Você está na página 1de 15

Federal Register / Vol. 65, No.

201 / Tuesday, October 17, 2000 / Rules and Regulations 61789

Diameter. U.S. Environmental Measurement Branch, Research Triangle 17.0 Tables, Diagrams, Flowcharts,
Protection Agency, Emission Park, North Carolina. January 1977. and Validation Data

Method 2Determination of Stack Gas 1.3 Data Quality Objectives. 22b) be between 0.48 and 0.95 cm (316
Velocity and Volumetric Flow Rate Adherence to the requirements of this and 38 inch). There shall be an equal
(Type S Pitot Tube) method will enhance the quality of the distance from the base of each leg of the
data obtained from air pollutant pitot tube to its face-opening plane
Note: This method does not include all of sampling methods. (dimensions PA and PB, Figure 22b); it
the specifications (e.g., equipment and is recommended that this distance be
supplies) and procedures (e.g., sampling) 2.0 Summary of Method.
between 1.05 and 1.50 times the
essential to its performance. Some material is 2.1 The average gas velocity in a external tubing diameter. The face
incorporated by reference from other stack is determined from the gas density
methods in this part. Therefore, to obtain
openings of the pitot tube shall,
and from measurement of the average preferably, be aligned as shown in
reliable results, persons using this method
velocity head with a Type S Figure 22; however, slight
should have a thorough knowledge of at
(Stausscheibe or reverse type) pitot tube. misalignments of the openings are
least the following additional test method:
Method 1. 3.0 Definitions [Reserved] permissible (see Figure 23).
6.1.2 The Type S pitot tube shall
1.0 Scope and Application. 4.0 Interferences [Reserved] have a known coefficient, determined as
5.0 Safety outlined in Section 10.0. An
1.1 This method is applicable for the identification number shall be assigned
determination of the average velocity 5.1 Disclaimer. This method may
involve hazardous materials, operations, to the pitot tube; this number shall be
and the volumetric flow rate of a gas permanently marked or engraved on the
stream. and equipment. This test method may
not address all of the safety problems body of the tube. A standard pitot tube
1.2 This method is not applicable at associated with its use. It is the may be used instead of a Type S,
measurement sites that fail to meet the responsibility of the user of this test provided that it meets the specifications
criteria of Method 1, Section 11.1. Also, method to establish appropriate safety of Sections 6.7 and 10.2. Note, however,
the method cannot be used for direct and health practices and determine the that the static and impact pressure holes
measurement in cyclonic or swirling gas applicability of regulatory limitations of standard pitot tubes are susceptible to
streams; Section 11.4 of Method 1 prior to performing this test method. plugging in particulate-laden gas
shows how to determine cyclonic or streams. Therefore, whenever a standard
swirling flow conditions. When 6.0 Equipment and Supplies pitot tube is used to perform a traverse,
unacceptable conditions exist, Specifications for the apparatus are adequate proof must be furnished that
alternative procedures, subject to the given below. Any other apparatus that the openings of the pitot tube have not
approval of the Administrator, must be has been demonstrated (subject to plugged up during the traverse period.
employed to produce accurate flow rate approval of the Administrator) to be This can be accomplished by comparing
determinations. Examples of such capable of meeting the specifications the velocity head (p) measurement
alternative procedures are: (1) to install will be considered acceptable. recorded at a selected traverse point
straightening vanes; (2) to calculate the 6.1 Type S Pitot Tube. (readable p value) with a second p
total volumetric flow rate 6.1.1 Pitot tube made of metal tubing measurement recorded after back
stoichiometrically, or (3) to move to (e.g., stainless steel) as shown in Figure purging with pressurized air to clean
another measurement site at which the 21. It is recommended that the external the impact and static holes of the
flow is acceptable. tubing diameter (dimension Dt, Figure standard pitot tube. If the before and

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00047 Fmt 4701 Sfmt 4700 C:17OCR2.SGM pfrm03 PsN: 17OCR2
61790 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

after p measurements are within 5 6.3 Temperature Sensor. A pitot tube with an NIST-traceable
percent, then the traverse data are thermocouple, liquid-filled bulb coefficient. Alternatively, a standard
acceptable. Otherwise, the data should thermometer, bimetallic thermometer, pitot tube designed according to the
be rejected and the traverse mercury-in-glass thermometer, or other criteria given in Sections 6.7.1 through
measurements redone. Note that the gauge capable of measuring 6.7.5 below and illustrated in Figure 2
selected traverse point should be one temperatures to within 1.5 percent of 5 (see also References 7, 8, and 17 in
that demonstrates a readable p value. the minimum absolute stack Section 17.0) may be used. Pitot tubes
If back purging at regular intervals is temperature. The temperature sensor designed according to these
part of a routine procedure, then shall be attached to the pitot tube such specifications will have baseline
comparative p measurements shall be that the sensor tip does not touch any coefficients of 0.99 0.01.
conducted as above for the last two metal; the gauge shall be in an 6.7.1 Standard Pitot Design.
traverse points that exhibit suitable p interference-free arrangement with 6.7.1.1 Hemispherical (shown in
measurements. respect to the pitot tube face openings Figure 25), ellipsoidal, or conical tip.
6.2 Differential Pressure Gauge. An (see Figure 21 and Figure 24). 6.7.1.2 A minimum of six diameters
inclined manometer or equivalent Alternative positions may be used if the straight run (based upon D, the external
device. Most sampling trains are pitot tube-temperature gauge system is diameter of the tube) between the tip
equipped with a 10 in. (water column) calibrated according to the procedure of and the static pressure holes.
inclined-vertical manometer, having Section 10.0. Provided that a difference 6.7.1.3 A minimum of eight
0.01 in. H20 divisions on the 0 to 1 in. of not more than 1 percent in the diameters straight run between the static
inclined scale, and 0.1 in. H20 divisions average velocity measurement is pressure holes and the centerline of the
on the 1 to 10 in. vertical scale. This introduced, the temperature gauge need external tube, following the 90 bend.
type of manometer (or other gauge of not be attached to the pitot tube. This 6.7.1.4 Static pressure holes of equal
equivalent sensitivity) is satisfactory for alternative is subject to the approval of size (approximately 0.1 D), equally
the measurement of p values as low as the Administrator. spaced in a piezometer ring
1.27 mm (0.05 in.) H20. However, a 6.4 Pressure Probe and Gauge. A configuration.
differential pressure gauge of greater piezometer tube and mercury- or water- 6.7.1.5 90 bend, with curved or
sensitivity shall be used (subject to the filled U-tube manometer capable of mitered junction.
approval of the Administrator), if any of measuring stack pressure to within 2.5 6.8 Differential Pressure Gauge for
the following is found to be true: (1) the mm (0.1 in.) Hg. The static tap of a Type S Pitot Tube Calibration. An
arithmetic average of all p readings at standard type pitot tube or one leg of a inclined manometer or equivalent. If the
the traverse points in the stack is less Type S pitot tube with the face opening single-velocity calibration technique is
than 1.27 mm (0.05 in.) H20; (2) for planes positioned parallel to the gas employed (see Section 10.1.2.3), the
traverses of 12 or more points, more flow may also be used as the pressure calibration differential pressure gauge
than 10 percent of the individual p probe. shall be readable to the nearest 0.127
6.5 Barometer. A mercury, aneroid, mm (0.005 in.) H20. For multivelocity
readings are below 1.27 mm (0.05 in.)
or other barometer capable of measuring calibrations, the gauge shall be readable
H20; or (3) for traverses of fewer than 12
atmospheric pressure to within 2.54 mm to the nearest 0.127 mm (0.005 in.) H20
points, more than one p reading is
below 1.27 mm (0.05 in.) H20. Reference
(0.1 in.) Hg. for p values between 1.27 and 25.4 mm
18 (see Section 17.0) describes Note: The barometric pressure reading may (0.05 and 1.00 in.) H20, and to the
commercially available instrumentation be obtained from a nearby National Weather nearest 1.27 mm (0.05 in.) H20 for p
for the measurement of low-range gas Service station. In this case, the station value values above 25.4 mm (1.00 in.) H20. A
(which is the absolute barometric pressure) special, more sensitive gauge will be
velocities. shall be requested and an adjustment for required to read p values below 1.27
6.2.1 As an alternative to criteria (1) elevation differences between the weather mm (0.05 in.) H20 (see Reference 18 in
through (3) above, Equation 21 station and sampling point shall be made at
a rate of minus 2.5 mm (0.1 in.) Hg per 30 Section 16.0).
(Section 12.2) may be used to determine
m (100 ft) elevation increase or plus 2.5 mm 7.0 Reagents and Standards [Reserved]
the necessity of using a more sensitive (0.1 in.) Hg per 30 m (100 ft.) for elevation
differential pressure gauge. If T is decrease. 8.0 Sample Collection and Analysis
greater than 1.05, the velocity head data
are unacceptable and a more sensitive 6.6 Gas Density Determination 8.1 Set up the apparatus as shown in
differential pressure gauge must be Equipment. Method 3 equipment, if Figure 21. Capillary tubing or surge
used. needed (see Section 8.6), to determine tanks installed between the manometer
the stack gas dry molecular weight, and and pitot tube may be used to dampen
Note: If differential pressure gauges other
Method 4 (reference method) or Method p fluctuations. It is recommended, but
than inclined manometers are used (e.g.,
magnehelic gauges), their calibration must be 5 equipment for moisture content not required, that a pretest leak-check be
checked after each test series. To check the determination. Other methods may be conducted as follows: (1) blow through
calibration of a differential pressure gauge, used subject to approval of the the pitot impact opening until at least
compare p readings of the gauge with those Administrator. 7.6 cm (3.0 in.) H20 velocity head
of a gauge-oil manometer at a minimum of 6.7 Calibration Pitot Tube. When registers on the manometer; then, close
three points, approximately representing the calibration of the Type S pitot tube is off the impact opening. The pressure
range of p values in the stack. If, at each necessary (see Section 10.1), a standard shall remain stable for at least 15
point, the values of p as read by the pitot tube shall be used for a reference. seconds; (2) do the same for the static
differential pressure gauge and gauge-oil The standard pitot tube shall, pressure side, except using suction to
manometer agree to within 5 percent, the
preferably, have a known coefficient, obtain the minimum of 7.6 cm (3.0 in.)
differential pressure gauge shall be
considered to be in proper calibration. obtained either (1) directly from the H20. Other leak-check procedures,
Otherwise, the test series shall either be National Institute of Standards and subject to the approval of the
voided, or procedures to adjust the measured Technology (NIST), Gaithersburg MD Administrator, may be used.
p values and final results shall be used, 20899, (301) 9752002, or (2) by 8.2 Level and zero the manometer.
subject to the approval of the Administrator. calibration against another standard Because the manometer level and zero

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00048 Fmt 4701 Sfmt 4700 C:17OCR2.SGM pfrm03 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61791

may drift due to vibrations and point. Conduct a post-test leak-check of 29.0. For other processes, other
temperature changes, make periodic (mandatory), as described in Section 8.1 methods, subject to the approval of the
checks during the traverse (at least once above, to validate the traverse run. Administrator, must be used.
per hour). Record all necessary data on 8.4 Measure the static pressure in 8.7 Obtain the moisture content
a form similar to that shown in Figure the stack. One reading is usually from Method 4 (reference method, or
26. adequate. equivalent) or from Method 5.
8.3 Measure the velocity head and 8.5 Determine the atmospheric 8.8 Determine the cross-sectional
temperature at the traverse points pressure. area of the stack or duct at the sampling
specified by Method 1. Ensure that the location. Whenever possible, physically
8.6 Determine the stack gas dry
proper differential pressure gauge is measure the stack dimensions rather
molecular weight. For combustion
than using blueprints. Do not assume
being used for the range of p values processes or processes that emit
that stack diameters are equal. Measure
encountered (see Section 6.2). If it is essentially CO2, O2, CO, and N2, use
each diameter distance to verify its
necessary to change to a more sensitive Method 3. For processes emitting
dimensions.
gauge, do so, and remeasure the p and essentially air, an analysis need not be
temperature readings at each traverse conducted; use a dry molecular weight 9.0 Quality Control

Section Quality control measure Effect

10.110.4 .................................. Sampling equipment calibration ....................... Ensure accurate measurement of stack gas flow rate, sample
volume.

10.0 Calibration and Standardization baseline coefficient value may or may calculated according to Equation 22
10.1 Type S Pitot Tube. Before its not be valid for a given assembly. The (see Section 12.3), to determine the
initial use, carefully examine the Type baseline and assembly coefficient values number of duct diameters. To ensure the
S pitot tube top, side, and end views to will be identical only when the relative presence of stable, fully developed flow
verify that the face openings of the tube placement of the components in the patterns at the calibration site, or test
are aligned within the specifications assembly is such that aerodynamic section, the site must be located at
illustrated in Figures 22 and 23. The interference effects are eliminated. least eight diameters downstream and
pitot tube shall not be used if it fails to Figures 24, 27, and 28 illustrate two diameters upstream from the
meet these alignment specifications. interference-free component nearest disturbances.
After verifying the face opening arrangements for Type S pitot tubes Note: The eight- and two-diameter criteria
alignment, measure and record the having external tubing diameters are not absolute; other test section locations
following dimensions of the pitot tube: between 0.48 and 0.95 cm (316 and 38 may be used (subject to approval of the
(a) the external tubing diameter in.). Type S pitot tube assemblies that Administrator), provided that the flow at the
fail to meet any or all of the test site has been demonstrated to be or
(dimension Dt, Figure 22b); and (b) the found stable and parallel to the duct axis.
base-to-opening plane distances specifications of Figures 24, 27, and
(dimensions PA and PB, Figure 22b). If 2-8 shall be calibrated according to the 10.1.2.3 The flow system shall have
Dt is between 0.48 and 0.95 cm 316 and procedure outlined in Sections 10.1.2 the capacity to generate a test-section
38 in.), and if P and P are equal and
A B
through 10.1.5, and prior to calibration, velocity around 910 m/min (3,000 ft/
between 1.05 and 1.50 Dt, there are two the values of the intercomponent min). This velocity must be constant
possible options: (1) the pitot tube may spacings (pitot-nozzle, pitot- with time to guarantee steady flow
be calibrated according to the procedure thermocouple, pitot-probe sheath) shall during calibration. Note that Type S
outlined in Sections 10.1.2 through be measured and recorded. pitot tube coefficients obtained by
10.1.5, or (2) a baseline (isolated tube) Note: Do not use a Type S pitot tube single-velocity calibration at 910 m/min
coefficient value of 0.84 may be assembly that is constructed such that the (3,000 ft/min) will generally be valid to
assigned to the pitot tube. Note, impact pressure opening plane of the pitot 3 percent for the measurement of
however, that if the pitot tube is part of tube is below the entry plane of the nozzle velocities above 300 m/min (1,000 ft/
an assembly, calibration may still be (see Figure 26B). min) and to 6 percent for the
required, despite knowledge of the 10.1.2 Calibration Setup. If the Type measurement of velocities between 180
baseline coefficient value (see Section S pitot tube is to be calibrated, one leg and 300 m/min (600 and 1,000 ft/min).
10.1.1). If Dt, PA, and PB are outside the of the tube shall be permanently marked If a more precise correlation between
specified limits, the pitot tube must be A, and the other, B. Calibration shall be the pitot tube coefficient, (Cp), and
calibrated as outlined in Sections 10.1.2 performed in a flow system having the velocity is desired, the flow system
through 10.1.5. following essential design features: should have the capacity to generate at
10.1.1 Type S Pitot Tube 10.1.2.1 The flowing gas stream least four distinct, time-invariant test-
Assemblies. During sample and velocity must be confined to a duct of definite section velocities covering the velocity
traverses, the isolated Type S pitot tube cross-sectional area, either circular or range from 180 to 1,500 m/min (600 to
is not always used; in many instances, rectangular. For circular cross sections, 5,000 ft/min), and calibration data shall
the pitot tube is used in combination the minimum duct diameter shall be be taken at regular velocity intervals
with other source-sampling components 30.48 cm (12 in.); for rectangular cross over this range (see References 9 and 14
(e.g., thermocouple, sampling probe, sections, the width (shorter side) shall in Section 17.0 for details).
nozzle) as part of an assembly. The be at least 25.4 cm (10 in.). 10.1.2.4 Two entry ports, one for
presence of other sampling components 10.1.2.2 The cross-sectional area of each of the standard and Type S pitot
can sometimes affect the baseline value the calibration duct must be constant tubes, shall be cut in the test section.
of the Type S pitot tube coefficient over a distance of 10 or more duct The standard pitot entry port shall be
(Reference 9 in Section 17.0); therefore, diameters. For a rectangular cross located slightly downstream of the Type
an assigned (or otherwise known) section, use an equivalent diameter, S port, so that the standard and Type S

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00049 Fmt 4701 Sfmt 4700 C:17OCR2.SGM pfrm03 PsN: 17OCR2
61792 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

impact openings will lie in the same the absolute value of the difference calibration velocity. Note also that it is
cross-sectional plane during calibration. between Cp(A) and Cp(B) is 0.01 or less. not necessary to draw an isokinetic
To facilitate alignment of the pitot tubes 10.1.4 Special Considerations. sample during calibration (see Reference
during calibration, it is advisable that 10.1.4.1 Selection of Calibration 19 in Section 17.0).
the test section be constructed of Point. 10.1.4.3 For a probe assembly
PlexiglasTM or some other transparent 10.1.4.1.1 When an isolated Type S constructed such that its pitot tube is
material. pitot tube is calibrated, select a always used in the same orientation,
10.1.3 Calibration Procedure. Note calibration point at or near the center of only one side of the pitot tube need be
that this procedure is a general one and the duct, and follow the procedures calibrated (the side which will face the
must not be used without first referring outlined in Section 10.1.3. The Type S flow). The pitot tube must still meet the
to the special considerations presented pitot coefficients measured or alignment specifications of Figure 22
in Section 10.1.5. Note also that this calculated, (i.e. Cp(A) and Cp(B)) will be or 23, however, and must have an
procedure applies only to single- valid, so long as either: (1) the isolated average deviation () value of 0.01 or
velocity calibration. To obtain pitot tube is used; or (2) the pitot tube less (see Section 10.1.4.4).
calibration data for the A and B sides of is used with other components (nozzle, 10.1.5 Field Use and Recalibration.
the Type S pitot tube, proceed as thermocouple, sample probe) in an 10.1.5.1 Field Use.
arrangement that is free from 10.1.5.1.1 When a Type S pitot tube
follows:
aerodynamic interference effects (see (isolated or in an assembly) is used in
10.1.3.1 Make sure that the
Figures 24, 27, and 28). the field, the appropriate coefficient
manometer is properly filled and that
10.1.4.1.2 For Type S pitot tube- value (whether assigned or obtained by
the oil is free from contamination and
thermocouple combinations (without calibration) shall be used to perform
is of the proper density. Inspect and
probe assembly), select a calibration velocity calculations. For calibrated
leak-check all pitot lines; repair or
point at or near the center of the duct, Type S pitot tubes, the A side
replace if necessary.
and follow the procedures outlined in coefficient shall be used when the A
10.1.3.2 Level and zero the Section 10.1.3. The coefficients so side of the tube faces the flow, and the
manometer. Switch on the fan, and obtained will be valid so long as the B side coefficient shall be used when
allow the flow to stabilize. Seal the pitot tube-thermocouple combination is the B side faces the flow. Alternatively,
Type S pitot tube entry port. used by itself or with other components the arithmetic average of the A and B
10.1.3.3 Ensure that the manometer in an interference-free arrangement side coefficient values may be used,
is level and zeroed. Position the (Figures 24, 27, and 28). irrespective of which side faces the
standard pitot tube at the calibration 10.1.4.1.3 For Type S pitot tube flow.
point (determined as outlined in Section combinations with complete probe 10.1.5.1.2 When a probe assembly is
10.1.5.1), and align the tube so that its assemblies, the calibration point should used to sample a small duct, 30.5 to 91.4
tip is pointed directly into the flow. be located at or near the center of the cm (12 to 36 in.) in diameter, the probe
Particular care should be taken in duct; however, insertion of a probe sheath sometimes blocks a significant
aligning the tube to avoid yaw and pitch sheath into a small duct may cause part of the duct cross-section, causing a
angles. Make sure that the entry port significant cross-sectional area reduction in the effective value of Cp(s).
surrounding the tube is properly sealed. interference and blockage and yield Consult Reference 9 (see Section 17.0)
10.1.3.4 Read pstd, and record its incorrect coefficient values (Reference 9 for details. Conventional pitot-sampling
value in a data table similar to the one in Section 17.0). Therefore, to minimize probe assemblies are not recommended
shown in Figure 29. Remove the the blockage effect, the calibration point for use in ducts having inside diameters
standard pitot tube from the duct, and may be a few inches off-center if smaller than 30.5 cm (12 in.) (see
disconnect it from the manometer. Seal necessary. The actual blockage effect Reference 16 in Section 17.0).
the standard entry port. will be negligible when the theoretical 10.1.5.2 Recalibration.
10.1.3.5 Connect the Type S pitot blockage, as determined by a projected- 10.1.5.2.1 Isolated Pitot Tubes. After
tube to the manometer and leak-check. area model of the probe sheath, is 2 each field use, the pitot tube shall be
Open the Type S tube entry port. Check percent or less of the duct cross- carefully reexamined in top, side, and
the manometer level and zero. Insert sectional area for assemblies without end views. If the pitot face openings are
and align the Type S pitot tube so that external sheaths (Figure 210a), and 3 still aligned within the specifications
its A side impact opening is at the same percent or less for assemblies with illustrated in Figure 22 and Figure 2
point as was the standard pitot tube and external sheaths (Figure 210b). 3, it can be assumed that the baseline
is pointed directly into the flow. Make 10.1.4.2 For those probe assemblies coefficient of the pitot tube has not
sure that the entry port surrounding the in which pitot tube-nozzle interference changed. If, however, the tube has been
tube is properly sealed. is a factor (i.e., those in which the pitot- damaged to the extent that it no longer
10.1.3.6 Read ps, and enter its nozzle separation distance fails to meet meets the specifications of Figure 22
value in the data table. Remove the the specifications illustrated in Figure and Figure 23, the damage shall either
Type S pitot tube from the duct, and 27A), the value of Cp(s) depends upon be repaired to restore proper alignment
disconnect it from the manometer. the amount of free space between the of the face openings, or the tube shall be
10.1.3.7 Repeat Steps 10.1.3.3 tube and nozzle and, therefore, is a discarded.
through 10.1.3.6 until three pairs of p function of nozzle size. In these 10.1.5.2.2 Pitot Tube Assemblies.
readings have been obtained for the A instances, separate calibrations shall be After each field use, check the face
side of the Type S pitot tube. performed with each of the commonly opening alignment of the pitot tube, as
10.1.3.8 Repeat Steps 10.1.3.3 used nozzle sizes in place. Note that the in Section 10.1.5.2.1. Also, remeasure
through 10.1.3.7 for the B side of the single-velocity calibration technique is the intercomponent spacings of the
Type S pitot tube. acceptable for this purpose, even though assembly. If the intercomponent
10.1.3.9 Perform calculations as the larger nozzle sizes (>0.635 cm or 14 spacings have not changed and the face
described in Section 12.4. Use the Type in.) are not ordinarily used for isokinetic opening alignment is acceptable, it can
S pitot tube only if the values of A and sampling at velocities around 910 m/ be assumed that the coefficient of the
B are less than or equal to 0.01 and if min (3,000 ft/min), which is the assembly has not changed. If the face

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00050 Fmt 4701 Sfmt 4700 C:17OCR2.SGM pfrm03 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61793

opening alignment is no longer within 11.0 Analytical Procedure Tstd = Standard absolute temperature,
the specifications of Figure 22 and Sample collection and analysis are 293 K (528 R).
Figure 23, either repair the damage or concurrent for this method (see Section Vs = Average stack gas velocity, m/sec
replace the pitot tube (calibrating the 8.0). (ft/sec).
new assembly, if necessary). If the W = Width.
intercomponent spacings have changed, 12.0 Data Analysis and Calculations p = Velocity head of stack gas, mm
restore the original spacings, or Carry out calculations, retaining at H2O (in. H20).
recalibrate the assembly. least one extra significant figure beyond pi = Individual velocity head reading
10.2 Standard Pitot Tube (if that of the acquired data. Round off at traverse point i, mm (in.) H2O.
applicable). If a standard pitot tube is figures after final calculation. pstd = Velocity head measured by the
used for the velocity traverse, the tube standard pitot tube, cm (in.) H2O.
12.1 Nomenclature.
shall be constructed according to the ps = Velocity head measured by the
criteria of Section 6.7 and shall be A = Cross-sectional area of stack, m2
(ft2). Type S pitot tube, cm (in.) H2O.
assigned a baseline coefficient value of
Bws = Water vapor in the gas stream 3600 = Conversion Factor, sec/hr.
0.99. If the standard pitot tube is used
as part of an assembly, the tube shall be (from Method 4 (reference method) 18.0 = Molecular weight of water, g/g-
in an interference-free arrangement or Method 5), proportion by mole (lb/lb-mole).
(subject to the approval of the volume. 12.2 Calculate T as follows:
Administrator). Cp = Pitot tube coefficient,
n
dimensionless.

10.3 Temperature Sensors.
10.3.1 After each field use, calibrate Cp(s) = Type S pitot tube coefficient, p i + K
i =1
dial thermometers, liquid-filled bulb dimensionless. = n Eq. 2-1
thermometers, thermocouple-
potentiometer systems, and other
Cp(std) = Standard pitot tube coefficient;
use 0.99 if the coefficient is p i
i =1
sensors at a temperature within 10 unknown and the tube is designed
percent of the average absolute stack according to the criteria of Sections 12.3 Calculate De as follows:
temperature. For temperatures up to 405 6.7.1 to 6.7.5 of this method.
2 LW
C (761 F), use an ASTM mercury-in- De = Equivalent diameter. De = Eq. 2-2
glass reference thermometer, or K = 0.127 mm H2O (metric units). 0.005 L+W
equivalent, as a reference. Alternatively, in. H2O (English units). 12.4 Calibration of Type S Pitot
either a reference thermocouple and Kp = Velocity equation constant. Tube.
potentiometer (calibrated against NIST L = Length. 12.4.1 For each of the six pairs of p
standards) or thermometric fixed points Md = Molecular weight of stack gas, dry readings (i.e., three from side A and
(e.g., ice bath and boiling water, basis (see Section 8.6), g/g-mole (lb/ three from side B) obtained in Section
corrected for barometric pressure) may lb-mole). 10.1.3, calculate the value of the Type
be used. For temperatures above 405C Ms = Molecular weight of stack gas, wet S pitot tube coefficient according to
(761 F), use a reference thermocouple- basis, g/g-mole (lb/lb-mole). Equation 23:
potentiometer system calibrated against n = Total number of traverse points.
NIST standards or an alternative
Pbar = Barometric pressure at pstd
reference, subject to the approval of the C p(s) = C p(std ) Eq. 2-3
Administrator.
measurement site, mm Hg (in. Hg). p
Pg = Stack static pressure, mm Hg (in.
10.3.2 The temperature data 12.4.2 Calculate Cp(A), the mean A-
Hg).
recorded in the field shall be considered side coefficient, and Cp(B), the mean B-
valid. If, during calibration, the absolute Ps = Absolute stack pressure (Pbar + Pg),
side coefficient. Calculate the difference
temperature measured with the sensor mm Hg (in. Hg),
between these two average values.
being calibrated and the reference Pstd = Standard absolute pressure, 760
12.4.3 Calculate the deviation of
sensor agree within 1.5 percent, the mm Hg (29.92 in. Hg).
each of the three A-side values of Cp(s)
temperature data taken in the field shall Qsd = Dry volumetric stack gas flow rate
from Cp(A), and the deviation of each of
be considered valid. Otherwise, the corrected to standard conditions,
the three B-side values of Cp(s) from
pollutant emission test shall either be dscm/hr (dscf/hr).
Cp(B), using Equation 24:
considered invalid or adjustments (if T = Sensitivity factor for differential
appropriate) of the test results shall be pressure gauges. Deviation = C p(s) Cp Eq. 2-4
made, subject to the approval of the Ts = Stack temperature, C (F). ( A or B)
Administrator. Ts(abs) = Absolute stack temperature, K 12.4.4 Calculate the average
10.4 Barometer. Calibrate the (R). deviation from the mean, for both the A
barometer used against a mercury = 273 + Ts for metric units, and B sides of the pitot tube. Use
barometer. = 460 + Ts for English units. Equation 25:

3
C p (s) C p( A or B)
i =1
A or B = Eq. 2-5
3

12.5 Molecular Weight of Stack Gas. M s = M d (1 Bws ) + 18.0 Bws Eq. 2-6 12.6 Average Stack Gas Velocity.

VerDate 11<MAY>2000 21:10 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00051 Fmt 4701 Sfmt 4700 E:\FR\FM\17OCR2.SGM pfrm04 PsN: 17OCR2
61794 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

Ts(abs)
Vs = K p C p p avg Eq. 2-7
Ps M s

1
m (g/g mole)( mmHg) 2
34.97 Metric
sec ( K )( mmH 2 O)

1
m (1b/1b - mole)(in.Hg) 2
85.49 English
sec ( R )(in. H 2 O)
12.7 Average Stack Gas Dry Volumetric Flow Rate.

T P
Q = 3600(1 Bws )vs A std s Eq. 2-8
Ts (abs) Pstd

13.0 Method Performance [Reserved] 8. Annual Book of ASTM Standards, Part Velocities. U.S. Environmental Protection
26. 1974. p. 648. Agency, Emission Measurement Branch,
14.0 Pollution Prevention [Reserved] 9. Vollaro, R.F. Guidelines for Type S Pitot Research Triangle Park, NC. November 1976.
15.0 Waste Management [Reserved] Tube Calibration. U.S. Environmental 15. Smith, Marvin L. Velocity Calibration
Protection Agency, Research Triangle Park, of EPA Type Source Sampling Probe. United
16.0 References N.C. (Presented at 1st Annual Meeting, Technologies Corporation, Pratt and Whitney
Source Evaluation Society, Dayton, OH, Aircraft Division, East Hartford, CT. 1975.
1. Mark, L.S. Mechanical Engineers September 18, 1975.) 16. Vollaro, R.F. Recommended Procedure
Handbook. New York. McGraw-Hill Book 10. Vollaro, R.F. A Type S Pitot Tube for Sample Traverses in Ducts Smaller than
Co., Inc. 1951. Calibration Study. U.S. Environmental 12 Inches in Diameter. U.S. Environmental
2. Perry, J.H., ed. Chemical Engineers Protection Agency, Emission Measurement Protection Agency, Emission Measurement
Handbook. New York. McGraw-Hill Book Branch, Research Triangle Park, N.C. July Branch, Research Triangle Park, NC.
Co., Inc. 1960. 1974. November 1976.
3. Shigehara, R.T., W.F. Todd, and W.S. 11. Vollaro, R.F. The Effects of Impact 17. Ower, E. and R.C. Pankhurst. The
Smith. Significance of Errors in Stack Opening Misalignment on the Value of the
Measurement of Air Flow, 4th Ed. London,
Sampling Measurements. U.S. Environmental Type S Pitot Tube Coefficient. U.S.
Pergamon Press. 1966.
Protection Agency, Research Triangle Park, Environmental Protection Agency, Emission
Measurement Branch, Research Triangle 18. Vollaro, R.F. A Survey of Commercially
N.C. (Presented at the Annual Meeting of the Available Instrumentation for the
Park, NC. October 1976.
Air Pollution Control Association, St. Louis, Measurement of Low-Range Gas Velocities.
12. Vollaro, R.F. Establishment of a
MO., June 1419, 1970). Baseline Coefficient Value for Properly U.S. Environmental Protection Agency,
4. Standard Method for Sampling Stacks Constructed Type S Pitot Tubes. U.S. Emission Measurement Branch, Research
for Particulate Matter. In: 1971 Book of Environmental Protection Agency, Emission Triangle Park, NC. November 1976.
ASTM Standards, Part 23. Philadelphia, PA. Measurement Branch, Research Triangle (Unpublished Paper).
1971. ASTM Designation D 292871. Park, NC. November 1976. 19. Gnyp, A.W., et al. An Experimental
5. Vennard, J.K. Elementary Fluid 13. Vollaro, R.F. An Evaluation of Single- Investigation of the Effect of Pitot Tube-
Mechanics. New York. John Wiley and Sons, Velocity Calibration Technique as a Means of Sampling Probe Configurations on the
Inc. 1947. Determining Type S Pitot Tube Coefficients. Magnitude of the S Type Pitot Tube
6. Fluid MetersTheir Theory and U.S. Environmental Protection Agency, Coefficient for Commercially Available
Application. American Society of Mechanical Emission Measurement Branch, Research Source Sampling Probes. Prepared by the
Engineers, New York, N.Y. 1959. Triangle Park, NC. August 1975. University of Windsor for the Ministry of the
7. ASHRAE Handbook of Fundamentals. 14. Vollaro, R.F. The Use of Type S Pitot Environment, Toronto, Canada. February
1972. p. 208. Tubes for the Measurement of Low 1975.

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00052 Fmt 4701 Sfmt 4700 C:17OCR2.SGM pfrm03 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61795

17.0 Tables, Diagrams, Flowcharts, and Validation Data

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00053 Fmt 4701 Sfmt 4725 C:17OCR2.SGM pfrm03 PsN: 17OCR2
61796 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00054 Fmt 4701 Sfmt 4725 C:17OCR2.SGM pfrm03 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61797

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00055 Fmt 4701 Sfmt 4725 C:17OCR2.SGM pfrm03 PsN: 17OCR2
61798 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00056 Fmt 4701 Sfmt 4725 C:17OCR2.SGM pfrm03 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61799

PLANT lllllllllllllllllllllllllllllllllllllllllllllllllllllllllllll
DATE llllllllllllllllllllllllllllllllllllllllllllllllllllllllllllll
RUN NO. llllllllllllllllllllllllllllllllllllllllllllllllllllllllllll
STACK DIA. OR DIMENSIONS, m (in.) lllllllllllllllllllllllllllllllllllllllllllllll
BAROMETRIC PRESS., mm Hg (in. Hg) llllllllllllllllllllllllllllllllllllllllllllll
CROSS SECTIONAL AREA, m2 (ft2) llllllllllllllllllllllllllllllllllllllllllllllll
OPERATORS llllllllllllllllllllllllllllllllllllllllllllllllllllllllll
PITOT TUBE I.D. NO. llllllllllllllllllllllllllllllllllllllllllllllllllllll
AVG. COEFFICIENT, Cp = llllllllllllllllllllllllllllllllllllllllllllllllllll
LAST DATE CALIBRATED llllllllllllllllllllllllllllllllllllllllllllllllllll

SCHEMATIC OF STACK CROSS SECTION

VerDate 11<MAY>2000 21:10 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00057 Fmt 4701 Sfmt 4706 E:\FR\FM\17OCR2.SGM pfrm04 PsN: 17OCR2
61800 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

Stack
Vel. Hd., temperature Pg
Traverse p mm Hg (p)12
Pt. No. mm (in.) Ts, Ts, (in. Hg)
H2O C (F) K (R)

Average(1)

Figure 26. Velocity Traverse Data

VerDate 11<MAY>2000 21:10 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00058 Fmt 4701 Sfmt 4725 E:\FR\FM\17OCR2.SGM pfrm04 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61801

VerDate 11<MAY>2000 10:28 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00059 Fmt 4701 Sfmt 4725 C:17OCR2.SGM pfrm03 PsN: 17OCR2
61802 Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations

PITOT TUBE IDENTIFICATION NUMBER: lllllllllllllllllllllllllllllllllllllllllllll


DATE: lllllllllllllllllllllllllllllllllllllllllllllllllllllllllllll
CALIBRATED BY: llllllllllllllllllllllllllllllllllllllllllllllllllllllll

A SIDE CALIBRATION
Pstd P(s) Deviation
Run No. cm H2O cm H2O Cp(s) Cp(s)Cp(A)
(in H2O) (in H2O)

Cp, avg
(SIDE A)

B SIDE CALIBRATION
Pstd P(s) Deviation
Run No. cm H2O cm H2O Cp(s) Cp(s)Cp(B)
(in H2O) (in H2O)

Cp, avg
(SIDE B)

3
C p(s) Cp (A or B)
i =1
A or B = Eq. 2-5
3

[Cp, avg (side A)Cp, avg (side B)]*


*Must be less than or equal to 0.01
Figure 29. Pitot Tube Calibration Data

VerDate 11<MAY>2000 21:10 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00060 Fmt 4701 Sfmt 4706 E:\FR\FM\17OCR2.SGM pfrm04 PsN: 17OCR2
Federal Register / Vol. 65, No. 201 / Tuesday, October 17, 2000 / Rules and Regulations 61803

Method 2ADirect Measurement of 1.0 Scope and Application 2.0 Summary of Method
Gas Volume Through Pipes and Small 2.1 A gas volume meter is used to
Ducts 1.1 This method is applicable for the
determination of gas flow rates in pipes measure gas volume directly.
Note: This method does not include all of and small ducts, either in-line or at Temperature and pressure
the specifications (e.g., equipment and exhaust positions, within the measurements are made to allow
correction of the volume to standard
supplies) and procedures (e.g., sampling) temperature range of 0 to 50 C (32 to
conditions.
essential to its performance. Some material is 122 F).
incorporated by reference from other 3.0 Definitions [Reserved]
methods in this part. Therefore, to obtain
1.2 Data Quality Objectives.
reliable results, persons using this method Adherence to the requirements of this 4.0 Interferences [Reserved]
should have a thorough knowledge of at least method will enhance the quality of the 5.0 Safety
the following additional test methods: data obtained from air pollutant
sampling methods. 5.1 Disclaimer. This method may
Method 1, Method 2.
involve hazardous materials, operations,
and equipment. This test method may

VerDate 11<MAY>2000 21:10 Oct 16, 2000 Jkt 194001 PO 00000 Frm 00061 Fmt 4701 Sfmt 4700 E:\FR\FM\17OCR2.SGM pfrm04 PsN: 17OCR2

Você também pode gostar