Você está na página 1de 1

A-30-673:TEMPLATE 12/13/10 11:09 AM Page 1

Bulletin A-30-673

Series 673 Pressure Transmitter


Specifications - Installation and Operating Instructions

1/4˝-18 (22.3 MM) Ø1-19/32


CONDUIT KNOCKOUT [40.48]
Ø1-13/32
[35.72]

3-5/8
[92.08]

2-23/64
[59.93]
2 1-11/16
[50.8] [42.86]
3/4 HEX 3/4 HEX
[19.05 HEX] [19.05 HEX]
Conduit Version 1/4 NPT 1/4 NPT
ø1-5/8 Ø1-5/8
[41.28] [41.28]
CABLE ANCHOR CONDUIT VERSION

The low cost Series 673 Pressure Transmitter is a fixed range SPECIFICATIONS
transmitter designed for harsh environments and suitable for high shock Service: Compatible, non-combustible liquid, gas, or vapor.
and vibration applications. Constructed of stainless steel, the Series 673 Wetted Materials: 17-4 PH stainless steel.
provides a 4-20 mA output signal with 0.25% accuracy. Use the Series Accuracy: RSS* (at constant temperature) ±0.25%.
673 in industrial OEM equipment, hydraulic systems, HVAC equipment, Non-Linearity: (BFSL) ±0.22% FS;
industrial engines and compressor controls. Hysteresis: 0.10% FS;
Non-Repeatability: 0.05% FS.
INSTALLATION Temperature Limits: -40 to 185°F (-40 to 85°C).
Electrical Connections Pressure Limits: 2 X full scale range.
The Series 673 Pressure Transmitter is a true 2-wire, 4-20 mA current Compensated Temperature Range: -40 to 176°F (-40 to 80°C).
output device and delivers rated current into any external load of 0-800 Thermal Effects: Zero Shift: ±2.0% FS/100°F (±1.8% FS/50°C); Span
ohms. The transmitter has a two wire cable where red is positive and Shift: ±1.5% FS/100°F (±1.3% FS/50°C); Warm-up Shift: ±0.1% FS
black is negative. Connect shield to system or earth ground. total.
The 4-20 mA current output units are designed to have current flow in Stability: 0.5% FS/year.
one direction only - please observe polarity. We suggest that the Response Time: 5 milliseconds.
electrical cable shield be connected to the system’s loop circuit ground Supply Voltage: 9-30 VDC.
to improve electrical noise rejection. Output: 4-20 mA**, 2-wire.
Zero and Span Adjustment: Fixed.
Media Compatibility Loop Resistance: 0 to 800Ω.
These transmitters are designed to be used with any gas or liquid Shock: 200 g.
compatible with 17-4 PH stainless steel. Vibration: 20 g.
Electrical Connections: 2 ft (61 cm) multi-conductor cable.
Pressure Fitting Conduit Models: 1/4˝-18 (22.3 mm) knockout.
Typically, standard pipe fittings and procedures should be used. Enclosure: Stainless steel and Valox.
However, for pressure ranges in excess of 500 psig, we suggest the use Weight: 2.3 oz (65 g).
of a sealant. Excessive torquing of metal fittings may cause a slight zero Agency Approvals: CE.
shift. The use of plastic fittings typically results in no noticeable zero
shift. Torquing does not appreciably affect linearity or sensitivity.

MAINTENANCE *RSS of non-linearity, non-repeatability and hysteresis


Upon final installation of the Series 673 Pressure Transmitter, no routine **Zero output factory set to within ±0.16 mA.
maintenance is required. A periodic check of system calibration is **Span (full scale) output factory set to within ±0.16 mA.
recommended. The Series 673 is not field serviceable and should be
returned if repair is needed (field repair should not be attempted and
may void warranty). Be sure to include a brief description of the problem
plus any relevant application notes. Contact customer service to receive
a return goods authorization number before shipping.

©Copyright 2010 Dwyer Instruments, Inc. Printed in U.S.A. 12/10 FR# RA-443786-00 Rev. 1

DWYER INSTRUMENTS, INC. Phone: 219/879-8000 www.dwyer-inst.com


P.O. BOX 373 • MICHIGAN CITY, INDIANA 46361, U.S.A. Fax: 219/872-9057 e-mail: info@dwyer-inst.com