Você está na página 1de 20

Technical Information

Liquiphant T FTL20H
Level limit switch
Level limit switch for liquids in the foodstuff industry,
compact design, housing made of corrosion-resistant stainless steel

Application
The Liquiphant T FTL20H is a level limit switch for
liquids in storage tanks, agitators and pipes which have
to meet particularly high hygiene standards internally
and externally.
It is used in particular in areas where other measurement
methods would probably fail: e.g. in the event of
viscosity, build-up, turbulences, flows, air bubbles, rash
temperature change when cleaning.
The Liquiphant T FTL20H is a hygiene version for fluid
temperatures up to 150 C.

TI379F/00/en

Your benefits
E.g. Stainless steel housing with round connector
M 12x1, degree of protection IP69K, always air-tight
even in the event of hour-long overflooding and
intensive cleaning
External test option using test magnet
On-site function control using external LED display
Large selection of process connections for hassle-free
installation in existing systems
Easy to install even at points difficult to access due to
compact design
Rugged stainless steel housing (316L)
CIP and SIP cleanability ensured
EHEDG certification

Liquiphant T FTL20H

Table of contents
Function and system design. . . . . . . . . . . . . . . . . . . . . 3

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . 9

Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Switching outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operating modes for variants AC and DC-PNP . . . . . . . . . . . . . . . 4

Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Performance characteristics. . . . . . . . . . . . . . . . . . . . . 8
Switching delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measured value resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Measuring frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Hysteresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Settling time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Influence of ambient temperature . . . . . . . . . . . . . . . . . . . . . . . . . 8
Influence of medium temperature . . . . . . . . . . . . . . . . . . . . . . . . . 8
Influence of medium pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

10
10
12
12
12
12

Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Function test with test magnet . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Light signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 16


CE approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overfill protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Marine approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . .

16
16
16
16
16

Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 16
Liquiphant T FTL20H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Socket wrench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Welding boss G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Welding boss G 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Welding neck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Coupling nut . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

17
17
17
17
18
18

Operating conditions: Installation instructions . . . . . . 8


Orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Supplementary documentation . . . . . . . . . . . . . . . . . 19
Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Operating conditions: Environment. . . . . . . . . . . . . . . 9


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Shock resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Operating conditions: Process . . . . . . . . . . . . . . . . . . . 9


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
State of aggregation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Viscosity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Gas content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Solids content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Endress + Hauser

Liquiphant T FTL20H

Function and system design


Measuring principle

The tuning fork of the FTL20H is brought to its resonance frequency by means of a piezoelectric drive.
If the tuning fork is covered by liquid, this frequency changes. The electronics of the FTL20H monitor the
resonance frequency and indicate whether the tuning fork is freely vibrating or is covered by liquid.

Measuring system

The measuring system comprises:


Liquiphant T FTL20H limit switch
Progammable logic control (PLC), miniature contactor, solenoid valve or AS-i bus

AS-interface

...
U

PLC

~ 230 V
L00-FTL20Hxx-14-05-xx-en-001

Example 1): Overfill protection or top level detection


Example 2): Lower level detection or dry running protection
Example 3): Dry running protection for pump

Endress + Hauser

Liquiphant T FTL20H

Input
Measured variable

Density

Measuring range

> 0.7 g/cm3


other density settings on request, e.g. 0.5 g/cm3

Output
Switching outputs
DC-PNP valve connector
Function

DC-PNP M 12x1

Positive voltage signal at the switch output of the


electronics (PNP)

AC 2-wire

AS-i

Switching the power


supply line

Switching the
D0 bit

Switch behaviour

ON/OFF

0/1
(free / covered)

Relay switching capacity

250 mA

D0 bit

Fail-safe mode

MIN/MAX (see below)

Switching delay

approx. 0.5 s on coverage / approx. 1.0 s on tuning fork becoming uncovered


other switching time on request

Switching threshold

with vertical orientation: 13.0 mm from top of fork


with horizontal orientation: 3.5 mm from fork centre

Hysteresis

Operating modes for variants


AC and DC-PNP

D1 bit
D1: 0 error

3 0.5 mm

The FTL20H can be connected in two operating modes. By choosing the suitable operating mode
(MAX or MIN safety), you ensure that the FTL20H switches safely even in the event of a fault (e.g. if the power
supply line is disconnected).
MAX - maximum safety
The FTL20H keeps the electronic switch closed as long as the liquid level is below the fork.
Example of an application: overfill protection
MIN - minimum safety
The FTL20H keeps the electronic switch closed as long as the fork is immersed in liquid.
Example of an application: dry running protection for pumps
The electronic switch opens if the limit is reached, if a fault occurs or the power fails.

Endress + Hauser

Liquiphant T FTL20H

Power supply
Cable entry

27.5 mm

2.)

27.5 mm

1.)

min. 40 mm

min. 40 mm

L00-FTL20Hxx-04-05-xx-xx-001

Pg 11 / NPT
QUICKON

M 12x1
Plastic

M 12x1
Metal

Accessories
1.) 4 x 0.34 M 12 elbowed (order number: 52010285)
2.) 4 x 0.34 + female angled connector M 12 with built-in LEDs (order number: 52018763)

Electrical connection

Variant DC-PNP (direct current) M 12x1 connector


Voltage source: shock-protected voltage or Class 2 circuit (North America)
Suitable for use in non-equivalent operation:
When both outputs are connected, the MIN and MAX outputs take on opposite states in trouble-free operation.
In the event of an alarm condition or a line break, both electronic switches are open. In addition to level
monitoring, function-dependent sensor monitoring can also be performed with the aid of 2-channel evaluation.

Operating mode MAX (NC contact)

Operating mode MIN (NO contact)

L00-FTL20xxx-04-05-xx-xx-002

1: BN

2: WT

3: BU
0.5 A

1: BN
3: BU
4: BK
0.5 A

L+

L
L00-FTL20xxx-04-05-xx-xx-009

Endress + Hauser

L00-FTL20xxx-04-05-xx-xx-003

L+
L00-FTL20xxx-04-05-xx-xx-010

Liquiphant T FTL20H

Variant DC-PNP (direct current) valve connector


Operating mode MAX (NC contact)

Operating mode MIN (NO contact)

3
L00-FTL20xxx-04-05-xx-xx-005

L00-FTL20xxx-04-05-xx-xx-004

+
3

0.5 A

PE

(Ground)

L+

PE

(Ground)

0.5 A

L+

L00-FTL20xxx-04-05-xx-xx-011

L00-FTL20xxx-04-05-xx-xx-012

Variant AC (alternating current) valve connector


Operating mode MAX

Operating mode MIN

2
L00-FTL20xxx-04-05-xx-xx-007

L00-FTL20xxx-04-05-xx-xx-006

3
> 19 V

> 19 V

R
0.5 A

0.5 A
PE

(Ground)

L1

PE

(Ground)
L00-FTL20xxx-04-05-xx-xx-013

L1

N
L00-FTL20xxx-04-05-xx-xx-014

Note!
Approved for relays with a holding power/rated power >2.5 VA (253 V) or > 0.5 VA (24 V).
Relays with lower holding power/rated power can be operated via a parallel-connected RC-element (option).

Endress + Hauser

Liquiphant T FTL20H

Connect AS-i bus

AS-i +
2

1: BN
3: BU

AS-i
L00-FTL20xxx-04-05-xx-xx-008

Programming instructions for the AS-i


AS-i profile: S-3.A.1
The address is defaulted to 0 (HEX). It is changeable via the bus master or programming unit.
Data bit:
D0:1 Sensor covered

D1:1 Status = O.K.

D0:0 Sensor free

D1:0 Status = error

D2 and D3 are not used.

Parameter bits (P0...P3) are not used.


Electrical connection

DC-PNP valve connector

DC-PNP M 12x1

AC 2 wire

AS-i

Supply voltage

10...35 V DC

10...35 V DC

19...253 V AC

24.5...31 V DC

Cable entry

Pg 11 / NPT

M 12x1

Pg 11 / NPT

M 12x1

Cable specification

Max 1.5 mm2 and


3.5...6.5

IEC 60947-5-2

Max 1.5 mm2 and


3.5...6.5

IEC 62026-2

Power consumption

< 825 mW

< 825 mW

< 810 mW

< 825 mW

Current consumption

< 15 mA

< 15 mA

< 3.8 mA

< 25 mA

Residual ripple

5 Vss at 0...400 Hz

5 Vss at 0...400 Hz

Endress + Hauser

Liquiphant T FTL20H

Performance characteristics
Switching delay

0.5 s when covering


1.0 s when becoming free
other switching time on request

Reference operating
conditions

Ambient temperature: 23 C
Process pressure: 1 bar
Medium: water
Medium density: 1
Medium temperature: 23 C
Installation from above /vertical
Density setting: > 0.7

Measured value resolution

< 0.5 mm

Measuring frequency

Approx. 1100 Hz in air

Maximum measured error

13.0 1 mm

Repeatability

0.5 mm

Hysteresis

3.0 0.5 mm

Settling time

<2s

Influence of
ambient temperature

Negligible

Influence of
medium temperature

29.6 x 10-3 mm/C

Influence of
medium pressure

55.2 x 10-3 mm/bar

Operating conditions: Installation instructions


Orientation

The Liquiphant T FTL20H can be installed in any position in a container or pipe. The formation of foam does
not impair its function.

3
L00-FTL20Hxx-11-05-xx-xx-001

Example 1): Overfill protection or top level detection


Example 2): Dry running protection for pump
Example 3): Lower level detection

Endress + Hauser

Liquiphant T FTL20H

Connecting cable

Up to 1000 m with AC/DC-PNP, AS-i to IEC 62026-2

Operating conditions: Environment


Ambient temperature range

40...+70 C
25...+70 C (AS-i)

Ambient temperature limits

Derating from 90.0 C process temperature: Reduction to max. 50.0 C ambient


Derating from 90.0 C process temperature: Reduction to max. 150 mA relay switching capacity

Storage temperature

40...+85 C

Degree of protection

IP65 with valve connector


IP66/67 with M 12x1 connector PPSU (plastic)
IP66/68 with M 12x1 connector 316L (metal);
IP69K with accessory 52018763 (signalling via connector with LEDs)

Shock resistance

to EN 60068-2-27 (30 g)

Vibration resistance

to EN 60068-2-64

Electromagnetic
compatibility

Interference emission to EN 61326, Electrical Equipment Class B,


interference immunity to EN 61326, Annex A (Industrial) and NAMUR Recommendation NE 21 (EMC).
AS-interface to EN 50295.

Overvoltage protection

Overvoltage category III

Operating conditions: Process


Medium temperature range

40...+150 C

Process pressure

40 bar

State of aggregation

Liquid

Density

> 0.7 g/cm3 (other density setting on request)

Viscosity

1...10000 cst

Gas content

Stagnant mineral water

Solids content

< 5 mm

Endress + Hauser

Liquiphant T FTL20H

Mechanical construction
Note!
All dimensions in mm

Valve connector

QUICKON

40

51.5

45.5

28.5

40

40

31.5

31.5

32 AF

*1

13

*1

38.2

32 AF

13

*1

32 AF

3.5 * 2

3.5 *
17.1

13

173

179.5

31.5

36

45

30

3.5 *
17.1

17.1

*2

*2

38.2

M 12x1

185.5

Design, dimensions

38.2

*2
L00-FTL20Hxx-06-05-xx-en-001

*1 Switch point with vertical installation


*2 Switch point with horizontal installation; the level increases in the direction of the arrow
Switch points at: density 1 / 23 C / 0 bar

Process connections

Process connection /
Dimensions

Accessories
(optional)

G A, G A
DIN ISO 228/I

Max. 40 bar
Max. 150 C
93

22.7

Pressure
Temperature

63.9

32 AF

16

38

L00-FTL20xxx-06-05-xx-en-003

10

Endress + Hauser

Liquiphant T FTL20H

Process connection /
Dimensions

Accessories
(optional)

Pressure
Temperature

GA
DIN ISO 228/I
for flush-mounted installation in welding boss

Welding boss
(without tuning fork
alignment)
with silicone O-ring
Endress+Hauser
52018765

Max. 25 bar
Max. 150 C

EHEDG with welding boss 52018765


93

22.7

Max. 40 bar
Max. 100 C

63.9

32 AF

16

38

FDA-listed materials
as per
21 CFR Part 175-178

L00-FTL20xxx-06-05-xx-en-003

G1A
DIN ISO 228/l

Max. 40 bar
Max. 150 C
74.5

93

22.7

32 AF

18.6

47.9

L00-FTL20xxx-06-05-xx-en-004

G1A
DIN ISO 228/I
with sealing surface for flush-mounted installation
in welding boss
EHEDG with welding boss 52001051
74.5

93

22.7

32 AF

18.6

47.9

Welding boss
(without tuning fork
alignment)
with silicone O-ring
Endress+Hauser
52001051

Max. 25 bar
Max. 150 C
Max. 40 bar
Max. 100 C

FDA-listed materials
as per
21 CFR Part 175-178

L00-FTL20xxx-06-05-xx-en-004

NPT
ANSI B 1.20.1

Max. 40 bar
Max. 150 C

R
DIN 2999
93

22.7

(63.9)

32 AF

47.9

L00-FTL20xxx-06-05-xx-en-005

Max. 40 bar
Max. 150 C

NPT
ANSI B 1.20.1
R
DIN 2999
93

22.7

(63.9)

32 AF

47.9

L00-FTL20xxx-06-05-xx-en-005

Endress + Hauser

11

Liquiphant T FTL20H

Process connection /
Dimensions

Accessories
(optional)

Pressure
Temperature

Triclamp
1" = 50.5 mm
2" = 64.0 mm
ISO 2852

Clamping ring and


front seal

Max. 16 bar
Max. 120 C

EHEDG only with 2" version and special seal *

* Seal from
manufacturer
Hyjoin Limited, UK

Max. 2 bar
Max. 150 C

108.8

47.9

22.7

L00-FTL20xxx-06-05-xx-xx-008

Threaded pipe joint


DN 25
DN 32
DN 40
DIN 11851

Sealing ring with collar

DN 50:
Max. 25 bar
Max. 140 C

With thread adapter nut


108.8

DN 25, DN 32, DN 40:


Max. 40 bar up to 100 C
Max. 25 bar up to 140 C

47.9

22.7

L00-FTL20xxx-06-05-xx-xx-009

Flush-mounted for welding neck 1"


Works standard
Endress+Hauser with silicone seal (included) and
coupling nut (accessory 52021715)

Welding neck
(tuning fork
can be aligned)
Endress+Hauser
52001047

Max. 40 bar
Max. 100 C
Max. 25 bar
Max. 150 C

EHEDG
116

22.7

52.9
47.9

FDA-listed materials
as per
21 CFR Part 175-178

L00-FTL20xxx-06-05-xx-xx-010

Weight

Approx. 300 g

Materials

Sensor and housing made of 316L, surface quality Ra < 1.5 m

Housing

Pipe housing

Terminals

Valve connector,
QUICKON,
M 12x1

12

Endress + Hauser

Liquiphant T FTL20H

Human interface
Function test with test magnet

Variants AC and DC-PNP:


On testing, the current state of the electronic switch is reversed.
Variant AS-interface:
On testing, D0 is inverted.
Performing test
Hold the test magnet against the mark on the nameplate:

L00-FTL20Hxx-19-05-xx-xx-001

The switching state changes.


Light signals

Variants AC and DC-PNP with valve connector/QUICKON

gn

rd

L00-FTL20Hxx-07-05-xx-xx-001

Green light (gn) lighting:


FTL20H is connected to the power supply and is operational.
Red light (rd) lighting:
Mode of operation MAX (overfill protection): sensor is immersed in liquid.
Mode of operation MIN (dry running protection): sensor is immersed in liquid.
Green light (gn) does not come on
Error:
No power supply.
Check plug, cable and power supply
Red light (rd) flashing:
Error:
Overload or short-circuit in load circuit.
Rectify the short-circuit
Reduce maximum load current to below 250 mA
Error:
Internal sensor error or sensor corroded.
Replace device

Endress + Hauser

13

Liquiphant T FTL20H

Variant AS-interface and DC-PNP with M 12x1 circular connector PPSU

gn

ye

rd

L00-FTL20Hxx-07-05-xx-xx-002

Green light (gn) lighting:


FTL20H is connected to the power supply and is operational.
Yellow light (ye) lighting:
Sensor is immersed in liquid.
Red light (rd) lighting with AS-interface:
Error:
Address 0 set or communication error.
Carry out addressing process
Parameterise slave
Or reduce line length (< 100 m total length)
Red light (rd) lighting with DC-PNP
Error:
Overload or short-circuit in load circuit.
Rectify the short-circuit
Reduce maximum load current to below 250 mA
Green light (gn) does not come on
Error:
No power supply.
Check plug, cable and power supply
Red light (rd) flashing (2 Hz):
Error:
Internal sensor error or sensor corroded.
Replace device

14

Endress + Hauser

Liquiphant T FTL20H

Variant DC-PNP with M 12x1 circular connector 316L

ye 1
ye 2

gn

L00-FTL20Hxx-07-05-xx-xx-003

Green light (gn) lighting:


FTL20H is connected to the power supply and is operational.
Yellow light (ye 1) lighting:
The sensor is not covered by liquid.
Yellow light (ye 2) lighting:
The sensor is covered by liquid.
Green light (gn) does not come on
Error:
No power supply.
Check plug, cable and power supply
Green light (gn) lighting, both yellow lights (ye 1+2) does not come on
Error:
short-circuit in load circuit.
Rectify the short-circuit
Error:
Internal sensor error or sensor corroded.
Replace device

Endress + Hauser

15

Liquiphant T FTL20H

Certificates and approvals

Note!
The specified certificates and approvals are available on www.endress.com/ftl20.

CE approval

The device is in conformity with the statutory requirements of the EC Directives.


Endress+Hauser confirms successful testing of the device by affixing the CE mark.

Sanitary compatibility

EHEDG (see process connections, Page 11),


Approval number: 3119/03/0445

Overfill protection

WHG and leakage

Marine approval

German Lloyd (GL),


Approval number: 42855-02HH

Other standards and


guidelines

AS-i profile S-3.A.1 as per EN 50295 (limit switch)

Ordering information
Liquiphant T FTL20H

10

Certificate *
0
3
9

20

Variant for non-hazardous areas,


CSA General Purpose,
Special version

Process connection, material


GCJ
GDJ
GEJ
RCJ
RDJ
RRJ
RSJ
UPJ
TCJ
TDJ
MNJ
MPJ
MQJ
YY9

30

Thread ISO228
Thread ISO228
Thread ISO228
Thread ANSI
Thread ANSI
Thread DIN2999
Thread DIN2999
Flush-mounted,
ISO2852
ISO2852
DIN11851
DIN11851
DIN11851
Special version

DN 25-38 (1...1-"),
DN 40-51 (2"),
DN 25 PN 40,
DN 32 PN 40,
DN 40 PN 40,

316L
316L
316L
316L
316L
316L
316L
316L
316L
316L
316L
316L
316L

installation in accessory: welding neck

installation in accessory: welding neck 1" 52001047


Tri-Clamp connection
Tri-Clamp connection
Hygienic connection
Hygienic connection
Hygienic connection

19...253 V AC, 2-wire


10... 35 V DC, PNP 3-wire
AS-i bus
Special version

Additional equipment
B
C
D
E
F
Y

FTL20H

G A,
G A,
G 1 A,
" NPT,
" NPT,
R ,
R ,

Electronics
1
2
3
9

40

WHG, leakage detection


CSA US

Valve connector
Pg 11,
ISO4400,
Valve connector
NPT ", ISO4400,
Connector
M 12x1, PPSU,
Valve connector with QUICKON connection,
insulation displacement connection,
Connector
M 12x1, 316L,
Special version

150 C,
150 C,
150 C,
150 C,
150 C,

IP65
IP65
IP66/67
IP65
I66/68/69K

Order code

* The specified certificates and approvals are available on www.endress.com/ftl20.

16

Endress + Hauser

Liquiphant T FTL20H

Accessories

!
Socket wrench

Note!
All dimensions in mm
Order number: 52010156

32 AF

Socket wrench AF 32

L00-FTL20xxx-00-05-xx-en-001

Welding boss G

Order number: 52018765

max. 25 bar
max. 150 C

26

For flush-mounted installation


and sealing

21

max. 40 bar
max. 100 C
55 0.4

G
ISO 228

Material: corrosion-resistant steel


1.4435 (AISI 316L)

32

Sensor cannot be aligned

Weight: 0.13 kg
Seal: silicone O-ring
Order number: 52019735

L00-FTL5xxxx-06-05-xx-xx-019

FDA-listed materials as per


21 CFR Part 175-178
Welding boss G 1

Order number: 52001051

max. 25 bar
max. 150 C

29.6

For flush-mounted installation


and sealing

24.6

max. 40 bar
max. 100 C
60 0.4

G1
ISO 228

Material: corrosion-resistant steel


1.4435 (AISI 316L)

41

Sensor cannot be aligned

Weight: 0.19 kg
Seal: silicone O-ring
Order number: 52001386
L00-FTL5xxxx-06-05-xx-xx-020

FDA-listed materials as per


21 CFR Part 175-178

Endress + Hauser

17

Liquiphant T FTL20H

Order number: 52001047


With 3.1.B material certificate: 52006909
for flush-mounted installation and sealing
of a Liquiphant FTL50H, FTL20H
with process connection EE2, UPJ

4.5

Material: corrosion-resistant steel


1.4435 (AISI 316L)

65

RD 52 x 1 6

Sensor can be aligned

42.3

Welding neck

Weight: 0.15 kg
Seal on Liquiphant
Order number: 942816-0000

~26
L00-FTL5xxxx-06-05-xx-xx-022

FDA-listed materials as per


21 CFR Part 175-178
Coupling nut

Order number: 52021715


for connection UPJ or
welding neck 52001047
RD 52 x 1/ 6

36

56

Weight: 0.17 kg

63

DIN 11851-F25-1.4301

21

L00-FTL20Hxx-06-05-xx-xx-007

Cable

Order number: 52010285


4 x 0,34 M 12 socket

Body: PUR (blue)

27.5

Cable: PVC (grey) 5 m length

Coupling nut: Cu Sn/Ni

min. 40

Protection: IP67
L00-FTL20Hxx-07-05-xx-xx-004

Temperature range: 25 C to +70 C

Order number: 52018763


4 x 0,34 M 12 socket with integrated LEDs

Body: PVC (transparent)


Coupling nut: 316L

27.5

Cable: PVC (orange) 5 m length

min. 40

Protection: IP69K (fully locked)


Temperature range: 25 C to +70 C

18

L00-FTL20Hxx-07-05-xx-xx-005

Endress + Hauser

Liquiphant T FTL20H

Supplementary documentation
Operating Instructions

FTL20H
KA214F/00
Order number: 52019486
Welding boss G
KA219F/00
Order number: 52020163

Certificates

Liquiphant FTL20, FTL20H


Allgemeine bauaufsichtliche Zulassung Z-65.11-311
ZE247F/00
Liquiphant FTL20, FTL20H (Leckage)
Allgemeine bauaufsichtliche Zulassung Z-65.40-312
ZE248F/00
Liquiphant FTL20, FTL20H
Number of the Certification Document 37102
ZE249F/00
Liquiphant FTL20, FTL20H
Certificate of Compliance No. 1238461
ZE250F/00

Endress + Hauser

Note!
The specified certificates and approvals are available on www.endress.com/ftl20.

19

Liquiphant T FTL20H

International Head Quarter


Endress+Hauser
GmbH+Co. KG
Instruments International
Colmarer Str. 6
79576 Weil am Rhein
Deutschland
Tel. +49 76 21 9 75 02
Fax +49 76 21 9 75 34 5
www.endress.com
info@ii.endress.com

TI379F/00/en/07.04
SL/FM+SGML6.0 ProMoDo

Você também pode gostar