Escolar Documentos
Profissional Documentos
Cultura Documentos
www.elsevier.com/locate/jelechem
Received 13 August 2001; received in revised form 6 December 2001; accepted 12 December 2001
Abstract
Boron-doped diamond (BDD) electrodes have been examined for the electrochemical detection of six tricyclic antidepressant
drugs (TCAs): imipramine, desipramine, clomipramine, amitriptyline, nortriptyline, and doxepin. Cyclic voltammetry, flow
injection analysis (FIA) and HPLC with electrochemical detection have been used to study the oxidation reactions and to detect
the TCAs. At diamond electrodes, well-defined and highly reproducible voltammograms were obtained for all six drugs with a
signal to background (S/B) ratio about 24 times greater than those at glassy carbon electrodes. Diamond is the first electrode
material to show well-defined voltammograms for nortriptyline due to its wide potential window. In the FIA-mode, at an
operation potential of 0.85 V versus Ag AgCl, diamond exhibited a background current of 7 nA with rapid stabilization (15 min)
conversely to the case of GC, which appeared to stabilize after 1 h, but again increased thereafter. The analytical peaks of HPLC
for the TCAs were well resolved. Linear calibration curves were linear over the ranges from 0.05 to 100 mM. The limits of
detection (S/N=3) were 3 nM for imipramine and desipramine, 0.5 nM for clomipramine, 163 nM amitriptyline, 1080 nM for
nortriptyline and 92 nM for doxepin. The electrodes have shown reproducible results over several days of analysis. This method
has been applied for the determination of imipramine and desipramine in plasma samples. The BDD surface was reproducible
with no adsorption of blood components during plasma analysis. This work shows the promising use of diamond as an
amperometric detector in HPLC, especially for TCA analysis. 2002 Published by Elsevier Science B.V.
Keywords: Tricyclic antidepressant drugs; Boron-doped diamond; HPLC; Plasma sample; Reproducibility
minescence or electrochemical detection [2 8]. For though few reports are available about the use of GC in
therapeutic drug monitoring, where specificity is re- the HPLC-ED for the analysis of TCAs, the stability
quired, chromatographic methods are the most ade- aspects are not mentioned. Surmann and Peter, on the
quate techniques. The advantages of HPLC for the other hand, have reported the influence of pretreatment
analysis of TCAs are its versatility and simplicity of of carbon fiber electrodes on the analysis of bamipine
sample preparation. However, some trialkylamines and and imipramine [13]. Therefore, a stable electrode mate-
related compounds are difficult to detect by use of UV rial with sensitive detection capabilities is a prime re-
absorption, because they do not adsorb very well in the quirement for wider application of electrochemical
UV visible region, since they have low molar absorp- detectors. Recently emerging diamond materials meet
tivities [2]. The chemiluminescence method is promis- all these requirements to a great extent.
ing, but the limit of detection was found to be relatively The use of highly boron-doped diamond (BDD) thin
high (500 nM) [2,46]. Electrochemical detection meth- films as electrode materials for several applications,
ods are very advantageous over other methods in terms including electroanalysis [14 20], energy storage
of simplicity, sensitivity, selectivity and cost. However, devices [21], and electrosynthesis [22] has gained great
these methods have not become as popular as other interest in the past few years due to the impressive
methods due to certain unavoidable problems such as properties of such films. The superiority of diamond
electrode deactivation, with the necessity of frequent thin films for electroanalysis results from their attrac-
pretreatment and other procedures to reactivate the tive properties, such as the wide electrochemical poten-
solid electrodes. Glassy carbon (GC), one of the widely tial window in aqueous solution [23,24], very low
used electrodes for electrochemical detection, due to its voltammetric background current ( 1 order of magni-
relatively wide potential window and low cost, is very tude lower than GC) [25,26] inertness of the surface to
susceptible to contamination and fouling [9 12]. Al- adsorption of reaction products [14], extreme electro-
chemical stability [21], and relative insensitivity to dis- confirmed by Raman spectroscopy. The results showed
solved oxygen [27]. Several research groups including the narrow, intense peak at 1332 cm 1, which is char-
our group have previously reported the uniqueness of acteristic of diamond and reflects a high degree of
diamond for the detection of various environmentally, crystallinity. In addition, a broad peak centered at
biologically and clinically important compounds using approximately 1200 cm 1 was observed, which is usu-
flow injection analysis or liquid chromatography, re- ally attributed to either amorphous diamond or ex-
sulting in sensitive and stable detection [15 20]. During tremely small diamond crystallites [21].
the use of other electrode materials such as metals, GC
and carbon paste, the sensitivity is decreased for the
2.2. Voltammetric in6estigation
detection of the compounds that oxidize at relatively
high potentials due to the interference of the oxygen
Cyclic voltammetric measurements were carried out
evolution reaction, or the requirement of frequent sur-
in a single compartment cell with a saturated calomel
face pretreatment processes due to the strong adsorp-
electrode (SCE) as the reference electrode and a plat-
tion of reactants and products on the electrode surface.
inum wire as the counter electrode. The BDD working
Recently, Rao et al. demonstrated the ascendancy of
electrode along with the conducting Si substrate or GC
diamond electrodes over GC for the detection of sulfa
electrode was pressed against the bottom of the glass
drugs [20]. Therefore, BDD, with its inert surface prop-
cell by a Viton O-ring. The electrical contact for BDD
erties could be a promising electrode for the electro-
was made through the backside of the scratched Si
chemical detection of TCAs using HPLC.
substrate by contacting the brass current collecting
In the present study, we report the detection of six
back plate. Before use as a working electrode, BDD
TCAs, imipramine and its metabolite desipramine,
film was pretreated by ultrasonication in 2-propanol for
clomipramine, amitriptyline and its metabolite nor-
about 10 min followed by rinsing with high purity-wa-
triptyline, and doxepin at neutral pH, by the use of
ter. The purpose of this pretreatment is to remove the
BDD electrodes for the electrooxidation. Cyclic voltam-
organic impurities that may have remained or formed
metry, flow injection analysis and HPLC with an am-
during the deposition of diamond in the CVD chamber.
perometric detector were used for the detection.
The GC electrode (GC-20 plate or GC-20 rod, Tokai
Comparison experiments were performed using GC
Carbon Co., Ltd) was pretreated by polishing with
electrodes.
diamond paste (Fujimi), followed by ultrasonication in
2-propanol before the experiment. The supporting elec-
trolyte was a mixture of 0.05 M KH2PO4 + 0.05 M
2. Experimental
K2HPO4 (KH2PO4/K2HPO4, pH 6.9 90.1). The geo-
metric area of the working electrode was estimated to
2.1. Preparation of highly boron-doped diamond
be 0.09 cm2.
electrode
The highly BDD films were grown by the microwave- 2.3. Flow injection analysis
assisted plasma chemical vapor deposition (CVD) tech-
nique with a commercial microwave plasma reactor The FIA system used in the present study, consisted
(ASTeX) Corp., Woburn, MA). Films were grown on of a micro-LC pump (Bioanalytical System, LC-100),
Si(100) substrates. The details of the preparation have an injector (Rheodyne) with a 20-ml injection loop, an
been described previously [14]. A mixture of acetone amperometric detector (Bioanalytical Systems, LC-4C),
and methanol in the ratio of 9:1 (v/v) was used as the and an XY recorder (Graphtec, WX400). The flow
carbon source. B2O3, the boron source was dissolved in rate set for the pump was 1 ml min 1 and was confi-
the acetone+methanol mixture at a B/C molar ratio of rmed before every experiment by measuring the volume
1:100. The boron concentration in the film prepared of the buffer collected at the outlet for 10 min. The
under these conditions was ca. 1.5 1021 cm 3 as wall-jet type flow cell consisted of the Ag AgCl 1 M
estimated by Notsu et al., based on nuclear reaction LiCl reference electrode and a stainless steel tube as the
measurements [11B(p,a)8Be], carried out by means of counter electrode, which also served as the tube for the
1-meV proton bombardment and subsequent compari- solution outlet. A 0.5-mm-thick silicon rubber gasket
son of the a-spectrum in the 6 8 MeV region with a was used as a spacer in the cell. The geometric area was
BN standard [28]. High purity hydrogen was used as estimated to be 0.64 cm2. The cell volume was esti-
the carrier gas. The bubbling of the acetone+ mated to be 24 ml by assuming a 25% compression of
methanol +B2O3 solution was carried out at 25 C. the gasket. Phosphate buffer (KH2PO4 + K2HPO4, pH
The deposition of the film was carried out at a mi- 6.99 0.1) was used as the mobile phase. The analyte
crowave power of 5 kW. A film thickness of 40 mm solutions were prepared with the same buffer. The
was achieved after 10 h deposition. The film quality was injection volume of the analyte was 20 ml.
120 T.A. I6andini et al. / Journal of Electroanalytical Chemistry 521 (2002) 117126
Hydrodynamic voltammograms were obtained for cyclic voltammetric behavior of the TCAs. At the BDD
each compound prior to amperometric determination electrode, the oxidation of imipramine occurred at
and the detection potential was chosen in the limiting slightly higher potentials ( 0.844 V vs. SCE) than that
current range of the voltammogram. at the GC electrode ( 0.800 V vs. SCE). It is estab-
lished that diamond exhibits higher overpotentials for
2.4. HPLC analysis inner sphere redox reactions in comparison to GC, with
the oxygen evolution and hydrogen evolution reactions
The HPLC system consisted of a reversed-phase C18 being the best examples. At both the electrodes, the
column (Inertsil ODS-3 1504.6 mm ID; particle size, voltammogram exhibited similar features, i.e. irre-
5 mm) connected to the FIA system. The chro- versible oxidation peaks. However, for BDD, the low
matograms were obtained at an applied potential 0.85 background current (0.19 nA at 0.844 V vs. SCE) at the
V versus Ag AgCl 1 M LiCl for the imipramine/de- oxidation potentials, compared to that at GC (1.39 nA
sipramine couple, at 0.93 versus Ag AgCl for at 0.8 V vs. SCE), resulted in a large S/B ratio. Table 1
clomipramine and at 1.3 V versus Ag AgCl for the presents a comparison of data for S/B ratios obtained
amitriptyline/nortriptyline couple and doxepin. The from the cyclic voltammetric data for the oxidation of
temperature was maintained at 26 C. All the experi-
ments were performed with a mixture of acetonitrile
and phosphate buffer 0.05 M (KH2PO4/K2HPO4 pH
6.99 0.1) in the ratio of 375:625 (v/v) as the mobile
phase for the detection of all TCAs, except for
clomipramine (50:50).
2.6. Materials
Imipramine, desipramine, clomipramine, amitri- Fig. 1. Cyclic voltammograms for 10 mM imipramine in 0.1 M
ptyline, nortriptyline, doxepin, KH2PO4, K2HPO4, phosphate buffer (pH 6.9) at: (A) at a BDD electrode (B) a GC
HClO4, NaOH (Wako Chemical Co.) and acetonitrile electrode. Dashed lines show the voltammograms after stirring the
(Nacalai Tesque) were used without further purifica- solution. The sweep rate was 100 mV s 1.
tion. All of these solutions were prepared using ultra-
Table 1
pure (18 MV) water. Comparison of S/B ratio obtained from the cyclic voltammetry data
for the oxidation of TCAs in 0.1 M phosphate buffer (pH 6.9) for
BDD and GC electrode
3. Results and discussion
BDD GC
3.1. Voltammetric studies Ep/V vs. SCE S/B Ep/V vs. SCE S/B
Fig. 1 shows the cyclic voltammograms obtained at a Imipramine 0.844 52.31 0.800 12.59
potential sweep rate of 100 mV s 1 for a deaerated Desipramine 0.833 55.84 0.833 33.00
Clomipramine 0.936 40.43 0.877 11.85
solution of 10 mM imipramine in 0.1 M phosphate
Amitriptyline 1.333 19.17 0.979 10.70
buffer (pH 6.9) together with the corresponding back- Nortriptyline 1.350 10.23 Undetectable
ground voltammogram, at BDD (Fig. 1A) and GC Doxepin 1.300 21.86 0.895 8.09
(Fig. 1B). In fact, deaeration has little effect on the
T.A. I6andini et al. / Journal of Electroanalytical Chemistry 521 (2002) 117126 121
100 mM concentrations of all five TCAs in 0.1 M surface. At GC, strong adsorption on the surface gener-
phosphate buffer at BDD and GC. The results show ally provides high sensitivity for the detection of TCAs,
that the S/B ratios were about 2 4 times higher for but the adsorbed reaction products form insulating
BDD compared to GC. The relatively low background layers on the surface resulting in electrode fouling [20].
current for diamond was attributed to the predominant The shoulder peak observed at higher potentials for
hydrogen termination of the surface as well as the both the electrodes may be due to the oxidation of a
compact nonporous surface [16,20]. Films used in this reaction product. Similar voltammograms were ob-
study are predominantly hydrogen terminated, as they tained for desipramine, a metabolite of imipramine. In
were prepared in hydrogen plasma. Such films showed the case of clomipramine, a higher potential (V0.94
an O/C ratio of 0.03. Earlier, Yagi et al. from our V vs. SCE) was needed to oxidize the compound, since
group reported that XPS results of as-deposited BDD clomipramine contains highly electronegative substi-
showed only a sharp C 1s peak and no discernable O 1s tuted chlorine that stabilizes the compound and makes
peak, which is typical for the H-terminated surface [29]. oxidation more difficult.
However, although the presence of oxygen functional At the BDD electrode the peak currents increased
groups on the GC surface is believed to be partly linearly with the square root of the sweep rate (w 1/2)
responsible for its high background current [9], our within the range of 5300 mV s 1 (r 2 0.9996) with a
previous studies have shown that surface oxygen does zero intercept. The linearity suggested that the current
not play a major role in the case of diamond in is limited by diffusion control in the interfacial area of
determining the background current. An increase of the electrode and the adsorption steps and specific
O/C ratio from about 0.03 to 0.18 changed its back- surface interaction can be neglected.
ground to about 3 5 times, which was still much lower The electrochemical oxidation of imipramine, de-
than that of glassy carbon [29]. sipramine and clomipramine are believed to occur at
Another important observation is the recovery of the the nitrogen atom in cyclohexane involving a two-elec-
voltammetric peak at a BDD electrode. The oxidation tron one-proton mechanism, in analogy to that for
curve of imipramine at BDD could be recovered by methyliminobibenzyl [30,31], as illustrated in Scheme 2.
stirring the solution for 2 min after the second cycle The dimer is formed, which is more easily oxidized than
(Fig. 1A). In the case of GC, the cyclic voltammogram the monomer. The redox-peak pair that appeared at
could not be retraced even after vigorous stirring, indi- 0.037 V after the first cycle may be due to the dimer,
cating partial deactivation of the electrode surface since after the first oxidation, the dimer is formed and
probably due to the adsorption of oxidation products is available near the electrode surface [32]. A simple
(Fig. 1B). The oxidation peak current at GC was about experiment was performed to investigate whether this
4 times larger than that obtained at BDD. This may be peak was due to the adsorption of reaction products on
due to the adsorption of the compound on the electrode the electrode surface. After stirring the solution for 2
3.2. Flow injection analysis Fig. 6. FIA-ED results for 20-ml injections of: (A) 10 nM imipramine
at a BDD electrode and (B) 10 nM imipramine at a GC electrode.
Hydrodynamic voltammograms obtained at the dia- The mobile phase was 0.1 M phosphate buffer pH 6.9. The flow rate
mond electrode for 20-ml injections of 10 mM TCAs in was 1 ml min 1.
124 T.A. I6andini et al. / Journal of Electroanalytical Chemistry 521 (2002) 117126
tions. In the case of GC electrodes, the response for of the electrode surface itself. Furthermore, at the end
100 mM decreased to 72% after 18 injections. For 100 of the experiment, a visible thin film formation on the
nM imipramine, the peak variability was 10%. Am- GC surface was observed, which is probably due to
perometric peaks obtained for 10 nM imipramine the adsorption of contaminants from the buffer. GC
were highly irregular due to the large noise in the is prone to such contaminant deposition [12]. No visi-
background current (Fig. 6B). The BDD electrode ble change was observed on the diamond electrode.
showed a smaller degree of noise in the baseline, with Similar observations were made by Sarada et al. for
a low background current, compared to those at GC GC at 1.28 V and for BDD at 1.52 V. versus
electrodes. Similar results were obtained for other Ag AgCl [35]. The noise in the background current
TCAs. A linear dynamic range is obtained from 0.01 after 10 h of experimentation was about one order of
to 100 mM (r 2 =0.995), with an experimental detec- magnitude higher compared to that for BDD. The
tion limit of 0.01 mM for all TCAs, and 0.1 mM for low noise at BDD shows that very low detection lim-
nortriptyline as an exception. At GC, the peak cur-
its can be obtained. For the clinical applications, dur-
rent was saturated at high concentrations and the ex-
ing therapeutic monitoring, the stability of the
perimental limit of detection was 0.1 mM for all
electrode surface is highly important and BDD sa-
TCAs, which is one order higher than those achieved
tisfies this requirement. Also, the background current
at BDD. Nortriptyline was not detectable at GC.
at the BDD electrode was about 3 5 times less than
3.3. Chromatographic detection that for the GC electrodes. This enables the detection
of TCAs at very low concentrations.
Prior to the chromatographic studies of the TCAs, Since desipramine is a metabolite of imipramine
the stabilities of electrodes were examined and com- and nortriptyline is a metabolite of amitriptyline, the
pared over a period of 10 h, for two types of com- chromatograms were obtained for imipramine and de-
mercial GC electrodes, and a diamond thin film at sipramine simultaneously (Fig. 8A) and similarly, for
the detection potentials, i.e. 800 mV versus Ag AgCl amitriptyline and nortriptyline (Fig. 8B). Chro-
for GC and 850 mV versus Ag AgCl for BDD (Fig. matograms for clomipramine and doxepin were ob-
7). At BDD, the background current stabilized within tained separately (Fig. 8C and D). The mobile phase
20 min after switching to the detection potential and was acetonitrile+ phosphate buffer at a ratio of
the current was very stable during the 10 h of experi- 375:625 for all TCAs except for clomipramine (50:50),
mentation. However, at both types of GC, longer since the Cl atom in its structure makes for high
times (\ 60 min) were required to obtain consider- polarity. The elution times for desipramine and
able stability and there was a continuous change in imipramine were 7.5 and 16.3 min, respectively, and
the current during the 10 h of experiment. The con- for clomipramine it was 10.55 min. Nortriptyline and
tinuous change in the current with time was either amitriptyline eluted at 7.66 and 18.5 min, respectively,
due to the adsorption of impurities or the oxidation and doxepin eluted at 10.39 min.
TCA calibration curves obtained for the chro-
matograms were linear over the range 1000.05 mM.
The limit of detection for imipramine and de-
sipramine was 3 nM (S/N = 3). The lowest limit of
detection that can be achieved is 0.5 nM for
clomipramine. The linear calibration equation and
limits of detection (S/N = 3) are shown in Table 2.
BDD electrodes showed good linearity and reproduci-
bility for the detection of TCAs. These detection lim-
its were 5 times less than those reported previously
with an electrochemical detector [36]. The results were
reproducible from film to film, and with the same
film, reproducible peaks were obtained for the 3
months over which the experiments were performed.
Ultrasonication in 2-propanol for 10 min was suffi-
cient for cleaning the electrode surface when it was
used after a long interval. No mechanical or electro-
Fig. 7. Current vs. time profiles for BDD (at 0.85 V vs. Ag AgCl),
GC20 Tokai and GC GL Tokai (at 0.80 V vs. Ag AgCl). The mobile
chemical pretreatment was required before each set of
phase was acetonitrile +0.5 M phosphate buffer (pH 6.9) = 37.5:62.5. experiments, whereas GC required pretreatments in-
The flow rate was 1 ml min 1. cluding mechanical and electrochemical treatments.
T.A. I6andini et al. / Journal of Electroanalytical Chemistry 521 (2002) 117126 125
Table 2
Calibration linear data for TCAs by using HPLC
Table 3 References
Recoveries from blank plasma spiked with 1.0 nmol of each de-
sipramine and imipramine (n= 3)
[1] J. Wang, M. Bonakdar, C. Morgan, Anal. Chem. 58 (1986)
1024.
Compound Recovery
[2] G.M. Greenway, S.J.L. Dolman, Analyst 124 (1999) 759.
[3] S. Ulrich, J. Materns, J. Chromatogr. B 696 (1997) 217.
Amount Percentage/% SD/nmol CV/%
[4] A. Mason, K.M. Rowan, B. Law, A.C. Moffat, Analyst 109
found/nmol
(1984) 1213.
[5] A. Bakkali, E. Corta, J.I. Ciria, L.A. Berrueta, B. Gallo, Ta-
Imipramine 0.9089 90.89 0.0209 2.29
lanta 49 (1999) 773.
Desipramine 0.9283 92.83 0.0252 2.71
[6] J.L. Paz, A. Townshen, Anal. Commun. 33 (1996) 31.
[7] F. Pommier, A. Sioufi, J. Godbillon, J. Chromatogr. B 703
(1997) 147.
for desipramine and imipramine, respectively. The data [8] R.F. Suckow, T.B. Cooper, J. Pharm. Sci. 70 (1981) 257.
indicated that there was no fouling at the BDD electrode [9] D.T. Fagan, I.-F. Hu, T. Kuwana, Anal. Chem. 57 (1985)
after plasma analysis. The result shows that this method 2759.
[10] I.-F. Hu, D.H. Karweik, T. Kuwana, J. Electroanal. Chem.
can be used for the rapid analysis of imipramine and
188 (1985) 59.
desipramine in blood samples. [11] J. Xu, Q. Chen, G.M. Swain, Anal. Chem. 70 (1998) 3146.
[12] S. Ranganathan, T.-C. Kuo, R.L. McCreery, Anal. Chem. 71
(1999) 3574.
4. Conclusions [13] P. Surmann, B. Peter, Electroanalysis 8 (1996) 692.
[14] T.N. Rao, I. Yagi, T. Miwa, D.A. Tryk, A. Fujishima, Anal.
Chem. 71 (1999) 2506.
Conductive boron-doped diamond electrodes exhib- [15] B.V. Sarada, T.N. Rao, I. Yagi, T. Miwa, D.A. Tryk, A.
ited an excellent performance for the electrochemical Fujishima, Anal. Chem. 72 (2000) 1632.
detection of TCAs. Cyclic voltammetric studies show the [16] M.D. Koppang, M. Witek, J. Blau, G.M. Swain, Anal. Chem.
superiority of diamond over GC in terms of reproduci- 71 (1999) 1188.
bility and sensitivity (S/B ratio). Diamond is the first [17] J. Xu, G.M. Swain, Anal. Chem. 70 (1998) 1502.
[18] S. Jolley, M. Koppang, T. Jackson, G.M. Swain, Anal. Chem.
material to yield well-defined cyclic voltammograms for 69 (1997) 4409.
nortriptyline on the unmodified surface due to its wide [19] M.C. Granger, J. Xu, J.W. Stojek, G.M. Swain, Anal. Chim.
potential window. FIA and HPLC results indicate that Acta 397 (1999) 145.
the TCAs examined in this study are detected ampero- [20] T.N. Rao, B.V. Sarada, D.A. Tryk, A. Fujishima, J. Elec-
metrically with high sensitivity and reproducibility. The troanal. Chem. 491 (2000) 175.
[21] K. Honda, T.N. Rao, D.A. Tryk, A. Fujishima, M. Watan-
limit of detection (S/N =3) ranges from 0.5 nM for
abe, K. Yasui, H. Masuda, J. Electrochem. Soc. 147 (2000)
clomipramine to 1.08 mM for nortriptyline. The back- 659.
ground current stability test for diamond indicated faster [22] F. Okino, H. Shibata, S. Kawasaki, H. Taouhara, T. Momota,
current stabilization after switching to the detection M. Nishitani-Garno, I. Sakaguchi, T. Ando, J. Electrochem.
potential (15 min) with a lower background current Solid-State Lett. 2 (1999) 382.
[23] N. Vinokur, B. Miller, Y. Avyigal, R. Kalish, J. Electrochem.
(7 nA) in comparison to GC electrodes, which exhibited
Soc. 143 (1996) L238.
a larger variation in background current during 10 h of [24] H.B. Martin, A. Argoitia, U. Landau, A.B. Anderson, J. An-
experimentation. The result obtained for the detection of gus, J. Electrochem. Soc. 143 (1996) L133.
imipramine and desipramine in plasma samples showed [25] R. Tenne, C. Levy-Clement, Isr. J. Chem. 38 (1998) 57.
that the use of a BDD electrode with HPLC for the [26] G.M. Swain, J. Electrochem. Soc. 141 (1994) 3382.
detection of TCAs is rapid and reproducible with no [27] T. Yano, D.A. Tryk, K. Hashimoto, A. Fujishima, J. Elec-
trochem. Soc. 145 (1998) 1870.
adsorption of the electroactive component of blood on [28] H. Notsu, I. Yagi, T. Tatsuma, D.A. Tryk, A. Fujishima, J.
the electrode surface. These results demonstrated the Electroanal. 492 (2000) 31.
possibility of the practical utility of diamond in amper- [29] I. Yagi, H. Notsu, T. Kondo, D.A. Tryk, A. Fujishima, J.
ometric detection coupled with FIA and HPLC systems Electroanal. 473 (1999) 173.
especially for TCAs. [30] S.N. Frank, A. Bard, J. Electrochem. Soc. 122 (1975) 898.
[31] M. Neptune, R.L. McCreery, A. Mania, J. Med. Chem. 22
(1979) 196.
[32] I. Biryol, B. Uslu, Z. Kucukyavuz, Pharm. Biomed. Anal. 15
Acknowledgements (1996) 371.
[33] E. Bishop, W. Hussein, Analyst 109 (1984) 73.
This research was supported by the Japan Society for [34] D.J. Turk, S.A. McClintock, W.C. Purdy, Anal. Lett. 18
(1985) 2605.
the Promotion of Science (JSPS), Research for the Future
[35] B.V. Sarada, T.N. Rao, N. Spataru, D.A. Tryk, A. Fujishima,
Program, Exploratory Research on Novel Artificial submitted for publication.
Materials and Substances for Next Generation Indus- [36] A.G. Chen, Y.K. Wing, H. Chiu, S. Lee, C.N. Chen, K.
tries. Chan, J. Chromatogr. B 693 (1997) 153.