X-Radiography Diagnostic
Medical Imaging.

Image Production
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03

It is believed that the ulnar artery is


repetitively damaged by blunt compression
against the hamate. Our patient was a postal
worker who frequently used a rubber stamper.
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03
Digital Radiography
MED PHYS 1E03
Computer number representation
MED PHYS 1E03

If we have n bits we have 2n values

8 bits gives us 256 values


16 bits gives us 65636 values
32 bits give more than 4x109 values
Sampling
MED PHYS 1E03
Matrix size
MED PHYS 1E03

512 x 512

256 x 256

128 x 128

64 x 64

32 x 32
Grey level quantisation
MED PHYS 1E03

Grey level (G) given to each pixel is also governed by G = 2n

n is usually 8 giving us 256 grey levels from 0 to 255.

The number of grey levels determines quality in an image.


Grey level quantisation
MED PHYS 1E03

256 128
Grey level quantisation
MED PHYS 1E03

64 128
Grey level quantisation
MED PHYS 1E03

64 32
Grey level quantisation
MED PHYS 1E03

16 32
Grey level quantisation
MED PHYS 1E03

16 8
Grey level quantisation
MED PHYS 1E03

4 8
Grey level quantisation
MED PHYS 1E03

4 2
Digital Radiography
MED PHYS 1E03
MED PHYS 1E03
MED PHYS 1E03
Flat Panel Detectors : Indirect

MED PHYS 1E03

Phosphor layer eg Gd2O2S:Tb


or scintillator eg CsI:Tl
Placed in intimate contact with
an active matrix array Phosphor or
scintillator
X-rays converted to light
photons and the charge stored
in each pixel element.

The magnitude of the charge in


each pixel is the latent image.

Active matrix array is read out.


Glass substrate
Photosensitive
storage element Switching
element
Flat Panel Detectors : Direct

MED PHYS 1E03

Layer of photoconductive
material (amorphos silicon
or amorphos selenium)
Conductive electrode added

Capacitive storage element High voltage Photoconductor


added to store charge electrode

Interacting x-rays produce charge


which is attracted to the
electrode and the capacitive
storage element.

Active matrix array is read out.

Glass substrate
Capacitive
storage element Switching
element
Flat Panel Detectors

MED PHYS 1E03 Multiplexer


Array made up of pixel
electrodes. Amplifier
Each pixel has a switch.

Each row has a single control


line to activate switch.

Each column has a single Switching control


signal line with a readout
amplifier.

This allows the imager to be


read out one line at a time.
Unlike CCDs there is no
charge transfer from one
pixel to the other.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Initialisation state ready to


acquire image.

Switching control
All array switches are held
off while exposure is made.

Charge collected in pixel


storage elements.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switch on the first row are


now switched on.

Switching control
Charge transferred down
appropriate signal lines,
digitised and stored via the
multiplexer.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switch on the first row are


now switched on.

Switching control
Charge transferred down
appropriate signal lines,
digitised and stored via the
multiplexer.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switch on the first row are


now switched on.

Switching control
Charge transferred down
appropriate signal lines,
digitised and stored via the
multiplexer.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer ADC

Switch on the first row are


now switched on.

Switching control
Charge transferred down
appropriate signal lines,
digitised and stored via the
multiplexer.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Switching control
Next row similarly treated.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Switching control
Next row similarly treated.

Switch is closed.
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Switching control
Next row similarly treated.

Switch is closed.

Charge is read out.


Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Switching control
Next row similarly treated.

Switch is closed.

Charge is read out.


Flat Panel Detectors

MED PHYS 1E03 Multiplexer ADC

Switches now returned to


the off position.

Switching control
Next row similarly treated.

Switch is closed.

Charge is read out.


Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Next row similarly treated.

Switching control
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Next row similarly treated.

Switching control
Flat Panel Detectors

MED PHYS 1E03 Multiplexer

Switches now returned to


the off position.

Next row similarly treated.

Switching control
Flat Panel Detectors

MED PHYS 1E03 Multiplexer ADC

Switches now returned to


the off position.

Next row similarly treated.

Switching control
Contrast

MED PHYS 1E03

The difference in density (contrast) will depend


E1 E2 where on the characteristic line of the detector the
two exposures fall.
Contrast
MED PHYS 1E03
Latitude
MED PHYS 1E03
MED PHYS 1E03

X-Radiography Diagnostic
Medical Imaging.

Image Production