Você está na página 1de 4

Proceedings of the 29th Annual International

Conference of the IEEE EMBS


FrD06.2
Cit Internationale, Lyon, France
August 23-26, 2007.

Wearable Seismocardiography
Paolo Castiglioni, Andrea Faini, Gianfranco Parati, and Marco Di Rienzo, Member, IEEE

Abstract Seismocardiogram (SCG) is the recording of body contraction of the heart muscle ejecting blood into the
vibrations induced by the heart beat. SCG contains vessels. These forces are responsible for body accelerations
information on cardiac mechanics, in particular heart sounds in the order of 5 mg. Since most studies on
and cardiac output. In this paper we present a new wearable ballistocardiogram (BCG) were aimed at measuring these
device for SCG recordings during long term monitorings, and
movements and their correlation with the cardiac output, in
the results of a validation test in 4 subjects. The system is based
on the integration of the MagIC smart shirt (i.e., a textile-based this paper BCG will refer to the SCG low-frequency
wearable system for the assessment of ECG and respiratory component reflecting recoil movements, while SCG will
movements), and an external triaxial MEMS accelerometer indicate the whole spectrum of cardiac vibrations.
positioned on the left clavicle. SCG was estimated as the Despite initial enthusiasms for its possible clinical
average of accelerations occurred in each heart beat. The SCG applications, however, the SCG technique has remained a
components due to the valve closure and to recoil forces
tool for physiologists, failing to be widely used in clinical
following the heart contraction (ballistocardiogram) were
extracted by high-pass (>18 Hz) and band-pass (0.6-20 Hz) practice. Possibly this is due to the need of using relatively
filters respectively. Then the difference between the I and J complex recording devices, which restricted its applicability
waves of the ballistocardiogram ( |I-J| index, possibly related to to laboratory tests, where similar clinical information can be
the cardiac output) was identified by an ad-hoc procedure and provided by other systems (echocardiographers and
compared with the model flow indirect estimation of cardiac phonocardiographers). In recent years, however, MEMS
output.
technology has provided small and cheap accelerometers
Validation on 4 volunteers showed that: 1) our wearable
system provides statistically consistent estimates of both heart-
that may became part of a new class of wearable systems for
sound related vibrations and recoil movements; 2) reliable long term SCG monitoring. Although such a class of
estimates of the |I-J| index can be obtained by considering systems will be inevitably more prone to noise and artefacts
about 1 minute of SCG recording in stationary conditions; and than laboratory devices, they might allow the assessment of
3) changes of the |I-J| index during exercise correlate well with SCG in a wider range of conditions, including sleep and
changes of cardiac output estimated by the model flow. daily life activities.
In this paper we describe a new wearable system we have
I. INTRODUCTION
recently developed for SCG assessment out of the laboratory

S EISMOCARDIOGRAPHY (SCG) is the study of body


vibrations induced by the heart beat. This term was
popularized by the works of Salerno and Zanetti at the
setting, through MEMS accelerometers. We also illustrate
the results obtained from the evaluation of the capability of
the system to monitor changes in cardiac output.
beginning of the 90s [1-3], but techniques for measuring
body movements associated to the cardiac activity are much
older. According to the measured quantity (displacement,
velocity or acceleration) and sensors position, this method
has been called apex-cardiography [4], kineto-cardiography
[5], cardio-kymocardiography [6] or ballistocardiography
[7], this latter being probably the more widely used term.
Vibrations induced by the cardiac activity are mainly
correlated to valve closure (heart sounds) and cardiac
systole. In particular, SCG movements at relatively lower
frequencies reflect the recoil forces that results from the

Manuscript received March 30, 2007.


P. Castiglioni is with Polo Tecnologico (Biomedical Technology
Department), IRCCS S.Maria Nascente, Fondazione Don Gnocchi
ONLUS, Milan, Italy (fax: +39-024048919; e-mail: pcastiglioni@
dongnocchi.it).
A. Faini is with DIMET, Universit degli Studi di Milano Bicocca, Italy
G. Parati is with Istituto Auxologico Italiano, Milan, Italy
M. Di Rienzo is with Polo Tecnologico (Biomedical Technology Figure. 1. Scheme of the wearable SCG system: A= 3D accelerometer;
Department), IRCCS S.Maria Nascente, Fondazione Don Gnocchi B=textile ECG electrodes; C=piezoelectric textile sensor for thoracic
ONLUS, Milan, Italy. movements; D= electronic board for signals storing and transmission;
BD= textile paths of conductive fibers; AD= external cable.

1-4244-0788-5/07/$20.00 2007 IEEE 3954


II. SYSTEM DESCRIPTION the low signal-to-noise ratio, filtered accelerations were split
into individual beats, synchronised with the R-peak, and
A. Sensors and Recording Device averaged. Figure 2 shows the averaged vertical accelerations
Our SCG system is based on the integration of a textile- after band-pass and high-pass filtering in a representative
based smart garment previously developed (the MagIC subject. Typical BCG waves [9] are clearly visible in the
system) and an external MEMS 3D accelerometer. In short, band-pass filtered SCG, while vibrations synchronous to
the MagIC system is composed of a vest of cotton and lycra heart sounds appear in the high-pass filtered SCG .
with embedded textile ECG electrodes, an embedded textile
piezoresistive transducer for respiratory movements and a
C. Assessment of I-J amplitude from BCG
portable electronic module. Electrical connections are made
of textile paths of conductive fibers. The electronic module The BCG feature more correlated to the cardiac
digitizes signals at 150 Hz, locally stores them on a memory hemodynamics is the difference between the I and J peaks.
card and, possibly, transmit data via bluetooth connection to This feature can be correctly quantified on the acceleration
a remote computer or PDA. Details on the system have been component oriented as the mechanical axis of the heart. The
provided elsewhere [8]. orientation of the cardiac axis may change with changes of
To the above smart garment we added an external triaxial posture [9]. This makes it difficult to identify an
MEMS accelerometer (LIS3L02AL, STMicroelectronics, accelerometer axis (even after rotations of the reference
size= 4.85 x 4.85 x 1.46mm; weight=0.08 grams; full scale= frame) for calculating the I-J amplitude in a wearable
2g, resolution= 0.5mg). The accelerometer was stuck on system. For this reason, we computed the I-J amplitude
the left clavicle. We selected this site because vibrations independently from the reference frame as follows. First, we
related to heart sounds are transmitted with low damping, plot the band-pass filtered acceleration components, after the
and pulse movements of arterial walls do not affect R-peak synchronized average, in the X-Y-Z frame of the tri-
importantly recoil movements. The accelerometer was axial accelerometer, obtaining the 3D trajectory of the
connected to the MagIC electronic module through an acceleration vector. Then we estimated the modulus of the
external cable (fig.1). difference between the acceleration vectors in I and J, |I-J|,
as the maximum distance between any couple of points of
the acceleration trajectory.
B. Assessment of SCG components Only the part of the trajectory within 50 ms and 250 ms
Each component (X, Y and Z) of the acceleration vector from the R peak was considered: in fact this is the time
was filtered twice to separate the BCG signal (band-pass period where I and J peaks are expected to occur. The
filter, from 0.6 to 20 Hz) from heart-sounds related distance calculated in this way does not dependent on
vibrations (high-pass filter, from 18 Hz). The R-peak of rotations of the reference frame (fig. 3).
ECG was identified through a threshold algorithm. Due to

Figure 3. Trajectory of the acceleration vectors obtained by plotting


the X, Y and Z components of BCG, for 450 ms after the R peak; the
Figure 2. Upper panels: average ECG wave over 600 heart beats in a thick line falls in the time window between 50 ms and 250 ms, i.e.,
subject sitting at rest for 10; lower panels: average vertical where I and J waves are expected to occur; the modulus of the I-J
acceleration after band-pass (left) and high-band (right) filtering; thin distance is estimated as the maximum distance between two points of
lines are the 95% confidence intervals of the mean. The band-pass the curve
filtered SCG shows typical BCG waves, while the high-pass filtered
SCG shows vibrations synchronous with heart sounds. 3955
III. SYSTEM VALIDATION
First, the system has been validated by assessing the
statistical consistency of the estimates of low and high
frequency components of SCG, i.e., BCG and heart-sounds
vibrations, during a 10 minutes recording in a subject sitting
at rest.
Then we focused on the validation of the |I-J| index. This
was done by evaluating the minimum number of heart beats
required for the calculation of |I-J|; and by quantifying the
relation between changes of cardiac output and changes of
|I-J|. For this validation, we recorded the SCG in four
healthy volunteers during two conditions of 5 minutes
duration each: sitting at rest (REST), and while performing
physical exercise at 25W on cycloergometer (EXE). We
derived an indirect measure of cardiac output, to be
compared with the |I-J| index, by simultaneously measuring
the finger arterial pressure by the Finometer device
(Finapres Medical Systems, Amsterdam, the Netherlands) in Figure 5. |I-J| estimated in REST condition in a typical subject, as a
function of the employed number of heart beats N. The dashed line is
both REST and EXE conditions. It was possible to indirectly the value calculated from the whole 5 minutes recording (N=317).
calculate changes of cardiac output by analyzing the pulse
contour of the blood pressure wave with the model-flow derived from the whole REST recording with values
method [10]. obtained by progressively increasing the number of heart
beats N used for the estimate, starting from N=10 beats. A
A. SCG components: BCG and high frequency vibrations
typical example is shown in fig.5: after an initial transient,
We verified the statistical consistency of the average the estimate quickly converges to the final value. In practice,
synchronized with the R-peak by computing again the band- the |I-J| estimate reaches a sufficiently stable value when N
pass and high-pass filtered components of SCG previously is greater than 60 heart beats, i.e., after about 1 minute.
shown in fig.2, but this time by means of a randomly Similar results were obtained in all the four subjects.
synchronized average. This was done by synchronizing the We also evaluated whether the increase of cardiac output
average of the acceleration components with the series of R- which is expected to occur from REST to EXE, produces
peaks after having added a random delay to the ECG. The comparable changes of the |I-J| index. For this aim, firstly
added delay removed the beat-by-beat synchronization we expressed the increase of cardiac output from REST to
between ECG and SCG. Results are shown in fig. 4. The EXE in percentage; then we similarly evaluated the percent
absence of the SCG features observed previously (fig.2) increase of the |I-J| index. Finally we computed the linear
indicates that they are not due to random fluctuations or to correlation between the two percent increases among the
methodological artifacts. four subjects.
Individual values of percent increases of cardiac output
and |I-J| index are shown in Table 1. Changes of the two
quantities are similar in all the subjects. The regression line
between cardiac output percent changes (CO) and |I-J|
percent changes is given by:
CO = 0.61 I J + 16.5 (1)
with correlation coefficient r2=0.89.

TABLE I
CHANGES OF CARDIAC OUTPUT, (CO), FROM REST TO EXE AS
ESTIMATED BY THE MODEL FLOW METHOD, AND CORRESPONDING
Figure 4. Average vertical acceleration after band-pass (left) and high- CHANGES OF |I-J|, IN PERCENTAGE
band (right) filtering of the same subject of fig.2, but in this case the
average was de-synchronized from the R-peaks (see text); thin lines Subject CO Model Flow |I-J|
show the 95% confidence intervals of the mean.
1 28.8% 26.5%
2 46.5% 43.3%
B. |I-J| index 3 55.5% 68.0%
To evaluate the minimum number of heart beats needed 4 43.6% 40.7%
for a valid estimate of the |I-J| index, we compared the value

3956
IV. CONCLUSION [4] A. Silvestre, G. Sandhu, K. B. Desser, and A. Benchimol, "Slow
filling period/rapid filling period ratio in the apexcardiogram:
This study showed the feasibility of monitoring the SCG relation to the diagnosis of coronary artery disease," Am. J Cardiol.,
by a wearable system based on textile technology for ECG vol. 42, no. 3, pp. 377-382, Sept.1978.
recordings plus an external MEMS triaxial accelerometer.
[5] A. Grandi, F. Barzizza, L. Bernardi, A. Venco, and G. Finardi,
The system we propose does not require the accelerometer "Kinetocardiographic detection of ventricular dyssynergy after
axes to be oriented precisely with respect to the cardiac myocardial infarction. Correlations with two-dimensional
mechanical axis, simplifying the subjects instrumentation. echocardiography," Acta Cardiol., vol. 39, no. 1, pp. 19-27, 1984.
Moreover we showed that it can provide reliable estimates
[6] R. A. Silverberg, G. A. Diamond, R. Vas, D. Tzivoni, H. J. Swan,
of the SCG components, in particular the main waves of
and J. S. Forrester, "Noninvasive diagnosis of coronary artery
BCG, just from one minute of recording during a steady- disease: the cardiokymographic stress test," Circulation, vol. 61,
state condition. Since we also showed that it is possible to no. 3, pp. 579-589, Mar.1980.
derive an index related to the changes of cardiac output from
the I and J waves, our system could potentially be used to [7] I. Starr, J. Rawson, H. A. Schroeder, and N. R. Joseph, "Studies on
the estimation of cardiac ouptut in man, and of abnormalities in
assess long term changes of cardiac hemodynamics, for cardiac function, from the heart's recoil and the blood's impacts; the
instance in sleep studies, in subjects during daily-life ballistocardiogram," Am. J. Physiol, vol. 127, pp. 1-28, July1939.
activities, or in cardiac patients under therapy or following
rehabilitation. [8] M. Di Rienzo, F. Rizzo, P. Meriggi, B. Bordoni, G. Brambilla, M.
Ferratini, and P. Castiglioni, " Applications of a Textile-Based
Wearable System for Vital Signs Monitoring,", 28th Annual
REFERENCES International Conference of the IEEE ed Engineering in Medicine
and Biology Society, 2006, pp. 2223-2226.
[1] D. M. Salerno and J. Zanetti, "Seismocardiography: A New
Technique for Recording Cardiac Vibrations. Concept, Method, [9] R. W. Soames and J. Atha, "Three-dimensional
and Initial Observations," J Cardiovasc Technol, vol. 9, pp. 111- ballistocardiographic responses to changes of posture," Clin. Phys.
118, 1990. Physiol Meas., vol. 3, no. 3, pp. 169-177, Aug.1982.

[2] D. M. Salerno and J. Zanetti, "Seismocardiography for monitoring [10] K. H. Wesseling, J. R. Jansen, J. J. Settels, and J. J. Schreuder,
changes in left ventricular function during ischemia," Chest, vol. "Computation of aortic flow from pressure in humans using a
100, no. 4, pp. 991-993, Oct.1991. nonlinear, three-element model," J Appl. Physiol, vol. 74, no. 5, pp.
2566-2573, May1993.
[3] J. Zanetti, M. O. Poliac, and R. S. Crow, "Seismocardiography:
waveform identification and noise analysis," 1991, pp. 49-52.

3957

Você também pode gostar