Você está na página 1de 6

Feature Report

Engineering Practice

Pressurized Piping:
Sampling Steam and Water
Without proper Isolation Source: Jonas
valves

systems,analysis Flow < 20 ft.


< 200 ft.
sloping
down
of steam and water Primary
cooler

chemistry can provide Isokinetic Sampling Expansion

erroneous results — sampling nozzle tubing coil


Secondary
cooler/chiller
with costly implications On-line
instruments
T P

Lee Machemer Analyzer


flowmeter
Sample filter
with bypass Total Pressure Thermal Pressure
Jonas, Inc. flowmeter gauge shut-off reduction
valve valve
Back

C
pressure Grab Temperature
orrosion and deposition in boil- regulator samples indicator
ers, steam turbines and many
types of process equipment Figure 1. This figure shows an example of a well-designed sampling system for
are among the most expensive extracting and conditioning a representative sample of steam or water. It includes
isokinetic sampling, rapid condensation and cooling, pressure reduction, and
causes of outages in utility and in- process indicators, as well as safety devices to protect online instruments
dustrial steam plants. Deposits and and plant personnel
scale buildup on heat transfer sur-
faces reduce efficiency, and when al- industrial steam plants do not or periodically. Proper design of the
lowed to accumulate on steam tur- have properly designed and oper- sampling systems is critical in order
bines, such buildup can reduce the ated sampling systems to monitor to produce samples and analytical
megawatt capacity. Corrosion-re- water and steam chemistry. In fact, results that are representative of
lated failures can result in outages in water chemistry and corrosion the sampled stream [3–9]. Problems
ranging from a few days to several control audits, sampling problems with sample withdrawal, transport,
months, depending on the affected are found in roughly 70% of all collection, and handling are often
systems, and can potentially cost plants. As a result, operating deci- major sources of errors that can
tens of millions of dollars. sions are often based on data that then lead to incorrect or unneces-
To reduce the risk of corrosion can have sampling errors as high sary corrective actions by the op-
and deposition in water and steam as +/- 1,000%. These errors, as well erators. A meticulously performed
systems, the standard practice is to as data inconsistencies and concen- chemical analysis is of little value
monitor cycle chemistry and control tration swings in the analytical re- if a bad sample is used. As shown
impurity levels within industry- sults, become commonplace and are in the Box on p. TK, there are many
and manufacturer-recommended often ignored by plant personnel, potential causes of sampling errors,
limits for the equipment. In steam preventing the timely identification some of which can cause analysis
plants, the chemical parameters of of actual chemistry excursions. This results to be orders of magnitude
interest include: pH, conductivity, article outlines the principles that higher or lower than the actual con-
sodium, calcium, magnesium, chlo- must be considered when designing centration in the process stream.
ride, sulfate, fluoride, phosphate, and operating water- and steam- In high-purity systems, the mea-
acetate, formate, propionate, total sampling systems. sured concentration of impurities in
organic carbon (TOC), silica, cop- many of the process streams is in the
per, and dissolved and suspended Sampling system design low parts-per-billion (ppb) range. At
iron (oxides). Typical target concen- To monitor systems for the ingress such low concentrations, the fluid
trations are in the range of <1 part of impurities and for the production being extracted is very sensitive
per billion (ppb) to several parts per and transport of corrosion products, to any deposition or chemical reac-
million (ppm) [1, 2]. several cycle streams are sampled tions within the sampling system.
Unfortunately, many utility and and analyzed, either continuously The extraction of non-representa-
2 Chemical Engineering www.che.com January 2014
Jonas
Causes of Sampling
System Errors
(in order of priority/impact)

O
perators should be mindful of these
common common sources of sam-
pling errors:
• Sample withdrawal — Sample does
not (1) represent the stream (due to
wall effects, stratification, not isoki-
netic, or mixing issues), and (2) rep-
resent all phases (solid, liquid, gas)
FIGURE 2. This weld-in style, single-port isokinetic sampling nozzle meeds cur- • Deposition in the sample line (could
rent ASTM standards for sampling water and steam. Flanged connections to the also result in plugging)
process pipe are also acceptable • Release of built-up deposits into the
sample stream (leading to spikes)
tive samples can lead to large sam- sample line blockage and can cause • High pressure drop leading to insuf-
pling errors — as high as 1,000% unacceptable time lags between ficient sample flow (typically due to
[3]. Even in lower-purity systems, sample collection and analysis. A long sample lines)
sampling errors due to improperly sample flowing at 2 ft/s through 500 • Changes in sample flowrate (for in-
designed sampling systems can be ft of tubing will take over four min- stance, sampling system can take up
significant. utes to reach the analyzers. to 6 hours to reach equilibrium from
the start of sample flow)
A well-designed sampling system
(Figure 1) consists of an isokinetic Why isokinetic sampling? • High sample temperature (can lead to
sampling nozzle (discussed below), Isokinetic sampling is the extraction pH and conductivity errors)
isolation valves, sample tubing, a of a representative portion of the • Chemical reactions in sample lines or
primary cooler (for steam and high- process stream without altering the coolers (reduction of oxygen concen-
temperature liquid samples), a physical and chemical properties of tration, change in pH and so on)
secondary sample cooler, pressure- the sample. In isokinetic sampling, • Corrosion of the sampling system may
reduction and total-flow-regulation all phases (solid oxides and precipi- lead to generation of corrosion prod-
valves, a thermal-shutoff valve (for tates, liquid droplets, and vapor) of ucts (as a result of improper materials
of construction)
process temperatures above 100ºF), the sampled fluid enter the sam-
back-pressure regulator and sam- pling nozzle with the same velocity • Filters in the system interfere with de-
ple drains. Because steam impuri- vector (meaning the same velocity sire to sample suspended solids
ties are easily adsorbed by magne- and direction of flow). The main • Sorption on sample tubing and sus-
tite (Fe3O4), the oxide buildup on reason isokinetic sampling is nec- pended oxides may remove a por-
the inner diameter of the sampling essary is that the sampled stream tion of the chemical species being
monitored
nozzle and tubing should be mini- is almost always a two-phase fluid
mized. For this reason, all wetted (gas-liquid, gas-solid, liquid-solid)
components of the sampling system and the second phase typically has
should be made from at least Type a very different chemistry composi- the pipe. Proper sampling nozzle
316 stainless steel, and carbon and tion than the steam or water [2]. In design must consider the effects of
low-alloy steels should be avoided. addition, the second phase (droplets flow-and vibration-induced forces
Deposit buildup in the sample or particles) typically has a different on the nozzle, as well as the design
lines can result in plugging of the density and inertia compared to the pressure and temperature.
sample line or seizing of sample primary phase (gas or liquid) and Prior to 2006, the ASTM Standard
isolation valves. Even when not therefore would not be proportion- D1066 “Standard Practice for Sam-
directly affecting sample flow, de- ally represented in a sample that pling Steam” [4] included a multi-
posits in the sampling system can was not withdrawn isokinetically port sampling nozzle which, in its
affect the sample accuracy. Deposits (Figure 3). The benefits of isokinetic most basic form, consisted of a piece
can act as ion exchange media and sampling have been verified during of pipe with multiple holes in it. The
adsorb or release impurities during an EPRI project [3] and through an sampling pipe extended most or all
changes in the flow conditions. Even independent analysis [10]. of the way across the process pipe
the best sampling system design is and was supposed to simultane-
still susceptible to deposition and Sampling nozzle design ously sample from several locations
plugging if the cycle chemistry at The design of the isokinetic sam- across the diameter of the pipe.
the plant is not maintained within pling nozzle (Figure 2) is a critical However, research has shown that
industry standards, particularly part of the sampling system, and such a multi-port design operates
when high concentrations of corro- should be performed prior to the se- non-isokinetically, is prone to plug-
sion products (such as iron oxide or lection of the other sampling system ging, and is susceptible to failure
copper oxide) are present. Lengthy components. As noted, if designed due to vibration [3].
sample lines (for instance >100 ft) or incorrectly, the sampling nozzle In many piping applications, the
low sample flowrates (for instance could provide a sample that is not flow in the process pipe is fully tur-
< 4 ft/s) increase the probability of representative of the conditions in bulent. This results in a uniform
Chemical Engineering www.che.com January 2014 3
Engineering Practice

VB VB VB
velocity profile across the pipe and Source: Jonas

the stream is well mixed, so the com-


position is uniform across the pipe.
This makes it unnecessary to sam-
ple at more than one location along
the pipe diameter. In light of this,
the Electric Power Research Insti-
tute (EPRI) developed a single port
sampling nozzle design in the early
1990s [3]. The EPRI-style isokinetic
sampling nozzles have a compact
design, which has the advantage VN VN VN
of being inserted only about 12% of
the way into the pipe, compared to
multi-port nozzles, which must tra-
verse most (if not all) of the diam-
eter of the pipe. In fact, the single- Isokinetic Non-isokinetic
VB = VN VB < VN VB > VN
port EPRI-style nozzle design was
included in the ASTM standard in FIGURE 3. Shown here is the ideal flow path of particulates and droplets into
1996, and in 2006, it became the isokinetic and non-isokinetic sampling nozzles. In isokinetic sampling, the extracted
only recommended sampling nozzle fluid is representative of the composition in the process pipe, including particles and
design included in the standard. droplets. When the sampling is non-isokinetic, the concentration of particles and
VBof
droplets can be higher or lower than that found in the process fluid. Vb = velocity
process fluid; Vn = velocity in the sampling nozzle
Transport of samples
Almost any fluid will leave or pick
up some residue, both while flowing in the sample lines and to reduce the isokinetic sampling nozzle, and
through a tube and being stored in the time required to achieve equi- the primary sample cooler should be
a container. As a result, any chemi- librium between impurities in the located as close to the sample point
cal analysis will become biased due flowing sample and the tubing, the as possible (less than 20 ft).
to the loss or gain of contaminants. sample tubing after the primary The total length of sample
VN tub- VN
Several factors contribute to deposi- cooler/condenser should be sized ing should be as short as possible
tion on the tubing wall, including: so that the sample flow velocity to limit both the pressure drop and
crystallization resulting from solu- is maintained around 5–6 ft/s [3– the lag time from when the sam-
Isokinetic
bility changes, settling due to grav- 8,12]. Several studies have shown ple enters the isokinetic Vsampling
B = VN VB
ity and hydrodynamic forces, and that both linear velocity and the nozzle to when it reaches the ana-
electrostatic attraction of charged Reynolds Number (Re; a unitless di- lyzers. Low sample residence time
particles [6]. mension that describes the amount in the tubing is also preferred, to
In any sampling system, there of fluid turbulence) control the net limit chemical reactions, such as
will be an exchange of contami- deposition of particulates in sample oxygen scavenging and sorption on
nants and particulates between the lines [13–17]. Therefore, the sample oxides. The sample velocity should
flowing sample and the sample line line should be designed to achieve be maintained as constant as pos-
surfaces. Eventually, an equilibrium both turbulent flow (Re>4,000) and sible in the interest of maintain-
state will be reached. Whenever the proper velocity (5-6 ft/s). ing equilibrium between deposition
sample is not in equilibrium with In steam-sampling systems, one and re-entrainment of particles,
the surface, the sample composition of the most critical design consider- and chemical equilibrium between
will be changed from its original ations is the size and length of the the sample and deposits. In one
state. In general, the time for new sample line from the sampling noz- research project [16], it was found
sample tubing to reach equilibrium zle to the primary sample cooler/ that it took less than 30 days for a
decreases with smaller tubing (due condenser. Long and oversized sam- newly installed sampling system to
to decreased surface area) and in- ple lines produce a significant pres- reach equilibrium when the sample
creased sample velocity. Even when sure drop and heat loss from the was flowing at 6 ft/s, compared to
a sufficient sample velocity (say, on extracted fluid. In general, both the several years for a sample flowing
the order of 6 ft/s) is maintained, temperature and pressure of the at 1 ft/s.
the equilibration process can take sample should be maintained right In addition to the effects on sam-
up to a month. It is for this reason up to the primary cooler/condenser ple purity, the use of larger-diame-
that sample streams should flow so that desuperheat and condensa- ter tubing can result in an unnec-
continuously rather than be peri- tion occur together. To achieve this, essary waste of sample water (and
odically started and stopped. the sample line should have approx- additional energy required to heat
In order to minimize deposition imately the same inside diameter as and cool the fluid), require impracti-
4 Chemical Engineering www.che.com January 2014
Table 1. Reynolds Number, sampling rate, annual volume, and pressure drop (ΔP) for water (T =
100°F) flowing through various sizes of tubing at 5 Ft/s
Outer dia. Inner dia. Wall thick- Reynolds Num- Required sampling Annual vol., Estimated ΔP per
(OD), in. (ID), in. ness, in. ber, unitless rate, cm3/min gal/yr) 100 ft of tubing, psi
0.250 0.120 0.065 6.8x103 670 93,000 57
0.250 0.152 0.049 8.6x103 1,070 148,000 42
0.375 0.245 0.065 1.4x104 2,780 386,000 24
0.500 0.370 0.065 2.1x104 6,340 879,000 14

cal sample-conditioning equipment, temperature and pressure, and pro- counterflow design and be sized to
and place an extra and expensive vide a minimum change of cross- ensure adequate cooling capacity,
burden on the makeup system. section between the inside diameter with allowances for reduced heat
Table 1 compares sampling rate, of the isokinetic sampling nozzle transfer due to scale buildup. The
Reynolds number, estimated pres- and the orifice of the valve. Large cooler tubing should be made from
sure drop, and the annual volume of changes in cross-section can result Type 316 stainless steel or Inconel.
water consumed for several tubing in deposition within the valve and Sample tubing after the primary
sizes with a sample flow velocity of may eventually lead to seizing of sample cooler. This tubing should
5 ft/s. Typically, 1/4-inch tubing with the valve, which can become a safety be sloping downward to allow for
a sampling rate of 1,000 to 1,200 issue if the sample line is damaged complete draining during outages,
cm3/min (condensed) is sufficient to during operation. Valves should be and have a minimum number of
provide for all online analyzers and made of Type 316 stainless steel or bends. It should be sized so that the
grab sampling while maintaining a higher alloy. Because valves in sample flow velocity is 5 to 6 ft/s.
the required flow velocities. steam and water service are suscep- Pressure-reduction valve. Such a
tible to deposition inside the valve valve is used to reduce pressure and
Additional considerations (particularly for steam service), it therefore control the flow of a cooled
When designing a sampling sys- is recommended to always have two sample in order to protect on-line
tem such as that shown in Figure isolation valves. instruments. For samples greater
1, follow these recommended prac- Sample tubing between the isoki- than 500 psig, the pressure reducer
tices for each of the components dis- netic sampling nozzle and pri- should be a rod-in-tube type orifice
cussed below: mary cooler. Such tubing should or capillary [5]. For samples less
Installation location for the be as short as possible (not longer than 500 psig, the pressure reducer
sampling nozzle. The preferred than 20 ft for steam systems) in should be a needle valve.
location is in long, vertical sections order to minimize the pressure drop Thermal shut-off valve. This valve
of pipe, away from all flow distur- and reduce the possibility of impu- protects personnel and downstream
bances (such as bends, valves, and rity deposition in the sample tubing. components by automatically inter-
so on) [4,6]. Ideally, the sampling The inside diameter of this sample rupting sample flow when the sam-
nozzle should be at least 35 internal tubing should be close to the inside ple temperature reaches a preset
pipe diameters downstream, and 4 diameter size of the isokinetic sam- limit, in the event of an insufficient
pipe diameters upstream, of any pling nozzle, to minimize changes in amount or loss of cooling water or a
flow disturbances. In many plants cross-sectional area. This normally fouled sample cooler.
where space is at a premium, this is requires ¼- or 3/8-in. tubing for Pressure and temperature
not possible, so it is recommended liquid water and medium- to high- gages and flow indicator. Such
that the sampling nozzle be located pressure steam systems, and 1/2-in. devices provide the operator with
where the ratio of its distance from tube or 1/2-in. pipe for low-pressure verification that the system is work-
the upstream disturbance to down- steam systems. ing properly.
stream disturbance is about 9:1. If The sample line should include a Back-pressure regulator. This
a long vertical section is not avail- series of bends or a coil to allow for regulator is used to maintain a
able, the sampling nozzle may be any movement or expansion of the slight pressure (~ 20 psig) in the
installed in a long horizontal sec- process pipe. Sharp radius bends sample tubing before the grab sam-
tion, provided the sampling nozzle should be avoided. The tubing ple location. This will ensure proper
is installed on the top of the pipe be- should be downward sloping along flow to the online, chemical-analysis
tween the “10 o’clock” and “2 o’clock” the entire length to eliminate any instruments.
positions to prevent the possibility sections where condensed steam or In-line sample filters. These fil-
of water accumulating around the water can accumulate and result in ters should be installed to protect
sampling nozzle during outages. water hammer during startup. on-line instruments during com-
Isolation valves. These valves Primary and secondary sample missioning, or any other time when
should be rated for the application coolers. The coolers should have a high concentrations of corrosion
Chemical Engineering www.che.com January 2014 5
Engineering Practice

products (iron, copper) are pres- the on-line instruments and grab • Check sample temperatures after
ent in the sample. They should be sample tap. A high pressure drop both primary and secondary sam-
installed downstream of the grab- through the system could result in ple coolers
sampling line (as shown in Figure insufficient sample flow at the sam- • Check sample pressure
1), or must be bypassed when ob- ple panel, or the deposition rate in • Ensure flowrate through on-line
taining grab samples for iron and steam sample lines could be high, instruments meets manufacturer
copper analysis. which could result in plugging of requirement
Online analyzers. The sample the sample line or a sample that is • Check for any vibration at the
flowrate, temperature, and pres- not representative of the conditions sampling nozzle and along the
sure must all be within the instru- in the pipe. The design must also length of the sample tubing.
ment manufacturers specifications. ensure that the maximum pressure
A chiller may be required in order recommended by the online instru- Operation and maintenance
to cool the sample streams to the ment makers is not exceeded. Once the sampling system is in-
proper temperature. ASTM D5127 stalled, proper operation and main-
requires that the sample tempera- Commissioning of the system tenance are required to ensure ac-
ture be 25±1°C when measuring After the sampling system is in- curate sampling, including:
pH, and ASTM D5391 requires that stalled, the following tasks should Total sampling rate and sam-
the sample temperature be con- be performed to ensure proper op- pling time. The total sampling
trolled to 25±0.2°C when measuring eration of all components: rate should be governed by the rate
conductivity if specialized tempera- • Check all sampling points to en- required for isokinetic sampling,
ture compensation is not available. sure proper location and sampling which is a function of the sampling
Such strict temperature control nozzle orientation nozzle design and the process mass
may not be practical; therefore, the • Verify that all sample tubing and flow rate. Even if this sampling rate
use of modern pH and conductivity cooling water tubing is properly exceeds the requirements of on-
analyzers that include temperature sized for the required flowrate line analyzers, the total sampling
compensation algorithms may be • Ensure all valves and flowmeters rate should be maintained by rout-
an acceptable alternative. operate properly ing excess flow either through the
Booster pumps. These pumps may • Confirm proper flowrate of cooling grab-sample location to drain or the
be required for long sample lines water to the primary and second- condenser hotwell. For high-purity
(high pressure drop) or low pres- ary sample coolers systems, it can take up to six hours
sure samples (condensate). • Check for leaks along entire of isokinetic sample flow to stabiliz
Once all of the sampling compo- length of sample tubing including the sample chemistry. This time can
nents are specified, the estimated the sample panel be shorter for lower-purity systems,
pressure drop through the system • Perform startup and calibration but for all sampling systems, con-
should be calculated. The pressure of all on-line instruments in ac- tinuous flow is preferred.
drop throughout the entire sam- cordance with the OEM’s instruc- Grab samples. There are many
pling system (including primary and tion manual opportunities for the grab sample
secondary coolers, tubing, valves • Verify that online instrument read- to degrade during collection and
and elbows) must be low enough to ings agree with reading on DCS or storage. This is especially critical
ensure that there is enough pres- other data recording system, and in samples for pH, conductivity, dis-
sure to provide adequate flow veloc- that alarms are working properly solved oxygen, and hydrazine analy-
ity (5-6 ft/s) through the tubing to • Check sample flowrates sis. Special preparation of grab sam-

References Palo Alto, Calif., CS-5164, April 1987. Proceedings of the Workshop on Corrosion-
8. O. Jonas and J. Mancini, Sampling savvy, Product Sampling from Hot-Water Systems,
1. O. Jonas, Corrosion and water chemistry EPRI, Palo Alto, Calif., NP-3402-SR, March
problems in steam systems — Root causes Power Engineering, May 2005.
1984.
and solutions, Materials Performance, De- 9. L. Eater, Make sure water chemistry samples
cember 2001. are representative, Power, July 1989. 14. Emory, B., Crud Sample-System Design Cri-
teria, Proceedings of the Workshop on Corro-
2. Interim Consensus Guidelines on Fossil 10. Binette, V., et. al. Impact of Sampling System sion-Product Sampling from Hot-Water Sys-
Plant Chemistry, EPRI, Palo Alto, Calif., Design on Superheated Steam Quality, Pro- tems, EPRI, Palo Alto, Calif., NP-3402-SR,
CS-4629 and Other Water Chemistry Guide- ceedings of the 5th International Conference March 1984.
lines, June 1986. on Fossil Plant Cycle Chemistry, EPRI, Palo
Alto, Calif., TR-108459, Nov. 1997. 15. Sundberg, L., Sampling of Metallic Impuri-
3. Development of a Steam Sampling System, ties in BWRs., Proceedings of the Workshop
EPRI, Palo Alto, Calif., TR-100196, Dec. 11. Daucik. K., Design of Sampling Devices for on Corrosion–Product Sampling from Hot-
1991. Water/Steam Cycle, Proceedings of the 9th Water Systems,EPRI, Palo Alto, Calif., NP-
4. Standard Practice for Sampling Steam,. International Conference on Fossil Plant 3402-SR, March 1984.
ASTM D1066, 2011. Cycle Chemistry, EPRI, TR-1020563, Jan.
2010. 16. Survey of Corrosion-Product Generation,
5. Standard Practices for Sampling Water from Transport, and Deposition in Light-Water
Closed Conduits, ASTM D3370, 2008. 12. McKinney, J., Analyzers and Steam Panels – Nuclear Reactors, EPRI, Palo Alto, Calif.,
A Perspective from Both Sides of the Fence, NP-522, March 1979.
6. Steam and Water Sampling, Conditioning, Proceedings of the 9th International Confer-
and Analysis in the Power Cycle, ASME Per- ence on Fossil Plant Cycle Chemistry,” EPRI. 17. Svoboda, R., et. al. Trace Analysis of Corro-
formance Test Code (PTC) 19.11, 2008. TR-1020563, Jan. 2010. sion Products by Integrated Sampling Tech-
niques, Water Chemistry 3, British Nuclear
7. Guidelines Manual on Instrumentation and 13. Bird, L., Requirements for Crud-Sampling Energy Systems, London, 1983.
Control for Fossil Plant Chemistry, EPRI, Systems for PWR Primary-Coolant Circuits,
6 Chemical Engineering www.che.com January 2014
ples and/or sample containers may ing all of engineering efforts to ob- nozzle, attachment to the process
be required, depending upon the tain representative samples. pipe, valves, and all welds should be
type of analysis being performed. In Maintaining clean coolers. Pe- periodically inspected for evidence
some cases, chemicals are added to riodic cleaning of the cooling water of cracking and other forms of dam-
the container before the sample is side of the coolers may be required, age. For sampling wet steam and
added to prevent sample degrada- again to maintain proper heat water, the section of process pip-
tion (for instance, samples used for transfer and sample temperature. ing immediately downstream of the
the analysis of iron or copper). The frequency of cleaning depends sampling nozzle should be periodi-
Also, collection methods for sam- upon the scaling properties of the cally inspected for thinning by flow-
ples to be analyzed for pH, conduc- water used for cooling. accelerated corrosion, for installa-
tivity, dissolved oxygen, ammonia, Sample tube cleaning. All sam- tions sampling liquid water should
hydrazine, or organics must exclude ple tubing should be periodically be checked for cavitation. n
any contact between the sample cleaned by flushing or acid clean- Edited by Suzanne Shelley
and the air. Storage methods and ing, or it should be replaced. The
holding times of samples from col- frequency of cleaning depends on Author
lection to analysis require special the amount of impurities in the Lee Machemer, P.E., is
consideration to avoid degradation sample streams. One “quick and president of Jonas, Inc.(4313
Nebraska Court, Pomfret, MD
of samples. dirty” method to test the cleanli- 20675, Phone: 301-934-5605;
Email: jonasinc@steamcycle.
Calibration and maintenance. ness of the sample line is to shut off com) and has worked for the
These steps should be routinely per- the sample flow at the sample panel company for 17 years as a
water chemistry and corrosion
formed on all on-line instruments and then quickly turn on the flow to consultant. Machemer has
per the manufacturers’ recommen- the maximum sampling rate. If the been involved with the design
and development of several
dations. Improperly calibrated and sample is brown or black, there are products used in fossil-fired,
maintained instruments will result deposits in the sampling system. nuclear, and geothermal power-generation facili-
ties. He holds a B.S.Ch.E. from the University
in inaccurate measurements, negat- Maintain safety. The sampling of Delaware and is a Professional Engineer.

Chemical Engineering www.che.com January 2014 7

Você também pode gostar