INDEX

Product Description ............................................................................................................... 03


Warranty of the Equipment ................................................................................................... 04
Important Safety Considerations .......................................................................................... 04
General Data ......................................................................................................................... 04
Technical Specifications ....................................................................................................... 08
Symbols ................................................................................................................................. 09
Dimensions ............................................................................................................................ 10
Installation .............................................................................................................................. 12
Operation ............................................................................................................................... 14
Maintenance .......................................................................................................................... 16
Precautions ........................................................................................................................... 16
Cleaning ................................................................................................................................. 16
Control of the sterilization process ...................................................................................... 19
Troubleshooting .................................................................................................................... 20
Final Considerations .............................................................................................................. 20

DEAR CUSTOMER
Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY
product comparable to the best products available in the World. This manual is a general presentation
of your product and it will give you important details to help you to solve possible problems.
Please, read it and keep this with you.

PRODUCT DESCRIPTION
Its an autoclave for sterilization by vapor, modern design, easy to operate, ideal for dentistry
offices (general practice, endodontia, pediatric dentistry, orthodontics, prosthesis, operative dentistry
and esthetics).
Its structure is made of carbon steel, with high-shine epoxi paint, polymerized at 250C, with
phosphate treatment resistant to corrosion and cleaning materials.
The chamber is made of stainless steel, offering durability and resistance to permanent temperature
changes, and has a capacity of 12 liters.
Its internally coated with heath isolating material, which optimizes its operation and reduces the
energy consumption.
It features an anti-vacuum valve, which prevents the door and safety seal from blocking, and its
automatically activated if an excessive pressure was detected. The doors sealing system is made of
high-performance silicone, resistant to temperature and pressure changes, and its flexibility allows the
perfect sealing between surfaces in contact, preventing leakage.
The control panel contains LED and function buttons, graduated manometer for temperature and
pressure, ergonomic handle with automatic depressurization safety system, which prevents accidents
in case of unintentional opening.
It comes with a hose for steam outlet, graduated glass, fuses for 127/220V~ and two perforated
stainless-steel trays, which allow the heat diffusion, thus ensuring the quality of the sterilization
process.
Note: Give preference to hand pieces and peripherals developed in accordance with requirements
of directive 93/42/EEC (with CE mark).
Quality Management System certificate in accordance to ISO 9001:2000 and ISO 13485:2003
standart, assuring that equipments are produced in base with standardized procedures, meeting the
quality continually improve.
Products manufactured in agreement with RDC 59 - ANVISA - (Sanitary Surveillance National
Agency) - with guarantees accomplishment hole to the sanitary legislation and according to BPF (Good Manufacturing Procedures), requested by Mercosul Sanitary authorities.
In order to comply with directive 93/42/EEC for CE Mark purpose, it is defined that the working life
of metallic structure this product is estimated in 10 years (life-cycle) provided that the end-user
follows the requirements and instructions of this manual.
IMPORTANT:
This equipment is for dental use use only. It must be operated and utilized by specialized
professional (certified professional, according to the legislation of the country) and following the
instructions of the manual. The operation of the equipment required, for the professional, the utilization
of correct instruments and it should to be in perfect conditions of the use, and to protect the
professional, the patients and others, in the eventual danger situation.
This equipment cant be used in the presence of inflammable anesthetics or products that may
cause explosion.
After finish life-cycle of the equipment, this equipment is throw way, so it should to be destroyed
in appropriate area (according to the legislation of the country).
Observe the countrys legislation pertinent to installation and waste disposal.
NOTE:
The safety degree in the presence of inflammable anesthetic mixtures with air, oxygen or nitrous
oxide is not applicable to this equipment.
3

WARRANTY OF THE EQUIPMENT


This warranty covers products manufacturing faults for the warranty period and standards
specified on the Warranty Certificate. The Warranty Certificate is attached to product, and must be
completed by the Authorized Technician when equipment is installed.
Doubts and information: GNATUS Call center (55-16) 2102-5000.

IMPORTANT SAFETY CONSIDERATIONS


- Never touch the steam external outlet (12) and, after the sterilization is finished, never touch the
internal autoclaves surfaces. The inappropriate use may cause burns. GNATUS may not be hold
responsible for accidents caused by inadequate operation.
- Dont allow patients, and most specially children, to touch the autoclave.
- Periodically check that the doors door latch is correctly in place before operation. If this procedure
were not observed, the sealing ring may come off.
- Before using the autoclave, verify that the internal chamber is clean, also check if the filter (17A)
is completely clean, free of residues (page 6).
- When raising the door latch, the autoclaves door must open without difficulty.
- Always make sure that the depressurization is complete: the manometers needle must be in the
0 (zero) position. Never use force to open the autoclave!
- If one of the safety devices were activated (unexpected steam leak) usually caused by the
internal steam outlet blockage or by an obstruction in the pressure control valve, wait until the complete
depressurization before switching the autoclave off.
- Before turning your autoclave on, install the fuses included with the equipment.
If the voltage were changed, these fuses may be changed.
- Use only the hose supplied by GNATUS, or one similar, so as to prevent if from melting when in
contact with steam, which may cause the external steam outlet obstruction.
- We recommend you to read this manual entirely. Use is as consultation manual.

GENERAL DATA
Maintenance of the equipment

The equipment doesnt need preventive maintenance, however when there is any problem that
this manual cant solve, ask for an authorized Gnatus technician.

List of pieces and circuit scheme

Gnatus Company declares that the supply of the circuit scheme, list of pieces or any other
information that propitiate technical attendance for the user, can be request if there is an agreement
between the user and Gnatus Company.

Protection against current overload

The equipment is protected from current overload by three fuses.


- For 127V~: 3 fuses, 2 of 15A, 1 of 0,5A;
- For 220V~: 3 fuses, 2 of 08A, 1 of 0,5A.
NOTE: : To install the fuses, see page 12.

GENERAL DATA
Caracteristics and operation
Manometer

Its used for temperature and pressure


control, and its graduation range from 0 kgf/cm
to 4.0 kgf/cm and from 100C to 152 C.
Manometers tolerance variation
Divide the manometers scale by three equal
parts. For the first part, the tolerance is 3%, for
the second part is 2%, and for the third is 3%.

Sealing ring

Besides sealing the door, it also works as a


safety device, relieving chambers internal
pressure in casa of overpressure.
If the sealing ring were moved from its place,
call a Gnatus authorized technician to carry out
the maintenance of the equipment.

Safety seal and anti-vacuum valve

Placed in the internal part of the door, they


diminish the pressure in the chamber, in case it
exceeds. The anti-vacuum valve has the same
function, and it also reduces the pressure of the
chamber if it were negative (vacuum).
NOTE:
If during the sterilization procedure the valves
were opened, the operator can reset them (page
16). However, the appropriate cooling time must
be observed in order to prevent accidents.
If the valves were opened repeatedly, call a
Gnatus authorized technician.
NOTE: Execute the replacement of the
safety seal and anti-vacuum valve within the
period of 36 months. If it is necessary to
do it before the period indicated, request
the presence of an authorized Gnatus
technician.

Vlvula anti-vaco

Sello de seguridad

Control panel

Found in the front part of the autoclave, it


contains all controls necessary to operate the
equipment.
4 A Cycle start
4 B Drying
4 C Cancel
4 D Equipment ON
4 E Heating / Sterilization
4 F Drying
4 G Cycle end

4A
4B
4C

4D
4E
4F
4G

GENERAL DATA
Caracteristics and operation
Steam internal outlets

17B

Theyre holes in the rear wall of the internal


chamber (17), they take the steam to the
pressure control valve.
Note: In the lower hole theres a protective
filter (17A), it prevents solid residues from
entering the steam outlet.

17
17A

External steam outlet

The 5/16" hose fitting, found in the autoclave


rear part, takes the steam outside.

External steam outlets hose

This accessory must be connected and


fixed to the external steam outlet using a clamp
(11). It takes steam out.

11

Trays and support

The support (18) and the two trays (16) keep


the materials to be sterilized apart from the
chambers side walls, from the bottom and from
the water.

16

WARNING:
The material to be sterilized cant be in contact
with any of the internal walls of the chamber.

18

Graduated glass

Its used to dose the quantity of distilled water


needed for each sterilization cycle.

Tray holder

Its a device used to handle trays, avoiding


direct contact with the operator.

GENERAL DATA

15

16

14
13

01

12

02

26

03

10

09

04

08
07
06

05

10 - Steam outlet hose


11 - Hoses fixing clamp
12 - Steam external outlet
13 - Anti-vacuum valve
14 - Sealing ring
15 - Safety seal
16 - Trays
26 - Main switch

01 - Autoclaves body
02 - Front door
03 - Manometer
04 - Control panel
05 - Door latch
06 - Plug
07 - Cord
08 - Voltage inverter switch
09 - Fuse holder

11

TECHNICAL SPECIFICATIONS
Tension of Feeding
127V~

+20%
-20%

220V~

+20%
-10%

Frequency
50/60Hz
Power
1600VA
Sterilization pressure
1,7 kgf/cm2 0,4 (1,3 to 2,1kgf/cm2)
Operation temperature
128C 5 (123 to 133C)
Sterilization time
16 minutos

Protection fuses
F1 - 127 / 220V~ = 0,5A
F2 and F3 - 127V~ = 15A
F2 and F3 - 220V~ = 08A
Diameter of the chamber
214 mm
Chambers depth
318 mm
Water consumption
150 - 200ml
Net Weight
23 Kg
Gross Weight
26 Kg

Degree of protection
Tipo BF
Water leak protection
IPX0
Complete cycle
60 minutes
These values are approximate,
they may change depending on
the materials type and volume,
energy oscillation, etc.
Manometers tolerance
variation
Divide the manometers scale
by three equal parts. For the
first part, the tolerance is 3%,
for the second part is 2%, and
for the third is 3%.

Volume
12 Liters

Standards
This equipment was designed, manufactured and tested in compliance with the following standards:
NBR-IEC srie 601-1 Equipamento Eletromdico - Parte 1: Prescries gerais para segurana;
EN 980:2003 (Ed. 2) - Graphical symbols for use in the labelling of medical devices;
ISO 14971 - Medical devices - application of risk management medical devices;
ISO 9687: 1993 - Dental equipment - graphical symbols;
ISO 7494 - Norma dental units;
ISO 13485-2 - Quality systems - medical devices;
ISO 780 - Packaging - pictorial marking for handling goods;
ISO 11144 - Norma dental equipment - connections for suply and waste lines;
EN 60601-1 (1990);
Amendment 1 EN 60601-1 (1992);
Amendment 2 EN 60601-1 (1995);
Amendment13 EN 60601-1 (1995);
EN 60601-1-2 (2001);
NBR IEC 601-1 (1994);
Emenda 1 NBR IEC 601-1 (1994);

CISPR 11, edio 3.1 (ago/1999);


IEC 61000-4-2 (1999);
IEC 61000-4-3 (1998);
IEC 61000-4-4 (1995);
IEC 61000-4-5 (1995);
IEC 61000-4-6 (1996);
IEC 61000-4-11 (1996);

Special precautions and installation conditions


- Install the autoclave on a firm and leveled surface, approximately at 80 cm. from the floor. Leave
enough space for handling the materials to be sterilized.
- Check that the electricity network is compatible with the equipments specification. If its not
compatible, change the voltage selection key and replace the fuses (see page 12).
- Before turning your autoclave on, install the fuses included with the equipment (see page 12).
- Check if the grounding conductor is correctly connected.
- Never use cord extensions, connectors or voltage transformers.
NOTE: This information are also contained in the Installation and maintenance manual owned by the
GNATUS authorized technician.

TECHNICAL SPECIFICATIONS
Transport and storage conditions

The equipment must be transported and stored with the following observations:
- Carefully, should not suffer drop and neither receive impact.
- With the side of the arrow upward.
- With humidity protection, not to expose to rains, sparkling of water or humidified floor.
- With maximum piling up of 04 units for storage.

Packing symbols

04

Packing to be transported and / or


stored with care (should not suffer
drop and neither receive impact).

Packing to be transported and / or


stored with care.

Packing to be transported and / or


stored with the harrows up.

The packing must be stored and


transported away from direct sun
light exposure.

Packing to be stored with a maximum


stack of 04 units.

Temperature limit for the packing


to be stored or transported.

Product symbols
Indicates that the device is on

Indicates drying

Indicates that the device is


in heating mode

Indicates cycle start

Indicates that the device is


in sterilization mode

Indicates cycle end

Indicates cycle cancellation

Warning See manual

TECHNICAL SPECIFICATIONS
Dimensions (mm)

10

TECHNICAL SPECIFICATIONS
Contents of the marks accessible and not accessible

11

TECHNICAL SPECIFICATIONS
Contents of the marks accessible and not accessible (Legends)
01 - Products identification label
02 - Attention: Hot surface
03 - Fuse 15A (F2-F3): Use this fuse for 127V power supply.
04 - Attention: Before turning Autoclave on, remove this cap and install the fuses.
05 - Attention: This equipment leaves the manufacturer with the selecting key of voltage at 220V
being so be sure that the net is compatible before connecting. If the voltage will be 127V please invert
the voltage selection switch position.
06 - Fuse 08A (F2-F3): Use this fuse for 220V power supply.
07 - Before first use, please remove the protector film, which wraps this stell tray.
08 - Fuse 0,5A (F1) - 127/220V~
09 - Maximum power per resistance: 800W - Warning: 2 Resistances

INSTALLATION

Preparing the autoclave (Steps previous to operation)


Voltage change
For your safety, the autoclave is factory-set to
220V. For this reason, you must check the suitability
of the electricity supply. If the electricity network
available is 127V, change the voltage switch.
Change the voltage switch (08) according to the
local electricity network (127 or 220 V).
WARNING:
Dont change the voltage while the device is on.
Then, connect the equipment to the electricity
supply and proceed as follows.

08

Installing or replacing fuses

NOTE:
For reasons of safety, the F2/F3 fuses are not
installed in factory. Therefore, these fuses must be
installed before turning the equipment on. Using a
screwdriver, loosen the fuse holders cap (09), and
place the fuses (19) in specified position. To replace
the fuses, repeat the procedure.
WARNING:
Before installing or replacing the fuses, disconnect
the equipment from the electricity supply.
NOTE:
Fuses that come with the autoclave:
- 2 fuses for 127V~ ( F2/F3 15A) + 1 reserve;
- 2 fuses for 220V~ ( F2/F3 08A) + 1 reserve;
- 1 fuse 0,5A *installed in the unit * + 1 reserve.
WARNING: After replacing the fuses for the first
time, we recommend to reserve some fuses for future
replacements.

12

19

09

INSTALLATION
Preparing the autoclave (Steps previous to operation)
Preparation
Connect and fix the rubber hose (10) using
the clamp (11) to the external steam outlet (12)
found in the autoclaves rear part. The other
end must be placed in a water container (20)
or in a drainage which can stand high
temperatures.
In order to prevent the obstruction of water
and steam flow, keep the hose straight,
avoiding the formation of bends and siphons.

12

11
10
06

WARNING:
Use only the hose supplied by GNATUS, or
one similar, so as to prevent if from melting
when in contact with steam, which may cause
the external steam outlet obstruction.
Open the front door raising the door latch,
then push it to the left and check if the trays
(16) are in their place. There must be two trays,
and the biggest must be places in the upper
part.

20

16

WARNING:
Before first use, please remove the
protector film, which wraps this stell tray (27).
The female socket must be of the type used
for computers, with three pins.
Check the ground connection.

WARNING:
The ground connection is very important to
the operators safety and for the equipments
warranty. For this reason, the middle pin (ground)
cant be removed or cut.

27

13

OPERATION
How to use de autoclave
Raise the door latch, push it to the left, and
check the cleanliness of the autoclave.
Connect the energy plug (06) to the socket,
the green led (4D) in the front panel will come on,
indicating that the autoclave is on.
Using the graduated glass (21), pour inside
the autoclave the following quantities of water:
Autoclave partially full 150ml
Autoclave completely full 200ml
WARNING:
Use only distilled water. If you use other
water than distilled, it may cause the obstruction
of the internal valves and stain the instruments,
and the warranty will not be applicable.
All material being sterilized must be previously
washed, rinsed and dried.

IMPORTANT: Dont use kraft paper for


wrapping the materials to be sterilized.
(see page 35 types of wrappings for
autoclave sterilization)
* Observe each countrys standards for
preparation of material to be sterilized.
Distribute the wrappings at equal distances
(22) leaving enough space between them so as
to allow the steam to circulate.
WARNING:
The material to be sterilized (22) cant be in
contact with any of the internal walls of the
chamber.

21
22

Close the autoclaves door pushing it against


the tank, pushing the door latch to the right and
then lowering it. Check that it is correctly closed
(23)
WARNING:
In order to avoid accidents and burns, its
very important that the door is well closed.

22

23

14

OPERATION
How to use the autoclave (Control panel)
Press the cycle start key (4A), the
autoclave will automatically start to heat, and
you will hear 2 beeps, and a green LED (V)
will come on. The heating time depends on
the quantity of the material to be sterilized.
After the heating has reached a
temperature of 128C 2, the autoclave will
enter in the pressure sterilization mode, with
a pressure from 1,7 0,4 kgf/cm (24), the
LED (4E) will change from green (V) to yellow
(A), and it will stay that way for 16 minutes.
After the sterilization cycle is finished,
the valve will automatically open and the
autoclave will be depressurized. After three
minutes you will hear ten beeps. The blue
LED (4F) will come on, indicating the drying
mode. In this moment, the door shall be opened
and it must remain ajar (25) so as to dry the
sterilized material.
IMPORTANT:
If it was necessary to stop the
sterilization cycle, press the cycle interruption
key (4C). Wait until the autoclave is completely
depressurized before opening the door.
WARNING:
When in drying mode, the autoclaves
door must be kept ajar (25). During this period,
the user mustnt touch the internal parts of
the autoclave, so as to prevent accidents
and burns. The drying cycle lasts half an
hour. Finally, the autoclave will emit three
beeps, the green (V) led (4G) will come on,
indicating the end of the cycle.
If the material were not completely dry,
you can select the extra dry function (4B).
Leaving the door ajar (25) press the (4B)
key, the blue (4F) LED will come on, indicating
the start of a new drying cycle that will last
30 minutes. If the operator notes that the
material is already dry, he can cancel the
cycle by pressing the (4C) key.

24

IMPORTANT:
In case of power cut, the operation of
the autoclave will be interrupted, the yellow
(4G) LED will come on and it will remain
blinking, activating the discharge valve for 3
minutes. During this time, no function will be
available. When the power were restored,
check the water level, if necessary, replenish
the specified quantity and start the cycle
again.

25

15

OPERATION
Resetting the valves (safety seal and anti-vacuum valve)
If the valve (13) were opened during the
sterilization process, the operator can reset it.
Open the autoclaves front door and wait until
its cold. Proceed as described in the drawing,
pressing them until they go back to their original
position.
If the valves were opened repeatedly, call
a Gnatus authorized technician.

WARNING
HIGH TEMPERATURE. Before carrying out
the procedure described above, wait until its
cold.

13

MAINTENANCE
Recommendations for the dental equipment maintenance

Your Gnatus equipment has been designed and developed according to the standards of modern
techology. Similarly to other kinds of equipment, it requires special care, which is many times neglected
due to several reasons and circunstances.
Therefore, here are some important reminders for your daily routine. Try to follow these simple
rules, which will save you a lot of time and will avoid unnecessary expenses once they start making
part of your working procedure.

Precautions
Check the voltage before connecting the device to electricity supply (see page 12)
- Install your apparatus on aproper place protected by solar rays and humidity.
- The plastic parts cant be in contact with most chemical substances used in dentistry treatments,
i.e. acids, mercury, acrylic liquids, amalgams, etc.
- Dont touch the autoclaves internal surfaces when its in drying mode.
- Before turning your autoclave on, install the fuses included with the equipment.
If the voltage were changed, these fuses may be changed.

To clean the equipment, we recommend the use of BactSpray (Reg n MS:


3.2079.0041.001-5) or any other similar product:
Active component: Benzalkonium chloride (tri-quaternary ammonium)
Solution 50%................................................. 0.329%
Chemical composition: Butyl Glycol, Decyl polyglucose, Sodium Benzoate, Sodium
Nitrate, Essence, Deodorized Propane / Butane, demineralized Water.
For more information concerning cleaning procedures, see manufacturers
instructions.
WARNING:
In order to prevent risks and damages to equipment, make sure that the liquid does
not enter into the unit.
The application of other solvent-based cleaning products or sodium hypochloride
isnt recommended, because they may damage the equipment.

16

MAINTENANCE
Internal cleaning (sterilization chamber and trays)
To clean the interior of your equipment, sterilization chamber and trays, we recommend the use of
SDS - 1000 n Reg. MS: 3.02053-1.
Follow manufacturers directions for cleaning.
Products composition: inorganic-acids, fluoride and anionic tense-active based products.
Important: The alteration of the visual aspect inside the Autoclave tank is a normal
feature of the product owing to the heating in the region.

Piping cleaning
Monthly the internal components of the autoclave must be cleaned and disinfected.
To clean the interior of your equipment, tubing, valves and filters, we recommend the use of INOX
CLEAN n Reg. MS: 3.02053-1.
Follow manufacturers directions for mixing the product.
Products composition: inorganic-acids, phosphate, acrilato de sdio. Mix the product following the
instructions given by the manufacturer.
Pour the solution in the chamber and start the sterilizing cycle. Repeat the sterilizing operation using
only distilled water. When these cycles are finished, remove the remnants of the cleaning solution
using a clean piece of cloth moistened with water.

Cleaning the filter

- The autoclaves filter (17A), (see page 06) shall be cleaned daily.
- In order to clean the filter, remove it via the handle (17 B), and then wash it in running water.
- Dont use sharp objects, which could damage the fabric.

Wrapping types for sterilization by autoclave

The materials used for wrapping the pieces to be sterilized must keep the sterilization after the
sterilization process and during storage.
These materials must:
- allow the sterilization process;
- be appropriate for the material and the sterilization process;
- allow the drying, ventilation and dispersion of the sterilizing agents;
- not have lighter and darker parts;
- be resistant to wetness, be free of holes and scratches;
- dont release particles.
NOTE:
GNATUS recommend the use of surgical paper, because the microbial barrier of this material is
above 90%, and it permits the operator to visually check the state
of the instruments. It also has chemical indicators of the process.
GNATUS prohibit the use of kraft paper for wrapping
materials to be sterilized, because it isnt manufactured with this
purpose. Its use has many disadvantages when compared to
other materials, such as irregularities and uneven texture, what
affects the resistance, turning it inadequate for its use as microbial
barrier. Besides, it frequently contains starch, colorings and other
toxic products that can stain the parts. It can also contain
Alkylthiophene, which, during the sterilization process may give
off an unpleasant smell, which causes nausea and headache if
breathed.

17

MAINTENANCE
And besides all these disadvantages, during the sterilization process, this kind of paper releases
pieces of fiber that can clog the autoclaves valves and piping system.
Observe each countrys standards related to wrapping preparation.

Suggested method for wrapping instruments and materials


The wrapping technique for wrapping instruments and/or materials before the sterilization process
is described in the following sequence of illustrations.
This folding sequence isnt unintentional. It makes its handling easier and it prevents contamination
when the wrapping is opened.

A - Place the piece diagonally and in the middle of the wrapping


material;
B - Hold the a corner, fold it and cover the piece;
C - Hold the b corner and fold it over the piece, covering it and
bending the end upwards;
D - Repeat the last step with the d corner;
E - Fold the c corner in the direction of the operator;
F - Take the c corner and insert them under the folds made in
3, 4 and 5;
G - Let the c corner outside, in order to prevent the sterilized
material from being contaminated and making the wrapping opening
easier;
H - If necessary, seal the wrapping using cord or crepe.
Observe each countrys norms related to wrapping
techniques application.

18

CONTROL OF THE STERILIZATION PROCESS


Control of the pressurized and saturated steam sterilization process
The BIOCLAVE 12L autoclave is tested in the factory in order to verify its sterilization capability. Its
used an ATTEST Biological Indicator (BI), in order to comply with requirements from ANVISA (Sanitary
Administration National Agency) in Brazil, and from international organizations, such as AAMI (Association
for the Advancement of Medical Instrumentation) and CDC (Centers for Disease Control and Prevention)
The BI test consists of an ampoule that contains and strip impregnated with 105 Geobacilus
sthearothermophillus spores. These organisms are particularly resistant to the steam sterilization
process.
All BIOCLAVE 12L autoclaves are tested at the production line following a procedure that ensures
the equipments efficiency to provide safe and sterilized articles.

Maintenance of the qualification

In order to maintain the quality and the sterilization efficiency during the equipments regular use,
the BI test must be carried out routinely. To perform this test in the same place where the autoclave is
installed, it must be used the ATTEST control system, which consists of BI test ampoules and an
incubator, which allows the incubation of the test ampoule and the verification of the final result, 48
hours later.

Testing procedure using biological indicators


1) Make a test package that allows the steam to penetrate into it, place a BI inside it, and place the
test package in the coldest part of the autoclave (in this case, the coldest part of the BIOCLAVE is the
upper part, near the door), with the load for a routine cycle.
2) Identify the test package that contains the BI, remember that this package shall be opened to
remove the BI.
3) Once the cycle is finished, wait until the package is cold, then open it and remove the BI; break
the BI ampoule inside the specified incubator and incubate it for 48 hours.
4) At the same time, place a BI ampoule (one that hasnt passed through the sterilization process)
into the incubator, in order to verify the incubator operation and also to control the viability of the spores
used in the test; check the BIs (both the positive control and the test ampoule), follow the directions
provided by the manufacturer.
5) Record test data and its results in a record system
6) If the BI tests results were positive, dont use the equipment until it has been checked and
repaired by the authorized technical service.

Attest biological monitoring system for steam sterilization process

For more information about biological and chemical control, access www.3M.com/br/hospitalar

19

TROUBLESHOOTING
Imprevisto
Problem:

Probable Cause:

Solution:

- T h e a u t o c l a v e d o e s n t - The cord is disconnected


switch on
- Power cut

- Connect the cord to the socket


- Wait until energy is back

- Burnt fuse(s)

- Replace fuse(s)

- Bad contact in the socket

- Replace socket

- The ON/OFF switch is OFF.

- Turn the ON/OFF button ON.

- The autoclave cant reach or - The voltage supplied is lower


than the one required by the
maintain the pressure high
autoclave

- An electrician must check and


adapt the electricity supply

- Leakage in the safety seal

- Disconnect the autoclave, and


remove the safety seal and the
anti-vacuum valve, clean them
and place them again. If it were
necessary, replace them.

- The door latch isnt in position,


causing leakage

- Place the door latch pushing it


downwards

- Not enough water

-Check that the volume of water


is the specified

- Overloaded chamber

- Use up to 70% of the chamber,


leaving spaces between the
packages

-The autoclaves door cant be -The autoclave


depressurized
opened

isnt

yet

-Wait
the
depressurization

autoclave

* In case of continous fails, request to the attendance section to the customer a


authorized technician Gnatus.
ATTENTION:
Any other problem that could happen with its equipment, should only be repaired by an Authorized
Technician GNATUS.
The bad use, negligence or maintenance not made by technician authorized by the maker, will imply
in the loss of the warranty.

FINAL CONSIDERATIONS
The most important aspect related to equipment care is that concerning spare parts.
To guarantee the life span of your equipment, use only original Gnatus spare parts. They are
sure to follow the technical specifications and standards required by Gnatus.
We must also point out to you our chain of authorized dealers. Only dealers that make part of this
chain will be able to keep your equipment constantly new for they count on technical assistants who
have been trained and on spedific tools for the correct maintenance of your equipment.
Doubts and information: GNATUS Call center (55-16) 2102-5000.

20