Temperature Probe
Enter/On Key
Quality Control
Liquid QC Procedure
Verify GEM Code, fill cuvette with the control, and
insert the cuvette into the instrument.
Press the Main Menu key, then Calibration #1,
press Enter/On.
Press Cuvette GEM Code #1, press Enter/On.
The currently defined GEM Code is displayed on
the Ready screen for visual verification each time
a sample is analyzed.
Quality Control
Liquid QC Procedure
Selecting the GEM Code option from the
Calibration Menu displays the GEM Code
currently defined. Press Cancel or OK to leave the
GEM Code as currently displayed, or change the
current value.
After entering a four digit numeric GEM Code, the
operator must confirm the entry in a subsequent
display.
GEM OPL presents an error message if the GEM
Code entered does not match one of those
defined by the manufacturer.
Quality Control
Liquid QC Procedure
Hold the ampule by the top above the break line and
shake the ampule with gentle tapping. Restore all
liquid to the bottom of the ampule with gentle tapping.
Use appropriate commercially available ampule
openers to protect fingers while opening.
Carefully snap open the ampule.
Quality Control
Liquid QC Procedure
Sample contents immediately after opening.
Verify that QC results are within manufacturers
package insert ranges.
If values are not within range, repeat QC.
If still not passing range, contact POCT Services.
Do not use instrument for patient testing if QC fails.
Document all QC results on the Liquid QC Log.
Document all corrective actions taken.
Specimen Collection and Stability
Verify patient identification with two identifiers (patients
name and date of birth) before drawing the blood
sample.
Observe universal precautions; wear gloves and other
personal protective equipment as appropriate.
If there is the potential for specimen mix-up, the
specimen must be labeled with patients full name and
medical record number.
Specimen Collection and Stability
Hold here
Specimen Collection and Stability
Verify that the GEM Code is correct for the cuvette bag
in use. If not, you must change/update the GEM Code
through the Calibration Menu (Refer to GEM OPL
Operators Manual on how to change GEM Code on the
GEM OPL instrument).
Collect and prepare the heparinized blood sample as
described.
Patient Sample Analysis Procedure
Discard if bulging
Patient Sample Analysis Procedure
Sample #1
tHb O2Hb COHb MetHb
17.0 39.7 56.6 1.1
g/dL % % %
Patient Sample Analysis Procedure