MEDIDAS COM RESISTORES

Ana Paula Mattos Costa


Matrcula 201610004714
Fsica Geral - Laboratrio
Turma 4

2016.1
1. Objetivo

A prtica consiste na medio de resistores atravs da utilizao de


instrumentos digitais e analgicos visando estimar o valor esperado e a definir a
compatibilidade e a discrepncia das medidas diretas de uma grandeza fsica
(resistncia).

2. Material Utilizado

Para a realizao da prtica, foram utilizados os seguintes materiais:

1. Conjunto com 110 resistores de mesmo valor;


2. Um multmetro digital;
3. Um multmetro analgico.

Figura 1: Resistores (esq.), multmetro digital (centro) e multmetro analgico (dir.).

3. Procedimento e Dados Obtidos

Primeiramente, foi obtido o valor nominal do conjunto de resistores atravs do


Cdigo de Cores. Por meio deste, o valor da resistncia identificado pela
comparao das cores das faixas presentes no corpo do resistor com uma tabela de
cores (figura 2).

O valor nominal foi obtido para um conjunto de resistores padro contendo


quatro faixas, sendo a primeira e a segunda faixas as responsveis pela identificao
do primeiro e do segundo dgitos do valor da resistncia, respectivamente. A terceira
faixa, chamada de multiplicador, indica a ordem de grandeza do valor. A quarta faixa,
a da tolerncia, dada em porcentagem e indica a preciso do valor da resistncia em
relao ao valor obtido atravs do Cdigo de Cores (Figura 2).

Figura 2: Cdigo de cores de resistores.

O conjunto de resistores utilizados apresentava as faixas nas seguintes cores:


1 faixa: vermelha; 2 faixa: vermelha; 3 faixa: laranja; 4 faixa: dourada.

De acordo com o Cdigo de Cores, identificou-se uma resistncia de valor


nominal 22 x 1k com uma tolerncia de 5%. Ento:

Rnominal = (22,0 1,1)k

Aps a identificao do valor nominal, foram medidos os valores das


resistncias de cada um dos 110 resistores utilizando-se o multmetro digital como
ohmmetro, tocando os terminais dos resistores com os fios ligados ao aparelho. Pelo
valor nominal obtido, a chave seletora do multmetro precisou ser colocada na posio
200k, de maneira a permitir a leitura das resistncias. O erro estabelecido pelo
fabricante do multmetro digital foi de 0,8% nessa faixa de leitura.

A Tabela 1 apresenta as medidas que obtivemos e sua distribuio pode ser


vista na Figura 3.

Tabela 1: Medidas de resistncia obtidas atravs da utilizao do multmetro digital.


Resistor Ri Resistor Ri Resistor Ri Resistor Ri Resistor Ri
(i) (k) (i) (k) (i) (k) (i) (k) (i) (k)
1 21,5 23 21,5 45 21,2 67 21,6 89 21,6
2 21,6 24 21,6 46 21,6 68 21,6 90 21,6
3 21,7 25 21,6 47 21,6 69 21,6 91 21,6
4 21,7 26 21,6 48 21,6 70 21,6 92 21,6
5 21,7 27 21,6 49 21,6 71 21,6 93 21,6
6 21,7 28 21,6 50 21,6 72 21,7 94 21,7
7 21,7 29 21,7 51 21,7 73 21,7 95 21,7
8 21,7 30 21,7 52 21,7 74 21,7 96 21,7
9 21,7 31 21,7 53 21,7 75 21,7 97 21,7
10 21,7 32 21,7 54 21,7 76 21,7 98 21,7
11 21,7 33 21,7 55 21,7 77 21,7 99 21,7
12 21,7 34 21,7 56 21,7 78 21,7 100 21,7
13 21,7 35 21,8 57 21,7 79 21,7 101 21,8
14 21,7 36 21,8 58 21,7 80 21,7 102 21,8
15 21,7 37 21,8 59 21,7 81 21,8 103 21,8
16 21,8 38 21,8 60 21,7 82 21,8 104 21,8
17 21,8 39 21,8 61 21,7 83 21,8 105 21,8
18 21,8 40 21,8 62 21,8 84 21,8 106 21,8
19 21,8 41 21,8 63 21,8 85 21,8 107 21,9
20 21,8 42 21,8 64 21,8 86 21,8 108 21,9
21 21,8 43 21,9 65 21,8 87 21,8 109 21,9
22 21,9 44 22,1 66 21,8 88 22,0 110 21,9

Figura 3: Histograma - Conjunto de 110 resistores. Medio com multmetro digital.


O conjunto de medidas foi dividido em cinco grupos de 22 resistores e, para
cada um deles, fez-se os histogramas (Figuras 4 a 8). A estimativa padro foi
calculada para todo o conjunto de medidas, bem como para cada um dos grupos.

Figura 4: Histograma - Grupo 1.

Figura 5: Histograma- Grupo 2.

Figura 6: Histograma - Grupo 3.


Figura 7: Histograma - Grupo 4.

Figura 8: Histograma - Grupo 5.

Analogamente, o mesmo procedimento foi realizado utilizando-se o multmetro


analgico. Este foi inicialmente calibrado, com a chave seletora na posio 1k
satisfatria para realizao das medies, provocando-se um curto circuito de modo a
ajustar o ponteiro no zero. Foi considerado um erro de escala de metade da menor
diviso da mesma correspondente a 1,0k. Aps a medio das resistncias (Tabela
2), foram feitos os histogramas (Figuras 9 a 14) e os clculos dos parmetros
estatsticos tanto para o conjunto completo de resistores quanto para os cinco grupos
em que o conjunto foi dividido.

Para as medies utilizando os diferentes multmetros, considerou-se tambm


o erro do fabricante dos aparelhos.
Tabela 2: Medidas de resistncia obtidas atravs da utilizao do multmetro analgico.
Resistor Ri Resistor Ri Resistor Ri Resistor Ri Resistor Ri
(i) (k) (i) (k) (i) (k) (i) (k) (i) (k)
1 23,0 23 22,5 45 22,0 67 22,5 89 22,5
2 23,0 24 23,0 46 22,5 68 23,0 90 22,5
3 23,0 25 23,0 47 23,0 69 23,0 91 22,5
4 23,0 26 23,0 48 23,0 70 23,0 92 22,5
5 23,0 27 23,0 49 23,0 71 23,0 93 23,0
6 23,0 28 23,0 50 23,0 72 23,0 94 23,0
7 23,0 29 23,0 51 23,0 73 23,0 95 23,0
8 23,0 30 23,0 52 23,0 74 23,0 96 23,0
9 23,0 31 23,0 53 23,0 75 23,0 97 23,0
10 23,0 32 23,0 54 23,0 76 23,0 98 23,0
11 23,0 33 23,0 55 23,0 77 23,0 99 23,0
12 23,0 34 23,0 56 23,0 78 23,0 100 23,0
13 23,0 35 23,0 57 23,0 79 23,0 101 23,0
14 23,0 36 23,0 58 23,0 80 23,0 102 23,0
15 23,0 37 23,0 59 23,0 81 23,0 103 23,0
16 23,0 38 23,0 60 23,0 82 23,0 104 23,0
17 23,0 39 23,0 61 23,0 83 23,0 105 23,0
18 23,0 40 23,0 62 23,0 84 23,0 106 23,0
19 23,0 41 23,0 63 23,0 85 23,0 107 23,0
20 23,0 42 23,0 64 23,0 86 23,0 108 23,0
21 23,0 43 23,0 65 23,0 87 23,0 109 23,0
22 23,0 44 23,5 66 23,0 88 23,0 110 23,0

Figura 9: Histograma - Conjunto de 110 resistores.


Figura 10: Histograma - Grupo 1.

Figura 11: Histograma - Grupo 2.

Figura 12: Histograma - Grupo 3.


Figura 13: Histograma - Grupo 4.

Figura 14: Histograma - Grupo 5.

4. Clculos

4.1. Amplitude (A)


=

4.2. Mdia das resistncias ( )

Utilizada para estimar o valor mdio representativo dos dados da amostra.



1
=

=1

onde Ri = valor da resistncia i (k) e N = nmero total de medies.


4.3. Desvio-padro ( )

Utilizado para estimar o erro em cada medida da amostra.


1
2
1
= ( 2 )

=1

onde = .

4.4. Erro na mdia ( )

Utilizado para estimar o erro associado mdia amostral.

4.5. Erro-padro ()

Utilizado para estimar o erro considerando, alm do erro na mdia, os


possveis desvios decorrentes dos aparelhos de medio e do valor nominal.
1
2
= (2 + 2
+ 2
+ 2
)

onde = 0,008 para o multmetro digital e = 0,04 e


= 1,0 para o multmetro analgico.

4.6. Estimativa-padro (Restimada)

4.7. Compatibilidade e Discrepncia

A discrepncia calculada de forma que se

| |
2

aceita-se que o valor estimado (mdia das resistncias) compatvel com o valor de
referncia , no caso o valor nominal encontrado com o Cdigo de Cores.

5. Resultados

O valor de referncia obtido atravs do Cdigo de Cores, conforme mostrado


no item 3 Rnominal = (22,0 1,1)k.

Tabela 3: Resultados obtidos a partir do conjunto de 110 resistores.

Parmetro Digital (k) Analgico (k)


21,7 0,0 23,0 0,0
0,1 0,2
0,2 0,9
- 1,0

1,1 1,7

Tabela 4: Resultados obtidos a partir dos grupos de 22 resistores.

Grupo 1 Grupo 2 Grupo 3 Grupo 4 Grupo 5 Resultado


Parmetro
(k) (k) (k) (k) (k) (k)
Digital
21,7 0,0 21,7 0,0 21,7 0,0 21,7 0,0 21,7 0,0 21,7 0,0
0,1 0,1 0,1 0,1 0,1 0,1
0,2
1,1 1,1 1,1 1,1 1,1 1,1
Analgico
23,0 0,0 23,0 0,0 22,9 0,0 23,0 0,0 22,9 0,0 23,0 0,0
0,0 0,2 0,2 0,1 0,2 0,0
0,9
1,0

1,7 1,7 1,7 1,7 1,7 1,4

Deve-se observar que, para o clculo do erro-padro no caso do multmetro


digital, no foi considerado o erro associado escala ( ), visto

que no faria sentido, j que no temos escala para visualizao.

Foi tambm realizado o clculo dos parmetros estatsticos envolvendo a


mdia das mdias dos grupos, como pode ser visto na ltima coluna da Tabela 4. Tais
clculos so anlogos aos realizados para as amostras individuais.
Tabela 5: Compatibilidade.

Parmetro Digital x Cdigo de Cores Analgico x Cdigo de Cores Analgico x Digital

| |
0,3 0,6 0,6

Compatibilidade Compatvel Compatvel Compatvel

6. Concluses

Analisando os resultados obtidos, no observamos uma grande discrepncia


entre o valor nominal e os valores experimentais da resistncia. No entanto, a
discrepncia maior quando consideramos os valores estimados por meio do
multmetro analgico em comparao aos obtidos com o uso do multmetro digital.
Ainda assim, os valores de discrepncia calculados sempre bem menores que 2
mostraram que as medies dos resistores so compatveis com o valor nominal dos
mesmos.

Nos resultados, podemos perceber que os erros das mdias so desprezveis.


Porm, algumas fontes de erros foram observadas durante o procedimento. No caso
do multmetro digital, a maior influncia no erro vem do erro do prprio aparelho de
medio utilizado (erro do fabricante). J para o multmetro analgico, observa-se um
erro mais considervel, uma vez que, alm do erro do fabricante (maior do que o
mesmo para o aparelho digital), tem-se tambm o erro visual na obteno do valor
medido (correspondente metade do valor da menor diviso da escala do aparelho).

Apesar das fontes de erros mencionadas, pode-se dizer que as medidas das
resistncias so bastante satisfatrias e os multmetros mostraram-se bastante
precisos, j que as discrepncias entre as medidas e o valor nominal dos resistores
so muito pequenas.