Você está na página 1de 4

Boditech Med Inc. Document No.

: INS-CA-CN

COMPOSITION OF REAGENTS
i-CHROMA hsCRP-All in one
i-CHROMA hsCRP-All in one consists of Test Device and
ImmunoAssay for Quantitative and Point of Care Detection Buffer. Test Device is individually sealed with a
desiccant in aluminum pouch, and Detection Buffer is packed and
Testing (POCT) Quick Measurement of high delivered separately from Test Device in a Styrofoam box filled
sensitivity C-Reactive Protein (hsCRP) and C- with ice pack.
Reactive Protein (CRP) in Human y Test Device contains a test strip in which murine monoclonal
Serum/Plasma/Whole blood with i-CHROMA antibody against human CRP and of rabbit IgG has been
Reader System. immobilized on the test and on the control line of strip,
respectively.
y Detection Buffer contains fluorescence-labeled anti-CRP
(Mouse monoclonal), fluorescence-labeled anti-rabbit IgG,
BSA as a stabilizer, and Sodium Azide as a preservative in PBS.
INTENDED USE
i-CHROMA hsCRP-All in one along with i-CHROMA Reader is
a fluorescence immunoassay that measures hsCRP and CRP in WARNINGS AND PRECAUTIONS
serum, plasma, and whole blood. The test is used as an aid to
predict future cardiovascular diseases (CVD) as well as to see y IVD For In Vitro Diagnostic Use.
infection and inflammation. y Carefully follow the instructions and procedures described in
this insert. REF Catalog No. i-CHROMA hsCRP-
25/50/100/200
INTRODUCTION
y Dont use Test Device if its lot # does not match with ID Chip
C-Reactive Protein (CRP) is synthesized by the liver .The CRP level # that is inserted onto the instrument.
may rise from a normal level of <5 mg/L to 500 mg/L during the
bodys general, non-specific response to infectious and other acute y i-CHROMA hsCRP-All in one is only operational in i-
inflammatory events. CHROMA Reader. And tests should be applied by
For some time, the measurement of CRP concentration has been used professionally trained staff working in certified laboratories at
as a clinical tool for monitoring autoimmune diseases and infectious some remove from the patient and clinic at which the sample(s)
processes, such as differential Diagnosis of bacterial infection and viral is taken by qualified medical personnel.
infection, therapy monitoring,and rheumatoid arthritis1,2.
It has recently been suggested that along with serum cholesterol, may y LOT Neither inter-change materials from different
be marker of diagnosis in cardiovascular diseases (acute myocardial product lots nor use beyond the expiration date. The use
infarction). And, hsCRP is emerging as the strongest and most of medical device beyond expiration date may affect on
independent predictive risk factor for CVD7,8. test result.
A statement regarding use of high sensitivity C-reactive protein to
assess risk of cardiovascular diseases without inflammatory events: y i-CHROMA hsCRP-All in one Test Device should
lower than 1.0 mg/L, low risk. remain in its original sealed pouch until ready to use. Do
between 1.0 and 3.0 mg/L, average risk. not use the Test Device if the pouch is damaged or the
higher than 3.0 mg/L, high risk. seal is broken. Discard after single use.
For infection and inflammatory events, CRP normal reference:
y CRP Test Device and Reader should be used away from
10.0mg/L.
vibration and magnetic field. During normal usage, i-
PRINCIPLE CHROMA hsCRP-All in one may introduce minute vibration,
which should be regarded normal.
i-CHROMA hsCRP-All in one is based on fluorescence
immunoassay technology9, 10. i-CHROMA hsCRP-All in one uses y Use separate clean pipette tips and sample vials for
a sandwich immunodetection method, such that by mixing different specimens. The pipette tips and sample vials
detection buffer with blood specimen in test vial, the fluorescence- should be used for one specimen only. Discard after
labeled detector anti-hsCRP antibody in buffer binds to hsCRP single use.
antigen in blood specimen. As the sample mixture is loaded onto y Blood specimens, used test devices, pipette tips and
the sample well of the test device and migrates the nitrocellulose sample vials are potentially infectious. Proper laboratory
matrix of test strip by capillary action, the complexes of detector safety techniques, handling and disposal methods should
antibody and hsCRP are captured to anti-hsCRP sandwich pair be followed in accordance with standard procedures and
antibody that has been immobilized on test strip. Thus the more relevant regulations observed by microbiological hazard
hsCRP antigen is in blood specimen, the more complexes are materials.
accumulated on test strip. Signal intensity of fluorescence of
detector antibody reflects amount of hsCRP captured and is y i-CHROMA hsCRP-All in one can be used as diagnosis and
microprocessed from i-CHROMA Reader to show hsCRP assessment for infection and inflammation or high risk of CVD.
concentration in blood specimen. The default result unit of i- The results should be interpreted by the physician along with
CHROMA hsCRP-All in one is displayed as an mg/L from i- clinical findings and other laboratory test results.
CHROMA Reader. The working range and the detection limit of i- y Do not smoke, eat, or drink in areas in which specimens or kit
CHROMA hsCRP-All in one system are 0.5-200 mg/L. and 0.1 reagents are handled.
mg/L, respectively.
Boditech Med Inc. Document No. : INS-CA-CN

Thermal Printer
STROAGE AND STABILITY
i-CHROMA hsCRP Control(L, H)
y Store the detector buffer in a refrigerator at 2 - 8C. The
detection buffer is stable up to 18 months. Lancet (if whole blood)

y Once removed from refrigerator, allow the Detection Buffer for


10 minutes to return to room temperature before testing.
PROCEDURE
y Store i-CHROMA hsCRP-All in one Test Device at 4-
30C in its sealed pouch. i-CHROMA hsCRP-All in one 1. Set a Test Device on a dust-free clean place.
test device is stable for 18 months (while in the sealed
pouch) if stored at 4-30C. 2. Check/insert ID Chip onto the instrument. Make sure that the
Test Device lot # matches with ID Chip #.
y If stored in a refrigerator, allow a minimum of 10 minutes for
the Test Device to reach room temperature while it is in the 3. Take out a tube containing Detection Buffer from refrigerator
sealed pouch. and leave it at room temperature for a couple of minutes.

y Do not remove the device from the pouch until ready to use. 4. Make a puncture on the top of the detector tube by inserting an
The Test Device should be used immediately once opened. empty blood collection capillary

y The storage and shipping of Test Kit should be complied as 5. Draw whole blood, serum, plasma, or control with a sample
indicated in manual. However, it is remotely possible that only collector (10 L).
part of Test Kit is affected by stability problems. 6. If necessary, wipe out the excess blood outside of the capillary
with paper towel or Kimwipes

SAMPLE COLLECTION AND PREPARATION 7. Assemble the capillary and the tube into one.

The test can be performed with either serum or plasma or whole 8. Shake the assembled tube gently 5 times by inversion to take the
blood. blood out of capillary.

y For serum sample, collect the blood in a tube without 9. Remove the cap off the top of tube. Discard two drops of
anticoagulant and allow to be clotted. Remove the serum from reagent onto the paper towel before applying to the cartridge.
the clot as soon as possible to avoid hemolysis. For plasma 10. Apply only two drops onto the sample well of a cartridge.
sample, collect the blood in a tube treated with EDTA.
Anticoagulants other than EDTA for plasma specimen have not 11. Leave it in Room Temperature 3 mins.
been evaluated. If testing cannot be conducted within an hour 12. Insert Test Device onto the holder of i-CHROMA Reader.
after preparation of specimen, the serum/plasma should be
Make sure direction of Test Device and push the device
stored at -20o C until tested. In case of use whole blood, apply it back all the way. Instrument will automatically scan the Test
immediately after specimen was taken. Device.
y The specimen must be at room temperature and be 13. Read the results on the display screen of i-CHROMA Reader
homogeneous before testing. Frozen specimens must be or tear out the print for record.
completely thawed, thoroughly mixed, and brought to room
temperature prior to testing. If specimens are to be shipped,
they should be packed in compliance with regulations.
y It is recommended to avoid using severely hemolyzed Refer to i-CHROMA Reader Operation Manual for the
specimens whenever possible. If a specimen appears to be complete instructions on use of the Test.
severely hemolyzed, another specimen should be obtained and
tested.
RESULT
i-CHROMA Reader calculates hsCRP test results automatically
MATERIALS PROVIDED and displays hsCRP concentration on the screen as form of mg/L.
Boditech Med Inc. i-CHROMA hsCRP-All in one For further information, refer to the Operation Manual for i-
CHROMA Reader.
REF Catalog No. i-CHROMA hsCRP-25/50/100/200
Box contains: Quality Control
Test Devices 25/50/100/200T/box
Quality Control
Detection Buffer 25/50/100/200 Tubes (500 L/tube)
y A quality control test using commercially available controls
Sample collector 25/50/100/200/box should be performed as a part of good testing practice, to
confirm the expected QC results, to confirm the validity of the
ID Chip 1/box assay, and to assure the accuracy of patient results. If you want
Insert 1/box to perform QC of Test Kit, we recommend using Kamiya CRP
control.
MATERIALS REQUIRED BUT NOT PROVIDED
y A quality control test should be performed at regular intervals,
i-CHROMA Reader REF Catalog No. FR-203 and before using a new kit with patient specimens, controls
Boditech Med Inc. Document No. : INS-CA-CN

should be tested to confirm the test procedure, and to verify the 3. Imprecision: For the intra-assay imprecision, 20 replicates were
tests produce the expected QC results. QC specimens should also tested at each control sample. For the inter-assay imprecision,
be run whenever there is any question concerning the validity of tests were conducted on 10 sequential days, with 10 runs per day
results obtained. Upon confirmation of the expected results, the and with 10 replicates at each CRP concentration.
test device is ready to use with patient specimens. Control
standards are not provided with this test kit. For information Imprecision of i-CHROMA hsCRP-All in one
about obtaining the controls, contact Boditech Med Incorporates
Technical Services for assistance. CRP (mg/L) Intra-assay Inter-assay

Mean S.D CV% Mean S.D CV%


Procedure Control
y Each i-CHROMA hsCRP Test Device contains internal   
control that satisfies routine quality control requirements. This
internal control is performed each time a patient sample is   
tested. This control indicates that the test device was inserted
and read properly by i-CHROMA Reader. An invalid result   
from the internal control causes an error message on i-
CHROMA Reader indicating that the test should be repeated. 4. Linearity: The high pool (~200 mg/L) was diluted with the very
low pool (~0.50 mg/L) to the following final percentages; 100%,
75%, 50%, 25%, 10%, 5% and 0%. Sample was assayed in
LIMITATIONS OF THE PROCEDURE triplicate in one analytical run at each CRP level. The coefficient
of linear regression was R=0.997.
y The results of i-CHROMA hsCRP-All in one should be
evaluated with all clinical and laboratory data available. If CRP 5. Comparability: The CRP concentrations of 150 clinical
Test results do not agree with the clinical evaluation, additional specimens were quantified independently with i-CHROMA
tests should be performed. hsCRP-All in one and TBA 200FR (Toshiba Inc.) and BNII
ProSpec (Dade Behring Inc.) automatic analyzer. While the
y The false positive results include cross-reactions with some whole blood was used for i-CHROMA hsCRP-All in one, the
components of serum from individual to antibodies; and non- serums were used for TBA 200FR and BNII ProSpec. The test
specific adhesion of some components in human blood that results were compared and their compatibilities were
have similar epitopes to capture and detector antibodies. In the investigated with linear regression and correlation of coefficient
case of false negative results, the most common factors are: (R). i-CHROMA hsCRP-All in one was comparable well to
non-responsiveness of antigen to the antibodies by that certain other methods (R=0.988 and R=0.989) 10.
unknown components are masking its epitope, such that antigen
cannot be seen by the antibodies; instability of CRP antigen,
resulting in degradation with time and, or temperature, such
that they become no longer recognizable by antibodies; and 50
y=1.05727X+0.26972
degraded other test components. The effectiveness of the test is R=0.989
highly dependent on storage of kits and sample specimens at 40

optimal conditions.
i-CHROMA (mg/l)

30

y Plasma using anticoagulants (e.g. heparin or citrate) other than


EDTA has not been evaluated in i-CHROMA hsCRP-All in 20

one and thus should not be used.


10
y Other factors may interfere with i-CHROMA hsCRP-All in
one and may cause erroneous results. These include technical 0
or procedural errors, as well as additional substances in blood
specimens. 0 10 20 30 40 50
BN II ProSpec (mg/l)

PERFORMANCE CHARACTERISTICS

50
1. Analytical Sensitivity: Analytical sensitivity means the lowest
concentration of CRP that the test system can detect with 40
y=0.9748X-0.14791
R=0.988
CV<20%. Analytical sensitivity of i-CHROMA hsCRP-All in
one was determined by testing 10 times with three lots of
i-CHROMA (mg/l)

30

reagents. Analytical sensitivity of i-CHROMA hsCRP-All in 20

one system was 0.5 mg/L.


10

2. Specificity: Other bio-molecules, such as Hb, CEA, AFP, ALT,


Troponin I, CK-MB, Albumin, and serum amyloid P component 0

were added to test specimen with much higher level than their 0 10 20 30 40 50

physiological level in normal blood. There was no significant TBA 200FR (mg/l)

interference with the CRP measurement, nor was their any


significant assay cross-reactivity with those bio-molecules tested. Tracebility: Traceable to CAP/BCR/IFCC Reference Preparation for CRM
470.
Boditech Med Inc. Document No. : INS-CA-CN

REFERENCES
Manufacture Address:
1. Pepys MB and Hirschfield GM. C-reactive protein: a critical update.
J Clin. Invest 2003; 111:1805-1812. #3-2A,56, Soyanggang-ro, Chuncheon-si, Gang-won-do,

2. Volanakis JE. Human C-reactive protein: expression, structure, and 200-957, Korea
function. Mol Immunol 2001;38:189-197.
T. +82-33-243-1400
3. Koenig W, Sund M, Frohlich M, et al. C-reactive protein, a sensitive
marker of inflammation, predicts future risk of coronary heart
F. +82-33-243-9373
disease in initially healthy middle-aged men. Circulation 1999; www.boditech.co.kr
99:237-242.
4. Rifai N, Ridker PM. Proposed Cardiovascular Risk Assessment
Algorithm Using High-Sensitivity C-reactive protein and Lipid i-CHROMATM is registered trademarks of Boditech Med
Screening. Clin. Chem. 2001; 47:28-30. Incorporated.
5. Rifai N and Ridker PM. High-Sensitivity C-Reactive Protein: A
novel and Promising Marker of Coronary Heart Disease. Clin. Chem.
2001; 47(3): 403-411. Revision No: 16

6. Biasucci LM, Liuzzo G, Grillo RL, et al. Elevated levels of C- Date of last revision: August. 17. 2015
reactive protein at discharge in patients with unstable angina predict
recurrent instability. Circulation 1999; 99:855-860.
7. Taubes G. Does inflammation cut to the heart of the matter? Science
2002; 296:242-245.
8. Ridker PM, Hennekens CH, Buring JE, and Rifai N. C-reactive
protein and other markers of inflammation in the prediction of
cardiovascular disease in women. N Engl J Med 2000:342(12): 836-
843.
9. Brooks DE, Devine DV, Harris PC, et al. RAMP(TM): A rapid,
quantitative whole blood immunochromatographic platform for
point-of-care testing. Clin Chem 1999; 45:1676-1678.
10. Oh SW, Moon JD, Park SY, et al. Evaluation of fluorescence hs-
CRP immunoassay for point-of care testing. Clin Chim Acta 2005;
356:172-177.
11. Claus DR, Osmond AP, Gewurz H. Radioimmunoassay of human
C-reactive protein and levels in normal sera. J. Lab. Clin Med
1976;87:120-128
12. Kindmark CO. The concentration of C-reactive protein in sera from
healthy indivisuals. Scand J Clin Lan Invest 1972;29:407-411

Boditech Med Inc. express and implied warranties (including


implied warranties of merchantability and fitness) are conditional
upon observance of Boditech Med Inc. published directions with
respect to the use of Boditech Med Inc. products.

For Technical Assistance call


Boditech Med Inc.s
iCHROMA Reader Technical Services at
Tel: +82-33-243-1400
E-mail: support@boditech.co.kr

Boditech Med Incorporated


Office Address:
43, Geodudanji 1-gil, Dongnae-myeon,
Chuncheon-si, Gang-won-do, Korea

Você também pode gostar