Você está na página 1de 12

ISLH

Laboratory Hematology 7:89-100


© 2001 Carden Jennings Publishing Co., Ltd.

Official Publication

The Differential Cell Count

BEREND HOUWEN
Department of Pathology and Laboratory Medicine, Loma Linda University School of Medicine, Loma Linda, California

ABSTRACT WBC differential count has become widely accepted and


used by clinicians, and is generally considered to yield clini-
Differential cell counts have provided extensive data cally useful information in health and disease.
through more than a century of laboratory hematology. As a laboratory test, the differential cell count is unlike
Routine bench morphology count continues to play a role almost all other tests performed in the clinical laboratory. It
in analyses of blood and other tissues such as bone marrow, provides a pathology report on the morphological appear-
but for many purposes it is being replaced by other tech- ance of blood cells and on the frequency in which they
nologies that provide greater precision and consistency. appear, not in a descriptive manner but in a highly stylized,
This discussion of the differential cell count traces the quantitative format. It is called a “differential” count because
development of electronic analysis and looks at the recent all enumerated cells are brought into a common mathemati-
impact of monoclonal antibody–based flow cytometry. The cal context, ie, a frequency distribution for the different cell
introduction of the extended differential count has led to a types present. The total number of cells counted may vary,
rapid increase in electronic analysis, with resultant but in most clinical situations it is 100, and specific cell types
increased accuracy, lower cost, and more efficient turn- are commonly expressed as percentages of the total count.
around times. Moreover, the extended differential is leading Thus there is within this context a particular distribution of
to acquisition of effective data on complex specimens and is cell types for healthy, normal individuals, and there are dis-
paving the way for new parameters in the field. Lab. tribution patterns associated with disease states.
Hematol. 2001;7:89-100. Over time the information derived from the WBC differ-
ential count has become a cornerstone in laboratory hema-
KEY WORDS: Differential cell count · Flow tology and is widely used for screening, case finding, diagnosis,
cytometry · Extended differential and monitoring of hematologic and nonhematologic disor-
cell count ders. It is used, for example, in diagnosis of bacterial or viral
infectious disease; evaluation of allergic conditions; diagnosis
and monitoring of malignant disease such as leukemia; and
INTRODUCTION staging of HIV infection. The WBC differential count is
often also used to monitor a patient’s progress and/or
Differential cell counts typically are considered enumera- response to treatment. For example, a decrease in elevated
tions of the different white blood cell (WBC) types circulat- neutrophil counts in patients treated for (bacterial) infectious
ing in the blood. However, differential counts can also apply disease is generally regarded as a positive response.
to other cell lineages such as red blood cells or to cells within However, the test is not without its problems. The tradi-
the same lineage but at different maturation levels. The tional procedure for the differential WBC count by manual
microscopy is time consuming and labor intensive and is
therefore one of the most expensive routine tests in the clini-
Correspondence and reprint requests: Berend Houwen, MD, PhD, Corpo- cal hematology laboratory. Furthermore, the 100-cell differ-
rate Medical Director, Beckman Coulter, Inc., 4300 N. Harbor Blvd., P.O. ential count is often criticized for its statistical shortcomings
Box 3100, Fullerton, CA 92834-3100; 714-773-3566; fax: 714-773-3570 because of its small sampling size. Despite these problems,
(e-mail: bhouwen@beckman.com). most clinicians consider the manual WBC differential count
Received November 17, 2000; accepted November 28, 2000 effective, perhaps not as a direct decision-making tool, but

89
90 B. Houwen

more as a source of collateral information, like taking a film preparation and staining, as well as the small number of
patient’s body temperature. The manual differential count cells analyzed and the resultant inherent statistical problems,
thus shows a combination of poor statistical reliability, high have led in most situations to the replacement of image analy-
expense, and definite clinical utility, making it a prime target sis systems by alternative technologies.
for automation and alternate approaches. Those efforts date Popularity for image analysis still persists, primarily in
back to the 1960s, and to date continue to push the bound- Japan and in certain European countries; but, given the cur-
aries of differential cell counting. rent sophistication of flow-based analyzers, its role in the
Newer technologies have enabled WBC differential clinical laboratory environment is much reduced. The future
counts to be obtained electronically from automated analyz- development of such systems could become more directed
ers, at much lower cost, with greater statistical reliability, toward high-quality image capture, storage, and distribution
often with greater overall accuracy, and with faster availability to other workstations for review.
of results to the clinician. Moreover, these technologies have
enabled laboratorians to measure cellular differentiation and Flow-Based Analysis Systems
to quantify cell types other than WBC, with sometimes sur- Another altogether different approach to differential
prising clinical utility as outcomes. counting was the development of a flow-based system that
could analyze blood cells in suspension [3]. Such a system
THE EVOLUTION OF THE DIFFERENTIAL CELL requires the use of a combination of a center stream, contain-
COUNT ing the sample’s cells, and an outside sheath fluid surround-
ing the sample stream. The sheath flow narrows the sample
The introduction of aniline dyes in the second half of the stream and forces the cells into single file, thus enabling pas-
nineteenth century made it possible to study individual sage of single cells through a sensing zone, optical or otherwise.
blood cells by microscopy after a small amount of blood had The next steps were the development of multiparameter
been placed onto glass slides, dried and fixed, and then analysis on a cell by combining the signals from several
stained. Eosin was the first of these dyes in 1856, followed by biosensors, for instance forward and side light scatter, and
hematoxylin in 1865 and later by the metachromatic the development of fluorescent signals [4] (Figure 1). These
Romanowsky dyes. Amazingly, the practice of manual leuko- steps led to the development of flow cytometers and later of
cyte differential counting essentially has not changed in more fluorescence activated cell sorters (FACS) [5]. The earliest
than a century, although a few refinements have been made. developments in automated flow-based hematology differen-
These consist mainly of improvements in dye quality, stain- tial cell counters were based on continuous-flow systems,
ing procedures, and microscopes and automation of slide where individual samples were separated by air bubbles, simi-
preparation and staining. lar to chemistry analyzers then in existence [6]. The combi-
nation of continuous flow with sample separation and
Image Analysis enzyme cytochemical treatment of patient samples led to the
Attempts to automate WBC differential counting began in first automated 5-cell differential introduced in the mid
the 1960s by image analysis using computer-based algorithms 1970s [7,8]. Although results showed impressive improve-
[1,2]. Building automated microscope systems capable of ment in statistical terms over manual WBC differential
morphological analysis appeared at the time an attractive counting, it was not practical to operate the analyzer on a
solution for replacing manual differential counting: a consis- 24-hour basis because of prohibitive reagent cost and because
tent algorithm to replace subjective observer interpretation of calibration of the continuous flow system proved delicate.
morphology, coupled with automated sample handling to cre-
ate a hands off, “walk-away” system. In practice, most image Developments in Automation
analysis systems were capable of identifying the majority if Many technological developments have occurred since,
not all blood cells in normal specimens, but failed to do so in aimed at making differential WBC counting accessible by
preparations from patients with significant disease. Conse- automated electronic analysis. The primary goal of these
quently, the operator was required to interact with the system developments has been to provide a differential count of the
to resolve identification of the remaining, typically difficult- 5 “normally appearing” WBCs (neutrophils, lymphocytes,
to-classify cells. The reality with image analysis systems was monocytes, eosinophils, and basophils) or, at a minimum, to
that they were slow, even compared to manual microscopy, provide a screening differential that could eliminate the need
and that they were not truly walk-away systems because of for manual microscopy. In addition to the combination of
often poorly executed automation and cell “pattern” recogni- continuous flow and enzyme cytochemistry, laboratories
tion algorithms requiring almost constant operator interac- developed systems that analyzed WBC based on light scatter-
tion. In addition, a number of countries have regulatory ing properties (optical flow systems) or on cell sizing (imped-
requirements that mandate review of all results before their ance). At first these systems yielded partial differential
release. These factors plus stringent requirements for blood counts: enumeration of lymphocytes and granulocytes, with
Differential Cell Count 91

FIGURE 1. Schematic drawing of a flow cytometer with detection capability of forward scatter (representing cell volume; laser wave-
length); side scatter (representing cellular contents; laser wavelength); and/or side fluorescence (light emitted by a fluorescent
probe; longer than laser wavelength). Depending on the use of beam splitters (dichroic mirrors) and band filters, emitted light of
different wavelengths can be measured separately. Light sources for flow cytometers are usually lasers, mostly emitting light at
488 nm. Semiconductor lasers, emitting at 633 nm, have recently become used in routine, clinical instruments.

a third population of monocytes, usually mixed in with other cal methods and combinations of technologies such as imped-
cells [9-11] (Figure 2). Eosinophils and basophils were not enu- ance, radio frequency, and optical sensors have been used for
merated separately by these systems. These so-called 3-part a variety of analyzers from different manufacturers. For repre-
differentials were used for screening purposes, typically in sentation of the data, the histograms used for the mostly single-
combination with their own flagging systems and review cri- parameter 3-part differential were largely abandoned and were
teria for selected parameters derived from the blood count. replaced by bivariate scatterplots, often in multiple represen-
Such rule-based systems were surprisingly good at predicting tations (Figure 3). Over the years major improvements have
abnormalities in patient specimens requiring manual review been made in the methods that analyze the 5-part differential
or enumeration (Table 1). In unflagged specimens, the dif- multiparameter signals. Adaptable cluster analysis algorithms
ferential count results would simply consist of the electronic have largely displaced the earlier use of fixed discriminators as
3-part count, with the inherent assumption that the eosinophil boundaries between cell populations. This change was made
and basophil count would be normal and therefore should possible by huge increases in computing power available at
not require separate reporting. The major benefit for the lab- low cost. The development of decision assist systems is seen as
oratory was that a significant reduction of 30% to 45% in further improvement in the laboratory’s ability to interpret
manual reviews and differential counts could be obtained electronic differential findings.
without compromising patient care. The results were lower
costs and higher throughput (shorter turn-around times). In
the study cited in Table 1 as well as in other studies, there
was no significant difference in outcome between optical- or
impedance-based systems [12,13].

Development of the 5-Cell Method


The next development came with further evolution of the
enzyme cytochemical 5-cell method, and with the availability FIGURE 2. Histogram of a 3-part WBC (white blood cell) dif-
of other systems capable of 5-cell differentials, using different ferential with 3 discriminators. The first discriminator sepa-
forms of analysis and cytochemical treatment [14-20]. These rates debris from lymphocytes; the second separates lympho-
developments were made possible by the introduction of cytes from a central, mixed-cell population, consisting most-
reagents that interact with WBC and make these recognizable ly of monocytes and eosinophils; and a third discriminator
on the basis of cell volume and cellular contents. Purely opti- separates granulocytes from the central population.
92 B. Houwen

TABLE 1. Review Criteria of the 3-Part Differential Cell Count*

Reference Interval Review Criteria

White blood count, 109/L 4.0-10.0 <3.5 or >12.0


Hemoglobin, g/L
Female 120-160 <100 or >180 g/L
Male 140-180
Mean cell volume, fL 80-100 <80 or >102 fL
Mean cell hemoglobin concentration, g/L 310-360 <310 or >360 g/L
Platelets, 109/L 140-440 <100 or >50
Lymphocytes, % 16-48 <16 or >48% FIGURE 3. Bivariate scattergram of a 5-cell WBC differential
Monocytes, % 1-10 <1 or >15% based on an adaptive cluster algorithm. These algorithms have
Granulocytes, % 35-75 <45 or >78% largely replaced older methods of separating cell populations
by fixed or moving discriminators. This scattergram contains
*These review criteria were used to screen for morphological abnor- the following populations: debris; lymphocytes; monocytes;
malities in addition to the analyzer’s own flagging criteria (2194 specimens basophils; neutrophilic granulocytes; and eosinophils. SFL
tested; 45.6% of specimens were without distributional or morphological indicates side fluorescence; SSC, side scatter.
abnormalities). In this study [12], evaluating 3 different instruments, the
false normal rate ranged from 1.5% to 2.1% and the false abnormal rate
ranged from 1.1% to 3.8%. The reduction in manual review of blood
specimens using these rules was 30%. The EDC will make immature granulocytes and nucleated
red blood cells an integral part of the electronic differential
and will replace flagging for these frequently observed cell
Both 3- and 5-part differential systems are equipped with types. While atypical or variant lymphocytes are also expected
flagging algorithms, alerting the user to possible morphologi- to be part of the EDC, at this moment left shift is not, due to
cal or other abnormalities that require review of the data or lack of agreement on the morphology of left-shifted cells.
some other form of intervention such as morphology review. Other cell types that can be included in the EDC are blast
The advantage of 5-part over 3-part systems is a potential for cells and hematopoietic progenitor cells. Once the EDC is
further reduction in manual review rates. In reality, however, fully implemented on hematology analyzers, manual review
laboratory adherence to a combination of the 5-part differen- rates are expected to drop to very low numbers and should be
tial, the analyzer’s own flagging system, and blood count cri- needed only in highly abnormal specimens, such as those
teria such as those as set forth in Table 1 often led to an from patients with hematological malignancies.
increased review rate. A more rational approach in analyzers An expected problem for EDC is validation of its perfor-
with a 5-part differential count plus flagging system would mance: many of the cell types represented in the EDC are
be to discontinue the use of review criteria based on blood often present in blood in low numbers. Yet these low num-
count parameters. Those were developed primarily as a safety bers can constitute the critical decision level which requires
net for 3-part differential systems that generally have rather validation of analyzer performance at 1 per 100 WBC or
primitive differential count flags. Several current electronic less. In these situations, it is not possible to use the manual
differential counters have flags that are user-adjustable or 400-cell reference count for validation because of the enor-
that flag at several levels. Thus the user is enabled to adapt mous imprecision at such levels (Table 3). Alternative refer-
the system to the laboratory’s specific caseloads and needs, ence methods, primarily monoclonal antibody based flow
and to decrease further unnecessary morphology reviews. cytometry, will have to be used.1

Extended Differential Count Replacement of Morphological Analysis


A logical next step in this evolution is the replacement of Although there remains a role for morphological analysis
instrument flagging of abnormal cell populations by an actual of blood and other tissues such as bone marrow, for many
count of such cells: the extended differential count (EDC).
The EDC includes cell types other than the 5 normally
present in blood (Table 2). A task force, initiated by the
International Society for Laboratory Hematology, with repre- 1
Manufacturers must validate electronic differential counts by comparison
sentation from experts, industry, and regulators, has formu- against a “reference” method or, at minimum, against a predicate device
lated an approach that can enable clinical implementation of that was validated earlier; this validation is part of the regulations for use as
a clinical device. Analyzer performance must be equal to or better than the
the EDC. The task force has also made recommendations performance of reference or predicate method, and must be tested in speci-
that blood cell counts from bone marrow and body fluid mens from normal, healthy individuals as well as in specimens from
specimens be included in the EDC. patients with diseases.
Differential Cell Count 93

TABLE 2. Cell Types Included in the Traditional and the Extended Electronic Differential Count (Proportional)

Traditional Reference Interval, no. and %* Extended Limit Frequency†

Neutrophils 1.89-5.32 38-68 Left shift Varies NA‡


Lymphocytes 1.39-3.19 22-50 Immature granulocytes >1% 11%
Monocytes 0.30-0.76 5-11.4 Nucleated red blood cells >1% 2%
Eosinophils 0.05-0.35 0.8-5.3 Blasts >1% 0.7%
Basophils 0.01-0.06 0.2-1.0 Atypical lymphocytes >10% 0.3%
Hematopoietic progenitor cells NA
Other 0.2%

*Based on 328 medical students, male (195) and female (133). Analyzed in the author’s laboratory by Sysmex XE-2100 hematology analyzer (Sysmex
Corp., Kobe, Japan).
†Clinical frequency in patient specimens (n = 24,579) [21].
‡Neither “band” or “stab” cells nor left shifted cells are currently included in the extended differential count, but are included in the neutrophil group.
NA indicates not available.

routine purposes it is gradually being replaced by other All these factors plus much improved instrumentation
technologies. In most developed countries the traditional and computer analysis have made flow cytometry a superb
investigation of WBC and red blood cell (RBC) morphology tool for clinical use. Additionally, flow cytometry is capable
on a blood film has largely been replaced by some form of of providing information that can offer insights into cellular
electronic analysis in combination with cytochemical or physiology, such as cell metabolism, apoptosis and drug-
enzyme cytochemical treatment of blood cells. Recent devel- response, not available by traditional morphological analysis
opments in monoclonal antibody–based flow cytometry [22,23]. Another methodology that is replacing parts of tra-
have made significant inroads in other areas of traditionally ditional morphology practices is molecular analysis of tissues
morphological blood, bone marrow, and lymph node analy- or even of individual cells. Examples include detection of
sis, particularly for malignancies. The degree to which flow minimal residual disease, multiple drug resistance detection
cytometry has replaced morphological analysis varies from by polymerase chain reaction (PCR) and diagnosis of chro-
institution to institution, even in a single country, but has mosomal abnormalities in individual cells by fluorescent
been substantial over the past decade and a half. This trend in situ hybridization (FISH) techniques [24,25].
has been supported by the development of a wide range of As the role for traditional manual microscopy–based
monoclonal antibodies and an increasingly better under- analysis diminishes, electronic cell analysis by hematology
standing of the cellular physiology of cluster designations analyzers, flow cytometry using monoclonal antibodies and
(CDs) and epitopes the antibodies are directed against. Not other probes and molecular diagnostics will play increasingly
only has the diagnostic potential for monoclonal antibodies important roles in diagnostic approaches involving blood
grown enormously, but the objectivity of flow-based results cells, whether in blood, bone marrow, or body fluids.
often is preferable over a morphological interpretation col-
ored by the observer’s own subjectivity. Lastly, the relative STATISTICAL CONSIDERATIONS
ease of analyzing large numbers of cells gives data obtained
by flow cytometry much greater precision than that As discussed earlier, low numbers of cells enumerated in a
obtained by manual microscopy. differential cell count lead to uncertainty about the results

TABLE 3. Statistical Uncertainty of the 100-Cell White Blood Cell (WBC) Differential Count and Its Effect on Assumptions of Cell Type Distri-
butions*

Cell type No.† 100 500 1000 10,000

Basophil 1 0-5.4* 0.3-2.3* 0.5-1.8* 0.8-1.3*


Eosinophil 4 1.1-9.9* 2.5-6.1* 2.9-5.4* 3.6-4.5
Monocyte 9 4.2-16.4* 6.6-11.9* 7.3-10.9 8.4-9.6
Lymphocyte 45 35-55.3* 40.6-49.5 41.9-48.1 44.0-46.0
Neutrophil 60 49.7-69.7* 55.6-64.3 56.9-63.1 59.0-61.0

*Results marked with an asterisk (*) show at least one confidence limit outside reference intervals. (For reference intervals see Table 2.)
†The proportional cell counts for the individual WBC types have been chosen arbitrarily, and these do not represent a 100-cell differential.
94 B. Houwen

obtained. Typically, in a manual microscopic WBC differen- THE REFERENCE WBC DIFFERENTIAL COUNT
tial count a total of 100 cells are counted. Therefore, almost The reference WBC differential count was introduced to
always far fewer than 100 cells are counted for each of the enable manufacturers and users to validate automated elec-
main 5 WBC categories as present in blood films from nor- tronic WBC differentials [28]. The method is used widely
mal, healthy adult individuals (listed in order of decreasing for evaluation of electronic differential cell counts and leads
frequency: neutrophils, lymphocytes, monocytes, eosinophils to satisfactory results for most normal cell types.2 When the
and basophils). Although more than 50 of 100 WBC may be manual reference method was developed, it was well recog-
neutrophils in a normal blood film, typically only 1 basophil nized that this could have all or at least some of the same
per 100 WBC will be found. Imprecision in routine, manual problems as the routine differential count. To contain some
differential counting is significant, and its wide confidence of the imprecision of the routine count, it was decided that
limits seriously affect clinical decision-making [26]. Even if 2 independent observers would each count 200 WBC on
the total number of cells enumerated were doubled or 2 separately prepared blood films. Instrument results should
quadrupled, reproducibility would be just mediocre even for not be available to the observers, and their manual results
the most frequently appearing cell types, and still poor for should be adjudicated (if necessary) by a third independent
cells that are less frequently observed. Assuming that the fre- observer who has access to all information, including clinical
quency of a certain cell type (N) remains constant regardless data. For this purpose the availability of 1 or 2 additional
of how many cells are counted on a blood film, the standard slides is recommended. Problems are frequently encountered
deviation of its estimate will show stochastic convergence with the monocyte count, in part due to distributional prob-
toward zero based on 1/√N with increasing sample size of N. lems, in part due to misidentification. The process of judg-
However, this calculation requires large numbers of cells to ment allows for resolving discrepancies between observers as
be counted, far exceeding the potential of even the most well as between instrument and observers, whether in mor-
persistent morphologist. The statistical uncertainty of the phology or counts, and overcomes issues of sample switch-
100-cell WBC differential count and its effect on assump- ing, clerical errors, and other errors.
tions of cell type distribution in a patient sample is shown in Despite these improvements, the method is still too
Table 3. The 5 WBC categories (with arbitrarily chosen cell imprecise to allow method comparison for cell categories
counts for each cell type) and 95% confidence limits are that are present in low frequency, such as basophils. When
shown at various total WBC differential counts. The cell applied to cell types with similar or lower frequency, the
counts represent typical levels within the reference interval method essentially falls apart and alternative methods must
for each cell type. When 100 WBC are counted, all cell types be used. These are not always readily available, however, and
show an upper confidence limit that is outside the reference there is a great need for the development and homologation
interval. Even at 10,000 cells counted, the cell type present of such methods, if they are to be made available at all.
with the lowest frequency (basophils) shows an upper confi- Several attempts have been made to develop monoclonal
dence limit exceeding the reference interval. Such results antibody–based flow cytometry methods for such purposes.
would be deemed abnormal and, if no morphological abnor- The rationale behind this is simple: when cells appear in the
malities were found, distributionally abnormal. Patients blood in low or very low frequency, the method comparison
might therefore be subjected to unnecessary further testing, becomes in fact rare event analysis. Poisson error will be large
driving up health care costs without any benefit to patients. when the sample size is small, and it will be impossible to
Similarly, actual distributional abnormalities can easily go judge from side-by-side comparison whether either method
undetected by manual microscopy. Many automated analyz- is correct. The solution is to increase the sample size, easily
ers, however, count around 8,000 WBC; the model in Table 3 done by using flow cytometry.
clearly shows the statistical superiority of the electronic cell Monoclonal antibody–based 5-cell differential counts can
count over the manual count. It is unlikely that electronic readily and reproducibly be obtained by using CD45, CD14
WBC differential counts result in significant errors in deter- and CD13 or other antibody combinations [29, 30]. Similar
mining distributional abnormalities. approaches have been developed for nucleated red blood cells
A classic study illustrates this point. The study involved (NRBC) and reactive immature granulocytes [31-34].
73 examiners each performing a 100-cell WBC differential An example of the problems caused by imprecision of the
count on 496 patient specimens with distributional abnor- manual count for low cell counts is shown in Figure 4. The
malities present in 169 specimens (34%). Distributional purpose of the illustrated investigation was to test low-range
abnormalities were “detected” in 1.6% of samples without a performance of NRBC identification (<5 per 100 WBC) by
distributional abnormality (false positives), while in 14.1%
of samples with such abnormality present none was found by
the examiners (false negatives) [27]. This is probably one of 2
This discussion focuses on reference WBC differential counting, while
the most poignant examples of the limitations of the manual recognizing that instrument validation comprises more than just a manual,
100-cell differential count. microscopic WBC differential count.
Differential Cell Count 95

the area of the blood film suitable for morphological analysis


because part of the blood film will be too thick and part will
be too thin (Figure 5). For a representative count of all cells,
it is necessary to cover the entire width of the examination
area, because WBCs do not distribute evenly over the glass
surface. Large cells such as monocytes and neutrophils are
pushed toward the tail end of the blood film (“feathered
edge”) and to the sides. Thus there can be under-representa-
tion in the differential count if the side areas of the film are
not included in the examination. A battlement track scan
will avoid such problems and at the same time will avoid
repeat analysis of the same cells.
It is important to realize that for all types of blood film
analysis cells, the following issues apply. The process by which
the cells are applied to the slide (spreading, smearing, spin-
ning) may have caused changes in their distribution and
appearance. Second, the cells have been flattened from their
original spherical appearance by air drying. Finally, the cells’
contents have undergone changes because of fixation and sub-
sequent staining. Thus, the morphological analyses of fixed
and stained blood cells carry a large number of artifacts, and
direct comparison with untreated live cells in suspension may
not always be possible, or is not very good. A typical example
is found in the fragile lymphocytes from patients with chronic
FIGURE 4. Scatterplots showing the correlation between lymphocytic leukemia often damaged beyond recognition by
instrument NRBC (nucleated red blood cell) enumeration (y- the methods used to apply blood cells onto the slides.
axis), visual NRBC counts from blood films (A, 400 white
blood cells [WBCs] counted) and flow cytometric NRBC
counts on the same specimens (B, 20,000 events counted).
CELL IDENTIFICATION ISSUES
The interval chosen is 0 to 10 NRBC per 100 WBC. Flow Manual differential counts are subject to a variety of cell
cytometric enumeration resulted in much better correlation
identification errors, some of which are attributable to the
than the visual method due to higher numbers of NRBC
counted and better NRBC identification.

a new hematology analyzer. On 37 blood samples with


known presence of NRBC, the results of the analyzer count
was compared with the results of the manual 400-cell refer-
ence count and with the results of a 20,000-cell flow cytome-
try count. It was impossible to decide on the basis of a
manual reference 400-WBC differential count whether the
analyzer was correct or not. However, the results from com-
parison with the flow cytometry method clearly showed that
the analyzer gave the correct counts for NRBC.

BLOOD FILM PREPARATION


Guidelines for blood film preparation and staining can be
found in many laboratory and hematology textbooks and
have recently been reviewed [35]. A well prepared and FIGURE 5. Schematic example of a blood film: the part closest
stained blood film is essential for good-quality morphological to the origin is usually too thick for morphological use, and
analysis of blood cells. In North America, the most commonly the tail end is too thin especially for evaluation of red blood
used method for blood film preparation is wedge-push type cell morphology. During a differential leukocyte count it is
with Wright or Wright-Giemsa staining. May-Grünwald- paramount that white blood cells be counted only once,
Giemsa type staining is more popular in Europe and else- hence the often-used method of a “battlement track,” cover-
where. For good-quality cell recognition it is necessary to use ing the entire width of the blood film.
96 B. Houwen

observer’s training level and experience, others to motivation, microscopic differential count. One of the results of a WBC
fatigue, stress, or level of distraction. Current practices of differential, often reviled by laboratorians but considered by
cross-training of technologists can also lead to problems, many clinicians as very valuable is expressed as a “left shift.”
mainly because of insufficient regular practice of microscopic The morphological identification of left-shifted cells has
analysis. Color blindness can play a role, although this been one of the most inconsistent and controversial aspects
should not a priory disqualify someone as a morphologist. A of the manual differential, and therefore is no longer used in
common practice of “skipping” counting cells the observer is many institutions [37]. Ironically, the term “left shift” finds
unsure about should be strongly discouraged, as doing so its origins in the manual count itself. The devices used to
may lead to ignoring significant morbidity. punch in the numbers for the various cell types had the keys
It should also be recognized that certain cell types lead to for immature WBC types on the left side of the key board,
greater discordance between observers: observers rarely dis- hence “left shift” in a blood specimen when immature WBC
agree in eosinophil counts, but quite often disagree about were found to be present.
monocyte or band neutrophil counts. This discrepancy When observing the hematopoietic “tree,” from primitive
occurs because morphological criteria can easily be defined stem cell to fully differentiated, functional “end” cells, one
for eosinophils, whereas it can be difficult to distinguish a can distinguish the differentiating pathways and can apply a
small monocyte from a large, granular lymphocyte (NK cell). differential count to the functional end products of the tree’s
Generally the poorest reproducibility and accuracy results are branches (Figure 6). For purposes of this discussion, the
obtained for counting cells with different maturation levels terms “differentiation” and “maturation” will be considered
within the same lineage. Examples are band and segmented loosely equivalent, and both are used to indicate an ongoing
neutrophils and, to a lesser degree, immature granulocytes: physiological process of cellular change, either by cell divi-
promyelocytes, myelocytes and metamyelocytes. It has sion plus cellular change, or by a cellular change alone, one
proven to be very difficult to apply morphological criteria that leads from immature to a mature, functional end cell
consistently by multiple observers and even more so across stage. A differential count of functional end cells separates
different institutions or countries. Thus the rather subjec- different lineages and can indicate changes in the distribu-
tive differentiation of cell types by morphology often leads tion of these lineages in a blood or bone marrow sample.
to inconsistent results: there is much less chance for incon- When compared to an established reference interval, this
sistency when electronic cell analysis, monoclonal anti- indication can provide clinically useful information. For
body–based flow cytometry, or image analysis algorithms are example, increased numbers of eosinophils may be the result
applied to WBC populations. of allergies. Increased numbers of lymphocytes, especially
with atypical or variant morphology, may be associated with
APPLICATIONS OF THE DIFFERENTIAL CELL certain viral infections such as infectious mononucleosis.
COUNT Other viral infections, for instance those associated with
human immunodeficiency virus (HIV), can result in low
The WBC differential count applies equally to blood, numbers of circulating lymphocytes.
bone marrow, or body fluids, although cell types may differ It is important to realize that the differential count is not
and the presence of various cell types also may be quantita- restricted to separation of blood cells into different types of
tively different. In most institutions, clinicians use the pro- functional end cells (the lineages); it can also give an indica-
portional cell count of peripheral blood WBC, despite efforts tion of the distribution of cells at different levels of matura-
that have demonstrated a laboratory preference for a count in tion within a single lineage. However, when the differential
absolute numbers. This reluctance to change is probably count is done by purely morphological identification there
caused by the origins of the WBC differential count: results are significant pitfalls. While lineages usually can be easily
obtained on enumeration of 100 WBC on a stained blood differentiated from each other (the “horizontal” differential
film, without a recalculation to absolute counts. count; Figure 6A), it is not so simple to apply stepwise mor-
The foundation for the WBC differential count dates phological criteria (ie, segmented neutrophils versus band or
back to the days of Paul Ehrlich and others, more than a cen- stab cells) onto a biological continuum of differentiation
tury ago, with the application of aniline dyes to blood cell within a single lineage (the “vertical” differential count; Fig-
preparations, rendering blood cells visible for microscopic ure 6B). For this reason, left shift is difficult to quantitate by
identification [36]. And although some of the WBC differ- morphological means. Proper identification of left shift can
ential count applications may seem antiquated now, and be useful in identifying bacterial infection, especially in
although reliance by clinicians on its numbers may not newborns and in the elderly population, but the lack of con-
always be fully warranted, there is no denying that the test sensus about the morphological characteristics of what con-
can deliver powerful supportive data for clinical information stitutes left shift has severely damaged its reputation among
and decision making. This ambivalence is further under- laboratorians [36]. Another form of vertical differential
scored by the morphological uncertainties of the manual, counting is represented by the identification of even more
Differential Cell Count 97

A B

C D

FIGURE 6. The hematopoietic “tree,” from primitive stem cell to fully differentiated “end cells.” A, Schematic representation of the
stem cell model for haematopoiesis, in this case the formation of eosinophils. Whereas the pluripotent stem cell has self-renewal
capacity without entering a differentiation pathway (“lineage commitment”), progenitor cells have a limited proliferative potential.
At first, expansion predominates gradually replaced by differentiation without mitotic activity, resulting in mature end cells enter-
ing the circulation. B, Although the exact processes of expansion and differentiation differ per lineage, the basic process remains
the same. Typically, under steady state conditions blood cells are not released from the haematopoietic marrow until mature.
Release of earlier stages points toward accelerated or disordered production. C, Under normal, steady-state conditions the differen-
tial count of cells in the peripheral blood is restricted to mature, functional end cells: the “horizontal” differential of blood cells at
the same level of maturation. D, During accelerated or disordered production of cell types of one or more lineages, immature cells
belonging to that/those lineages may be found in the peripheral blood. The enumeration of immature cells in relation to its
mature, functional end product is deemed “vertical” differential count. Although rarely used extensively, normal and pathological
haematopoietic marrow will lend itself by nature to vertical differential counts. The vertical differential count of a particular lin-
eage is a reflection of the dynamics of cell production within that lineage.

immature neutrophils, such as promyelocytes, myelocytes, Recently, several efforts have been made to analyze bone
and metamyelocytes. This can lead to difficulties among marrow by hematology blood cell analyzers. One applica-
these specific cell types themselves, but as a group (“imma- tion is to determine more effectively the myeloid to ery-
ture granulocytes”) they are sufficiently different in morphol- throid ratio (M:E). Others include blast cell counts and
ogy that they can be separated from mature neutrophils and myeloid differentiation patterns [38,39]. New technological
band forms (Figure 6C). developments have made it possible to study maturation
98 B. Houwen

of the most immature reticulocytes as a proportion of all


reticulocytes, themselves immature red blood cells. While it is
well established that reticulocyte counts represent the produc-
tion rate of RBC (erythropoiesis), to many the exact role of
IRF has not been as clearly defined. Based on the kinetic rela-
tionship between changes in IRF and reticulocyte counts it is
evident that the IRF is dissimilar from the reticulocyte count:
changes often do not occur at the same time or on the same
scale. This fact is illustrated in a patient receiving an autolo-
gous bone marrow reinfusion as part of cancer treatment (Fig-
ure 7). After the patient receives the cytotoxic conditioning
regimen, the IRF values drop to below normal before the
reticulocyte counts become below normal. Conversely, as
engraftment takes place the IRF increases to normal before
the reticulocyte counts respond. When such changes and rela-
tionships between parameters are applied to generally accept-
ed interpretations of what governs a production process, it is
clear that the reticulocyte count represents the first derivative
of erythropoiesis, ie, the RBC production rate. Changes in
the IRF thus reflect the rate of change in the RBC production
FIGURE 7. Bivariate plot showing the number of reticulocytes rate, ie, the second derivative of the production process.
in relation to the “vertical” differential, in this case the imma- Greater immaturity indicates an increase in RBC production
ture reticulocyte fraction (IRF). The data are represented on rates and less immaturity a decrease. Similarly, left shifted
a log/log scale for easier interpretation and contain reference neutrophils indicate an increase in the neutrophil production
intervals for the reticulocyte count and IRF. The resulting
rate, typically caused by infection or by administration of
division of the scatterplot into 6 quadrants means that the
hematopoietic growth factors such as G- or GM-CSF.
central quadrant represents normal values for both reticulo-
cyte count and IRF. The data were obtained from a patient When expressed appropriately, ie, as the proportion of
undergoing autologous marrow reinfusion and represents the immature cells within any specific lineage, the vertical differ-
patient’s course during 25 days from initiation of treatment ential ratio can thus become an indication of the second
to full recovery. The first noticeable effect from high-dose derivative of the production process for that lineage: the rate
myeloablative therapy was a drop in IRF, not reticulocytes, of change in its production rate. Second derivative measure-
indicating a severe reduction in marrow erythroid prolifera- ments are important because changes occur in advance of
tion. Reticulocytes continued to mature without mitosis changes in the actual bone marrow output, ie, the specific
from the remaining reservoir of red blood cell precursors cell counts themselves, and in the case of IRF have a strong
until depleted (day 2). After a period of bone marrow sup- predictive value [42,43]. It is exciting that such measure-
pression where both parameters are suppressed, recovery is
ments may become available for myeloid and other lineages,
indicated by (first) a rise in IRF, followed almost 10 days later
although they are certainly not yet mainstream.
by a return of the reticulocyte count to normal. The vertical
differential of peripheral blood cells (reticulocytes), in this Vertical differential counting for the erythroid and
case IRF, can thus reflect events in the hematopoietic marrow myeloid lineage will likely become routinely possible on
relating to cell production in the erythroid lineage. It is hematology analyzers, and the challenge is going to be incor-
important to note that in situations such as this changes in porating such parameters into the laboratory and particularly
IRF precede those in actual reticulocyte counts. making these accessible and useful to clinicians.

CONCLUSIONS
patterns in the erythroid series as well [40-41]. With the
introduction of the EDC in the clinical laboratory it is Electronic cell identification and enumeration has reached
expected that electronic analysis of bone marrow and body a point of sophistication that often matches or surpasses that
fluid specimens will rapidly increase and will replace to a of the routine bench morphology count. WBC differential
major extent manual microscopy. counts obtained electronically have much greater precision
Vertical differential counting might become an important than their manual counterparts, and their accuracy is mostly
tool in assessing maturation and differentiation processes in equal or better. Both cost and turn-around times are lower
the hematopoietic marrow. One of the roles of vertical differ- for electronic differentials. The extended differential will
ential counting can be viewed as similar to that of the imma- enable laboratories to perform counts on complex specimens,
ture reticulocyte fraction (IRF). The IRF expresses the number including bone marrow and body fluid samples. Manual
Differential Cell Count 99

review is likely to become greatly reduced, and new parame- 14. Cornbleet PJ, Myrick D, Levy R. Evaluation of the Coulter STKS
ters may emerge. These could include methods to measure five-part differential. Am J Clin Pathol. 1993;99:72-81.
the kinetics of hematopoiesis without ever doing a bone mar- 15. Kessler C, Sheridan B, Charles C, et al. Performance of the Coul-
row aspirate or biopsy. ter Gen•S System’s white cell differential on an abnormal speci-
Compared to that obtained by electronic differential men data set. Lab Hematol. 1997;3:32-40.
counts, relatively little information is obtained by routine 16. Brigden ML, Page NE, Graydon C. Evaluation of the Sysmex
morphological analysis of blood cell populations and their NE-8000 automated hematology analyzer in a high-volume out-
precursors. Although manual, optical (morphological) WBC patient laboratory. Am J Clin Pathol. 1993;100:618-625.
differential counting has been the mainstay for the analysis 17. Houwen B, Mast BJ, Lebeck LK, Chang L, Hull B. Performance
of WBC subpopulations for much of the twentieth century, evaluation of the Sysmex SE-9000 hematology workstation. Sys-
it has become clear that the limitations posed by morpholog- mex J Int. 1994;4:5-18.
ical analysis of Romanowsky stained blood films are impact- 18. Fournier M, Gireau A, Chretien MC, et al. Laboratory evaluation
ing its potential for clinical use. of the Abbott Cell DYN 3500 5-part differential. Am J Clin
Pathol. 1996;105:286-292.
19. Jones RG, Faust A, Glazier J, Matthews R, Potter B. Cell-DYN
ACKNOWLEDGMENT
4000: utility within the core laboratory structure and preliminary
This article is in large part based on a paper written by the comparison of its expanded differential with the 400-cell manual
author for the International Council for Standardization in differential count. Lab Hematol. 1998;4:34-44.
Haematology. 20. Corberand JX, Segonds C, Fontanilles AM, Cambus JP, Fillola G,
Laharrague P. Evaluation of the Vega haematology analyzer in a
university hospital setting. Clin Lab Haematol. 1999;21:3-10.
REFERENCES
21. Konno T, Kanda T. The total hematology system based on labora-
1. Mahouy G, Lund PR, Chinn S, Barnes RD. The use of automated tory information system and quick turn around time. Sysmex J Int.
image analysis in differential white cell counting. Scand J Haema- 1996;6:113-119.
tol. 1973;10:315-318. 22. Selleri C, Maciejewski JP. The role of FAS-mediated apoptosis in
2. Mui JK, Fu KS, Bacus JW. Automated classification of blood cell chronic myelogenous leukemia. Leuk Lymphoma. 2000;37:283-297.
neutrophils. J Histochem Cytochem. 1977;25:633-640. 23. Garrido SM, Willman C, Appelbaum FR, Banker DE. Three-
3. Crossland-Taylor PJ. A device for counting small particles sus- color versus four-color multiparameter cell cycle analyses of
pended in fluid through a tube. Nature. 1953;171:37. primar y acute myeloid leukemia samples. Cytometr y.
4. Kamentsky LA, Melamed MR, Derman H. Spectrophotometer: a new 2000;42:83-94.
instrument for ultrarapid cell analysis. Science. 1965;150: 630-631. 24. Kobayashi H, Takemura Y, Kawai Y, et al. Competitive reverse
5. Bonner WA, Hulett HR, Sweet RG, et al. Fluorescence activated transcription-polymerase chain reaction assay for quantification of
cell sorter. Rev Sci Instrum. 1972;43:404-409. human multidrug resistance 1 (MDR 1) gene expression in fresh
6. Skeggs LT. An automatic method for colorimetric analysis. Am J leukemic cells. J Lab Clin Med. 2000;135:199-209.
Clin Pathol. 1957;28:311. 25. Seong CM, Giralt S, Kantarjan H, et al. Early detection of relapse
7. Saunders AM. Development of automation of differential leukocyte by hypermetaphase in situ hybridization after allogeneic bone
counts by the use of cytochemistry. Clin Chem. 1972;18:783-788. marrow transplantation for chronic myeloid leukemia. J Clin
8. Mansberg HP, Saunders AM, Groner W. The Hemalog-D white Oncol. 2000;18:1831-1836.
cell differential system. J Histochem Cytochem. 1974;22:711-724. 26. Rümke CL. The imprecision of the ratio of two percentages
9. Johnston CL. Leukocyte screening using the ELT-800WS. Blood observed in differential white blood cell counts. Blood Cells.
Cells. 1985;11:241-255. 1985;11:137-140.
10. Cox CJ, Habermann TM, Payne BA, Klee GG, Pierre RV. Evalua- 27. Koepke JA, Dotson MA, Shifman MA. A critical evaluation of the
tion of the Coulter counter model S-Plus IV. Am J Clin Pathol. manual/visual differential leukocyte counting method. Blood Cells.
1985;84:297-306. 1985;11:173-186.
11. Burgi W, Marti HR. Automated blood count analysis by trimodal 28. Reference Leukocyte Differential Count (Proportional) and Evalu-
size distribution with the Sysmex E-5000. J Clin Chem Clin ation of Instrument Methods. Approved Standard. NCCLS docu-
Biochem. 1989;27:365-368. ment H20-A. Villanova, Pa; 1992.
12. Houwen B, Tisdall PA, Croucher B, McDonald MJ, Fettes G, 29. Lebeck LK, Mast BJ, Houwen B. Flow cytometric white blood cell
Pinto A. Automated histogram differential analysis: a comparison differential: A proposed alternate reference method. Sysm J Int.
of 3 instruments and strategies for clinical implementation. Sysmex 1993;3:61-69.
J. 1985;8:14-21. 30. Hubl W, Wolfbauer G, Andert S, et al. Toward a new reference
13. Pierre RV. Leukocyte differential counting. In: Koepke JA, ed. method for the leukocyte five-part differential. Cytometry.
Practical Laboratory Hematology. New York, NY: Churchill Living- 1997;30:72-84.
stone; 1991:131. 31. Tsuji T, Sakata T, Hamaguchi Y, Wang F-s, Houwen B. A new
100 B. Houwen

rapid flow cytometric method for the enumeration of nucleated 38. D’Onofrio G, Zini G, Tommasi M, et al. Automated analysis of
red blood cells. Cytometry. 1999;37:291-301. bone marrow: Routine implementation and differences from
32. Paterakis G, Kossivas L, Kendall R, Anagnostopoulos N, peripheral blood. Lab Hematol. 1998;4:71-79.
Stavropoulos-Giokas A, Tassiopoulou A. Comparative evaluation 39. Wang F-s, Tsuji T, Sakata T, Hamaguchi Y, Hagbloom R, Houwen
of the erythroblast cell count generated by three-color fluorescence B. Counting and staging of peripheral and cord blood nucleated
flow cytometry, the Abbott Cell DYN 4000 hematology analyzer, red blood cells by flow cytometry [abstract]. Lab Hematol.
and microscopy. Lab Hematol. 1998;4:64-70. 1998;4:96.
33. Lund-Johansen F, Terstappen LW. Differential expression of cell 40. Zini G, Mistretta G, Crollari L, et al. Automated counting of
adhesion molecules during granulocyte maturation. J Leukoc Biol. nucleated red blood cells (NRBC): Evaluation of the Sysmex
1993;54:47-55. XE-2100 system [abstract]. Lab Hematol. 1999;5:94.
34. Fujimoto H, Sakata T, Hamaguchi Y, Shiga S, Wang F-s, 41. Spanish Multicentric Study Group for Hematopoietic Recovery.
Houwen B. Flow cytometric method for enumeration and classi- Flow cytometric reticulocyte quantification in the evaluation of
fication of reactive immature granulocyte populations. Cytometry. hematologic recovery. Eur J Haematol. 1994;53:293-297.
2000;42:371-378. 42. Davis BH. Immature reticulocyte fraction: By any name, a useful
35. Houwen B. Blood film preparation and staining procedures. Lab clinical parameter of erythropoietic activity. Lab Hematol.
Hematol. 2000;6:1-7. 1996;1:3-12.
36. Cornbleet PJ, Novak RW. Classifying segmented and band neu- 43. Wang F-s, Itose Y, Sagoaga O, et al. Development and clinical
trophils. CAP Today. 1994;8:37-41. application of nucleated red blood cell staging with hematology
37. Schumacher H, Glazier J, Ismail S, Mazella F. Bone marrow analy- analyzer XE-2100. 13th International Symposium on Technologi-
sis provided by the Cell DYN 4000 hematology analyzer cal Innovations in Laboratory Hematology; April 2000; Banff,
[abstract]. Lab Hematol. 1998;4:94. Alberta, Canada.

Você também pode gostar