Tel:
+420-2-8384-2063, -2064, -3022
Fax: +420-2-8384-1067
e-mail: ndt@starmans.cz
EMAT probe
EMAT probe transducer consists of a coil with appropriate parameters.
Common coils are used for frequencies 2 to 5 MHz. The coil is placed into
magnetic field of permanent magnets made of rear earths.
Protective layer of the probe
Protective layer, in this application, is made of corundum, which prevents
probe from damage even after hard impact. Thus, EMAT probe may have
virtually unlimited service life.
Preamplifier cable length
There is a built-in preamplifier in the EMAT probe, because generally
EMAT signal is very weak compared to classic piezoelectric transducer.
Preamplifier amplifies the weak signal, to make it possible to connect the
probe using 5 m cable without any problem.
Pulse separator
Pulse separator separates initial pulses from the measured signal and
protects preamplifier against high voltage of power pulser. Probe works as single element transducer.
EMAT cables
Cables shall have appropriate impedance to transfer high power and prevent energy losses.
EMAT parameters for thickness measurement
Dead zone
15 s
Range of measured thickness
0.8 mm 150 mm
Distance between probe and material 1 to 2 mm !!!
Thickness of protective layer
0.5 mm
Material of protective layer
corundum plate
Type of probe
shear wave
Frequency of probe
2 7 MHz
Probe holder
For linear and area testing, the EMAT probe can be mounted
into special carriage with small rollers for easy and continuous
probe movement. Distance between probe and material is
maintained in the range to 1 to 2 mm no contact with
measured surface.
Movement co-ordinates may be detected for B-scan presentation on the display.
Preamplifier
Description of Functioning
EMAT system based on DIO2000 electronics:
Pulse separator
Coax. cables
Probe coil
DIO2000 US card
DIO2000 Pulser
Protective layer
- pulser DIO2000-BP;
- EMAT probe with bulit-in preamplifier.
Measured tube
Personal Computer
EMAT Signal
With switched on function AutoGain, DIO2000
automatically compensates attenuation changes within
the material. Thickness can be red by the echo position
or utilizing correlation thickness meter automatically,
without intervention of operator.
Example:
Measuring SDH 3 mm in the calibration block section of railway wheel.