLT2
FLT3
FLT4
NEGA
CHAR
28
10

SW DIP-8/SM

J20

U26G

1
2
3
4
5
6
7
8

7
9

IN
COM

10

OUT

ULN2004A
TL084
13

CON8
12

TLP521-2_0

14

U28D

1
2
3
4
5
6
7
8
9

WRITE_0
CW_0
CCW_0
D_PC3
INV_LOW_SPEED
INV_HIGH_SPEED
D_PA2
D_PA3

J19

ULN2803

MEM1
MEM2
MEM3
MEM4
MEM_BANK_SEL
LIH
RH
CCD_RESET
WRITE
CW
CCW

18
17
16
15
14
13
12
11
10

Q1
Q2
Q3
Q4
Q5
Q6
Q7
Q8
COM

I1
I2
I3
I4
I5
I6
I7
I8
GND

IN
COM

ULN2004A

U51
1
2
3
4
5
6
7
8
9

14

OUT

ULN2004A

CON4

MEM1_0
MEM2_0
MEM3_0
MEM4_0
MEM_BANK_SEL_0
LIH_0
RH_0
CCD_RESET_0

15

OUT

ULN2004A

8
1

10uF/25V

16

OUT

D29
1N4148

R165

13

IN
COM

+ C113

12

R164

1
9

R178
ABC_IN

14

U26A

+12V
D28
1N4148

ADG201A

ABC_ANL
TP10

IN
D

C35
MF104K

ULN2803

8
7

iPSU Interface Connector

05.11.21

HML

03.02.19

MYK

99.08.20

JCK

DATE

DRW

DESCRIPTION

REVISION
A

MODEL

APPR

HT CONTROL BOARD1

DWG

ABBE100005

REX R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

HM LEE

ES JEON

YM SEO

TITLE

LISTEM CORPORATION
D

REV

TL082
7

+
8

U34B
+12V

U26E
5
9

R/F

R78

10K

R77

IN
COM

12

OUT

R84

10K

4.7K

U31D
7
8

ULN2004A

+12V

R85
C12

R184 10K
-KV1

KSEN

1
R50

R52 1M

R54 154K

R57

34K

10
9

1M

C14 100P

C26

40KV/1V

-KV

10K

U33B
U31A
ADG201A
3

C11 100P

VR1
10K

D
IN

R79

R55 154K

14

VR2
50K

U31B
ADG201A

+KV1

C6
472

R60

-MA

1K
C21
103

L2

15
16

D
IN

100uH

ZD5
1N4752

C23

D13
1N4148

C22

104
+MA

1N4148

R86
10K TP13

U34A
+12V

C128
103
D16
1N4148

-12V
C86
104

C25

MSEN

R87
10K
200mA/1V(R)
2mA/1V(F)

103

R82

R80
1K

D14

TL082
1

1N4148

-12V

ZD6
1N4752

C24

103

D15

104
+12V

4.99K

-12V

R81

1
2

R53 1M

ADG201A

C127
103

-MA1
2
1

100K
3

D6
1N4148

+KV
R51
10K

11

C13
104

TL082

D5
1N4148

-12V

D
IN

R58 100K
R56
10K

C10
102

ZD4
1N4744

C8
100P

D4
1N4148

5
+12V

100uH

D3
1N4148

ZD3
1N4744

C9
102

L1

C7
100P

-KVSEN
+KVSEN
GND
-MASEN
+MASEN
GND

1
2
3
4
5
6

103

J13

ADG201A
U31C

TP6

100P

D
IN

R83
10K

100K

+MA1

R63

C34
104

-12V
D10
1N4148

24.9K
U33A
-12V

1N4148

C84
104

D8
1N4148

C16
104

C15
100P
R64
61.9K

R67

R61

R62

10K

4.99K

C17
103

TP20

C18

100K

R68

R69

1K

1K

+IN1
-IN1
FB
+IN2
-IN2
GND

+5V

12
11
8
9
10
14
13
4
TP7

U41

R70 10K
1

U38
TL594

J14

C94
104

VR3
10K

R72
1K

C19
0.33

104

R71
1K

C20
104

1
2
3
4
5
6
7
8

I1
I2
I3
I4
I5
I6
I7
GND

Q1
Q2
Q3
Q4
Q5
Q6
Q7
COM

R73
R74
R75
R76

16
15
14
13
12
11
10
9

KDR1
KDR2

R
R
R
R

KDR3
KDR4

1
2
3
4
5
6
7

+
C55
47uF

ULN2004

D11
U37A

11

EXP
KSY

16
15
3
1
2
7

VCC
C2
C1
E1
E2
REF
CTL
DTC

+12V

D7

D9
1N4148

CT

TL082

KOUT

RT

TP9

+12V

+12V

C85
104

TP21

R65

R66

1K

1K

2
1

1N4148
D12

D
IN

ADG201A

1N4148

+5V

+12V

C81

U37D
7
8

D
IN

+12V

C93

13
S

U31

13
104

ADG202A

ADG201A

U37

104

5
ADG201A

C80
104

-12V

C92
104

-12V

8
7

iPSU Interface Connector

05.11.21

HML

03.02.19

MYK

99.08.20

JCK

DATE

DRW

DESCRIPTION

REVISION
A

MODEL

APPR

HT CONTROL BOARD1

DWG

ABBE100004

REX R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

HM LEE

ES JEON

YM SEO

TITLE

LISTEM CORPORATION
D

REV

D[0..7]

D[0..7]

U14

D0
D1
D2
D3
D4
D5
D6
D7

34
33
32
31
30
29
28
27

D0
D1
D2
D3
D4
D5
D6
D7

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

RD
WR
A0
A1
RES
85A

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

I1
I2
I3
I4
I5
I6
I7
I8
GND

Q1
Q2
Q3
Q4
Q5
Q6
Q7
Q8
COM

R42
R43
R44
R45

ULN2803
U21
1
2
3
4
5
6
7
8
9

14
15
16
17
13
12
11
10

I1
I2
I3
I4
I5
I6
I7
I8
GND

Q1
Q2
Q3
Q4
Q5
Q6
Q7
Q8
COM

+5V

82C55

I5
I6
I7
I8
I4
I3
I2
I1
GND

RT10
2.2K*9

Q5
Q6
Q7
Q8
Q4
Q3
Q2
Q1
COM

1
2
3
4
5
6
7
8

R46
R47
R48

CON8
J8
220/60 RELAY
TSEL RELAY
MAIN
CHARGE
PRECHARGE
DISCHARGE

18
17
16
15
14
13
12
11
10

1
2
3
4
5
6
7
8
9

R/F
R32
R33
R34
R35

ULN2803
U20
5
6
7
8
4
3
2
1
9

J6
POS1
POS2
POS3
RESET
EXP SIGNAL
NORM
MAG1
MAG2

18
17
16
15
14
13
12
11
10

R36
R37
R38
R40

+5V

R154
10k

14

U19
1
2
3
4
5
6
7
8
9

R28
R29
R30
R31

R156 100
B_PC0

4
U44B
74LS14

CON9

EXT_SW0

3
C106
104

R24
R25
R26
R27

J7
MAG3
INVFWD
FOCAL
FSAFTY
BUCKY2
BUCKY1
BUCKY START
POWER

14
13
12
11
15
16
17
18
10

1
2
3
4
5
6
7
8

+5V

R157
10k

14

R158 100
B_PC1

EXT_SW1

U44C
74LS14

ULN2803

C107
104

2
3
4
5
6
7
8
9

+12V
U18

C5
104

U14

U5

C53

U15

C54

104

RD
WR
A0
A1
RESET
CS

C64

104

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

104

82C55

16
15
14
13
12
11
10
9

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

104

4MHZ

680

IGBT FAULT
EXT_SW1
EXT_SW0

1A
1C
2A
2C
3A
3C
4A
4C

R/F

J17

1
2
3
4
5
6
7
8

R16

680

R17

680

R18

680

R19

680

1
2
3

D0
D1
D2
D3
D4
D5
D6
D7

R136
2.2K

I1
I2
I3
I4
I5
I6
I7
GND

Q1
Q2
Q3
Q4
Q5
Q6
Q7
COM

16
15
14
13
12
11
10
9

+5V

EXP
KSY
MSY
8253INT

RT11
2.2K*5
+12V
U54

8
7
6
5
4
3
2
1

D0
D1
D2
D3
D4
D5
D6
D7

RD
WR
A0
A1

22
23
19
20

RD
WR
A0
A1

85C

21

CLK0
G0
OUT0

9
11
10

CLK1
G1
OUT1

15
14
13

CLK2
G2
OUT2

18
16
17

+5V

RD
WR
A0
10k A1
RES
85D

R7

U15
D0
D1
D2
D3
D4
D5
D6
D7

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

CS

82C55_3

8253

+5V

AEC_END
U50
1
2
3
4
5
6
7
8

34
33
32
31
30
29
28
27

10
9
8
7
6
5
4
3
2
1

CON3

16
15
14
13
12
11
10
9

U5
D0
D1
D2
D3
D4
D5
D6
D7

R23

ULN2004

680

B_PC0
B_PC1

10K

R22

J5
TEMP
CHARGE SEN

R6

Y2
4

680

TLP521-4

+5V

C105

680

R21

+5V

TP2

R20

U4

RD
WR
A0
A1
RES
85B

5
36
9
8
35
6

D0
D1
D2
D3
D4
D5
D6
D7

1
2
3
4
5
6
7
8

2
3
4
5

+5V

34
33
32
31
30
29
28
27

1A
1C
2A
2C
3A
3C
4A
4C

TLP521-4
U17

U4
D0
D1
D2
D3
D4
D5
D6
D7

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

16
15
14
13
12
11
10
9

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

1A
1C
2A
2C
3A
3C
4A
4C

1
2
3
4
5
6
7
8

TLP521-4

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

INV_LOW_SPEED
INV_HIGH_SPEED
D_PA2
D_PA3
PT_AEC_RESET

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

MEM1_0
MEM2_0
MEM3_0
MEM4_0
MEM_BANK_SEL_0
LIH_0
RH_0
CCD_RESET_0

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

WRITE_0
CW_0
CCW_0
D_PC3
CH_SEL0
CH_SEL1
CH_SEL2

R185 680
R186 680
R138 680
R140 680

+kVP
-kVP
VSYNC

J15
1
2
3

CON3

8
7

iPSU Interface Connector

05.11.21

HML

03.02.19

MYK

99.08.20

JCK

DATE

DRW

DESCRIPTION

REVISION
A

MODEL

APPR

HT CONTROL BOARD1

DWG

ABBE100003

REX R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

HM LEE

ES JEON

YM SEO

TITLE

LISTEM CORPORATION
D

REV

+5V

+12V

J9
1
2
3
4

+5V

RT7
2.2K*9

+12V
C

+5V

49
36
37
3
58
60

LD1
POWER

26
27
28
29
2
61
59
6

C_INHIBIT1

LD2
ERROR

to PSU
3

RSEN
KSEN
MSEN
FSEN_ANL

LD3
READY
LD4
X-RAY

U16
18
17
16
15
14
13
12
11
10

+12V
H +12V
H-READY
H-XRAY
H-TRANSFER
H-INHIBIT

1
2
3
4
5
6
7
8

+kVP
-kVP

R8
R9
R10
R11
R12
R13
R14
R15

GND1

16
17
18
19
20
21
22
23

G1
G2

1
19

4
5

39
40
41
42
43
44
45
46

22
20

OE/VPP
CE

PSEN

P3.7/RD
P3.6/WR

1
2
3
4
5
6
7
8
9
11

47

31
30

P4.0/CMSR0
P4.1/CMSR1
P4.2/CMSR2
P4.3/CMSR3
P4.4/CMSR4
P4.5/CMSR5
P4.6/CMSR6
P4.7/CMSR7

7
8
9
10
11
12
13
14

P3.0/RXD
P3.1/TXD
NC0
NC1
NC2

24
25
32
33
38

C_READY1
C_XRAY1
C_TRANSFER1
C_INHIBIT1
/485RE
485DE

I1
I2
I3
I4
I5
I6
I7
I8
I9
I10

O1
O2
O3
O4
O5
O6
O7
O8

19
18
17
16
15
14
13
12

F800H
F900H
FA00H
FB00H
FC00H
FD00H
FE00H

DA
7SEG
KEY

85C

RD
WR

PWM0
PWM1

CON8

+5V

T2IN

R1IN

R1OUT

19

R2IN

R2OUT

20

T2OUT

11
15
16
10
9
6

J11

T1IN

C2+
C2+
C2C2GND1
GND2

JM3

1
2
3

1
2
3

FSEN
+ C74

MSY

D0
D1
D2
D3
D4
D5
D6
D7

C75
47uF/35

47uF/35

FOUT

KEY

2
3
4
5
6
7
8
9

A1
A2
A3
A4
A5
A6
A7
A8

19
1

/OE
DIR

CON3

CON3

R2 1K

14
13
12
11
10
9
8
7

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

A0
A1

17
16

A0
A1

DA

15

WR

14

VOUT_A

KOUT

VOUT_B

FOUT

VOUT_C

20

VOUT_D

19

AOUT
+12V
R41

VREF

R4 100K
RES
ZD1
BX03

+ C4
1uF

Q5
2N2222A

DEG1
DEG2
DEG3
DEG4
DEG5
DEG6

B1
B2
B3
B4
B5
B6
B7
B8

D1
1N4148

A
B
C
D
E
F
G
DP

SND-3360

/485RE
485DE
R1
820

4.7K

Q1
LM336-5.0V

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

18
17
16
15
14
13
12
11

+5V
RT8
103

SW1

18
17
16
15
14
13
12
11

FND1

1
2
3
4
5
6
7

DEG1 14
DEG2 11
DEG3 10
DEG4 6
DEG5 8
DEG6 7

R5
1K

+ C3
100uF
N.C.
RO
/RE
DE
DI
GND
GND

G1
G2

R3 8.25K

U42
VCC
N.C.
A
B
Z
Y
N.C.

1
19

Y1
Y2
Y3
Y4
Y5
Y6
Y7
Y8

R39
820

MAX489
R194
760

13
9
4
2
1
12
5
3

SWBP

1
2
3
4

C76
C102
C103
C79
C98
C99
C100
C101

Q4
2N2907A

JM2

+5V

R193
120

A1
A2
A3
A4
A5
A6
A7
A8

1uF/10V

CON6
14
13
12
11
10
9
8

2
3
4
5
6
7
8
9

+5V

8
13
12
17
14

DB0
DB1
DB2
DB3
DB4
DB5
DB6
DB7

U13
D0
D1
D2
D3
D4
D5
D6
D7

74HC245

MAX233

R192
760

+12V
-12V

8
7
6
5
4
3
2
1

+12V
C1+
C1VVV+

U10
MX7226

U11

R135 10K

T1OUT

TP8 TP14TP15
+12V

74HC541

U7

D0
D1
D2
D3
D4
D5
D6
D7

85B
85A
85D

16L8

+5V

18

11
12
13
15
16
17
18
19

27C512

80C552

VCC

D0
D1
D2
D3
D4
D5
D6
D7

A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15

10
9
8
7
6
5
4
3
25
24
21
23
2
26
27
1

U9

EA/VP
VSS0
VSS1
STADC
AVREFAVSS

U3

A1

19
18
17
16
15
14
13
12
1

A0
A1
A2
A3
A4
A5
A6
A7
OC

12
74LS14

D0
D1
D2
D3
D4
D5
D6
D7
C

13

2
3
4
5
6
7
8
9
11

U44F

10

18

48

P2.0/A8
P2.1/A9
P2.2/A10
P2.3/A11
P2.4/A12
P2.5/A13
P2.6/A14
P2.7/A15

VDD
AVDD
AVREF+
EW

P1.0/CT0
P1.1/CT1
P1.2/CT2
P1.3/CT3
P1.4/T2
P1.5/RT2
P1.6/SCL
P1.7/SDA

2
3
4
5
6
7
8
9

ALE/P

D0
D1
D2
D3
D4
D5
D6
D7

D0
D1
D2
D3
D4
D5
D6
D7

J4

1
2
3
4
5
6

P3.2/INT0
P3.3/INT1
P3.4/T0
P3.5/T1

P5.0/ADC0
P5.1/ADC1
P5.2/ADC2
P5.3/ADC3
P5.4/ADC4
P5.5/ADC5
P5.6/ADC6
P5.7/ADC7

A1
A2
A3
A4
A5
A6
A7
A8

57
56
55
54
53
52
51
50

R134 10K

H +5V
TXD
RXD
GND

11

D[0..7]

74LS14

74HC573

1
68
67
66
65
64
63
62

74HC541

J2

Y1
Y2
Y3
Y4
Y5
Y6
Y7
Y8

ULN2803

+5V

1
2
3
4
5
6
7
8

U8
18
17
16
15
14
13
12
11

1
2
3
4
5
6
7
8
9

I1
I2
I3
I4
I5
I6
I7
I8
GND

ABC_ANL

RESET

P0.0/AD0
P0.1/AD1
P0.2/AD2
P0.3/AD3
P0.4/AD4
P0.5/AD5
P0.6/AD6
P0.7/AD7

J3

Q1
Q2
Q3
Q4
Q5
Q6
Q7
Q8
COM

KOUT
FOUT
AOUT

30P

C_TRANSFER1

C63

30P

8253INT

C_XRAY1

2
3
4
5
6
7
8
9

C +12V
C-READY
C-TRANSFER
C-GND

30P

AGND

C_READY1

16
15
14
13
12
11
10
9

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

TLP521-4

1
2
3
4

X2

U2

1
1A
1C
2A
2C
3A
3C
4A
4C

30P

U44

C32

VDD

14

GND2

J22

35

15

RES

U6
1
2
3
4
5
6
7
8

30P

U42

C61

X1

7
5
4
3
2

RT6
680*5
1

30P

U13
C60

VSS

C +12V
C-READY
C-XRAY
C-TRANSFER
C-INHIBIT
C-GND

1
2
3
4
5
6

34

TP18
C

J1

30P

U11

C59

U44E

U1

C125
103

U43
TLP521-1

30P

U10

C58

A0

Y1
11.0592MHZ

C2 30P

R145 100

30P

U9

C57

DGND

30P

U8

C56

30P

U7

C52

D27
DIODE

CON2

U44A
74LS14
2

U3

C51

1
2
3

R144
1K
4

C45
10uF/35V

U2

C50

D[0..7]
C1 30P

R143 4.3K

C47
10uF/35V

C
JM7
CON3

+5V
1
2

C49
10uF/35V

U1

C44
10uF/35V

14

C46
10uF/35V

C126
103

J16

+12V

C48
10uF/35V

D0
D1
D2
D3
D4
D5
D6
D7

2
3
4
5
6
7
8
9

2
3
4
5
6
7
8
9

U44D
74LS14
8

+5V

14

+
+

CCU_VSYNC

-12V

+5V

RT2
10K*9
+5V

SW2
1

SWBP
2

SW3
1

SWBP
2

SW4 SWBP
1
2
SW5
8
7
6
5
SW DIP-4
2

A
B
C
D
E
F
G
DP

DEG4
DEG6
DEG3
DEG1
D6
D5
D7
7SEG
A0
D4
D1
D0
D2
D3

U12
1
2
3
4
5
6
7
8
9
10
11
12
13
14

DGT4
DGT6
DGT3
DGT1
ID6
ID5
ID7
WRMODE
IO4
IO1
ID0
ID2
ID3

ICM7218B
GND
DGT7
DGT5
DGT2
DGT8
SEG G
SEG F
SEG E
SEG C
V+
SEG D
SEG B
SEG A
D.P

28
27
26
25
24
23
22
21
20
19
18
17
16
15

DEG5
DEG2
G
F
E
C
D
B
A
DP

C62
104

+5V

8
7

iPSU Interface Connector

05.11.21

HML

03.02.19

MYK

99.08.20

JCK

DATE

DRW

DESCRIPTION

REVISION
A

MODEL

APPR

REX R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

HM LEE

ES JEON

YM SEO

TITLE

HT CONTROL BOARD1

DWG

ABBE100002
LISTEM CORPORATION

REV

:QP-0410-1(R0)

J?
1
2
3
4
5
6
7
8

SW1

SW2

SW3

SW4

SW5

SW6

SW9

SW10

SW11

SW12

SW13

SW14

CON8

SW

FUNCTION

SW

FUNCTION

SW

FUNCTION

SW

FUNCTION

SW

FUNCTION

SW

FUNCTION

SW1

KVP DEC

SW2

KVP INC

SW3

mA DEC

SW4

mA INC

SW5

TIME RESET

SW6

Function

SW9

NORMAL

SW10

MAG1

SW11

MAG2

SW12

MAG3

SW13

ABC ON/OFF

SW14

C
B
A
0
REVISION

MODEL
02.07.31

SW6,12

DESCRIPTION

DATE

JCK
DRW

APPR

FLUORO KEYBOARD
REV

APPR. BY

DWG AAZE100006

JC. KIM SH LEE KD KIM

LISTEM CORPORATION

DRW. BY
99.08.20

REX R/F SERIES

TITLE

CHK. BY

:QP-0410-1(R0)

+5V

VR1
10K

J1

J2
1
2
3
4
5
6
7
8

PC4
PC5
PC6
PC7

1
2
3
4
5
6
7
8
CON8

1
2
3
4
5
6
7
8
9
10
11
12
13
14
CON14

D1
1

2
1N4007

+12V
U2
R2

CON30

7407

D5
2

ABC
1
+5V

7407

U1C

D6
2

5
U1D

6
7407

9
U1E

D3
2

10

MAG2
1

2
3
4
5
6
7
8
9

VR2
10K

D4
2

22
D2
1N4007

+ C1
10uF

R1
1K

A
1

TP3
A

TP4
A

MAG1
1

1
1

7407

11

VCC

LM317/TO220
MAG3
1

7407
D7

VOUT

VIN

U1B

22/2W

U1A

PC3

1
2
3
4
5
6
7
8
9
10
11
12
13
14

ADJ

LCD1

+12V

+5V
1

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30

NORM
1

TP1
K
TP2
K

C
B

U2,R2,D1,C1,VR2, D2 22

06.05.09

ESJ
MYK

02.12.20

99.08.16

JCK

DATE

DRW

REVISION

DESCRIPTION

MODEL REX R,R/F SERIES


DRW. BY

ES JEON
APPR

CHK. BY

HM LEE

APPR. BY

TITLE FLUORO DISPLAY BOARD

DWG AAZE100004

YM SEO
LISTEM CORPORATION

REV

3
Q8
BUS13A

5A 250V
+ C14
2200uF/100WV

3
Q10
BUS13A
3

R55

0.5/3W
R56

0.5/3W

Q11
BUS13A
+12V
-12V

J4

+12V

+ C3
100uF/35V

C16
104

0.2/5W

10K

2.2K/0.5W

+12V

1K/0.5W

D8

Q6
2N3440

D10

-12V

VR38
100

R42

R58
100K

47K

C12

47K

8
100K

R60
2k

C21
104

R40
680

1N4148

3
7

EA
OSCOUT
U1
CT
LM3524

11

COMP

RT

SHTDWN
+SENSE
-SENSE
GND
2

Q2
IRF630

R7
10

R8
1K

8
-

C2

+12V
C23
104

2
4

14

EB

R9
10

VR4

R18
2.2K/2W

C5
103

D1
1N4745

R20
J3

100/3W

C9

6
8

D14
1N4148

Q5
IRFP250

R16

100uF/35V

R10
1K

7
50T:50T:50T

Q1
IRF630

1
2
3

68/0.5W

D4
1N4745

R17
470/2w

T2
222/2KV

1: 10

R19
2.2K/2W

+ C8
470uF/200V

C6
103

D3
1N4745

0.1/5W

R5
5K

1K

R28

R27

10K

20K
D7

T1

+ C7
470uF/200V

R21

10K

1K

R15
470/2w

10
4
5
8

C18
104
CA
CB

V+

C1 2.7K

R3

Q4
IRFP250

D2
1N4745

+12V

12
13

C11
104

68/0.5W
U4B
TL082

C24
104

47K

D13
1N4148

R6

U5A
TL082

0.01

U5B
TL082

1
U3A
+
LM358 4

104

R43

R41
780

R14

R2
5K

D9

V-

VR1

C13

47K

R32
2.7K

104

R49

R48
10K

R51
3.4k

R30

120K

R45

104

R50
1K

104

C17

8
1
R52

15

U4A
TL082

C20

1N4148

1N4148
R33
2.4K

+
4

VIN

10K

VREF

R46

U3B
LM358
8

R31
5.11K

104

R34

C19

R35

2
Q7
TIP42C

R47

R39
12K
+12V

RXE250
40A 60V

R37
12K

C15
104

0.5/3W

16

1
2
3
4
5
6
7
8

R36

1
3 2
2

J1

F2
R57

R44
15K/3W

3
Q9
BUS13A

2
1

0.5/3W

F1

J2

R54

Q3
2P4M

R23
20K

R26
10K

-12V

R24
20K

C4
104

R25
1N4148
C10

C22
104

D6
1N4148

R12
R11
100

560

R13
0.2/5W

20K

D5
224
4

7
3
U2A
LF353

100/2W

U2B
LF353

-12V

R22

1N4148

TITLE

C
B
A
0
REVISION

R51-3,4K /R60,2K/R58,C24,F2,R29

J2-1,2

DESCRIPTION

01.11.14 MYK
01.04.28 MYK
99.08.20 JCK
DATE DRW APPR

MODEL
DRW.BY
JC
KIM

NEW FILAMENT BOARD

REX-R,R/F SERIES
CHK.BY
CH
LEE

APPR.BY
KD
KIM

DWG

AAZE100005
LISTEM CORPORATION

D[0..7]
RT5
10K

+12V
U15

D0
D1
D2
D3
D4
D5
D6
D7

RD
WR
A0
A1
RES
85C

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

82C55

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

UDN2981

U18
8
7
6
5
4
3
2
1
9

I8
I7
I6
I5
I4
I3
I2
I1
GND

11
12
13
14
15
16
17
18
10

Q8
Q7
Q6
Q5
Q4
Q3
Q2
Q1
COM

8
7
6
5
4
3
2
1
9

I8
I7
I6
I5
I4
I3
I2
I1
GND

+12V

ULN2803
U17

GND

VCC

34
33
32
31
30
29
28
27

+12V
11
12
13
14
15
16
17
18
10

Q8
Q7
Q6
Q5
Q4
Q3
Q2
Q1
COM

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30

GND

+5V
1

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30

U21

SW2

D0
D1
D2
D3
D4
D5
D6
D7

34
33
32
31
30
29
28
27

D0
D1
D2
D3
D4
D5
D6
D7

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

RD
WR
A0
A1
RES
85E

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

CON30
82C55

1
2
3
4
5
6
7
8

16
15
14
13
12
11
10
9

4
+5V
C33

10uF
LCD0
LCD1
LCD2
LCD3
LCD4
LCD5
LCD6
LCD7
RS
R/W
E1
E2

TO LCD DRIVER

+5V
+5V

4
3
2
1
40
39
38
37

RD
WR
A0
A1
RES
85D

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

2
BZ

CON30A3P

CN3
PB0
PB1
PB2
PB3
PB4
PB5

PC6
PC7
OFF
ON
COM

+5V

34
33
32
31
30
29
28
27

D0
D1
D2
D3
D4
D5
D6
D7

RD
WR
A0
A1
RES
85F

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

TO KEY BOARD

+5V
RT10
4.7K

82C55

CN1
PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

+5V
+12V

PB0
PB1
PB2
PB3
PB4
PB5
PC6
PC7

RT6
10K

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20

D0
D1
D2
D3
D4
D5
D6
D7

GND

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

D0
D1
D2
D3
D4
D5
D6
D7

CON20

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30

34
33
32
31
30
29
28
27

GND

D0
D1
D2
D3
D4
D5
D6
D7

U1

1
2
3

J10
R33
R

VCC

BU1

26

+5V
U19

82C55

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20

+12V

ULN2803

CN4

D0
D1
D2
D3
D4
D5
D6
D7

18
17
16
15
14
13
12
11
10

I1
D1
I2
D2
I3
D3
I4
D4
I5
D5
I6
D6
I7
D7
I8
D8
+12V GND

+5V

GND

U16

26

1
2
3
4
5
6
7
8
9

CN2
ROW0
ROW1
ROW2
ROW3
ROW4
ROW5
ROW6
ROW7
CS0
CS1
CS2
CS3
CS4
CS5
CS6
CS7
Small-Focus
Large_Focus
Over_Load
Error_Lamp
Ready
X-Ray

+5V

2
3
4
5
6
7
8
9

TO FL MODE

J8
1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

R34

1.5K

R35

1.5K
D7
1N4007

R36

1.5K

R37

1.5K

R38

1.5K

D9
1N4007
1
2
3
4
5
6
7
8

R44

100

C27
105

R45

100

C28
105

R46

100

C29
105

R47

100

C30
105

R48

100

C31
105

R49

100

C32
105

R50

100

C34
105

U28

D8
1N4007

1
2
3
4
5
6
7
8

100

D6
1N4007

CON8

J9

R43

1
2
3
4
5
6
7
8

1A
1C
2A
2C
3A
3C
4A
4C

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

16
15
14
13
12
11
10
9

TLP521-4

U29
D10
1N4007
R39

1.5K

R40

1.5K

D11
1N4007
D12
1N4007

CON8
R41

1.5K

1
2
3
4
5
6
7
8

1A
1C
2A
2C
3A
3C
4A
4C

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

16
15
14
13
12
11
10
9

TLP521-4

+5V

U16

D13
1N4007

+5V

26
C18
7

+5V

26

104

U19

C19
7

+5V

26

104

U21

26
C20

104

U1

C9
7

104

C35
105

1
C
B

MODEL

DESCRIPTION

REVISION

03.01.06
99.08.20

JCK

DATE

DRW

DISPLAY CONSOLE BOARD 2


REV

MYK
CHK.BY

TITLE

REX R,R/F SERIES


DRW.BY

JC KIM

SH LEE

DWG

APPR.BY

KD KIM

LISTEM CORPORATION

APPR
C

AAZE100002

RT7
1K

+5V

+12V

RT1
10K
+5V
18K
U14A

+ C4
10uF

10K

R8

+5V

+ C5
10uF

T1OUT

T1IN

18

T2OUT

T2IN

R1IN R1OUT

19

R2IN R2OUT

20

C1+
C1VVV+

10
10
10
10

C-READY
C-XRAY
C-TRANSFER
C-INHIBIT

3
74HC14

+5V

R4 330

8
13
12
17
14

18
17
16
15
14
13
12
11
10

1
2
3
4
5
6
7
8

I1
Q1
I2
Q2
I3
Q3
I4
Q4
I5
Q5
I6
Q6
I7
Q7
I8
Q8
COM GND

1
2
3
4
5
6
7
8
9

+FKV
-FKV

ULN2803

R5
10K

Q1
Q2
Q3
Q4
Q5

R6
4.7K +5V

MAX233

+ C8

28

20
VCC
GND

VCC
GND

14

+5V

+5V

C16

C17

104

+5V

104

18
19
20
21
22
23
24
25

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

14
15
16
17
13
12
11
10

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

U22
D0
D1
D2
D3
D4
D5
D6
D7

34
33
32
31
30
29
28
27

RD
WR
A0
A1
RESET
CS

5
36
9
8
35
6

D0
D1
D2
D3
D4
D5
D6
D7

D0
D1
D2
D3
D4
D5
D6
D7
RD
WR
A0
A1
RES
85A

RD
WR
A0
A1
RES
85B

34
33
32
31
30
29
28
27

D0
D1
D2
D3
D4
D5
D6
D7

5
36
9
8
35
6

RD
WR
A0
A1
RESET
CS

82C55

+5V

U23
PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

4
3
2
1
40
39
38
37

PB0
PB1
PB2
PB3
PB4
PB5
PB6
PB7

18
19
20
21
22
23
24
25

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

14
15
16
17
13
12
11
10

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

10
9
8
7
6
5
4
3
25
24
21
23
2
26
1

A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14

20
22
27

CE
OE
WE

D0
D1
D2
D3
D4
D5
D6
D7

11
12
13
15
16
17
18
19

PC0
PC1
PC2
PC3
PC4
PC5
PC6
PC7

+5V
J5
1
2
3
4

VDD VCC +5V


+12V
1

CON4

+ C26
100uF/25V

U25
2 PP10-12-5

4.7K

4
5

JP1
CON3

U13B
74HC00

10

100uF/16V

C+12V

6
9

U14C
U13A
74HC00

6
74HC14

U13C
74HC00

12

74HC00
U13D

11

RT9

4.7K

U14E
13

P1.7

U14D
10

J6
R29

74HC14

74HC14

1
2
3

JP2
CON3

MAG1
MAG2
U26
H+5V

R25
1.5K
R27

1.5K

R28

1.5K

1.5K

1
2
3
4
5
6
7
8

14
13
12
11
10
9
8

VCC NC
NC RO
A
REB
DE
Z
DI
Y
GND
NC GND

1
2
3
4
5
6
7

MAG3

R42
4.7k

MAG4

1
2
3
4
5
6
7
8

1A
1C
2A
2C
3A
3C
4A
4C

1
2
3
4
5
6
7
8
CON8

U27
H+5V

R24
1.5K

1.5K

11

R30

H+5V

+ C25

KM28C256

+ C23
22uF

104

PC[0..7]
1

P1.6
R10

2
1
2
3

PA0
PA1
PA2
PA3
PA4
PA5
PA6
PA7

82C55

18K

104

100uF

4
3
2
1
40
39
38
37

R9
D2
1N4148

6N137A

C15

104

C22

U23

85A
85B
85C
85D
85E
85F
85G
85H

+5V

U9

MAG1
MAG2
MAG3
MAG4

6N137A
U11

+5V

PC[0..7]

C2+
C2+
C2C2GND1
GND2

R20
R21
R22
R23

TP2

C21

U22

104

11
15
16
10
9
6

CON10

U14B

R3 4.7K

5
4
3
2

H+5V
TXD
RXD
GND

2
1

VCC

C14

U8

PC[0..7]

J3

TP1

+5V

D[0..7]

U8

16
15
14
13
12
11
10
9

330
1

U12
7

O1
O2
O3
O4
O5
O6
O7
O8

SW1

R1
U10

I1
I2
I3
I4
I5
I6
I7
I8
I9
I10

19
18
17
16
15
14
13
12

+5V
13

74HC14

104

26

14

RT4
10K

U14F

R2 10K

14

RT3
2.2K

12

C13

U7

104

28

16L8

+5V

C12

+5V

+FKV

+5V

+5V

10

+5V

-FKV

TLP521-2

C-READY
C-XRAY
C-TRANSFER
C-INHIBIT
COM
ON
OFF
C-GND

CON10

1
2
3
4
5
6
7
8
9
10

8
7
6
5

QC1
QE1
QC2
QE2

2.2K

104

28

U6

GND

R12 680

C+12V
C+12V

1A
1C
2A
2C

R14

+5V
20

A[8..15]

U6
1
2
3
4
5
6
7
8
9
11

10

26

27C512
A[8..15]

A15
A14
A13
A12
A11
A10
A9
A8

C11

U5

104

14

1
2
3
4

OE/VPP
CE

20
D0
D1
D2
D3
D4
D5
D6
D7

26

680

2.2K

22
20

C10

U4
11
12
13
15
16
17
18
19

D0
D1
D2
D3
D4
D5
D6
D7

VCC

1
2
3
4
5
6
7
8
9
10

R13

74HC573

RD
WR

+5V

TLP521-4
U24

+12V

J2

8032

A8
A9
A10
A11
A12
A13
A14
A15

A0
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
A11
A12
A13
A14
A15

GND

R11

+5V
20

HGND

16
15
14
13
12
11
10
9

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

1A
1C
2A
2C
3A
3C
4A
4C

17
16
29
30
11
10

U7
10
9
8
7
6
5
4
3
25
24
21
23
2
26
27
1

D[0..7]

1
2
3
4
5
6
7
8

P1.0/T2
P1.1/T2X
P1.2
RD
WR
P1.3
PSEN
P1.4
ALE/P
P1.5
TXD
P1.6
RXD
P1.7

A0
A1

19
18
17
16
15
14
13
12
1

A0
A1
A2
A3
A4
A5
A6
A7
OC

26

H-+12V
H-READY
H-XRAY
H-TRANSFER
H-INHIBIT
+KV
-KV

INT0
INT1
T0
T1

1
2
3
4
5
6
7
8

P1.6
P1.7

U2

CON10

12
13
14
15

ULN2004

RY24W-K

A8
A9
A10
A11
A12
A13
A14
A15

D0
D1
D2
D3
D4
D5
D6
D7
C

VCC

RT8
680

Q1
Q2
Q3
Q4
Q5
Q6
Q7
COM

21
22
23
24
25
26
27
28

2
3
4
5
6
7
8
9
11

GND

+ C6
10uF

10K

I1
I2
I3
I4
I5
I6
I7
GND

Q1
Q2
Q3
Q4
Q5

16
15
14
13
12
11
10
9

P2.0
P2.1
P2.2
P2.3
P2.4
P2.5
P2.6
P2.7

RESET

D0
D1
D2
D3
D4
D5
D6
D7

1
2
3
4
5
6
7
8

R19

J1

RES

D0
D1
D2
D3
D4
D5
D6
D7

A[8..15]

+ C7
10uF

X2

39
38
37
36
35
34
33
32

VSS

5
6
7
8

7
8

18

30P

+5V

10K

D5
1N4001

X1

U3

RY24W-K

19

U5
P0.0
P0.1
P0.2
P0.3
P0.4
P0.5
P0.6
P0.7

10

4
3
2
1

R18

K3

C1

RT2
2.2K

+5V

20

7
8

EA/VP

30P

D4
1N4001

U4
31

VCC

C2

+5V

10

10K

RY24W-K
K2

+ C3
10uF

GND

Y1
11.0592MHz

4
7
8

+5V

40

R17

D[0..7]

40

VCC

10K

D[0..7]

D3
1N4001

1
2
3
4
5
6
7
8
9
10

D[0..7]

RES

74HC14

+ C24
10uF

R16

K1
3
5

SW3
SW DIP-4

4.7K

CON10

R7

20

1
2
3
4
5
6
7
8
9
10

R15
D1
1N4148

READY SW1
X-RAY SW1
GND
READY SW2
X-RAY SW2
GND
+24v
SPOT X-RAY SW
SPOT READY SW
FL MODE SW
24V

D0
D1
D2
D3
D4
D5
D6
D7

5 4 3 2

J4

QC1
QE1
QC2
QE2
QC3
QE3
QC4
QE4

16
15
14
13
12
11
10
9

J7
R31

1.5K

R32

1.5K

TLP521-4

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8
CON8

MAX489
R26
1.5K

1
C
B

MODEL

DESCRIPTION

REVISION

03.01.06
99.08.20

JCK

DATE

DRW

DISPLAY CONSOLE BOARD 1


REV

MYK
CHK.BY

TITLE

REX R,R/F SERIES


DRW.BY

JC KIM

SH LEE

APPR.BY

DWG

KD KIM

LISTEM CORPORATION

APPR
C

AAZE100001

1
2
3
4
5

VIA/5
1
2
3
4
5
VIA/5
1
2
3
4
5
VIA/5
1
2
3
4
5
VIA/5

3
1
2
3
4
5
VIA/5
1
2
3
4
5
VIA/5
1
2
3
4
5
VIA/5
1
2
3
4
5

VIA/5

1
2
3

VIA/3

Title
<Title>
Size
A3
Date:

Document Number
<Doc>
Monday, June 16, 2003

Rev
<RevCode>
Sheet

of

1
2
3

+5V

VIA/3

4
1
2
3
VIA/3
1
2
3
VIA/3

1
2
3
4
5
VIA/5

1
2
3
VIA/3
1
2
3
VIA/3

1
2
3
4
5
VIA/5

Title
<Title>
Size
A3
Date:

Document Number
<Doc>
Monday, June 16, 2003

Rev
<RevCode>
Sheet

of

1
2
VIA/2

Title

+12V

<Title>
Size
A3
Date:

Document Number
<Doc>
Monday, June 16, 2003

Rev
<RevCode>
Sheet

of

:QP-0410-1(R0)

A
J1
+
-

1
2

R1
330K/3W

NOTE1
R2
33k/3w

R3
100K/3W

R4
68K/3W

R5

R6

D1
IN4764

R7
68K/3W

R8

R9
68K/3W

Q1
IRFP450

LED1
D3
1N4733

D2
IN4764

U1
1

P3
6

R10

1
2

3.3M/0.25W

TO DISCHARGE RELAY

P2
1
2
3

CHARGE SEN

TLP531

NOTE1 : AC220/1PHASE 33K/3W


AC380/3PHASE 22K/3W
1PHASE-1EA
3PHASE-2EA

D
C

MODELREX R, R/F SERIES

DRW. BY

A
REVISION


DESCRIPTION

03.06.09 ES. JEON


ES. JEON
DATE REVISER

CHK. BY
SH. LEE

APPR. BY
KD.KIM

TITLE CHARGE/DISCHARGE MONITOR

DWG AZAE100001
LISTEM CORPORATION

REV

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

RP2_1
RP2_2
RP2_3
RP2_4
RP2_5
RP2_6
RP2_7
RP2_8

RP3_1
RP3_2
RP3_3
RP3_4
RP3_5
RP3_6
RP3_7
RP3_8

RP4_1
RP4_2
RP4_3
RP4_4
RP4_5
RP4_6
RP4_7
RP4_8

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8
RP1_1
RP1_2
RP1_3
RP1_4
RP1_5
RP1_6
RP1_7
RP1_8

ION CHAMBER Cable

TB1

ION CHAMBER #1
J1

9
1
2
3
4
8
5
6
7
10

+12V

10

RP3_7

RP3_8

R11
1K

R9 1K
14

C9
0.1uF

RP3_2

RP3_1

12

RP3_3

RP3_4

R8
1K

IC4_OUT

-12V U1A
LF347

RP2_2

RP2_1

RP2_3

RP2_4

JP11

IC1_OUT

R2
1K

Preamplifier

screen

ION CHAMBER Cable


ION CHAMBER #2

+12V -12V

J2
1
7
8
9
5
3
2
6
4
CON9

IC2_OUT

-12 VDC
+12 VDC
GND
IC1 OUTPUT
CENTER AREA
LEFT AREA
RIGHT AREA
RESET

1
7
8
9
5
3
2
6
4

1
TP5
GND

IC2_OUT
FLD2-IC2
FLD1-IC2
FLD3-IC2
STRT-IC2

+12V

3
2
1

3
2
1

JP10

CON9

R1
1.3K

RP1_2

R3 1K

C7
0.1uF

C4
100uF/35V
+12V

RP1_4

5
7
1
6
2
8
3
4
9

RP1_1

NOTE1

13

3
2
1

JP9

11
RP3_5

RP4_4

3
2
1

JP12
J4

RP3_6

C10
0.1uF

11

11
+
4

R10 1K

C2
C3
0.1uF
100uF/35V

RP1_3

RP2_8

R5
1K

U1D
LF347

RP4_3

RP2_7

U1C
LF347

R7
1.3K

RP4_2

RP4_1

TP4
IC3

NOTE1

RP4_8

shield

RP4_6

RP4_5

RP4_7

RP2_5

RP1_8

GND
IC4 INPUT
IC1 INPUT
IC2 INPUT
IC3 INPUT
-FLD 2 DR
-FLD 1 DR
-FLD 3 DR
-STRT DR

RP2_6
RP1_7

C8
0.1uF

R12
1.3K

U1B
LF347

NOTE1

RP1_6

R4 1K

AEC INTERFACE Cable

R6
1.3K

C1
0.1uF

+12V

TP2
IC2

NOTE1
RP1_5

+12V

TP1
IC1

+12V

TP6
IC4

11

CON10

1
7
8
9
5
3
2
6
4

IC1_OUT
FLD2-IC1
FLD1-IC1
FLD3-IC1
STRT-IC1

-12V

JUMPER

RP4
10K

IC GND
-12 VDC
+12 VDC
+24 VDC
GND

RP3
10K

RP2
10K

Power Interface Cable

1
2
3
4
5
6
7
8

RP1
10K

1
2
3
4
5
6
7
8

1
2
3
4
5
6
7
8

-12 VDC
+12 VDC
GND
IC1 OUTPUT
CENTER AREA
LEFT AREA
RIGHT AREA
RESET

1
7
8
9
5
3
2
6
4

Preamplifier

screen

IC3_OUT

ION CHAMBER Cable

ION CHAMBER #3

CON9
K10

K6
4

CR2
1N4148
C5
0.1uF

JP8

8
1

2
2

1
1

8
1

JP4
2
2

1
1
JP4-1

2
2

1
1

8
1

JP7-1
1
R13
10K
8
1

1
1
JP5-1

2
2

1
1
JP3-1

STRT-IC1

8
1

JP1

2
2

1
1
JP1-1

STRT-IC2

2
2

1
1

CON9

8
1

2
2

1
1

8
1

ION CHAMBER Cable

JP15-1

ION CHAMBER #4

K16

J5

4
JP13
2
2

1
1

8
1

1
7
8
9
5
3
2
6
4

2
IC4_OUT

JP13-1

STRT-IC3

JUMPER JP9,JP10,JP11,JP12
POS.A - NO OFFSET ADJUSTMENT
POS.B - OFFSET ADJUSTMENT

JUMPERS
JP1,JP2,JP5,JP6

JUMPERS
JP13,JP14,JP15,JP16

IC1=IC2=IC3=IC4

IC1=IC2=IC3

IC1=IC2

IC1IC2IC3IC4

screen

CON9

JUMPERS
JP3,JP4,JP7,JP8

Preamplifier

K13
JP15

ION CHAMBER
Type

-12 VDC
+12 VDC
GND
IC1 OUTPUT
CENTER AREA
LEFT AREA
RIGHT AREA
RESET

1
7
8
9
5
3
2
6
4

JP16-1

8
1

IC3_OUT
FLD2-IC3
FLD1-IC3
FLD3-IC3
STRT-IC3
4

JP16

FLD3-IC3
K1
4

4
JP3

8
1

K14

8
1

K5

4
CR1
1N4148

2
2

FLD3-IC2

FLD3-IC1

K9

JP5

1
1

1
7
8
9
5
3
2
6
4

JP14-1

FLD1-IC3
K4
4

4
2

2
2

8
1

JP2-1

FLD1-IC2

JP7

8
1

1
1

K8

4
CR4
1N4148

2
2

JP14

FLD2-IC3
K3
4

JP2

J3
4

8
1

JP6-1

8
1

FLD1-IC1

K12

TP3
ST

1
1

8
1

2
2

K7

4
CR3
1N4148

JP6

FLD2-IC2

FLD2-IC1

K11

K15
4

JP8-1

C6
100uF/35V

K2
4

-12 VDC
+12 VDC
GND
IC1 OUTPUT
CENTER AREA
LEFT AREA
RIGHT AREA
RESET

1
7
8
9
5
3
2
6
4

Preamplifier

screen

NOTE1 : U1 Not Used

C
B

MODEL

0
REVISION
5

DESCRIPTION

02.12.20

MYK

DATE

DRW

APPR

TITLE

REX R,R/F SERIES

DRW. BY

CHK. BY

APPR. BY

MYKIM

SH LEE

KD KIM

DWG

AEC ADAPTATION BOARD

AAZE100027
LISTEM CORPORATION

REV