Você está na página 1de 10

C H A P T ER 6

Clinical Masking

William S. Yacullo

In the rst edition of this text, Sanders (1972) wrote the fol- a masking stimulus must be applied to the nontest ear to
lowing introduction to his chapter on clinical masking: eliminate its participation.
Of all the clinical procedures used in auditory assess-
ment, masking is probably the most often misused
and the least understood. For many clinicians the Air-Conduction Testing
approach to masking is a haphazard hit-or-miss bit Cross hearing occurs when a stimulus presented to the test
of guesswork with no basis in any set of principles. ear crosses over and is perceived in the nontest ear. There
(p 111) are two parallel pathways by which sound presented through
an earphone (i.e., an air-conduction transducer) can reach
Unfortunately, this statement may still hold true today.
the nontest ear. Specically, there are both bone-conduction
The principles of clinical masking are difcult for many
and air-conduction pathways between an air-conduction
beginning clinicians to understand. Although the clinician
signal presented at the test ear and the sound reaching the
can apply masking formulas and procedures appropriately
nontest ear cochlea (Studebaker, 1979). First, the earphone
in most clinical situations, a lack of understanding of the
can vibrate with sufcient force to cause deformations of
underlying theoretical constructs becomes evident dur-
the bones of the skull. An earphone essentially can function
ing cases where modication of a standard procedure is
required. A lack of understanding of the underlying con- as a bone vibrator at higher sound pressures. Because both
cepts of masking often leads to misuse of clinical procedures. cochleas are housed within the same skull, the outcome is
Theoretical and empirical bases of masking provide a stimulation of the nontest ear cochlea through bone con-
strong foundation for the understanding of applied clini- duction. Second, sound from the test earphone can travel
cal masking procedures. It will become evident throughout around the head to the nontest ear, enter the opposite ear
this chapter that there is not a single correct approach to canal, and nally reach the nontest ear cochlea through an
clinical masking. Any approach to clinical masking that is air-conduction pathway. Because the opposite ear typically
based on sound theoretical constructs and veried through is covered during air-conduction testing, sound attenuation
clinical experience is correct. One approach will not meet provided by the earphone will greatly minimize or elimi-
all clinical needs. A strong foundation in the underlying nate the contribution of the air-conduction pathway to the
concepts of clinical masking serves three purposes. First, process of cross hearing. Consequently, cross hearing dur-
ing air-conduction testing is considered primarily a bone-
it allows the clinician to make correct decisions about the
conduction mechanism.
need for masking. Second, it allows the clinician to make a
Cross hearing is the result of limited interaural attenu-
well-informed decision when selecting a specic approach
ation (IA). IA refers to the reduction of energy between
to clinical masking. Finally, it allows the clinician to apply
ears. Generally, it represents the amount of separation or
and modify a clinical masking procedure appropriately.
the degree of isolation between ears during testing. Speci-
cally, it is the decibel difference between the hearing level
THE NEED FOR MASKING (HL) of the signal at the test ear and the HL reaching the
nontest ear:
A major objective of the basic audiologic evaluation is
assessment of auditory function of each ear. There are situ- IA = dB HL Test Ear dB HL Nontest Ear
ations during both air-conduction and bone-conduction
testing when this may not occur. Although a puretone or Consider the following hypothetical examples pre-
speech stimulus is being presented through a transducer to sented in Figure 6.1. You are measuring puretone air-
the test ear, the nontest ear can contribute partially or totally conduction threshold using traditional supra-aural
to the observed response. Whenever it is suspected that the earphones. A puretone signal of 90 dB HL is presented to the
nontest ear is responsive during evaluation of the test ear, test ear. Because of limited IA, a portion of the test signal can

77
78 SECTION I Basic Tests and Procedures

contact with the skin covering the cranial skull. Insert ear-
phones are coupled to the ear by insertion into the ear canal.
Generally, IA increases as the contact area of the trans-
ducer with the skull decreases (Zwislocki, 1953). More
specically, IA is greater for supra-aural than circumaural
earphones. Furthermore, IA is greatest for insert earphones
(Killion et al., 1985; Sklare and Denenberg, 1987), partly
because of their smaller contact area with the skull. (The
reader is referred to Killion and Villchur, 1989; Zwislocki
A et al., 1988, for a review of advantages and disadvantages of
earphones in audiometry.) Because supra-aural and insert
earphones are most typically used during audiologic testing,
they will be the focus of this discussion.
There are different approaches to measuring IA for air-
conducted sound (e.g., masking method, compensation
method, method of best beats; the reader is referred to
Zwislocki, 1953, for discussion). The most direct approach,
however, involves measurement of transcranial thresholds
(Berrett, 1973). Specically, IA is measured by obtaining
unmasked air-conduction (AC) thresholds in subjects with
B unilateral, profound sensory/neural hearing loss and then
FIGURE 6.1 Interaural attenuation (IA) is calculated as calculating the threshold difference between the normal and
the difference between the hearing level (HL) of the sig- impaired ears:
nal at the test ear and the HL reaching the nontest ear
cochlea. A puretone signal of 90 dB HL is being presented IA = Unmasked AC Impaired Ear Unmasked AC Normal Ear
to the test ear through traditional supra-aural earphones.
Example A: If IA is 40 dB, then 50dB HL is reaching the For example, if unmasked air-conduction thresholds are
nontest ear cochlea. Example B: If IA is 80 dB, then 10 obtained at 60 dB HL in the impaired ear and 0 dB HL in
dB HL is reaching the nontest ear cochlea. (From Yacullo the normal ear, then IA is calculated as 60 dB:
WS. (1996) Clinical Masking Procedures. 1st ed. Boston,
MA: Allyn & Bacon, 1996, p 3. Adapted by permission of IA = 60 dB HL 0 dB HL
Pearson Education, Inc., Upper Saddle River, NJ.) = 60dB
There is the assumption that air- and bone-conduction
thresholds are equal (i.e., no air-bone gaps) in the ear with
reach the nontest ear cochlea. If IA is 40 dB, then 50 dB HL normal hearing.
theoretically is reaching the nontest ear: Figure 6.2 illustrates the expected unmasked puretone
air-conduction thresholds in an individual with normal
IA = dB HL Test Ear dB HL Nontest Ear hearing in the left ear and a profound sensory/neural hearing
= 90 dB HL 50 dB HL loss in the right ear. Unmasked bone-conduction thresholds,
regardless of bone vibrator placement, are expected at HLs
= 40 dB
consistent with normal hearing in the left ear. If appropriate
If IA is 80 dB, then only 10 dB HL is reaching the nontest contralateral masking is not used during air-conduction test-
ear. It should be apparent that a greater portion of the test ing, then a shadow curve will result in the right ear. Because
signal can reach the nontest ear when IA is small. Depend- cross hearing is primarily a bone-conduction mechanism,
ing on the hearing sensitivity in the nontest ear, cross hear- unmasked air-conduction thresholds in the right ear will
ing can occur. shadow the bone-conduction thresholds in the left (i.e.,
IA during earphone testing is dependent on three fac- better) ear by the amount of IA. For example, if IA for air-
tors: Transducer type, frequency spectrum of the test sig- conducted sound is equal to 60 dB at all frequencies, then
nal, and individual subject. There are three major types of unmasked air-conduction thresholds in the right ear theo-
earphones currently used during audiologic testing: Supra- retically will be measured 60 dB above the bone-conduction
aural, circumaural, and insert (American National Stan- thresholds in the better ear. The shadow curve does not
dards Institute/Acoustical Society of America [ANSI/ASA], represent true hearing thresholds in the right ear. Rather, it
2010). Supra-aural earphones use a cushion that makes reects cross-hearing responses from the better (i.e., left) ear.
contact solely with the pinna. Circumaural earphones use When using supra-aural earphones, IA for puretone
a cushion that encircles or surrounds the pinna, making air-conducted signals varies considerably, particularly across
CHAPTER 6 Clinical Masking 79

FIGURE 6.2 Expected unmasked puretone air- and bone-conduction thresholds in an


individual with normal hearing in the left ear and a profound sensory/neural hearing
loss in the right ear.Without the use of appropriate contralateral masking, a shadow
curve will result in the right ear. Unmasked air-conduction thresholds in the right ear
will shadow the bone-conduction thresholds in the better (i.e., left) ear by the amount
of interaural attenuation.(From Yacullo WS. (1996) Clinical Masking Procedures. 1st ed.
Boston, MA: Allyn & Bacon, 1996, p 7. Adapted by permission of Pearson Education,
Inc., Upper Saddle River, NJ.)

subjects, ranging from about 40 to 85 dB (Berrett, 1973; ied length, a nipple adaptor, and a disposable foam eartip.
Chaiklin, 1967; Coles and Priede, 1970; Killion et al., 1985; A major advantage of the 3A insert earphone is increased
Sklare and Denenberg, 1987; Smith and Markides, 1981; IA for air-conducted sound, particularly in the lower fre-
Snyder, 1973). Your assumption about IA will inuence the quencies (Hosford-Dunn et al., 1986; Killion et al., 1985;
decision about the need for contralateral masking. The use Sklare and Denenberg, 1987; Van Campen et al., 1990). This
of a smaller IA value assumes that there is smaller separa- is clearly illustrated in the results of a study by Killion et al.
tion between ears. Consequently, contralateral masking will (1985) (Figure 6.3).
be required more often. When making a decision about the Increased IA with 3A insert earphones is the result of
need for contralateral masking during clinical practice, a two factors: (1) Reduced contact area of the transducer with
single value dening the lower limit of IA is recommended the skull and (2) reduction of the occlusion effect (OE).
(Studebaker, 1967a). Zwislocki (1953) evaluated IA for three types of earphones:
Based on currently available data, a conservative estimate circumaural, supra-aural, and insert. Results suggested that
of IA for supra-aural earphones is 40 dB at all frequencies. IA for air-conducted sound increased as the contact area of
Although this very conservative estimate will take into the earphone with the skull decreased. When an acoustic
account the IA characteristics of all individuals, it will result signal is delivered through an earphone, the resultant sound
in the unnecessary use of masking in some instances. pressure acts over a surface area of the skull determined by
Commonly used insert earphones are the Etymotic the earphone cushion. The surface area associated with a
Research ER-3A (Killion, 1984) and the E-A-RTONE 3A small eartip will result in a smaller applied force to the skull,
(E-A-R Auditory Systems, 1997). The ER-3A and the E-A- resulting in reduced bone-conduction transmission.
RTONE 3A insert earphones are considered functionally Chaiklin (1967) has also suggested that IA may be
equivalent because they are built to identical specications increased in the low frequencies with a deep insert because
(Frank and Vavrek, 1992). Each earphone consists of a of a reduction of the OE. ANSI/ASA (2010) denes the OE
shoulder-mounted transducer, a plastic sound tube of spec- as an increase in loudness for bone-conducted sound at
80 SECTION I Basic Tests and Procedures

the button transducer (Blackwell et al., 1991; Hosford-


Dunn et al., 1986). Blackwell et al. (1991) compared the IA
obtained with a standard supra-aural earphone (TDH-50P)
and a button transducer tted with a standard immittance
probe cuff. Although greater IA was observed with the but-
ton transducer, the difference between the insert and supra-
aural earphone did not exceed 10 dB at any frequency.
There are only limited data available regarding IA of
3A insert earphones using deeply or intermediately inserted
foam eartips. IA values vary across subjects and frequency,
ranging from about 75 to 110 dB at frequencies of 1,000 Hz
and about 50 to 95 dB at frequencies >1,000 Hz (Killion
et al., 1985; Sklare and Denenberg, 1987; Van Campen et al.,
1990). Based on Studebakers (1967a) recommendation, we
will again use the smallest IA values reported when making
a decision about the need for contralateral masking. To take
advantage of the signicantly increased IA proved by the 3A
insert in the lower frequencies, a single value of IA will not
be employed across the frequency range.
FIGURE 6.3 Average and range of interaural attenu- Based on currently available data, conservative estimates
ation values obtained on six subjects using two ear- of IA for 3A insert earphones with deeply inserted foam eartips
phones: TDH-39 encased in MX-41/AR supra-aural are 75 dB at 1,000 Hz and 50 dB at frequencies >1,000 Hz.
cushion () and ER-3A insert earphone with deeply The IA values recommended clinically for 3A earphones
inserted foam eartip (). (From Killion MC, Wilber LA, assume that deeply inserted foam eartips are used. Maximum
Gudmundsen GI. (1985) Insert earphones for more inte- IA is achieved in the low frequencies when a deep eartip
raural attenuation. Hear Instrum. 36, 34, 36. Reprinted
insertion is used (Killion et al., 1985). The recommended
with permission from Hearing Instruments, 1985, p 34.
deep insertion depth is achieved when the outer edge of the
Hearing Instruments is a copyrighted publication of
Advanstar Communications Inc. All rights reserved.) eartip is 2 to 3 mm inside the entrance of the ear canal. Con-
versely, a shallow insertion is obtained when the outer edge
of the eartip protrudes from the entrance of the ear canal
frequencies below 2,000 Hz when the outer ear is covered (E-A-R Auditory Systems, 1997). An intermediate insertion
or occluded. There is evidence that the OE inuences the is achieved when the outer edge of the eartip is ush with the
measured IA for air-conducted sound (e.g., Berrett, 1973; opening of the ear canal (Van Campen et al., 1990). There
Chaiklin, 1967; Feldman, 1963; Killion et al., 1985; Littler are limited data suggesting that IA is similar for either inter-
et al., 1952; Van Campen et al., 1990; Zwislocki, 1953). In mediate or deep insertion of the foam eartip. However, a
fact, there is an inverse relationship between magnitude shallow insertion appears to signicantly reduce IA (Killion
of the OE and the measured IA in the lower frequencies. et al., 1985; Sklare and Denenberg, 1987; Van Campen et al.,
Specically, an earphone that reduces the OE will exhibit 1990). Remember that a major factor contributing to supe-
increased IA for air-conducted sound. Recall that cross rior IA of the 3A insert earphone is a signicantly reduced
hearing occurs primarily through the mechanism of bone OE. There is evidence that the OE is negligible when using
conduction. When the nontest ear is covered or occluded either deeply or intermediately inserted insert earphones.
by an air-conduction transducer, the presence of an OE will In fact, the advantage of a greatly reduced OE is lost when
enhance hearing sensitivity for bone-conducted sound in a shallow insertion is used (Berger and Kerivan, 1983). To
that ear. Consequently, the separation between ears (i.e., achieve maximum IA with 3A insert earphones, deeply
IA) is reduced. The increased IA for air-conducted sound inserted eartips are strongly recommended.
observed in the lower frequencies when using 3A insert More recently, E-A-R Auditory Systems (2000a, 2000b)
earphones (with deeply inserted foam eartips) is primarily introduced a next-generation insert earphone, the E-A-
related to the signicant reduction or elimination of the OE. RTONE 5A. The lengthy plastic sound tube that conducted
The OE is presented in greater detail later in this chapter sound from the body-level transducer of the 3A has been
in the section on clinical masking procedures during bone- eliminated in the 5A model; rather, the foam eartip is cou-
conduction audiometry. pled directly to an ear-level transducer. Very limited data
If increased IA is a primary goal when selecting an obtained with only two subjects (unpublished research by
insert earphone, then the 3A is the transducer of choice. Evi- Killion, 2000, as cited in E-A-R Auditory Systems, 2000b)
dence suggests that the 3A insert earphone provides signi- suggest that the average IA for puretone stimuli ranging
cantly greater IA, particularly in the lower frequencies, than from 250 to 4,000 Hz is equivalent (within approximately
CHAPTER 6 Clinical Masking 81

5 dB) to the average values reported for the 3A insert ear- ognition thresholds. Rather, a different response task when
phone (Killion et al., 1985). measuring different speech thresholds in each ear (i.e., SDT
IA for speech is typically measured by obtaining speech in one ear and SRT in the other) can affect the measured
recognition thresholds (SRTs) in individuals with unilateral, IA for speech. Comparison of SRTs between ears or SDTs
profound sensory/neural hearing loss. Specically, the dif- between ears generally should result in the same measured
ference in threshold between the normal ear and impaired IA. Smith and Markides (1981) measured IA for speech in
ear without contralateral masking is calculated: 11 subjects with unilateral, profound hearing loss. IA was
calculated as the difference between the SDT in the better
IA = Unmasked SRTImpaired Ear SRTNormal Ear ear and the unmasked SDT in the poorer ear. The range
of IA values was 50 to 65 dB. It is interesting to note that
Recall that SRT represents the lowest HL at which speech the lowest IA value reported for speech using a detection
is recognized 50% of the time (ANSI/ASA, 2010; Ameri- task in each ear was 50 dB, a value comparable to the lowest
can Speech-Language-Hearing Association [ASHA], 1988). minimum reported IA value (i.e., 48 dB) for spondaic words
IA for spondaic words presented through supra-aural ear- (e.g., Martin and Blythe, 1977; Snyder, 1973).
phones varies across subjects and ranges from 48 to 76 dB There is also some evidence that it may be appropri-
(Martin and Blythe, 1977; Sklare and Denenberg, 1987; Sny- ate to use a more conservative estimate of IA when making
der, 1973). Again, a single value dening the lower limit of a decision about the need for contralateral masking during
IA is recommended when making a decision about the need assessment of suprathreshold speech recognition. Although
for contralateral masking (Studebaker, 1967a). A conserva- IA for the speech signal remains constant during measure-
tive estimate of IA for spondees, therefore, is 45 dB when ment of threshold or suprathreshold measures of speech
using supra-aural earphones (Konkle and Berry, 1983). The recognition (i.e., the decibel difference between the level of
majority of audiologists measure SRT using a 5-dB step size the speech signal at the test ear and the level at the non-
(Martin et al., 1998). Therefore, the IA value of 48 dB is typ- test ear cochlea), differences in the performance criterion
ically rounded down to 45 dB. for each measure must be taken into account when select-
There is considerable evidence that speech can be ing an appropriate IA value for clinical use. SRT is dened
detected at a lower HL than that required to reach SRT. relative to a 50% response criterion. However, suprathresh-
Speech detection threshold (SDT) is dened as the lowest old speech recognition performance can range from 0% to
HL at which speech can be detected or discerned 50% 100%.
of the time (ASHA, 1988). The SRT typically requires an Konkle and Berry (1983) provide an excellent ratio-
average of about 8 to 9 dB greater HL than that required nale for the use of a more conservative estimate of IA when
for the detection threshold (Beattie et al., 1978; Chaiklin, measuring suprathreshold speech recognition. They suggest
1959; Thurlow et al., 1948). Given this relationship between that the fundamental difference in percent correct criterion
the two speech thresholds, Yacullo (1996) has suggested requires the specication of nontest ear cochlear sensitiv-
that a more conservative value of IA may be appropriate ity in a different way than that used for threshold measure-
when considering the need for contralateral masking dur- ment. If suprathreshold speech recognition materials are
ing measurement of SDT. presented at an HL equal to the SRT, then a small percent-
Consider the following hypothetical example. You are age of the test items can be recognized. It should be noted
measuring speech thresholds in a patient with normal hear- that the percentage of test words that can be recognized
ing in the right ear and a profound, sensory/neural hear- at an HL equal to SRT is dependent on the type of speech
ing loss in the left ear. If the patient exhibits the minimum stimuli, as well as on the talker and/or recorded version of
reported IA value for speech of 48 dB, then an SRT of 0 dB a speech recognition test. Regardless of the type of speech
HL would be measured in the right ear and an unmasked stimulus (e.g., meaningful monosyllabic words, nonsense
SRT of 48 dB HL would be measured in the left ear. If an syllables, or sentences) and the specic version (i.e., talker/
unmasked SDT is subsequently measured in the left ear, it is recording) of a speech recognition test, 0% performance
predicted that the threshold would occur at an HL of about may not be established until an HL of about 10 dB rela-
8 to 9 dB lower than the unmasked SRT. An unmasked SDT tive to the SRT. Konkle and Berry (1983) recommend that
would be expected to occur at about 39 to 40 dB HL. Com- the value of IA used for measurement of suprathreshold
parison of the unmasked SDT in the impaired ear with the speech recognition should be estimated as 35 dB. That is,
SRT in the normal ear theoretically would result in mea- the IA value of 45 dB (rounded down from 48 dB) based
sured IA of approximately 39 to 40 dB. When an unmasked on SRT measurement is adjusted by subtracting 10 dB. This
SDT is measured and the response is compared to the SRT adjustment in the estimate of IA reects differences in per-
in the nontest ear, a more conservative estimate of IA for cent correct criterion used for speech threshold and supra-
speech may be appropriate. threshold measurements.
It should be noted that the actual IA for speech does not The majority of audiologists use an IA value of 40 dB
change during measurement of speech detection and rec- for all air-conduction measurements, both puretone and
82 SECTION I Basic Tests and Procedures

speech, when making a decision about the need for con- into vibration, both cochleas can be potentially stimulated.
tralateral masking (Martin et al., 1998). The use of a single Consequently, an unmasked bone-conduction threshold
IA value of 40 dB for both threshold and suprathreshold can reect a response from either cochlea or perhaps both.
speech audiometric measurements can be supported. Given Although a bone vibrator may be placed at the side of the
the smallest reported IA value of 48 dB for spondaic words, test ear, it cannot be assumed that the observed response is in
a value of 40 dB is somewhat too conservative during mea- fact from that ear.
surement of SRT. However, it should prove adequate during Consider the following example. You have placed a bone
measurement of SDT and suprathreshold speech recogni- vibrator at the right mastoid process. A puretone signal of
tion when a more conservative estimate of IA (by approxi- 50 dB HL is presented. If IA is considered to be 0 dB, then it
mately 10 dB) may be appropriate. should be assumed that a signal of 50 dB HL is potentially
Unfortunately, there are only very limited data available reaching both cochleas. It should be apparent that there is
about IA for speech when using insert earphones. Sklare essentially no separation between the two cochleas during
and Denenberg (1987) reported IA for speech (i.e., SRT unmasked bone-conduction audiometry.
using spondaic words) in seven adults with unilateral, pro- Based on currently available data, a conservative estimate
found sensory/neural hearing loss using ER-3A insert ear- of IA for bone-conducted sound is 0 dB at all frequencies.
phones. IA ranged from 68 to 84 dB. It should be noted that
the smallest reported value of IA for spondaic words (i.e.,
68 dB) is 20 dB greater when using 3A insert earphones with WHEN TO MASK
deeply inserted foam eartips (Sklare and Denenberg, 1987) Contralateral masking is required whenever there is the pos-
than when using supra-aural earphones (i.e., 48 dB) (Mar- sibility that the test signal can be perceived in the nontest ear.
tin and Blythe, 1977; Snyder, 1973). Therefore, a value of IA is one of the major factors that will be considered when
60 dB represents a very conservative estimate of IA for evaluating the need for masking. The basic principles under-
speech when using 3A insert earphones. This value is derived lying the decision-making processes of when to mask during
by adding a correction factor of 20 dB to the conservative IA puretone and speech audiometry will now be addressed.
value used with supra-aural earphones (i.e., 40 dB) for all
threshold and suprathreshold measures of speech.
Based on currently available data, conservative estimates Puretone Audiometry: Air
of IA for all threshold and suprathreshold measures of speech Conduction
are 40 dB for supra-aural earphones and 60 dB for 3A insert
When making a decision about the need for masking dur-
earphones with deeply inserted foam eartips.
ing puretone air-conduction testing, three factors need to
be considered: (1) IA, (2) unmasked air-conduction thresh-
Bone-Conduction Testing old in the test ear (i.e., HL at the test ear), and (3) bone-
conduction hearing sensitivity (i.e., threshold) in the non-
There are two possible locations for placement of a bone
test ear. Recall that when cross hearing occurs, the nontest
vibrator (typically, the Radioear B-71) during puretone
ear is stimulated primarily through the bone-conduction
threshold audiometry: The mastoid process of the tempo-
mechanism. When a decision is made about the need
ral bone and the frontal bone (i.e., the forehead). Although
for contralateral masking, the unmasked air-conduction
there is some evidence that a forehead placement produces
threshold in the test ear (ACTest Ear) is compared to the bone-
more reliable and valid thresholds than a mastoid place-
conduction threshold in the nontest ear (BCNontest Ear). If
ment (see Dirks, 1994, for further discussion), the major-
the difference between ears equals or exceeds IA, then air-
ity (92%) of audiologists in the United States continue to
conduction threshold in the test ear must be remeasured
place a bone-conduction transducer on the mastoid process
using contralateral masking. The rule for when to mask dur-
(Martin et al., 1998).
ing puretone air-conduction testing can be stated as follows:
IA is greatly reduced during bone-conduction audiom-
Contralateral masking is required during puretone air-
etry. IA for bone-conducted sound when using a bone vibra-
conduction audiometry when the unmasked air-conduction
tor placed at the forehead is essentially 0 dB at all frequencies;
threshold in the test ear equals or exceeds the apparent bone-
IA when using a mastoid placement is approximately 0 dB at
conduction threshold (i.e., the unmasked bone-conduction
250 Hz and increases to about 15 dB at 4,000 Hz (Studebaker,
threshold) in the nontest ear by a conservative estimate of IA:
1967a). Regardless of the placement of a bone vibrator (i.e.,
mastoid vs. forehead), it is generally agreed that IA for bone- AC Test Ear BC Nontest Ear IA
conducted sound at all frequencies is negligible and should
be considered 0 dB (e.g., Dirks, 1994; Hood, 1960; Sanders This rule is consistent with the guidelines for manual
and Rintelmann, 1964; Studebaker, 1967a). When a bone puretone threshold audiometry recommended by ASHA
vibrator, regardless of its location, sets the bones of the skull (2005).
CHAPTER 6 Clinical Masking 83

FIGURE 6.4 Audiogram illustrating the need for contralateral masking during
puretone air-conduction audiometry. See text for discussion.

Note that the term apparent bone-conduction thresh- First consider the need for contralateral masking
old is considered when making a decision about the need assuming that air-conduction thresholds were measured
for masking. Remember that an unmasked bone-conduc- using supra-aural earphones. A conservative estimate of IA
tion threshold does not convey ear-specic information. It is 40 dB. We will use the following equation when making a
is assumed that the bone-conduction response can originate decision about the need for contralateral masking:
from either or both ears. Therefore, the unmasked bone-
conduction response is considered the apparent or possible AC Test Ear BC Nontest Ear IA
threshold for either ear.
Consider the unmasked puretone audiogram* pre- Because it is not possible to measure bone-conduction
sented in Figure 6.4. Because IA for bone-conducted sound threshold at 8,000 Hz, it is necessary to predict an unmasked
is considered 0 dB, unmasked bone-conduction thresholds threshold given the ndings at other test frequencies. In this
are traditionally obtained at only one mastoid process. particular example, unmasked bone-conduction threshold
During air-conduction threshold testing, the potential for at 8,000 Hz will probably have a similar relationship with
cross hearing is greatest when there is a substantial differ- the air-conduction thresholds in the better (i.e., left) ear.
ence in hearing sensitivity between the two ears and when a Because there is no evidence of air-bone gaps at the adja-
stimulus is presented at higher HLs to the poorer ear. Con- cent high frequencies, we will assume that a similar relation-
sequently, there is greater potential for cross hearing when ship exists at 8,000 Hz. Therefore, our estimate of unmasked
measuring puretone thresholds in the right ear. bone-conduction threshold is 45 dB HL.
It will be necessary to remeasure puretone thresholds at
*The puretone audiogram and audiometric symbols used throughout all test frequencies in the right ear using contralateral mask-
this chapter are those recommended in ASHAs (1990) most recent ing because the difference between ears equals or exceeds
guidelines for audiometric symbols (see Chapter 3). our estimate of IA.
84 SECTION I Basic Tests and Procedures

Right Ear Masking unmasked bone-conduction thresholds before obtaining


(Test Ear) Needed? unmasked air-conduction thresholds. Decisions about the
need for masking during air-conduction testing then can be
250 Hz 550 40? Yes
500 Hz 605 40? Yes
made using the important bone-conduction responses.
1,000 Hz 8025 40? Yes 3A insert earphones are often substituted for the supra-
2,000 Hz 9030 40? Yes aural conguration during audiometric testing. We now
4,000 Hz 9540 40? Yes will take a second look at the audiogram in Figure 6.4 and
8,000 Hz 10045 40? Yes assume that air-conduction thresholds were obtained with
3A insert earphones. Recall that conservative estimates of IA
However, contralateral masking is not required when for 3A insert earphones with deeply inserted foam eartips
testing the left ear. The difference between ears does not are 75 dB at 1,000 Hz and 50 dB at frequencies >1,000 Hz.
equal or exceed the estimate of IA. Previously, we determined that contralateral masking was
not required when testing the better (i.e., left) ear using
Left Ear Masking supra-aural earphones. Given the greater IA offered by 3A
(Test Ear) Needed? insert earphones, it is easy to understand that contralateral
masking again should not be required when testing the left
250 Hz 200 40? No ear. However, a different picture results when considering
500 Hz 255 40? No the need for contralateral masking when testing the right ear.
1,000 Hz 3025 40? No
2,000 Hz 3530 40? No Right Ear Masking
4,000 Hz 4040 40? No (Test Ear) Needed?
8,000 Hz 4545 40? No
250 Hz 550 75? No
500 Hz 605 75? No
Many audiologists will obtain air-conduction thresh-
1,000 Hz 8025 75? No
olds prior to measurement of bone-conduction thresholds.
2,000 Hz 9030 50? Yes
A preliminary decision about the need for contralateral
4,000 Hz 9540 50? Yes
masking can be made by comparing the air-conduction 8,000 Hz 10045 50? Yes
thresholds of the two ears.
Contralateral masking is required during puretone air- Because of the greater IA provided by 3A insert ear-
conduction audiometry when the unmasked air-conduction phones in the lower frequencies, the need for contralateral
threshold in the test ear (ACTest Ear) equals or exceeds the air- masking is eliminated at 250, 500, and 1,000 Hz. It should
conduction threshold in the nontest ear (ACNontest Ear) by a con- be apparent that the process of evaluating the need for con-
servative estimate of IA: tralateral masking when using either supra-aural or insert
AC Test Ear AC Nontest Ear IA earphones is the same. The only difference is the substitu-
tion of different values of IA in our equations.
It is important to remember, however, that cross hearing for
air-conducted sound occurs primarily through the mecha- Puretone Audiometry:
nism of bone conduction. Consequently, it will be necessary
to re-evaluate the need for contralateral masking during air-
Bone Conduction
conduction testing following the measurement of unmasked Remember that a conservative estimate of IA for bone-
bone-conduction thresholds. conducted sound is 0 dB. Theoretically, masked bone-
Consider again the audiogram presented in Figure 6.4. conduction measurements are always required if ear-specic
Let us assume that we have not yet measured unmasked bone- information is needed. However, given the goal of bone-
conduction thresholds. We can make a preliminary decision conduction audiometry, contralateral masking is not always
about the need for contralateral masking by considering the required. Generally, bone-conduction thresholds are pri-
difference between air-conduction thresholds in the two ears. marily useful for determining gross site of lesion (i.e., con-
Based on the air-conduction responses only, it appears that ductive, sensory/neural, or mixed). The presence of air-bone
contralateral masking is needed only when testing the right gaps suggests a conductive component to a hearing loss.
ear at octave frequencies from 1,000 through 8,000 Hz. Yet, The major factor to consider when making a decision
once unmasked bone-conduction thresholds are measured, about the need for contralateral masking during bone-
it becomes apparent that contralateral masking will also be conduction audiometry is whether the unmasked bone-
required when testing the right ear at 250 and 500 Hz. conduction threshold (Unmasked BC) suggests the presence
It is conventional to obtain air-conduction thresholds of a signicant conductive component in the test ear.
prior to bone-conduction thresholds. However, an alter- The use of contralateral masking is indicated whenever
native (and recommended) approach involves obtaining the results of unmasked bone-conduction audiometry suggest
CHAPTER 6 Clinical Masking 85

the presence of an air-bone gap in the test ear (Air-Bone mal distribution of the relationship between air- and bone-
GapTest Ear) of 15 dB or greater: conduction thresholds in individuals without signicant
air-bone gaps, then an air-bone difference of 10 dB is not
Air-Bone Gap Test Ear 15 dB
unexpected.
where If unmasked bone-conduction thresholds suggest air-
bone gaps of 10 dB or less, then contralateral masking is not
Air-Bone Gap = AC Test Ear Unmasked BC required. Although unmasked bone-conduction thresholds
do not provide ear-specic information, we have accom-
ASHA (2005) recommends that contralateral masking plished our goal for bone-conduction testing. If unmasked
should be used whenever a potential air-bone gap of 10 dB bone-conduction thresholds suggest no evidence of sig-
or greater exists. When taking into account the variability nicant air-bone gaps, then we have ruled out the presence
inherent in bone-conduction measurements (Studebaker, of a signicant conductive component. Consequently, our
1967b), however, a criterion of 10 dB may be too stringent. assumption is that the hearing loss is sensory/neural in
There is a certain degree of variability between air- and nature.
bone-conduction threshold, even in individuals without Figure 6.5 provides three examples of the need for
conductive hearing loss. If we assume that there is a nor- contralateral masking during bone-conduction audiometry.

Frequency (Hz) Frequency (Hz)

Hearing level in decibles (dB HL)


Hearing level in decibles (dB HL)

A B
Frequency (Hz)
Hearing level in decibles (dB HL)

FIGURE 6.5 Audiograms illustrating the need for con-


tralateral masking during bone-conduction audiometry.
Example A: Masked bone-conduction thresholds are not
required in either ear. Example B: Masked bone-conduction
thresholds are required only in the right ear. Example C:
Masked bone-conduction thresholds are potentially required
C in both ears. See text for further discussion.
86 SECTION I Basic Tests and Procedures

Unmasked air- and bone-conduction thresholds are pro- for contralateral masking during speech audiometry: (1) IA,
vided in each case. (2) presentation level of the speech signal (in dB HL) in the
Example A. Contralateral masking is not required dur- test ear, and (3) bone-conduction hearing sensitivity (i.e.,
ing bone-conduction testing in either ear. When we com- threshold) in the nontest ear.
pare the unmasked bone-conduction thresholds to the Contralateral masking is indicated during speech audi-
air-conduction thresholds in each ear, there are no potential ometry whenever the presentation level of the speech signal
air-bone gaps of 15 dB or greater. For example, consider the (in dB HL) in the test ear (Presentation LevelTest Ear) equals or
thresholds at 2,000 Hz. Comparison of the unmasked bone- exceeds the best puretone bone-conduction threshold in the
conduction threshold to the air-conduction thresholds sug- nontest ear (Best BCNontest Ear) by a conservative estimate of IA:
gests a potential air-bone gap of 5 dB in the right ear and
0 dB in the left ear. Because the unmasked bone-conduction Presentation Level Test Ear Best BC Nontest Ear IA
threshold does not suggest the presence of signicant air-
bone gaps in either ear, our conclusion is that the hearing Because speech is a broadband signal, it is necessary to
loss is sensory/neural bilaterally. Obtaining masked bone- consider bone-conduction hearing sensitivity at more than
conduction thresholds, although they would provide ear- a single puretone frequency. Konkle and Berry (1983) and
specic information, would not provide additional diag- Sanders (1991) recommend the use of the bone-conduction
nostic information. puretone average of 500, 1,000, and 2,000 Hz or some other
Example B. Comparison of unmasked bone-conduc- average that is predictive of the SRT. ASHA (1988) recom-
tion thresholds to the air-conduction thresholds in the left mends that the puretone bone-conduction thresholds at
ear does not suggest the presence of signicant air-bone 500, 1,000, 2,000, and 4,000 Hz should be considered. Mar-
gaps. Consequently, masked bone-conduction thresholds tin and Blythe (1977) suggest that 250 Hz can be eliminated
are not required in the left ear. Our conclusion is that the from any formula for determining the need for contralat-
hearing loss is sensory/neural. eral masking when measuring the SRT. Yet, the nontest ear
However, masked bone-conduction thresholds will be bone-conduction threshold at 250 Hz may be an important
required in the right ear. Comparison of unmasked bone- consideration when measuring the SDT. Olsen and Matkin
conduction thresholds to the air-conduction thresholds (1991) state that the SDT may be most closely related to the
in the right ear suggests potential air-bone gaps ranging best threshold in the 250 to 4,000 Hz range when audio-
from 25 to 35 dB. The unmasked bone-conduction thresh- metric conguration steeply rises or slopes. Therefore, fol-
olds may reect hearing in the better (i.e., left) ear. Bone- lowing the recommendation of Coles and Priede (1975), the
conduction thresholds in the right ear may be as good as the most conservative approach involves considering the best
unmasked responses. They also may be as poor as the air- bone-conduction threshold in the 250- to 4,000-Hz fre-
conduction thresholds in that ear. Because we do not have quency range.
ear-specic information for bone-conduction thresholds, The examples presented in Figures 6.6 and 6.7 illus-
the loss in the right ear can be either mixed or sensory/neu- trate the need for contralateral masking during threshold
ral. To make a denitive statement about the type of hearing and suprathreshold speech audiometry, respectively. First
loss, it will be necessary to obtain masked bone-conduction consider the audiogram presented in Figure 6.6. Audiom-
thresholds in the right ear. etry was performed using supra-aural earphones. Puretone
Example C. There is evidence that contralateral masking testing (using appropriate contralateral masking dur-
will be required when measuring bone-conduction thresh- ing both air- and bone-conduction audiometry) reveals a
olds in both ears. Comparison of unmasked bone-conduc- severe-to-profound, sensory/neural hearing loss of gradu-
tion thresholds to the air-conduction thresholds suggests ally sloping conguration in the right ear. There is a very
potential air-bone gaps ranging from 30 to 35 dB in each mild, sensory/neural hearing loss of relatively at cong-
ear. As in the previous example, bone-conduction thresh- uration in the left ear. Given the difference between ears
olds in each ear may be as good as the unmasked responses. observed during puretone audiometry, it is anticipated that
They may also be as poor as the air-conduction thresholds contralateral masking may be needed during assessment of
in that ear. To make a denitive statement about the type SRT in the poorer ear.
of hearing loss, it will be necessary to obtain masked bone- There are different approaches that can be used when
conduction thresholds in both ears. determining the need for contralateral masking during
measurement of SRT. The most efcient and recommended
approach involves predicting the speech threshold using
Speech Audiometry the puretone threshold data in the poorer ear and, on that
Because speech audiometry is an air-conduction procedure, basis, determining the need for contralateral masking. For
the rules for when to mask will be similar to those used dur- example, SRT is measured at 20 dB HL in the left ear, a nd-
ing puretone air-conduction audiometry. There are three ing consistent with the puretone results. Given the relatively
factors to consider when making a decision about the need low HL at which the SRT was established in the better (i.e.,

Você também pode gostar