Escolar Documentos
Profissional Documentos
Cultura Documentos
1.1 Switch
1.2 Transformer
AC 33V
AC 220V
AC 13V
AC 13V
Structure
-- The analog single enters circuit magnification unit through BNC connector.
-- The power of front circuit magnification is A 12v, A -12v and constant
current power A 33v.
-- Left side is RBC circuit unit and the right side is WBC circuit unit.
Working principle
The signal magnification unit has channels, namely WBC channel and
RBC/PLT channel.
WBC channel: under the constant source flow, a certain value of pulse
voltage is produced when the WBC pass through the micropore. The voltage
is first magnified by OP37 and then TL072. User can adjust the VR9 to add
magnification rate.
RBC/PLT channel: under the constant source flow, a certain value of pulse
voltage is produced when the WBC pass through the micropore. The voltage
is first magnified by OP37 and then TL072. User can adjust the VR7 to add
magnification rate.
Problem and Solution
Structure
-- HGB analog signal is connected with the signal test unit of WBC cup.
-- The A 12v is transformed into 5v through module 317 providing current for
HGB signal test unit.
Working Principle
The HGB circuit receives and controls HGB signal from WBC cup. The signal
is transformed and magnified and then sent to A/D unit in CPU.
A certain current is provided for LBD under the control of VR3.
Structure
-- Negative Pressure Circuit.
Working Principle
-- Negative pressure circuit: Signal is magnified through circuit. The pressure
sensor detects value in the vacuum chamber. The pressure signal is
magnified through differential magnifier and comparison magnifier and then
output through opt-iso.
-- Liquid level circuit: the circuit consists of lyse unit, detergent unit, diluent
unit and waste unit. In the circuit the signal is magnified and output and then
processed through CPLD.
-- Reagent Test Circuit: In the circuit the liquid level signal is processed
5V Input
Working Principle
The analog signal of WBC, RBC and PLT is respectively transformed by 3
A/D chips. The acquired data is directly sent to FIFO after which CPU reads
related value from FIFO.
HGB analog signal is processed through an A/D transformer inside PSP.
Figure 1-6
Working Principle
The power for mother board has 2 modes, namely analog power and digital
Structure
1.5 Recorder
Figure 1-9
Structure
Temperature test
Working Principle
The recorder receives signal from mother board. The signal is processed and
then sent to printer.
The recorder consists of data buffering circuit, control circuit, motor driving
circuit, thermal printing.
2 Tubing System
Probe
Volume Control
Sampling Board
Assembly
Probe Washing Diluent Adding
Cover Assembly
Counting Cup
Valves
Filter
Figure 2-1
2.2.1 Valves
Specification
-- Voltage: DC 24V
-- Power: 2.5W - 3.4W
-- Ambient Temperature: 0 - 45
-- Fluid Pressure: -90 - 200kpa
Functions
Valves functions are listed as Table 2-1.
Table 2-1 Valves function
No. Function
V1 Switch among RBC adding, mixing and discharging
V2 Switch among WBC adding, mixing and charging
V3 Switch between cup cleaning and measurement
V4 WBC measurement
V5 RBC measurement
V6 RBC mixing
V7 RBC measurement tube cleaning
V8 WBC measurement tube cleaning
V9 RBC measurement tube discharging
V10 WBC measurement tube discharging
V11 Liquid discharging after probe external wahsing
V12 Probe external surface washing with diluent
V13 Diluent adding
V14 Probe inner washing with diluent
V15 Diluent adding to WBC or RBC cup
V16 Lyse adding
V17 Negative pressure access
Figure 2-3
Principle
Black
White ns
+V White
Q1 Q2 Q3 Q4
2.2.3.1 Structure
The WBC cup is used for HGB measurement and WBC totals and diffs. RBC
is used for RBC and PLT totals.
The structure of cups assembly is shown as Figure 2-6.
Chamber
HGB Receiver
Cover
RBC Cup
Figure 2-8
Replacement of Mat
Step:
-- Discharge all liquid and use a screwdriver to loose the two screws. It is a
better method to loose one a little bit and then the other to reach a parallel
screw out.
-- Carefully drag out the back assembly.
-- Use a nipper to get out the mats and the sensor as shown in Figure2-9.
Figure 2-9
-- Installation:
Use a nipper to pick up a mat and pack it into the hole, and then install the
sensor and the two mats. Pack the back assembly and tighten the screws.
-- Check out if any liquid leakage.
-- Use control to make modifications.
Cover
Left Cover
Screwdriver
Figure 2-10
-- Take out the cover with screwdriver.
WBC Signal
RBC Signal
Figure 2-12
-- Take out ground cable.
Ground Cable
Figure 2-13
-- Take out silica tube under sample cup with hand.
Figure 2-14
-- Push sample fixed holder gently with hand, take out sample cup.
RBC Cup
WBC Cup
Figure 2-15
-- Take out hard tube of cup back chamber connector. Attention: hard tube
can not be folded.
NOTE: RBC cup is on the left side and WBC on the right side.
-- OUT of V1 connected to upside connector of WBC cup back chamber with
hand, NC of V2 connected to the downside connector of RBC cup back
chamber.
-- OUT of V13 connected to upside connector of WBC cup back chamber,
NC
connected to downside connector of WBC cup back chamber.
Upper Connector
Down Connector
Figure 2-16
-- Make RBC, WBC signal line through rectangle aperture subjacent sample
Figure 2-17
Function
Main function is sampling, dilute and rinse during testing.
Structure
Sampling unit is composed of sample probe assembly (sample probe,
probe
Figure 2-18
Problem and Reason
-- Problem: MC or MD error
Reason: KD failure
-- Problem: In pre-diluent mode, PLT data over-high
Reason: Interior of sample probe is dirty. Take out and dip in detergent.
-- Problem: Sample probe noisy during running.
Reason: MC failure
-- Problem: V15 spray
Reason: sample probe block
-- Problem: MC or MD error
Reason: Longitude belt of sample unit looses.
-- Problem: Repeatability of RBC test result is not so good.
Reason: There is blood on probe wiper.
Function
The assembly provides diluent and lyse for sampling, counting and cleaning.
Stepping Motor
Check out the rotating shaft and connectors. Make sure that they can work
in a good status.
Installation of Pistons
-- Diluent piston: Pick up a seal ring and put it around the piston rod and then
insert the rod into cylinder. This way can automatically set the seal ring into its
position. Cover the cylinder with 4 screws.
-- Lyse piston: Insert one of the head of glass cylinder into A T-shape
connector and then fill the other space inside the connector with glue. Insert
the piston rod into the glass cylinder.
Figure 2-20
-- Vacuum chamber
The vacuum chamber is an enclosure from which gas is removed by a
vacuum ump. On its upper side there are two ports. One is linked to the
pressure sensor on control board to test vacuum pressure inside. When
abnormal vacuum pressure is read, the pump starts to work to add or
decrease pressure. The other port is linked to related liquid flows of counting
or cleaning.
-- Valve
Valve6 is used to switch vacuum and waste drain.
2.3.1 Startup
Needle reposition
System initialization
Adding detergent Adding detergent
Absorb detergent
into WBC cup into RBC cup
Priming lyse
2.3.2
Cleaning RBC cup
Shut
Do wn
2.3.3 Priming
Soaking
Liquid discharging
2.3.4 Counting
Lyse
Valve 16
Sensor
WBC cup
Diluent
Valve 13
Valve 12
piston)
Valve 14
Valve 15
Sample Sampling
Valve10 Valve10
Sample Cup
Valve8 Valve8
Valve11 Valve11
Predilute
Valve 13
Valve 12
piston)
Valve14
Injector
Sampling Needle
Diluent
Valve 13
Valve 12
piston)
Cleaning fit
Valve 11
Vacuum pump
Waste Container
Pressure pump
2.3.11 Declog
Sample valve
Bubble
Valve 10
Bubble
Valve 9
Valve 8
WBC glass tube
Back chamber
of RBC cup Valve5
Valve 17
Pump
Waste container
3 Adjustment
Condition
If vacuum pump works and no clog or leakage exists inside tubing, negative
pressure adjustment has to be performed when:
-- Clog due to too long counting time
-- The system sends alarm of low vacuum pressure while the vacuum pump
is working normally.
Method
Adjust VR4 and VR5 on mother board.
Condition
The HGB voltage has to be adjusted when:
-- After maintenance or replacement of mother board.
Method
Adjust VR1 and VR3 on mother board referring to specifications.
Condition
The voltage plus has to be adjusted when:
-- After replacement of sample cup.
-- After replacement of cell sensor.
-- After replacement of mother board.
Method
-- Press Menu and select 7 System Set and then press and Ok. The
system will require password 3333 and press Ok. Select Graphic Debug.
After running test, the plus value can be readout as shown in Figure 3-1.
WBC Plus
RBC Plus
Figure 3-1
-- Use control to obtain a plus value and make comparisons.
-- If the plus value obtained from control is much different from real value,
the VR9 and VR7 potentiometer have to be adjusted by the following steps:
a. Open the left door and find a metal cover as shown in Figure 3-2.
Figure 3-3
Press Menu and select 7 System Set and then press and Ok. The system
will require password 3333 and press Ok. Now press Shift and Menu all
motor adjustment data will be readout as listed in Table 3-2.
Table 3-2 Motor adjustment
Item Description Set Value
X2wbc Step of the probe to WBC cup 1215
X2rbc Step of the probe to RBC cup 865
Ymbld The depth of the probe out of the machine 750
Y2wbc1 The fall steps of probe to WBC cup 1 550
Y2wbc2 The fall steps of probe to WBC cup 2 550
Smbld Steps of sample syringe taking the blood 400
Smwbc Absorb steps of sample syringe to WBC cup 585
Dmin The volume of diluent syringe dispense to the 1462
Dmout Volume of diluent syringe pore outside the 480
Dmpin Volume of diluent syringe take from 230
Note:
-- The above value is only for reference, subject to change of different wares.
-- Rotation speed: 180steps/sec.
4 Final Test
After maintenance the engineer must run the device to check its function.
4.1 Startup
-- Turn on power supply.
-- Observe mechanical movement of all parts and make sure that steps
are
normal and smooth.
-- Check all lights. Green indicates the device is ready while orange indicates
the system is counting cells.
-- Check screen brightness.
-- Set up system date and time.
-- Check the function of transmission between Whole blood mode and
predilute mode. System will send out Di if the mode transmission is
successful. Blue light indicates whole blood mode and off light indicates
predilute mode.
4.4 Calibration
-- Run control under whole blood mode and predilute mode.
-- Compare real results and control results. The difference has to meet
specifications.
-- 15 tests are necessary for bother whole blood mode and predilute mode.
Accuracy and repeatability have to meet specifications.
6 HB-7021-YH-006 Transformer
7 HB-7021-YH-007 LCD
18 HB-7021-YH-018 Switch
27 HB-7021-YH-027 O-ring
6 Appendix