SITRANS F M MAGFLO 9.

Service

9.3
Check list MAG sensor ATTENTION!
If there is leakage from MAG 1100/3100/3100 W or MAG 5100 W and the unit has been used
to measure inflammable/explosive liquids, there might be a risk of explosion when checking
with a megger.

Disconnect all leads to MAG 1100/3100/3100 W or MAG 5100 W


MAG 1100/3100/3100 W or MAG 5100 W installed and filled with the medium:

Check coil resistance with a multimeter NO


R 85-86 Nominal value: See table

YES

NO
Check coil insulation with a megger (500 V) Sensor may be defective.
R 85 - Gnd Nominal value: Infinite resistance Contact Siemens Flow
Instruments
YES

> 500 kohm


Check electrode circuit with a moving-coil instru-
ment without amplifier (measurement unstable)
R 82 - Gnd Nominal value: 5-500 kohm
R 83 - Gnd Nominal value: 5-500 kohm Dirty electrodes or
conductivity too low
< 500 kohm

Sensor apparently OK.


Remove sensor, dry it and check

MAG 1100/3100/3100 W or MAG 5100 W removed from system empty and dry:

Check coil resistance with a multimeter NO


R 85-86 Nominal value: See table

YES

Check coil insulation with a megger (500 V) NO


R 85 - Gnd Nominal value: Infinite resistance
YES

Check electrode insulation with megger (500 V)


R 82 - Gnd Nominal value: Infinite resistance NO
R 83 - Gnd Nominal value: Infinite resistance

YES Commissioning

Check continuity between electrode and electrode


terminal NO
R 82 - Electrode Nominal value: 0 kohm
R 83 - Electrode Nominal value: 0 kohm

YES

MAG 1100/3100/3100 W or MAG 5100 W OK Sensor may be defective.


Contact Siemens Flow
Instruments

SFIDK.PS.027.W4.02 77