COMBINED MEETING
WFHSS / AIOS
KSVAS LTD 1
L Kingston. Registered Authorised
Person {Sterilizers}
Member of :-
BSI - CH198 Sterilization
CEN - TC 102 wg 2/3 Steam Sterilizers
ISO - TC 198 wg 3 Moist Heat
Sterilization.
EPL/66 Safety of Measuring, Control and
Laboratory Equipment.
BSI LBI 35- 1
KSVAS LTD 2
SCOPE OF THE PRESENTATION
FOR THE CHANGES WITHIN BS EN 1360
KSVAS LTD 4
REVIEW OF THE CHANGES WITHIN
BS EN 1360
Hollow load B :- single ended open space
ratio of length to diameter of cavity greater
than 1 less or equal to 5 (1 L/D 5) Where
diameter greater than or equal to 5mm(D
5mm or double ended where ratio length /
diameter greater than equal to 2 less than10
(2 L/D 10) or (D 5mm)
Temperature measurement reference
position:- identified by manufacture
condition in useable space of chamber
KSVAS LTD 5
BS EN 1360 Structure
SMALL STEAM STERILIZER
TYPE
REQUIREMEMENTS
B N S
Type Type Type
All wrapped or
non wrapped All non All wrapped or
products wrapped non wrapped
Solid products products
KSVAS LTD 6
BS EN 1360 Structure
SMALL STEAM STERILIZER
TYPES WITHIN S
LOADS IN TYPE TESTS
S1 S2 S3
Type Type
Type
All wrapped or All non All wrapped or
non wrapped wrapped non wrapped
products Solid products products
KSVAS LTD 7
Operational Services
Points to be aware of :
Water Supply Steam Generation.
Designed to function with water
free of contaminants
Cause no harm to sterilizer or the
sterilizer load
KSVAS LTD 8
Operational Services
Points to be aware of :
Cooling Purposes
Vacuum systems
Total hardness value 0,7mmol/l 2,0mmol/l
KSVAS LTD 9
Operational Services
Points to be aware of :
KSVAS LTD 11
BS EN 1360 PROCESS PERFORMANCE DATA
Values Type testing
KSVAS LTD 12
BS EN 1360 PROCESS DATA
limiting values to be recorded
Start Time for Time for Sterilization Holding Sterilization Drying Drying
vacuum pressure start time time end time start end
pulse pulse time time
KSVAS LTD 13
BS 1360 Sterilizers Process Evaluation System
If fitted
1. Compare with validated cycle any change pressure or
temperature time period that change occurs
1. Change beyond programme limits
Shall cause an
indicate
Fault
KSVAS LTD 14
BS 1360 Sterilizers Process Evaluation System
If fitted
1. Compare two independent temperature sensors sterilizer
temperature instrument chamber and recorder ; or
Shall cause an
indicate
Fault
KSVAS LTD 15
BS 1360 Sterilizers Process System
KSVAS LTD 16
BS 1360 Sterilizers Process System
TIME SECONDS E
-1BAR
KSVAS LTD 19
TYPE B cycle
P (bar)
15 min
30-40 min
P3
P2
P1 300sec 600sec
f0 f1 f2 f3
Time in seconds
KSVAS LTD 21
BS 13060 STERILIZER
Leak Test Reports
FAIL PASS
KSVAS LTD 22
BS 1360 Sterilizers Performance
ATTAINMENT STERILIZATION
CONDITIONS
All temperatures measured in useable
space and load
KSVAS LTD 23
BS 1360 Sterilizers Performance
TYPE TEST PROCEDURE
Thermometric measurement
Eight sensors shall be used
Test equipment
requirements One external pressure sensor
KSVAS LTD 24
BS 1360 Sterilizers Testing
Process Cycle
Process Challenge Device & Chemical
Indicator system for Hollow Load A
Key
1 Capsule
2 Sealing
3 Indicator system
4 Connector
5 Open end
6 Tube
KSVAS LTD 25
BS 1360 Sterilizers Steam
Testing
Dimensions in millimetres
KSVAS LTD 26
BS EN 1360 Porous load
Non- Condensable Gases
KSVAS LTD 27
BS 1360 Sterilizers Steam Testing
Steam Service supply main steam
KSVAS LTD 28
BS 1360 Sterilizers Steam Testing
Steam Service supply main steam
Apparatus
for sampling
Steam
Condensate
for chemical
analysis
KSVAS LTD 29
BS 1360 Sterilizers
Definition hollow space A
Not hollow
Objects which are not hollow
Less than
Within standard cavity to diameter 1
less than 1
Hollow A objects
Cavity with length to diameter greater 1 Greater
than 1 and greater than 5 for all objects than 5
KSVAS LTD 30
BS 1360 Sterilizers
Definition hollow space B
Hollow B objects
KSVAS LTD 31
BS EN 1360 Porous load
WORKS / INSTALLATION TEST
Small load textiles cotton sheets Reduced test
450mm x 300mm pack
x Volume of
more than
Small porous
54 litres
load test single Diameter
x
wrapped of at least
35 cm
x
Thermal
Sensors x x
positions
KSVAS LTD 32
BS EN 1360 Porous load
TYPE TEST
x Volume of
more than
Small porous load test
54 litres
single wrapped x
Diameter
of at least
35 cm
Thermal x
Sensors x
positions
KSVAS LTD 33
SMALL STERILIZERS
Moist Heat Sterilization
The future may be a combined
standard applicable to all moist
heat sterilizers
EN 13060
EN 285
In accordance with the Vienna
agreement
KSVAS LTD 34
The Vienna Agreement
An agreement made between CEN and ISO to ensure
harmonisation of standards published by the two
organisations.
KSVAS LTD 36