Escolar Documentos
Profissional Documentos
Cultura Documentos
Second Supplement to USP 36–NF 31 General Information / 〈1251〉 Weighing on an Analytical Balance 6217
microorganisms that pass the 0.45-µm filter are of interest ment, Validation, and Routine Control of a Sterilization
because they present the greatest potential challenge to the Process for Medical Devices. Arlington, VA: AAMI;
sterilizing filtration process. They should be evaluated 2006.
against the upstream bioburden. Bioburden of greatest con- 7. AAMI/ISO. 11737. Sterilization of Medical Devices—Mi-
cern includes Pseudomonas, Brevundimonas, Ralstonia, and crobiological Methods—Part 1: Estimation of Population
Mycoplasma. of Microorganisms on Products. Arlington, VA: AAMI;
2006.
Identification ■2S (USP36)
6. ANSI/AAMI/ISO. 11137. Sterilization of Health Care testing top-loading, direct-reading laboratory scales and balances. 2005.
http://www.astm.org/Standards/E898.htm. Accessed 21 March 2012.
Products—Radiation—Part 1: Requirements for Develop-
6218 〈1251〉 Weighing on an Analytical Balance / General Information Second Supplement to USP 36–NF 31
The location should be free of drafts and should not be near diation, magnetic fields, or changes in temperature or hu-
ovens, furnaces, air conditioner ducts, or cooling fans from midity, the assessment should be conducted with those sys-
equipment or computers. The balance should be positioned tems operating in order to duplicate a worst-case scenario.
away from outside windows so that direct sunlight does not
strike the balance. The balance should not be installed near
sources of electromagnetic radiation such as radio-frequency Operational Qualification
generators, electric motors, or hand-held communication
devices (including cordless telephones, cellular telephones, An operational qualification should be performed either
and walkie-talkies). The balance should not be located near by the user or by a qualified third-party vendor after the
magnetic fields induced by laboratory instrumentation or equipment has been installed.
other equipment. As a minimum, the power should be turned on and the
The performance of the balance should be assessed fol- balance should be allowed to equilibrate according to the
lowing installation and before use in order to demonstrate manufacturer’s instructions (1–24 h, depending on the type
adequate performance. In some situations, it may not be of balance) before use. Depending on the balance, analysts
possible to position the balance in an optimum environ- should include the following procedures in the operational
ment. Examples of potential facility issues include the qualification:
following: 1. Mechanical mobility of all moveable parts
1. Air currents sometimes are present in the laboratory. 2. Control of stable indication
2. Temperatures in the laboratory vary excessively (check 3. Manually triggered or automatic adjustment by
the manufacturer’s literature about temperature means of built-in weights
sensitivity). 4. Operation of ancillary equipment
3. Humidity is either very low or very high. Either condi- 5. Tare function
tion may increase the rate at which the sample 6. Initial calibration
weight varies because of pickup or loss of water. Low Several types of electronic analytical balances use built-in
humidity increases the buildup of static electricity. weights for manually triggered or automatic adjustment.
4. Adjacent operations are causing vibration. This adjustment usually is applied to reduce the drift of the
5. Corrosive materials are used nearby or are routinely balance over time and to compensate for drifts caused by
weighed. variations in the ambient temperature.
6. The balance is located within a fume hood because it Calibration normally is performed as part of the opera-
is used to weigh corrosive or hazardous materials. tional qualification, but it also can be performed periodically
7. The balance is adjacent to equipment that produces a thereafter. Calibration should be performed at the location
magnetic field (e.g., a magnetic stirrer). where the balance is used in normal operation.
8. Direct sunlight strikes the balance.
In situations when the balance is located near equipment
or systems that induce vibration, drafts, electromagnetic ra-
Second Supplement to USP 36–NF 31 General Information / 〈1251〉 Weighing on an Analytical Balance 6219
6220 〈1251〉 Weighing on an Analytical Balance / General Information Second Supplement to USP 36–NF 31
2. Different operators may weigh differently on the bal- with a small opening, and the enclosure should be replaced
ance—i.e., the minimum weight determined by dif- rapidly following material transfer. Special precautions
ferent operators may be different. should be taken to be certain that the receiver and the en-
3. The standard deviation of a finite number of replicate closure are constructed from a material that is compatible
weighings is only an estimation of the true standard with the liquid sample. The receiver and enclosure must
deviation, which is unknown. have a seal that is sufficient to prevent leaks from a liquid
4. The determination of the minimum weight with a that is of low viscosity or has low surface tension or a low
test weight may not be completely representative for boiling point.
the weighing application.
5. The tare vessel also may influence minimum weight
because of the interaction of the environment with Types of Weighing
the surface of the tare vessel.
For these reasons, when possible, weighings should be Weighing for quantitative analysis: The initial step for
made at larger values than the minimum weight. many quantitative analyses is to accurately weigh a specified
amount of a sample. Section 6.50.20 in the General Notices
stipulates that solutions for quantitative measures must be
OPERATION OF THE ANALYTICAL BALANCE prepared using accurately weighed analytes: i.e., analysts
must use a balance that meets the criteria in 〈41〉. Errors
Select the appropriate balance for the quantity and per- introduced during the weighing of a sample can affect the
formance needed. General chapter 〈41〉 provides require- accuracy of all subsequent analytical measurements.
ments for balances used for materials that must be accu- Addition weighing: Addition weighings typically are
rately weighed.The balance user should check the balance used for solid samples or liquid samples for which volatility
environment (vibration, air currents, and cleanliness) and is not an issue. The receiver is placed on the balance. After
status of calibration before use. the balance display stabilizes, the analyst should tare the
balance; add the desired amount of material to the receiver;
Receivers allow the balance display to stabilize; record the weight;
and quantitatively transfer the material to an appropriate
To ensure suitable performance in measuring the weight vessel or, if it cannot be guaranteed that the entire amount
of a specimen, analysts should consider selection of a proper has been transferred, weigh the receiver again and note the
receiver for the material. weight difference.
General characteristics: All receivers must be clean, Dispense weighing: Dispense weighing typically is
dry, and inert. The total weight of the receiver plus the used for weighing emulsions or viscous liquids such as oint-
specimen must not exceed the maximum capacity of the ments. In these situations it is not practical to weigh the
balance. With a properly maintained and adjusted labora- material into a typical receiver. Accordingly, the analyst
tory balance, weighing uncertainty for small samples, i.e., should tare the balance; place the sample on the balance in
net weights with a mass not exceeding typically a few per- a suitable container (e.g., a bottle, tube, transfer pipet, or
cent of the capacity of the balance, essentially is determined syringe) that has been wiped clean on the outside; record
by the repeatability. However, repeatability depends on the the weight after the balance display stabilizes; transfer the
size and surface area of the weighed object. For this reason desired amount of sample to an appropriate receiving ves-
large or heavy receivers introduce a deviation from the con- sel, such as a volumetric flask; and place the pipet or sy-
ditions under which the repeatability was determined with- ringe back onto the balance. The difference in the two
out considering the receivers. Therefore, either receivers of a weighings is equal to the weight of the transferred speci-
low mass and small surface should be used (especially in men.
cases when specimens of low weight are being measured) Gravimetric dosing: Gravimetric dosing typically is
or the repeatability test should be performed with the re- used for sample and standard preparations or capsule filling.
ceiver placed on the weighing pan as a preload. Receivers For such weighing the analyst places the volumetric flask,
should be constructed from nonmagnetic materials in order vial, or capsule shell on the balance; tares the balance after
to prevent magnetic interference with electronic balance the balance display stabilizes; adds the solid or liquid com-
components. Receivers should be used at ambient tempera- ponents into the receiver by means of dosing units; and
ture in order to prevent the formation of air currents within records the respective weights.
the weighing chamber.
Solid samples: Receivers for weighing solid materials in- Problem Samples
clude weighing paper, weighing dishes, weighing funnels,
or enclosed vessels, including bottles, vials, and flasks. Hy- Electrically charged samples and receivers: Dry, finely
groscopic papers are not recommended for weighing be- divided powders may be charged with static electricity that
cause they may have a detrimental effect on the observed can make the powder either attracted to or repelled by the
results. receiver or the balance, causing inaccurate weight measure-
Weighing dishes typically are constructed from a polymer ments and specimen loss during transfer. A drift in the bal-
or from aluminum. Antistatic weighing dishes are available ance readings should alert the operator to the possibility
for measuring materials that retain static electricity. Weigh- that the material has a static charge. Commercially available
ing funnels typically are constructed from glass or from a balances with a built-in antistatic device can be used to
polymer. The design of this type of receiver combines attrib- remedy the problem. Such devices may use piezoelectric
utes of a weighing dish and a transfer funnel, which can components or a very small amount of a radioactive ele-
simplify the analytical transfer of a weighed powder to a ment (typically polonium) to generate a stream of ions that
narrow-necked vessel such as a volumetric flask. For solid dissipate the static charge when passed over the powder
samples that are volatile or deliquescent, analysts must being weighed. Antistatic weigh boats, antistatic guns, and
weigh the material into an enclosed vessel. Where practical, antistatic screens also are commercially available. The static
analysts should use an enclosed vessel with a small opening charge depends also on the relative humidity of the labora-
in order to reduce sample weight loss from volatilization or tory, which in turn depends on atmospheric conditions.
weight gain from the adsorption and absorption of atmos- Under certain conditions, static charge is caused by the type
pheric water. of clothing worn by the operator and this charge can cause
Liquid samples: Receivers for liquid samples typically large errors in the weighing. Borosilicate glassware and
are inert, enclosed vessels. For liquid samples that are vola- plastic receivers have a well-known propensity for picking
tile or deliquescent, analysts should use an enclosed vessel
Second Supplement to USP 36–NF 31 General Information / 〈1660〉 Durability of Glass Containers 6221
up static charge, especially at low relative humidity. The Add the following:
gloves used to protect the operator also may increase the
potential for a static charge problem. Placing the container
in a metal holder may help to shield the static charge, and
antistatic gloves also can help to alleviate the problem.
〈1660〉 EVALUATION OF THE
■