Você está na página 1de 5

Accessed from 208.85.168.

103 by mettler2 on Mon Jun 03 12:43:04 EDT 2013

Second Supplement to USP 36–NF 31 General Information / 〈1251〉 Weighing on an Analytical Balance 6217

microorganisms that pass the 0.45-µm filter are of interest ment, Validation, and Routine Control of a Sterilization
because they present the greatest potential challenge to the Process for Medical Devices. Arlington, VA: AAMI;
sterilizing filtration process. They should be evaluated 2006.
against the upstream bioburden. Bioburden of greatest con- 7. AAMI/ISO. 11737. Sterilization of Medical Devices—Mi-
cern includes Pseudomonas, Brevundimonas, Ralstonia, and crobiological Methods—Part 1: Estimation of Population
Mycoplasma. of Microorganisms on Products. Arlington, VA: AAMI;
2006.
Identification ■2S (USP36)

Recovered microorganisms can be identified to an appro-


priate level using the methods described in Microbial Identifi-
cation, Characterization, and Strain Typing 〈1113〉. Only those
microorganisms that present a potential risk to the product
or the patient require identification to the species level or
beyond. For a detailed discussion of the methods used to
grow and identify microorganisms, see 〈1113〉. It is not nec- 〈1251〉 WEIGHING ON AN
essary for the purposes of evaluating presterilization bi-
oburden to identify all isolates to the species level, although ANALYTICAL BALANCE
this can be helpful in some investigations. Analysts must
conduct all evaluative work with pure cultures, and they
must apply normal microbiology laboratory procedures for
the selection and maintenance of pure cultures.
Change to read:
Bioburden Control
A typical bioburden-control program includes review and ■ INTRODUCTION
analysis of potential sources of contamination as well as
sound process design and preventive and monitoring meas- Weighing is a frequent step in analytical procedures, and
ures. The microbiological contamination-control program the balance is an essential piece of laboratory equipment.
should be developed to identify and control bioburden and The general information described here applies directly to
to assess product risk based on a formal assessment of risk electronic balances used in analytical procedures. Although
modalities. The bioburden risk assessment should result in many portions of the chapter are applicable to all balances,
the establishment of critical control points and should in- some are applicable only to analytical balances. This chapter
clude consideration of the following elements: should not be considered all-inclusive, and other sources of
• Microbiological attributes of materials before steriliza- information (e.g., the US National Institute for Science and
tion and the manufacturing process used for the materi- Technology and balance manufacturers) may be useful and
als (if applicable) applicable when analysts perform a weighing operation or
• Inherent antimicrobial properties of the materials implement a weighing procedure.1,2 The information given
• Time limits for process execution in this chapter is applicable not only to balances used for
• Water activity of the material materials that must be accurately weighed (see Balances
• Environmental conditions within the facility 〈41〉) but also to balances used in all analytical procedures.
• Equipment design and cleaning
• Sanitization, decontamination, and other active micro-
bial control processes (such as prefiltration, tempera- QUALIFICATION
ture, pH, osmolarity, etc.)
Controlling the bioburden of materials and products to be Users should consult Analytical Instrument Qualification
sterilized will ensure conformance to the levels required by 〈1058〉, standard operating procedures, and recommenda-
the sterilization process validation. Additionally, controlling tions from manufacturers when they devise qualification
the bioburden levels of the items to be sterilized assures that plans.
residuals (e.g., allergens, endotoxins, and exotoxins) from
that population will also be controlled. This is important be- Installation
cause direct detection of these materials is challenging.
The balance’s performance depends on the conditions of
the facility where it is installed. Analysts should consult infor-
REFERENCES mation provided by the manufacturer before they install a
balance.
1. 21 CFR 807, Subpart E. Premarket Notification
Procedures. Support surface: The balance should be installed on a
2. FDA. Updated 510(k) sterility review guidance K90-1; solid, level, nonmagnetic surface that minimizes the trans-
final guidance for industry and FDA. 2002. http:// mission of vibration (e.g., a floor-mounted, granite weigh
www.fda.gov/MedicalDevices/DeviceRegulationandG- bench). If a metallic support surface is used, the surface
uidance/GuidanceDocuments/ucm072783.htm. should be grounded in order to prevent the buildup of
3. Murray CJ. Sampling and data analysis for environ- static electricity.
mental microbiology. In: Hurst CJ, Knudsen GR, McIn- Location: If possible, the balance should be located in a
erney J, eds. Manual of Environmental Microbiology. room that is temperature and humidity controlled. The loca-
2nd ed. Washington, DC: ASM Press; 2002:166–177. tion should have a clean, consistent electrical power supply.
4. Pflug IJ. Microbiology and Engineering of Sterilization 1 National Physical Laboratory. Good practice guide No. 70, weighing in the
Processes. 7th ed. Minneapolis, MN: Environmental pharmaceutical industry. 2004. http://resource.npl.co.uk/docs/
Sterilization Laboratory; 1990. science_technology/mass_force_pressure/clubs_groups/instmc_weigh-
5. Block, S. Disinfection, Sterilization & Preservation. 5th ing_panel/pharmaweigh.pdf. Accessed 21 March 2012. [NOTE—Nomenclature
in this chapter tends to follow this document, except where it conflicts with
ed. Philadelphia, PA: Lippincott, Williams and Wilkins; USP terms.]
2001. 2 American Society for Testing and Materials. E898 Standard test method of

6. ANSI/AAMI/ISO. 11137. Sterilization of Health Care testing top-loading, direct-reading laboratory scales and balances. 2005.
http://www.astm.org/Standards/E898.htm. Accessed 21 March 2012.
Products—Radiation—Part 1: Requirements for Develop-

Official from December 1, 2013


Copyright (c) 2013 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 208.85.168.103 by mettler2 on Mon Jun 03 12:43:04 EDT 2013

6218 〈1251〉 Weighing on an Analytical Balance / General Information Second Supplement to USP 36–NF 31

The location should be free of drafts and should not be near diation, magnetic fields, or changes in temperature or hu-
ovens, furnaces, air conditioner ducts, or cooling fans from midity, the assessment should be conducted with those sys-
equipment or computers. The balance should be positioned tems operating in order to duplicate a worst-case scenario.
away from outside windows so that direct sunlight does not
strike the balance. The balance should not be installed near
sources of electromagnetic radiation such as radio-frequency Operational Qualification
generators, electric motors, or hand-held communication
devices (including cordless telephones, cellular telephones, An operational qualification should be performed either
and walkie-talkies). The balance should not be located near by the user or by a qualified third-party vendor after the
magnetic fields induced by laboratory instrumentation or equipment has been installed.
other equipment. As a minimum, the power should be turned on and the
The performance of the balance should be assessed fol- balance should be allowed to equilibrate according to the
lowing installation and before use in order to demonstrate manufacturer’s instructions (1–24 h, depending on the type
adequate performance. In some situations, it may not be of balance) before use. Depending on the balance, analysts
possible to position the balance in an optimum environ- should include the following procedures in the operational
ment. Examples of potential facility issues include the qualification:
following: 1. Mechanical mobility of all moveable parts
1. Air currents sometimes are present in the laboratory. 2. Control of stable indication
2. Temperatures in the laboratory vary excessively (check 3. Manually triggered or automatic adjustment by
the manufacturer’s literature about temperature means of built-in weights
sensitivity). 4. Operation of ancillary equipment
3. Humidity is either very low or very high. Either condi- 5. Tare function
tion may increase the rate at which the sample 6. Initial calibration
weight varies because of pickup or loss of water. Low Several types of electronic analytical balances use built-in
humidity increases the buildup of static electricity. weights for manually triggered or automatic adjustment.
4. Adjacent operations are causing vibration. This adjustment usually is applied to reduce the drift of the
5. Corrosive materials are used nearby or are routinely balance over time and to compensate for drifts caused by
weighed. variations in the ambient temperature.
6. The balance is located within a fume hood because it Calibration normally is performed as part of the opera-
is used to weigh corrosive or hazardous materials. tional qualification, but it also can be performed periodically
7. The balance is adjacent to equipment that produces a thereafter. Calibration should be performed at the location
magnetic field (e.g., a magnetic stirrer). where the balance is used in normal operation.
8. Direct sunlight strikes the balance.
In situations when the balance is located near equipment
or systems that induce vibration, drafts, electromagnetic ra-

Table 1. Suggested Performance Tests and Acceptance Criteria


Property Definition Examples Acceptance Criteria
Change in weighing value divided by
the change in load, usually measured NMT 0.05% deviation where 〈41〉 is
between zero and the capacity of the The test load at or sufficiently close to applicable. For other uses, respective
Sensitivity balance. the capacity of the balance. tolerance requirement divided by 2.
Ability of a balance to follow the linear
relationship between a load and the
indicated weighing value. Nonlinearity
usually is expressed as the largest NMT 0.05% deviation where 〈41〉 is
magnitude of any linearity deviation From 3 to 6 points over the range of applicable. For other uses, respective
Linearity within the test interval. the balance. tolerance requirement divided by 2.
Deviation in the measurement value
caused by eccentric loading—in other
words, the asymmetrical placement of Performed in the center of gravity and
the center of gravity of the load rela- the four quadrants (for rectangular
tive to the load receiver. Eccentricity platter shapes) or at analogous loca-
usually is expressed as the largest tions for other platter shapes. Test
magnitude of any of the deviations load usually should be 30% of the ca-
between an off-center reading and pacity of the balance or higher (refer NMT 0.05% deviation where 〈41〉 is
the center reading for a given test to the manufacturer’s manual for any applicable. For other uses, respective
Eccentricity load. possible upper limit). tolerance requirement divided by 2.
Ability of a weighing instrument to dis-
play identical measurement values for
repeated weighings of the same ob-
jects under the same conditions, e.g.,
the same measurement procedure,
same operator, same measuring sys-
tem, same operating conditions, and
same location over a short period of
time. Repeatability usually is expressed 10 replicate weighings (using a test Requirement from 〈41〉 where applica-
as the standard deviation of multiple weight that is a few percent of the ble. For other uses, user specified re-
Repeatability weighings. nominal capacity of the balance). quirements will apply.

Official from December 1, 2013


Copyright (c) 2013 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 208.85.168.103 by mettler2 on Mon Jun 03 12:43:04 EDT 2013

Second Supplement to USP 36–NF 31 General Information / 〈1251〉 Weighing on an Analytical Balance 6219

Depending on the acceptance criterion, it may be suffi-


Performance Qualification cient to consider only the nominal weight value of the test
weights. If the nominal value of the test weight is consid-
Table 1 provides a list of the most important balance ered, analysts should ensure that the maximum permissible
properties that should be assessed during performance qual- error does not exceed one-third of the acceptance criterion.
ification. Depending on the risk of the application and the Alternatively, if the certified value of the test weight is con-
required weighing process tolerance, some of these tests sidered, its calibration uncertainty should not exceed one-
may be omitted. Tests also can be omitted if there is evi- third of the acceptance criterion. If more than one weight is
dence that the property in question has only minimal effect used to perform the test, the calibration uncertainties of the
on the weighing performance. Any procedures used should weights must be summed and the sum should not exceed
be consistent with in-house standard operating procedures, one-third of the acceptance criterion. For tests such as ec-
applicable for the specific balance, and adequately justified. centricity or repeatability, the use of certified weights is op-
Performance qualification should be performed periodically tional, but analysts must ensure that the mass of the weight
as described in standard operating procedures, and the fre- does not change during the test.
quency of each of the individual tests can vary depending The tests described above also can be included in formal
on the criticality of the property. periodic calibration in order to fulfill applicable cGMP
The weights that are used to perform the tests should be requirements.
stored and handled in a manner that minimizes contamina-
tion. Before executing the tests, the analyst should place the Balance Checks
weights in the vicinity of the balance for an appropriate
time to reach sufficient thermal equilibrium. If possible, all A balance check using an external weight helps ensure
tests should be carried out with a single test weight in order that the balance meets weighing tolerance requirements.
to minimize handling errors, but multiple test weights are The balance check is performed at appropriate intervals
permitted. based on applicable standard operating procedures. The fre-
The tests should be recorded in such a manner that the quency of the balance check depends on the risk of the
data can be used to easily track balance performance and to application and the required weighing tolerance. Checks
assist in laboratory investigations as needed. Meaningful ac- with external weights can be replaced partially using auto-
ceptance criteria can be set depending on the required matic or manually triggered adjustment by means of built-in
weighing tolerance, i.e., the maximum allowed deviation weights. When analysts perform the balance check with an
permitted by specifications, regulations, etc., of a quantity external weight, the same acceptance criteria may apply as
to be weighed from its target value. Procedures should be described in the sensitivity test above.
in place to address test results that are outside acceptable
ranges and to provide assurance that balance cleanliness
and environment have not affected the result. Also, a proce- Minimum Weight
dure should be in place for removing a balance from opera-
tion when observed results fall outside acceptable ranges. The minimum net sample weight, mmin, of a balance can
Sensitivity, linearity, and eccentricity all account for sys- be expressed by the equation:
tematic deviations; i.e., they limit the accuracy of the bal-
ance (based on the definition of accuracy in Validation of mmin = k × s/required weighing tolerance
Compendial Procedures 〈1225〉 and ICH Q2). In the Interna-
tional Vocabulary of Metrology (VIM) and documents of the where k is the coverage factor (usually 2 or larger) and s is
International Organization for Standardization, this concept the standard deviation (in a mass unit, e.g., in mg) of not
is referred to as trueness. Because deviations are largely in- fewer than 10 replicate measurements of a test weight. The
dependent from each other, it is not likely that all deviations minimum weight describes the lower limit of the balance
occur simultaneously and have the same algebraic sign. below which the required weighing tolerance is not adhered
Therefore the arithmetic addition of all individual deviations to. The equation above takes into account that the perfor-
to assess the balance accuracy would constitute a rather mance of laboratory balances at the lower end of the meas-
conservative approach. A quadratic addition of the individ- urement range is limited by the finite repeatability.
ual deviations is a more realistic approach. By allocating For materials that must be accurately weighed, 〈41〉 stipu-
50% of the weighing tolerance budget to the acceptance lates that repeatability is satisfactory if two times the stan-
criteria of the individual properties, e.g., sensitivity, linearity, dard deviation of the weighed value, divided by the nomi-
and eccentricity, analysts ensure adherence to the required nal value of the weight used, does not exceed 0.10%. For
weighing tolerance. Therefore, the acceptance criteria for this criterion the equation above simplifies to:
the individual properties that account for the systematic de-
viations are set to weighing tolerance divided by 2. These mmin = 2000 × s
properties—or a subset of them—also can be taken to fulfill
the accuracy requirement described in 〈41〉. In this case the If not subject to the requirements of 〈41〉, the minimum
acceptance criteria thus allow a maximum deviation of weight value may vary depending on the required weighing
0.05% for sensitivity, linearity, and eccentricity. Repeatability tolerance and the specific use of the balance.
preferably is tested with a test weight of a few percent of To facilitate handling, the test weight that is used for the
the balance capacity. At the lower end of its measurement repeatability test does not need to be at the minimum
range, the performance of laboratory balances is limited by weight value but can be larger because the standard devia-
the finite repeatability, and limitations induced by system- tion of repeatability is only a weak function of the test
atic deviations normally can be neglected. Therefore, the weight value.
whole weighing tolerance budget can be allocated to the In order to satisfy the required weighing tolerance, when
acceptance criterion of the repeatability test. samples are weighed the amount of sample mass (i.e., the
For the sensitivity and linearity tests as described above, net weight) must be equal to or larger than the minimum
the analyst should use certified weights with an appropriate weight. The minimum weight applies to the sample weight,
weight class (e.g., according to International Organization not to the tare or gross weight.
of Legal Metrology R111 or American Society for Testing Factors that can influence repeatability while the balance
and Materials E617, available from www.oiml.org and www. is in use include:
astm.org, respectively). [NOTE—If a differential method is 1. The performance of the balance and thus the mini-
used for the linearity test, certified weights may not be mum weight can vary over time because of changing
required.] environmental conditions.

Official from December 1, 2013


Copyright (c) 2013 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 208.85.168.103 by mettler2 on Mon Jun 03 12:43:04 EDT 2013

6220 〈1251〉 Weighing on an Analytical Balance / General Information Second Supplement to USP 36–NF 31

2. Different operators may weigh differently on the bal- with a small opening, and the enclosure should be replaced
ance—i.e., the minimum weight determined by dif- rapidly following material transfer. Special precautions
ferent operators may be different. should be taken to be certain that the receiver and the en-
3. The standard deviation of a finite number of replicate closure are constructed from a material that is compatible
weighings is only an estimation of the true standard with the liquid sample. The receiver and enclosure must
deviation, which is unknown. have a seal that is sufficient to prevent leaks from a liquid
4. The determination of the minimum weight with a that is of low viscosity or has low surface tension or a low
test weight may not be completely representative for boiling point.
the weighing application.
5. The tare vessel also may influence minimum weight
because of the interaction of the environment with Types of Weighing
the surface of the tare vessel.
For these reasons, when possible, weighings should be Weighing for quantitative analysis: The initial step for
made at larger values than the minimum weight. many quantitative analyses is to accurately weigh a specified
amount of a sample. Section 6.50.20 in the General Notices
stipulates that solutions for quantitative measures must be
OPERATION OF THE ANALYTICAL BALANCE prepared using accurately weighed analytes: i.e., analysts
must use a balance that meets the criteria in 〈41〉. Errors
Select the appropriate balance for the quantity and per- introduced during the weighing of a sample can affect the
formance needed. General chapter 〈41〉 provides require- accuracy of all subsequent analytical measurements.
ments for balances used for materials that must be accu- Addition weighing: Addition weighings typically are
rately weighed.The balance user should check the balance used for solid samples or liquid samples for which volatility
environment (vibration, air currents, and cleanliness) and is not an issue. The receiver is placed on the balance. After
status of calibration before use. the balance display stabilizes, the analyst should tare the
balance; add the desired amount of material to the receiver;
Receivers allow the balance display to stabilize; record the weight;
and quantitatively transfer the material to an appropriate
To ensure suitable performance in measuring the weight vessel or, if it cannot be guaranteed that the entire amount
of a specimen, analysts should consider selection of a proper has been transferred, weigh the receiver again and note the
receiver for the material. weight difference.
General characteristics: All receivers must be clean, Dispense weighing: Dispense weighing typically is
dry, and inert. The total weight of the receiver plus the used for weighing emulsions or viscous liquids such as oint-
specimen must not exceed the maximum capacity of the ments. In these situations it is not practical to weigh the
balance. With a properly maintained and adjusted labora- material into a typical receiver. Accordingly, the analyst
tory balance, weighing uncertainty for small samples, i.e., should tare the balance; place the sample on the balance in
net weights with a mass not exceeding typically a few per- a suitable container (e.g., a bottle, tube, transfer pipet, or
cent of the capacity of the balance, essentially is determined syringe) that has been wiped clean on the outside; record
by the repeatability. However, repeatability depends on the the weight after the balance display stabilizes; transfer the
size and surface area of the weighed object. For this reason desired amount of sample to an appropriate receiving ves-
large or heavy receivers introduce a deviation from the con- sel, such as a volumetric flask; and place the pipet or sy-
ditions under which the repeatability was determined with- ringe back onto the balance. The difference in the two
out considering the receivers. Therefore, either receivers of a weighings is equal to the weight of the transferred speci-
low mass and small surface should be used (especially in men.
cases when specimens of low weight are being measured) Gravimetric dosing: Gravimetric dosing typically is
or the repeatability test should be performed with the re- used for sample and standard preparations or capsule filling.
ceiver placed on the weighing pan as a preload. Receivers For such weighing the analyst places the volumetric flask,
should be constructed from nonmagnetic materials in order vial, or capsule shell on the balance; tares the balance after
to prevent magnetic interference with electronic balance the balance display stabilizes; adds the solid or liquid com-
components. Receivers should be used at ambient tempera- ponents into the receiver by means of dosing units; and
ture in order to prevent the formation of air currents within records the respective weights.
the weighing chamber.
Solid samples: Receivers for weighing solid materials in- Problem Samples
clude weighing paper, weighing dishes, weighing funnels,
or enclosed vessels, including bottles, vials, and flasks. Hy- Electrically charged samples and receivers: Dry, finely
groscopic papers are not recommended for weighing be- divided powders may be charged with static electricity that
cause they may have a detrimental effect on the observed can make the powder either attracted to or repelled by the
results. receiver or the balance, causing inaccurate weight measure-
Weighing dishes typically are constructed from a polymer ments and specimen loss during transfer. A drift in the bal-
or from aluminum. Antistatic weighing dishes are available ance readings should alert the operator to the possibility
for measuring materials that retain static electricity. Weigh- that the material has a static charge. Commercially available
ing funnels typically are constructed from glass or from a balances with a built-in antistatic device can be used to
polymer. The design of this type of receiver combines attrib- remedy the problem. Such devices may use piezoelectric
utes of a weighing dish and a transfer funnel, which can components or a very small amount of a radioactive ele-
simplify the analytical transfer of a weighed powder to a ment (typically polonium) to generate a stream of ions that
narrow-necked vessel such as a volumetric flask. For solid dissipate the static charge when passed over the powder
samples that are volatile or deliquescent, analysts must being weighed. Antistatic weigh boats, antistatic guns, and
weigh the material into an enclosed vessel. Where practical, antistatic screens also are commercially available. The static
analysts should use an enclosed vessel with a small opening charge depends also on the relative humidity of the labora-
in order to reduce sample weight loss from volatilization or tory, which in turn depends on atmospheric conditions.
weight gain from the adsorption and absorption of atmos- Under certain conditions, static charge is caused by the type
pheric water. of clothing worn by the operator and this charge can cause
Liquid samples: Receivers for liquid samples typically large errors in the weighing. Borosilicate glassware and
are inert, enclosed vessels. For liquid samples that are vola- plastic receivers have a well-known propensity for picking
tile or deliquescent, analysts should use an enclosed vessel

Official from December 1, 2013


Copyright (c) 2013 The United States Pharmacopeial Convention. All rights reserved.
Accessed from 208.85.168.103 by mettler2 on Mon Jun 03 12:43:04 EDT 2013

Second Supplement to USP 36–NF 31 General Information / 〈1660〉 Durability of Glass Containers 6221

up static charge, especially at low relative humidity. The Add the following:
gloves used to protect the operator also may increase the
potential for a static charge problem. Placing the container
in a metal holder may help to shield the static charge, and
antistatic gloves also can help to alleviate the problem.
〈1660〉 EVALUATION OF THE

Volatile samples: When weighing a liquid that has a


low boiling point, analysts must receive the specimen in a
INNER SURFACE DURABILITY OF
vessel with a gas-tight enclosure of small diameter. The ana-
lyst then tares the vessel and enclosure, adds the desired
GLASS CONTAINERS
amount of sample, and replaces the enclosure. After the bal-
ance display stabilizes, the analyst records the specimen
weight.
Warm or cool samples: Samples that are warm or cool PURPOSE
should be equilibrated in the laboratory, or the weight read-
ings may be erroneous. With regard to warm samples, the This general information chapter provides information
apparent weight is smaller than the true weight because of about factors that affect the durability of the inner surface
heat convection. For example, a flask that is warmer than of glass containers. Recommended approaches are provided
ambient air warms up this air, which then flows upward to evaluate the potential of a drug product to cause forma-
along the flask and reduces the apparent weight of the con- tion of glass particles and delamination of the inner surface.
tents by viscous friction. Screening methods are provided to detect glass particles
Hygroscopic samples: Hygroscopic materials readily ab- and delamination, allowing a comparison to be made of
sorb moisture from the atmosphere and steadily gain weight glass durability on a lot-to-lot basis or between different
if left exposed. Therefore, hygroscopic samples must be ei- glass manufacturers.
ther weighed promptly or placed in a vessel with a gas-tight
enclosure. For a gas-tight vessel, analysts should tare the SCOPE
vessel and enclosure, add the desired amount of sample,
and replace the enclosure. After the balance display stabi- This chapter addresses bottles and vials manufactured by
lizes, the analyst can record the specimen weight. molding and ampuls, cartridges, vials, and prefillable sy-
Aseptic or biohazardous samples: The weighing of ringes manufactured from tubing glass. Glass for pharma-
sterile or biohazardous samples should take place within the ceutical packaging is classified as Type I borosilicate glass,
confines of a clean bench, biosafety cabinet, isolator, or sim- Type II treated soda-lime-silica glass, or Type III soda-lime-
ilar containment device. Air flow within the hood potentially silica glass on the basis of the hydrolytic resistance of the
can cause balance instability, so after a balance has been glass, as defined in Containers—Glass 〈660〉. Type I glass
installed under a hood, analysts should perform a rigorous containers are suitable for most products for parenteral and
qualification study with suitable weight artifacts (see 〈41〉) in nonparenteral use. Type II glass containers are suitable for
order to determine the acceptability of the balance perfor- most acidic and neutral aqueous products for parenteral and
mance in this environment. nonparenteral uses, and can be used for alkaline parenteral
Weighing corrosive materials: Many chemicals, such products when stability data demonstrate their suitability.
as salts, are corrosive, and materials of this nature should Type III glass containers usually are not used for parenteral
not be spilled on the balance pan or inside the balance products or for powders for parenteral use, except when
housing. Extra care is essential when materials of this nature suitable stability test data indicate that Type III glass is satis-
are weighed. Analysts should consider the use of sealed con- factory. This chapter focuses primarily on Type I glass, be-
tainers such as weighing bottles or syringes. In the event of cause it is the most widely used in the pharmaceutical and
a spill, requalification of balance may be necessary, depend- biopharmaceutical industry for parenteral products although
ing on the nature of the spill. the guidance can be equally applied to Type II and Type III
glass used for parenteral products.
The chapter should be useful for the following:
Safety Considerations When Weighing • Molded and tubular glass container manufacturers and
converters
During a weighing, the analyst may be exposed to high • Pharmaceutical and biopharmaceutical companies
concentrations of a pure substance. The analyst must care- • Contract manufacturing and filling organizations
fully consider this possibility at all times and should be fa- Glass delamination may be described as the appearance
miliar with the precautions described in the substance′s Ma- of thin flexible flakes of glass (or lamellae) that can range in
terial Safety Data Sheet before weighing it. Hazardous size from <50 µm to 200 µm in a drug product solution.
materials should be handled in an enclosure that has appro- This is a serious quality issue and can result in a product
priate air filtration. Many toxic—and possibly allergenic— recall. The appearance of glass lamellae is a lagging indica-
substances present as liquids or finely divided particles. tor of a strong interaction between the drug product and
When weighing these substances, analysts should use a the inner surface of the glass. Although delamination is the
mask that covers the nose and mouth to prevent any inhala- most obvious visual indicator, it represents the final stage of
tion of the substance, and they should use gloves to prevent a complex glass corrosion reaction, and can be observed
any contact with the skin. [NOTE—The use of gloves is good only at a point where prevention is no longer an option.
practice for handling any chemical. If it is necessary to han- Adding further complexity to detection, mechanical energy
dle the container being weighed, the analyst should wear from shaking or vial-to-vial contact during transportation
gloves not only for self-protection but also to prevent mois- may be required to dislodge the lamellae from the internal
ture and oils from being deposited on the weighed surface of a filled vial and facilitate observation.
container.]■2S (USP36) Tests for delamination combine the visual examination of
the solution, an examination of the vial’s internal surface
and analysis of an aggressive test solution to assess the pro-
pensity of the internal glass surface of vials to delaminate.
These examinations and the use of an aggressive test solu-
tion are intended to be conducted by the pharmaceutical
manufacturer, not the glass manufacturer or converter.

Official from December 1, 2013


Copyright (c) 2013 The United States Pharmacopeial Convention. All rights reserved.

Você também pode gostar