GENESYS 30 Visible Spectrophotometer Lockout Specifications

Instrument
Must meet all of the technical and performance specifications outlined in the specifications table
Must guarantee a minimum stray light specification of <0.1% at 340nm and 400 nm
Must be a scanning monochromator-based instrument
Must be a standalone instrument with on-board control and analysis software
Must have a high-resolution color screen, at least 800 x 480 pixels
Must have a rubberized keypad with at least 1mm key travel and an audible click upon pressing each key
Sample compartment must be accessible from the front and top
Sample compartment liner and sample stage must be removable without the use of tools
Removable sample compartment liner must be washable
Sample compartment lid must be user-replaceable without removing the instrument cover
Must have a pre-aligned, user-replaceable lamp that can be changed without removing the instrument cover
Must have option to add a directly connected printer accessory without use of tools
Must include a factory warranty of at least two years

Local Control Software


Must include a Live Display mode for instantaneous, real-time single wavelength measurements of absorbance, transmittance and concentration
Must support single and multi-wavelength fixed measurements, wavelength scanning and quantitative analysis application modes
Quant mode must support calibration curves with up to six standards
Quant mode must offer linear through zero and linear with intercept fit options
Must include automatic setting of scan speed and integration time to yield optimal data quality
Must support result data tables with up to 10 entries
Must be able to save methods on the instrument
Must be able to save methods and data to a USB memory device
Must export data in a text file format that is readable by common text editing and spreadsheet programs
Must have SmartStart feature to allow the instrument to boot up to a specified methods list
Must offer the capability to upload a method which combines absorbance readings at up to six wavelengths with up to six factors and mathematical
operations including addition, subtraction, multiplication, division and raising the absorbance value to the 2nd and 3rd power. Such methods shall
be created offline using an application on a personal computer.

Accessories
Must support option for a pass through lid accessory to accommodate tubing for flow cells and temperature accessories
Test tube holder accessory must
- Have the capability to support measurement of test tube cuvettes from 12 to 25 mm OD
- Offer a tall test tube cover accessory to support making measurements with test-tubes up to 150mm (6 inches) tall with the sample
compartment lid open
- Be equipped with a depth gauge to aid users in ensuring that sufficient sample is in the tube to make a valid measurement
Must have the capability to support measurement of rectangular and cylindrical cells up to 100 mm pathlengths
Must have the capability to support measurement of glass or plastic filters up to 50 mm long x 80 mm tall x 5 mm thick
Must have the capability to support temperature control of one 10 mm pathlength square cuvette from 5 100 C using a water thermostatted
cuvette holder

System Software (Optional) for control from an attached personal computer


Must be compatible with Microsoft Windows 7, 8.1, and 10 (64-bit OS in all cases)
Must support single and multi-wavelength fixed, wavelength scanning, quantitative and Rate/Kinetics application modes
Must support enzyme kinetics calculations
Must offer performance verification tests with pre-programmed pass/fail limits specific to the instrument
GENESYS 30 Spectrophotometer Guaranteed SPECIFICATIONS
Specifications
Optical Design Single Beam
Spectral Bandwidth 5nm
Light source (Typical Service Life) Tungsten-Halogen (>1000 Hrs)
Detector Silicon Photodiode
Wavelength
Range 325nm 1100nm
Accuracy 2nm
Repeatability < 1nm
Scan speed Automatic up to 1200nm/min
Data resolution 1nm
Photometric
Range -3A to +3A
Display -3A to +3A, 0 to 200 000 %T, 0 to 9 999 999 C
Accuracy 0.002A (0 - 0.3A)
0.5% of ABS reading (0.301A - 2.5A)
Repeatability 1 0.002 A
Noise 2 0.001A at 0A
0.001A at 1A
0.002A at 2A
Drift 3 <0.002A/Hr
Stray Light 4 <0.1%T at 340nm and 400nm
Baseline Flatness 5 <0.003A
Display 5 diagonal, 32 bit color display, 800 x 480 pixels
Keypad Tactile rubber 23 keys with numeric keypad
Sample compartment Accessible from top, front or side
Able to accommodate cells up to 100mm pathlength
Available accessory for test-tubes up to 25mm diameter
and 150mm height
Removable, washable sample compartment liner with
magnetic placement and hold-down
Printer Thermal printer accessory with 80mm width paper
Connectivity USB-A on front face supports flash memory devices
for method and data storage
USB-B on side supports connection to a Windows
computer running optional remote control software.
Languages English, Spanish, German, French, Italian,
Portuguese, Russian,Chinese, Japanese, Korean,
Thai, Arabic
Dimensions 35.5cm x 38.5cm x 19.5cm (L x W x H)
Weight 7.5 kg
Power Requirements External AC to DC converter. Voltage and Hz selected
automatically, 100-240 volts, 50-60Hz.
Warranty 2 years standard with one year extension upon
registration of the instrument with Thermo Fisher
Scientific within 6 months of purchase
1
Measured at 1.0 A @546nm
2
RMS at 500nm. 60 consecutive measurements
3
At 500nm after warmup.
4
Measured using glass cutoff filters or NaNO2 at 340nm
5
325nm 1000nm smoothed

www.thermofisher.com
2016 Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. Specifications,
terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Africa +27 11 822 4120 Denmark +45 70 23 62 60 India +91 22 6742 9434 New Zealand +64 9 980 6700
Australia +61 3 9757 4300 Europe-Other +43 1 333 50 34 0 Italy +39 02 950 591 Russia/CIS +43 1 333 50 34 0
Austria +43 1 333 50 34 0 Finland/Norway/Sweden Japan +81 45 453 9100 Spain +34 914 845 965
Belgium +32 53 73 42 41 +46 8 556 468 00 Latin America +1 561 688 8700 Switzerland +41 61 716 77 00
Canada +1 800 530 8447 France +33 1 60 92 48 00 Middle East +43 1 333 50 34 0 UK +44 1442 233555
China +86 10 8419 3588 Germany +49 6103 408 1014 Netherlands +31 76 579 55 55 USA +1 800 532 4752

Você também pode gostar