TECHNICAL SERVICE REPORT

PRODUCT NEWS
SUBJECT: GUIDE LINES SERVICE SOFTWARE V 3.1.0

This is a summary of the main new features included in 2. Incorporation of the help menu
the service software version 3.1.0. Please, carefully read
the User Manual to obtain more detailed information. We The help menu has been incorporated into the main bar.
explain how to use the new software hereinbelow. In order From this menu, you can select 2 options:
to visualize the improvements, the analyzer needs to
First: a link to the preventive maintenance document,
have loaded the new firmware version v 3.0 which
which is saved in pdf format. This document shows
comes along with the CD-ROM.
the analyzers preventive maintenance to be carried
In this version, we have improved the sensitivity adjustment
out periodically in order to get a proper operation of the
and a new menu has been included, the help menu.
analyzer.
Second: the about option, this screen shows
1. Sensitivity adjustment per rack information on the service program version and on the
type. firmware version loaded in the analyzer.

With this version and the firmware version 3.0, the


sensitivity adjustment has become independent according
to the rack type. Now the sensitivity of each type of rack
will be able to be adjusted separately.
In order to make the adjustment, first of all you have to
select a rack and place on it the following configuration of
racks and samples:
Paediatric Rack: place 4 paediatric glasses in positions
5, 6, 17 and 18 with 400 uL of distilled water.
Rack of 13/15mm sample: place 4 tubes in positions
5, 6, 17 and 18 with 400 uL of distilled water.
Rack of reagents: place 2 bottles of 20mL in positions
5 and 6 with 2mL of distilled water.
Press the adjust button and follow the process shown by
the program.
Repeat this adjustment per each rack type.
Help menu screen
The sensitivity value can also be introduced manually for
each rack in the corresponding box.

Sensibility adjustment screen