Você está na página 1de 7

Differences in implant stability

associated with various methods of


preparation of the implant bed: An in
vitro study
Su-Jin Ahn, DMD, MSD, PhD,a Richard Leesungbok, DMD,
MSD, PhD,b Suk-Won Lee, DMD, MSD, PhD,c Young-Ku Heo,
DMD, MSD, PhD,d Kyung Lhi Kang, DMD, MSD, PhDe
Kyung Hee University Dental Hospital at Gangdong, Seoul,
Korea; Boston University School of Dental Medicine, Boston, Mass
Statement of problem. It is difficult to achieve the primary stability necessary for immediate loading in the posterior
maxilla because of thin cortical bone, low density trabecular bone, and inadequate bone height due to the presence of
the maxillary sinus.

Purpose. The purpose of this study was to examine the primary stability of dental implants placed by using different
methods of preparation for in vitro monocortical and bicortical models of the posterior maxilla.

Material and methods. Sixty screw-shaped implants (4.0 10 mm) were inserted into solid rigid polyurethane blocks.
The implants were divided into 6 groups (n=10) to test 2 variables: 1) location (monocortical or bicortical block) and
2) preparation method (standard preparation, underpreparation, or the osteotome technique). The insertion and
removal torques were measured and resonance frequency analysis (RFA) was performed to determine the primary
stability of each implant. Insertion and removal torque data were analyzed by 2-way ANOVA, followed by the post hoc
Tukey HSD multiple comparison test. RFA data were analyzed by 2-way and 1-way ANOVAs and the Tukey HSD mul-
tiple comparison test (=.05). The Pearson correlation analysis was also performed to examine correlations among
the values.

Results. The preparation method had a significant effect on insertion torque, RFA value, and removal torque; how-
ever location had a significant effect only on the removal torque (P<.001). There was a significant interaction be-
tween location and preparation method for RFA values (P=.045) and a significant difference in standard preparation
method according to the location (P=.039); however, there was no significant difference in underpreparation (P=1.00)
and osteotome technique (P=1.00). Statistically significant correlations were found between insertion torque and RFA
values (r=0.529, P< .001), insertion torque and removal torque values (r=0.517, P< .001), and removal torque and
RFA values (r=0.481, P<.001).

Conclusions. Underpreparation and bicortical fixation significantly increased implant stability and the osteotome
technique decreased implant stability in synthetic bone models that mimicked the posterior maxillary region. The pri-
mary stability values had statistically significant correlations to each other. (J Prosthet Dent 2012;107:366-372)

Clinical Implications
Based on the results of this in vitro study, standard preparation and
the bicortical fixation method produce greater primary stability than
the various other surgical methods evaluated.

Supported by grant No. KHU-20091453 from the Kyung Hee University, Seoul, Korea.

a
Clinical Assistant Professor, Department of Biomaterials & Prosthodontics, Kyung Hee University Dental Hospital at Gangdong.
b
Professor & Chairman, Department of Biomaterials & Prosthodontics, Kyung Hee University Dental Hospital at Gangdong.
c
Assistant Professor, Department of Biomaterials & Prosthodontics, Kyung Hee University Dental Hospital at Gangdong.
d
Adjunct Assistant Clinical professor, Center for Implantology, Boston University School of Dental Medicine.
e
Assistant Professor, Department of Periodontics, Kyung Hee University Dental Hospital at Gangdong.
The Journal of Prosthetic Dentistry Ahn et al
June 2012 367
Several critical factors are neces- implants, the increase in bone density lution and high variability of these in-
sary for successful osseointegration is actually limited to the periapical struments during examination.25 Cur-
of dental implants, including the area of the entire periimplant area, rently, there is no gold standard for
primary stability and surface charac- and in the pericylinder area there was the accurate measurement of implant
teristics of the implant, anatomical no increase in bone density with the stability, and studies have cast doubt
conditions, bone metabolism, design osteotome technique. In addition, upon the correlation between the val-
of the interim prosthesis, and the oc- many studies have suggested that the ues of insertion and removal torques
clusion pattern during the healing use of the osteotome decreases or and RFA.21,26-29
phase.1 The primary stability of the does not affect primary stability.18-20 This study had 2 objectives: to
implant, which results from the initial According to Nkenke et al,21 use of compare the primary implant stabil-
interlocking between alveolar bone the osteotome to condense the bone ity associated with different prepara-
and the body of the implant, affects results in longitudinal cracks and gaps tion methods in both monocortical
the secondary stability of the implant in the region of the bone collar, in- and bicortical models of the posterior
because the latter results from subse- creasing the rate of implant failure. To maxilla and to examine the correla-
quent contact osteogenesis and bone date there is insufficient scientific and tion between biomechanical testing
remodeling.2,3 As a consequence, a clinical evidence to support immedi- (insertion and removal torque) and
high degree of primary implant stabil- ate loading in the posterior maxillary RFA. The null hypotheses were that
ity is a key prerequisite for immediate region.22 location (monocortical block or bicor-
or early loading.4,5 The primary de- Several methods can be used to tical block) and preparation method
terminants of the primary stability of measure primary implant stability; (standard preparation, underprepara-
an implant are the surgical technique these include biomechanical tests, tion, or osteotome technique) would
used, the design of the implant, and which are represented by measure- not affect the primary stability of im-
the mechanical properties of the bone ment of the insertion and removal plants and that there would be no cor-
tissue.6 torque and nondestructive measure- relation between the values of inser-
The posterior region of the maxilla ments such as resonance frequency tion and removal torques and RFA.
is characterized by thin cortical bone analysis (RFA). Biomechanical test-
and trabecular bone of low density. In ing, such as measurement of the MATERIAL AND METHODS
addition, in many instances the height insertion and removal torque, is
of the bone in this region is insuffi- more accurate than nondestructive Polyurethane specimens
cient to achieve high primary stability measurements such as RFA and the
because of the presence of the maxil- Periotest.23 However since biome- Solid rigid polyurethane blocks
lary sinus. Therefore, dental implants chanical testing is destructive and can (Sawbones; Pacific Research Labora-
in this region show the highest rate be applied only once, its clinical util- tories Inc, Vashon, Washington) were
of failure, and surgical techniques ity is limited. Therefore, nondestruc- used to simulate monocortical and
have been proposed to increase their tive measurements such as RFA are bicortical conditions in the posterior
primary stability.7-10 The most widely commonly used in clinical practice.24 maxillary region (Fig. 1). To model
used methods include preparation The use of RFA and the Periotest is cancellous bone, a density of 0.32 g/
of the site with tools one size smaller also limited because of the low reso- cm3 was selected because the mean
than the diameter of the implant,11
bone condensation using an osteo-
tome,12-14 and the use of bicortical
fixation.15 Among these methods, the
osteotome technique was introduced
to increase the primary stability and
success rate of implants in areas of
poor bone density, such as the pos-
terior maxillary region.16 Theoretically,
the osteotome condenses the bone
to increase primary stability by lateral
osseocompression. However, accord-
ing to Blanco et al,17 who studied the
placement of implants using the osteo-
tome in the maxillary tuberosities of 1 Polyurethane synthetic bone blocks and implant specimens. Left: bone
human cadavers and performed histo- blocks with monocortical layer. Right: bone blocks with bicortical layer.
morphometric assessment around the
Ahn et al
368 Volume 107 Issue 6
drills were used in accordance with
the manufacturers instructions (Ne-
obiotech Co Ltd). The implants were
then placed in the block. For the UM
group, the 3.4 mm-diameter drill was
not used before the implants were
inserted. In the OM group, a round
bur and the 2 mm-diameter twist
drill were used, and the area was ex-
panded by using 2.2 mm, 2.8 mm,
and 3.5 mm-diameter osteotomes
(Straumann AG, Basel, Switzerland)
before placement of the implants in
the block. For the SB, UB, and OB
groups the implants were inserted in
bicortical blocks with the same surgi-
cal protocols as those described for
the SM, UM, and OM groups.
2 Schematic representation of different preparation methods for monocorti-
cal and bicortical blocks. Measurement of insertion torque

When all implant specimens had


bone mineral density of the cancellous resorbable blast media (RBM), whose been inserted, peak insertion torques
bone in the posterior maxillary region stability has been demonstrated.33-35 were measured with a digital torque
is 0.31 g/cm3.30 To mimic the cortical gauge instrument (MG series; Mark-
layer, epoxy sheets (Sawbones; Pacific Preparation of the implant bed 10 Corporation, New York, NY) with
Research Laboratories Inc) filled with a measuring range of 0 to 135 Ncm.
short fibers were used to laminate the Implant sites were prepared on the
material that represented the cancel- polyurethane blocks with a handpiece Resonance Frequency Analysis
lous bone. Given that the mean thick- (INTRAsurg 300; KaVo Dental, Bib-
ness of cortical bone in the maxillary erach, Germany). The procedure was RFA was performed by using the
region is 1.49 0.34 mm31 and the performed at 1200 rpm with external latest version of the Osstell Mentor
maxillary sinus inferior border thick- cooling. The implants were divided (Osstell AB, Gteborg, Sweden). This
ness is 0.86 0.21 mm,32 a thickness into 6 groups with a combination of device uses a wireless measurement
of 1 mm was selected for the cortical location and preparation method. with a magnetic post (Smartpeg; Os-
layer in the model. For the monocor- Different surgical protocols were used stell AB) connected to the implant.
tical block, one side of a 15 mm sec- for each of the 6 groups: the SM group The manufacturer states that vibra-
tion that represented cancellous bone (Standard preparation + Monocorti- tion of the magnetic post, which is
was laminated with 1 mm of cortex cal fixation); UM group (Underprepa- excited by magnetic pulses from a
(to produce a bone block that was ration + Monocortical fixation); OM handheld computer, generates reso-
16 mm thick in total). For the bicorti- group (Osteotome technique + Mono- nance frequencies, and the resulting
cal block, both sides of a 6 mm sec- cortical fixation); SB group (Standard RF value, in hertz, is converted auto-
tion that represented cancellous bone preparation + Bicortical fixation); UB matically into implant stability quo-
were laminated to provide a bone group (Underpreparation + Bicortical tient (ISQ). Values range from 1 to
block that was 8 mm thick. fixation); and OB group (Osteotome 100, and a high value indicates high
technique + Bicortical fixation) (Fig. implant stability.
Implants 2). Each group consisted of 10 im-
plants, for a total of 60 implants. Measurement of removal torque
Sixty screw-shaped implants (Si- For the SM, UM and OM groups,
nusQuick; Neobiotech Co Ltd, Seoul, the implants were inserted into mono- The bone block was placed in a
Korea) were used. All implants measured cortical blocks. For the SM group, the locking vice and fixed to transmit the
4.0 mm in diameter and 10 mm in length. procedure began with a round bur removal torque longitudinally. The
These implants were tapered, screw- (Neobiotech Co Ltd), followed by a torque was increased gradually (dis-
shaped implants of the self-tapping 2 mm-diameter twist drill; then, 2.8 placement; 0.5 mm/min), and the
type. Their surfaces were treated with mm, 3.2 mm, and 3.4 mm-diameter point at which the specimen became
The Journal of Prosthetic Dentistry Ahn et al
June 2012 369
unfastened was recorded. The re- osteotome techniques (P<.05). nique had the lowest mean (SD) RFA
moval torque was measured with the The mean RFA values (SD) of values (monocortical block 58.90
same torque gauge instrument (MG each group are presented in Table III. (3.63), bicortical block 58.80 (4.08))
series; MARK-IO Corporation) that Among the preparation methods, the (Table III). Based on the 2-way ANO-
was used to measure the insertion under-preparation method had sig- VA, the preparation method had a
torque. nificantly higher mean (SD) RFA val- significant effect on the RFA values
Insertion and removal torque ues in the monocortical block (66.50 (Table IV) (P<.001); however, there
data were analyzed by 2-way analysis (2.59) and the standard preparation was no significant effect on the loca-
of variance (ANOVA) with location method had significantly higher RFA tion (P=.103). The Tukey HSD mul-
and preparation method, followed values in the bicortical block (68.40 tiple comparison test showed there
by multiple comparisons using the (5.87)) than other methods evalu- were significant differences among
Tukey Honestly Significant Difference ated (P<.05). Similar to the insertion standard preparation, underprepara-
(HSD) multiple comparison test. RFA torque values, the osteotome tech- tion, and the osteotome technique
data were analyzed by 2-way ANOVA
to examine the effect of location and Table I. Mean values (SD) of insertion torque (Ncm)
preparation method and the 1-way
ANOVA, followed by the post hoc Standard Under Osteotome
Tukey HSD multiple comparison test Preparation Preparation Technique
to evaluate differences among the
testing groups. The Pearson correla- Monocortical fixation 89.45 (10.03)a 104.57 (18.16)b 58.92 (9.53)c
tion analysis was also performed to Bicortical fixation 76.98 (15.84) a
104.62 (18.58) b
60.58 (15.49)c
examine correlations between inser- Different lower case letters in same column indicate significant differences (P<.05)
tion torque, RFA value, and removal
torque. An alpha value of .05 was Table II. Two-way ANOVA for insertion torque
used for statistical analysis. All calcu-
Sum of Mean
lations were performed with statisti-
cal software (SAS version 9.1 for Win- Source Squares df Squares F P
dows; SAS Institute Inc, Cary, NC).
Corrected model 20917 5 4183 18.41 <.001
Data are presented as mean (SD).
Location (A) 193 1 193 0.85 .3601
RESULTS Preparation method (B) 20125 2 10063 44.28 <.001
AB 598 2 299 1.32 .277
The mean insertion torques (SD) Error 12271 54 227
of each group are presented in Table I.
Among the preparation methods, un-
derpreparation had significantly high-
Table III. Mean values (SD) of RFA values (ISQ)
er mean (SD) insertion torque values Standard Under Osteotome
(monocortical block, 104.57 (18.16): Preparation Preparation Technique
bicortical block, 104.62 (18.58)) than
the other methods evaluated (P<.05). Monocortical fixation 63.30 (2.5)a 66.50 (2.59)b 58.90 (3.63)d
The osteotome technique had the Bicortical fixation 68.40 (5.87)c 66.30 (2.54)b 58.8 (4.08)d
lowest mean (SD) values (monocor-
tical block, 58.92 (9.53); bicorti- Different lower case letters in same column indicate significant differences (P<.05)

cal block, 60.58 (15.49)). Two-way


ANOVA revealed that the preparation Table IV. Two-way ANOVA for RFA values
method had a significant effect on the
Sum of Mean
insertion torque (Table II) (P<.001).
However, there was no significant ef- Source Squares df Squares F P
fect for the location (P=.361). The in-
Corrected model 839 5 168 12.02 <.001
teraction between preparation meth-
od and location was not significantly Location (A) 38 1 38 25.39 .103
different (P=.277). The Tukey HSD Preparation method (B) 709 2 354 2.75 <.001
multiple comparison test showed sig- AB 92 2 46 3.29 .045
nificant differences among standard Error 754 54 14
preparation, underpreparation, and
Ahn et al
370 Volume 107 Issue 6
cal stability by the presence of bone
Table V. Mean values (SD) of removal torque (Ncm) regeneration and remodeling in the
Standard Under Osteotome latter. In addition, primary stability,
rather than secondary stability, is a
Preparation Preparation Technique
critical factor when immediate load-
ing is considered because it develops
Monocortical fixation 40.53 (12.17)aa 38.32 (7.60)aa 17.02 (7.48)ba
within 1 week of implantation. For
Bicortical fixation 54.64 (12.04)ab 53.28 (11.86)ab 26.23 (9.88)bb
immediate loading, single implants
Different lower case letters in same column indicate significant differences (P<.05) must be inserted with a torque greater
than or equal to 30 Ncm, and splinted
Table VI. Two-way ANOVA for removal torque implants with a torque greater than or
equal to 20 Ncm.5 In the present study,
Sum of Mean
the insertion torque exceeded 30 Ncm
Source Squares df Squares F P in all groups. These values are higher
than those reported by Nkenke et al,26
Corrected model 10949 5 2190 20.37 <.001
who measured insertion torque in the
Location (A) 2442 1 2442 22.72 <.001 posterior maxillary region of human
Preparation method (B) 8410 2 4205 39.13 <.001 cadavers. These findings suggest that
AB 96 2 48 0.45 .641 measured insertion torque in polyure-
Error 5804 54 107 thane specimens is higher than that
in the posterior maxillary region in
(P<.05). There was significant in- torque values than the monocortical humans, even though the mean den-
teraction between the location and fixation method (P<.05) (Table V). sity was similar. In addition, for all the
preparation methods for the RFA Statistically significant correla- groups in this study, the mean values
values (P=.045). Therefore, based on tions were found between insertion of removal torque were lower than the
the preparation method and location, torque and RFA values (r=0.529, mean values of insertion torque. This
modified 1-way ANOVA analysis and P<.001), insertion torque and remov- can be explained by the restricted vis-
the Tukey HSD multiple comparison al torque values (r=0.517, P<.001) coelastic properties of the surround-
test were conducted. These revealed and removal torque and RFA values ing artificial bone, resulting in less
that there was a significant difference (r=0.481, P<.001). resistance during removal. These find-
in standard preparation according to ings are in agreement with other re-
the location (P=.039), but there was DISCUSSION cently published data using polyure-
no significant difference in under- thane foam blocks in which removal
preparation (P=1.00) and osteotome This study demonstrated that 2 torques were lower than insertion
technique (P=1.00). variables affect the primary stability torques;36 similar observations have
The mean removal torque values of implants: location (monocortical been made in animal studies.37,38
(SD) of each group are presented in block or bicortical block) and prepa- The preparation method had a
Table V. Two-way ANOVA revealed ration method (standard prepara- significant effect on insertion torque,
that location and preparation method tion, and underpreparation or the RFA values, and removal torque. The
had a significant effect on the removal osteotome technique). The correla- underpreparation method had a sig-
torque (Table VI) (P<.001). There was tion between biomechanical testing nificant effect on insertion torque
no significant interaction between (insertion and removal torque) and compared to standard preparation
preparation method and location on RFA were also evaluated. The results and the osteotome technique in either
removal torque (P=.641). Among the lead to a rejection of the first null monocortical or bicortical blocks.
preparation methods, there was no hypothesis that location and prepa- However, the standard preparation
significant difference between stan- ration method do not affect the pri- method had a more significant effect
dard preparation and underprepara- mary stability of implants; the second on RFA values than the underprepara-
tion (P=.897); however, there was a hypothesis that there was no correla- tion method in bicortical blocks, and
significant difference between stan- tion between the values of insertion for removal torque values, there was
dard preparation and osteotome and removal torque and RFA was also no difference between the standard
technique and between underprepa- rejected. preparation and underpreparation
ration and osteotome technique Primary stability is the result of method. These findings are some-
(P<.05). As for the difference of the mechanical engagement with the sur- what different from those of another
location, bicortical fixation meth- rounding bone tissue and can be dis- recent study on the influence of surgi-
od had significantly higher removal tinguished from secondary or biologi- cal technique and surface roughness
The Journal of Prosthetic Dentistry Ahn et al
June 2012 371
on the primary stability of implants in and removal torque, even though the 3. Esposito M, Grusovin MG, Willings M,
Coulthard P, Worthington HV. Interven-
artificial bone with a density equiva- correlation coefficient value was low- tions for replacing missing teeth: differ-
lent to maxillary bone.23 The authors er than another study with a smaller ent times for loading dental implants.
found that both the insertion and number of specimens.29 Nevertheless, Cochrane Database Systematic Reviews
2007;18:CD003878
removal torque values obtained with a similar study with a larger number 4. Kim SK, Lee HN, Choi YC, Heo SJ, Lee CW,
the undersized technique were higher of specimens showed low coefficient Choie MK. Effects of anodized oxidation
than with the press-fit technique. Per- values with statistically significant dif- or turned implants on bone healing after
using conventional drilling or trabecular
haps these differences were due to the ferences.39 In the present study, inser- compaction technique: histomorphometric
presence of the cortical portion of the tion torque, RFA value. and removal analysis and RFA. Clin Oral Implants Res
synthetic bone models of the present torque were correlated but only with a 2006;17:644-50.
5. Roccuzzo M, Aglietta M, Cordaro L. Im-
study. Tabassum et al,36 showed that poor linear relationship for strength. plant loading protocols for partially eden-
for implant sites with a cortical thick- A limitation of the present study is tulous maxillary posterior sites. Int J Oral
ness of 2 mm or more, undersized that only the mechanical characteris- Maxillofac Implants Suppl 2009;24:147-57.
6. Sennerby L, Meredith N. Implant stability
drilling had no significant effect on tics of the primary stability of an im- measurements using resonance frequency
insertion and removal torque. Over plant were considered. Furthermore, analysis: biological and biomechanical as-
pects and clinical implications. Periodontol
tightening of the implant may result the model only simulated the oral
2000 2008;47:51-66.
in micro fractures of the cortical parts environment, which is more complex. 7. Jaffin RA, Berman CL. The excessive loss of
around the implants. The osteotome In vivo studies are required to under- Branemark fixtures in type IV bone: a 5-year
analysis. J Periodontol 1991;62:2-4.
technique had significantly low values stand the actual clinical situation in 8. Balshi TJ, Lee HY, Hernandez RE. The use
for insertion torque, RFA, and remov- which many biological factors influ- of pterygomaxillary implants in the partially
al torque, which demonstrates that ence the primary stability of implants. edentulous patient: a preliminary report. Int
J Oral Maxillofac Implants 1995;10:89-98.
the use of the osteotome decreases 9. Balshi TJ, Wolfinger GJ, Balshi SF. Analy-
implant stability in polyurethane CONCLUSIONS sis of 356 pterygomaxillary implants in
blocks. edentulous arches for fixed prosthesis
anchorage. Int J Oral Maxillofac Implants
In addition, the outcomes of the Within the limitations of this in 1999;14:398-406.
present study also revealed that the vitro study, the following conclusions 10.Herrmann I, Lekholm U, Holm S, Kultje C.
use of a bicortical block significantly were drawn: Evaluation of patient and implant charac-
teristics as potential prognostic factors for
increased removal torque when com- 1. Preparation method and lo- oral implant failures. Int J Oral Maxillofac
pared with the use of a monocortical cation had a significant effect on Implants 2005;20:220-30.
11.Friberg B, Ekestubbe A, Sennerby L. Clinical
block. However there were no signifi- implant stability in synthetic bone
outcome of Brnemark system implants of
cant effects of the bicortical block on models that mimicked the posterior various diameters: a retrospective study. Int J
insertion torque or RFA values. The maxillary region. Oral Maxillofac Implants 2002;17:671-7.
12.Nedir R, Nurdin N, Szmukler-Moncler S,
insertion torque and RFA values were 2. The underpreparation and bi- Bischof M. Placement of tapered implants
affected by the preparation method cortical fixation method significantly using an osteotome sinus floor elevation
and RFA values interacted significantly increased implant stability compared technique without bone grafting: 1-year
results. Int J Oral Maxillofac Implants
between preparation method and lo- to standard preparation and mono- 2009;24:727-33.
cation. Among the preparation meth- cortical fixation. 13.Nedir R, Bischof M, Vazquez L, Nurdin N,
ods, the standard preparation caused 3. Low values of insertion torque, Szmukler-Moncler S, Bernard JP. Osteo-
tome sinus floor elevation technique with-
a significant difference between bicor- RFA, and removal torque were found out grafting material: 3-year results of a
tical and monocortical blocks, which in groups in which the osteotome was prospective pilot study. Clin Oral Implants
means that with bicortical blocks, the used, which demonstrates that use of Res 2009;20:701-7.
14.Crespi R, Cappar P, Gherlone E. Osteo-
standard preparation method is more the osteotome decreases implant sta- tome sinus floor elevation and simul-
effective in increasing RFA values than bility in polyurethane blocks. taneous implant placement in grafted
with monocortical blocks. 4. There were significant correla- biomaterial sockets: 3 years of follow-up. J
Periodontol 2010;81:344-9.
Primary stability can be measured tions between insertion torque, RFA 15.Ivanoff CJ, Grndahl K, Bergstrm C, Lek-
during the placement of an implant value, and removal torque. holm U, Brnemark PI. Influence of bicorti-
cal or monocortical anchorage on maxillary
by means of various parameters such
implant stability: a 15-year retrospective
as insertion and removal torque or by REFERENCES study of Brnemark system implants. Int J
using RFA or the Periotest. Although Oral Maxillofac Implants 2000;15:103-10.
1. Chiapasco M. Early and immediate restora- 16.Summers RB. A new concept in maxillary
individual parameters have been in- tion and loading of implants in completely implant surgery: the osteotome technique.
vestigated, few studies of the cor- edentulous patients. Int J Oral Maxillofac Compendium 1994;15:152-8.
relations between parameters have Implants Suppl 2004;19:76-91.
2. Berglundh T, Abrahamsson I, Lang NP,
been done. In this study, statistically Lindhe J. De novo alveolar bone formation
significant correlations were found adjacent to endosseous implants. Clin Oral
Implants Res 2003;14:251-62.
between insertion torque, RFA value,
Ahn et al
372 Volume 107 Issue 6
17.Blanco J, Surez J, Novio S, Villaverde G, 26.Nkenke E, Hahn M, Weinzierl K, Radespiel- 35.Piattelli M, Scarano A, Paolantonio M, Iezzi
Ramos I, Segade LA. Histomorphometric Trger M, Neukam FW, Engelke K. Implant G, Petrone G, Piattelli A. Bone response
assessment in human cadavers of the peri- stability and histomorphometry: a correla- to machined and resorbable blast material
implant bone density in maxillary tuber- tion study in human cadavers using stepped titanium implants: an experimental study in
osity following implant placement using cylinder implants. Clin Oral Implants Res rabbits. J Oral Implantol 2002;28:2-8.
osteotome and conventional techniques. 2003;14:601-9. 36.Tabassum A, Meijer GJ, Wolke JG, Jansen
Clin Oral Implants Res 2008;19:505-10. 27.Akkocaoglu M, Uysal S, Tekdemir l, Akca K, JA. Influence of surgical technique and sur-
18.Bchter A, Kleinheinz J, Wiesmann HP, Cehreli MC. Implant design and intraos- face roughness on the primary stability of
Kersken J, Nienkemper M, Weyhrother H, et seous stability of immediately placed an implant in artificial bone with different
al. Biological and biomechanical evaluation implants: a human cadaver study. Clin Oral cortical thickness: a laboratory study. Clin
of bone remodeling and implant stability Implants Res 2005;16:202-9. Oral Implants Res 2010;21:213-20.
after using an osteotome technique. Clin 28.da Cunha HA, Francischone CE, Filho HN, 37.Shalabi MM, Wolke JG, Jansen JA. The
Oral Implants Res 2005;16:1-8. de Oliveira RC. A comparison between cut- effects of implant surface roughness and
19.Fanuscu MI, Chang TL, Aka K. Effect of ting torque and resonance frequency in the surgical technique on implant fixation in
surgical techniques on primary implant assessment of primary stability and final an in vitro model. Clin Oral Implants Res
stability and peri-implant bone. J Oral torque capacity of standard and TiUnite 2006;17:172-8.
Maxillofac Surg 2007;65:2487-91. single-tooth implant under immediate 38.Shalabi MM, Wolke JG, de Ruijter AJ, Jan-
20.Proff P, Bayerlein T, Rottner K, Mai R, loading. Int J Oral Maxillofac Implants sen JA. A mechanical evaluation of implants
Fanghnel J, Gedrange T. Effect of bone 2004;19:578-85. placed with different surgical techniques
conditioning on primary stability of 29.Aka K, Akkocaoglu M, Cmert A, Tek- into the trabecular bone of goats. J Oral
FRIALIT-2 impalts. Clin Oral Implants Res demir I, Cehreli MC. Bone strains around Implantol 2007;33:51-8.
2008;19:42-7. immediately loaded implants support- 39.Jun SH, Chang BM, Weber HP, Kwon JJ.
21.Nkenke E, Lehner B, Fenner M, Roman FS, ing mandibular overdentures in human Comparison of initial stability param-
Thams U, Neukam FW, et al. Immediate cadavers. Int J Oral Maxillofac Implants eters and histomorphometric analysis of
versus delayed loading of dental implants 2007;22:101-9. implants inserted into extraction sockets:
in the maxillae of minipigs: follow-up of 30.Devlin H, Homer K, Ledgerton D. A human fresh cadaver study. Int J Oral Maxil-
implant stability and implant failures. Int J comparison of maxillary and mandibular lofac Implants 2010;25:985-90.
Oral Maxillofac Implants 2005;20:39-47. bone mineral densities. J Prosthet Dent
22.Weber HP, Morton D, Gallucci GO, Roc- 1998;79:323-7. Corresponding author:
cuzzo M, Cordaro L, Grutter L. Consensus 31.Miyamoto I, Tsuboi Y, Wada E, Suwa Dr Richard Leesungbok
statements and recommended clinical H, Iizuka T. Influence of cortical bone Department of Biomaterials & Prosthodontics
procedures regarding loading protocols. thickness and implant length on implant Kyung Hee University Dental Hospital at
Int J Oral Maxillofac Implants Suppl stability at the time of surgery--clinical, Gangdong
2009;24:180-3. prospective, biomechanical, and imaging Sangil-Dong 149, Kangdong-Gu
23.Tabassum A, Meijer GJ, Wolke JG, Jansen study. Bone 2005;37:776-80. Seoul134-727
JA. Influence of the surgical technique and 32.Cakur B, Smbll MA, Durna D. Rela- KOREA
surface roughness on the primary stabil- tionship among schneiderian membrane, Fax: +82-2-440-7549
ity of an implant in artificial bone with a Underwoods septa, and maxillary sinus E-mail: sbykmw@yahoo.co.kr
density equivalent to maxillary bone: a inferior border. Clin Implant Dent Relat Res
laboratory study. Clin Oral Implants Res 2011 Apr 19. [Epub ahead of print] Acknowledgments
2009;20:327-32. 33.Wennerberg A, Albrektsson T, Johansson C, This study was supported by the program of
24.Chang PC, Lang NP, Giannobile WV. Andersson B. Experimental study of turned Kyung Hee University for the young researcher
Evaluation of functional dynamics during and grit-blasted screw-shaped implants in medical science. (KHU-20091453)
osseointegration and regeneration associ- with special emphasis on effects of blasting
ated with oral implants. Clin Oral Implants material and surface topography. Biomate-
Res 2010;21:1-12. rials 1996;17:15-22. Copyright 2012 by the Editorial Council for
25.Aparico C, Lang NP, Rangert B. Validity 34.Wennerberg A, Albrektsson T, Lausmaa J. The Journal of Prosthetic Dentistry.
and clinical significance of biomechanical Torque and histomorphometric evaluation
testing of implant/bone interface. Clin Oral of c.p. titanium screws blasted with 25-
Implants Res Suppl.2 2006;17:2-7. and 75-microns-sized particles of Al2O3. J
Biomed Mater Res 1996;30:251-60.

Availability of Journal Back Issues

As a service to our subscribers, copies of back issues of The Journal of Prosthetic Dentistry for the preceding 5 years are
maintained and are available for purchase from Elsevier, Inc until inventory is depleted. Please write to Elsevier, Inc,
Subscription Customer Service, 6277 Sea Harbor Dr, Orlando, FL 32887, or call 800-654-2452 or 407-345-4000 for
information on availability of particular issues and prices.

The Journal of Prosthetic Dentistry Ahn et al

Você também pode gostar