Você está na página 1de 139

Medical Imaging modalities

X-Ray Radiography

Gizeaddis L.
Jimma University
Jimma Institute of Technology
Biomedical Engineering Department
EM waves for medical Imaging
X-rays
are invisible

penetrate matter

can be diffracted

ionize gases

Change a photo emulsion

create light emission in different substance

induce biological changes in living tissue


X-ray Production
X-rays are produced when highly energetic electrons interact
with matter and convert their kinetic energy into
electromagnetic radiation.
A device that accomplishes such a task consists of
an electron source,
an external energy source to accelerate the electrons
an evacuated path for electron acceleration,
a target electrode,
X-ray Production
Requirements:
a source of fast moving electrons
must be a sudden stop of the electrons motion
in stopping the electron motion, kinetic energy (KE) is
converted to EMS energies
Infrared (heat),
light
x-ray energies
X-ray Production
Positive voltage (kVp) is applied to ANODE

Negative electrons from CATHODE = attracted across the


tube to the positive ANODE.

Electrons slam into anode suddenly stopped.

X-RAY PHOTONS ARE CREATED


X-ray Production
Electron beam is focused from the cathode to the anode target by
the focusing cup
Electrons interact with the electrons on the tungsten atoms of
target material
PHOTONS sent through the window PORT
Tube Interactions
possible tube interactions
Heat
Most kinetic energy of projectile e- is converted into heat
Projectile e- interact with the outer-shell e- of the target atoms
but do not transfer enough energy to the outer-shell e- to
ionize.

x-rays
X-rays = Characteristic (20%) or Bremsstrahlung (80%)
Heat is an excitation
rather than an ionization
Characteristic Radiation
The incident electron interacts with the K-shell electron via a
repulsive electrical force.

The K-shell electron is removed leaving a vacancy in the K-shell.

An electron from the adjacent L-shell (or possibly a different shell)


fills the vacancy.

A characteristic x-ray photon is emitted with an energy equal to


the difference between the binding energies of the two shells.
Characteristic Radiation
It is called characteristic
because it is characteristic
of the target element in the
energy of the photon
Produced
characteristic x-rays require
a tube potential of at least
70 kVp
Characteristic Radiation

Only K-
characteristic
x-rays of
tungsten are
useful for
imaging
Characteristic Radiation
has discrete
energies
based on the
e- binding
energies of
tungsten
Bremsstrahlung Radiation
Bremsstrahlung radiation arises from energetic electron
interactions with an atomic nucleus of the target material.

In a "close" approach, the positive nucleus attracts the


negative electron, causing deceleration and redirection,
resulting in a loss of kinetic energy that is converted to an x-
ray.

The x-ray energy depends on the interaction distance


between the electron and the nucleus; it decreases as the
distance increases.
Bremsstrahlung Radiation
Brems can be
produced at any
projectile e- value
Major factors that
affect x-ray
production
efficiency are
the atomic number
of the target
material and
the kinetic energy
of the incident
electrons
Bremsstrahlung Radiation
Brems x-rays have a range of energies and form a continuous
emission spectrum
x-ray tube
The x-ray tube provides an environment for x-ray production
via bremsstrahlung and characteristic radiation mechanisms.
Major components are
the cathode,
anode,
rotor/stator,
glass (or metal) envelope,
and tube housing
x-ray tube
Cathode
Cathode is a helical filament of
tungsten wire surrounded by a
focusing cup
This structure is electrically
connected to the filament circuit.
Electrical resistance heats the
filament and releases electrons
via a process called thermionic
emission.
Anode
The anode is a metal target electrode that is maintained at a
positive potential difference relative to the cathode.
Electrons striking the anode deposit the most of their energy
as heat, with a small fraction emitted as x-rays.
Tungsten is the most widely used anode material because of its
high melting point and high atomic number.
The high atomic number of tungsten provides better
bremsstrahlung production efficiency compared with low-Z
elements
Rotating Anode X-ray Tube
Anode angle and focal spot size
Heel effect (angle dependent attenuation )

It is a loss of
intensity on the
anode side of the x-
ray field of view.
It is caused by
attenuation of the x-
ray beam by the
anode.
Off-focus radiation
X-ray tube collimator assembly
Interaction of X-ray with matter
Several modes of interaction of x-rays with matter:
Scattering

Thompson or Rayleigh scattering,

Compton scattering)

Photoelectric absorption (transfer its energy to atoms


of the target material)

Pair production
Rayleigh (Coherent) scattering
It is the collision of a photon with an electron such that the
photon is deflected into a new directionelastic scattering

Energy of incident photon same as the scattered photon


Compton scattering
Compton scattering
The energy of the scattered photon is

The wavelength of the scattered photon increases over the


incident photon, and thus its energy decreases.

Compton scattering is the major source of tissue damage due


to X-rays. For these reasons, this phenomenon is very
undesirable
Photoelectric effect
All of the incident photon energy is transferred to an
electron, and eject the electron from the atom.
Photoelectric effect
If this electron has a binding energy Eb, then this is the
amount of energy required to remove the electron from its
shell.

If E>Eb, then the remainder of the energy ends up as kinetic


energy.

The moving electron is the photoelectron.


Photoelectric effect
Pair production
Occurs when a photon of high energy (>1.02MeV) interacts
with a nucleus.
Different modes of interaction
At low photon energies (<26 keV), the photoelectric effect

dominates the attenuation processes in soft tissue.

When higher energy photons interact with low Z materials (e.g.,

soft tissue), Compton scattering dominates.

Rayleigh scattering occurs with low probability, 10% of the

interactions in mammography and 5% in chest radiography.

Only at very high photon energies (>1.02 MeV), pair production

contribute to attenuation.
Different modes of interaction
Different modes of interaction
Attenuation of photons in matter
Linear attenuation coefficient

The linear attenuation coefficient is the sum of the individual


linear attenuation coefficients for each type of interaction:

linear attenuation coefficient is proportional to the density of


the material.
For instance
Attenuation of photons in matter
Mass attenuation coefficient

For a given thickness, the probability of interaction is


dependent on the number of atoms per volume.

This dependency can be overcome by normalizing the linear


attenuation coefficient for the density of the material, called
the mass attenuation coefficient:
Half value layer

The thickness of material required to reduce the intensity


of an x- or gamma-ray beam to one-half of its initial
value.
Mean free path

One cannot predict the range of a single photon in matter.

However, the average distance a photon traveled before


interaction is called the mean free path (MFP) of the photon
beam,
Example 1
A narrow beam containing 2000 monoenergetic photons is
reduced to 1000 photons by a slab of copper 0.01m thick.
What is the total linear attenuation coefficient of the copper slab for
these photons?

What is the HVL?


Solution

The HVL of a monoenergetic beam of x


or -rays in any medium is
Example 2
A 2mm thickness of material transmits 25% monoenergetic
beam of photons, calculate the HVL of the beam and mean
free path of a photon.
Planer Radiography/Projection
imaging
the acquisition of a
2D image of the
patient's 3D
anatomy
Instrumentation of Planer
Radiography
Basic components of a planar X-ray radiography
system are:
X-ray tube
a collimator
reduce the patient dose and amount of Compton scattered X-rays,

an anti-scatter grid
reduce further the contribution of scattered X-rays to the image,

a detector
converts the energy of the transmitted X-rays into light
X-ray Detectors

Traditional x-ray film


Screen-film radiography

Digital detectors
Computed radiography

Digital radiography
Screen-film cassette

The screen-film detector system used for general


radiography consists of
a cassette,

one or two intensifying screens,

a sheet of film.

The film is a sheet of thin plastic with a photosensitive


emulsion coated onto one or both sides.
A typical screen film cassette and its cross-section
Intensifying Screens
Screens are made of a scintillating material, which is called a

phosphor.

When x-rays interact in the phosphor, visible or ultraviolet (UV)

light is emitted.

a radiographic transducer - converts x-ray energy into light.

It is the light given off by the screens that principally causes the

film to be darkened.
Principles Intensifying Screens
Radiographic film processing
When the film is processed, it is exposed to several
different chemicals solutions for controlled periods of time.
Development - The developing agent gives up electrons to
convert the silver halide grains to metallic silver.
Stopping the development - The stop bath simply stops
the development process by diluting and washing the
developer away with water.
The Latent Image
The silver halide grains (emulsion), fixed in a thin gelatin
film, oxidize when exposed to photons.

The exposed film precipitates the fine silver particles, making


them opaque to visible light.

Careful attention to exposure time and development produces


a degree of blackening across the film which varies with the
original variation in photon flux and thus produces a 2D
latent image.
X-ray film
X- film processing
Fixing - Unexposed silver halide
crystals are removed by the fixing
bath.
The fixer dissolves only silver halide
crystals, leaving the silver metal behind.

Washing - The film is washed with


water to remove all the processing
chemicals.
Drying - The film is dried for viewing
Digital Detectors
Computed Radiography
CR differs from analog in that the CR cassette contains a
phosphor plate instead of a sheet of film
Computed Radiography
When the x-ray energy is absorbed by the BaFBr:Eu phosphor, the
absorbed energy excites electrons associated with the europium
atoms,
The excited electrons become mobile, and some fraction of them
interact with a so called F-center.
The F-center traps these electrons in a higher-energy, metastable
state, where they can remain for days to weeks.
The number of trapped electrons per unit area of the imaging
plate is proportional to the intensity of x-rays incident at each
location during the exposure.
Imaging plate
CR Reader

The next step is


the cassette is
placed into reader
and processor unit.

The plate is
scanned with a
very small laser
beam.
CR Reader

The laser beam stimulates


the plate causing light to
be produced.

The light that is produced


is proportional to the x-ray
exposure to that specific
spot.
Reading Phase

The light is amplified and


converted to electrical
signals by photomultiplier
tubes (PMT) and digitized.

The digital data is stored as


a digital image.
CR Plate
Digital Radiography
Unlike CR, DR does not need a separate read-out device.
X-RAYS
DR
DETECTORS

ADC
11011101
X-RAYS
CR
CR PLATE

SCANNER

ADC
11011101
Digital radiography detectors

There are two types of digital radiography (DR) detectors,


Indirect flat panel detectors

Direct flat panel detectors.

Indirect-conversion is most commonly used,


X-ray energy is first converted into light by a CsI:TI scintillator, and
then the light is converted into a voltage using a two-dimensional
array of photodiodes.
Indirect flat panel detector
When an X-ray is absorbed in a CsI rod light will be produced.
The light is converted to an electrical signal by the photodiodes in the
thin film transistors (TFT) array and stored in capacitors which are
formed at the junction of the photodiodes.
This signal is then read out line-by-line in parallel using a multiplexer.
The signals are amplified and digitized using analogue-to-digital
(A/D) converter.
Indirect flat panel detector
Indirect flat panel detector
Direct flat panel detector

Eliminates the intermediate step of converting X-ray energy


into light,

Uses direct absorption of the X-ray photons to produce an


electrical signal.

Materials such as amorphous selenium (alloyed with arsenic)


have been used.

X-ray absorption is not as efficient as for CsI:Tl


Scattered radiation in Projection
Radiography
Scattered radiation in Projection
Radiography
Contrast
Contrast
Anti-scatter grid
Anti-scatter grid is composed of a
series of lead grip strips aligned with
the x-ray source.
They are used to reduce the amount
of scattered radiation reaching the
detector by utilizing geometry
Grid is placed between patient and
detector
Anti-scatter grid
Grid ratio: is the ratio of the
height to the width of the inters
paces (not the grid bars) in the
grid.

Grid ratios of 8: 1, 10: 1, and 12:


1 are most common in general
radiography, and a grid ratio of
5: 1 is typical in mammography.
Anti-scatter grid
The grid is essentially a one-dimensional collimator, and
increasing the grid ratio increases the degree of collimation.

Higher grid ratios provide better scatter cleanup, but they


also result in greater radiation doses to the patient.

A grid is quite effective at attenuating scatter that strikes the


grid at large angles (where 0 degrees is the angle normal to
the grid), but grids are less effective for smaller-angle scatter.
parameters that
describe anti-scatter grid properties
Example
Specialized X-ray imaging
techniques

1. Mammography
2. Fluoroscopy
3. Digital Subtraction
Angiography
1. Mammography
Mammography is a radiographic examination that is
specially designed for detecting breast pathology.

It uses a low-dose x-ray system to examine breasts.

A mammography exam, called a mammogram, is used to


aid in the early detection and diagnosis of breast diseases
in women.
Distinguishing features of mammography equipment
from other x-ray imaging

These features are due to:-

Cancer produce very small physical changes in the breast that


are difficult to visualize with conventional x-ray imaging.

Mammography examination requires the highest image quality


than other x-ray procedures.
This is because the breast consist soft tissues with relatively small
differences in density (or atomic number).
Attenuation of breast tissues

the attenuation differences


between normal tissue and
cancerous tissue is highest
at very low x-ray energies
(10 to 15 keV) and is poor
at higher energies (>35
keV).
A mammography system
A mammography system
Mammography X-ray tube design

Cathode
The mammographic x-ray tube is typically configured with
dual filaments in a focusing cup that produce small focal spot
sizes.

A small focal spot minimizes geometric blurring and


maintains spatial resolution necessary for microcalcification
detection.
Mammography X-ray Tube Anode

Most x-ray tubes use tungsten as the anode material,

But Mostly mammography equipment uses molybdenum


anodes with an atomic number (Z) of 42 or in some designs, a
dual material anode with an additional rhodium with an atomic
number (Z) of 45 .

These materials are used because they produce a characteristic


radiation spectrum that is close to optimum for breast imaging.
Tube assembly
The x-ray tubes are arranged such that the cathode side of the
tube is adjacent to the patients chest wall,
since the highest intensity of x-rays is available at the cathode side, and
the attenuation of x-rays by the patient is generally greater near the
chest wall of the image.
Tube Filtration

Most x-ray machines use aluminium or "aluminium equivalent" to


filter the x-ray beam to reduce unnecessary exposure to the patient,

Mammography uses filters that work on a different principle and


are used to enhance contrast sensitivity.

Molybdenum (same as in the anode) is the standard filter material.

Added tube filters of the same element as the target reduce the
low- and high energy x-rays in the x-ray spectrum and allow
transmission of the characteristic x-ray energies
Tube Filtration
Mammography system
Collimation

Fixed-size metal apertures or variable field size shutters


collimate the x-ray beam.

For most mammography examinations, the field size matches


the film cassette sizes (e.g., 18 X 24 cm or 24 X 30 cm).

The exposure switch is enabled only when the collimator is


present.

Many new mammography systems have automatic


collimation systems that sense the cassette size
The automatic exposure control (AEC)

It is also called a phototimer, employs a radiation sensor, an

amplifier, and a voltage comparator, to control the exposure.

Unlike most conventional x-ray machines, the AEC detector is

located underneath the cassette.

This sensor consists of a single ionization chamber or an array of

three or more semiconductor diodes.

The sensor measures the residual x-ray photon flux transmitted

from the patient.


The automatic exposure control (AEC)

During the exposure, x-ray interactions in the sensor

release electrons that are collected and charge a

capacitor.

When the voltage across the capacitor matches a preset

reference voltage in a comparator switch, the exposure is

terminated.
The automatic exposure control (AEC)
Compression
Breast compression is a necessary part of the mammography
examination. Benefits of compression include:
A more uniform breast thickness resulting in a better fit of the
exposure into the film latitude or dynamic range.

Reduced blurring from patient motion.

Reduced scattered radiation and improved contrast sensitivity.

Reduced radiation dose.

Firm compression reduces overlapping anatomy and decreases


tissue thickness of the breast
Compression
Compression is achieved with a compression paddle, a flat plate
attached to a pneumatic or mechanical assembly.

Suspicious areas often require "spot compression to eliminate


superimposed anatomy by further spreading the breast tissues over
a localized area.
Dense opacity with specular border in the cranial part of the right breast;
Cluster of irregular microcalcification suggesting a low differentiated
carcinoma. (Courtesy of Dr. Van Ongeval, Department. of Radiology.)
Mammography Detector Technology
The two types of digital mammography are:

1. 2D mammography
With 2D digital mammography, the radiologist is
viewing all of the complexities of breast tissue in a
one flat image.

Disadvantage
Sometimes breast tissue can overlap, giving the illusion
of normal breast tissue looking like an abnormal area
2. 3D mammography/ tomosynthesis
is a mammography system where the x-ray tube and
imaging plate move during the exposure.

creates a series of thin slices through the breast that allow


doctors to examine breast tissue detail one slice at a time to
help find breast cancer at its earliest stages.

allows radiologists to view the breast tissue in one


millimeter slices, so that they can provide a more confident
assessment.

finds cancers missed with conventional 2D mammography


2. Fluoroscopy
Fluoroscopy
It is an x-ray imaging procedure that allows real time
imaging of a patient with high temporal resolution.
Allows continuous viewing of a time varying x-ray image
and permits live visual evaluation of dynamic events.
Uses TV technology which provides 30 frames per second
imaging.
Allows acquisition of a real time digital sequence of images
(digital video) that can be played back as a movie loop.
Fluoroscopy
Fluoroscopy uses continuous X-ray imaging,
The image output of a fluoroscopic imaging system is a projection
radiographic image, but in a typical 10-minute fluoroscopic
procedure a total of 18,000 individual images are produced.
fluoroscopy is used
to monitor interventional surgery,
for the placement of catheters, guide-wires and pacemakers in cardiac
catheterization laboratories,
for dynamic studies of the GI tract and cardiovascular system using
contrast agents.
Fluoroscopy chain components

Image intensifier tube


Image intensifier (II)
The principal component of the imaging chain that
distinguishes fluoroscopy from radiography is the image
intensifier.

Image intensifiers are used to convert the x-ray spectrum to


light energy.

The fluoroscopy image intensifiers are several thousand more


sensitive than the screen-film cassette image intensifiers.
Image intensifier
There are four principal components of an II:
the photocathode (in the input
screen),
the three focusing electrodes
(G1. G2, and G3),
G2
the anode (part of the output G2
G1
window) .
Output phosophor
The input screen

The input screen of the II consists of four different layers:


Vacuum window keeps air out of II

Support

CsI needles

photocathode
The input screen

X-rays must pass through the vacuum window and support, before
striking the cesium iodide (Csl) input phosphor.

CsI forms in long crystalline needles that act like light pipes,
limiting the lateral spread of light and preserving spatial
resolution.
It absorb the x-rays and convert their energy into visible light.

Light strikes the photocathode causing electrons to be liberated


into the electronic lens system of the II.
Electron optics
The five-component ("pentode") electronic lens system of the II.
The G1, G2, and G3 electrodes

the input screen substrate (the cathode)

the anode near the output phosphor

under the influence of the ~25,000 to 35,000 V electric field, electrons are
accelerated and arrive at the anode with high velocity and considerable
kinetic energy.

The intermediate electrodes (G1, G2, and G3) shape the electric field,
focusing the electrons properly onto the output layer.

After penetrating the very thin anode, the energetic electrons strike the
output phosphor and cause a burst of light to be emitted.
The output phosphor
The electrons strike the output phosphor, causing emission of light.
The thick glass output window allows light to escape the top of II.
Light that is reflected in the output window is scavenged to reduce
glare by the addition of a light absorber around the circumference
of the output window.
Electron optics
Image intensifier video
The output phosphor

Each electron causes the emission of approximately 1,000


light photons from the output phosphor.
The image is much smaller at the output phosphor than it is
at the input phosphor, because the 23- to 35-cm diameter
input image is focused onto a circle with a 2.5-cm diameter.
The reduction in image diameter leads to amplification
(minification)

Minification gain of II =

Characteristics of II Performance

Parameters that are useful in specifying the capabilities of the


II,
Conversion factor

Brightness gain

These characteristics are useful in troubleshooting lIs when


they are not performing properly
Characteristics of II Performance
Conversion factor
Defined as a measure of the gain of an II
( / . )
=
(mR/see.)
100 to 200 for new II
Degrades over time, ultimately can lead to II replacement
The conversion factor is the proper quantity for expressing
image intensification.
Brightness gain (BG)

It is the product of the electronic and minification gains of the II.

The electronic gain of an II is roughly about 50, and the


minification gain changes depending on the size of the input
phosphor and the magnification mode.
= ( )

As the effective diameter of the input phosphor decreases


(increasing magnification), the brightness gain decreases.
Automatic Brightness Control

The purpose of ABC is to keep the brightness of the image


constant at monitor.
By regulating the x-ray exposure rate (control kVp, mA or both)

ABC triggers with changing patient size and field modes.


Automatic Brightness Control
Monitor

Camera
Iris/Diaphragm
Optics

Image
Image Intensifier
Brightness
Feedback
Grid

Table

Reference Brightness

Collimator
Filter
Automatic
X ray Tube KkV,mA Brightness Generator
Control
Fluoroscopic Data Acquisition-image
Intensified System

1. X-ray Tube
2. Patient
3. Image Intensifier
4. Output Phosphor
5. Camera
6. Monitor
Video Camera
Two methods are used to
electronically convert the
visible image on the output
phosphor of the image
intensifier into an electronic
signal:
Television camera tube

Thin film transistors (TFT)


The television camera

At the TV camera , an electron beam is swept in raster fashion on


the TV target.

The TV target is a photoconductor, whose electrical resistance is


modulated by varying levels of light intensity.

In areas of more light, more of the


electrons in the electron beam pass
across the TV target and reach the
signal plate, producing a higher
video signal in those lighter
reasons.
The closed circuit TV system used in fluoroscopy.
The television camera

The video signal is amplified


and is transmitted by cable to
the television monitor, where it
is transformed back into a
visible image.
Flat panel digital fluoroscopy

Flat pane devices are thin film transistor (TFT) pixelated arrays
that are rectangular in format and are used as x-ray detectors.
CsI scintillator is used to convert the incident x-ray beam into
light.
TFT systems have a photodiode at each detector element which
converts the light energy to electronic signal
Flat panel digital fluoroscopy

Flat panel detectors replace the image intensifier, video


camera and directly records the real-time fluoroscopic image
sequence.

The flat panel detector is substantially less bulky than the


image intensifier and TV system, but provides the same
functionality.
Troubleshooting/repair of a faulty x-ray
machine
Causes of X-ray tube failures
Troubleshooting/repair of a faulty x-ray
machine

Causes of X-ray tube failures


1. Cathode failure:

Prolonged heating of the filament by


normal current or due to repeated
exposure causes evaporation of
filament metal causing its progressive
thinning which in turn renders it
vulnerable to break
Troubleshooting/repair of a faulty x-ray
machine

Causes of X-ray tube failures


1. Anode failure:

Melting of anode results due to excessive heat production by


the bombardment of electrons on it.

This damage to anode causes uncontrolled x-ray production.

It also causes vibration on the rotor due to imbalance disk


which increases the possibility of anode stem fracture.
Troubleshooting/repair of a faulty x-ray
machine

Causes of X-ray tube failures


1. Glass envelope failure:

It may crack due to secondary arcing from the filament to the


metal deposits on the glass wall as a result of tungsten
evaporation.
Troubleshooting/repair of a faulty x-ray
machine
Steps to extend tube life
Anode should be warmed up before actual exposure is made.
Use high KVp and low Map settings to avoid overheating of the
anode
Consult tube rating chart before to avoid overheating of the anode
Do not allow the overheating of filament by repeated exposure in a
short time.
Do not run rotationary anode unnecessarily as it shortens the life
of bearings
Troubleshooting/repair of a faulty x-ray
machine

Fault Diagnosis
Problem Solving Techniques
Preparation
Step 1 Observation
Step 2 Define Problem Area
Step 3 Identify Possible Causes
Step 4 Determine Most Probable Cause
Step 5 Test and Repair
Troubleshooting/repair of a faulty x-ray
machine

Fault Diagnosis
Observation of a problem Safety
Possibilities that may be observed are: first !!!
Is there an operator error?

A burning smell? Where does it come from?

Is there an increase in temperature? For example:


The X-ray tube housing has become very hot to touch

A lock coil, in the area where a burning smell is observed, is very


hot to touch.
Troubleshooting/repair of a faulty x-ray
machine
Fault Diagnosis
Observation of a problem
Possibilities that may be observed are:
Unusual sound. What sort of sound? Where from?
Absence of sound. For example:
No anode rotation noise from the X-ray tube.
Ventilation fans are quiet

Visual observation. For example:


Smoke rising from equipment, or a HT cable end
Troubleshooting/repair of a faulty x-ray
machine

Fault Diagnosis
Observation of a problem

Possibilities that may be observed are:

Wrong mechanical operation. Look for obstructions, or


loose sections. For example:
A film is jammed in the processor.

Manufacturers manual should be referred at all times!!


Troubleshooting/repair of a faulty x-ray
machine

Fault Diagnosis
Typical Problems
Operator error.
Equipment incorrectly calibrated.
Faulty connecting plugs, sockets, or cables.
A safety interlock is preventing equipment operation.
Electrical or electronic failure.
High-tension cable or X-ray tube failure.
Mechanical problems.
Alignment adjustments
Common X-Ray Machines Problems and possible cause
Assignment
QUIZZ 1
Discuss on the following topics.
X-ray generators, their function and circuit design.

Image quality parameters in fluoroscopy.

Digital subtraction Angiography

Submission Deadline: Nov 24, 2015

Você também pode gostar