Optimax User Manual

Hand Piece
Lubrication System

764017A
2

Instructions for use


Please read these instructions before using the Contents.
autoclave.

Section Page
Keep these “Instructions for use” in a safe place
1 Overview of Control 3
close by the unit for future reference.
2 Installation Instruction 4

Service Helpline: Telephone: 01254 844 116

e.mail: customerservice@prestigemedical.co.uk
Prestige Medical Service staff are available to provide
advice and assistance during normal office hours. To avoid
delays when making contact please have the unit’s Model
and Serial Numbers at hand. These can be found on the
rating plate located at the rear of the unit or on the right
hand side of the casing when the door is open.
For additional information visit
www.prestigemedical.co.uk

UK Customers Overseas
Prestige Medical Limited Customers
East House, Duttons Way, Contact your local
Shadsworth Business Park, distributor. In case of
Blackburn BB1 2QR doubt contact Prestige
Medical.
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606

www.prestigemedical.co.uk
sales@prestigemedical.co.uk

Destination Kit Date of purchase

Model no.
0086

Prestige Medical Limited


East House, Duttons Way,
Shadsworth Business Park,
Serial no. Blackburn BB1 2QR
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606
www.prestigemedical.co.uk
sales@prestigemedical.co.uk
Registered in England
764017A Reg No. 2826793
3

Section 1: Overview of Controls

Lubrication reservoir fill cap. Error light will illuminate when Air button, only runs air through
the door is opened or there is the hand pieces. No lubrication
a malfunction in the system is being discharged.

Start button: Starts the cycle of


the H6000 when pressed.

Door switch. When the door


is not closed the unit will not
operate.

Three handpiece locations.

Mist Filter
(6 packs of 12 filters)
Part no. 769001

Absorbent Door Strips


(3 packs of 20 strips)
Part no. 769002

Lubrication chamber
level indicator.

Door, Keeps lubrication in the


chamber when in operation.

764017A
4

Section 2: Installation Instructions

Step 1.
Place the Optimax in a location where you have an air line.
Generally this location will be located near your
sterilization area. You will also need an electrical outlet,
Connect your air line and electrical.

Step 2.
Fill lubrication reservoir with the supplied lubrication fluid.
Make sure not to over fill the reservoir or spillage will
occur. After filling, be sure to tighten the oil reservoir cap.

Step 3.
Place absorbent pads into place. These pads will absorb
any excess lubrication that is released in the chamber.
The pads need to be replaced every tank cycle.

764017A
5

Step 4.
Open the door and attach up to 3 handpieces to the unit
· When attaching high speed handpieces be sure to
tighten the handpiece securely to the nut.
· When attaching straight noise cone and contra angles,
make sure that the latch attach is secure.

Notice how all the latches are almost Notice the latch is not flush with
flush with the connector. the stem of the E-type connector.

CORRECT ✔ NOT CORRECT ✗

Step 5.
After attaching handpieces, close the door. The unit will
not operate if the door is not properly closed.

764017A
6

Step 6.
Select the S (Short-35 second cycle), L (Long-45
second cycle) or E (Extra long-50 second cycle) for each
individual handpiece. If no light for 1,2 or 3 is illuminated
the position is turned off.
· SHORT CYCLE For high speed handpieces
· LONG CYCLE For low speed handpieces & air motors
· EXTRA LONG CYCLE For speed reduction handpieces
· NO LIGHT ILLUMINATED Hand Piece position is in the
off position

Step 7.
Press the start button. In less than 3 minutes all
handpieces are correctly cleaned and lubricated.
(3 minutes for 3 handpieces)

Step 8.
Press the air button if you wish to run a cycle of only air
through your handpieces. This process removes any
excess lubrication fluid from the turbine.

764017A
7

Step 9.
Remove your handpieces from the unit. Lubrication and
cleaning is now complete. To remove the handpiece turn
the HP nut counter clockwise to un-screw it.
For E-type attachments press the button release to
unlock and remove the straight noise cone or contra
angle. You may proceed to sterilization (if
necessary).

Step 10.
You may adjust the pressure in the reservoir to adjust
the amount of air pressure that is supplied to the hand
piece. Pull the knob outward to unlock it. Turn the knob
counter clockwise to lower the pressure.
Turn the knob clockwise to increase the pressure.
Reference the air pressure gauge on the side of the unit
to verify the set pressure.
The silver knob located next to the air pressure
adjustment knob is used to regulate the amount of oil
that is dispensed in a cycle. If you wish to increase the
amount of oil, turn the valve counter clockwise. If you
wish to decrease the amount turn the valve clockwise.
This must be adjusted prior to starting a cycle.

764017A
8

Contact us for advice on infection control

Prestige Medical Limited


East House, Duttons Way,
Shadsworth Business Park,
Blackburn BB1 2QR

Tel : +44 (0) 1254 682 622


Fax : +44 (0) 1254 682 606

www.prestigemedical.co.uk
sales@prestigemedical.co.uk
@PrestigeMedica1
Due to a continuous process of
Registered in England development, Prestige Medical reserves
Reg No. 2826793 the right to amend the specification of
Vat No. 633707835 products and services without notice.

Registered Office:
Barlow Road,
Aldermans Green Industrial Estate,
Coventry CV2 2LD

Prestige Medical is a registered trademark