ETHOS D ETHOS D CONFIGURATION


ETHOS D WITH TERMINAL 20

P/N Description Power


supply
44022 ETHOS D with 230V/50Hz
Terminal 20 for storage
of up to 30 10-step
programs
44024 ETHOS D with 230V/60Hz
Terminal 20 for storage
of up to 30 10-step
programs

ETHOS D Control Options

P/N Description
INTRODUCTION 33684P Infra-Red External Temperature
Control kit for continuous
The Ethos D is the most technologically advanced temperature control of all vessels
instrumentation for total microwave sample 86303P APC-55E Automatic Pressure Control
preparation. up to 55 bar (800 psi)
In 1994, the ETHOS project was started to develop 86008 Automatic magnetic stirrer cpl
a new generation of microwave chemistry systems
based on the experience of several thousands users TECHNICAL SPECIFICATIONS
worldwide, who process some of the most difficult
samples (ores, coal, fly ash, ceramics, and plastics). Microwave labstation for quality control for
Milestone R&D engineers were given the goal to digestion, drying and evaporation/concentration
develop the safest, highest performing and most procedures.
flexible microwave platform for laboratory Patented self-sealing explosion resistant door.
applications. Dual independent magnetrons with rotating
No limits were set, only goals. microwave diffuser for homogeneous
This is ethical. microwave distribution.
This is ETHOS. Installed microwave power 1.600Watt.
Delivered microwave power 1.000Watt in
10Watt increments.
Variable microwave pulse width from 0,1 to
5.
Microwave cavity all made of 18/8 stainless
steel housing with multilayer Teflon plasma
coating applied at 350C.
All hardware protected against acid/organic
solvents with polymer coating both inside and
outside surfaces.
Dimensions 35cm x 35cm x 35cm (ca. 45L
volume).
Built-in exhaust system with automatic constant
flow control, acid and organic solvent resistant;
with flow rate 5m3/minute.
Terminal 20, 2 lines 20 characters (LDC display
black/white, flat keyboard, alphanumeric
display, 12 special function keys
Storage of up to 30 10-step programs; includes
microwave power, time, pressure, internal and
external temperature

ETHOS D Page 1
ETHOS D User Manual - Rev. 0/2001

RS-232 for connecting PC (for data base


reporting and programming of the unit),
System control options includes the following
devices.
ATC-400CE Automatic Temperature Control
for direct monitoring of vessel content, reaction
temperature up to 300C
APC-55 Automatic Pressure Control for direct
sensing monitor of vessel content, reaction
pressure control up to 55bar.
IRTC-500 Infra Red Temperature Control,
advanced non-contact monitor to protect rotors
and vessels with optional magnetron regulation
or system shutdown.
Dimensions of the microwave unit: 55cm large
x 55cm long and 70cm high.
Power supply 230V, 50 or 60Hz, 10,6A,
2.400Watt.
Weight approximately 90 kg.

ETHOS D Page 2
ETHOS D User Manual - Rev. 0/2001

A LOOK AT THE HIDDEN SIDE OF THE SAFETY THROUGH PATENTED DOOR


ETHOS UNIT DESIGN

The most important safety features are the ones you The door design is the most critical issue when
dont see from the outside. developing new microwave hardware.
For this reasons Milestone lets you see under the Mechanical strength and perfect sealing of the
cabinet cover of the ETHOS. cavity to avoid microwave leaks are prerequisites.
Milestones engineers, on the basis of the movable
SAFETY THROUGH MECHANICAL DESIGN door principle used in high pressure autoclaves,
designed a door with an innovative opening/self
The ETHOS is not a kitchen type nor a resealing mechanism.
commercially modified kitchen microwave unit; it The door frame is mounted on four spring-loaded
is a rugged, heavy duty professional system high strength steel bars (picture).
designed to guarantee years and years of trouble
free and safe operation.
The hardware The microwave cavity is quite large,
to accommodate different types of accessories.
Dimensions are 35cm x 35cm x 35cm, for a volume
of approximately 43L.
The door is, as standard feature, made of stainless
steel; however, on request, it is possible to have a
door with a high impact shock resistant 27cm x
27cm glass window for easy viewing of the
illuminated cavity (picture).
is manufactured with 18/8 stainless steel, 1,5mm
gauge (picture).

In the event of a sudden overpressure inside the


cavity, the entire door opens horizontally (moves
out) presenting a large surface area for immediate
and safe release of the overpressure.
Microwave power is instantaneously cut off.
Immediately afterward, the door is pulled back by
the springs, resealing the cavity.
The microwave cavity door opens downwards to a
flat horizontal position, forming a stable work
platform for handling microwave accessories before
and after digestion procedures.

ETHOS D Page 3
ETHOS D User Manual - Rev. 0/2001

SAFETY THROUGH ELECTRICAL DESIGN PERFORMANCE THROUGH MICROWAVE


HARDWARE DESIGN
The ETHOS microwave platform is equipped with
a total of 7 microswitches to assure the highest The heart of the ETHOS platform is the
safety conditions. innovative and patented microwave
Four (two on each side of the door) fast release feed/distribution system to the working cavity.
microswitches instantaneously interrupt the Two 800Watt industrial magnetron, each with its
microwave emission whenever the door is opened own high voltage power supply, assure a powerful
(previous picture). 1.600Watt of microwave installed power (picture).
One safety switch on the cabinet cuts off
microwave emission whenever the cabinet is Two high flow cooling fans guarantee peak
detached (picture). performance of the magnetrons.
The microwaves are first fed into a pre-mixing
chamber.
From here an ingenious pyramid shaped rotating
metal diffuser evenly distributes the microwaves in
the cavity (picture).

Two thermal switches on magnetrons body cut off


power in case of overheating (picture).

ETHOS D Page 4
ETHOS D User Manual - Rev. 0/2001

The ETHOS microwave platform has been


designed with the microwave emission into the
cavity from the back wall instead of the
conventional position on top of the cavity.
This unique construction allows both the top and
the bottom of the cavity to be free.
A free top allows the installation of batch reactors,
while the free bottom makes possible the
installation of a continuous flow reactor or of an
advanced magnetic stirring device.
The ETHOS microwave platform can be upgraded
with suitable modules/accessories to carry out: acid
digestion, vacuum drying,
evaporation/concentration, protein/peptide
hydrolysis, solvent extraction, batch organic
synthesis and continuous flow organic synthesis.

DURABILITY THROUGH CORROSION


RESISTANCE

One of the key issues was assuring a long lifetime


of our unit when performing digestions with
aggressive acids (HNO3, H2SO4, HCl, H3PO4, HF).
The all stainless steel hardware undergoes a
pretreatment (sandblasting plus heat cycles) to
Surfaces are Teflon coated with the diffuser assure best bonding conditions between the
protected by an inert high density polypropylene Teflon coating and metal surfaces.
plate. The microwave cavity, the top cavity, the door and
This unique microwave design and the advanced even the external surface of the air inlet (grid) are
microprocessor control assures optimal microwave plasma coated with 5 layers of Teflon applied at
distribution in the cavity, thus preventing hot 350C (picture).
spots.
The installed power of 1.600Watt has been limited
by design to operate at approximately 2/3 of the
maximum power (1.000Watt) to assure longer
magnetrons lifetime.
The 1.000Watt of delivered power are delivered in
10Watt increments to fit all types of applications.

PERFORMANCE THROUGH VERSATILITY

To fulfill the goals of making the ETHOS the most


versatile microwave platform, Milestones
engineers built 12 microwave leakage protected
(inlet/outlet) connectors.
Temperature and pressure sensor, connectors for
vacuum and inert gas supply, can be all easily
installed without any hardware modifications.

DURABILITY THROUGH CHASSIS DESIGN

PERFORMANCE THROUGH MODULARITY The cabinet chassis is manufactured with magnetic


18/8 stainless steel (1,0mm gauge) to act as a

ETHOS D Page 5
ETHOS D User Manual - Rev. 0/2001

Faraday cage to eliminate potential interference to DURABILITY THROUGH USER-FRIENDLY


other instruments (picture). SERVICE

As perfection is not of this world, easy access in


case of servicing is a must.
Please refer to the separate ETHOS Service Manual
It is protected against acid/organic solvent vapors
for further information.
through high temperature polymer coating on both
More details about this subject are available in the
ETHOS Service Manual.

the outside and inside surfaces.

DURABILITY THROUGH ADVANCED


EXHAUST/COOLING SYSTEM

This is another important factor for assuring long


life of digestion vessels and safe working
conditions.
The ETHOS platform is supplied with a heavy
duty, 5m3/minute exhaust fan placed above the
cavity and installed separately from the electronics
to eliminate the danger of acid corrosion of the
electronics (previous picture).
The air flow from the grids at the bottom of the
cavity cools the external surfaces of the rotor and
carries away any acid vapors released in the cavity.
An acid resistant flexible hose connects the exhaust
fan to a fume hood assuring a safe working
environment.

ETHOS D Page 6
ETHOS D User Manual - Rev. 0/2001

MMR4/12 MMR-12 REFLUX SET-UP

ETHOS D COMPATIBILITY
WITH DIGESTION, DRYING
The MCR/MMR series rotors require a vacuum
AND CONCENTRATION source.
ROTORS Below are the options available.

The ETHOS D can accommodate the following P/N Description


acid digestion vessels and rotors: VAC1000 VAC-2000 Vacuum Solvent
Recovery Module
P/N Description VAC2000 VAC-1000 Scrubber for
SK6T HPR-1000/6M High Pressure neutralization of acids
Monoblock for temperature
SK6TP HPR-1000/6M High Pressure
Monoblock for temperature and
pressure
SK-10T HPR-1000/10S High Pressure
Segmented rotor for temperature
control
SK-10TP HPR-1000/10S High Pressure
Segmented rotor for temperature and
pressure
SK-12T MPR-600/12S Medium Pressure
Segment rotor for temperature
SK-12TP MPR-600/12S Medium Pressure
Segment rotor for temperature and
pressure
SK-6LV MPR-600/6S Low Pressure Rotor
cpl for temperature control

All the SK starter kits include the complete rotor


with work station and tension wrench to
open/closed vessels.

The ETHOS D can accommodate the following


drying and evaporation/concentration rotors and
pre-digestion reflux rotors:

P/N Description
40103 MCR-1 Drying rotor in
polypropylene 6,5L
34649 MCR-6 Concentration rotor 6
position empty
MMR1/8 MMR-8 PTFE EVAPORATION
SET-UP
MMR2/8 MMR-8 WEFLON
EVAPORATION SET-UP
MMR3/8 MMR-8 GLASS BOTTLES SET-
UP
MMR4/8 MMR-8 REFLUX SET-UP
MMR1/12 MMR-12 PTFE EVAPORATION
SET-UP
MMR2/12 MMR-12 WEFLON
EVAPORATION SET-UP
MMR3/12 MMR-12 GLASS BOTTLES SET-
UP

ETHOS D Page 7
ETHOS D User Manual - Rev. 0/2001

INSTALLATION OF THE ETHOS Remove the carton box located in the upper part of
D WITH TERMINAL 20
WARNING
Never install the ETHOS D under a fume hood!

In fact, if you should install the microwave system


under a fume hood, the acid vapors might
contaminate the air used for ventilating the
electronic components, exposing them to a possible
corrosion.

UNPACKING AND CHECKING LIST

When unpacking, verify that all parts correspond to the box (picture) and keep it apart.
the packing list included in the shipment.
Utmost care must be taken during unpacking of the
ETHOS D to avoid any scratch or damage to the It contains accessories which will be taken during
special acid and organic solvent resistant external the installation of the ETHOS D.
coating. The carton box should include the following:
n. 1 Milestone ETHOS D microwave
Carefully cut the protective belts, as shown in the labstation;
next picture. n. 1 Power cord for ETHOS D;
n. 1 Plastic hose 2m long;
n. 1 Terminal 20
n. 1 Interface cable 25-pin for connecting the
ETHOS D to the l terminal model 20

ETHOS D Page 8
ETHOS D User Manual - Rev. 0/2001

Lift-up the main carton box to facilitate the removal WARNING


of the ETHOS D (picture). The weight of the ETHOS D is approximately 90
The ETHOS D comes protected by a plastic bag to kg.
avoid it getting wet during transportation. At least 3 persons should remove the system from
its box and locate it on the bench.

Provide a bench of at least 1m length and 80cm


wide and locate the ETHOS D on the bench. Open now the boxes containing the accessories of
A proper location of the ETHOS D is next to a the ETHOS D.
fume hood, since the microwave unit should be
connected to the fume hood by using the hose
supplied (details will follow).

Notice the ETHOS D power cord (picture).

ETHOS D Page 9
ETHOS D User Manual - Rev. 0/2001

Notice the plastic hose 2ml long (picture).

Notice also the Terminal 20 with its 25 pin


interface cable (picture).

ETHOS D Page 10
ETHOS D User Manual - Rev. 0/2001

ETHOS D OVERVIEW

Notice some of the most important details of the


ETHOS D.
Main switch (picture).

This is the air inlet for the air exhaust system,


located on the side of the unit. Dont worry about
microwave leakages! A metal grid, which can be
seen through the plastic connector, reflects
microwaves back to the cavity, not allowing any
leakage through the opening. Do not block or
obstruct the opening since air circulation through
the microwave cavity is an important safety aspect
of microwave closed vessel acid digestion.

Overview of the back side of the ETHOS D


WARNING
During the ETHOS D installation the main switch
should be in the OFF position.

Exhaust module main switches (picture).

The ETHOS D comes with the exhaust module as


standard feature.
(picture). Notice in the upper part (left) the open
connection for the plastic for a second optional
blower.
WARNING
During the ETHOS D installation both the switches
should be in the OFF position.
Detail of the interfaces of the ETHOS D (picture);
at left a standard RS-232 9 pin interface port, to be
used for connecting the microwave to external PC;

ETHOS D Page 11
ETHOS D User Manual - Rev. 0/2001

at right the special 25 pin interface port to be used Detail of the left back panel of the ETHOS D
for connecting the ETHOS D to the Terminal 20. (picture), showing the connection to the power cord
(at right, down), the power connection to external
devices (right, up), and the fuse holder with two
fuses.

More details about this subject are available in the


ETHOS Service Manual.

ETHOS D Page 12
ETHOS D User Manual - Rev. 0/2001

TERMINAL 20 INTERFACE PORTS ETHOS D SETUP

The picture below shows the back side of the Connect the 25 pin interface cable supplied to the
Terminal 20. back side port of the ETHOS D and use a
It is possible to see the Special 25 pin interface port screwdriver to fix it well (picture) To the left is the
for connecting the terminal to the ETHOS D main RS-232 9 pin for connecting the ETHOS D to an
unit. external PC.

Connect the other end of the 25 pin cable to the


Terminal 20 and again use a screwdriver to fix it
well (picture).

ETHOS D Page 13
ETHOS D User Manual - Rev. 0/2001

WARNING
Both the 25 pin connections must be very tight;
damages may occur when the ETHOS and
Terminal 20 are not properly connected.

Plug-in the supplied power cord at the back side of


the ETHOS (picture).

Introduce now the L shaped adapter to the


exhaust outlet located at the left of the ETHOS
back panel. Your new ETHOS D is now ready to work!
Connect the exhaust hose to the L shaped adapter
(picture).

The free end of the exhaust hose should be placed


under a fume hood or to an existing exhaust
(picture).

ETHOS D Page 14