Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

1. DC Voltage$ 100V to 1mV 3.47% to 0.37% Using Fluke 5500 A


1 mV to 10 mV 0.37% to 0.04% Calibrator by Direct Method
10 mV to 100 mV 0.04% to 0.01%
100 mV to 1 V 0.01% to 0.01%
1 V to 100 V 0.01% to 0.01%
100 V to 1000 V 0.01%

2. AC Voltage$ 10 Hz to 1 kHz
1 mV to 10 mV 2.7% to 0.64% Using Fluke 5500 A
10 mV to 100 mV 0.64% to 0.35% Calibrator by Direct Method
100 mV to 1 V 0.35% to 0.20%
1 V to 10 V 0.20% to 0.05%

1 kHz to 100 kHz


10 mV to 100 mV 0.08% to 0.79%
100 mV to 1 V 0.08% to 0.47%

1 kHz to 90 kHz
1 V to 10 V 0.04% to 0.47%

45 Hz to 1 kHz
10 V to 100 V 0.2 % to 0.07%
100 V to 1000 V 0.07%

1 kHz to 18 kHz
10 V to 100 V 0.05% to 0.14%

1 kHz to 10 kHz
100 V to 1000 V 0.07% to 0.3%

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 2 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

3. DC Current$ 10 A to 100 mA 0.8% to 0.02% Using Fluke 5500 A


100 mA to 1 A 0.02% to 0.04% Calibrator by Direct Method
1A to 10 A 0.04% to 0.07%

10A to 550 A 0.88 % to 0.5% With Current Coil

4. AC Current$ 45 Hz to 1 kHz
30A to 10mA 1.13% to 0.2% Using Fluke 5500A
10 mA to 100 mA 0.2% to 0.5% Calibrator
100 mA to 1 A 0.5% to 0.3%
1 A to 10 A 0.3% to 0.41%

1 kHz to 5 kHz
10 mA to 1 A 0.2% to 0.9%

50 Hz
10 A to 100 A 1.6 % to 0.51%
100 A to 550 A 0.51 % With Current Coil

5. AC Power 50 Hz
Single Phase$ 120V/240V Using Fluke 5500A
0.01A to 10A 1.15 % to 0.15 % Calibrator by Direct Method
0.2PF to UPF
Lead/Lag

6. Power Factor$ 50 Hz
0.2 -1-0.2 PF lead/lag 0.001-0.003PF Using Fluke 5500A
Calibrator by Direct Method

7. Frequency$ 10Hz to 1MHz 0.6 % to 0.06 % Using Fluke 5500A


Calibrator by Direct Method

8. Capacitance$ 1 kHz
0.35 nF to 10 nF 4.21% to 0.7% Using Fluke 5500 A
10 nF to 10uF 0.7 % to 0.4 % Calibrator / Decade by
Direct Method

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 3 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

100 Hz 0.4 % to 1.3 % Using Capacitance Box


1.09 F to 1.1 mF Agronic CDB6 by Direct
Method

9. Inductance$ 1 kHz Using Decade Inductance


1mH to 10 H 4.8 % to 2.6 % Box Agronic CDB 6 by
Direct Method

10. DC Resistance$ 100 m to 1 0.12% Using Tinsley ZX-74 E


1 to 100 0.12 to 0.03% Decade Resistance Box
100 to 10 k 0.03% to 0.02% Fluke 5500 A Calibrator /
10 k to 1 M 0.02%
1 M to 10 M 0.02% to 0.08% Cropico Decade Tera ohm
10M to 100M 0.08% to 0.6% Box
100 M to 1 G 0.6% to 2.4%
1 G to 1 T 2.4%

11. Oscilloscope 5mV to 55V @1kHz 1.8 % to 2.3% Using Fluke 5500 A
Amplitude$ (p-p) (Square / Sine Calibrator by Direct Method
/Triangle
Wave Signal)

Timer Marker 5 ns to 5 s 1.2 %


Bandwidth$ upto 200 MHz 3.6%
(50 kHz Ref)

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 4 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

12. Temperature Simulaton$


(Temperature Indicators/ Recorders/ Controllers)
E Type Thermocouple (-) 30C to 1000C 0.25C Fluke 5500 A Calibrator
J Type Thermocouple (-)30C to 1200C 0.3C
K Type Thermocouple (-)30C to 1372C 0.47C
N Type Thermocouple (-)30C to 1300C 0.32C
R Type Thermocouple 50C to 1767C 0.7C
S Type Thermocouple 50C to 1767C 0.5C
T Type Thermocouple (-)30C to 400C 0.3C
RTD (-)199C to 800C 0.3C

II. MEASURE

1. DC Voltage$ 10 mV to 100 mV 0.05% to 0.01% Using Keithley 6 DMM


100 mV to 1 V 0.01 to 0.004% 2000 by Direct Method
1 V to 100 V 0.004% to 0.006%
100 V to 1000 V 0.006% to 0.01%

1 kV to 12 kV 2.75% Using Fluke HV Probe with


4 DMM by Direct Method

2. AC Voltage$ 50 Hz Using Keithley 6 DMM


5 mV to 10 V 0.42% to 0.1% 2000
10 V to 100 V 0.1%
100 V to 750 V 0.1 %

0.75 kV to 28 kV 6.1% Using Fluke HV Probe with


4 DMM by Direct Method

3. DC Current$ 1 mA to 100 mA 0.1% Using Keithley 6 DMM


100 mA to 3 A 0.14 % 2000 by Direct Method

3 A to 15 A 0.14 % to 0.08% Using Ballantine Current


Shunt with Keithley 6
DMM 2000 by Direct
Method

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 5 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

4. AC Current$ 50Hz Using Keithley 6 DMM


10 mA to 100 mA 4.8% to 0.45% 2000 by Direct Method
100 mA to 3 A 0.45% to 0.24%

3A to 10 A 0.24% to 0.14% Using Keithley 6 DMM


2000, Ballantine Current
Shunt by Direct Method

5. DC Resistance$ 10 to 100 0.5% to 0.02% Using Keithley 6 DMM


100 to 10 k 0.02 % to 0.01% 2000 by Direct Method
10 k to 1 M 0.01 %
1 M to 10 M 0.01% to 0.05%
10M to 100M 0.05% to 0.2%

6. Time$ 1 sec to 10 sec 0.15 sec Using Aplab Universal Time


10 sec to 90 min 0.15 sec and Frequency Counter 119
by Direct Method

7. Frequency$ 10 Hz to 100 Hz 0.06% to 0.01% Using Keithley 2000 6


100Hz to 500kHz 0.01% to 0.02% DMM by Direct Method

8. Temperature Simulaton$
(Temperature Indicators/ Recorders/ Controllers)
E Type (-) 30C to 1000C 0.22C Using Keithely 2000 6
Thermocouple DMM, Digital Temperature
J Type (-) 30C to 1200C 0.32C Readout 1529 by Direct
Thermocouple Method
K Type (-)30C to 1372C 0.53C
Thermocouple
N Type (-)30C to 1300C 0.35C
Thermocouple
R Type 50C to 1767C 0.80C
Thermocouple
S Type 50C to 1767C 0.80C
Thermocouple

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 6 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

T Type (-) 30C to 400C 0.3C


Thermocouple

9. DC Voltage 1kV to 12kv 2.8% Using Fluke HV Probe with


4 DMM by Direct Method

10. AC Voltage 0.5kV to 28kV 6.0% Using Fluke HV Probe with


4 DMM by Direct Method

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 7 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

MECHANICAL CALIBRATION

I. Dimension (Basic Measuring Instrument, Gauge etc.)

1. Calipers$
(External/ Internal)
Vernier / Digital / Dial )
L.C. 0.01 mm 0 to 200 mm 7.2 m Using Caliper Checker /
0 to 300 mm 7.6 m Long Slip Gauges K Grade
0 to 600 mm 8.7 m and Slip Gauges 0 Grade

L.C. 0.02 mm 0 to 1000 mm 14.6 m Using Caliper Checker /


0 to 2000 mm 20.7 m Long Slip Gauges K Grade
and Slip Gauges 0 Grade

2. Depth Gauge$
(Vernier / Digital / Dial)
L.C. 0.0 mm 0 to 200 mm 6.4 m Using Long Slip Gauges K
0 to 300 mm 6.6 m Grade and Slip Gauges 0
0 to 600 mm 8.6 m Grade

3. Vernier And Dial /


Digital Height Gauge$
L.C. 0.001 mm 0 to 600 mm 7.8 m Using Height Master / Slip
0 to 1000 mm 9.7 m Gauges 0 and Long Slip
Gauges K Grade

4. Pistol Calipers$
L.C. 0.01 mm 0 to 100 mm 6.0 m Using Grade0 and 1 Slip
Gauges

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 8 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

5. Micrometers$
(Mechanical / Digital /
Digimatic )
L.C. 0.01 mm 0 to 100 mm 5.9 m Using Grade 0 and 1 Slip
100 mm to 300 mm 6.1 m Gauges / Long Slip
300 mm to 600 mm 7.5 m Gauges K Grade / ULM
600 mm to 1000 mm 9.0 m

L.C. 0.001 mm 0 to 25 mm 0.7 m Using Grade 0 and 1 Slip


25 mm to 50 mm 0.8 m Gauges / Long Slip
50 mm to 100 mm 1.1 m Gauges K Grade / ULM
100 mm to 150 mm 1.5 m
150 mm to 200 mm 1.9 m

6. Master Rods / Setting 0 to 100 mm 1.2 m Using ULM / Grade 0 Slip


Rods$ 100 mm to 300 mm 2.0 m Gauges / Long Slip
300 mm to 600 mm 4.8 m Gauges K Grade / Digital
600 mm to 1000 mm 8.7 m Dial Gauge

7. Internal / Stick
Micrometers$
L.C. 0.01 mm 0 to 200 mm 6.1 m Using ULM / Slip Gauges
200 mm to 500 mm 7.1 m 0 and Long Slip Gauges
K Grade

8. Depth Micrometers$
L.C. 0.01 mm 0 to 50 mm 5.8 m Using Grade 0 and 1 Slip
50 mm to 150 mm 5.9 m Gauges / Long Slip
150 mm to 300 mm 6.1 m Gauges K Grade

9. Three Pin Micrometer$


L.C. 0.001 mm 15 mm to 100 mm 3.3 m Using Plain Ring Gauges

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 9 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

10. Gauge Blocks 0 to 25 mm 0.14 m Using Electronic Slip


Grade 0, Gade1 25 mm to 100 mm 0.2 m Gauge Comparator & Slip
and Grade 2$ 100 mm to 500 mm 4.5 m Gauges K Grade / ULM &
Long Slip Gauges K
Grade

11. Bore Dial Gauge$


L.C. 0.001 mm 0 to 150 mm 1.3 m Using ULM / Long Slip
150 mm to 400 mm 2.9 m Gauges KGrade

12. Thickness Gauge$


(Digital/Dial)
L.C. 0.001 mm 0 to 25 mm 0.7 m Using Grade0 and 1 Slip
Gauges

13. Dial Gauges Plungers


Type$
(Dial /Digital )
L.C. 0.01 mm 0 to 25 mm 1.52 m Using Digital Dial Gauge
comparator / ULM

14. Lever Dial Gauges$


L.C. 0.001 mm 0 to 1 mm 1.5 m Using Digital Dial Gauge
L.C. 0.002 mm 0 to 0.2 mm 1.8 m comparator / ULM

15. Thread Plug Gauges$ 5 mm to 50 mm 1.4 m Using ULM


50 mm to 100 mm 1.4 m
100 mm to 150 mm 1.45 m
150 mm to 200 mm 1.56 m

16. Thread Ring Gauges$ 5 mm to 50 mm 1.8 m Using ULM


50 mm to 100 mm 1.78 m
100 mm to 150 mm 2.79 m
150 mm to 200 mm 2.8 m

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 10 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

17. Plain Plug Gauges$ Up to 100 mm 1.0 m Using ULM


100 mm to 150 mm 1.1 m
150 mm to 200 mm 1.2 m

18. Plain Ring Gauges$ 5 mm to 50 mm 1.76 m Using ULM


50 mm to 100 mm 1.8 m
100 mm to 150 mm 2.79 m
150 mm to 200 mm 2.80 m

19. Snap Gauges$ Up to 50 mm 1.76 m Using ULM


50 mm to 100 mm 1.78 m
100 mm to 150 mm 2.79 m
150 mm to 200 mm 2.80 m

20. Thread Measuring 0.17 mm to 6.35 mm 1.0 m Using ULM


Wires$

21. Feeler Gauges$ Up to 1 mm 0.91 m Using ULM

22. Foils$ Up to 1 mm 0.91 m Using ULM

23. Radius Gauge Set$ Up to 25 mm 6.7 m Using Profile Projector

24. Pitch Gauge$ 0.25 mm to 10 mm 2.8 m Using Profile Projector

25. Bevel Protractor$ 0 to 180 3.6 min Using Profile Projector

26. Combination Set$ 0 to 180 34.7 min Using Profile Projector

27. Test Sieves$ Up to 2 mm 2.8 m Using Profile Projector

28. Sprit Level$


Base Length Up to
Sensitivity 29 m / meter Using Electronic Frame
300 mm
L.C.: 0.02m/m Level and Surface Plate

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 11 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

29. Surface Plate Up to 5000 mm longer


L W
side 2.5 Using Electronic Frame
B Level
L = Length in mm
W = Width in mm
B = Base length = 200
mm

30. Digital Comparator


Probe / Dial
Calibration Tester$
L.C. 0.0001 mm 0 to 25 mm 0.94 m Using ULM

31. Universal Length 0 to 100 mm 0.54 m Using Grade K Slip


Measuring Machine$ 100 mm to 200 mm 1.21 m Gauges & Long Slip
Gauges K Grade

II. WEIGHTS

1. Spring Balance$
L.C. 20 g 500 g to 5 kg 0.02 kg Using M1 Class
Standard weights
L.C. 50 g >5 kg to 10 kg 0.03 kg

L.C. 100 g >10 kg to 20 kg 0.06 kg

L.C. 200 g >20 kg to 50 kg 0.12 kg

L.C. 500 g >50 kg to 100 kg 0.30 kg

L.C. 1 kg >100 kg to 150 kg 0.60 kg

III. ACCELERATION AND SPEED

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 12 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

1. Tachometer$ 1000 rpm to 3000 rpm 0.57 % rdg Using Digital Tachometer
( Contact Type ) Calibration Unit

2. Tachometer / Rpm 1000 rpm to 20000 rpm 0.06 % rdg Using Digital Tachometer
Indicator With Sensor$ Calibration Unit
( Non Contact Type )

IV. PRESSURE INDICATING DEVICES

1. Pressure Gauges & 0 to 70 bar 1.65 % rdg Using Digital Pressure


Pressure Transmitter# Calibrator
(Hydraulic) 50 bar to 1000 bar 0.31 % rdg

4 bar to 34 bar 0.32 % rdg

68 bar to 686 bar 0.19 % rdg Using Hydraulic Dead


Weight Tester

2. Vacuum Transmitter$ 0.11 bar to 0.87 bar 0.09 % rdg Using Pneumatic Dead
weight tester
V. WEIGHING SCALE AND BALANCE

1. Electronic Weighing
Balance
L.C. 1 g 0 to 5 kg 0.30 g Using M1 Class
L.C. 5 g > 5 kg to 30 kg 0.67 g Standard weights
L.C. 10 g > 30 kg to 60 kg 1.30 g
L.C. 0.02 kg > 60 kg to 150 kg 12.0 g

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 13 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

THERMAL CALIBRATION

I. TEMPERATURE

1. Liquid In-Glass (-) 20 C to 150C 0.41 C Using Four-wire PRT with


Thermometer, RTD, 150 C to 600 C 0.42 C Indicator with High/Low
Thermocouples, Temperature Liquid Bath
Temperature Sensors
With and without
Indicator$

2. RTD, Thermocouple, 600 C to 1000 C 2.06 C Using S-type


Temperature Sensors Thermocouple and Read
with and without Unit using Dry Block
Indicator$ Furnace

3. Deep Freezer, (-) 40 C to 600 C 1.71 C Using PRT with Indicator &
Incubator, Oven, 600 C to 1200 C 3.11 C S-Type Thermocouple with
Furnace, Autoclave, Read Unit
Temperature Baths
etc. at Single
Specified Position

Dheeraj Chawla Avijit Das


Convenor Program Director
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalyam Road, K. R. Puram P.O., Coimbatore, Tamil Nadu

Accreditation Standard ISO/IEC 17025: 2005


Certificate Number CC-2164 (in lieu of C-0169, C-0135, C-0912) Page 14 of 14
Validity 11.12.2016 to 10.12.2018 Last Amended on 27.02.2017

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability ()

4. Deep Freezer, (-) 40 C to 400 C 1.92 C Using RTD with Data


Incubator, Oven, Logger
Furnace,
Temperature Baths
Etc. At Five Positions
Including Centre
(Spatial Mapping)
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%
$
Only in Permanent Laboratory

Only for Site Calibration

Laboratory can also calibrate instruments/devices of coarser resolution / least count within the accredited
range using same reference standard/ master equipment under the scope of accreditation.
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing
actual environmental conditions and master equipment used.