I. SOURCE
2. AC Voltage$ 10 Hz to 1 kHz
1 mV to 10 mV 2.7% to 0.64% Using Fluke 5500 A
10 mV to 100 mV 0.64% to 0.35% Calibrator by Direct Method
100 mV to 1 V 0.35% to 0.20%
1 V to 10 V 0.20% to 0.05%
1 kHz to 90 kHz
1 V to 10 V 0.04% to 0.47%
45 Hz to 1 kHz
10 V to 100 V 0.2 % to 0.07%
100 V to 1000 V 0.07%
1 kHz to 18 kHz
10 V to 100 V 0.05% to 0.14%
1 kHz to 10 kHz
100 V to 1000 V 0.07% to 0.3%
4. AC Current$ 45 Hz to 1 kHz
30A to 10mA 1.13% to 0.2% Using Fluke 5500A
10 mA to 100 mA 0.2% to 0.5% Calibrator
100 mA to 1 A 0.5% to 0.3%
1 A to 10 A 0.3% to 0.41%
1 kHz to 5 kHz
10 mA to 1 A 0.2% to 0.9%
50 Hz
10 A to 100 A 1.6 % to 0.51%
100 A to 550 A 0.51 % With Current Coil
5. AC Power 50 Hz
Single Phase$ 120V/240V Using Fluke 5500A
0.01A to 10A 1.15 % to 0.15 % Calibrator by Direct Method
0.2PF to UPF
Lead/Lag
6. Power Factor$ 50 Hz
0.2 -1-0.2 PF lead/lag 0.001-0.003PF Using Fluke 5500A
Calibrator by Direct Method
8. Capacitance$ 1 kHz
0.35 nF to 10 nF 4.21% to 0.7% Using Fluke 5500 A
10 nF to 10uF 0.7 % to 0.4 % Calibrator / Decade by
Direct Method
11. Oscilloscope 5mV to 55V @1kHz 1.8 % to 2.3% Using Fluke 5500 A
Amplitude$ (p-p) (Square / Sine Calibrator by Direct Method
/Triangle
Wave Signal)
II. MEASURE
8. Temperature Simulaton$
(Temperature Indicators/ Recorders/ Controllers)
E Type (-) 30C to 1000C 0.22C Using Keithely 2000 6
Thermocouple DMM, Digital Temperature
J Type (-) 30C to 1200C 0.32C Readout 1529 by Direct
Thermocouple Method
K Type (-)30C to 1372C 0.53C
Thermocouple
N Type (-)30C to 1300C 0.35C
Thermocouple
R Type 50C to 1767C 0.80C
Thermocouple
S Type 50C to 1767C 0.80C
Thermocouple
MECHANICAL CALIBRATION
1. Calipers$
(External/ Internal)
Vernier / Digital / Dial )
L.C. 0.01 mm 0 to 200 mm 7.2 m Using Caliper Checker /
0 to 300 mm 7.6 m Long Slip Gauges K Grade
0 to 600 mm 8.7 m and Slip Gauges 0 Grade
2. Depth Gauge$
(Vernier / Digital / Dial)
L.C. 0.0 mm 0 to 200 mm 6.4 m Using Long Slip Gauges K
0 to 300 mm 6.6 m Grade and Slip Gauges 0
0 to 600 mm 8.6 m Grade
4. Pistol Calipers$
L.C. 0.01 mm 0 to 100 mm 6.0 m Using Grade0 and 1 Slip
Gauges
5. Micrometers$
(Mechanical / Digital /
Digimatic )
L.C. 0.01 mm 0 to 100 mm 5.9 m Using Grade 0 and 1 Slip
100 mm to 300 mm 6.1 m Gauges / Long Slip
300 mm to 600 mm 7.5 m Gauges K Grade / ULM
600 mm to 1000 mm 9.0 m
7. Internal / Stick
Micrometers$
L.C. 0.01 mm 0 to 200 mm 6.1 m Using ULM / Slip Gauges
200 mm to 500 mm 7.1 m 0 and Long Slip Gauges
K Grade
8. Depth Micrometers$
L.C. 0.01 mm 0 to 50 mm 5.8 m Using Grade 0 and 1 Slip
50 mm to 150 mm 5.9 m Gauges / Long Slip
150 mm to 300 mm 6.1 m Gauges K Grade
II. WEIGHTS
1. Spring Balance$
L.C. 20 g 500 g to 5 kg 0.02 kg Using M1 Class
Standard weights
L.C. 50 g >5 kg to 10 kg 0.03 kg
1. Tachometer$ 1000 rpm to 3000 rpm 0.57 % rdg Using Digital Tachometer
( Contact Type ) Calibration Unit
2. Tachometer / Rpm 1000 rpm to 20000 rpm 0.06 % rdg Using Digital Tachometer
Indicator With Sensor$ Calibration Unit
( Non Contact Type )
2. Vacuum Transmitter$ 0.11 bar to 0.87 bar 0.09 % rdg Using Pneumatic Dead
weight tester
V. WEIGHING SCALE AND BALANCE
1. Electronic Weighing
Balance
L.C. 1 g 0 to 5 kg 0.30 g Using M1 Class
L.C. 5 g > 5 kg to 30 kg 0.67 g Standard weights
L.C. 10 g > 30 kg to 60 kg 1.30 g
L.C. 0.02 kg > 60 kg to 150 kg 12.0 g
THERMAL CALIBRATION
I. TEMPERATURE
3. Deep Freezer, (-) 40 C to 600 C 1.71 C Using PRT with Indicator &
Incubator, Oven, 600 C to 1200 C 3.11 C S-Type Thermocouple with
Furnace, Autoclave, Read Unit
Temperature Baths
etc. at Single
Specified Position
Laboratory can also calibrate instruments/devices of coarser resolution / least count within the accredited
range using same reference standard/ master equipment under the scope of accreditation.
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing
actual environmental conditions and master equipment used.