RESULTS

2.1 TABLE 1 : LEVEL MEASUREMENT (Tank T1 as OPEN tank, Heater OFF)

Run I Set I Set II Average


A Suggested Level (mm) 700.0 700.0 700.0
B Actual Level (LG) (mm) 693.0 694.0 693.5
C Actual Level (LI) (mm) 698.0 697.0 697.5
D Temperature (TG) (C) 28.2 28.4 28.3
E Deviation = |B-C| (mm) 5.0 3.0 4.0

Run II Set I Set II Average


A Suggested Level (mm) 600.0 600.0 600.0
B Actual Level (LG) (mm) 595.0 596.0 595.5
C Actual Level (LI) (mm) 600.0 599.0 599.5
D Temperature (TG) (C) 28.2 28.0 28.1
E Deviation = |B-C| (mm) 5.0 3.0 4.0

Run III Set I Set II Average


A Suggested Level (mm) 500.0 500.0 500.0
B Actual Level (LG) (mm) 500.0 499.0 499.5
C Actual Level (LI) (mm) 504.0 502.0 503.0
D Temperature (TG) (C) 28.3 28.5 28.4
E Deviation = |B-C| (mm) 4.0 3.0 3.5

Run IV Set I Set II Average


A Suggested Level (mm) 400.0 400.0 400.0
B Actual Level (LG) (mm) 399.0 398.0 398.5
C Actual Level (LI) (mm) 401.0 400.0 400.5
D Temperature (TG) (C) 28.2 28.0 28.1
E Deviation = |B-C| (mm) 2.0 2.0 2.0

Run V Set I Set II Average


A Suggested Level (mm) 300.0 300.0 300.0
B Actual Level (LG) (mm) 298.0 299.0 298.5
C Actual Level (LI) (mm) 300.0 300.0 300.0
D Temperature (TG) (C) 28.2 28.4 28.3
E Deviation = |B-C| (mm) 2.0 1.0 1.5
2.2 TABLE 2 : LEVEL MEASUREMENT (Tank T1 as CLOSED tank, Heater OFF)

Run I Set I Set II Average


A Suggested Level (mm) 700.0 700.0 700.0
B Actual Level (LG) (mm) 688.0 690.0 689.0
C Actual Level (LI) (mm) 691.0 692.0 691.5
D Temperature (TG) (C) 28.2 28.3 28.3
E Deviation = |B-C| (mm) 3.0 2.0 2.5

Run II Set I Set II Average


A Suggested Level (mm) 600.0 600.0 600.0
B Actual Level (LG) (mm) 599.0 600.0 599.5
C Actual Level (LI) (mm) 603.0 602.0 602.5
D Temperature (TG) (C) 28.2 28.4 28.3
E Deviation = |B-C| (mm) 4.0 2.0 3.0

Run III Set I Set II Average


A Suggested Level (mm) 500.0 500.0 500.0
B Actual Level (LG) (mm) 484.0 486.0 485.0
C Actual Level (LI) (mm) 486.0 488.0 487.0
D Temperature (TG) (C) 28.2 28.2 28.2
E Deviation = |B-C| (mm) 2.0 2.0 2.0

Run IV Set I Set II Average


A Suggested Level (mm) 400.0 400.0 400.0
B Actual Level (LG) (mm) 392.0 394.0 393.0
C Actual Level (LI) (mm) 393.0 394.0 393.5
D Temperature (TG) (C) 28.2 28.3 28.3
E Deviation = |B-C| (mm) 1.0 1.0 0.5

Run V Set I Set II Average


A Suggested Level (mm) 300.0 300.0 300.0
B Actual Level (LG) (mm) 295.0 297.0 296.0
C Actual Level (LI) (mm) 297.0 298.0 297.5
D Temperature (TG) (C) 28.2 28.0 28.1
E Deviation = |B-C| (mm) 2.0 1.0 1.5
2.3 TABLE 3 : LEVEL MEASUREMENT (Tank T1 as OPEN tank, Heater ON)

Run I Set I Set II Average


A Suggested Level (mm) 700.0 700.0 700.0
B Actual Level (LG) (mm) 689.0 690.0 689.5
C Actual Level (LI) (mm) 691.0 691.0 691.0
D Temperature (TG) (C) 48.9 49.0 49.0
E Deviation = |B-C| (mm) 2.0 1.0 1.5

Run II Set I Set II Average


A Suggested Level (mm) 600.0 600.0 600.0
B Actual Level (LG) (mm) 598.0 600.0 599.0
C Actual Level (LI) (mm) 600.0 600.0 600.0
D Temperature (TG) (C) 48.8 48.8 48.8
E Deviation = |B-C| (mm) 2.0 0.0 1.0

Run III Set I Set II Average


A Suggested Level (mm) 500.0 500.0 500.0
B Actual Level (LG) (mm) 494.0 495.0 494.5
C Actual Level (LI) (mm) 496.0 496.0 496.0
D Temperature (TG) (C) 48.6 48.5 48.6
E Deviation = |B-C| (mm) 2.0 1.0 1.5

Run IV Set I Set II Average


A Suggested Level (mm) 400.0 400.0 400.0
B Actual Level (LG) (mm) 401.0 402.0 401.5
C Actual Level (LI) (mm) 404.0 403.0 403.5
D Temperature (TG) (C) 48.5 48.4 48.5
E Deviation = |B-C| (mm) 3.0 1.0 2.0

Run V Set I Set II Average


A Suggested Level (mm) 300.0 300.0 300.0
B Actual Level (LG) (mm) 300.0 301.0 300.5
C Actual Level (LI) (mm) 303.0 302.0 302.5
D Temperature (TG) (C) 48.3 48.2 48.3
E Deviation = |B-C| (mm) 3.0 1.0 2.0
2.4 TABLE 4 : LEVEL MEASUREMENT (Tank T1 as CLOSED, tank, Heater ON)

Run I Set I Set II Average


A Suggested Level (mm) 700.0 700.0 700.0
B Actual Level (LG) (mm) 690.0 689.0 689.5
C Actual Level (LI) (mm) 690.0 690.0 690.0
D Temperature (TG) (C) 49.0 49.0 49.0
E Deviation = |B-C| (mm) 0.0 1.0 1.0

Run II Set I Set II Average


A Suggested Level (mm) 600.0 600.0 600.0
B Actual Level (LG) (mm) 590.0 591.0 590.5
C Actual Level (LI) (mm) 590.0 590.0 590.0
D Temperature (TG) (C) 48.8 48.9 48.9
E Deviation = |B-C| (mm) 0.0 1.0 0.5

Run III Set I Set II Average


A Suggested Level (mm) 500.0 500.0 500.0
B Actual Level (LG) (mm) 497.0 498.0 497.5
C Actual Level (LI) (mm) 497.0 498.0 497.5
D Temperature (TG) (C) 48.7 48.8 48.8
E Deviation = |B-C| (mm) 0.0 0.0 0.0

Run IV Set I Set II Average


A Suggested Level (mm) 400.0 400.0 400.0
B Actual Level (LG) (mm) 399.0 400.0 399.5
C Actual Level (LI) (mm) 399.0 399.0 399.0
D Temperature (TG) (C) 48.5 48.6 48.6
E Deviation = |B-C| (mm) 0.0 1.0 0.5

Run V Set I Set II Average


A Suggested Level (mm) 300.0 300.0 300.0
B Actual Level (LG) (mm) 294.0 295.0 294.5
C Actual Level (LI) (mm) 294.0 294.0 294.0
D Temperature (TG) (C) 48.4 48.6 48.5
E Deviation = |B-C| (mm) 0.0 1.0 0.5
SAMPLE CALCULATIONS

Example -

2.1 TABLE 1 : LEVEL MEASUREMENT (Tank T1 as OPEN, Heater OFF)

For (RUN I)

(Run I) - A

+
Average Suggested Level (mm) =
2
700 + 700
=
2
= 700.0 mm

(Run I) B

+
Average Actual Level (LG) (mm) =
2
693 + 694
=
2
= 693.5 mm

(Run I) C

+
Average Actual Level (LI) (mm) =
2
691 + 692
=
2
= 691.5 mm
(Run I) D

+
Average Temperature (TG)(C) =
2
28.2 + 28.4
=
2
= 28.3 C

(Run I) E

At Set I Deviation (mm) = | |

= | 693 698 |

= 5.0 mm
NO. CONTENTS PAGES

1 Cover page -

2 Objectives 1

3 Summary 2

4 Introductions 3

5 Theory

6 Results

7 Discussions

8 Conclusions

9 Recommendations

10 References

11 Appendices
SUMMARY

The main objective of this experiment of level measurement and calibration is to measure

and differentiate the actual level (LG- sight glass with mm scale) and actual level (LI- level

indicator with mm scale) between open and closed tank based on given suggested level by using

level transmitter and indicator. As we know, level is measured at position of interface between

phase (height) and it is a vertical measurement which measure from surface to reference point,

P= pgh where pressure at base is height minus hydrostatic pressure. Level measuring devices in

this experiment is differential device and float device. Other objectives are to measure two type

of measurement, direct measurement and indirect measurement where in indirect measurement,

involves converting measurement which is temperature that can affect the accuracy of indirect

measurement. In this experiment, it was run 5 times with different suggested level starts with 700

mm , 600 mm , 500 mm, 400 mm and 300 mm and the level is measured by four parts of

experiment, where in Experiment 1, is to measure actual level of opened tank with direct

measurement as tank T1 is opened and the heater off. In Experiment 2, the tank T1 is closed and

the heater off as we want to observed the closed tank with direct measurement. However, in

Experiment 3, temperature is used as a factor that can affected accuracy of actual level in indirect

measurement where tank T1 as open tank, heater on and same goes in Experiment 4 but the tank

T1 is closed (close tank). By obtaining correct actual level at level transmitter and indicator and

using correct formula for its deviation, the result is obtained where in Experiment 1 and 2,

employs with physical properties such as fluid motion, buoyancy, optical, thermal, electrical

properties and does not require compensation for changes level causes by temperature changes.

Then, in Experiment 3 and 4, temperature that are used in that experiments are 50 C and the

heater will be turn on until the temperature reach 50C and will turn off the heater automatically,
when the temperature drops the heater will be turned on again as long as the heater stays on at

the panel. So the deviation in this two experiments shows the lowest ranges between LG and LI.

Meanwhile, Experiments 2 and 4, the pressure is open as the experiment conducted was in closed

tank so that the tank is under pressure. In closed tank system, the transmitter location is restricted

by maximum allowable distance above the lower tap. In pressurized system, the transmitter

should be mounted at or below the lower tap to ensure the transmitter always sees a positive

pressure on both the measurements and references sides. The experiment is successful because of

the objectives is reached and based on the actual level (LG), actual level (LI) and the deviation is

observed nearly same with the theory.


SAMPLE CALCULATIONS

Example -

2.2 TABLE 2 : LEVEL MEASUREMENT (Tank T1 as CLOSED, Heater OFF)

For (RUN I)

(Run I) - A

+
Average Suggested Level (mm) =
2
700 + 700
=
2
= 700.0 mm

(Run I) B

+
Average Actual Level (LG) (mm) =
2
688 + 690
=
2
= 689.0 mm

(Run I) C

+
Average Actual Level (LI) (mm) =
2
691 + 692
=
2
= 691.5 mm
(Run I) D

+
Average Temperature (TG)(C) =
2
28.2 + 28.3
=
2
= 28.3 C

(Run I) E

At Set I Deviation (mm) = | |

= | 688 - 691 |

= 3.0 mm
SAMPLE CALCULATIONS

Example -

2.3 TABLE 3 : LEVEL MEASUREMENT (Tank T1 as OPEN, Heater ON)

For (RUN I)

(Run I) - A

+
Average Suggested Level (mm) =
2
700 + 700
=
2
= 700.0 mm

(Run I) B

+
Average Actual Level (LG) (mm) =
2
689 + 690
=
2
= 689.5 mm

(Run I) C

+
Average Actual Level (LI) (mm) =
2
691 + 691
=
2
= 691.0 mm
(Run I) D

+
Average Temperature (TG)(C) =
2
48.9 + 49.0
=
2
= 49.0 C

(Run I) E

At Set I Deviation (mm) = | |

= | 689 691 |

= 2.0 mm
SAMPLE CALCULATIONS

Example -

2.4 TABLE 4 : LEVEL MEASUREMENT (Tank T1 as CLOSED, Heater ON)

For (RUN I)

(Run I) - A

+
Average Suggested Level (mm) =
2
700 + 700
=
2
= 700.0 mm

(Run I) B

+
Average Actual Level (LG) (mm) =
2
690 + 689
=
2
= 689.5 mm

(Run I) C

+
Average Actual Level (LI) (mm) =
2
690 + 690
=
2
= 690.0 mm
(Run I) D

+
Average Temperature (TG)(C) =
2
49.0 + 49.0
=
2
= 49.0 C

(Run I) E

At Set I Deviation (mm) = | |

= | 690 - 690 |

= 0.0 mm