xendos 1800 Oxygen Analyser

The xendos 1800 is a stable,


accurate and highly specific oxygen
analyser for safe areas.

Designed for safe area oxygen


analysis
Low maintenance and
re-calibration requirements
Range of alarm outputs to aid
integration with other systems
Easy to set up and operate
Special versions for solvent
bearing samples
Special high flow rate cell
option

Specification
Gases Measured: O2

PERFORMANCE
Technology: Paramagnetic transducer
Range: 0-100% O2
Intrinsic Error (accuracy): <0.2% of reading or 0.05% O2 1
Linearity: <0.05% O2 2
Repeatability: <0.1% of reading or 0.05% O2 1
Response time (T90) 4 to 8 seconds 3
Zero drift/ week: <0.05% O2
Span drift/ week: <1% of reading or 0.05% O2 1
SIGNAL OUTPUTS
Analogue: One 4-20 mA, isolated & one 0-1V dc, non-isolated. Ranges selectable from 0 - 2.5, 5, 10, 25 and 100% O2
Alarms: Concentration Two volt free changeover relays rated 250Vac/3A or 28Vdc (non inductive) maximum and 5V/10mA ac/dc minimum
Sample flow fail One volt free changeover relay rated 250Vac/3A or 28Vdc (non inductive) maximum and 5V/10mA ac/dc minimum
PHYSICAL
Dimensions:(W x D x H) 448mm x 229mm x 235 mm / 17.6 x 9 x 9.25
Weight: 26 kg / 57 lbs
Hazardous area
Classification: Non-Hazardous Areas only
Ingress Protection: IP 66 / NEMA 4X
Mounting: Wall or Panel

1 whichever is the greater


2 inherently linear, value dependant on calibration gases
3 dependant on configuration

LEADERS IN GAS ANALYSIS


Ambient Conditions Sample Wetted Materials
Temperature:

Stainless Steel Pipework


Stainless Steel Pipework
Solvent Reistant Cell +

+Hastelloy pipework
Solvent Resistant Cell
Operating: -10C to 50C/14F to 122F

Standard Cell + Flow

High Flow rate cell /


Storage: -20C to 55C/-4F to 131F

Standard Cell +
Standard Cell +

Standard Cell +
Sample Pump
Atmospheric Pressure: Material Of

Analyser
79 to 124 kPaa/11 to 18psia Construction

AFCD
Alarm
Basic

BPR
(for operation up to 2000m altitude.)
Warm Up Time Beryllium-Copper 4
Borosilicate Glass 4 4 4 4 4 4 4 4
4 hours at an ambient Bonded Borosilicate 4
temperature of 20C (68F) Glass Fibre
Brass 4
Phospher Bronze 4
Fluorocarbon Rubber 4 4 4 4 4
Hastelloy C-276 4
Nickel (electroless) 4 4 4 4 4 4 4 4
Neoprene Rubber 4
Glass Filled Nylon 12 4
Polysulphone 4
Power Supply Platinum 4 4 4 4 4 4 4 4
Platinum / Iridium alloy 4 4 4 4 4 4 4 4
100 to 240V ac 10% - 50/60Hz -50VA max. Glass Filled 4
Note: The internal pumps are AC voltage Polypropylene
Polypropylene 4 4 4 4 4
dependant and are supplied as 110Vac 50, PVC 4
110Vac 60Hz, or 230Vac 50Hz. Ensure that the PVDF 4 4
correct AC input voltage is supplied to an Gold Plated Silver 4
302 / EN58A SSteel 4
analyser fitted with a sample pump. 303 Stainless Steel 4 4 4 4 4
316 Stainless Steel 4 4 4 4 4 4 4 4
Viton (325 cell) 4 4 4 4 4 4
Viton - A 4
Chemraz (364 cell) 4 4
PTFE 4 4

Sample Gas Conditions


Basic with with with S.Steel or
Configuration Analyser AFCD AFCD and AFCD and Hastelloy/PFA
Sample Pump BPR High Flow cell or bypass
Inlet Pressure 0.04 psig/ 1 to 5 psig/ -0.03 to 1 psig/ 17 to 22 psia/ 0.05 psig/
0.3 kPag minimum 7 to 35 kPag -0.2 to 7kPag 119 to 154 kPaa 0.4 kPag minimum
Flow Rate 50 to 250ml/min 1.2 to 3.5 l/min 1.6 to 1.8 l/min 1.0 to 2.0 l/min 50 to 70 l/hour (60 l/hour nominal)
Vent Pressure 11.5 to 18.0 psia (80.5 to 126kPaa) - DO NOT RESTRICT ANALYSER VENT
Dew Point Non-condensing at ambient temperature
Temperature -10C to +50C (+14F to +122F)
Particulates <3m (micron)
Condition Clean, non-flammable * and free from oil/condensate**
Connections NPT. INT Inlet/Outlet Connectors (Female). (6mm option available)
Adjust pressure and sample flow externally to provide sample flow rate
* For Flammable samples use the xendos 1900
** For Corrosive samples use a solvent resistant cell option
AFCD - Automatic Flow Control Device BPR - Back Pressure Regulator

Front View Side View


448mm / 17.6 229mm / 9

407mm / 16

174.2mm
/ 6.86 235mm
/ 9.25
Analyser
The xendos 1800 is fitted as standard Analogue outputs: Concentration Alarms:
with the following features: Isolated 4-20mA, 600 ohm maximum Two volt free changeover relay outputs
impedance, non-isolated 0-1V d.c. and LED. User configurable as high or
minimum impedance 470 ohms. The low, set point field adjustable over the
analogue output signals can be set to one range 0-100%O2
of two pre-selectable measurement ranges. Range Indication:
Volt free changeover relay outputs and
LED

Please note: this analyser is not suitable for use with flammable samples,
use the xendos 1900

Sample Conditioning Options


The xendos 1800 is available with a Where the sample is non-corrosive analogue output can be pre-configured to
number of sampling options, to suit a choose from the columns headed: be driven high (>20mA), or low (<4mA,
variety of applications. Your choice will Standard Cell. default).
affect both sample wetted materials, and Choose Automatic Flow Control Device Choose a sample pump if the sample
the required sample condition. (AFCD) c/w internal sample filter to simplify supplied to the analyser is at an
sample flow control. insufficient pressure to generate a suitable
Choose Automatic Flow Control Device flow rate.
(AFCD) and Back Pressure Regulator Where enhanced pipework integrity is
(BPR) to reduce the effects of changes in
required, or where the sample is corrosive
barometric or vent pressure when
measuring high concentrations of oxygen. or solvent bearing, or where a high
sample flow rate is required, choose from
Choose the sample flow fail alarm option the column headed Special Cell & Metal
to provide a volt free changeover relay Pipework.
output and LED indication when the flow
falls below the minimum required level
(during a Sample Flow fail condition, the

Display
The xendos 1800 is available with either a 00.0% to 100.0% O2, or a 4 1/2 digit 100.00% O2
3 1/2 digit display, capable of displaying display capable of displaying 00.00% to

Sample Inlet
Sample Inlet fittings can be supplied as: /4 NPTF 316 stainless steel fittings,
1

6mm OD 316 stainless steel fittings,


6mm OD Hastelloy fittings

Enclosure Fittings
Three options are available, the choice Purge fittings: 1/4 NPTF Inlet/Outlet Breather port: A Breather Port with a 1/4
depends on application. connectors can be fitted to the sample NPT connector can be fitted to the
compartment of the enclosure to provide sample compartment of the enclosure to
None: sealed enclosure.
protection against internal toxic and/or provide pressure relief protection resulting
corrosive sample gas leaks from internal sample gas leaks.

Conduit Entry
The xendos 1800 is supplied with a 3
/4 NPT conduit entry
choice of cable entry. M20
Choose from: Pg13.5

Mounting
The xendos 1800 is available for either The flush panel mounting analyser is additional mounting brackets.
wall or panel mounting. supplied with a panel cutout and

Quickstart Manual
The xendos 1800 is supplied with a
QuickstartTM manual in a variety of
languages.

Service Manual
A service manual containing technical tool and test equipment lists, and electrical
descriptions, fault diagnosis information, drawings is available. It is intended for use
parts removal, refitting and test instructions, by Servomex trained service personnel.
Enquiry & Ordering Information
To Specify the analyser you require, enter your details below, tick one
Analyser option in each box and post or fax to your local Servomex company,
xendos 1800 Analyser o agent, or representative.
(01800B1GEN)
Name:
Company:
Address:

Order/Enquiry Number

Please send me details of planned maintenance services &


recommended spares o

1
Standard Cell, Standard Cell, Special Cell
(Viton Pipework) (Viton Pipework with Pump) & Metal Pipework
Basic Analyser o (01) with AFCD & Pump (110V/50Hz) o (08) Standard cell, stainless steel pipework o(02)
with flow alarm o (03) with AFCD & Pump (110V/60Hz) o (09) Solvent resistant cell, stainless steel
pipework o(14)
with AFCD o (04) with AFCD & Pump (230V/50Hz) o (10)
Solvent resistant cell, with Hastelloy
with AFCD & BPR o (05) with AFCD, flow alarm & Pump pipework o(15)
with flow alarm & AFCD o (06) (110V/50Hz) o (11) High flow rate cell, stainless steel
with flow alarm, AFCD & BPR o (07) with AFCD, flow alarm & Pump pipework o(16)
(110V/60Hz) o (12) High flow rate, solvent resistant cell,
with AFCD, flow alarm & Pump stainless steel pipework o(17)
(230V/50Hz) o (13) High flow rate solvent resistant cell, Hastelloy
pipework o(18)
Standard cell, stainless steel bypass o(19)
Solvent resistant cell, stainless steel bypass o(20)
Solvent resistant cell, Hastelloy bypass o(21)

Display 3
3 1/2 digits o (001)
4 1/2 digits o (002)

Sample Inlet 4
/4 NPT
1
o (0)
6mm OD stainless steel o (1)
6mm OD Hastelloy o (2)

Enclosure 5
None o (0)
Purge Fittings o (1)
Breather port o (2)

Conduit Entry 6
/4 NPT
3
o (1)
M20 o (2)
PG 13.5 o (3)

Mounting 7
Wall mounted o (0)
Flush panel mounted o (1)

Quickstart Manual 8
English o (1)
French o (2)
German o (3)
Spanish o (4)
Portuguese o (5)
Dutch o (6)
Italian o (7)
Japanese o (8)
Chinese o (9)

Service Manual 9
None o (0)
Service manual (English) o (1)

LEADERS IN GAS ANALYSIS


Performance Approval EC Directive Compliance
The xendos 1800 complies with
EN50104:1999 Electrical apparatus for the The xendos 1800 complies with the Low
detection and measurement of oxygen. Voltage Directive 73/23/EEC and the EMC
Directive 89/336/EEC (as amended by
Directive 92/31/EEC), both as amended by
Directive 93/68/EEC.

It conforms to the following harmonised


European standards for product safety and
electromagnetic compatibility:

EN 50081-1: Generic emission standard

EN 50082-2: Generic immunity standard

EN 61010-1: Safety requirements for


electrical equipment for measurement,
control and laboratory use.

This product is rated for Installation


Category II in accordance with IEC 664.

This product is rated for Pollution


Degree 2 in accordance with IEC 664.

Performance Specification Continued


Configuration Units Base with with AFCD with AFCD High Flow Transducers,
Analyser AFCD and Sample Pump and BPR Standard or Solvent resistant

Response time (T90) Seconds <4 <7 <8 <7.5 <5


(200ml/min) (60 l/hour)
Noise (peak to peak) % O2 <0.04 <0.05 <0.05 <0.04 <0.04

Ambient Pressure % of reading for a 1% 1 1 1 <0.13 1


Coefficient change in ambient pressure
Sample Flow Rate % O2 over 50 to <0.1 N/A N/A N/A <0.2
Effect 250ml/min (over 50 to 70 liters/hour)
Ambient Temperature /10C 0.2 % O2 0.5% of reading
Coefficient

The performance specification has been written, and verified, in accordance with the international standard IEC 1207-1:1994
Expression of performance of gas analysers.
Servomex companies, agents and representatives are located throughout the world. Your nearest contact is:

Visit www.servomex.com for technical data sheets, application and technology


information for all Servomex analysers.

Servomex has a policy of constant product improvement and therefore reserves


the right to change specifications without notice.

Certificate No.______Q05166
BS EN ISO 9001 LEADERS IN GAS ANALYSIS

Servomex Group Limited, Jarvis Brook, Crowborough, East Sussex, TN6 3DU, England % (44) 1892 652181. Fax: (44) 1892 662253
Servomex B.V., Stephensonstraat 20, 2723 RN Zoetermeer, Netherlands % (31) 79-346 42 42. Fax: (31) 79-342 08 19
Servomex S.A., 8 Rue Proudhon, B.P. 50, 93212 St Denis La Plaine Cedex, France % (33) 1 49 46 22 50. Fax: (33) 1 49 46 22 51
Servomex GmbH, Mnsterstrae 5, 59065 Hamm, Germany % (49) 23 81 68 82 13. Fax: (49) 23 81 68 81 75
Servomex Company, Inc., 90 Kerry Place, Norwood, MA 02062, USA % (1) 781-769-7710. Fax: (1) 781-769-2834
Servomex Asia Pacific Ltd, 3F-1, No.88, Sec. 6, Chung Shan N. Rd., Taipei, Taiwan % (886) 2-2833 8848. Fax: (886) 2-2833 8844
Servomex Asia Pacific Ltd, (Spectris Group China Ltd-Shanghai Representative Office)
Room 1805 Central Plaza,227 North Huang Pi Rd,Shanghai 200003, PR China % (86) 21 6375 8137 Fax: (86) 21 63758139
Global email: info@servomex.com Website: http://www.servomex.com

Servomex Group Limited. 2001. A Spectris.company. All rights reserved. TDS 1800 Rev. 1 09/01 E