C R E AT I N I N E ( J A F F E ' S M e t h o d )
Example: If the Absorbance of sample is 0.200 and the Performance Pack Presentation
Absorbance of standard is 0.18. The calculation shall be: Product Code/ KGCRE KGCRE KGCRE
Linearity Limit: 20mg/dl Catalogue No. 101.6.1 101.6.2 101.6.3
Abs of Sample Pack Size * 2X50ml 4X50ml 4X100ml
Creatinine= X Conc. of Std. Precision: Reagent 1 1x50ml 2X50ml 2X100ml
(mg/dl) Abs of Standard (mg/dl) Reagent 2 1X50ml 2X50ml 2X100ml
Within run
Reagent 3 1X2ml 1X3ml 1X3 ml
Control Control 1 Control 2 *Pack size may vary on market demand.
If the creatinine concentration exceeds 20mg/dl, dilute the No. of samples 20 20
sample with normal saline and repeat the assay. The Revision No: (Ver: KGCRE101.6/1)
Mean (mg/dl) 1.32 7.17
reportable results in this case shall be calculated by multiplying Date of Issue: 1st April 2010
the results obtained with dilution factor. S.D. 0.02 0.10
C.V. % 1.74 1.34 Symbols
0.200 Following symbols are used in the labeling of KEE GAD
x 2.0 =2.22 mg/dl
0.180 Between run kits:
REF Catalogue No. Batch No.
Reference value Control Control 1 Control 2
No. of samples 60 60 Read instructions
Serum Mean (mg/dl) 1.32 7.18
CE Mark
Male : 0.6 1.5mg/dl
S.D. 0.02 0.10
Female : 0.5 1.3mg/dl In Vitro Diagnostics Storage temperature
C.V. % 1.70 1.35
Reference range varies from population to population;
Expiry Date CONT Content
therefore, each laboratory should establish its own References
normal range.
1. Gross JL, de Azevedo MJ, Silveiro SP, Canani LH, Caramori CREATININE Product Name Manufactured By
Limitations ML, Zelmanovitz (2005)."Diabetic nephropathy: diagnosis,
prevention, and treatment". Diabetes Care 28 (1): 16476.
1. The reagent and sample volumes can be altered doi:10.2337/ diacare.28.1.164. PMID 15616252.
proportionately so that the sample:reagent ratio remains
same. 2. Delanghe J; De Slypere JP, De Buyzere M, Robbrecht J,
2. Temperature has to be maintained constant throughout the Wieme R, Vermeulen A (Aug 1989). "Normal reference
reaction as the rate of color development is highly values for creatine, creatinine, and carnitine are lower
temperature sensitive. vegetarians" (PDF). Clin. Chem. 35 (8): 18023. PMID
2758659
Quality Control
3. Lamb, E; Newman, D. J. & Price, C. P. 'Kidney Function Tests'
The patient results obtained for each batch can be validated by in Tietz Textbook of clinical chemistry and Molecular Manufactured by:
using normal and abnormal control sera with assayed values Diagnostics. Burtis, C. A., Ashwood, E. R. & Bruns, D. E. KEE GAD Biogen Pvt. Ltd.
for creatinine. (Eds). Elsevier Inc., Fourth Edition. 2006; 24:797-835. A-8, Third Floor, Naraina Industrial Area,
Phase-II, New Delhi-110028 (India)