Program installation
If you are installing overa n existing versin of Recoil, it is recommended
that you bakup the settings files (Recoil.ini, isotopes.ini and calibraty.ini) if
you have manually modified them for your purposes and unistall the versin
of Recoil you already have You can then edit the new settingd files to reflect
your modifications.
Recoil is supplied with a Windows95c compatible install and unistall proces.
To install Recoil, double click.
General analysis
For a single given electronic environment of the probe nucleus, the static
interactions are characterized by the following set of hyperfine parameters,
which, in general produce a set eight Lorentzian lines :
Analysis models
Lorentzian multiplet analysis
The Lorentzian multiplet analysis is a standard analysis that permits the
user to fit several Lorentzian singlets, doublets or sextets, corresponding to
paramagnetic sites with or without a quadrupole splitting and sites with a
magnetic hyperfine field and quadrupole shift (within the first-order
perturbation limit).
Domains of applicability
Model parameters
Global parameters
Background (in counts): the off-resonance background of the spectrum
Singlet sites (parameters for each site)
CS (in mm/s): the center shift
Area (in counts mm/s): total Spectral rea of the site (proportional to the site
population)
W (in mm/s): the HWHM of the peak.