Você está na página 1de 17

ESCUELA SUPERIOR POLTECNICA DEL

LITORAL

FACULTAD DE INGENIERIA MECANICA Y CIENCIAS DE LA

PRODUCCION

INFORME DE LABORATORIO

PRACTICA # 6: MEDICIN DE TEMPERATURA


ABSTRACT

For practical measurement of temperature there are a number of devices that can be mechanical as well
as electronic, during the practice the temperature of a single fluid was measured using various
measuring instruments such as glass thermometer, thermocouple and thermopile. The results were
compared by means of a calibration curve with respect to a pattern which in this case was a K-type
thermometer. Of all the instruments used it was concluded that the thermopile is the closest to the
measurements of the standard thermometer. It was also determined that the speed of response of each
instrument played an important role in practice since the measurements taken were instantaneous.

Keywords: temperatura, thermocouple, thermopile.

RESUMEN

Para la prctica medicin de temperatura se cuenta con una serie de dispositivos que pueden ser
mecnicos como electrnicos, durante la prctica se midi la temperatura de un solo fluido utilizando
varios instrumentos de medicin como termmetro de vidrios, termopares y termopila. Los resultados
se compararon mediante una curva de calibracin con respecto a un patrn que en este caso era un
termmetro tipo K. De todos los instrumentos utilizados se concluy que la termopila es el ms cercano
a las medidas del termmetro patrn. Se determin tambin que la velocidad de respuesta de cada
instrumento jug un papel importante en la prctica ya que las mediciones tomadas fueron instantneas.

Palabras clave: temperatura, termocupla, termopila.

I. INTRODUCCIN

La medida de temperatura constituye una de las mediciones ms comunes y ms


importantes que se efectan en los procesos industriales. Las limitaciones del sistema de
medida quedan definidas en cada tipo de aplicacin por la precisin, por la velocidad de
captacin de la temperatura, por la distancia entre el elemento de medida y el aparato
receptor y por el tipo de instrumento indicador.

Los instrumentos de temperatura utilizan diversos fenmenos que son influidos por la
temperatura y entre los cuales figuran:
variaciones en volumen o en estado de los cuerpos (slidos, lquidos o gases);
variacin de resistencia de un conductor (sondas de resistencia);
variacin de resistencia de un semiconductor (termistores);
f.e.m. creada en la unin de dos metales distintos (termopares);
intensidad de la radiacin total emitida por el cuerpo (pirmetros de radiacin);
otros fenmenos utilizados en laboratorio (velocidad del sonido en un gas,
frecuencia de resonancia de un cristal...).
Medicin de Temperatura

Las escalas de medicin de la temperatura, se dividen fundamentalmente en dos tipos:


relativas y absolutas.

Escalas Relativas:
Escala Celsius
Escala Fahrenheit
Escala Raumur,etc

Escalas Absolutas:
Escala Kelvin
Escala Rankine

Termmetro de vidrio

El termmetro de vidrio (fig. 6.1) consta de un depsito de vidrio que contiene, por
ejemplo, mercurio y que al calentarse se expande y sube en el tubo capilar. Los
mrgenes de trabajo de los fluidos empleados son:
Mercurio.....-35 hasta +280C
Mercurio (tubo capilar lleno de gas)...-35 hasta +450C
Pentano.-200 hasta + 20C
Alcohol-110 hasta +50C

Figura 6. 1: Termmetro de vidrio


Termmetro Bimetlico

El termmetro bimetlico es un instrumento utilizado para medir temperatura mediante


la contraccin y expansin de dos distintas aleaciones metlicas de alto y bajo
coeficiente de dilatacin.

Los termmetros bimetlicos estn constituidos por un tubo en cuyo interior est
colocada una espiral helicoidal bimetlica. Dicha espiral est soldada por un extremo a
la parte inferior del tubo y por otra a una varilla de transmisin, a su vez conectada a una
aguja indicadora.
Figura 6. 2: Termmetro bimetlico
Funcionamiento

Las aleaciones metlicas, la de alto y bajo coeficiente de dilatacin, al ser expuestas a


determinada temperatura, transmiten un movimiento giratorio a una aguja indicadora, es
decir, las variaciones de temperatura causan que el bimetal sufra una deformacin, esta
se transmite a la aguja indicadora
Los termmetros bimetlicos usan la diferencia de coeficiente de expansin trmica de
metales dismiles para poder determinar cul es el cambio de temperatura. ste se
proporciona mediante un movimiento mecnico, o sea, por el giro de una aguja sobre
una escala graduada.

Figura 6. 3: Elementos de un termmetro bimetlico

Termopar

El termopar se basa en el efecto descubierto por Seebeck en 1821, de la circulacin de


una corriente en un circuito formado por dos metales diferentes cuyas uniones (unin de
medida o caliente y unin de referencia o fra) se mantienen a distinta temperatura. Esta
circulacin de corriente obedece a dos efectos termoelctricos combinados, el efecto
Peltier que provoca la liberacin o absorcin de calor en la unin de dos metales
distintos cuando una corriente circula a travs de la unin y el efecto Thomson que
consiste en la liberacin o absorcin de calor cuando una corriente circula a travs de un
metal homogneo en el que existe un gradiente de temperaturas. Existen diferentes tipos
de termopares que son utilizados en la industria, algunos de ellos se pueden apreciar en
la tabla 6.1.

Figura 6. 4: Termopar

Tabla 6.1: Tipo de termopares

Termopila

Consiste en varias termocuplas conectadas en serie, de tal forma que los voltajes se
suman. Teniendo una gran cantidad de termocuplas y un buen gradiente de temperatura
es posible obtener algunos voltios.

La f.e.m. que proporciona la termopila depende de la diferencia de temperaturas entre la


unin caliente (radiacin procedente del objeto enfocado) y la unin fra. Esta ltima
coincide con la de la caja del pirmetro, es decir, con la temperatura ambiente.
Figura 6. 5: Termopila

Detectores de temperatura tipo resistencia (RTD)

El principio de operacin de los detectores de temperatura tipo resistencia est basado en


el hecho de que la resistencia elctrica de los metales varia directamente con la
temperatura. La magnitud de este cambio frente a 1C de cambio en la temperatura se
conoce como el coeficiente de resistencia de temperatura (). Para la mayora de los
metales puros, este coeficiente es constante dentro de un rango de temperatura.

El cambio en la resistencia es una funcin del coeficiente de resistencia de temperatura y


puede ser expresado por la ecuacin:

= 0 (1 + )
Donde,

Rt = Resistencia en ohms a la temperatura T


R0 = Resistencia en ohms a la temperatura de referencia (generalmente 0C)
= Coeficiente de resistencia de temperatura

II. OBJETIVOS

Conocer la temperatura de una sustancia por medio de diferentes instrumentos.


Aprender a utilizar la tabla de termopares.
Hallar la curva de calibracin de los diferentes instrumentos de medicin de
Temperatura con respecto al termmetro patrn.

III. MATERIALES Y EQUIPOS

Termmetro de vidrio
Termmetro bimetlico
Multmetro
Termopila tipo J
Termocupla tipo E
Termmetro tipo K (patrn)
Termmetro digital RTD
Recipiente con agua
Equipo: Multmetro digital escala mvil
Marca: Acoustic
Modelo: DT-890C
Voltaje DC: 200mv-2v-20v-200v-1000v
Voltaje AC: 2v-20v-200v-700v
Corriente DC: 2ma-20ma-200ma-20a
Corriente AC: 200ma-20a
Resistencia: 200 -2k -20k -200k -2m -20m -200m
Capacimetro: 2000pf-20 f-200 f-2mf-20mf
Temperatura: -40cto+400c-400cto+1000c-0cto40c
Zumbador: menor de 500ohm.
Dimensiones: ancho x alto 9.5 x 18.3 cm.

Equipo: Termmetro
Marca: EXTECH
Modelo: SDL-200
Caracteristicas:
Registro de datos de 4 canales con 6 tipo termopar (J,
K, E, T, R,S) y registro de datos de 2 canales con sondas RTD
(detector termomtrico de resistencia) (Pt de 100 ohmios)
El ajuste de compensacin se utiliza para que la
funcin cero realice mediciones relativas
Almacena manualmente 99 lecturas y lecturas de 20 M
con la tarjeta de memoria SD de 2 G
Registra las lecturas con reloj fechador y fecha real
Frecuencia de muestreo programable por el usuario: 1 a
3600 segundos

Equipo: Termmetro de mano de alta precision


Marca: OMEGA
Modelo: 450-AET
Pantalla: 0,3 "(7,62 mm)
Indicacin LCD: F / C seleccionable
Rango de funcionamiento: 0 a 50 C ambiente
Precisin en el tiempo: 90 das: aadir 0,05% rdg; 1 ao:
aadir 0.1% rdg
Alcance: 0,02% / C
Conexin de la sonda: Conector subminiatura (termopar);
Conector de acoplamiento TA3F (RTD) suministrado con
unidades RTD y termistor
Temperatura de almacenamiento: -20 a + 70 C
Descripcin grfica del banco de pruebas

IV. PROCEDIMIENTO

1. Calentamos el agua del recipiente hasta que observemos que el termmetro


bimetlico marque una temperatura de 180F.
2. Se procede a introducir los diferentes instrumentos de medicin.
3. Leemos las temperaturas indicadas en los diferentes instrumentos.
4. Agregamos agua fra y mezclamos hasta que el termmetro bimetlico
descienda 10F para nuevamente proceder a registrar los datos.
5. Realizamos el paso 4 hasta obtener mnimo seis mediciones.

V. TABLA DE DATOS

Temperatura Termmetro Termmetro Termocupla Termmetro Termmetro Termopila


Ambiente Patrn tipo K Bimetlico tipo E digital RTD de vidrio tipo J
(C) (C) (F) (C) (F) (F) (mV)
28.5 81.3 180 82 176 182 7.8
28.4 73.5 170 76 166 170 7.1
28.4 69.0 160 71 153 160 6.3
28.6 61.8 150 64 141 150 5.1
28.4 58.3 140 59 134 140 4.5
29.0 54.3 130 55 127 132 4.0
Con ayuda de la tabla de termocupla tipo J transformamos los C a mV

Tabla de termocupla tipo J

Temperatura Ambiente (C) Temperatura Ambiente (mV)


28.5 1.458
28.4 1.453
28.4 1.453
28.6 1.463
28.4 1.453
29.0 1.484
VI. CALCULOS REPRESENTATIVOS

Transformamos de F a C con la siguiente ecuacin:

( 32)
=
1,8
Termmetro Bimetlico

(18032)
1) = = 82.2
1,8

(17032)
2) = = 76.7
1,8

(16032)
3) = = 71.1
1,8

(15032)
4) = = 65.6
1,8

(14032)
5) = = 60.0
1,8

(13032)
6) = = 54.4
1,8

Termmetro digital RTD

(17632)
1) = = 80.0
1,8

(16632)
2) = = 74.4
1,8

(15332)
3) = = 67.2
1,8

(14132)
4) = = 60.6
1,8

(13432)
5) = = 56.7
1,8

(12732)
6) = = 52.8
1,8
Termmetro de vidrio

(18232)
1) = = 83.3
1,8

(17032)
2) = = 76.7
1,8

(16032)
3) = = 71.1
1,8

(15032)
4) = = 65.6
1,8

(14032)
5) = = 60.0
1,8

(13232)
6) = = 55.6
1,8

Termopila tipo J

Para transformar de mV a C primeros calculamos el voltaje de la junta caliente, despus


con el voltaje obtenido buscamos en la tabla tipo J los C que corresponden. Hacemos
uso de la siguiente ecuacin para calcular el voltaje de junta caliente:

Donde VJc es el voltaje generado por las juntas calientes y VJf es el voltaje generado por
las juntas fras y V es el voltaje que se pudo obtener con el multmetro, para la prctica
se realizaron tres conexiones para las juntas fras y calientes.

+
=

.+(.)
1) = = .

Con ayuda de la tabla tipo J buscamos la temperatura para el voltaje calculado, en


caso de no encontrar el valor en la tabla procedemos a interpolar.

T = 77.6 C
.+(.)
2) = = .

T = 73.2 C

.+(.)
3) = = .

T = 68.2 C

.+(.)
4) = = .

T = 60.9 C

.+(.)
5) = = .

T = 56.9 C

.+(.)
6) = = .

T = 54.4 C

VII. TABLA DE RESULTADOS

Termmetro Termmetro Termocupla Termmetro Termmetro Termopila


Patrn tipo K Bimetlico tipo E digital RTD de vidrio tipo J
(C) (C) (C) (C) (C) (C)
81.3 82.2 82 80.0 83.3 77.6
73.5 76.7 76 74.4 76.7 73.2
69.0 71.1 71 67.2 71.1 68.2
61.8 65.6 64 60.6 65.6 60.9
58.3 60.0 59 56.7 60.0 56.9
54.3 54.4 55 52.8 55.6 54.4
VIII. GRFICAS

1) Termmetro Patrn vs Termmetro Bimetlico

2) Termmetro Patrn vs Termocupla Tipo E


3) Termmetro Patrn vs Termmetro Digital RTD

4) Termmetro Patrn vs Termmetro de Vidrio


5) Termmetro Patrn vs Termopila J

IX. ANALISIS DE RESULTADOS

Analizando las grficas de calibracin de cada uno de los instrumentos se procede a


juzgar cual es el que ms se acerca a los datos de entrada representados por el
termmetro patrn que fue un termmetro tipo K, como podemos observar la grfica
nmero 5 presenta una buena relacin con la grfica linealizada esperada, por lo tanto
podemos confiar en este instrumento, adems se concluy que se us correctamente la
tabla de la termocupla tipo J.

Seguido de la termopila los instrumento que ms se acercan al termmetro patrn son el


termcupla tipo E y el termmetro de vidrio, aunque no hay mucha variacin con
respecto a la termopila. Al anotar los datos del termmetro de vidrio se tuvo que
transformar a C esto pudo provocar un cambio en la realizacin de la grfica para evitar
esto se deba anotar directamente en los grados a trabajar. Aunque tambin podemos
hacer uso de los otros instrumentos debido a que no presentan una gran variacin con
respecto al instrumento patrn.

X. CONCLUSIONES

Se pudo determinar la temperatura de una sustancia utilizando varios instrumentos de


medicin, se aprendi a utilizar la tabla de termopares y analizando las grficas de
calibracin de cada instrumento con respecto al instrumento patrn, se puede decir que
la termopila tipo J es el ms eficiente, adems los otros instrumento de medicin
tambin presentan una gran eficiencia por ende podemos decir que la prctica medicin
de temperatura fue realizada con xito y las conversiones de mV a C fueron realizadas
correctamente.

XI. RECOMENDACIONES

Usar el equipo de proteccin personal para evitar accidentes.


Al introducir agua fra mezclar para no tener error en las mediciones.
Verificar si los instrumentos se encuentran en C o F.
Verificar que todos los instrumentos se encuentren en buenas
condiciones.
Ante cualquier duda preguntar al profesor.

XII. BIBLIOGRAFIA

DT-890C - Cotzul. (2017). Cotzul. Recuperado de http://cotzul.com/herramientas/1073-


dt890c.html

SDL200 - Termmetro con registro de datos de 4 canales. (2017). Extech.com.es. Recuperado de


http://www.extech.com.es/instruments/product.asp?catid=64&prodid=673

Thermometers, H. (2017). High Accuracy Handheld Thermometers. Mx.omega.com.


Recuperado de http://mx.omega.com/pptst/450_HANDHELD.html

Marketizer.com, Q. (2017). Los termmetros bimetlicos y su modo de funcionamiento |


QuimiNet.com. Quiminet.com. Recuperado de https://www.quiminet.com/articulos/los-
termometros-bimetalicos-y-su-modo-de-funcionamiento-3516632.htm

ANTONIO CREUS SOLE. (1997). Instrumentacin Industrial (6ta edicin), Barcelona,


Marcombo.

Você também pode gostar