Valid as o V4.0
Version
Software
Operating Instructions
2
Table of Contents
Page
2 Safety First 9
2.1 Indications
2.2 Health and safety instructions
3 Product Description 14
3.1 Components
3.2 Hazardous areas and safety equipment
3.3 Functional description
3.4 Accessories
6 Practical Use 26
6.1 Switching on/off
6.2 Standard programs
6.3 Firing programs
8 What If 30
8.1 Error messages and notifications (alarm)
8.2 Technical malfunctions
8.3 Repair
9 Product Specifications 33
9.1 Delivery form
9.2 Technical data
9.3 Acceptable operating conditions
9.4 Acceptable transportation and storage conditions
10 Miscellaneous 35
10.1 Press table
10.2 Firing table
10.3 Menu structure of the EP600 Combi
10.4 1-step firing program
10.5 2-step firing program
3
List of Parts
4
4
55
52
53
54
1
73
2
4
51
6
11
79
14 54
7
20
8 9 10 12
5
59
55
58
62
61
62 64
66
57
60 63
81
65
56 67
73
56
68
80
71
69 15
70
72
103
114
24
50
19
100 112
113
101
111
74 102 110
6
5 75
11
22
54
79
73
7
18
23
21 13
17 16
1
25
140
76
77
78
122 164
163
121
150
162
123 160 161
165
G F
170
7
1. Introduction / Signs and Symbols
1.1 Preface 1.2 Signs and symbols Display 1.3 Notes regarding the
Operating Instructions
Dear Customer, The signs and symbols in these FIRING Mode
Thank you for having purchased Operating Instructions and on These Operating Instructions
the EP600 Combi Press and the furnace facilitate the finding facilitate the correct, safe, and
Ceramic Furnace. This furnace of important points and have economic use of the EP600
with the intelligent press drive the following meanings: PRESSING Mode Combi Press and Ceramic
has been especially developed Furnace. They are divided into
for the Ivoclar Vivadent all- Risks and dangers several, clearly structured
ceramic systems (IPS e.max, This symbol marks chapters. This should enable you
IPS Empress). It is the latest of safety instructions that IPS e.max
to locate specific topics quickly
our technical high-quality must be followed to and easily.
products. Inappropriate use may prevent injury or death.
damage the equipment and be Furthermore, damage to the IPS Empress Esthetic
You must read these
harmful to personnel. Please furnace and/or laboratory may Operating Instructions
observe the relevant safety be avoided.
instructions in Chapter 2. IPS Empress
Important information Layering Technique To inform you about risks,
The EP600 Combi is designed This symbol marks dangers, important information,
according to EN61010-1 and additional information and contraindications, these
thus complies with the relevant for correct and IPS Empress 2
Instructions contain correspon-
EU regulations. economic use of the Layering Technique
ding signs/symbols to mark
EP600 Combi. important paragraphs.
Fuses
A max. 2.1
Ableitstrom / Leakage current
max. 0.75 mA
Timer active,
programs cannot be
Vacuum started
Energy-saving mode
K-Module active
(Communication
module) Program memory card
active
Alternating current
Data transfer from/to
(IEC 417)
Program memory card
active
Burn hazard
(IEC 417)
Risk of crushing
Outlet
RS 232
(SELV, max. 24 VDC/A1)
PC printer connection
8
2. Safety First
This chapter is especially important for individuals 2.1 Indications Further instructions to assure
who work with the EP600 Combi or who have to proper use of the furnace:
carry out maintenance or repair work. This chapter The EP600 Combi has been designed for pressing The instructions, regulations,
must be read and the corresponding instructions IPS e.max and IPS Empress ingots and for firing and notes in these Operating
followed! dental ceramic materials and should be used for Instructions must be observed.
these purposes only. Uses other than the ones The instructions, regulations,
stipulated, e.g. cooking of food, firing of other and notes in the vacuum
materials, etc., are contraindicated. The manufac- pump Operating Instructions
turer does not assume any liability for damage must be observed.
resulting from misuse. The user is solely responsi- The furnace must be operated
ble for any risk resulting from failure to observe under the indicated environ-
these instructions. mental and operating
conditions (see Chapter 9)
The EP600 Combi must be
properly maintained.
2.1.3
Never touch the electronic
components, since a static
Risks and dangers discharge may damage them.
9
2.1.4
Never reach under the furnace
head during operation. There is
a risk of crushing and a burn
Risk of crushing / burn hazard hazard.
2.1.5
Never touch the press drive
during operation. There is a risk
Risk of crushing of crushing.
10
2.1.9
Press cycles must not be
conducted without the firing
Contraindication
plate (50) in place.
2.1.10
The firing plate (50) must be
correctly positioned in the firing
Contraindication
chamber.
2.1.12
The air vents must be kept free
from obstruction and clean at all
Contraindication times. If this is not done, there is
a risk of overheating the
furnace.
11
2.1.14
If the firing tray (150) is in
place, press cycles must not
Contraindication
be conducted.
12
2.2 Health and safety It is important that the user After maintenance, the The furnace is tested for use
instructions become familiar with the required safety tests (high at altitudes of up to 4000 m
warnings and operating voltage resistance, protective (13,123 ft) above sea level.
conditions to prevent injury conductor, etc.) must be
This furnace has been designed to personnel or damage to carried out. The furnace may only be
according to EN 61010-1 and materials. The manufacturer used indoors.
has been shipped from the is not responsible for damage Ensure that only fuses of the
manufacturer in safe operating resulting from misuse or fai- indicated type and current
condition. To maintain this lure to observe the Operating rating are used. Risks and dangers
condition and to ensure risk-free Instructions. Warranty claims Do not work with
operation, the user must observe cannot be accepted in such If it is assumed that safe liquids near the
the notes and warnings in these cases. operation is no longer furnace. Should a
Operating Instructions. possible, the power must be liquid accidentally enter the
Before switching on the disconnected to avoid furnace, disconnect power and
Place furnace on a fire-proof furnace, make sure that the accidental operation. consult the Customer Service.
table, observe local regulation voltage indicated on the Safe operation is no longer Do not operate the furnace.
(e.g. distance to combustible rating plate complies with possible if:
objects, etc.). the local power supply. the furnace is visibly Risks and dangers
damaged Any disruption of the
Always keep the air vents at The mains socket must be the furnace does not work protective conductor
the rear of the furnace free equipped with a residual the furnace has been either inside or outside
from obstruction. current operated device (FI) stored under unfavourable the furnace or any loosening of
conditions over an the protective conductor may
Do not touch any parts that The power plug may only be extended period of time lead to danger for the user in
become hot during operation inserted into sockets with case of a malfunction. Deliberate
of the furnace. protected contacts. Use only original spare parts. interruptions are not tolerated.
There is a burn hazard. Material developing harmful
Before calibration, main- The temperature range for gases must not be fired.
Clean furnace only with a dry tenance, repair, or change of faultless operation is +5 C
or slightly damp cloth. Do parts, the power must be to + 40 C (41 F to 104 F).
not use any solvents. disconnected if the furnace
Disconnect power before has to be opened. If the furnace has been stored
cleaning. at very low temperature or
If calibration, maintenance, high atmospheric humidity,
Use original packaging for or repair has to be carried the unit must be dried or left
transportation purposes. out with the power connected to adjust to the room
and furnace open, only temperature for approx.
qualified personnel, who are 1 hour prior to connecting
familiar with the risks and power.
dangers, may perform the
procedures.
13
3. Product Description
The electronic controls and the Description of the safety equipment of the
mechanical components for the furnace:
opening mechanism are located
in the furnace base. The heater Safety equipment Protective effect
is located in the muffle in the
furnace head, where it is Protective conductor Protection from electrical shock
embedded in the stone lining Fuses Protection from electrical shock and damage to
segments. Operation and control the furnace
of the furnace is done with
state-of-the-art electronic Interlock switch Protection from electrical shock
components.
3.4 Accessories
14
4. Installation and Initial Start-Up
4.1 Unpacking and Packing and shipping of 4.2 Selecting the location 4.3 Assembly and initial
checking the contents individual components start-up
The packaging of the Place the furnace on a flat table
EP600 Combi permits simple using the rubber feet. Make sure
The new packaging provides the Make sure the voltage indicated
and safe shipping of individual that the furnace is not placed in
following advantages: on the rating plate complies
components. Simply use the two the immediate vicinity of heaters
Reusable packaging with the local power supply. If
corresponding inserts. Fold the or other sources of heat.
Closing mechanism with inte- this is not the case, the furnace
side flaps (2) and combine the Furthermore, protect the furnace
grated transportation grips must not be connected.
two packaging parts by means from direct sunlight. Make sure
Ideal protection by Styrofoam of the transportation flaps. that air may properly circulate
inserts
between the wall and the
Easy handling / optimum The packaging may be discarded furnace.
unpacking with the regular household
The packaging may be used refuse. Also ensure that there is enough
in several ways (modules)
space between the furnace and
the user, as the furnace releases
Check the delivery for complete-
heat during the opening of the
ness (see delivery form in
furnace head.
Chapter 9) and transportation
damage. If parts are damaged
The furnace should neither be
or missing, contact your local
placed nor operated in areas
Ivoclar Vivadent Service Center.
where there is an explosion
hazard.
Remove the furnace components
from their packaging and place
the unit on a suitable table.
Please observe the instructions
on the outer packaging.
Important information
We recommend
keeping the original
packaging for future
transportation purposes. The
unit has been subjected to
extensive checks by the manu-
facturer. It may therefore exhibit
slight signs of testing (hairline
cracks, slight discoloration).
15
The furnace components are
assembled as follows:
1
Step 1:
Assembling the tray and the
25
firing plate
Mount the supports for the
sagger tray (25)
The sagger tray (1) is slipped
onto the screws (2). If
desired, the screws may be
1 2
tightened using a screwdriver.
Remove the firing plate (50)
from the protective paper
and place it in the stone
lining.
Clean the sealing surface
(19).
Step 2:
Assembling the furnace head
The furnace head is best 75
assembled with the rear of the 76 77
furnace base placed as shown in
the corresponding figure. Lift
the furnace head with both
hands and slip the plug-on
console (77) onto the
corresponding pegs (75) in the
furnace base until the leaf spring
(76) audibly snaps into place.
Step 3:
Mounting the connections
Connect the cables of the
furnace head with the
furnace base.
Insert the heater plug (66)
into the heater socket (17)
64
and secure it with the
retention screw (67) with a 18
right-hand turn. 78
Insert the thermocouple plug 23 65
(64) into the thermocouple 66
socket (18). 16
Make sure the polarity of the 17
plug is correct (+ signs
together).
67
Insert the press drive plug
(65) into the press drive
sockets (16) until it audibly
snaps into place.
Mount the grounding band
(78) of the furnace head
onto the sip-on tongue of
the furnace base (23).
16
Step 4:
54
Mounting the furnace head
connections cover
Once the connections between
the furnace head and the
furnace base are established,
assemble the furnace head
connections cover (54) and
secure it with the corresponding
screws (51).
Step 5:
Establishing additional
connections
79
Make sure the voltage indicated
on the rating plate of the
apparatus complies with the
local power supply. If everything
is in order, plug the power cord
(21) into the power socket (6). 6
Vacuum pump
Insert the plug of the vacuum 21
pump into the socket for the 14
7
vacuum pump (7) and slide the
vacuum hose onto the vacuum
hose connection (12).
12
We recommend using 12
the Ivoclar Vivadent
VP3 Vacuum Pump
with this furnace.
If another pump is used, the
maximum power consumption
and the maximum final vacuum
of the corresponding pump have
to be observed.
Step 6:
Setting-up the cooling grid for
investment rings
The cooling grid for investment
rings is supplied fully assembled.
Please position the grid in a
suitable place.
17
Step 7: same way as the ones stored in
164
Connecting the card reader the furnace. They may be
Connect the adapter (165) with accessed, edited, and run.
the RS232 PC/Printer interface Insert the program memory card
(14) and insert the plug (161) or into the card reader with the
the card reader (Type 'ECO 5000 microchip facing downwards.
KB1' from ORGA) into the The storage system is now ready
adapter. for operation. 163
Initial start-up Once the preheating phase has procedures. As a result, the Remove the furnace head
Connect the power cord (21) been completed, the unit is firing chamber may exhibit slight connections cover (54)
with the power supply. Put the ready for operation. signs of testing. These signs
Removing the secured furnace
On/Off switch (5) at the rear of The 'furnace selection' menu is show that the unit has been
head
the furnace on position "I". displayed to select whether the tested. These checks are
Before the furnace head is
The language selection menu furnace is used as a ceramic or important for quality assurance.
removed, the heater plug,
appears on the display. Select press furnace. After that, the
Disassembling the furnace head thermocouple plug, press
the language by means of the main menu appears on the
Note: Before the furnace head drive plug, and grounding
F4 key and confirm by pressing display.
connections cover is removed, band should be disconnected
ENTER. Complete the selection
Important information the furnace should be switched from the furnace base.
procedure using the function
During the test of the off and the power cord (21) Release leaf spring (76) and
key F1. After selecting the
press drive, noise from unplugged. lift off furnace head using
language, the display shows the
the press drive is both hands.
EP600 logo. After that, the unit
audible (approx. 2 seconds). Removing the furnace head
conducts a self-test to check the
During the self-test, this sound is connections cover
basic functions. The course of
normal. If this is not the case, a Remove screws (51) of the
the self-test is indicated by
technicalmalfunctionhas occurred. furnace head connections
means of a status bar. Once the
Please refer to Chapter 8. cover.
self-test has been completed
successfully, the preheating The power plug may
phase will start. This phase only be inserted into a
prepares the unit for optimum grounded socket. The
temperature control. The power cord must not
self-test and preheating phase come into contact with the hot
together usually take 30 min. furnace head and should be
protected accordingly.
To reduce the moisture
The EP600 Combi is
in the press chamber
equipped with a
(stone lining) the
special electronic
vacuum pump is
device that can bridge
switched on during the pre-
a power outage of approx.
heating phase, thus removing
10 seconds.
the moist air.
The EP600 Combi
As of Software Version V3.0 the Press and Ceramic
vacuum pump is kept on during Furnace undergoes
the complete pre-heating phase. extensive checks prior
to delivery. In the process, the
unit is subjected to special test
18
5. Operation and Standard Settings
5.1 Starting the furnace 5.2.1 Main menu "Pressing" 5.2.3 Menu "Program"
This menu appears after selection of the
menu item "Press furnace" in the main
menu.
Page 1
5.1.1 Start-up
After the furnace has been switched on, the In this menu, a graphical representation of
unit undergoes a self-test, and if necessary, a the currently selected program is displayed.
preheating cycle. The furnace selection The temperature curve indicates the set
menu is shown on the display. temperature values. The progress of the
Page 2
running program is shown by a bold line.
To reduce the moisture in the press The remaining time until the start of the
chamber (stone lining) the vacuum actual press process is also indicated. As
pump is switched on during the soon as the press process has commenced,
pre-heating phase, thus removing the time passed since the beginning of the
the moist air. press process is indicated.
As of Software Version V3.0 the vacuum Furthermore, the display indicates the
pump is kept on during the complete pre- distance travelled by the press plunger in the
heating phase. selected unit of measure since the beginning
of the press process, as well as the quality of
5.1.2 Main menu the vacuum in percent. If no program is in
In this menu, you may select whether the progress, F2 and F3 can be used to change
EP600 Combi is used as a press furnace or a between the various programs.
ceramic furnace. The 'Combi Furnace' Page 3 With F4, the parameters of the currently
software is required for the latter option. selected program can be viewed (see menu
Additionally, the menu item "Miscellaneous" "Program parameters").
can be accessed from here. This menu
automatically appears, once the preheating
cycle has been completed. 5.2.4 Menu "Program parameters"
Page 4
Important information
For the layering technique, we Higher values (stop speed of e.g.
recommend a stop speed of 300 mm/min) results in the press cycle
300 mm/min, for the staining being stopped sooner.
technique 150 mm/min. Lower values (stop speed of e.g.
100 mm/min) results in the press cycle
being stopped later. Consequently, the
press cycle is prolonged.
Page 2
20
Internal programs Text input
With this menu item, you may access
50 individual firing programs that are
stored in the furnace.
21
Important note SO: Quick opening 5.3.3 Main menu "Program manager"
Vacuum: If this option is switched off, the furnace This program permits copying of the firing
If the value 0 is entered for V1 and V2, head opens within one minute after the end programs to various storage media (EP600
no vacuum will be created. of the program. If this option is switched on, program memory, or program memory card).
the furnace head opens within 15 seconds
Long-term cooling after the end of the program.
Values between 50 and 1200 C (122 and
2192 F) may be entered. If 0 is entered, no NP: Overnight program
long-term cooling will be conducted. This option corresponds to the one offered
by the Programat P100 and Programat X1
Special programs furnaces. If this option is switched off, the
If "No" has been selected in the "Special furnace head opens after the end of a firing
program" option, Page 2 of the program program. If this option is switched on, the
parameters cannot be accessed. procedure is as follows:
If "Yes" is selected (press F4, "yes" appears, The firing program is completed, the
confirm with ENTER), Page 2 is opened to heater is switched off.
set the special values. The furnace head opens without the
buzzer sounding.
Once the furnace temperature drops Example:
Program parameters, Page 2 below 150 C (302 F), the furnace head Copy "Program" from "Internal
closes. The heater remains off. programs" No. 23 to "Program" on
"Memory card" No. 24
Program course Duplicating programs: If the source and
Programs in progress can be stopped by target media are identical, the program on
pressing STOP. An interruption of the this medium is duplicated.
programs is not possible.
Since copying always results in an existing
Empty programs program being overwritten (possibly an
Upon delivery, all the internal firing program empty program), the following inquiry will
spaces are filled with 'empty' programs. An be shown:
empty firing program does not have a name.
Empty programs are treated like any other
program. They can be selected, run, or
edited:
Use F2 and F3 to change from one Holding temperature 700 C
parameter to the next or the previous one.
Temperature increase 30 C/min.
F1 is used to change to the firing curve
display. Once the firing curve display is Closing time 18 seconds
visible, the selected program is started by Stand-by temperature 403 C
pressing START. Holding time 1 minute
Vacuum on / Vacuum off both 0
Additional information regarding the special (i.e. no vacuum)
programs Long-term cooling none
HV: Share of holding time with vacuum Special program no This inquiry should prevent the programs
With the parameter HV, the user selects if from being overwritten accidentally.
and when the vacuum is switched off during
the holding time H. Important information:
If V2 < T, the vacuum is switched off Only firing programs can be copied.
before the holding temperature is Press programs cannot be copied.
reached. In such cases, the parameter
HV is irrelevant.
If V2 >= T, the parameter HV is taken
into consideration.
If HV = 0, the vacuum is not switched
off during the holding time.
If HV > 0, the vacuum is switched off
after period HV has passed since the
beginning of the holding time, but at the
end of the holding time H at the latest.
22
5.4 Introduction to the menu Parameters for the card reader Page 2
"Miscellaneous" Baud rate = 9600
Data bits = 8
This menu is called up via the main menu. Parity = even
It permits the selection of a number of Stop bits = 1
lower-level programs with which the furnace Output device = ECO5000
can be configured, calibrated, tested, etc. Code page = CP 437
23
Direct >Press/Firing programs are 5.4.9 Menu "Protocols" 5.4.10 Menu "Timer"
protocolled (Press/ Firing program After selecting this menu the table with the In this menu, the timer may be programmed.
protocol table) and are shown or printed calibration data is shown. With this feature, the furnace may be
immediately on the serial interface heated up to stand-by temperature or
RS232. Please note that therefore, a switched off at any given time.
corresponding output device (printer or In the interval between the switch-on and
PC) has to be selected in the menu the switch-off times, the unit is kept at
'Interface'. stand-by temperature.
Pager 1 To change the setting, select the
Yes/No corresponding switch-on/switch/off time
Laboratory name with the cursor (F2/F3) and use the numeric
Name which will appear in the header keypad for selecting the desired time.
or footer of the various protocols. The timer can also be switched on or off
with the key timer active-yes/no.
The times are typically typed in via the
5.4.6 Menu "Maintenance" numeric keypad. F3 is used to change the
This menu provides the commands necessary Key F4 can be used to switch to the press selected menu item. ESC is used to undo
during the maintenance of the furnace. program protocol table (keys to be pressed: changes. ENTER is used to save the changes.
Lower press plunger manually F4, ENTER). The mains switch must not be switched off
Lift press plunger manually during this procedure.
5.4.7 Menu "Enter new program" When the timer is active, i.e. when
The "Main menu pressing" also provides the heater is switched off, the
three individual programs (Program No. 57) symbol is shown on the display. If
in addition to the standard programs. If a program has not been completely
these programs are not shown in the menu, finished at the time when the furnace is
they are still concealed. They can be inserted scheduled to turn off, the program will
using the following function: nonetheless be completed before the heater
To insert one of the programs (program No. is deactivated. While the timer is active,
57), select this menu item (F4), type in the programs can only be started once the time
correct code, and confirm with ENTER. settings have been adjusted accordingly.
24
5.5 Operating the menu / 5.5.2 Numeric keypad 5.6 Help function
key functions The numeric keypad is used to enter
numbers. This keypad also contains the ESC Pressing HELP will display a help text for the
and ENTER keys. current menu.
5.5.1 Navigation
1
1 Information line
This line always shows the most current
temperature and operating mode. If the help text consists of several pages, the
Additionally, useful program information is 'Up' (F2) and 'Down' (F3) symbols are
ESC: ENTER: Used to confirm
displayed. Discarding values entered numbers, for shown. They indicate that these keys may be
example, or the change used to navigate between the individual
of the temperature mode
2 Firing curve and dialog field pages. F1 is used to return to the last active
from C to F.
In this field, the dialogues and the various menu.
information are displayed. The different 5.5.3 Text input
parameters may be changed here.
5.7 Protocolling /
3 Function line Output of protocols
Change of level
1 level higher
From the press/firing program protocol
Selector up charts the protocols saved there can be
Selection within the same level output. For that reason use key F2 to set the
Selector down cursor on position 'Print'. Subsequently,
Selection within the same level enter the position number of the protocol to
be printed with the numbers block and
Change of level
confirm with the ENTER key. Using key F4
1 level lower
the selected protocol can be sent via the
Selection key serial interface RS232 to the printer or PC
Selection from various possibilities (Prograsoft).
The input of text is enabled by a virtual
The selection must always be confirmed
keyboard shown on the display. F2 and F3 Example press program protocol table:
with ENTER.
are used to select the characters. F4
transfers the selected, highlighted character
The four function keys below the display are
to the text field. F1 can be used to delete a
used to navigate through the various menus.
character from the text field. ENTER is used
The functions of the keys are represented by
to leave the text field and confirm the text.
symbols. These symbols are shown for those
If the user wants to quit this menu without
keys, which have a certain function at the
confirming the text, ESC is used.
current stage in the program. For example,
the symbol for the 'page down' key is only
shown if the contents of the current screen
5.5.4 Numbering of the various menus
take up more than one page.
To facilitate support, each menu is given a
number, which will become visible as soon
as the help function of the menu in question
is called up.
Example firing program protocol table:
25
6. Practical Use
6.1 Switching on/off 6.2 Standard press programs 6.3 Firing programs
Put the I/O switch at the rear of the furnace The firing programs are selected in the main
on position "I". The furnace is now switched menu "Firing".
on and the EP600 logo appears on the
display. Subsequently, the unit conducts a Without program memory card
self-test to check the basic functions. The Page 1
course of the self-test is shown by means of
a status bar. If the self-test does not reveal
any errors or malfunctions, the preheating
phase is started. The preheating phase
prepares the furnace for optimum
temperature control. The self-test and
preheating phase together take 30 minutes.
After the preheating cycle has been
completed, the unit is ready for operation. 1 Select the desired program and type of
The main menu appears. investment ring.
Important information
Please use the separate cooling
grid and not the sagger tray of the
furnace.
Important information
For further information about the
firing programs (one-step, two-step
programs), please refer to Chapter
10.4 and 10.5 on page 3738.
26
7. Maintenance, Cleaning, Diagnosis
This chapter describes the 7.1 Monitoring and This apparatus has These spare parts are not
maintenance and cleaning maintenance been developed for covered by the warranty.
procedures that may be typical use in the
performed by the users. All dental laboratory. Please also observe the shorter
The time for these maintenance
other tasks must be performed service and maintenance
procedures depends on the
by qualified service personnel at If the product is used in a pro- intervals.
frequency of use and the
a certified Ivoclar Vivadent duction facility, for industrial
working habits of the users. For
Service Center. applications, or in continuous
that reason, the recommended
firing operation, premature
times are only approximates.
ageing of certain spare parts
have to be expected.
Turn off the furnace
and disconnect power These spare parts are e.g.:
before maintenance Heating muffle
and cleaning, since Insulation material
there is a risk of Lamps
electrical shock.
Check all plug-in connections for correct fit Various external connections weekly
Check if the furnace head opens smoothly and without Furnace head opening mechanism monthly
excessive noise
Check if the thermocouple is straight and in the right place Thermocouple weekly
Check the stone lining inserts for large cracks or damage. Stone lining inserts weekly
If the stone linings are worn down, they must be replaced
by a certified Ivoclar Vivadent Service Center. Slight
hairline cracks on the stone lining surface do not have a
negative influence.
Check if the sealing rims of the furnace head and furnace Sealing rims of the furnace head weekly
base are clean and undamaged and furnace base
Check the keypad for visible damage. If the keypad is Keypad weekly
damaged, it must be replaced by a certified
Ivoclar Vivadent Service Center.
Check furnace temperature (calibration). Press chamber Every 34 months, or after
approx. 50 press cycles.
7.2 Cleaning
The furnace may only be cleaned when
it is cool, since there is a burn hazard.
Do not use any cleaning solutions.
27
7.3 Menu "Access alarm table" 7.5 Temperature calibration 4. Remove the firing plate (50) from the
furnace with the firing tongs and place it
In this menu, up to 20 of the last alarm The accuracy of the thermocouple changes on the sagger tray.
messages are listed in a table. In case of during use. In order for it to function
inquiries, please provide the service optimally, it needs to be recalibrated on a
technician with this information. regular basis. The automatic furnace
calibration set (Automatic Temperature
The enclosed error list contains Checking Set 1) has been especially
further information. developed for this purpose. We recommend
that you check your furnace with the
Automatic Temperature Checking Set 1 after 50
approximately 50 press cycles or after 3 to 4
months of use. The corresponding message
74
7.4 Menu "Diagnostic programs" will be displayed. Calibration should be
conducted while the furnace is at operating
Vacuum and system test temperature (stand-by temperature). It takes
approx. 2 hours. For that reason, we 5. Use tongs to insert the completed test
The program checks the final value (in
recommend that you calibrate the furnace assembly into the recesses for the ATK 1
mbar) that the vacuum reaches after a
overnight in order to save time. (124).
defined period of time and how much
time passes until 50 mbar are reached.
Important information
Muffle test The automatic request for
By measuring the maximum power calibration after 50 press cycles is
consumption of the heating muffle, the only displayed in the "Press
'age' of the heating muffle is determined furnace" mode. If the furnace is in the
and presented by a bar diagram. If the "Ceramic furnace" mode, the message will
muffle has a high power consumption not appear after 50 firing cycles. However,
(good, new), the bar is solid. At a certain we recommend conducting a calibration
stage, the user is informed that the after approx. 3-4 months if the furnace is in
heating muffle has to be replaced shortly. frequent use.
Keypad test
6. Place the tongs at the center of the test
If the user conducts this test, all the keys
123 assembly and press lightly until it
are represented on the display by a
perceptibly clicks into place.
matching symbol. Pressing the
corresponding key results in the symbol
of the display being inverted to change
to non-inverted (depending on the actual
condition.
123 121 122
F1 is not represented, since this key
is used to quit the menu.
28
7.6 Changing the press plunger 6. Reconnect power and switch on the
furnace with the I/O switch.
In order to facilitate the changing of the
press plunger, we recommend lowering the Never reach into the press drive
press plunger manually by using the during operation. There is a risk of
corresponding option in the menu crushing.
"Maintenance".
55
150
50
29
8. What If
This chapter will help you to recognize 8.1.1 Technical errors 8.1.2 Operating errors
malfunctions and take the appropriate In case of an operating error (incorrect data
measures or, if possible, to perform minor entry, etc.) an operating error message is
repairs. displayed, which has to be acknowledged
with F1.
8.1 Error messages and
notifications (alarm) 8.1.3 Notifications
In special cases, notifications are displayed,
which will provide you with important
Alarm notifications are directly displayed and
information. The notifications also have to
can be classified as follows:
be acknowledged with F1.
8.1.1 Technical errors
(The furnace has noticed a
technical defect.)
In case of a technical malfunction, a
corresponding message is displayed and an
8.1.2 Operating errors
alarm signal sounds. The alarm signal can be
(The user has tried to make an
switched off by pressing F2.
incorrect entry.)
As long as the malfunction is active, the
8.1.3 Notifications "Minimize" symbol (F3) is indicated,
Helpful information. otherwise the "Acknowledge" symbol (F1) is
shown.
Depending on which symbol is shown, the
error message has to be minimized to a
symbol by pressing F3 (see "Minimized error
messages") or acknowledged with F1 after it
has been read.
30
8.2 Technical malfunctions
Microswitch of the furnace head Is the opening for the operating pin Remove dirt (foreign objects) from the opening.
opening mechanism does not engage (item 15 in parts list) obstructed?
The calibration sample assembly Is the recess obstructed by a foreign object? Clean the recess with a vacuum cleaner. The
cannot be properly inserted EP600 should be cold for that purpose.
Loud noises in the press drive Is the press plunger (56) dirty or damaged? Clean press plunger or replace it, if necessary
Are the motion rods (81) of the press Clean motion rods with a dry cloth.
mechanism dirty? Do not apply grease.
Display incomplete Run the display test program and contact your
local Ivoclar Vivadent Service Center, if necessary.
Text on the display is difficult to read Is the contrast set correctly? Press "Contrast" key (101) until the contrast
is ideal.
Display is not illuminated Is the fuse (10) for the electronic Check fuse and replace it, if necessary.
controls OK?
Buzzer does not sound Is the buzzer switched off (tune 0)? Select tune (19)
Furnace head does not open Is the fuse (10) OK? Check fuse and replace it, if necessary.
Was the furnace head opened manually? Use only the corresponding key to open the
furnace head. Switch furnace off and on again.
Has the vacuum been released? Is the program still running? Wait until the
program is completed. Switch furnace off and on
again. If it still does not work, contact your local
Ivoclar Vivadent Service Center.
Press plunger fractured in the area Was the terminal screw of the press Replace defective press plunger.
of the terminal screw plunger (57) screwed too tightly?
Press plunger slips out of its holder Is the press plunger correctly secured? Tighten the terminal screw of the press
plunger (57).
Press plunger is too long Was the press plunger of the EP500 with Use the red press plunger designed for the
a length of 200 mm accidentally used? EP600 Combi (175 mm).
Vacuum pump does not start working Is the fuse for the vacuum pump (8) Check fuse and replace it, if necessary.
defective?
Was the max. power consumption Use only the vacuum pump recommended
(2.2 A) surpassed? by Ivoclar Vivadent.
Is the vacuum pump plug correctly Correctly connect the vacuum pump plug with
inserted? the furnace base.
31
Description Double-Check Action
Final vacuum is not reached Is the vacuum hose OK? Check vacuum hose (13) and hose connection.
Was the absolute value for the vacuum set Set a lower absolute value in the menu
incorrectly or not in accordance with the "Configuration" (Chapter 5.4.5)
pump capacity?
Is the press plunger OK, and in place? Make sure that the press plunger is mounted
and not fractured.
Wrong or illogical temperature Is the thermocouple (71) buckled or Contact your local Ivoclar Vivadent Service Center.
indication fractured?
Is the plug for the thermocouple correctly Insert the plug correctly.
inserted?
Is the plug for the thermocouple defective? Contact your local Ivoclar Vivadent Service Center.
Cracks in the heating muffle Are the cracks small and neglectable Small cracks in the muffle are normal and do not
(hairline cracks)? negatively influence the performance of the
furnace.
Are the cracks very large or are parts Contact your local Ivoclar Vivadent Service Center.
of the muffle broken off?
32
9. Product Specification
33
9.3 Acceptable operating 9.4 Acceptable
conditions transportation and
storage conditions
Acceptable temperature
range: Acceptable temperature
+5 C to +40 C range
(+41 F to +104 F) -20 C to +50 C
(-4 F to +122 F)
Acceptable humidity range
Maximum relative humidity Acceptable humidity range
80 % for temperatures up to Max. relative humidity 80 %
31 C (87.8 F), gradually
decreasing to 50 % relative Acceptable ambient
humidity at 40 C (104 F). pressure
Condensation excluded. 500 mbar 1060 mbar
Important
information
If condensation water
or ice is formed after
unpacking, we
recommend allowing
the furnace to dry at
room temperature for
at least 24 hours.
Do not connect
power cord. Even if
the furnace seems to
be dry on the outside,
humidity may still be
present inside.
34
10. Miscellaneous
35
36
10.2 Firing table
4)
4) These values are related to the possible set values. The actually achieved values however may differ and depend on various factors (e.g. starting temperature,
heating of the ring chamber, vacuum yes/no, condition of the heating muffle, local power supply, type of object holder, object size).
10.3 Menu structure of the EP600 Combi
Start
Preheating
Main menu
"Furnace type"
F4, (F1)
F1 or F2, (F1)
Firing program
groups
F1...F3, (F1)
F4, (F1)
Program-Manager
List of firing
programs
F4, (F1)
F2...F3 F2...F3
Press program Firing program
Parameters Parameters
Description:
37
10.4 1-step firing program
38
10.5 2-step firing program
39
Ivoclar Vivadent worldwide
Ivoclar Vivadent AG Ivoclar Vivadent Marketing Ltd. Ivoclar Vivadent s.r.l. Ivoclar Vivadent AB
Bendererstrasse 2 Rm 603 Kuen Yang Via dellIndustria 16 Dalvgen 14
FL-9494 Schaan International Business Plaza I-39025 Naturno (BZ) S-169 56 Solna
Liechtenstein No. 798 Zhao Jia Bang Road Italy Sweden
Tel. +423 235 35 35 Shanghai 200030 Tel. +39 0473 67 01 11 Tel. +46 8 514 93 943
Fax +423 235 33 60 China Fax +39 0473 66 77 80 Fax +46 8 514 93 940
www.ivoclarvivadent.com Tel. +86 21 5456 0776 www.ivoclarvivadent.it www.ivoclarvivadent.se
Fax. +86 21 6445 1561
Ivoclar Vivadent Pty. Ltd. www.ivoclarvivadent.com Ivoclar Vivadent S.A. de C.V. Ivoclar Vivadent UK Limited
1 5 Overseas Drive Av. Mazatln No. 61, Piso 2 Ground Floor Compass Building
P.O. Box 367 Ivoclar Vivadent Marketing Ltd. Col. Condesa Feldspar Close
Noble Park, Vic. 3174 Calle 134 No. 13-83, Of. 520 06170 Mxico, D.F. Warrens Business Park
Australia Bogot Mexico Enderby
Tel. +61 3 979 595 99 Colombia Tel. +52 (55) 5062-1000 Leicester LE19 4SE
Fax +61 3 979 596 45 Tel. +57 1 627 33 99 Fax +52 (55) 5553 1426 United Kingdom
www.ivoclarvivadent.com.au Fax +57 1 633 16 63 www.ivoclarvivadent.com.mx Tel. +44 116 284 78 80
www.ivoclarvivadent.com Fax +44 116 284 78 81
Ivoclar Vivadent GmbH Ivoclar Vivadent Ltd www.ivoclarvivadent.co.uk
Bremschlstr. 16 Ivoclar Vivadent SAS 12 Omega St, Albany
Postfach 223 B.P. 118 PO Box 5243 Wellesley St Ivoclar Vivadent, Inc.
A-6706 Brs F-74410 Saint-Jorioz Auckland, New Zealand 175 Pineview Drive
Austria France Tel. +64 9 914 9999 Amherst, N.Y. 14228
Tel. +43 5552 624 49 Tel. +33 450 88 64 00 Fax +64 9 630 61 48 USA
Fax +43 5552 675 15 Fax +33 450 68 91 52 www.ivoclarvivadent.co.nz Tel. +1 800 533 6825
www.ivoclarvivadent.com www.ivoclarvivadent.fr Fax +1 716 691 2285
Ivoclar Vivadent Polska Sp. z.o.o. www.ivoclarvivadent.us.com
Ivoclar Vivadent Ltda. Ivoclar Vivadent GmbH ul. Jana Pawla II 78
Rua Maestro Joo Gomes de Dr. Adolf-Schneider-Str. 2 PL-01-501 Warszawa
Arajo 50; Salas 92/94 D-73479 Ellwangen, Jagst Poland
Sao Paulo, CEP 02332-020 Germany Tel. +48 22 635 54 96
Brazil Tel. +49 (0) 79 61 / 8 89-0 Fax +48 22 635 54 69
Tel. +55 11 69 59 89 77 Fax +49 (0) 79 61 / 63 26 www.ivoclarvivadent.pl
Fax +55 11 69 71 17 50 www.ivoclarvivadent.de
www.ivoclarvivadent.com Ivoclar Vivadent S.A.
Ivoclar Vivadent Marketing Ltd c/Emilio Muoz, 15
Ivoclar Vivadent Inc. 114, Janki Centre Esquina c/Albarracn
2785 Skymark Avenue, Unit 1 Shah Industrial Estate E-28037 Madrid
Mississauga Veera Desai Road, Spain
Ontario L4W 4Y3 Andheri (West) Tel. + 34 91 375 78 20
Canada Mumbai 400 053 Fax + 34 91 375 78 38
Tel. +1 905 238 57 00 India www.ivoclarvivadent.com
Fax +1 905 238 5711 Tel. +91 (22) 673 0302
www.ivoclarvivadent.us.com Fax. +91 (22) 673 0301
www.ivoclarvivadent.firm.in
Version: 4
Issued: 6/2005
For software V4.0