13.50.3
START
MANU. MEAS.
ESC
3427
CAL
SET
MODE
ENT.
DATA IN DATA OUT
of storage tanks
OIL CONTENT ANALYZER
DRAIN
OCMA-300 Automotive Monitoring waste-water discharges
4.30.1
OCMA-300
3154
12.40.2 Quality Assurance Measuring residual oil on textiles, metal
parts, etc.
SPECIFICATIONS
Application: Oil contamination of fresh Sample/Solvent Volume: 20 mL/10 mL Model SR-300 Solvent
and salt water; soil analysis Extraction Method: Built-in extractor for Reclaimer (option)
Principle: Solvent extraction, NDIR oily water samples Designed especially for recycling
analysis (Infrared spectroscopy) Display: S-316 solvent, it features a 2-
Detector: Pyroelectric sensor Measured value; 3 digits LCD with back- layer column of activated carbon
Measuring Range: 0 to 200 mg/L light and activated aluminum. High
Resolution: Message; Character display LCD with capacity, simple operation, and
0 to 99.9 mg/L; 0.1 mg/L back-light (16 2 characters) needs no electricity.
100 to 200 mg/L; 1 mg/L Functions: Data memory up to 50 data,
Repeatability: Self diagnostics, Stable type display,
0 to 9.9 mg/L; 0.2 mg/L 1 digit Interactive operation, Calendar clock
10.0 to 99.9 mg/L; 2 mg/L 1 digit Output: Digital display, RS-232C and
100 to 200 mg/L; 4 mg/L 1 digit parallel printer port
Measurement: Automatic after Ambient Temperature: 0 to 40
introduction of the oily water sample or Power Requirement:
manual operation, selectable 100 to 240 V AC 10%, 50/60 Hz
Calibration: Automatic zero and span 100 to 120 V AC; approx. 80 VA
calibration after the oily water sample is 200 to 240 V AC; approx. 120 VA
introduced to the instrument. Dimensions:
Extraction Solvent: 200(W) 315(D) 342(H) mm
HORIBA S-316 solvent 7.8(W) 12.4(D) 13.5(H) in
Weight: Approx. 7 kg/15.5 lb
Please read the operation manual before using this product to assure safe and proper handling of the product.
The contents of this catalog are subject to change without prior notice, and without any subsequent liability to this company.
The color of the actual products may differ from the color pictured in this catalog due to printing limitations.
It is strictly forbidden to copy the content of this catalog in part or in full.
All brand names, product names and service names in this catalog are trademarks or registered trademarks of their respective companies.
Transmission %
B-Heavy Oil
are, then, measured by the infrared-absorption method. leum and (2) Type S-316 Solvent.
60
All types of petroleum absorb
40
infrared radiation in the range of
20
3.4 m. Type S-316 Solvent does
The new HORIBA OCMA-300 oil content analyzer comes loaded with a welcome range of innovative features. Add Type S-316 Agitate C-Heavy Oil
Separation not absorb infrared radiation in 0
Solvent thoroughly 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
A Powerful built-in computer makes the new version a joy to use. sample water sample water
this range. As a result, if mea-
100
m
Transmission %
lcomponents 80
oil m range, all the oil components S-316
solvent 60
everything for you from extracting the oil components to measurement and drain. The whole process is fast, (10 mL) components
extracted
that have been extracted can be
40
sure, super-accurate and, of course, no more troublesome range-switching. Now anyone can make precision measured precisely, with abso- 20
lutely no influence from the pres- 0
analyses without special training. ence of the S-316 Solvent.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
m
For environmentally-friendly measurements, HORIBA has developed Type S-316 extraction solvent for use with
the OCMA. The S-316 solvent can also be recycled by using HORIBA's optional Solvent Reclaimer.
The OCMA-300 is much more versatile, even simpler to use. HORIBA is proud to introduce to the world
a truly user-friendly oil content analyzer.