Você está na página 1de 8

Infection Control

Bacterial contamination ofthe water supply in newly installed dental units


Henry N, Williams*/Alicia Johnson**/Jaqueline 1. Kel]ey***/Marcie L. Baer****/
Tracy S. King*****/Barbara Mitchell******/John F. Hasler*******

Bacterial contamination ofthe water stipply of newly installed dental units was investigated.
Water .samples were collected from water suppiy Unes to the dental operatories prior to
connection ofthe dental units. Within hours following connection, and continuing for up
to the 6 months ofthe study water samples were obtained from the air-water syringe ofthe
units. The samples were .•serially diiuted lO-foid and piated on cuiture media for qttantitative
analysis. The formation of bacterial biofiim in tiie dentai water stipply tubing was tnonitored
by scanning electron microscopy The results of these studies revealed that tiie buiiding's
water supply to the dental units was contaminated prior to connection to the units. The
water stipply from the air-water syringe was therefore contatninated as well. The number of
contaminant bacteria in the dentai unit water suppiy increased for several weeks and then
stabilized. The lumen of the dental tubing became progressively contaminated with bacterial
biofiim, which sttbsequently became the primar}- reservoirfor maintenance ofthe contamination
ofthe dentai unit water supply. (Quintessence Int 1995:26:331-337.)

Introduction addressed by several investigators.'"' However, the


original source and time of initial contatnination in
It is weil established that the water supply of dental new units have not been reported.
units is grossly contaminated with bacteria. Some Some ofthe earlier studies on the dental unit water
potential sources of such contamination have been supply (DUWS) have reported the recovery of oral
bacterial species. However, more recent reports in
* Associate Professor, Department ofMicrobioiogy. Utiiversity of
which the bacteria were identified revealed that the
MarN'land at Baltimore. Baltimore College of Dental Surgery predominant organisms recovered are nearly all aer-
Dental School. Baltimore. Maryland. obic or facultative Gram-negative bacilli of the type
** Graduate Student, Department of Microbiology. University of conimonly found in water distribution systems.'' The
Maryland at Baltitnore, Baltimore College of Dental Surgery
Dental School. Baltimore. Mar>'land. predominance of Gram-negative organisms suggests
' " Research Assistant. Department of Microbiology. University of that the bacteria contaminating the DUWS originate in
Marj'land at Baltimore, Baltimore College of Dental Surgery the community water source that supplies the dental
Dental School. Baltinrore, Maryland
units. This is also suggested by resuits of a study in
***' Graduate Research Assistant, Department of Microbiology.
University of Maryland at Baltimore. Baltimore College of which typical levels of bacterial contamination were
Dental Surgery Dental Scliool. Baltimore. Maryland. found in dental units that were used by students to
*•-*• Private Practice, Baltimore. Maryland. learn basic dental practices but had not been used with
***-'"' Private Practice. Fort Bennitig, Georgia. patients.'
- " * " * Associate Dean, Clinicai and Hospital Affairs, University of The DUWS has been reported to harbor as many as
Maryland at Baltimore, Baltimore College of Dental Surger>'
Dental School. Baltimore. Marylatid. 1 million bacteria per mL of sample;'' some of these
Reprint requests: Dr Henry N. Williams. Associate Professor, Depart- bacteria have been implicated in opportunistic infec-
ment of Microbiology. University of Maryland at Baltimore, Baltimore tions.' This is at least 1,000-fold greater than the
College of Dental SLirgery Dental School. 666 West Baltimore Street,
Baltimore Maryland 21201. number recovered from potable water from fountains

Q u i ntesaonoo .Intornatic Number 5/1995 331


Infection Control

and sink faucets.'"' There is increasing agreement neetion samples. Following connection of the utitts.
among denial authorities that this represents an water samples ( postconnection samples ) were collect-
unacceptable bacterial load to introduce into the ed from the dental A/W syringe for up to a 6-month
mouths of patients. This is supported by the fact that period. Collected samples were taken immediately to
bacteria in the DUWS have been reported to have been the microbiology laboratory, where they were serially
a source of infection in immunocompromised pa- diluted 10-fold and spread plated onto dilute peptone
tients.' Clearly, the aggressive pursuit ofthe study of aii medium.*'^ The plates were incubated for 4 weeks at
aspects of the contamination of the water supply of 25°C. The number of colony-forming units (CFU)
dental units is warranted. This approach represents the were counted and the mean number for duplicate
best prognosis for finding a solution to this problem. plates was calculated. Although the general methods
During the complete renovation of all the dental used in collection and culturing of samples were as
clinics at the Baltimore College of Dental Surgery' and described above, the specific protocols and method-
Dental School, the authors had the opportunity to ology for each clinic are presented separately.
investigate the quality ofthe DUWS in newly installed
dental units. The results ofa preliminary investigation Chnic !
in which water samples were collected from the The disinfectant treatment of the water lines was
air-water (A/W) syringe and handpiece 1 month after completed on February 21, 1991. The first sample was
the installation ofthe first units revealed high levels of collected on February 22 (day I). Samples were
bacterial contamination. On the basis of these results, a collected daily through February 26 (day 5). The
more comprehensive study was designed to investigate remaining preconnection samples were usually collect-
the source and initation of contamination of the ed at 2- to 4-day intervals during the next 24 days
DUWS of newiy installed dental units in two clinics. before the new dental units were installed. The initial
The study monitored bacterial contamination of water postconnection sample was collected from the A/W
fi'cm the inhouse water lines prior to connection to the syringe within hours of installation ofthe new dentai
new dentai units and from the A/W syringe for 6 units on March 19 (day 26). Sampling ofthe water
delivered from the A/W syringe continued over the
months foilowing installation of the new equipment.
next 35 days; the last sample was collected on April 23.
Knowledge of the source and time of initiation of
Two sampling protocols were used in this clinic. One
contamination can enable clinicians to better under- was designed to sample a single dental unit (unit 2)
stand the nature ofthe contamination in their dental continuously through the entire study period from the
units and guide researchers and manufacturers in first preconnection sample to the last postconnection
developing strategies to overcome the problem. sample. The identical protocol was used for a second
unit (unit 11 ). with the exception that sampling ofthe
preconnection water line was initiated only 9 days
Method and materials before connection. Postconnection sampling of the
unit continued for the duration of the study as
Prior to the connection of the new dental units to
described for unit 2. Sampling the same unit offered
newly installed building water lines in two clinics, the
the advantage of monitoring the water quality from a
lines were treated overnight for disinfection with a single source throughout the study period. However,
bleach solution. This treatment is a standard practice the disadvantage of this approach was that each time
to reduce the number of microorganisms to a level the unit was activated and water was removed for
considered safe in accordance with public health sampling, the environment ofthe water in the lines was
standards. Following treatment, the bleach was Hushed disturbed. To compensate for this, a second protocol
from the lines. Water samples were collected from the was used: a different unit was sampled on each
water supply lines within 24 hours after flushing and at occasion for the first time without any prior disturb-
predetermined intervals up to the time of connection ance (with a single exception, unit 8). The sampling
to the new dental units. To collect these samples, the schedule was similar, but not identical, to that de-
water line shut-off valve, which extended into the scribed for unit 2.
cut-out opening in the floor or wall, was turned on and
the water coming from the pipe was collected into
sterile 15 -mL centrifuge tubes. Samples collected prior Clinic 2
to installation of dental tinits were designated precon- Following overnight disinfection and flushing ofthe

332 Quintessence Internationai Volume 26. Niimhcr 5/1995


Infection Control
water system, on July 9, 1992. 12 operatories were Table I Numbers of bacterial CFU (x \0\ except
randomly selected for study. On this date, samples where noted by *) per mL from clinic I
were taken from the in-house water lines ( preconnec-
tion samples) of six operatories. A sample was also
collected from the A/W syringe (postconnection Sample
samples) of one unit that had just been installed. day' Unit 2 Unit 11 Other units No.
Connection of remaining units was on July 10 (except 1* 15.0' 0.0'(1)
unit 8, which was connected on July 29). Sampling
from the A/W syringe of these units commenced on 2 0.0" 25.0'(3)
this date and continued daily through July 17. Subse- 3 15.0' 0.1 (4)
quently, samples were collected every 2 to 4 days 4 30.0' 0.0'(5)
through the end of July and on August 3, 10, 31, and 5 0.0' 260.0" (6)
September 4. A final sample was taken on Februaiy 1, 11 > 2.0
1993. The units were put in use for patient treatment 14 0.2 230.0'(9)
on July 29, 1992, Typically, all samples (5 mL) were 16 12.0 13.0 (10)
taken from the A/W syringe at the beginning ofthe day 17 0.2 15.0
and prior to use ofthe unit. Three sink faucets in the 26* 260,0 150.0 250.0 (19)
clinic were also routinely sampled in the same manner 640.0 (20)
to monitor the contamination level of water flowing 360.0 (8)
through the building's plumbing system. All samples 460,0 (17)
were processed and plated as described earlier. 28 11.0 260.0
30 980.0 510,0 420.0 (8)
In one dental unit from clinic 2, the lumen of tubing 33 520.0 310.0
supplying water to the A/W syringe was examined for 35 28.0 25.0
the colonization of bacteria hy scanning electron 40 800.0
microscopy (SEM). At 1 week, 2 weeks, 1 month, and 44 190.0 200-0 120.0 (8)
6 months postconnection ofthe unit, a section ( 10 cm) 47 220.0 160.0
ofthe water tubing was excised and cut into tour equal 50 560.0 140.0
pieces. Three of the pieces were placed in 2% 54 200.0 22.0
glutaraldehyde for examination by SEM. The fourth 61 260.0 680.0
piece was sliced in half lengthwise and scraped with a
sterile scalpel. The biofilm recovered on the blade was * Actual value.
suspended in water. The suspension was diluted and t After disinfection,
plated as described previously to determine the viabil- Í Preconnection samples obtained directly from the in-house
water lines prior to connection to dental units,
ity ofthe biofilm bacteria within the tube. The A/W
tj Posiconneclion samples obtained Trom the air-water syringe
syringe was replaced on the remaining tubing ofthe following connection of dental units to the in-house water
unit and normal usage was continued. system.
Tubing samples previously fixed in 2% glutaraldehyde
for electron microscopic examination were removed
from the fixative, cut in half lengthwise, and washed in
phosphate buffer solution for 10 to 15 minutes to
remove the fixative. The specimens were dehydrated
(for 10 minutes each) through a series of alcohol Results
washes (30%. 50%, 70%, 90%, and 100%) with a final Following treatment ofthe in-house water lines ofthe
treatment in hexamethyldisilazane (for 10 minutes) dental clinics with a disinfectant, the numbers of
and air dried overnight. The specimens were mounted bacterial contaminants were immediately reduced to
on SFM aluminum mounts and coated with gold- undetectable or low levels. However, this effect was
palladium in a Hummer VI Sputtering System (Tech- short lived, as the level of contamination increased
tiics) for 4 minutes. Examination by SEM (JSM-T200 rapidly within days (Tables 1 and 2). At the time of
Scanning Microscope. JEOL) was done at varying connection ofthe new dental units to the water lines,
magnifications and appropriate photographs were the water supply had already become contaminated
taken. with large numbers of bacterial colony-forming units.

Quintessence International Volume is. Number 5/1995 333

Í
Infection Control

Table 2 Numbers of bacterial CFU (X IÛ\ except where noted by *) per mL from in-house water lines and newly
installed dental units from clinic 2

Sample Unit Number


day
4 6 8 12 14 17 22 25 28 30 34 39
1+ -í
-
— _
-
0,7 > 1,0* > o,r 0,2' 0,2' - > 3.0 - > 3,0
2^ 2,6 > 3,0 0.2 > 3.0 > 3.0 2,4 > 3,0 > 3.0 0,4 > 3.0
3 ;> 3,0 0.7 - > 3,0 0,6 1.8 > 3,0 > 3,0 > 3.0 > 3.0 > 3,0 > 3.0
4 ;> 3.0 0.2 - > 3.0 1.3 > 3.0 > 3.0 > 3.0 > 3,0 > 3,0 > 3,0 > 3,0
5 ;> 3.0 0,7 - > 3,0 > 3,0 > 3,0 > 3,0 > 3.0 > 3,0 > 3,0 > 3,0 > 3.0
6 ;> 3,0 > 3,0 - > 3.0 > 3,0 > 3,0 > 3,0 > 3.0 > 3,0 > 3,0 > 3.0 > 3,0
7 ;> 3,0 > 3,0 - > 3,0 > 3,0 > 3,0 > 3,0 > 3.0 > 3.0 > 3.0 > 3.0 > 3,0
8 ;> 3,0 > 3,0 - > 3.0 > 3,0 > 3,0 > 3,0 > 3,0 > 3,0 > 3,0 > 3,0 > 3,0
9 ;> 3.0 > 3,0 - > 3,0 > 3,0 > 3,0 > 3,0 > 3.0 > 3.0 > 3.0 > 3,0 > 3,0
12 82,0 47,0 - 47,0 95,0 150,0 94,0 100,0 200,0 270,0 260,0 240.0
14 42,0 13,0 - 25,0 48,0 45.0 22,0 66,0 22,0 92,0 87,0 90,0
16 33,0 110,0 - 20,0 120,0 58,0 47.0 36,0 55,0 130,0 110,0 76.0
19 26,0 170,0 - 54,0 130.0 68.0 26.0 18,0 62,0 75,0 38,0 50,0
23 96,0 74,0 1,3 38.0 35,0 45,0 14,0 48,0 46,0 74.0 72.0 80.0
26 46,0 220.0 7,6 44,0 120.0 110.0 48,0 46,0 42.0 150,0 180,0 60,0
33 28,0 92,0 19.0 37.0 37,0 78,0 43,0 42,0 9.5 74,0 54.0 44,0
40 58,0 22.0 91,0 82.0 62.0 150.0 54,0 48,0 42.0 62,0 41,0 38,0
54 36,0 120,0 52,0 100,0 46,0 37,0 52,0 31.0 66,0 66,0 34,0 56,0
58 24,0 65,0 54,0 15,0 36.0 42,0 59,0 41,0 66.0 34.0 23,0 57,0
206 8,2 12,0 30,0 300,0 30,0 10,0 30,0 110,0 13,0 30,0 11.0 12.0

t Preconnection samples,
% Sample not taken,
S AU samples from day 2 through day 206 were pnstconnection samples.

The first water samples from the new dental units, days postconnection, although unusually high and low
collected within hours of installation, were observed in counts were sporadically recovered.
all cases to be grossly contaminated. In clinic 2, the number of colony-forming units
In clitiic 1. following the disinfection treatment, the recovered from water from the plumbing system had
level of contamination of water samples obtained exceeded 1,000 CFU/mL for most units by 1 day
directly from the water lines remained at undetectable following disinfection. The counts had exceeded
to low levels for the first 2 weeks. However, the 300,000 CFU/mL in the majority of units by 2 days
number had increased to 2.000 CFU/mL in unit 2 and postconnection. Such high counts so soon following
to 150,000 CFU/mL in unit 11 by day 17. when the disinfection exceeded expectations (based on the
previous results of ciinic 1 ) and the dilutions made of
last sample was taken before connection was made to
the samples. This caused the number of colony-
the new dental units (Table 1). The pattern and level of
forming units on the plates ofthe greatest dilution to be
contamination of units 2 and 11, which were disturbed
too many to count, or greater than 300, The results
by continuous samphng, were remarkably similar to
from these samples are expressed in Table 2 as > 3,0
those ofthe other units, which were typically sampled
(x 10') CFU/mL, Peak counts were recovered for the
only once and therefore remained undisturbed (until
majority of units between days 12 and 26. At the end of
the units were placed in routine use}. In units 2 and 11.
54 days in clinic 2, the level of contamination in most
the counts appeared to stabilize somewhat within 4 units had stabilized, typically to within a range of 3,1 to

334 Quintessence Int


Infection Control

6.6 X 10'^ CFL)/mL, At the 6-month sampling interval, Figs la to 1c Sequentially developing biofilm within the
tlie range for most units was 1,0 to 3,0 « 10' CFU/mL- déniai tubing (Original magnification X 1,600 )
By contrast, the mean number of bacterial colony-
forming units recovered from three sink faucets prior
to routine use ofthe clinics was 2,6 >* 10^ CFU/mL,
Following the opening of the chnics and start of
routine use, the counts from the faucets decreased to a
mean of 5.0 >i lO^* CFU/mL,
Examination by SEM of the lumen walls of the
tubing from a dental unit in clinic 2 revealed the
presence of attached individual bacterial cells and
microcolonies in samples obtained in the first weeks
following installation of the units and of developed
biofilm in samples collected 6 months postconnection
(Figs la to lc). The mean numbers of bacterial colonv-
forming units recovered from biofilm scrapings re- Fig la Biofilm at 2 weeks.
moved from the lumen ofthe tubing from one unit at 1
week, 2 weeks, 1 month, and 6 months were 4,1 x 10%
1,1 X 10\ 3.6 X 1O^ and 3.6 x lO" CFU/mL, respec-
tively.

Discussion
Bacterial contatttination of water delivered by the
dental A/W syringe and handpiece is a persistent and
universal problem for dental units connected to
community water supply systems. In this study, the
contamination of newly installed dental units was
investigated. The results revealed that the in-house
water lines supplying water to the dental uitits became
heavily contaminated, with microorganisms even after
bleach treatment, before connection to the new units.
The water going into the units following connection
immediately introduced hundreds of thousands of Fig 1b Biotilm at 1 month.
bacteria throughout the dental unit water delivery
system. Some of these bacteria undoubtedly attached
to the luminal walls ofthe plastic tubing within the
dental units, ultimately establishing a biofilm.
The development ofa biofilm on the lumen walls
over a period of 6 months was observed by SEM.
Durhig that time, the biofilm developed from a few
individual cells and microcolonies scattered along the
lumen walls after I or 2 weeks to a mature biofilm with
a dense population of coalescing microcolonies. Sev-
eral reports have implicated bacterial biofilms as the
primary source of microbiai contaminants in the
DLTWS,''^''"" This has been confirmed by recent
studies in our laboratory (unpublished data). Once
formed, these biofilms shed bacteria into the water
occupying the tubing. Unless treated with special
agents for its removal, the biofilm will persist as long as
it is exposed to an aqueous environment. For this Fig lc Biof im at 6 monttis.

SHyilUti llllöfliütlonal Number 5/1995 335


Infection Control

reason, it has been neariy impossible to eliminate or with the use of a water supply that contains thousands
reduce bacterial contamination of the DUWS in and millions of contaminant bacteria. To ensure the
conventional dental units. The larger numbers of delivery of water that is essentially ^''^'^ '^^ bacteria,
bacteria recovered fromthedental units in the days and some manufacturers of dental equipment have devel-
weeks following connection to the water lines in this oped dental units thai are designed to have a sterile
study were a direct result ofthe biofiim formation and water supply. In combination with a regular chemical
tiie shedding of biotllm oi^anisms into the water flush, this provides the units with the capacity to
supply. The difference between the relatively low deliver water that is essentially free of detectible
number of colony-forming units recovered from the bacteria. Several filtration apparatuses have also been
sink faucets and the large numbers from the A/W described. Dentists are encouraged to consider use of
syringe of dental units is tlirther evidence that the such systems. In the interim, further study on the
source of contamination of the DUWS is within the effects of flushing are warranted to resolve the flush
dentai units. The dental units contained built-in time controversy and establish uniform recommenda-
antiretraction valves to prevent the backflow of con-
tions.
taminated oral lluid from the A/W syringe or hand-
piece during use.
Although not a long-term remedy to the develop-
ment of bioHim In dental unit tubing, the numbers of Summary
bacteria transported by the building's water supply into
newly installed units can be nearly eiiminated by The results of this first such study on new dental units
proper disinfection and flushing ofthe water supply revealed that the source of inital bacterial contamina-
lines just before the connection to dental units is made. tion of the water delivered by newly installed new
Tliis was observed in chnic 1. in which the bacterial dental units was the public water supply. Further, the
counts remained low for at least 2 weeks. The reason contamination occurred immediately when the incom-
for the results in clinic 2, in which liigh levels of ing water supply, which carries large numbers of
contamination were observed within a day following colony-forming units, was introduced into the new
disinfection, are not known. It may indicate that dental units. Individual bacterial cells adhered to the
improper disinfection procedures were used. How- lumen walls ofthe tubing, where subsequent growth
ever, preconnection treatment will not eliminate the resulted in formation of a biofiim. This biofiim became
eventual contamination ofthe tubing in the dental unit the primary source and sustaining agent for the
with bacterial biofiim and the perpetual problem of bacterial contamination ofthe DUWS. Within days or
contamination of the DUWS. Until there is some weeks following connection to the building's water
means of preventing the development of biofiim in the lines, the level of contamination in the new units was
tubing of dental units, there is little hope of effectively typical ofthat observed in used dental units.
reducing or eh mina ting the associated bacteria! con- Recognizing this, dental clinicians are encouraged
tamination of dental units that receive nonsterile water. to fiush the water lines of new dental units for at least 2
The higher numbers recovered in this study from the minutes, just as recommended for used units, at the
sink faucets, particuiariy at the beginning ofthe study start of each business day and between patients.
prior to use (more than 200,000 CFU/mL), was sur- However, fiushing, which is currently the best availa-
prising, because previously water from these sources ble and most practicai procedure to reduce micro-
typically contained fewer than 100 CFU/mL and organisms, should not be considered an acceptable
rarely in excess of 500 CFU/mL of heterotrophic solution to the problem of DUWS contamination.
bacteria. A representative sample ofthe water from the Even after flushing, high numbers (more than 1,000
dental units was tested and confirmed by a commercial CFU/mL) of bacteria often remain. Also, the optimal
laboratory to be free (fewer than 1 per 100 mL) of duration of fiushing is somewhat conti"oversial, based
detectable coliform bacteria and, by this standard, to on widely disparate results that have ranged from 2
be safe for human consumption. However, evidence minutes' to 20 minutes.' To provide water that
suggests that other bacteria in the DUWS, such as the contains few, if any, potential pathogens, dentists are
pseudomonades, may be a source of infection, espe- encouraged to use equipment that provides water of
cially in immunocompromised individuals. similar quality to that which comes from water taps,
Dental professionals should not remain comfortable typically fewer than 500 CFU/mL.

336 Quintessence International Voliirpfj •iiimliBX'"''1O3::


Infection Control

Acknowledgments Prosthetic
The aiilhors express appréciât ion lo Ms Judy Pennington for assistance in
the preparation ofttiis manuscript. Ms Eleanor B, V^de for tiie eiectron
tnicroscopic work, and Mr Vernon Leciincr for leehnical assislance.
Rehabilitation
KctUi F.
References
1, WhildiouseRLS, Peters e.LiicmeJ.LMgeC [nliiience ofbiofilms

T
on microbiEii contamination in dental iinit water JDent l99i;l9;290- liis comprelieii.sive book presents
295.
s l r | ) - | j \ - ^ l i ' | i l c i ' l n i i i | i K ' s ill liicial a m i
2. Lewis DL, Boe [ÍÜ, Cross inrectLOn risks associated with current
procedures for using higli-speod dental handpieces, J Ciin Microbiül l i o i l i l y [)r<)stlHílic. ifiííihilitiilioii. In uildi-
liort to teohniqtic, thcuiy and mateiials are
S. BaggaBSR.MurpiiyRA. Anderson AW, Punwani i. ContLimimiiion
of dentai unit couhng water with oral micro organ isms and its (liscussed, and twcniy-cighf ciisi' liislories
prevention. J Am Dent Assoc I984i IO9:7i2-7l6. are featured. Tliis highly piariiral li'xt
4. Wlhams J, Johnston B. Johtison B, Huntington M. Mackenzie CD
Microbial contaminalioti of dentat unit wateriinesr Prevalence,
will hr of interest to not ordy prosllu-tic
intensity and micro bio iogical ciiaracteristics. J Am Dent Assoc icflniologists. bttt to plastic surgeons and
1993;i:4:59-65, {jîlicrs who seek better I
5. Sogaard P, Gatirn Hansen B. Frederiksen S. An investigation ofa
collection of yctiow-pigmcnted Pseiidomonas. Acta Pathol Micro- eoMinittnieation in prosthetics.
bioi Imiminoi Scand |B1 Í986;94:145-t53.
6. Gross A. Devine MJ, Cutright DE. Microbial contamination of
dental units and ultrasonic sealers. J Periodontul i976í47;67O-673.
7. Martin MV. The significance ofthe bacterial contamination of dental us s 1.30.
unit water systems Br Dent J 19a7;16î:I52-l54.
8. MakiJS. L.aCrtjis SJ, Hopkins BS, StaleyJT. Recovery and diversity
{
of heterotrophic bacteria from chlorinated driniiing waters. AppI Prosthetic Rehabilitation Order Mil. \V.V¿\Ü
Environ Microbiol I9S6;51 1047-1055.
9. Wiiliams HN, Quiiigy H, Romberg E. Evaluation and use ofa iow
nutrient medium and reduce, incubation temperature to study
bacterial contamination in the vater supply of dentai units. Can J
Microbiol Í994;4O;I2?-1-1I.
10. MayoJA. OertlingKM. Andrii ] SC. Bacterial bioftims: A source
contamination in dental aii ivater syringes. Clin Prev Dc
l990;i2:t3-20,
H. Keistrup J. Funder-Nieison TD, Theiiade J. Microbiai aggregate
contamination of wafer iines in dental equipment and its control.
Acta Pathoi Microbiol Scand | B | l977;85;l77-iSJ. =
C|b

Contents include:
Psytholügi<;al Considerations; Pationi Assessment^
Inipi-essiori aiit! Facial Materials; Impression Teeh-
niques; Ear PTO.'ÍI iiesps; Nasal Prostlicscs; Orliital
Pfostheses; Partial Prostheses; Color Mati;liing;
Alternative Method;, of I'^biation and Ri'tciition;
Direct Adhesive Fixation Lsinj; Aillit-sivfs aitd
Removers; Tissue Conditioinn'i; Killing the
Piii>dieiis; Os.scoitt1.egrated Implonls^, liiiniakes;
Pinlili'iiid.; Glossary;, Skin Grafts, Fliips, and
Tissue [v\pansion: and Anatomy.

Order Now
Tollírce i-800-621-0387
Fax 708-682-3288 (|Ulnlaf/ancs
boohf

Quintessence Publishing Co, Inc


QuintgAënce Intemaiionäl Volume '¿ä. Number 5/1995

Você também pode gostar