Escolar Documentos
Profissional Documentos
Cultura Documentos
IM/8002 Issue 11
8002
ABB
The Company
We are an established world force in the design and manufacture EN ISO 9001:2000
of instrumentation for industrial process control, flow
measurement, gas and liquid analysis and environmental
D
RE
GI
E
STER
applications.
Cert. No. Q5907
As a part of ABB, a world leader in process automation
technology, we offer customers application expertise, service and
support worldwide. EN 29001 (ISO 9001)
We are committed to teamwork, high quality manufacturing,
advanced technology and unrivalled service and support.
The quality, accuracy and performance of the Company’s Lenno, Italy – Cert. No. 9/90A
products result from over 100 years experience, combined with a
continuous program of innovative design and development to
incorporate the latest technology.
The UKAS Calibration Laboratory No. 0255 is just one of ten flow
calibration plants operated by the Company and is indicative of
our dedication to quality and accuracy.
Information in this manual is intended only to assist our customers in the efficient operation of our
equipment. Use of this manual for any other purpose is specifically prohibited and its contents are not to
be reproduced in full or part without prior approval of the Technical Publications Department.
1
1 INTRODUCTION
The Ammonia Probe, Model 8002, is a high The probe may also be used as a sensor in
sensitivity gas sensing probe developed by industrial monitors, for example the Model 8232
ABB. It measures the partial pressure, and Ammonia Monitor, an industrial continuous flow
hence concentration of ammonia in aqueous analyser incorporating automatic
solution. Ammonium concentrations can be standardisation.
measured after the sample has been treated
with alkaline buffer to generate ammonia. The Model 8002 is widely used for the determination
probe has a Nernstian response up to 1000mg of ammonia in boiler water, river water, treated
NH3l–1 or 10,000mg NH3l–1 with modification of and untreated sewage, etc., (see References 2,
the internal filling solution. The lower limit has not 3, 5, 6, 7, 9 and 10). Total nitrogen in the forms of
been determined as it is set by the purity of ammonia in the residue from the Kjeldahl
available water. digestion may also be measured quickly and
easily with the probe without the need for the
Model 8002 should be used in conjunction with usual final distillation step, (see References 4
an expanded scale pH meter such as Model and 8). The internal filling solution of the probe
7046; for convenience the direct reading must be modified in this application and in the
concentration scales may be used with the measurement of other samples containing a
function switches set to monovalent anion. The high concentration of dissolved species.
probe is supplied with membranes and filling
solution and a flow-through cap which enables
continuous analyses of up to 60 samples per
hour, when incorporated into a suitable flow
system.
SP
PART AR
ES
8002 KIT
No
265
Ma
de
in En
gla
nd
t
FIL IN
AM LING TER
voile Colou MO SO AL
N
t w r ch NIA LU
he an
PA n so ge P TION
MAD RT lutio from RO
E INNo. 80n mad yello BE
EN 02-2 e al to w
GLA 40 ka
ND line
2
2 GENERAL INFORMATION
Flow-through cap, Part No. 8002-830. This item f) Hold the probe upright and fill the probe
is supplied with the kit. It allows samples to be through the filling hole with the internal filling
pumped directly onto the sensing membrane of solution provided until there is a 50 to 60mm
the probe. A simple closed flow system may be depth of solution inside. Wipe any excess
constructed using the cap. Further details are filling solution from the body.
given in Section 3.
g) Tap the end of the probe with a finger to
2.3 Probe Assembly Instructions dislodge any air bubbles which may have
The probe takes only a few minutes to become trapped between the end of the
assemble, but for the best results the glass electrode and the membrane. Then
instructions must be followed carefully. screw the glass electrode down four turns to
the flush position again and then a further 1/3
a) Unscrew the end cap from the probe body. to 1/2 turn. The tip of the electrode will now be
Rinse out the probe body with distilled or pressing against the membrane. In some
deionised water and allow to drain. circumstances, particularly if the electrode
response is found to be unusually slow, the
b) Remove the teat from the glass electrode and screw may be given a further 1/4 turn.
rinse the electrode with distilled or deionised
water. Dry with a paper tissue. Note. Screwing the electrode down too far
may puncture the membrane.
c) Insert the glass electrode into the body and
screw it in until the top of the electrode is flush h) Push the probe cap onto the top of the
with the end face of the probe. Note the probe body.
number on the electrode cap in line with the
mark on the probe body. Unscrew the
electrode cap four full turns.
3
…2 GENERAL INFORMATION
Filling
Solution
Aperture
Body
Body Seal
Membrane
Washer
pH Electrode
End Cap
Reference
Element
Filling
Solution
Membrane
W asher
From Exhaust
mixing
coil
Clear PVC tubing
3 mm bore x 4.5 mm O/D
4
2 GENERAL INFORMATION 3 PROBE MEASUREMENTS
On no account should the end of the probe be Alkaline samples slowly lose ammonia so
allowed to dry out, should this accidentally measurements in open beakers are not always
occur the performance of the probe may satisfactory. Thus it is suggested that the
sometimes be restored by loosening and samples are contained in 100ml conical flasks to
tightening the electrode retaining nut to allow reduce losses. The probe body will fit closely into
more filling solution to flow between the glass the top of such a flask but losses may be further
electrode and the membrane. If this procedure is reduced by slipping the larger of the ‘O’ rings
not successful the membrane and filling solution from the spares kit over the probe stem so that it
need replacement. makes a seal with the neck of the flask.
If the probe is to be returned to its box it must be All samples should be stirred during
dismantled: firstly the glass electrode should be measurement; this may conveniently be done
unscrewed, then the rest of the probe emptied, using a magnetic stirrer. With some stirrers it will
rinsed and drained. The glass electrode should be necessary to put a sheet of thermal insulating
be stored with its end in a teat containing a material, such as polystyrene sheet, underneath
neutral buffer solution, taking care not to cover the beaker to prevent heating of the sample by
the reference element. the stirrer motor.
2.5 Internal Filling Solution After immersing the tip of the probe in a stirred
The internal filling solution incorporates a sample check that no air bubbles are trapped on
coloured indicator which is normally yellow. The the end of the probe. Allow time for the probe
indicator changes to blue if the probe potential to reach a steady value; the
membrane, or membrane seal, allows alkaline equilibration time will be dependent on the
reagent to leak into the probe. If this occurs the concentration of the ammonia in the sample. At
membrane and the filling solution should be a sample concentration of 10–3M (17mg NH3l–1)
replaced. Take care to tighten the end cap the time taken by the probe to reach its
sufficiently to provide a good seal. The solution equilibrium potential should be under two
keeps indefinitely. minutes, although it may take longer with certain
types of sample.
5
…3 PROBE MEASUREMENTS
The probe is sensitive to the osmotic pressure of 3.2 Buffer Solution – Sample pH
samples and special attention needs to be paid The two functions of the buffer solution are to
to this in the analysis of samples containing a adjust the pH of the samples to greater than 11
high total concentration of dissolved species so that all the ammonium ions are converted to
(i.e. the sum of the concentrations of all anions, ammonia and to fix the total concentration of
or cations and non-dissociated species). dissolved species in the buffered sample at
If this total concentration in the buffered sample approximately 0.2M. In order to ensure virtually
presented to the probe is greater than 0.25M, complete conversion of the ammonium ions it is
the probe potential will drift. To overcome this necessary to have a sufficient excess of hydroxyl
problem either the samples should be diluted or ions over the ammonium ions: to ensure the
the internal filling solution of the probe modified, correct pH the final hydroxyl concentration
as explained in a later section. should be at least two or three times the
maximum ammonium ion concentration
The probe is sensitive to changes in both expected (both concentrations expressed as
ambient temperature and the temperature of molarities).
samples, hence it is recommended that the
probe is not used in direct sunlight or in a series Thus a 1M sodium hydroxide solution (NaOH)
of samples or standards at different added to samples at a volume ratio of 1 : 10, is
temperatures. For the most accurate and a suitable buffer for samples containing up to 5 x
reproducible results the probe should be kept at 10–2M NH4+ (ca. 1000mgl–1), provided the
a constant temperature by use of a jacketed cell samples do not contain a high concentration of
or by immersion of the lower half of the probe in dissolved species. However for a sample which
a water bath (using the probe, fitted with a flow- contains less than 5 x 10–3M NH4+ (ca.100mgl–
through cap, in a flow system). 1) but has a total concentration of dissolved
6
3 PROBE MEASUREMENTS…
100
90
80
70
60
% Free Ammonia
+
pK of NH4 50
40
30
20
10
pH 0
6 7 8 9 10 11 12 13 14
-200
mV -100
7
…3 PROBE MEASUREMENTS
8
3 PROBE MEASUREMENTS…
9
…3 PROBE MEASUREMENTS
10
3 PROBE MEASUREMENTS…
90%
Ammonia 10mg/l
Probe Model
8002 80%
70%
60%
Chart
100mV f.s.d.
50%
60cm/h
2mg/l
40%
30%
1mg/l 1mg/l
20%
10%
0%
Traces Obtained from an Automatic Flow System
Probe
Pump
Air
Mixer
Buffer
W ash
W aste
Sampler
11
…3 PROBE MEASUREMENTS
3.8 The Known Addition Method 3.9 Modification of the Internal Filling
There are some advantages in using the Known Solution
Addition Method of analysis for occasional For analysis of samples containing very high
samples with ammonia concentrations greater concentrations of ammonia the internal filling
than the lower Nernstian limit in the experimental solutions must be modified to prevent the
conditions used. This method reduces the osmotic transfer of water across the membrane
number of standard solutions required to one; (see Reference 1). The internal filling solution
the concentration of this standard solution provided in the kit consists principally of 0.1M
should be between ten and a hundred times that ammonium chloride saturated with silver
of the sample. The experimental slope factor (S chloride.
mV/pNH3) of the probe should be known.
In the case of measurements in strong ammonia
To analyse a sample by this method, pipette an solutions the molarity of ammonium chloride in
aliquot, Vs ml, of alkaline sample into a beaker the internal filling solution should be increased
and measure the potential of the probe in this so that it is at least two or three times the
solution, E1 mV. Add a known volume, Va ml, of molarity of the strongest ammonia solution to be
the standard solution (where 10Va VS) of analysed. Sufficient silver nitrate solution should
concentration Ca mol l–1 and after stirring be added to ensure that the solution is saturated
measure the new potential of the probe, E2 mV. with silver chloride. The buffer for treating the
It should be so arranged that the potential samples should be chosen so that the buffered
change is several millivolts. The concentration of samples have pH greater than 11 and the total
ammonia in the original alkaline sample, Cs mol concentration of dissolved species matches that
l–1, may then be calculated from the formula: in the new filling solution.
CaV a
For measurements of samples containing a high
V a +V s total concentration of dissolved species,
Cs =
sufficient inert electrolyte, e.g. potassium nitrate,
[
antilog10 E -E
2
S
1
[
Vs
(V a +V s ) must be added to the filling solution so that the
concentrations in the internal filling solution and
buffered samples are equal. Such modification
of the filling solution is necessary in the analysis
of, for example, seawater or Kjeldahl digests. An
example of the calculation of the correct
concentration of inert electrolyte to add is given
in the subsequent sections.
12
3 PROBE MEASUREMENTS…
a) The initial dilution of the residue after e) If the digestion mixture contains a mercury
digestion should be as large as possible to catalyst, sodium iodide should be added to
simplify the procedure. If the dilution factor the alkaline buffer to prevent interference;
is such that the total concentration of for a copper catalyst EDTA should be
dissolved species in the solution presented added to prevent precipitation.
exceeds 0.25M then the probe filling
solution should be modified to prevent Note. Many of these requirements may be
osmotic transfer of water through the met by the use of a continuous flow system,
membrane. as already noted.
13
…3 PROBE MEASUREMENTS
0.90M Na2SO4
0.20M NaOH
14
3 PROBE MEASUREMENTS…
The above calculation assumes that the • ageing of the glass electrode – see Section
3.12.
solutions exhibit ideal behaviour, but this will not
be the case in practice. Thus, in general, some
3.12 Ageing of the Glass Electrode
drift in probe potential will still be observable and
After the probe has been in use for some
further adjustment of the probe filling solution months, the performance of the glass electrode
may be necessary. If the probe potential drifts in may eventually deteriorate due to its continued
the direction of lower ammonia concentration, use in weakly buffered solutions at near neutral
then the filling solution requires more Na2SO4 pH. The response of the probe becomes
and vice versa. It is recommended that ci for sluggish and the response slope drops. The
the filling solution be changed in 5% increments electrode may frequently be restored to its initial
until the drift is adequately small. condition by soaking for 12 hours in 0.1M
hydrochloric acid. The performance of the
The matching of total concentration of dissolved electrode may be checked as described below.
species on either side of the probe membrane is
more critical as the solutions become more
concentrated; thus some preliminary dilution of
the residue after digestion is advantageous.
15
…3 PROBE MEASUREMENTS 4 THEORY
3.13 Checking the Performance of The probe responds to the partial pressure of
the Glass Electrode ammonia in the sample solution. When the
The glass electrode may be checked probe is immersed in a sample, ammonia is
independently of the probe with a laboratory transferred across the gas-permeable
calomel reference electrode (e.g. Model 1370 membrane until the partial pressure of the gas in
210) by testing the pair in pH buffer solutions. the thin film of solution between the glass
electrode and the membrane is equal to the
The glass electrode should only be immersed to partial pressure in the sample. The internal filling
a depth of 5 to 10mm. There should be no liquid solution contains ammonium ions. The ratio of
contact with the reference element. the amount of ammonia to the activity of these
ions in the thin film gives rise to a characteristic
Connect the electrodes to a pH meter and pH which is measured by the glass electrode/
calibrate with pH buffers in the usual way. The reference electrode system. The value of this pH
glass electrode may be found to be somewhat is a function of the chemical reactions in the
slower in response than a conventional bulb solution. As a result of the equilibria the probe
type electrode, but adequate scale length exhibits a potential in ammonia solutions which
should be obtained, typically better than 90% of may be represented by:
the theoretical slope value.
2.3RT log [NH ]
E = E –
3.14 Aging of the Reference Element F 10 3
16
5 SPECIFICATION 6 SPARES & ACCESSORIES
Life
Under normal operating conditions the probe life
is at least 60 days without renewal of membrane
Dimensions
Total Length 152mm
Length below cap 105mm
Stem diameter 17.5mm
17
APPENDIX
To prepare, dilute 100ml A.R. concentrated When clean, rinse with distilled water and
ammonia solution, s.g. 0.88, with100ml distilled immediately transfer to the plating solution such
water and stir. that the reference element is completely
immersed.
Nitric acid, 25% vol./vol. (200ml)
A.1.3 Chloridisation
To prepare, cautiously pour 50ml A.R. Attach the screen of the combination electrode
concentrated nitric acid, s.g. 1.42, into 150ml cable to the negative terminal of the constant
distilled water, stirring continuously. Allow to current supply and immerse the lower 5cm of
cool before use. the electrode in the plating solution so that the
reference element is covered. Immerse a silver
Caution. Strong ammonia solutions and counter electrode in the plating solution and
strong nitric acid are both very poisonous connect to the positive terminal of the supply.
and corrosive. Care should be taken not to
inhale the fumes of either or to get any on Pass a current of 2mA for approximately 30
skin or clothing. Any slight spillage on skin, seconds. Carefully tap the electrode to remove
clothing or working surface should be all bubbles and then reverse the connections to
washed away with plenty of water. the supply (i.e. the electrode to be plated is
connected to the positive terminal).
2) Hardware:
Constant current supply, 2mA d.c. output Pass a current of 2mA for 30 minutes after which
time the electrode will be a dark brown or grey
Silver wire (counter electrode) colour. A more uniform coating may be obtained
by employing moderate stirring during the
1 beaker process
18
REFERENCES
19
NOTES
20
PRODUCTS & CUSTOMER SUPPORT
Products Customer Support
Automation Systems We provide a comprehensive after sales service
• for the following industries: via a Worldwide Service Organization. Contact
– Chemical & Pharmaceutical one of the following offices for details on your
– Food & Beverage nearest Service and Repair Centre.
– Manufacturing
– Metals and Minerals UK
– Oil, Gas & Petrochemical ABB Limited
– Pulp and Paper Tel: +44 (0)1453 826661
Fax: +44 (0)1453 829671
Drives and Motors
• AC and DC Drives, AC and DC USA
Machines, AC motors to 1kV ABB Inc.
• Drive systems Tel: +1 215 674 6000
• Force Measurement Fax: +1 215 674 7183
• Servo Drives
Flexible Automation
• Industrial Robots and Robot Systems
Flow Measurement
• Electromagnetic Flowmeters
• Mass Flowmeters
• Turbine Flowmeters
• Wedge Flow Elements
Process Analytics
• Process Gas Analysis
• Systems Integration
Transmitters
• Pressure
• Temperature
• Level
• Interface Modules