Você está na página 1de 14

 

 
BRSM 
  Systems:QMS ISO 9001 (2008) 

 
Assessment Report/Checklist 
Form 009 

 
CLIENT‐ORGANIZATION (NAME):    Parseh Pathobiology And Genetics Laboratory 

PHYSICAL LOCATION(s):   Parseh Medical Building, Jenah St., Sadeghieh 
Sq., Tehran, Iran 

Phase II Audit Date(s):    24‐NOV‐2016 

Phase I Audit Date(s):     

NAICS (or NACE) CODE 86.90

ISO 9001 EXCLUSIONS:    7‐3 

 
Assessment objectives: to audit requirements of iso9001:2008 
Iran Refrence lab regulations in ms of the 
company 
 
 
ASSESSMENT TEAM INFORMATION 
Assessment Team Leader, ATL Gita Zarsazi 
Assessor 2  
Assessor 3  
Assessor 4
Assessor 5
Assessor 6
 
 
Note: Adjusting the language within this form to the type of activity and nature of the organization is acceptable.
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

We pursuit to assure that client‐organization adheres to the protection of communities and consumers. This is to 
1  follow the intent to the substitution of the word “identify” to the word “determine” as part of the importance to 
the processes being capable of achieving desired outputs (including consumer’s well‐being). 

2  Scope of activities  Performing Clinical, Pathological and Genetic Tests 

BRSM requires that the organization’s legal obligations and regulatory 
3  Legal status  are met as part of granting certification.   
Certificate of lab 1391-12354-‫ ب‬valid until end of Shamsi Year
4  Quality Management System 
There was no change in products, locations and scope. Company uses BRS logo 
on packages and head letters as work instruction. 
Initial Audit: 
PL‐01‐15:
There was no record of "Bioscientia laboratory" in Germany, in the case, records of 
supplier evaluation have been observed.
PL‐02‐15:
General  Corrective action is not done regarding non conformity No. 03 in 
Requirements –  microbiology section about exhaust of hood, but there is no corrective 
4.1  What is the  action regarding to this.  √
implementation /  The action plan is effective, correction has been observed and in audit 
transition date?  no case founded regarding this NC
PL‐03‐15:
In some cases organization did not do trend analysis such as post test process 
The action plan is effective, correction has been observed and in audit 
no case founded regarding this NC
Such as trend of customer satsifaction or trend of reception

Documentation 
4.2  Requirements  No change happened in scope, methods and places √
(includes records) 

4.2.2  Quality Manual No change happened in scope, methods and places √

Control documents procedure PL-PR-14. Approval documentation or


supervisor's responsibility is responsible for the process or management
Control of 
4.2.3  representative. Technical director's approval documentation or √
Documents  documentation
Accesibility to documents are in job description of QA manager/MR

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 2 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

whom is Mrs. Soleimani she give access to all staff via the internal
network is provided. The records document the change is in form "
Development and Change Document"
Control  of  Record control procedure PL-PR-05. Distribution, maintenance, location 
4.2.4  Records  –  Are  and method of elimination are in records list PL-RF-107 is specified. √
records…  Such as insurance documents and genetics records duration is 5 years.
purchased items to one year. Records of internal quality control tests a
year.

All NC and observation has been enhanced


Management 

Commitment 
Software of reference lab has been enhanced √
Customer satisfaction and customer reception rate increased

Customer  Focus  – 
Actions taken regarding Customer feedback shows customer focus and 
5.2  (may  relate  to 
refer to 7.2.1 and 8.2.1 we did find evidences of customer focus  

7.2.1 and 8.2.1) 

Quality policy review has been observed in management review on


5.3  Quality Policy √
20/6/1395 (September 2016)

Planning;5.4.1 Objective table of past year and new year has been observed 
Quality
PL‐RE‐12 are the form which objectives break down in it, such as: 
Objectives/5.4.2
5.4 
quality Promotion of customer satisfaction criteria is the number of complaints 
management that is 6 and the objective is to decrease it to 4 in this year 
system planning

5.5 

Organization chart, PL‐RE‐24 contains all necessary departments and 
based on chart, job description PL‐JD‐29 have been provided, I checked 
Pathology department operator job descriptions such as: 
1‐Tasks related to the availability and operating the devices, and other 
Responsibility, 
5.5.1  technical factors for laboratory tests  √
Authority  
2‐ Survey and control different parts of laboratory devices and devices 
with violations to the appropriate authorities 
Management representative is Mrs. Soleimani from two years with a 
letter from  Mir Majid Mosallaiee 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 3 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Management  representative  is  Mrs.  Soleimani  from  three  years  with  a 


Management 
5.5.2  letter from  Mir Majid Mosallaiee  √
Representative 
She is Quality assurance manager too. 

Organization chart, PL‐RE‐24 contains all necessary departments such as:
Molecular. The reception and response, and sampling and write. 
Internal Department of Microbiology .biochemical, Fertility and Infertility. 
5.5.3  √
Communication Molecular section. Department of Pathology and Cytology 
Communication is done by internal telephone system, email and formal 
letter or forms. 
5.6 

Management review procedure PL‐PR‐02. The frequency is annual 
5.6.1  General  management review. All registered inputs to the meeting procedure.  √
Output of meeting the specified procedures, corresponds to the 
standard requirements. 
The last minutes of management review dated management review on 
20/6/1395 (September 2016) and the period specified 12 month in the 
5.6.2  Review Inputs  procedure.  √
Outputs of management review observed Such as: 
Customer satisfaction 
Improving the staffs health 
5.6.3  Review Output Increasing the due date of tests answers   √
Increase level of awareness of each department 

6  Resource Management

6.1  Refer to 6.2 ,6.3, 7.4 ,8.2.1


6.2  Human Resources 

There is PL‐PR‐01 "training procedure" for this clause of standard, which 
6.2.1  General 
contains  Training  needs  assessment,  training  calendar,  evaluate  the  √

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 4 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

effectiveness of training 
Ms. Mohamadipor knew about transferring the sample of patient  
There  was  a  training  courses  for  Mahanz  Babai  to  improve  the  ISO 
9001:2008  to  2015,  training  course  was  internal  and  done  by  Mrs. 
Soleimani, effectiveness of course is weighted by written exam and her 
grade id 18 ,the acceptance grade is consider 12 
Mrs  Gholami  is  working  in  reception  for  almost  2  year,  course  of 
“Dealing  with  clients”  is  hold  for  hir,  effectiveness  of  the  course  was 
reducing  the  complaints  of  clients,  complaints  of  the  September  was  8 
but the complaint in October was 2  

PL-PR-01
Competence,  There  was  a  training  about  “Infection  and  Disinfection”  for  Mr.  Akbari  
6.2.2  Awareness  and  on august 2016 effectiveness of course is weighted by written exam and  AR
Training  his grade is 15 ,the acceptance grade is consider 10 
Proper  sampling  techniques  course  is  hold  for    all  technical  experts  on 
July  2016.  Documents  of  Mr.  Semnani  observed  and    effectiveness  of 
course is weighted by written exam and his grade is 19 ,the acceptance 
grade is consider 10. 
In  case  of  job  Improvement,  there  was  no  need  assessment  for  job 
holder, such as Ms. Fahime Mohammadpour which was QC Department 
but now is in QA department 
 
 
Records  of  PM  for   Conventional  thermal  cycler  Corbett  has  been 
observed such as: 
Daily: cleaning after and before work 
Annually:  overhaul  by  supplier  which  is  done  on  5/5/95  (august  2016) 
with an accredited company by reference lab.  
Verification and validation of overhaul done on same date on MOM no 
95‐102. 
record of log of photometer A33 observed  

PL‐R‐2‐00 
Record of overhaul and calibration of photometer on June 2016 
6.3  Infrastructure  OBS 
Lamp adjusting 
Cleaning sucking  box. 
Compressed air passages cleaning. 
Calibration and validation is done by "ASA calibration company" includes 
accuracy and repeat test 
Average 96.2 
SD 1.58 
Result approve 
 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 5 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Work environment clean, Record of resource management checklist (all 
resources except human resource) checked monthly by Mr. Semnani.  
For records checklist on end of May 2016 has been observed. There are 
two corrective action serial 01 15/3/1395 and 02 18/5/1395 regarding 
this checklist to change test tubes tray and air‐condition, the root cause 
Work  was test tubes tray and air‐condition are rusted during time. 
6.4  √   
Environment   Correction: buying test tubes tray and air‐condition 
 Surface testing is done monthly due to reference lab regulation, plan of 
random test has been observed in PL‐CL‐25. Surface test of reception 
desk and laminar hood has been observed for August 2016. 
 

7  Product Realization 
There are SOPs in laboratory which are based work instruction,
refer to 6.2.2 samples and performance of staff, equipments are
validate annually (refer to 6.3) and based on 6.4 of this report,
company controls infrastructure monthly.
Planning of 
Reception unite asks information of patient due to Drs. Prescription 
7.1  Product 
of patients or in some cases such as FBS verbal inquires will be accepted 
√   
Realization 
For records: 
Ms. Bakshhaee, with admission No. 41229
Company has a list on reception desk as a guide for subcontracted
laboratories.
Ms. Bakshhaee, with admission No. 41229
Company uses an ERP so after appointing system patient waits
Determination  of  until his/her turn. so reception gives 2 days for answering except
Requirements  there is a exception (which are described in WI-RP-05) such as
7.2.1 
Related  to  the  "normal molecule" test which takes 7 working days √   
Product   For above three cases I saw records of sampling information in
sampling unit on the touch computer. Company uses automatic
labeling system on samples. 

7.2  Customer‐Related Processes 

There are rare situation for change in contract, parse lab has no
Review of 
record of change in recent years
Requirements 
7.2.2 
Related to the  Reviewing the tests required for patient is the duty of the Supervisor and √   
Product   reception section staff. They have to reconfirm the name of test with
patient verbally. The tests specifies by the prescription of the physician. 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 6 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Communication with customers (patient and Dr) is executed through the


ERP system such as:
Digital board
Customer  Appointing system
7.2.3 
Communication  SMS panel √  
Emails
telephone for emergency cases
face to face communication.
7.3  Design and Development 

Design and 
D&D is excluded because of nature of organization, labs shall
7.3.1  Development 
use national-international approved methods 
√   
Planning 

7.3.2  D & D Inputs  NA     

7.3.3  D & D Outputs  NA     

D & D 
7.3.4  NA     
Reviewing 

7.3.5  Verification  NA     

7.3.6  Validation  NA     

Control of (D & D) 
7.3.7  NA     
Changes 
7.4  Purchasing 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 7 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Procedure PL‐Pr‐21 is for purchasing, verifying and supplier evaluation, 
Form of assessment of subcontractors is PL‐RF‐91 
Purchasing 
7.4.1  For case Ms. Bakshhaee, with admission No. 41229, test tube lot is   
Process 
160410 that has been bought in May 2016, provider is “AVA Pezeshk” 
Grade and score is 92, acceptable score is more than 85 
another subcontractor is Biosceintia in Germany, as procedure PL‐RF‐91 
records of evaluation has are available in laboratory, grade is 83 and 
certificate of this lab "Biosceintia" from American pathologist has been 
observed which was valid to June‐10‐2018 

Purchasing 
7.4.2  √

 
Information  

Verification of purchase is done in two steps: 
1‐ Visionary control include date, quantity and packaging 
2‐ Verifying with control and standards before first use (refer to 
7.5.1 and 8.2.4) 
Verification of 
7.4.3 
Purchases 
Record of verification of Vitamin D3 Kit has been observed,  √
Date of arrival: 28/5/1395, lot numbers: 
5415413 expiration date march 2018 
7541541 expiration date June 2019 
6094854 Expiration date march 2016 
Responsible is Dr. Mr Mohseni 

7.5  Production and Service Provision 

Control of 
7.5.1  Production and  According to the developed SOP’s, all tests which are executed are based √  
Service Provision  on SOPs. The record of executed tests till result preparing were observed

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 8 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Validation of  such as admission numbers:


Processes for  For case Ms. Bakshhaee, with admission No. 41229
7.5.2 
Production and  All results will be delivered just to the patient or the individual who has
Service Provision  the admission receipt.

Identification  and  Its traceable to test tube lot 41229


7.5.3  Vitamin D3 Kit lot 187364732
Traceability 
Mindray SOP has been observed in Farsi
Samples and reports are considering as customer property
Reports are in ERP program, back up period is every hour on HDD
and daily on server. Server uses raid technology.
Customer  There are Non Conformity for customer property such as:
7.5.4 
Property  Patient with admission Number WR 58, sample became slippery,
√  
Mr Hosseini called patient and asked him to come to re-test, he
didn’t claim anything regarding Non Conformity notebook of
reception. 
Preservation of 
7.5.5  -  
Product 
Control of 
Monitoring and 
7.6   
Measuring  -
Devices 
8  Measurement, Analysis and Improvement 
8.1  General  As 8.2&8.3&8.4     
8.2  Monitoring and Measuring 
There are three kinds of customers: 
Patients feedback is via verbal, surveys, SMS, Telephone, Email and 
website 
Doctors verbal, surveys, Telephone, SMS 
Laboratories which sent samples to Parse lab annually surveys 
Surveys include:  
Variety of test 
Consistent results with the patient's clinical status 
Customer 
8.2.1 
Satisfaction 
Manner of staff:  √  
Reception 
Sampling 
Answering 
Technical supervisor 
Company will take action regarding to weak or unpleasant situation via 
telephone, such as case 19/4/1395, there were no compliant recorded 
from Initial audit to now.  
It show effectiveness of company objectives. 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 9 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Internal audit shall be hold at least once a year, audit checklist are base 
on reference lab requirements, ISO 9001, ISO15189, EMS and OHSMS. 
Internal audit is done by  a team, Mrs. Soleimani,  Mr. Semnani and Mr. 
Maleki as Consultant and lead auditor on 9&8‐ 6‐95(August 2016) 
8.2.2  Internal Audits  "sending sample to reference laboratories checklist" has observed   √  
Non Conformity was as below: 
4  corrective  action  for  serology,  12  for  training  effectiveness, 
effectiveness    of  corrective  action  was  14/07/95,  correction  were 
followed effectivley. 

Reception process: 
Criteria: Duration when patients wait 
Measurement period 3 month 
Acceptance level: 14 minutes and in first 3 month of 1395 result became 
12  minutes.  Corrective  action  of  last  year  was  changing  lining  program 
and  now  it  is  effective.  Ro  reach  12  minutes  they  recruited  a  new 
sampler. 
Monitoring  and 
 
8.2.3  Measurement 
Post‐test process:  √  
Process 
Criteria: Number of Non Conformities 
Measurement period 6 month  
Acceptance level: maximum 10 Non Conformities 
Result is 3 in first 6 months of 1395. 
 
Data  analysis  output  is  a  plan  to  enhance  phisycal  area  of  test  lab  and 
buy new equipments. 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 10 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Monitoring  and  measurement  of  final  product  is  done  for 


every case by supervisors (Dr. Mohseni and or Dr. Mir Majid 
Mosallaiee)
For Processes as SOPs, setup is kit and equipment test, for out
of range results there should be a repeated test which confirms
results
Monitoring  and  For the record:
8.2.4  Measurement  of  Mc fareland QC test has been observed in Microbiology √  
Product  department
Cr CV is 1.3 for august 2016 in form pl-rf-55 for lot 95003

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 11 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

Every department has Non Conformity notebook, I wrote Non


Conformity of reception above, for record:
Control of Non‐ Sampling
8.3  Conforming  Sampling test was not enough  √

 
Product   Corrective action:
Calling to client for re-test

Reception process: 
Criteria: Duration when patients wait 
Measurement period 3 month 
Acceptance level: 14 minutes and in first 3 month of 1395 result became 
12  minutes.  Corrective  action  of  last  year  was  changing  lining  program 
and  now  it  is  effective.  Ro  reach  12  minutes  they  recruited  a  new 
sampler. 

PL-RE-15 
 
8.4  Analysis of Data 
Post‐test process:  AR
Criteria: Number of Non Conformities 
Measurement period 6 month  
Acceptance level: maximum 10 Non Conformities 
Result is 3 in first 6 months of 1395. 
 
Data analysis output is a plan to enhance phisycal area of test lab and 
buy new equipments.

8.5  Improvement 

Refer to 8.2.1 
Continual 
8.5.1 
Improvement 
Refer to 5.4.1  √  
The percentage of accomplished MOM output to all output is 100% 
There are 46 CA and 8 PA forms in period of surveillance, such as CA#36 
NC: Mr. chegini transferred urine cup in hand and not in special baskets 
Root Cause: there is just one basket in organization, negligence of Mr. 
chengini 
8.5.2  Corrective Action  Company bought one more basket with another color to distinguish the  √  
place basket shall be placed 
Re‐training of Mr. chegini by me semnani 
 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 12 of 14 
REFERENCE, AS
This Clause

APPLICABLE1 
Has been
REFERENCE CRITERIA FOR GUIDANCE TO FOLLOW UPS – PLEASE WRITE ANNOTATION OR X TO INDICATE SPECIAL Verified,
  Reference Clause  ATTENTION WILL BE REQUIRE   please Tick
mark √ 
 

PA#8 to preventing customers from cold, adding 
8.5.3  Preventive Action  √  
automatic door to reception
 

FORM‐009_V4_QMS                                      Issued: July 9, 2003                                                               Revised: March 2013                                                       Page 13 of 14 
By exception, we indicate the specific Strengths and weakness of the organization in light of the management system, as we perceive and seen through
processes in reference to requirements of the criteria being ascribed as these apply and relate to the assessment plan and planning of the assessment.
This aspect relates to Accreditation Body as it does to the protection of consumers in the client-organization’s role within their respective supply – chain.

Disclaimers: (1) herein, we reiterate that any suggestion or opinion that the assessment team expresses does not construe to be a contractual
requirement. (2) The degree of certainty in conclusions increases by the competence of the assessment team as well as the attestation within the
assessment protocol, values, mission and vision under which we operate for the protection of consumers, still remains a certain level of uncertainty.

Strength Weakness

 
 

action Request Type I Form-008 – as we indicate and that each of the RA addresses the basic four (4) questions
as seen in Form-008:

action Request Type II Form-008 – as we indicate and that each of the RA addresses the basic four (4) questions
as seen in Form-008:
PL-01-16: In case of job Improvement, there was no need assessment for job holder, such as Ms. Fahime
Mohammadpour which was QC Department but now is in QA department

Observation Form-018 – with uniquely specific identification:

ATL: Gita Zarsazi DATE: 24‐NOV‐2016

END of REPORT
FORM-014 – Opening Meeting Checklist FORM-009 – MS Reporting,
FORM-017 – Closing Meeting Checklist FORM-008 – RA / AR, as requires
FORM-005 – Participants of opening & closing meetings FORM-018 – OBS, as requires
FORM-004 – Assessment Plan FORM-010 – As requires, recommendation for CoR
FORM-007 – Planning onto -004, as process determined FORM-026 – Assessment documents check-sheet, optional

Você também pode gostar