Testing procedure
Materials provided
Working solutions and calibrators as described above
Additional materials required
Controls as indicated below
o1-Microglobulin Anti-o1-micro- 0.9% NaCl
antigen globulin antibody General laboratory equipment
Absorbance
Assay
Refer to the appropriate operators manual and/or Instrument Settings
section of this insert for analyzer-specific assay instructions. The
performance of applications not validated from Roche is not warranted
and must be defined by the user.
Conc. Calibration
Calibration is performed with Calibrators 3ae via a standard series.
S15: Calibrators 3ae
Antigen/antibody Turbidimetric Key the assigned values for the calibrators into the instrument.
complex determination Calibration frequency
Anti-1-Microglobulin antibodies react with antigen in the sample to Full calibration is recommended:
form an antigen/antibody complex which, after agglutination, can be after lot change
determined turbidimetrically. as required following quality control procedures
Calibration verification: Not necessary
Working solution concentration
R1 Buffer Quality control
Acetate buffer: 35 mmol/l, pH 5.3; PEG; detergent For quality control, use Precinorm 1-Microglobulin, Precipath 1-
R2 Anti-1-microglobulin antibody Microglobulin, or other suitable control material. The control intervals
Polyclonal anti-human 1-microglobulin antibody (sheep), depen- and limits must be adapted to the individual laboratory requirements.
dent on titer; acetate buffer: 50 mmol/l, pH 5.3 Values obtained should fall within established limits. Each laboratory
should establish corrective measures to be taken if values fall outside
the limits.
o1-Microglobulin
Calculation References
Roche/Hitachi systems automatically calculate the 1-microglobulin 1 Kurrle-Weittenhiller A et al. AACC Poster Abstract. Clin Chem
concentration of each sample. 1992;38:10901091.
2 Weber M H, Scholz P, Stibbe W et al. 1-Mikroglobulin im Urin
W Limitations interference6 und Serum bei Proteinurie und Niereninsuffizienz. Klin Wschr
Criterion: Recovery within 10% of initial values. 1985;63:711713.
Icterus: No significant interference up to an I index of 15 (approx. 3 Colombo JP, ed. Klinisch-chemisch Urindiagnostik. Rotkreuz:
conjugated and unconjugated bilirubin concentration: 15 mg/dl or 256 LABOLIFE-Verlagsgemeinschaft, 1994.
mol/l). 4 Guder W, Zawta B. Basiswissen Labordiagnostik Niere (Boehringer
Hemolysis: No significant interference up to an H index of 600 (approx. Mannheim 1994).
hemoglobin concentration: 600 mg/dl or 372 mol/l) 5 Greiling H, Gressner AM, eds. Lehrbuch der Klinischen Chemie
Creatinine: < 0.5 mg/dl does not interfere. und Pathobiochemie, 3rd ed. Stuttgart/New York, 1995.
6 Data on file at Roche.
Carry-over: To avoid probe carry-over to 1-microglobulin when 7 Thomas L, ed. Labor und Diagnose, 4th ed. Marburg: Die Medizin-
performing GGT determinations with GGT Szasz liquid, use the ische Verlagsgesellschaft, 1992:470.
Evasion software with SMS/Acid wash on the Roche/Hitachi 717/902/ 8 Passing H, Bablok W. A New Biometrical Procedure for Testing
904/911/912/914/917/MODULAR P analyzers. Assays on the Roche/ the Equality of Measurements from Two Different Analytical
Hitachi 704 should be run in batch mode. Methods. J Clin Chem Clin Biochem 1983;21:709720.
9 Bablok W et al. A General Regression Procedure for Method
W Measuring/reportable range Transformation. J Clin Chem Clin Biochem 1988;26:783790.
2200 mg/l**
Manually dilute samples with higher concentrations with NaCl (e.g.
1 + 1). Multiply the result by the appropriate dilution factor (e.g. 2).
** The maximum reportable range is dependent on the highest standard concentration.
Method comparison6
A comparison of the 1-microglobulin determination using the Roche
Tina-quant U 1-Microglobulin assay (y) with a nephelometric 1-
microglobulin test (x) in human urine gave the following correlation
(mg/l):
Passing/Bablok8,9 Linear regression
y = -0.535 + 1.002 x y = 2.415 + 0.939 x
r = 0.981 r = 0.981
SD (md 95) = 14.067 Sy.x = 5.235
Number of samples measured: 133
The sample concentrations were between 3.3 and 193.9 mg/l.
1935950001 02 03
1316
o1-Microglobulin
Instrument settings
English 2001-05 1935950001 03 03
Roche/Hitachi 902
US users
Refer to application sheet for additional operating information.
No. <Chemistry>
Roche/Hitachi 914 1 Test Name 1M
Refer to application sheet for parameters.
2 Assay Code (Mthd) 2 Point End
Roche/Hitachi 904, 911, 912, 917, and MODULAR users
3 Assay Code (2. Test) 0
Read in the application parameters from the application diskette,
settings sheet or barcode sheet, as appropriate. 4 Reaction Time 10
5 Assay Point 1 17
Roche/Hitachi 704 6 Assay Point 2 35
Temperature: 37C 7 Assay Point 3 0
8 Assay Point 4 0
PROGRAM 2 CHEMISTRY PARAMETERS 9 Wavelength (SUB) 700
TEST [1M] 10 Wavelength (MAIN) 340
ASSAY CODE [2(2 POINT)] [15] [32] 11 Sample Volume 15
SAMPLE VOLUME [20] 12 R1 Volume 250
13 R1 Pos. .....
R1 VOLUME [350] [20] [NO]
14 R1 Bottle Size Small
R2 VOLUME [70] [20] [NO]
15 R2 Volume 0
WAVELENGTH [700] [340]
16 R2 Pos. 0
CALIB. METHOD [NON-LINEAR] [1] [5] 17 R2 Bottle Size Small
STD. (1) CONC.-POS. [ ---- ] [ ---- ] 18 R3 Volume 50
STD. (2) CONC.-POS. [ ---- ] [ ---- ] 19 R3 Pos. .....
STD. (3) CONC.-POS. [ ---- ] [ ---- ] 20 R3 Bottle Size Small
STD. (4) CONC.-POS. [ ---- ] [ ---- ] 21 Calib. Type (Type) Logit-Log (4P)
STD. (5) CONC.-POS. [ ---- ] [ ---- ] 22 Calib. Type (Wght) 0
STD. (6) CONC.-POS. [0] [0] 23 Calib. Conc. 1 .....
UNIT [ ---- ] 24 Calib. Pos. 1 .....
SD LIMIT [200] 25 Calib. Conc. 2 .....
DUPLICATE LIMIT [300] 26 Calib. Pos. 2 .....
SENSITIVITY LIMIT [5000] 27 Calib. Conc. 3 .....
ABS. LIMIT (INC/DEC) [0] [INCREASE] 28 Calib. Pos. 3 .....
PROZONE LIMIT [32 000] [UPPER] 29 Calib. Conc. 4 .....
EXPECTED VALUE [ ---- ] [ ---- ] 30 Calib. Pos. 4 .....
INSTRUMENT FACTOR [1.00] 31 Calib. Conc. 5 .....
32 Calib. Pos. 5 .....
---- Data entered by the operator 33 Calib. Conc. 6 0
34 Calib. Pos. 6 0
Roche/Hitachi 717 35 S1 ABS. 0
Temperature: 37C 36 K Factor 10000
PROGRAM 2 CHEMISTRY PARAMETERS 37 K2 Factor 10000
38 K3 Factor 10000
TEST [1M] 39 K4 Factor 10000
ASSAY CODE [2(2 POINT)] [24][50] 40 K5 Factor 10000
SAMPLE VOLUME [15] [4] 41 A Factor 0
R1 VOLUME [250] [20] [NO] 42 B Factor 0
R2 VOLUME [50] [20] [NO] 43 C Factor 0
WAVELENGTH [700] [340] 44 SD Limit 200
CALIB. METHOD [NON-LINEAR] [1] [5] 45 Duplicate Limit 300
STD. (1) CONC.-POS. [ ---- ] [ ---- ] 46 Sens. Limit 5000
STD. (2) CONC.-POS. [ ---- ] [ ---- ] 47 S1ABS. Limit (L) -32000
STD. (3) CONC.-POS. [ ---- ] [ ---- ] 48 S1ABS. Limit (H) 32000
STD. (4) CONC.-POS. [ ---- ] [ ---- ] 49 ABS. Limit 0
STD. (5) CONC.-POS. [ ---- ] [ ---- ] 50 ABS. Limit (D/I) Increase
STD. (6) CONC.-POS. [0] [0] 51 Prozone Limit 32000
SD LIMIT [200] 52 Proz Limit (Upp/Low) Upper
53 Prozone (End Point) 35
DUPLICATE LIMIT [300]
54 Expect. Value (L) .....
SENSITIVITY LIMIT [4000]
55 Expect. Value (H) .....
ABS. LIMIT (INC/DEC) [0] [INCREASE]
56 Instr. Fact. (a) 1
PROZONE LIMIT [32 000] [UPPER] 57 Instr. Fact. (b) 0
EXPECTED VALUE [ ---- ] [ ---- ] 58 Key Setting .....
PANIC VALUE [ ---- ] [ ---- ]
INSTRUMENT FACTOR [1.00] .... Data entered by the operator
---- Data entered by the operator
For detailed information, consult the operator manuals for Roche/Hitachi systems, the
respective application sheets and the package inserts for the control sera.
Tina-quant, Precinorm and Precipath are trademarks of a member of the Roche Group.
Roche Diagnostics GmbH, D-68298 Mannheim
W = Significant additions or changes Roche Diagnostics Corporation, Indianapolis, IN, USA
2001 Roche Diagnostics US Customer Technical Support 1-800-428-2336 2001-05
o1-Microglobulin
1935950001 04 03
1317
o1-Microglobulin
Tina-quantQ U 1-Microglobulin
Deutsch 2001-05 1935950001 05 03
Packungsgren fr Roche/Hitachi-Gerte
704 717 747 902 904 911 914 917 MODULAR W
Best.-Nr. Flasche Inhalt
912 P D
1360566 1 Puffer, 2 x 20 ml
2 1-Microglobulin-Antikrper, 2 x 5 ml
3ae Kalibratoren, 5 Flaschen, je 2 ml
Nicht alle Packungen und Gerte sind in jedem Land verfgbar. Fr weitere gertespezifische Arbeitsanleitungen wenden Sie sich bitte an Ihre Roche-Vertretung.
Testverfahren
Gelieferte Materialien
Gebrauchsfertige Lsungen und Kalibratoren wie vorher angegeben
o1-Microglobulin- o1-Microglobulin- Zustzlich bentigte Materialien
Antigen Antikrper Kontrollmaterial wie nachfolgend beschrieben
Natriumchlorid-Lsung (0,9%)
Allgemein bliche Laborausrstung
Extinktion
Testdurchfhrung
Anleitungen zur Testdurchfhrung finden Sie in den gertespezifischen
Anweisungen des entsprechenden Handbuchs und/oder unter den
Gerteeinstellungen dieser Packungsbeilage. Fr Arbeitsanleitungen,
die nicht von Roche validiert wurden, wird keine Gewhr bernommen.
Konz. Kalibrierung
Die Kalibrierung erfolgt mit den Kalibratoren 3ae ber eine Standardreihe.
S15: Kalibratoren 3ae
Antigen-Antikrper- Turbidimetrische Die Sollwerte der Kalibratoren gibt man in das Gert ein.
Komplex Messung Kalibrationshufigkeit
Eine Vollkalibration wird empfohlen:
1-Microglobulin-Antikrper reagieren mit dem Antigen aus der Probe Bei Chargenwechsel
unter Bildung eines Antigen-Antikrper-Komplexes, der nach Agglutination Wenn Qualittskontrollverfahren dies erfordern
turbidimetrisch gemessen wird. Kalibrationsverifikation: Nicht erforderlich
o1-Microglobulin
Qualittskontrolle Methodenvergleich6
Zur Qualittskontrolle Precinorm 1-Microglobulin, Precipath 1- Ein Vergleich der 1-Mikroglobulinbestimmung mit dem Reagenz
Microglobulin oder anderes geeignetes Kontrollmaterial einsetzen. Tina-quant U 1-Microglobulin (y) ergab mit einem nephelome-
Die Kontrollintervalle und Kontrollgrenzen sind den individuellen An- trischen 1-Microglobulin-Test (x) in Humanurin folgende Korre-
forderungen jedes Labors anzupassen. Die Ergebnisse mssen in- lationen (mg/l):
nerhalb der festgesetzten Bereiche liegen. Jedes Labor sollte Passing/Bablok8,9 Lineare Regression
Korrekturmanahmen fr den Fall beschreiben, da Werte auerhalb y = -0,535 + 1,002 x y = 2,415 + 0,939 x
des Bereichs liegen. r = 0,981 r = 0,981
SD (ma 95) = 14,067 Sy.x = 5,235
Berechnung Anzahl gemessener Proben: 133
Die Roche/Hitachi-Gerte berechnen automatisch die 1-Micro- Die Probenkonzentrationen lagen zwischen 3,3 und 193,9 mg/l
globulinkonzentration jeder Probe.
Literatur
W Einschrnkungen des Verfahrens Interferenzen6 1 Kurrle-Weittenhiller A et al. AACC Poster Abstract. Clin Chem
Als Bewertung gilt: Wiederfindung 10 % vom Ausgangswert. 1992;38:10901091.
Ikterus: Keine wesentliche Beeinflussung bis zum Index I von 15 2 Weber MH, Scholz P, Stibbe W et al. 1-Mikroglobulin im Urin
(ca.15 mg/dl bzw. 256 mol/l konjugiertes und unkonjugiertes Biliru- und Serum bei Proteinurie und Niereninsuffizienz. Klin Wschr
bin). 1985;63:711713.
Hmolyse: Keine wesentliche Beeinflussung bis zum Index H von 600 3 Colombo JP (Hrsg.). Klinisch-chemisch Urindiagnostik. Rotkreuz:
(ca. 600 mg/dl bzw. 372 mol/l Hmoglobin). LABOLIFE-Verlagsgemeinschaft, 1994.
Creatinin < 0,5 mg/dl strt nicht. 4 Guder W, Zawta B. Basiswissen Labordiagnostik Niere (Hrsg.
Verschleppung: Um bei GGT-Bestimmungen mit GGT Szasz liquid Boehringer Mannheim 1994).
eine Nadelverschleppung auf 1-Microglobulin zu vermeiden, ist auf 5 Greiling H, Gressner AM (Hrsg.). Lehrbuch der Klinischen Chemie
den Roche/Hitachi 717/902/904/911/912/914/917/MODULAR P Gerten und Pathobiochemie, 3. Auflage. Stuttgart/New York, 1995.
die Evasion Software mit SMS/Acid wash einzusetzen. Bestimmungen 6 Dokumentation Roche.
auf dem Roche/Hitachi 704 sollten in Serie (Batch) durchgefhrt 7 Thomas L (Hrsg.). Labor und Diagnose, 4. Auflage. Marburg: Die
werden. Medizinische Verlagsgesellschaft, 1992:470.
8 Passing H, Bablok W. A New Biometrical Procedure for Testing
the Equality of Measurements from Two Different Analytical
W Mebereich Methods. J Clin Chem Clin Biochem 1983;21:709720.
Mebereich: 2200 mg/l**
Proben mit hheren Konzentrationen werden manuell mit Natrium- 9 Bablok W et al. A General Regression Procedure for Method
chlorid-Lsung (0,9%) verdnnt (z. B. 1+1). Das Ergebnis mit dem Transformation. J Clin Chem Clin Biochem 1988;26:783790.
entsprechenden Verdnnungsfaktor multiplizieren (z. B. Faktor 2).
**Abhngig von der hchsten angegebenen Standardkonzentration
Vorlufiger Referenzbereich
2. Morgenurin7: < 14 mg/g Creatinin bzw.
< 1,58 g/mol Creatinin
Sammelurin: < 12 mg/l
< 20 mg/24 h
Jedes Labor sollte die bertragbarkeit der Referenzbereiche fr die
eigenen Patientengruppen berprfen und gegebenenfalls eigene
Referenzwerte ermitteln. Fr diagnostische Zwecke sind die 1-
Microglobulinergebnisse stets im Zusammenhang mit der Anamne-
se, der klinischen Untersuchung und anderen Untersuchungsergeb-
nissen zu werten.
1935950001 06 03
1318
o1-Microglobulin
Gerteeinstellungen Roche/Hitachi 902
Deutsch 2001-05 1935950001 07 03
US-Anwender
Weitere gertespezifische Informationen entnehmen Sie der Bedie- Zeile <Eingabe des Parameters>
nungsanleitung und dem Applikationsblatt. 1 Test Name 1M
Roche/Hitachi 914 2 Messmethode 2 Punkt End
Die Angaben ber die Gerteeinstellung entnehmen Sie der gerte- 3 Twin 0
spezifischen Anleitung. 4 Reaktionszeit 10
Roche/Hitachi 904/911/912/917/MODULAR 5 Mepunkt 1 17
Die Angaben ber die Gerteeinstellung entnehmen Sie der jeweiligen 6 Mepunkt 2 35
Applikationsdiskette, dem Gerteeinstellungsblatt oder dem Barcode. 7 Mepunkt 3 0
8 Mepunkt 4 0
Roche/Hitachi 704 9 Nebenwellenlnge 700
Temperatur: 37C 10 Hauptwellenlnge 340
11 Probenvolumen 15
PROGRAMM 2 CHEMISCHE PARAMETER 12 Volumen R 1 250
TEST [1M] 13 Position R 1 .....
ASSAY CODE [2(2 PUNKT)] [15] [32] 14 Flaschengre R 1 Klein
PROBENVOLUMEN [20] 15 Volumen R 2 0
R1 VOLUMEN [350] [20] [NO] 16 Position R 2 0
R2 VOLUMEN [70] [20] [NO] 17 Flaschengre R 2 Klein
WELLENLAENGE [700] [340] 18 Volumen R 3 50
19 Position R 3 .....
KALIB. ART [NICHTLINEAR] [1] [5]
20 Flaschengre R 3 Klein
STD. (1) CONC.-POS. [ ---- ] [ ---- ]
21 Kalibrationsart Logit-Log(4P)
STD. (2) CONC.-POS. [ ---- ] [ ---- ]
22 Wichtung 0
STD. (3) CONC.-POS. [ ---- ] [ ---- ] 23 Konz. Standard 1 .....
STD. (4) CONC.-POS. [ ---- ] [ ---- ] 24 Position Standard 1 .....
STD. (5) CONC.-POS. [ ---- ] [ ---- ] 25 Konz. Standard 2 .....
STD. (6) CONC.-POS. [0] [0] 26 Position Standard 2 .....
EINHEIT [ ---- ] 27 Konz. Standard 3 .....
S-LIMIT [200] 28 Position Standard 3 .....
ABW.-LIMIT [300] 29 Konz. Standard 4 .....
EMPFINDL.-LIMIT [5000] 30 Position Standard 4 .....
EXT. LIMIT (ST/FA) [0] [STEIGEND] 31 Konz. Standard 5 .....
PROZONE LIMIT [32 000] [UEBER] 32 Position Standard 5 .....
NORMALBEREICH [ ---- ] [ ---- ] 33 Konz. Standard 6 0
INSTRUMENT FAKTOR [1.00] 34 Position Standard 6 0
35 S1 ABS 0
---- Dateneingabe durch den Anwender
36 K Faktor 10000
37 K 2 Faktor 10000
Roche/Hitachi 717 38 K 3 Faktor 10000
Temperatur: 37C 39 K 4 Faktor 10000
40 K 5 Faktor 10000
PROGRAMM 2 CHEMIE PARAMETER
41 A Faktor 0
TEST [1M] 42 B Faktor 0
MESS METHODE [2(2 PUNKT)] [24][50] 43 C Faktor 0
PROBEVOLUMEN [15] [4] 44 S-Grenze 200
R1 VOLUMEN [250] [20] [NO] 45 Abweichungsgrenze 300
R2 VOLUMEN [50] [20] [NO] 46 Empfindlichkeitsgrenze 5000
WELLENLAENGEN [700] [340] 47 Untere S1 Ext. Grenze -32000
KALIBRATION [NICHTLINEAR] [1] [5] 48 Obere S1 Ext. Grenze 32000
STD. (1) KONZ.-POS. [ ---- ] [ ---- ] 49 Ext. Grenze Wert 0
50 Ext. Grenze fallend/steigend Steigend
STD. (2) KONZ.-POS. [ ---- ] [ ---- ]
51 Prozonengrenze Wert 32000
STD. (3) KONZ.-POS. [ ---- ] [ ---- ]
52 Prozonengrenze unter/ober Obere Grenze
STD. (4) KONZ.-POS. [ ---- ] [ ---- ]
53 Proz. grenze (Endpunkt) 35
STD. (5) KONZ.-POS. [ ---- ] [ ---- ]
54 Unterer Referenzwert .....
STD. (6) KONZ.-POS. [0] [0] 55 Oberer Referenzwert .....
SD GRENZE [200] 56 Instr. Faktor (a) 1
ABWEICHUNGSGRENZE [300] 57 Instr. Faktor (b) 0
EMPFINDL. GRENZE [4000] 58 Methodentaste .....
EXT. GR. (STEI/FALL) [0] [STEIGEND]
PROZONE GRENZE [32 000] [UEBER] .... Dateneingabe durch den Anwender
NORMALBEREICH [ ---- ] [ ---- ]
ALARMBEREICH [ ---- ] [ ---- ] Weitergehende Informationen siehe Bedienungshandbcher fr Roche/Hitachi-Gerte,
die gertespezifischen Applikationsbltter sowie die Packungsbeilage der Kontrollseren.
GERAETEFAKTOR [1.00]
---- Dateneingabe durch den Anwender W = Signifikante Ergnzungen oder nderungen
1935950001 08 03
1319
o1-Microglobulin
Tina-quantQ U 1-Microglobuline
Franais 2001-05 1935950001 09 03
Rf. Flacon Contenu 704 717 747 902 904 911 914 917 MODULAR
912 P D
1 360 566 1 Tampon, 2 x 20 ml
2 Anticorps anti-1-microglobuline, 2 x 5 ml
3ae Calibrateurs, 5 x 2 ml
Tous les appareils ne sont pas disponibles dans tous les pays. Pour lanalyse sur dautres analyseurs automatiques veuillez vous adresser la socit distributrice
des produits Roche de votre pays.
Performances analytiques
Les rsultats indiqus ci-dessous ont t obtenus avec un appareil
Roche/Hitachi (les rsultats obtenus au laboratoire peuvent diffrer
de ceux-ci):
Prcision6
La reproductibilit a t dtermine laide durines humaines et de
solution de contrle selon un protocole interne (n= 21). Les rsultats
suivants ont t obtenus:
Prcision intra-srie Prcision inter-srie
Echantillon E s CV E s CV
mg/l mg/l % mg/l mg/l %
Urine humaine I 16,9 0,99 5,9 18,6 1,40 7,5
Urine humaine II 45,2 1,15 2,5 28,9 1,29 4,5
Solution de contrle 19,3 0,82 4,3 19,8 1,17 5,9
1935950001 10 03
1320
o1-Microglobulin
Franais 2001-05 1935950001 11 03
1935950001 12 03
1321
o1-Microglobulin
Tina-quantQ U 1-Microglobulina
Espaol 2001-05 1935950001 13 03
Calibracin
La calibracin se realiza con los calibradores 3ae a travs de una
serie standard.
Conc. S15: Calibradores 3ae
Los valores tericos de los calibradores se introducen en el analizador.
Complejo Medicin Intervalo de calibraciones
antgeno-anticuerpo turbidimtrica Se recomienda una calibracin completa:
al cambiar el lote
si es necesaria para el control de calidad
Los anticuerpos anti-1-microglobulina reaccionan con el antgeno No se requiere ninguna verificacin de la calibracin.
de la muestra formando un complejo antgeno-anticuerpo que se
mide turbidimtricamente despus de la aglutinacin.
o1-Microglobulin
Control de calidad Comparacin de mtodos6
Emplear Precinorm 1-Microglobulina, Precipath 1-Microglobulina La comparacin entre la determinacin de 1-microglobulina con el
u otro material de control apropiado para el control de calidad. Los reactivo Tina-quant U 1-Microglobulina (y) y una determinacin
intervalos y lmites del control han de adaptarse a los requisitos nefelomtrica de 1-microglobulina (x) en orina humana ha dado las
individuales del laboratorio. Los resultados deben hallarse dentro de correlaciones siguientes (mg/l):
los intervalos definidos. Cada laboratorio debera detallar sus propias Passing/Bablok8,9 Regresin lineal
medidas de correccin en caso de valores fuera del intervalo indicado. y = -0,535 + 1,002 x y = 2,415 + 0,939 x
r = 0,981 r = 0,981
Clculo DE (dm) = 14,067 Sy.x = 5,235
Los analizadores Roche/Hitachi calculan automticamente la con- Nmero de muestras medidas: 133
centracin de 1-microglobulina de cada muestra. Concentraciones de las muestras: entre 3,3 y 193,9 mg/l
W Limitaciones del anlisis interferencias6 Bibliografa
Criterio: recuperacin 10% del valor inicial. 1 Kurrle-Weittenhiller A y cols. AACC Poster Abstract. Clin Chem
Ictericia: sin interferencias significativas por bilirrubina conjugada y 1992; 38:1090-1091.
no conjugada hasta aprox. 15 mg/dl o 256 mol/l (lo que corresponde, 2 Weber MH, Scholz P, Stibbe W y cols. a1-Mikroglobulin im Urin
en los EEUU, al ndice I de 15). und Serum bei Proteinurie und Niereninsuffizienz. Klin Wschr
Hemlisis: sin interferencias significativas por hemoglobina hasta 1985;63:711-713.
aprox. 600 mg/dl o 372 mol/l (lo que corresponde, en los EEUU, al 3 Colombo JP (ed.). Klinisch-chemisch Urindiagnostik. Rotkreuz:
ndice H de 600). LABOLIFE-Verlagsgemeinschaft, 1994.
Creatinina < 0,5 mg/dl no interfiere en el test. 4 Guder W, Zawta B. Basiswissen Labordiagnostik Niere (eds.
Arrastre: Para evitar el arrastre a las pipetas utilizadas para las Boehringer Mannheim 1994).
determinaciones de 1-microglobulina despus de efectuar deter- 5 Greiling H, Gressner AM (eds.). Lehrbuch der Klinischen Chemie
minaciones de GGT con GGT Szasz liquid, debe usarse el software und Pathobiochemie, 3a edicin. Stuttgart/Nueva York, 1995.
Evasion con SMS/Acid wash en los analizadores Roche/Hitachi 717/ 6 Documentacin de Roche.
902/904/911/912/914/917/MODULAR P. Las determinaciones en el 7 Thomas L (ed.). Labor und Diagnose, 4 a edicin. Marburgo: Die
Roche/Hitachi 704 deben realizarse en serie (batch). Medizinische Verlagsgesellschaft, 1992;470.
8 Passing H, Bablok W. A New Biometrical Procedure for Testing
W Intervalo de medicin the Equality of Measurements from Two Different Analytical Methods.
Intervalo de medicin: 2200 mg/l** J Clin Chem Clin Biochem 1983;21:709720.
Las muestras con concentraciones superiores se diluyen manualmente 9 Bablok W y cols. A General Regression Procedure for Method
con solucin de cloruro sdico al 0,9% (p.ej. 1+1). Multiplicar el Transformation. J Clin Chem Clin Biochem 1988;26:783790.
resultado por el correspondiente factor de dilucin (p.ej. factor 2).
**Dependiente de la ms alta concentracin standard indicada
6
Sensibilidad analtica (lmite de deteccin inferior)
Lmite de deteccin: 2 mg/l
El lmite de deteccin inferior equivale a la concentracin mnima
medible de 1-microglobulina que puede distinguirse de cero. Se
calcula a partir de las tres desviaciones estndares de 21 mediciones
del estndar ms bajo.
1935950001 14 03
1322
o1-Microglobulin
Espaol 2001-05 1935950001 15 03
1935950001 16 03
1323
o1-Microglobulin
Tina-quantQ U 1-Microglobulina
Italiano 2001-05 1935950001 17 03
Precisione6
La ripoducibilit stata determinata con campioni umani e con con-
trolli, eseguiti in base ad un protocollo interno (n = 21), con i seguenti
risultati:
Prec. nella serie Prec. fra le serie
Campione Media DS %CV Media DS %CV
mg/l mg/l mg/l mg/l
Urina umana I 16,9 0,99 5,9 18,6 1,40 7,5
Urina umana II 45,2 1,15 2,5 28,9 1,29 4,5
Siero di controllo 19,3 0,82 4,3 19,8 1,17 5,9
Sensibilit analitica6
Limite di sensibilit: 2 mg/l
Il limite di sensibilit inferiore rappresenta la concentrazione minima
di 1-microglobulina misurabile che pu essere distinta dallo zero.
Viene calcolato come 3 deviazioni standard di 21 replicati dello
standard pi basso.
1935950001 18 03
1324
o1-Microglobulin
Italiano 2001-05 1935950001 19 03
1935950001 20 03
1325
o1-Microglobulin
Tina-quantQ U 1-Microglobulina
Portugus 2001-05 1935950001 21 03
Produto registado no INFARMED Indica os analisadores Roche/Hitachi em que os dispositivos podem ser utilizados
Componentes do teste
Material fornecido
Solues de trabalho conforme descritas acima
Outros materiais necessrios
Antignio o1- Anticorpo anti-o1- Controlos conforme indicado abaixo
microglobulina microglobulina NaCI a 0,9%
Absorvncia Equipamento normal de laboratrio
Realizao do ensaio
Consulte o manual do operador apropriado e/ou a seco relativa s
definies do analisador neste folheto informativo para obter instru-
es mais especficas sobre o analisador. Quando se executam en-
saios no validados pela Roche, esta no garante os resultados, pelo
Conc. que esses ensaios devem ser definidos pelo utilizador.
Calibrao
Complexo Leitura A calibrao realizada com os Calibradores 3ae atravs de uma
antignio/anticorpo turbidimtrica srie de padres.
Os anticorpos anti-1-microglobulina reagem com o antignio na S15: Calibradores 3ae
amostra e formam um complexo antignio/anticorpo. Aps a aglu- Introduza no analisador os valores atribudos aos calibradores.
tinao, a determinao feita por turbidimetria. Frequncia das calibraes
Recomenda-se a realizao de uma calibrao completa:
Concentrao da soluo de trabalho aps mudana de lote
R1 Tampo conforme necessrio, de acordo com os procedimentos de controlo
Tampo acetato: 35 mmol/l, pH 5,3; PEG; detergente de qualidade
R2 Anticorpo anti-1-microglobulina Verificao da calibrao: no necessria.
Anticorpo policlonal (carneiro) anti-1-microglobulina humana,
dependente do ttulo; tampo acetato: 50 mmol/l, pH 5,3
264
o1-Microglobulin
Controlo de qualidade Passing/Bablok8,9 Regresso linear
Para o controlo de qualidade, utilize o Precinorm 1-Microglobulin, y = -0,535 + 1,002 x y = 2,415 + 0,939 x
o Precipath 1-Microglobulin ou outros materiais de controlo adequa- r = 0,981 r = 0,981
dos. Os intervalos e os limites de controlo devero ser adaptados SD (md) = 14,067 Sy.x = 5,235
s exigncias especficas de cada laboratrio e aos requisitos espec- Nmero de amostras medidas: 133
ficos de cada pas. Os valores obtidos devem situar-se dentro dos As concentraes das amostras variaram entre aprox. 3,3 e 193,9 mg/l.
limites estabelecidos. Cada laboratrio dever estabelecer as suas
prprias normas no que diz respeito s medidas correctivas a tomar Bibliografia
no caso de os valores se situarem fora dos limites. 1 Kurrle-Weittenhiller A et al. AACC Poster Abstract. Clin Chem
1992;38:10901091.
Clculo 2 Weber M H, Scholz P, Stibbe W et al. 1-Mikroglobulin im Urin
Os analisadores Roche/Hitachi calculam automaticamente a concen- und Serum bei Proteinurie und Niereninsuffizienz. Klin Wschr
trao de 1-microglobulina. 1985;63:711713.
3 Colombo JP, ed. Klinisch-chemisch Urindiagnostik. Rotkreuz:
W Limitaes - interferncias6 LABOLIFE-Verlagsgemeinschaft, 1994.
Critrio: recuperao dentro de 10% do valor inicial. 4 Guder W, Zawta B. Basiswissen Labordiagnostik Niere (Boehringer
Ictercia: Nenhuma interferncia significativa at a um ndice I de 15 Mannheim 1994).
(concentrao aprox. de bilirrubina conjugada e no-conjugada: 15 5 Greiling H, Gressner AM, eds. Lehrbuch der Klinischen Chemie
mg/dl ou 256 mol/l). ). und Pathobiochemie, 3rd ed. Stuttgart/New York, 1995.
Hemlise: Nenhuma interferncia significativa at a um ndice H de 6 Documentao da Roche.
600 (concentrao aprox. de hemoglobina: 600 mg/dl ou 372 mol/l). 7 Thomas L, ed. Labor und Diagnose, 4th ed. Marburg: Die Medizin-
ische Verlagsgesellschaft, 1992:470.
Creatinina: < 0,5 mg/dl no causa interferncias no teste. 8 Passing H, Bablok W. A New Biometrical Procedure for Testing
Contaminao (carry-over): Para evitar a contaminao da 1- the Equality of Measurements from Two Different Analytical
microglobulina pela sonda ao efectuar as determinaes da GGT Methods. J Clin Chem Clin Biochem 1983;21:709720.
com GGT Szasz (lquido), utilize o software Evasion com SMS/Acid 9 Bablok W et al. A General Regression Procedure for Method
Wash nos analisadores Roche/Hitachi 717/902/904/911/912/914/ Transformation. J Clin Chem Clin Biochem 1988;26:783790.
917/MODULAR P. Os ensaios no Roche/Hitachi 704 devem ser
efectuados em modo de lote (batch).
W Intervalo de medio/referncia
2200 mg/l**
Dilua manualmente com NaCI as amostras com concentraes su-
periores (p. ex., 1 + 1). Multiplique o resultado pelo factor de diluio
adequado (p. ex. 2).
**O intervalo mximo registado depende da concentrao padro mais elevada.
1935950001 22 03
462
o1-Microglobulin
Definies do analisador
Portugus 2001-05 1935950001 23 03
Roche/Hitachi 902
Utilizadores dos EUA
Para mais informaes sobre o funcionamento, consulte a folha da
aplicao. No. <Chemistry>
1 Test Name 1M
Clientes do Roche/Hitachi 914
Para mais informaes sobre os parmetros, consulte a folha da 2 Assay Code (Mthd) 2 Point End
aplicao. 3 Assay Code (2. Test) 0
Utilizadores do Roche/Hitachi 904, 911, 912, 917 e MODULAR 4 Reaction Time 10
Introduza os parmetros da aplicao a partir da disquete de apli- 5 Assay Point 1 17
cao, da folha com os cdigos de barras ou da folha das defi-nies, 6 Assay Point 2 35
conforme adequado. 7 Assay Point 3 0
8 Assay Point 4 0
Roche/Hitachi 704 9 Wavelength (SUB) 700
Temperatura: 37C 10 Wavelength (MAIN) 340
PROGRAM 2 CHEMISTRY PARAMETERS 11 Sample Volume 15
12 R1 Volume 250
TEST [1M] 13 R1 Pos. .....
ASSAY CODE [2(2 POINT)] [15] [32] 14 R1 Bottle Size Small
SAMPLE VOLUME [20] 15 R2 Volume 0
R1 VOLUME [350] [20] [NO] 16 R2 Pos. 0
R2 VOLUME [70] [20] [NO] 17 R2 Bottle Size Small
WAVELENGTH [700] [340] 18 R3 Volume 50
CALIB. METHOD [NON-LINEAR] [1] [5] 19 R3 Pos. .....
STD. (1) CONC.-POS. [ ---- ] [ ---- ] 20 R3 Bottle Size Small
STD. (2) CONC.-POS. [ ---- ] [ ---- ] 21 Calib. Type (Type) Logit-Log (4P)
STD. (3) CONC.-POS. [ ---- ] [ ---- ] 22 Calib. Type (Wght) 0
STD. (4) CONC.-POS. [ ---- ] [ ---- ] 23 Calib. Conc. 1 .....
STD. (5) CONC.-POS. [ ---- ] [ ---- ] 24 Calib. Pos. 1 .....
STD. (6) CONC.-POS. [0] [0] 25 Calib. Conc. 2 .....
UNIT [ ---- ] 26 Calib. Pos. 2 .....
SD LIMIT [200] 27 Calib. Conc. 3 .....
28 Calib. Pos. 3 .....
DUPLICATE LIMIT [300]
29 Calib. Conc. 4 .....
SENSITIVITY LIMIT [5000]
30 Calib. Pos. 4 .....
ABS. LIMIT (INC/DEC) [0] [INCREASE]
31 Calib. Conc. 5 .....
PROZONE LIMIT [32 000] [UPPER] 32 Calib. Pos. 5 .....
EXPECTED VALUE [ ---- ] [ ---- ] 33 Calib. Conc. 6 0
INSTRUMENT FACTOR [1.00] 34 Calib. Pos. 6 0
35 S1 ABS. 0
---- Dados introduzidos pelo operador
36 K Factor 10000
Roche/Hitachi 717 37 K2 Factor 10000
Temperatura: 37C 38 K3 Factor 10000
39 K4 Factor 10000
PROGRAM 2 CHEMISTRY PARAMETERS 40 K5 Factor 10000
41 A Factor 0
TEST [1M] 42 B Factor 0
ASSAY CODE [2(2 POINT)] [24][50] 43 C Factor 0
SAMPLE VOLUME [15] [4] 44 SD Limit 200
R1 VOLUME [250] [20] [NO] 45 Duplicate Limit 300
R2 VOLUME [50] [20] [NO] 46 Sens. Limit 5000
WAVELENGTH [700] [340] 47 S1ABS. Limit (L) -32000
CALIB. METHOD [NON-LINEAR] [1] [5] 48 S1ABS. Limit (H) 32000
STD. (1) CONC.-POS. [ ---- ] [ ---- ] 49 ABS. Limit 0
STD. (2) CONC.-POS. [ ---- ] [ ---- ] 50 ABS. Limit (D/I) Increase
STD. (3) CONC.-POS. [ ---- ] [ ---- ] 51 Prozone Limit 32000
STD. (4) CONC.-POS. [ ---- ] [ ---- ] 52 Proz Limit (Upp/Low) Upper
STD. (5) CONC.-POS. [ ---- ] [ ---- ] 53 Prozone (End Point) 35
STD. (6) CONC.-POS. [0] [0] 54 Expect. Value (L) .....
SD LIMIT [200] 55 Expect. Value (H) .....
DUPLICATE LIMIT [300] 56 Instr. Fact. (a) 1
57 Instr. Fact. (b) 0
SENSITIVITY LIMIT [4000]
58 Key Setting .....
ABS. LIMIT (INC/DEC) [0] [INCREASE]
PROZONE LIMIT [32 000] [UPPER]
.... Dados introduzidos pelo operador
EXPECTED VALUE [ ---- ] [ ---- ]
PANIC VALUE [ ---- ] [ ---- ]
INSTRUMENT FACTOR [1.00]
Fabricado por:
Dados introduzidos pelo operador Roche Diagnostics GmbH, D-68298 Mannheim
Para mais informaes, consulte os manuais do operador dos sistemas Roche/Hitachi, Distribuidor em Portugal:
as folhas da aplicao respectiva e os folhetos informativos dos soros de controlo. Roche Farmacutica Qumica, Lda, 2700 Amadora 2001-05
265
o1-Microglobulin
1935950001 24 03
562