Check the above mentioned probable causes. If the alarm still remains, check
the cable Cb4.
The voltage measured at Converting unit terminals Z9 and Z10 should be
approx. +22V and 0V respectively relative to ground. The current in cable Cb4
should be 4 - 20mA.
If the pressure is zero the values LRef, HRef and Pres in line 8.2 should be
greater than:
The Pres value should increase proportionally to the pressure, be equal to LRef
at 0 bar and equal to HRef at the high end of the pressure transmitter range.
Abnormal sample water pressure in the analysing unit. When this alarm is
detected, measuring is stopped.
Probable cause in case of low-pressure failure: The sample inlet probe valve is
closed, the pipe between sample probe and the sample pump is clogged or the
overboard discharge line is empty.
Probable cause in case of high-pressure failure: The analysing unit outlet valve
or the sample outlet probe valve is closed.
Check the above mentioned probable causes. If the alarm still remains check
the cable Cb4. The voltage measured at Converting unit terminals Z9 and Z10
should be approx. 22V and 0V respectively relative to ground. The current in
cable Cb4 should be 4 - 20 mA.
If the pressure is zero the values LRef, HRef and Pres in line 8.2 should be
greater than:
60
JOWA CLEANTOIL 9000 Operations and Technical Manual.
The Pres value should increase proportionally to the pressure, be equal to Lref
at 0 bar and be equal to Href at the high end of the pressure transmitter range.
Unexpected value detected, while monitoring fresh water, in the measuring unit
during cleaning.
Probable cause: The measuring unit is dirty or dry. Detector data is displayed in
order to simplify trouble shooting.
Clean the measuring unit according to the instruction in paragraph 2.8 “Manual
Cleaning of the Analysing Unit” at page 51. The alarm remains as long as the
measuring cell is dry or the sample flow is missing.
This indicates low overboard discharge flow. Probable causes: The overboard
discharge pump is not running. Overboard and slop tank valves are closed.
Defective flow meter or flow meter cables.
Check all conditions for overboard pumping. If the alarm still remains, check the
flow meter by measuring the current signal 4-20mA that appears in cable Cb3.
Check the Converting unit flow-meter power supply, if a Zener supply barrier is
used, by measuring the voltage between terminals ZF1/2+ to ZF1/2-. The
voltage should be between 20 and 25 VDC.
61
JOWA CLEANTOIL 9000 Operations and Technical Manual.
Page headline.
Static pressure measured in the analysing unit with closed water valve and
closed analysing valve. Consequently the static pressure is the pressure
obtained without flow in the analysing unit. The proper sample point valves are
of course supposed to be opened.
Low and high pressure limits. The value to check is calculated from the
difference between the pressure measured while fresh water is flowing and the
static pressure, during periods of cleaning. If any of the limit values are
exceeded this alarm is generated:
74
JOWA CLEANTOIL 9000 Operations and Technical Manual.
Low and high pressure limits. The value to check is calculated from the
difference between the pressure measured while sample water is flowing and
the static pressure. If any of the limit values are exceeded this alarm is
generated:
Line 6.5 displays the input signal from the pressure transmitter. The range 20 -
100% usually corresponds to 0 - 25 bar pressure.
The second line displays a formula that calculates the pressure that
corresponds to a 25 bar transmitter input signal. Before you enter a new formula
check the range of your transmitter because the range has varied during the
years that has passed.
75