QUALITY CONTROL

OF NUCLEAR
MEDICINE
INSTRUMENTATION
AND PROTOCOL
STANDARDISATION
EANM TECHNOLOGISTS GUIDE

Produced with the kind support of


TAB LE O F
CO N T E N TS
Foreword 4
Pedro Fragoso Costa
Introduction 6
Sebastijan Rep
Chapter 1 Principles of quality control and quality assurance 9
Goran Vuleta
Chapter 2 Internal and external auditing of quality 17
management systems: the approach of the
International Atomic Energy Agency
Maurizio Dondi, Thomas Pascual and Diana Paez
Chapter 3 Planar camera and SPECT system quality control 29
Mario Medvedec
Chapter 4 PET system quality control 43
Claudiu Peștean
Chapter 5 CT system quality control 53
Urban Zdešar and Dean Pekarovič
Chapter 6 Optimisation of SPECT and SPECT/CT – 67
acquisition and reconstruction
Maximilien Vermandel and Hélène Lahousse
Chapter 7 Optimisation of PET/CT — acquisition and reconstruction* 85
Leesa Ross and Dusty York
Chapter 8 Accreditation for clinical trials: the EANM EARL Project 103
Giorgio Testanera and Michel de Groot
Chapter 9 Radionuclide calibrator 117
Aljaz Socan
Chapter 10 Quality control of nuclear medicine instrumentation 131
and protocol standardisation
Kristof Baete
Chapter 11 Well counters 141
Petra Kolenc Peitl and Marko Krošelj
Chapter 12 Radiation Protection Measurement Equipment 153
Søren Holm

*Articles were written with the kind support of and in cooperation with the

QUALIT Y CONTROL OF NUCLEAR MEDICINE 3


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
FOREWORD

Foreword
In the ever-evolving field of nuclear medicine (NM),
technologists are at the intersection between the clinical,
research and academic domains, embodying the bridge
towards patients. Quality has become an unavoidable word
in NM practice. In the last two decades, NM has earned an
established place in various clinical areas on the basis of the
advances in respect of evidence-based practice and high-level
research. Naturally, nuclear medicine technologists (NMTs) have
been and will be involved in the clinical, research and academic
domains and therefore require the necessary tools to carry out
their tasks in compliance with best practice. This is the main
motivation for choosing Quality Control of Nuclear Medicine
Instrumentation and Protocol Standardisation as the topic for
our annual Technologist’s Guide.

4 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
FOREWORD

The Technologist’s Guide is an annual pub- work to achieve the same outcome in
lication envisioned and edited by mem- the domain of NM: maintenance of the
bers of the EANM Technologist Committee best practice standards to ensure opti-
(EANM-TC). It is one of the many EANM mal implementation of patient-focussed
educational initiatives and has the aim of diagnostic and therapeutic procedures.
completing the training of NMTs and en- I am extremely grateful to all authors for
couraging scientific exchange within the sharing their expertise, which has been
NM community. The Technologist’s Guide fundamental to the successful completion
was and continues to be a reference for of this Technologist’s Guide. I would like to
educational standards inside and outside thank the EANM Physics Committee, the
Europe. SNMMI-TS (Society of Nuclear Medicine
and Molecular Imaging Technologist Sec-
As the topic of this guide encompasses a tion) and the International Atomic Energy
very broad range of applications, we de- Agency (IAEA) for their help in ensuring
cided to divide the book into three parts. the outstanding quality of this book. I am
The first part focusses on the principles very much indebted to the EANM-TC edi-
of quality and standardisation, an under- torial and language revision group for their
standing of which is needed in order to dedication in reviewing and editing this
completely grasp the more practically guide. Finally, thanks are due to the EANM
oriented parts 2 and 3, which are devoted Board, the EANM Technologist Committee
to imaging procedures and non-imaging and all of those involved in the Technolo-
instrumentation, respectively. gist’s Guide project.

This book is a multidisciplinary effort in- Pedro Fragoso Costa


volving different professional groups that Chair, EANM Technologist Committee

QUALIT Y CONTROL OF NUCLEAR MEDICINE 5


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
INTRODUCTION

Introduction
Technologists are members of the team required for
implementation of diagnostic imaging in nuclear medicine
(NM). In many hospitals, the technologists are responsible
for the quality assurance (QA) duties. The development of
hybrid imaging has increased further the need for strict
implementation of quality control (QC) and also rendered
QC more demanding. These new guidelines from the EANM
Technologist Committee address the tasks necessary for the
smooth implementation of QC in NM departments.

6 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
INTRODUCTION

Quality control is required to ensure that QC tests for conventional NM modalities


NM equipment is functioning properly such as planar gamma camera imaging,
and constitutes an important part of the SPECT and PET and also for hybrid meth-
quality management in an NM depart- ods such as SPECT/CT and PET/CT. An
ment. The described QC tests are designed individual chapter is devoted to CT sys-
to detect problems before they affect tem QC as this constitutes an important
clinical patient studies. They are intended element in the optimisation of acquisition
to provide a full evaluation of equipment protocols. The second part covers image
performance and to ensure that equip- optimisation protocols for SPECT/CT and
ment is performing properly after service PET/CT modalities and accreditation for
or adjustment. clinical trials. The third part describes QC
Quality control is important due to the of non-imaging instrumentation, such as
need to optimise patient exposure and radionuclide dose calibrators, intraopera-
image quality during NM imaging exam- tive probes, body uptake probes and well
inations. The image quality is dependent counters.
upon the data acquisition parameters, This overview of QC and protocol opti-
which must be adapted to the detector misation will be a valuable tool for tech-
system and also the reconstruction algo- nologists and all clinical staff involved in
rithm, on the basis of which the acquisi- this particular field.
tion time can be shortened or the adminis- In the name of the EANM Technologist
tered activity of the radiopharmaceuticals Committee, I would like to thank all the au-
can be decreased. thors who have taken the time to prepare
These guidelines cover the principles and write the chapters and to all the edi-
of QC and QA, including QC and improve- tors who have helped to create this book.
ment of imaging protocols for both im-
aging and non-imaging instrumentation. Sebastijan Rep
The first part describes separately the on behalf of the editors

QUALIT Y CONTROL OF NUCLEAR MEDICINE 7


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
1
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
A N D Q UA L I T Y
O F Q UA L I T Y

A SSU R A N C E
PRINC I PL E S
CON T RO L

by Goran Vuleta
CHAPTER 1

Quality is an important factor for any product or process in medicine. It


A ND Q UA L I T Y A S S URA NC E
PRINC I PL E S O F Q UA L I T Y CO NTR O L

is widely recognised that the attainment of high standards of efficiency


and reliability in the practice of nuclear medicine, as in other specialties
based on advanced technology, requires an appropriate quality
assurance programme[1].

In order to maintain or improve the quali- isfactory. QC is a product-focussed con-


ty of equipment and processes, careful at- cept. It comprises a set of procedures
tention must be paid to the two standard intended to ensure that a manufactured
concepts of quality assurance and quality product or performed service adheres
control: to a defined set of performance criteria
»» Assurance: the act of giving confidence. (Table 1).
Quality assurance (QA) is a way of pre-
venting mistakes or failure in products. Hence, QA in nuclear medicine should
QA is a process-focussed concept. It is a cover all aspects of clinical practice. Spe-
systematic process implemented within cifically, QC is necessary in the submission
a quality system (Table 1). of requests for procedures; the prepa-
»» Control: the act of guiding a process. ration and dispensing of radiopharma-
Quality control (QC) is a process involv- ceuticals; the protection of patients, staff
ing the inspection of equipment to en- and the general public against radiation
sure that the quality of all aspects is sat- hazards and accidents caused by faulty

Quality assurance Quality control

Activities Set of activities for ensuring quality in processes Set of activities for ensuring quality in products
Aim To prevent defects To identify (and correct) defects
Process Proactive Reactive
Means of Planned and systematic activities, including Activities or techniques to achieve and maintain
achievement documentation the product quality, process and service
Responsible Everyone involved in the process Specific team that
persons tests the product
Tool QA is a managerial tool QC is a corrective tool

Table 1: Quality assurance versus quality control

10 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 1

Figure 1

Equipment specification Period of use

Acceptance tests Routine constancy test

Comparison of results
Performance check (status test) with reference values

Initial constancy test Are criteria


met?

Reference values for constancy test


Corrective action

End of warranty test


Post-service/corrective action tests

QA and QC cycle for a medical imaging device (based on information in [3] and [4])

equipment; the scheduling of patients; Routine QC testing starts after installa-


the setting-up, use and maintenance of tion of the instrument and continues until
electronic instruments; the methodology end use.
of the actual procedures; the analysis and After installation, and before it is put to
interpretation of data; the reporting of re- clinical use, a nuclear medicine instrument
sults and, finally, the keeping of records[1]. must undergo thorough and careful ac-
A fundamental principle in the QC of ceptance testing, the aim being to verify
nuclear medicine instruments is that it that the instrument has been installed and
should be undertaken as an integral part of performs in accordance with its specifica-
the work of the nuclear medicine unit and tions and its clinical purpose[5]. For each
by members of the unit staff themselves[2]. instrument, a set of basic specifications
All equipment used in nuclear medicine is produced by the manufacturer. These
for examinations (diagnostic procedures) specifications should be traceable.
or treatment (therapy) must be subject to Acceptance testing verifies the manu-
internal QC (Fig. 1). facturer’s specifications.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 11


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 1

Figure 2
At the same time, reference tests should
A ND Q UA L I T Y A S S URA NC E
PRINC I PL E S O F Q UA L I T Y CO NTR O L

be undertaken. These reflect operating


conditions under clinical conditions and
Quality provide results against which to test the
Control ongoing performance of the equipment
by routine testing at weekly, monthly,
quarterly or yearly intervals.
Rountine
OK?
YES
clinical A basic level of routine QC is required to
practice ensure that nuclear medicine equipment
NO is functioning properly. Routine QC tests
are intended to detect problems before
they impact on clinical patient studies
Correction (Fig. 2). Once the instrument has been
possible
localy? installed and accepted for clinical use on
the basis of the acceptance testing re-
sults, its performance needs to be tested
NO
routinely with simple QC procedures that
Call for are sensitive to changes in performance.
service
Tests must be performed by appropriate-
ly qualified and trained staff, and detailed
local operating procedures should be
written for this routine work. All test re-
YES Limited
Limited use clinical sults must be recorded and monitored for
possible? practice variations, and appropriate actions taken
when changes are observed. The QC tests
NO are an important part of the routine work,
and sufficient equipment time and staff
Service time must be allocated for routine QC[5].
All test results must be recorded and
monitored.
Example of a flowchart for a QC The QC of each instrument should have
programme (based on information
as its starting point the selection and pro-
in [2])
curement of the instrument itself, because
instruments may differ in their characteris-

12 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 1

tics and performance. The recommended »» Availability of space (sufficient space for

A ND Q UA L I T Y A S S URA NC E
PRINC I PL E S O F Q UA L I T Y CO NTR O L
frequency of QC tests depends on the sta- the instrument, for the clinical practice
bility of the equipment. The routine test- and for QC and maintenance procedures)
ing can be performed on a daily, monthly, »» Electrical power supplies (which must
quarterly or annual basis. follow instrument specifications regard-
The siting of an instrument in the de- ing voltage and frequency)
partment is largely determined by its ex- »» Temperature, humidity and air pollution
pected use. The selection of a location for (stable temperature and slow tempera-
mounting the instrument can affect the ture gradients, low humidity and clear air)
performance of the instrument (e.g. crystals »» Background radiation levels (location of
in scintillation detectors), and therefore the the hot cell, the storage and movement
QC. Further parameters that can affect the of radioactive materials, and the move-
proper functioning of the instrument are: ment of patients)

Figure 3

MAINTENANCE

PREVENTIVE CORRECTIVE
MAINTENANCE MAINTENANCE

DIRECT INDIRECT PLANNED UNPLANNED

Fixed Time Condition - Planed - Break down


- Service Based maintenance - Emergency
- Calibrate - Inspections - Control of the repairs
instruments - Analysis euqipment - Not possible
- Alignment - Detection of - Possible to to control
failures before control
- Fixed time
replacement break down

Types of maintenance

QUALIT Y CONTROL OF NUCLEAR MEDICINE 13


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 1

All of the instruments used in nuclear results of QC, there is a written description
A ND Q UA L I T Y A S S URA NC E
PRINC I PL E S O F Q UA L I T Y CO NTR O L

medicine are complex systems built from of all QC procedures, acceptable levels of
mechanical, electrical and electronic parts. tolerance and corrective measures in the
Any of these components can fail at some event that results are not within the level
point in time. For this reason, the mainte- of tolerance.
nance of instruments is necessary. Records must be maintained to provide
The goal of maintenance is to avoid the evidence of conformity to requirements
consequences of equipment failure. Main- and of the effective operation of the qual-
tenance of instruments is divided into two ity management system[6]. All records
categories (Fig. 3): must remain legible, readily identifiable
»» Preventive maintenance: a fundamen- and retrievable. They must be permanent
tal, routine and planned maintenance and non-erasable, as must changes to a
activity to keep equipment in operating record[6]. Record keeping is a main com-
status and to avoid unplanned mainte- ponent of an internal QC programme. Re-
nance activity. cords showing frequent malfunction and
»» Corrective maintenance: a set of activi- degradation of equipment performance
ties to detect and rectify a defect so that provide evidence of the need for com-
the equipment is returned to its normal plete instrument repair or replacement.
state.
Record keeping may include (depending
Preventive maintenance in nuclear med- on the equipment):
icine means the maintenance of equip- »» Instrument condition (physical,
ment in a given functional state through mechanical and electronic)
continued overviews, QC and detection »» All calibration records
and elimination of possible failures. Cor- »» QC results
rective maintenance in nuclear medicine »» Instrument maintenance records
means the restoration of equipment to a
functional status by means of repairs. The record keeping can be used to:
As already noted, all obtained re- »» Monitor compliance with QC procedures
sults, from the installation of equipment »» Educate employees
through to acceptance tests and then »» Help prevent instrument breakdowns
QC procedures, need to be recorded and »» Evaluate service personnel
stored. It is necessary that, apart from the »» Help ensure reliable patient results

14 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 1

Recognition by the head of department If the results of certain tests show a de-

A ND Q UA L I T Y A S S URA NC E
PRINC I PL E S O F Q UA L I T Y CO NTR O L
and the management of the institution viation from the allowed tolerance and ac-
of the need for QC is essential to its sat- ceptability it is necessary to decide wheth-
isfactory implementation and adequate er the instrument is fit to be used or needs
funding. to be put out of operation. This decision
It is necessary to clearly define who is must be the responsibility of people with
responsible for each aspect of QC as well clearly defined responsibilities.
as who will supervise the entire QC plan. The choice of tests and the frequency
That person must know all the technical of their performance must be specified
details and should be involved in the eval- for each device in nuclear medicine in
uation of the results. It is important that order to take account of their condition
tests for certain instruments are carried and their status. Protocols should be
out by people familiar with their use, and adapted to suit individual instruments. It
responsibility for daily and operational is essential that these protocols are strictly
tests should rest with the operators who followed.
regularly use these devices.

REFERENCES

1. International Atomic Energy Agency. Quality control surance for PET and PET/CT systems. IAEA Human
of nuclear medicine instruments 1991. IAEA-TEC- Health Series No. 1. Vienna: IAEA; 2009.
DOC-602. Vienna: IAEA; 1991. 5. Busemann Sokole E, Plachcinska A, Britten A,
2. International Atomic Energy Agency. Quality assur- Georgosopoulou ML, Tindale W, Klett R. Routine
ance for SPECT systems. IAEA Human Health Series quality control recommendations for nuclear med-
No. 6. Vienna: IAEA; 2009. icine instrumentation. Eur J Nucl Med Mol Imaging
3. International Electrotechnical Commission. Evalua- 2010;37:662–671.
tion and routine testing in medical imaging depart- 6. ISO 13485. Medical devices – Quality management
ments - Part 1: General aspects. IEC 61223-1. Geneva: systems – Requirements for regulatory purposes, 3rd
IEC; 1993. ed. Switzerland; 2016.
4. International Atomic Energy Agency. Quality as-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 15


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
2 I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
IN T ER N A L
AND EXTERNAL
AUDI T I N G
O F Q UA L I T Y
MA N AGE M E N T
SYSTE M S: T H E
A PPROAC H O F T HE
IN T ER N AT I O N A L
ATO M I C E N E RG Y
AGENC Y

by Maurizio Dondi,
Thomas Pascual
and Diana Paez
CHAPTER 2

INTRODUCTION
INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y
MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y

The health care sector is highly regulated and relies on state-of-


the-art diagnostic technologies. Additionally, health care costs are
usually covered by a third party, such as an insurance company or a
government programme. Third party payers request suppliers to show,
and document, that their products are not only controlled, to ensure
consistency, but also provided in adherence with national regulations
regarding patient and worker safety, and that medical practice is based
on evidence.

This requires the identification of quality quality measurement systems and reward
policies and objectives and the produc- mechanisms.
tion of a documentation system with
clearly defined processes, procedures
and responsibilities. Such a system is usu- NUCLEAR MEDICINE AND
ally referred to as a quality management QUALITY MANAGEMENT
system (QMS) and its purpose is to help Nuclear medicine services (NMS) are mul-
coordinate and direct activities in order tidisciplinary by nature as several different
to meet customer and regulatory require- professional competencies are involved,
ments and to improve effectiveness and each with its own regulations, processes
efficiency on a continuous basis. In health and outputs. However, they all contribute
care, effective quality management is fo- to the success of the discipline. For this rea-
cussed on the needs of patients because son, a comprehensive QMS is not limited
they are the ones who judge the effec- to quality assurance/quality control (QA/
tiveness of treatments and the appropri- QC) but has to involve all aspects of NMS,
ateness of the service. including but not limited to: clinical appli-
Quality management in health care cations, including machinery handling and
requires the close cooperation of people their QA/QC; radiopharmaceutical prepa-
with diverse expertise and is essentially rations and, again, their QA/QC; radiation
about delivering consistent quality, which, protection of both patients and staff and
in turn, is dependent upon reliable pro- of the environment; and the ability of final
cesses. Reliability requires the existence reports to satisfy clinical questions.
of performance goals, risk reduction pro- Several factors may influence the struc-
cedures, quality improvement policies, ture of a QMS in nuclear medicine, includ-

18 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 2

ing the size and structure of the NMS and »» Written standard operating procedures

INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y


MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y
the financial resources available. The latter (SOPs) for both primary (diagnosis and
in turn have an effect on the complexity therapy) and supporting processes and
of clinical practice, which may vary from their reference documents, such as
limited applications to more sophisticat- guidelines from scientific societies.
ed ones. But, whatever the situation, any »» Records of indicators and parameters.
NMS should implement, document and »» Identification of a quality manager, to be
maintain a QMS. Its effectiveness should appointed by and to report to the head
be continuously improved in accordance of the organisation. The quality manager
with the requirements of professional, reg- will be in charge of the operations and
ulatory and also standardisation or accred- application of the QMS. This role can be
iting bodies. assumed by a person who has other du-
ties within the organisation.
A QMS should include: »» A documentation control procedure, to
»» A quality manual (QM) with a clearly de- keep under control all documents con-
fined “Policy of Quality” confirming the cerning the QMS, including their stor-
willingness to act appropriately to attain age, updating and distribution to staff.
objectives and describing the tools to »» Procedures for human resources devel-
be employed in order to achieve the opment, concerning the recruitment of
stated goals. The creation of a QM is the staff and their continuous professional
first step: here the organisation’s strat- development, where necessary.
egies are clearly identified, with state- »» The QM should also provide evidence
ment of goals and description of plans that internal audit reviews are being
to achieve those objectives. Main pro- periodically carried out, with the aim of
cesses, as well as supporting process- verifying that the procedures are being
es, are defined, including roles and re- followed, ensuring that they are con-
sponsibilities. When possible, objectives gruous with the established goals and
should be described in quantitative or establishing whether there are possibili-
qualitative terms with measurable indi- ties to improve them.
cators of processes and monitoring of
routine activities. The QMS standardises the processes
»» A departmental organisation chart with to guarantee consistency in providing
clear definition of roles and responsibili- high-level services to patients, referring
ties and reporting lines. physicians and other stakeholders in a

QUALIT Y CONTROL OF NUCLEAR MEDICINE 19


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 2

safe environment. The NMS management nuclear medicine practitioners in low- to


INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y
MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y

ensures the availability of necessary re- middle-income countries with a tool that
sources and information to support the would help them identify areas of weak-
operation and for monitoring of process- ness in their practices, raise awareness of
es. The management also ensures the international standards and eventually en-
effectiveness of the QMS through self-as- courage the implementation of an annual
sessments, data analysis, verification of ac- systematic audit process for the nuclear
tivities and management reviews. medicine practice as a whole. Following
various interactions among meeting par-
ticipants, a manual entitled Quality Man-
THE QUANUM PROJECT agement Audits in Nuclear Medicine Practic-
OF THE IAEA es (QUANUM) was published in 2008.
The IAEA (www.iaea.org) is one of the or- For the first time, a unique and holistic
ganisations of the United Nations (UN) programme covering all the disciplines
system and has the mission of supporting involved in the delivery of nuclear medi-
peaceful applications of nuclear technolo- cine was made available to practitioners
gies in the UN member states, which num- worldwide, and in the subsequent years
ber more than 170. Therefore, in addition to the QUANUM tool was successfully ap-
activities in many other fields, it has a long plied across the world. Lessons were
history of providing assistance in the spe- learned from the first IAEA expert mis-
cialty of nuclear medicine. In many of these sions, and throughout this period the spe-
countries, NMS are rare and often isolated cialty of nuclear medicine continued to
and practitioners have serious difficulties develop rapidly. Consequently, the IAEA
in exchanging experiences with their peers recognised that there was a need to up-
and even in accessing scientific journals. date the manual so that it would reflect
On the basis of its awareness of the very current and best practices in NMS. The
different levels of quality of practice and new edition was published in 2015 under
the need to raise them to internationally the title “Quality Management Audits in
recognised standard levels, the Nuclear Nuclear Medicine Practices, Second Edi-
Medicine Section of the IAEA planned tion. IAEA Human Health Series No. 33”. It
the preparation and the implementation is accessible at http://www-pub.iaea.org/
of a project called Quality Management books/IAEABooks/10714/Quality-Manage-
Audits in Nuclear Medicine (QUANUM). ment-Audits-in-Nuclear-Medicine-Practic-
The aim of the project was to provide es-Second-Edition.

20 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 2

The QUANUM methodology »» Radiation protection aspects relating to

INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y


MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y
The QUANUM methodology aims at defin- patients, staff, public and environment
ing a comprehensive auditing procedure »» Evaluation of the quality system
that covers all aspects of nuclear medicine »» Quality control of equipment
and is designed to be applicable in a vari- »» Computer system and data handling
ety of settings (bearing in mind the global »» Clinical services (assessment of diagnos-
discrepancies in availability of resources tic procedures and therapy)
and the great diversity in economic cir- »» Assessment of non-imaging procedures
cumstances). Adopting this methodology Assessment of hospital radiopharmacy
will allow the NMS to demonstrate the (three different possible levels) and labo-
level of efficiency, quality, safety and reli- ratories (hormones and tumour markers)
ability in delivering clinical services. The
overall quality depends on the inventory Internal and external audits and
of strengths and weaknesses together audit review procedure
with the critical appraisal of the ‘variables’ A quality audit process has to be patient
as observed in practice. The primary goal is oriented, systematic and outcome based.
to raise the standards of nuclear medicine It should include regular internal checking,
practices by fostering the introduction of assessment and review. It will further rein-
a culture of quality management into rou- force the system of documentation in a busy
tine daily work. clinical setting and should be carried out on
To this purpose, and taking into account a regular basis to ensure adequate quality of
the multidisciplinary aspects of nuclear practice in nuclear medicine (internal audits).
medicine, the QUANUM process is accom- For the internal audit, the head of the
plished by completing a comprehensive NMS selects the audit team leader, usu-
quality checklist which focusses on the ally the quality manager, who will be in
following key areas and is available at charge of the audit and selects the other
https://humanhealth.iaea.org/HHW/Nu- members. The audit team consists of staff
clearMedicine/QUANUM_2.0_Excel_Tool_ members with extensive knowledge of
and_QNUMED/index.html: the current procedures of the NMS. An
audit team may include the following
»» Strategies and policy members: nuclear medicine physician,
»» Administration and management medical physicist, radiopharmacist, nucle-
»» Human resources development ar medicine technologist/radiographer, ra-
»» Radiation regulation and safety diation safety officer, delegates of nuclear

QUALIT Y CONTROL OF NUCLEAR MEDICINE 21


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 2

Figure 1
INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y
MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y

Regular
Auditing Circle

Programmed
Internal audit Entrance Assessments
team is formed Briefing
External
audit
team is
Managerial review formed
Radiation safety
Clinical review
QA/QC euipment External
Radiopharmacy Organisations

Routine
YES
Nuclear Standard
Medicine met?
Activities
NO

Preventive/
Corrective Action NO

Need external
assessment? YES

Follow-up
YES Standard met? NO

The internal and external auditing cycle (adapted from: Quality Management Audits in Nuclear
Medicine Practices, Second Edition. IAEA Human Health Series No. 33. Vienna, 2014)

22 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 2

medicine administrative and nursing staff auditors should assure the staff that

INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y


MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y
and a representative from the hospital confidentiality (including patient con-
administration and QMS. It is advisable to fidentiality) will be respected, and that,
include independent persons from other if required by the host, an appropriate
services of the institution who represent document to this effect will be signed.
the end-user group (e.g. oncologists, car- Audit teams nominated by the IAEA
diologists, endocrinologists, nephrologists, have signed such a confidentiality doc-
administrators). ument before the audit.
Following the internal audit, an external
audit by international experts may be car- B. Actual assessment, which includes:
ried out if necessary. The flow chart of the a. A complete tour of the premises
cycle is shown in Fig. 1. b. Review and evaluation of procedures
For the external audit, the composition and all relevant documentation, in-
of the team is agreed among the parties: cluding review of treatment records
the criteria of multidisciplinarity, auditing c. Observation of practical implemen-
competencies and independence should tation of working procedures
be adopted as indicated above for the in- d. Staff interviews
ternal audit team. e. Meeting with the management of
Besides self-assessment based on QUA- the institution and/or associated ed-
NUM, the completion of the IAEA web- ucational institution
based nuclear medicine database (http:// f. Review of the previous audit (self-as-
nucmedicine.iaea.org/) is a prerequisite for sessment according to QUANUM)
IAEA external audits. g. Filling the audit checklists

Components of the audit and C. Operational information: As part of


responsibilities of the audit team their responsibilities, the audit team
As part of the procedure, standardised au- collect all management and opera-
dit practices include: tional information, including (but not
limited to):
A. Entrance briefing: The entrance brief- a. Updated copies of licenses/accredi-
ing is required to introduce the audit tation documents
team and to present the staff, finalise b. Organisational flow chart and func-
the agenda and discuss the objectives, tion descriptions
methods and details of the audit. The c. Samples of SOPs

QUALIT Y CONTROL OF NUCLEAR MEDICINE 23


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 2

d. Samples of study reports cerned activity might need to be


INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y
MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y

e. Copies of data regarding patients’ considered.


waiting times b. Major: issues impacting the capacity
f. Updated information on waiting lists of the NMS to adequately perform its
g. Copies of QC data for relevant equip- activities that should be addressed
ment and radiopharmaceuticals in a timely manner (e.g. 3–6 months).
h. Radiation safety records c. Minor: issues that may be the object
i. Copies of letters of appraisal/com- of optimisation, to be accomplished
plaints within a defined time period and
j. List of deviations and non-confor- re-evaluated during the next audit.
mances
k. Customer/stakeholder satisfaction Particularly, where a critical non-con-
surveys formance has been found, the action plan
should be sent to the audit team for fur-
D. Exit briefing: The preliminary feed- ther interaction. If appropriate, the service
back from the auditors is document- has the responsibility to notify the regula-
ed and presented to the staff of the tory authorities.
NMS and any other relevant key per-
son during an interactive exit briefing. E. Conclusion of the audit and report:
This includes time for questions and The audit report contains conclusions
an open discussion on all the findings identifying critical, major and minor pri-
of the auditors. The institution is then orities with clear and practical recom-
encouraged to give an immediate re- mendations.
sponse to the assessment. The steps
intended by the institution to react to Minimum requirements and confor-
the recommendations should be part mance and non-conformance scoring
of the action plan. With the aim of de- A scoring system has been designed to
fining priorities, non-conformances are evaluate the level of conformance as indi-
scored and then prioritised as: cated in Table 1, where the system is ex-
a. Critical: issues impacting the safe- plained using the example of documenta-
ty of patients, staff, caregivers and/ tion of clinical procedures.
or environment that should be
promptly addressed (within days or Items marked as NA will not be included
weeks). Discontinuation of the con- in the assessment of the final scores. The

24 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 2

Score Description Example

INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y


MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y
NA Not applicable When a particular activity is not in place (e.g. laboratory
determinations for tumour markers)
0 Absent or inappropriate No documents available
1 Planned or approximate Documentation is planned or exists as an informal draft
2 Partially conforming or partially A limited number of SOPs or most SOPs exist, but important
implemented parts are lacking
3 Largely conforming or largely Most SOPs are complete but some information is missing (e.g.
implemented reference to guidelines, dosimetry data) or documents are not
being updated regularly
4 Fully conforming or fully All documentation is in place and SOPs are complete and
implemented subjected to review

Table 1: Use of the scoring system for evaluation of the level of conformance: the example of
documentation of clinical procedures

items ‘Absent or inappropriate’, ‘Planned or internal audit is demanded before an


or approximate’ and ‘Partially conforming NMS engages in the external peer review
or partially implemented’ fall into the cat- process arranged by the IAEA through
egory of ‘Non-Conformance’, whereas the expert missions. However, an emerging
elements ‘Largely conforming or largely trend in clinical practice is the “compulsi-
implemented’ and ‘Fully conforming or fication” of the audit process for verifica-
fully implemented’ are classified as ‘Con- tion of mandatory quality operations. As
formance’. The scores are used to build a mentioned by Pascual (2016) “This pro-
radar plot to enable visual presentation of cess of compulsification of clinical audits
the overall results (Fig. 2). is not entirely new and exclusive to nu-
clear medicine practitioners worldwide.
Globalization and the neoliberal environ-
IMPLEMENTATION CHALLENGES ment have slowly influenced imposing
TO THE QUANUM PROCESS: clinical audits not only in the medical
FROM VOLUNTARY TO imaging field but in other medical fields
COMPULSORY as well.” As an example, in the Europe-
As previously mentioned, QUANUM is an Union, the EN ISO 8402 directive has
a voluntary process and self-evaluation mandated that clinical audits shall be im-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 25


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 2

Figure 2

1. Strategies
INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y
MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y

100
17. HT Markers 2. Adm&Man

75

16. RP Lev 3 3. Human Res


50

25
15. RP Lev 2 4. Radiat Reg

14. RP Lev 1 5. Patient R.Prot

12. Ther Serv 6. QA System

9. Clin Serv 7. Equip. QA/QC

8. IT Syst

Example of a radar plot

plemented and requires that countries in practices assessed and audited by col-
the European Union formally establish, leagues from other countries. Particularly
as a policy, that clinical audits are per- appreciated is the exchange of informa-
formed under published and regulated tion and advice received during the visits.
national guidelines. In almost all cases, the visit from inter-
Following this directive, the QUANUM national experts for an external audit and
programme, in slightly modified form, has the work done in preparing for this audit,
become the reference standard in some including filling of the checklist, has trig-
European countries, such as Belgium. gered the implementation of a quality
system; this is in itself represents an ex-
cellent outcome. Indeed, the QUANUM
CONCLUSIONS programme was conceived exactly for this
The QUANUM programme has been very purpose.
well received by counterparts, who wel- In the majority of the externally audit-
come the opportunity to have their daily ed practices, the audit has shown that

26 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 2

the quality of practice is already satisfac- a centre to achieve a remarkable improve-

INTE RNATI O NA L ATO M I C E NE R GY AGE N C Y


MA NAGE M E NT S YS TE M S: THE A PPR OAC H O F THE
INTE RNA L A ND E X TE RNA L AUD I TI NG O F Q UA L I T Y
tory. On average, almost no difference ment in its performance after 3 years.
was found between the scores achieved This case emphasises the importance of
on pre-mission self-assessments and follow-up missions: not only do such mis-
post-mission evaluation by the external sions help in assessing the outcome of the
QUANUM team. The average level was QUANUM programme but also the very
75%, which is indeed a good performance fact that they are going to happen em-
at international level. powers audited centres to take corrective
In one case, the implementation of a actions and pursue the full implementa-
QMS after the QUANUM mission enabled tion of their quality system.

SELECTED BIBLIOGRAPHY
» Applying radiation safety standards in nuclear med- » IAEA Human Health Series No. 1. Quality Assurance
icine. Jointly sponsored by the International Atomic for PET AND PET/CT systems. International Atomic
Energy Agency, International Labour Office, Interna- Energy Agency, Vienna, 2009.
tional Organization for Medical Physics, Pan American » IAEA Human Health Series no. 6. Quality assurance for
Health Organization, World Federation of Nuclear SPECT systems. International Atomic Energy Agency,
Medicine and Biology and World Health Organization. Vienna, 2009.
International Atomic Energy Agency, Vienna, 2005. » IAEA Safety Standards Series No. GSR Part 3. Radiation
» Dondi M, Kashyap R, Pascual T, Paez D, Nunez-Miller protection and safety of radiation sources: Interna-
R. Quality management in nuclear medicine for bet- tional basic safety standards. General safety require-
ter patient care: the IAEA program. Semin Nucl Med ments. Jointly sponsored by: EC; FAO; UN; IAEA; ILO;
2013;43:167–171. OECD; PAHO; UNEP; WHO; IAEA. International Atomic
» D  ondi M, Andreo P. Developing nuclear medicine in Energy Agency, Vienna, 2011.
developing countries: IAEA’s possible mission. Eur J » Operational guidance on hospital radiopharmacy: a
Nucl Med Mol Imaging 2006;33:514–515. safe and effective approach. International Atomic En-
» http://www.eanm.org/publications/guidelines/in- ergy Agency, Vienna, 2008.
dex.php?navId=37 » Pascual TN. Examining quality management audits
» https://humanhealth.iaea.org/HHW/NuclearMedicine/ in nuclear medicine practice as a lifelong learning
QUANUM_2.0_Excel_Tool_and_QNUMED/index.html process: opportunities and challenges to the nuclear
» http://nucmedicine.iaea.org/ medicine professional and beyond. Nucl Med Com-
» https://www.iaea.org/services/key-programmes/hu- mun 2016;37:785–791.
man-health-programme » Radiation protection and safety of radiation sources:
» http://www-pub.iaea.org/books/IAEABooks/10714/ international basic safety standards. Interim edition.
Quality-Management-Audits-in-Nuclear-Medi- General safety requirements. International Atomic
cine-Practices-Second-Edition. Energy Agency, Vienna, 2011.
» http://www.snmmi.org/ClinicalPractice/content.as- » Strategies for Clinical Implementation and Quality
px?ItemNumber=6414&navItemNumber=10790 Management of PET Tracers. International Atomic
» h ttp://www.who.int/management/quality/assur- Energy Agency, Vienna, 2009.
ance/QualityCare_B.Def.pdf?ua=1 » www.iaea.org

QUALIT Y CONTROL OF NUCLEAR MEDICINE 27


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
3
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
A N D S PE C T

CON T RO L
Q UAL I T Y
C AME R A
PL AN A R

SYSTE M

by Mario Medvedec
CHAPTER 3

INTRODUCTION
SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T

In the twenty-first century, healthcare systems are undergoing a significant


change in that the emphasis is now on ‘quantitative quality’ through
targeting of evidence-based outcomes together with patient safety and
satisfaction. In this context, every (medical) product should be subjected
to certain tests in order to confirm its quality, performance, efficacy, safety,
reliability, stability, etc.

Planar and SPECT scintillation cameras quarterly, half-yearly and annually: colli-
should also undergo different evaluation mator hole angulation, tilt-angle check,
steps during their life-cycle: factory tomographic spatial resolution, SPECT/
testing before shipment, acceptance CT alignment;
and reference testing after their on-site »» describe the Jaszczak phantom and un-
installation and before their clinical use, derstand and explain the procedure for
and routine periodic quality control (QC) testing of overall SPECT system perfor-
testing thereafter. mance;
After reading this chapter, readers should »» describe the most common artefacts and
be able to: explain how to proceed in such cases.
»» define acceptance testing on planar and
SPECT scintillation cameras, and under-
stand the reasons for performing such ACCEPTANCE TESTING
testing; Acceptance testing is a set of standard
»» define, understand and explain the procedures intended to verify that the im-
need for and the procedures and peri- aging equipment performs in accordance
odicity of recommended QC tests on with the manufacturer’s specifications and
planar and SPECT scintillation cameras: intended clinical use without any deficien-
daily: visual and physical inspections, cies or defects. The standard procedures
collimator touch pads and emergency and performance measurements usual-
stop buttons, energy window settings, ly employed in acceptance testing are
uniformity and sensitivity; those published by the National Electrical
weekly and monthly: centre of rotation, Manufacturers Association (NEMA), In-
bar phantom spatial resolution and lin- ternational Electrotechnical Commission
earity, high-count flood uniformity/sen- (IEC) or other international authorities. In
sitivity correction map; addition, some patient studies should be

30 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

performed as a part of the acceptance is committed to quality-related practices.

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
procedure. In this way, acceptance and For many years, QC testing was perceived
reference testing provides baseline per- to be the responsibility of the individual,
formance data to be referred to in future to be performed at the individual’s dis-
QC tests, and supports the final user’s de- cretion. In the 1990s, however, aspects of
cision to accept or reject a particular piece QC testing were incorporated within the
of equipment for safe routine clinical use. European legal framework by European
The warranty period for the imaging sys- Council Directives, with subsequent merg-
tem should begin only when the system ing and updating in 2013. Now, as stated
has passed acceptance testing by achiev- above, the requirement for QC is to be fully
ing at least the minimum acceptable re- implemented in European Union Mem-
sults under clinical conditions. It should be ber States through the introduction, by
noted not only that good clinical practice 6 February 2018, of laws, regulations and
for medical devices entails compliance administrative provisions necessary in or-
with professionally agreed and widely ac- der to comply with European Council Di-
cepted technical standards, but also that, rective 2013/59. In this context, QC means
at the time of writing, acceptance testing the set of operations (programming, co-
is becoming a legal requirement within ordinating, implementing) intended to
the European Union: Member States are maintain or to improve quality. It includes
to bring into force European Council Di- monitoring, evaluation and maintenance
rective 2013/59-compatible laws, regula- at the required levels of all characteristics
tions and administrative provisions by 6 of performance of equipment that can be
February 2018 and ensure that acceptance defined, measured and controlled. Mem-
testing is carried out before the first use of ber States shall ensure that performance
equipment for clinical purposes[1–15]. testing is carried out on a regular basis, and
after any maintenance procedure liable to
QUALITY CONTROL TESTING affect performance. Furthermore, Member
A basic requirement for the successful es- States shall ensure that the competent
tablishment of quality management sys- authority takes steps to ensure that the
tems, quality assurance programmes and necessary measures are taken by the un-
QC procedures is that the leadership of dertaking to improve inadequate or defec-
healthcare institutions, including nuclear tive performance of medical radiological
medicine departments, recognises and equipment in use, and also adopt specific
understands quality-related principles and criteria for the acceptability of equipment

QUALIT Y CONTROL OF NUCLEAR MEDICINE 31


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 3

in order to ascertain when appropriate Start-up


SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T

corrective action is necessary, including As the first operational check it is rec-


taking the equipment out of service. ommended to initialise the planar and
Decisions on the types and frequencies SPECT scintillation imaging system by
of QC tests should take into account overall performing the daily reset/start-up pro-
circumstances and resources in the individ- cedure, which should complete without
ual nuclear medicine department in rela- any warnings or error messages. It saves
tion to its instrumentation. QC procedures time if all power switches are not turned
for planar and SPECT scintillation cameras off, so that the detector system does not
usually include visual and physical inspec- need to warm up. Ideally, all clocks with-
tions and tests in relation to emergency in the nuclear medicine department,
stop buttons, collimator touch pads, back- including those of all imaging, count-
ground, photopeak and energy window ing and computer systems, should be
settings, intrinsic/extrinsic uniformity and synchronised and checked daily for the
sensitivity, energy and uniformity calibra- purpose of ensuring accurate activity ad-
tion, high-count flood uniformity, spatial ministration and quantitative analysis of
resolution and linearity, multiple window acquired data.
spatial registration, whole-body scan and Visual and physical inspection of planar
tomographic spatial resolution, pixel size, and SPECT scintillation cameras should de-
detector head tilt, centre of rotation cali- tect external mechanical or electrical de-
bration, overall SPECT system performance fects or damage, particularly with respect
and attenuation correction. QC tests are to the detector heads and collimators,
performed on a daily, weekly, monthly, which may compromise imaging quality
quarterly, half-yearly or yearly basis. Local and patient or staff safety. If any deficien-
regulations may require additional checks, cies are detected, the imaging equipment
but in every case QC testing should be as- should not be used until the problems
sociated with careful record keeping[1–15]. have been resolved. If the detector’s col-
limator and yoke have a touch pad that
halts all motion when contact is made, a
DAILY QC TESTS touch pad test should be performed on a
Daily QC tests are usually performed daily basis and after each change of col-
on the morning of each working day or, limators. An additional operational check
at least, the day on which the imaging should be performed on emergency stop
equipment is planned to be used. buttons, if available, which should light

32 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

and shut down all motor-driven system are then due to the patient only and not

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
movements when pressed [1–3, 6, 7, 11–13, 15]. the scintillation camera itself. QC testing
of intrinsic or extrinsic uniformity and sen-
Energy window sitivity of the imaging equipment should
Daily operational checks of energy win- be performed daily in order to check the
dow settings should be performed to system’s response to spatially uniform flux
confirm that all preset pulse height anal- of 99mTc or 57Co photons. Such a flood field
yser energy windows are properly centred uniformity may be tested qualitatively
around the energy photopeaks of the ra- by visual inspection or quantitatively by
dionuclides to be used with the scintilla- calculation of the integral and differen-
tion camera for clinical imaging purposes, tial image uniformity within the camera’s
thus suggesting correct energy calibration central field of view and useful field of
of the system [1–7, 11–13, 15]. view. If a daily intrinsic low-count unifor-
mity test is selected, then each collimator
Background should be checked weekly or monthly by
Operational check of the background count an extrinsic high-count uniformity test.
rates with or without collimators and with- Overall sensitivity of the detection system
in one or more energy windows should be is calculated as count rate per unit activ-
performed daily to detect radiation caused ity (cps/MBq) of the imaged radioactive
by possible radioactive contamination of source [1–9, 11–13, 15].
the scintillation camera, floor or walls, ra-
diation from some neighbouring unshield- CT checkup and quality
ed source or an excess of electronic noise. Daily X-ray CT QC testing of the SPECT/CT
Under constant measuring conditions the system should be performed according to
background count rates should be approx- the manufacturer’s recommended proce-
imately constant in all detector directions dures and medical physics expert advice.
used for clinical imaging [1–4, 11, 13]. For instance, it may be recommended to
perform daily CT checkup and CT quality
Uniformity and sensitivity procedures which automatically execute a
One of the basic assumptions in nuclear set of CT tube warm-up acquisitions, au-
medicine imaging is that the response of tomatic function checks, and different air
the imaging system to a uniform irradia- and water calibration steps for all available
tion is uniform within defined limits. Ob- voltage settings, in order to guarantee op-
served differences in activity distribution timum image quality [2, 11, 12, 15].

QUALIT Y CONTROL OF NUCLEAR MEDICINE 33


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 3

WEEKLY QC TESTS usually by simultaneously using three or


SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T

more point sources of similar activities of


Centre of rotation 99m
Tc and collecting a specified number of
Centre of rotation (COR) is by definition a image counts. The sources are placed in
single point around which detectors rotate the same plane in the air, on and off the
and which ideally should also be the cen- axis of rotation and the centre of the field
tre of the projections recorded by detec- of view. Each detector must be positioned
tors at all angles. In other words, the COR is parallel to the axis of rotation and must ac-
the point at which the axis of rotation and quire an image at 0° and 180°. Point sourc-
the perpendicular from the centre of the es are imaged at an even number of detec-
detector plane intercept. The transaxial tor angular positions equally distributed
alignment of acquired projection images over 360°. COR offsets are easily corrected
with the system’s mechanical centre of ro- if the equipment manufacturers provide
tation is critical for accurate generation of alignment measurements and software
tomographic images reconstructed from which calculates and includes corrections
acquired projection images. Similarly, for for COR variations in tomographic acquisi-
the multi-head SPECT system it is crucial tion and reconstruction processes [1–9, 11–15].
that the electronic centre of each angular
projection used in the image reconstruc- Spatial resolution and linearity
tion process is consistently aligned with Spatial resolution is the ability of the scin-
the centre of mechanical rotation. The tillation camera to accurately resolve spa-
COR offsets principally vary with collima- tially separated radioactive sources. The
tor type, detector orbit as a function of quantitative measure of spatial resolution
angle and radius, detector configuration is given in millimetres as the full width at
and image zoom factor. The alignments half maximum (FWHM) or full width at
should be checked in both the x- and the tenth maximum (FWTM) of the peak of the
y-axis and should stay within acceptable imaged point or line radioactive sources.
limits given in millimetres. Any error re- Spatial linearity is the ability of the scin-
lated to COR will lead to image distortion tillation camera to accurately determine
and loss of tomographic spatial resolution, the position of photons without displace-
or even the appearance of ring artefacts in ment in relation to the actual position
reconstructed point source images. where these photons enter the detector.
The COR QC test can be performed Spatial linearity is quantitatively expressed
weekly to monthly by using one, but more in millimetres related to the displacement

34 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

of the measured peak location from the mination of extrinsic spatial resolution and

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
best-fit peak location. spatial linearity, the bar phantom is placed
The bar phantom is a rectangular or directly on the collimated detector. The de-
circular sheet of plastic material in which tector is irradiated either by the flood source
a number of lead bars are embedded in a placed directly on top of the bar phantom
pattern of parallel stripes, usually arranged or by the point source placed several metres
into four quadrants of parallel bars. Lead away from the bar phantom. After collect-
bars of a given thickness are supposed ing the required number of counts with an
to stop radiation, whereas plastic stripes appropriately set-up camera for the imaged
are supposed to be transparent to radi- radionuclide, spatial resolution is expressed
ation. The width of the lead bars and the in terms of the quadrant pattern, with the
distance between two bars is equal with- narrowest stripes still resolvable on the ac-
in one quadrant, but different for each of quired images. When used for determina-
the four quadrants (e.g. 2, 2.5, 3 and 3.5 tion of intrinsic spatial resolution and spatial
mm or 3.2, 4.0, 4.8 and 6.4 mm). The incre- linearity, the bar phantom is placed directly
ment of bar separations and widths from on the uncollimated detector and irradiated
one to another quadrant should be small by the point source placed away from the
enough to provide reasonably accurate fi- bar phantom at a distance which is at least
nal semi-quantitative estimation of spatial five times the largest dimension of the de-
resolution, for instance FWHM ≤8 mm. tector. After collecting the required number
The bar phantom should match the spatial of counts with an appropriately set-up cam-
resolution of the scintillation camera in such era for the imaged radionuclide, the intrinsic
a way that at least one quadrant of stripes spatial resolution can be approximated as
cannot be fully resolved in the acquired bar FWHM = 1.75·B, where B is the width of the
phantom image. Bar phantoms can be used narrowest bars that the scintillation camera
weekly, biweekly or more infrequently in can still resolve. For the purpose of thor-
routine QC testing for visual determination ough evaluation of spatial resolution and
of extrinsic or intrinsic spatial resolution of linearity, the bar phantom can furthermore
the scintillation camera, whichever appears be rotated and inverted in such a way that
more convenient. The same high-count im- the quadrant of the bar phantom with the
ages of bar phantoms can also serve for the narrowest bars is imaged in each quadrant
evaluation of the scintillation camera spatial of the detector in each direction, i.e. with
linearity, which is normally measured by us- the imaged stripes parallel to the x- and
ing the slit phantom. When used for deter- the y-axis of the detector. Moreover, the bar

QUALIT Y CONTROL OF NUCLEAR MEDICINE 35


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 3

phantom can be imaged in air at a certain rection of its detection system after it has
SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T

distance from the detector equipped with been properly tuned and adjusted. This cor-
different parallel-hole collimators or in tis- rection is basically applied by multiplying
sue-equivalent material added between each particular pixel in acquired images by
the bar phantom and the collimator. a factor calculated as the ratio of the aver-
The purpose of checking spatial reso- age counts in the high-count flood image
lution and linearity is to detect gradual to the counts in the corresponding pixel in
long-term deterioration of spatial resolu- the high-count flood image. Intrinsic uni-
tion, and to display imaged linear objects formity correction corrects for non-unifor-
as exactly linear as possible, as compared mities in the detector only, whereas extrin-
with acceptance and reference measure- sic uniformity correction corrects for both
ments. Bar phantom image acquisition detector and collimator non-uniformities.
may or may not be required by imaging The total number of counts to be collected
equipment manufacturers, but is done at in high-count flood images depends upon
the discretion of the user [1–9, 11–13, 15]. the particular procedure and equipment
but is typically in the range of tens to one
High-count flood and uniformity or a few hundred million. Modern scintil-
A flood source is typically a rectangular lation cameras include on-line corrections
source of uniformly distributed radioactiv- for detection system variations in energy
ity, in the form of either a sealed 57Co sheet response, spatial linearity and spatial unifor-
source or a plastic phantom fillable with mity across the field of view of the scintilla-
a solution of the selected radionuclide. tion camera, but these should be periodi-
Detector irradiation can also be consid- cally verified and re-created if necessary to
ered uniform if a point radioactive source assure acceptable integral and differential
is placed away from the detector face at uniformity [1–9, 11–13, 15].
a distance five or more times greater than
the largest linear dimension of the detector.
Ideally, a uniform irradiation of a scintillation QUARTERLY AND
camera detector should produce an image ANNUAL QC TESTS
of homogeneously distributed radioactivi-
ty. The purpose of acquiring a high-count Collimator hole angulation
flood image is to verify uniformity within Collimator hole angulation is the geomet-
the field of view of the scintillation camera ric relationship of the actual collimator
and to provide a uniformity/sensitivity cor- holes and septa to the crystal face of the

36 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

planar or SPECT scintillation camera or to Tilt-angle check

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
the axis of rotation of the SPECT scintilla- The angle of tilt of the SPECT scintillation
tion camera. In the case of an ideally col- camera detector is the angle between
limated and rotating scintillation camera the detector plane and the axis of rota-
detector, all edges of the holes and septa in tion, measured along the axis of rota-
a parallel hole collimator should be parallel tion. Assuming that the axis of rotation
to each other and exactly perpendicular to is horizontal, the parallel hole collimators
the crystal and the axis of rotation. Any dif- should also be levelled exactly horizon-
ferences in these angles are referred to as tally. This adjustment is usually done by
collimator angulation error. Non-orthogo- careful use of a spirit level or an angle
nality in the x-direction (perpendicular to gauge. Head tilt should normally be 0°
the axis of rotation) actually represents a at the beginning of tomographic acqui-
centre of rotation offset, whereas non-or- sition of the correctly set up system, and
thogonality in the y-direction (parallel to should remain 0° for all angles of rotation.
the axis of rotation) represents scintillation The angle of tilt can be determined from
camera head tilt. Both non-orthogonalities summed projection images over 360° of
can deteriorate the quality of reconstruct- a radioactive point source placed off the
ed images. axis of rotation.
Quality control testing of the collimator If there is no head tilt, the amplitude of
hole angulation checks the septal align- the sinusoidal motion of such a radioac-
ment and angulation for all parallel hole tive point source in the y-direction will be
collimators used. It is performed by using equal to zero, showing a constant rather
a point radioactive source placed a few than a sinusoidal pattern in the y-direction
metres from the face of the collimator, in of the projection images and a flat rather
the centre of each parallel hole collima- than ellipsoidal shape (i.e. a short ellipse
tor and in four or more other positions axis equals zero, while a long ellipse axis
approximately halfway to the edge of the equals a distance 2r off the axis of rota-
field of view. Acquired images should be tion) when all projection data in the x- and
visually inspected and checked for any y-directions are taken together. If there
asymmetries, streaks and distortions. If the is a head tilt, it can be determined from
collimator holes and septa do not appear the length of the short ellipse axis and
appropriately aligned and angulated, the the known radius of the radioactive point
manufacturer should provide a new colli- source [1–6, 11, 13, 14].
mator [1–4, 6, 11, 13, 14].

QUALIT Y CONTROL OF NUCLEAR MEDICINE 37


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 3

Tomographic spatial resolution order to allow correction of possible me-


Tomographic spatial resolution is the abil- chanical offset between the SPECT and CT
ity of the scintillation camera to accurate- gantry positions. Calibration of the SPECT
ly resolve spatially separated radioactive and CT fields of view must be performed
sources in the images acquired in tomo- every time the SPECT gantry and CT gan-
graphic mode and reconstructed from the try are separated, and after each major
raw data using the filter backprojection service or upgrade. Thereafter, SPECT/CT
technique with a ramp filter. The main alignment should be tested periodically,
quantitative measure of tomographic with the frequency of testing depending
spatial resolution is the FWHM of the peak on the stability of the particular SPECT/CT
of the reconstructed radioactive point or system. This QC test is done by performing
line source, which is given in millimetres a SPECT/CT scan using the manufacturer’s
for the x-, y- and z-directions. It can be test objects or a phantom that contains
measured in air or in scatter medium, usu- radioactivity (and sometimes contrast
ally by using three point or line sources of agent, too) and is visible on both SPECT
similar activity, well centred in the field of and CT. Following SPECT/CT scanning, the
view and imaged over the range of 0° to image fusion software is used to complete
360° evenly covered by collimated scintil- the calibration process and ensure accu-
lation camera detector(s) rotating at a ra- rate alignment of the SPECT and CT fields
dius of about 15 cm and collecting a suffi- of view. The same procedure should be re-
cient number of counts in each step. After peated for each collimator set and detec-
data acquisition and image reconstruction tor configuration used in bimodal SPECT/
as outlined above, the corresponding to- CT imaging [2, 11, 12, 15].
mographic spatial resolution quantities
should be calculated according to the Jaszczak phantom and overall SPECT/
prescribed methods [1–8, 11–13, 15]. CT system performance
Different phantoms are designed for dif-
SPECT/CT alignment ferent purposes. Some total performance
One of the prerequisites for accurate over- phantoms are used to check the best
all registration, attenuation correction performance characteristics of the SPECT
and anatomical localisation by SPECT/CT imaging systems achievable by time-con-
hybrid imaging systems is determination suming and high-count non-clinical stud-
of the three-dimensional alignment vec- ies. Other phantoms are used to simulate
tor of the SPECT and CT fields of view in typical clinical conditions and to show

Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

how the imaging system performs in such SPECT data to be reconstructed with fil-

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
situations. Cylindrical phantoms, used in tered backprojection and a ramp filter. Dif-
QC testing of system performance, are ferent total performance phantoms and
plastic cylindrical tanks that have differ- studies are used in acceptance, reference
ent shapes, dimensions, inner structures, or QC testing at less frequent intervals to
inserts and other physical characteristics check for possible slow degradation in the
and are fillable with solutions containing performance characteristics of different
different radionuclides. Image quality pa- low- to ultra-high resolution SPECT sys-
rameters which may be evaluated during tems. Total performance phantom studies
system performance tests include tomo- are also useful to assess the performance
graphic uniformity, contrast, resolution, characteristics of SPECT system hardware
attenuation, noise, linearity and lesion and software after significant preventive
detectability. Two examples of commer- or corrective maintenance and upgrades,
cially available cylindrical tomographic or when conducting research activities [1–4,
phantoms are the Jaszczak phantom and 6, 11, 13, 15]
.
the Carlson phantom. A Data Spectrum
ECT (emission computed tomography) Artefacts
phantom, usually known as the Jaszczak Artefacts in biomedical imaging are
phantom, consists of a main plastic cir- misperceptions and misrepresentations of
cular or elliptical tank which contains few the imaged objects caused by the imaging
parts: a segment of homogeneous radio- equipment or the employed image acqui-
activity, a segment of non-radioactive sol- sition and processing techniques. Various
id spheres of different sizes and a segment sudden or gradually developing problems
of (non-)radioactive rods of different sizes. may become evident at any time, but
These segments are used after image re- taking appropriate preventive measures
construction in order to detect possible – planning, preparing and organising the
ring artefacts and distorted spheres and nuclear medicine facility, conditioning the
rods, to evaluate the contrast and spatial electrical power supply, performing ac-
resolution of objects of a known size and ceptance, reference and QC testing, carry-
to calculate the linear attenuation coeffi- ing out regular maintenance, conducting
cient for attenuation correction if related overall clinical practice competently, etc.
software is available. The phantoms are – decreases the likelihood of artefacts.
typically filled with 99mTc and are imaged The most common imaging artefacts
for tens of minutes to acquire high-count are: full or partial ring and bull’s eye arte-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 39


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 3

facts; different forms of distortion; blur- requirement before proceeding with rou-
SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T

rings; variations in intensity; lines and tine clinical practice. If full correction of
stripes or other discontinuities; local spots detected problems is possible locally and
of higher or lower intensity; artefacts due in a timely way, this should be implement-
to patient movements or metallic objects; ed, with subsequent successful repetition
artefacts caused by energy spectrum dis- of QC procedures before continuation
tortions and energy resolution degrada- with daily clinical practice. If correction is
tion; and artefacts due to decreases in de- not possible, limited use of the imaging
tector sensitivity, poor spatial uniformity, system should be considered. For this pur-
poor contrast and spatial resolution, inad- pose a call for service is usually made. De-
equate image acquisition and processing pending on the outcome of this call, lim-
etc. All of these artefacts are extensively ited clinical practice may be continued. If
illustrated elsewhere. immediate service is needed, it should be
According to the general troubleshoot- followed by successful QC testing before
ing flowchart (including but not limited again proceeding with full routine clinical
to artefacts), the acquisition of QC results practice [1, 3, 4, 6, 13].
within acceptable limits is a necessary

40 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 3

SYSTE M Q UA L I T Y CO NTR O L
PL A NA R C AM E RA A ND S PE C T
REFERENCES

1. International Atomic Energy Agency. Nuclear medi- Guide. Vienna: EANM Technologist Committee and
cine resources manual. Vienna: International Atomic Technologist Education Subcommittee; 2007.
Energy Agency; 2006. 10. Council Directive 2013/59/Euroatom of 5 December
2. EANM Physics Committee, Busemann Sokole E, 2013 laying down basic safety standards for protec-
Płachcínska A, Britten A. Acceptance testing for nu- tion against the dangers arising from exposure to
clear medicine instrumentation. Eur J Nucl Med Mol ionising radiation, and repealing Directives 89/618/
Imaging 2010;37:672–681. Euratom, 90/641/Euratom, 96/29/Euratom, 97/43/
3. International Atomic Energy Agency. Quality assur- Euratom and 2003/122/Euratom. Official Journal of
ance for SPECT systems. Human Health Series No. 6. the European Communities 17.1.2014.; No. L-13: 1-73.
Vienna: International Atomic Energy Agency; 2009. 11. EANM Physics Committee, Busemann Sokole E, Płach-
4. Bailey DL, Humm JL, Todd-Pokropek A, van Aswegen cínska A, Britten A, EANM Working Group on Nuclear
A. Nuclear medicine physics - a handbook for teach- Medicine Instrumentation Quality Control, Lyra Geor-
ers and students. Vienna: International Atomic Ener- gosopoulou M, et al. Routine quality control recom-
gy Agency; 2014. mendations for nuclear medicine instrumentation.
5. Cherry SR, Sorensen JA, Phelps ME. Physics in nuclear Eur J Nucl Med Mol Imaging 2010;37:662–671.
medicine. 4th ed. Philadelphia: Saunders/Elsevier; 2012. 12. Siemens Medical Solutions. Operating Instructions
6. International Atomic Energy Agency. Quality control Symbia System T16/T6/T2/T Series. Erlangen: Sie-
of nuclear medicine instruments 1991. Technical mens Medical Solutions; 2009.
document 602. Vienna: International Atomic Energy 13. International Atomic Energy Agency. Quality con-
Agency; 1991. trol atlas for scintillation camera systems. Vienna:
7. Murphy PH. Acceptance testing and quality control International Atomic Energy Agency; 2003.
of gamma cameras, including SPECT. J Nucl Med 14. Busemann-Sokole E. Measurement of collimator
1987;28:1221–1227. hole angulation and camera head tilt for slant and
8. National Electrical Manufacturers Association. NEMA parallel hole collimators used in SPECT. J Nucl Med
NU 1. Performance measurement of scintillation 1987;28:1592–1598.
cameras. Rosslyn, VA: National Electrical Manufactur- 15. Jones DW, Hogg P, Seeram E. Practical SPECT/CT in
ers Association; 2001, 2007, 2013. nuclear medicine. London: Springer; 2013.
9. European Association of Nuclear Medicine. Best
Practice in Nuclear Medicine. Part 2. A Technologist’s

QUALIT Y CONTROL OF NUCLEAR MEDICINE 41


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
4
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
CON T RO L
Q UAL I T Y
SYSTE M
PE T

by Claudiu Peștean
CHAPTER 4

INTRODUCTION
Q UA L I T Y CO NTR O L
PE T SYS TE M

Since the first prototype of a PET/CT scanner was installed in Pittsburg


by Townsend in 1998, this technology has been developed in a way that
even the most optimistic specialists could never have anticipated.

This trend is due not only to the contribu- stand the basic principles of acquisition
tion that PET/CT has made to medicine and reconstruction. Like all diagnostic
by virtue of its combined morphological procedures in nuclear medicine, PET
and functional perspective, but also to its technology uses the ability of scintilla-
wide application in various pathologies, tor detectors to record signals produced
including above all oncologic diseases, by the interaction of gamma photons
against a background of increasing prev- with the detector crystals. PET is a tech-
alence. In relation to the PET part of the nique based on the administration of a
PET/CT technology, it is necessary to un- radiotracer containing positron-emitting
derline the importance of quantification isotopes. With respect to QA/QC, it is
of radiotracer uptake in pathological tis- not of great importance and is beyond
sues, commonly achieved by calculation the purpose of this chapter to detail as-
of the standardised uptake value (SUV) pects relating to the radiopharmaceutical
for fluorine-18 fluorodeoxyglucose (FDG), properties of the radiotracers used in PET.
which reflects the metabolic rate of the This information may be found in differ-
tissue and, implicitly, the aggressivity of ent papers, including the technologist’s
the tumour. Importantly, PET/CT has also guide published by EANM Technologist
started to be used for radiation therapy Committee on the subject: Principles and
planning. As PET/CT has developed into Practice of PET/CT. Part 2 – A technologist’s
an indispensable imaging procedure in di- guide[2]. The QC and QA for PET are direct-
agnostic and therapeutic strategies, it has ly related to the nuclear physics exploited
become ever more important that reliabil- in this technique and to the technology
ity of the produced images is ensured. This used to create PET/CT images. After ad-
is achieved by the quality control (QC) and ministration of the radiopharmaceutical
quality assurance (QA) procedures em- and following allowance for the uptake
ployed to ensure correct scanner set-up time, the patient is placed in the scan-
and operation[1]. ner and the data are acquired. First, the
To understand the concepts of QC and CT data are recorded based on measure-
QA tests for PET, it is essential to under- ment of the attenuation in the tissues of

44 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 4

the X-rays emitted by an X-ray tube. For data are recorded after the CT acquisition.

Q UA L I T Y CO NTR O L
PE T SYS TE M
this purpose, the patient is placed in the The table moves the patient inside the PET
gantry and X-rays are emitted from dif- detector. The PET detector (consisting in
ferent angles, traversing the patient and a multitude of scintillation crystals organ-
being recorded on the opposite side by ised in blocks and modules distributed on
a solid state detector. The CT image is a ring with several rows) detects, very close
produced on the basis of the recorded at- to the point of release, the annihilation
tenuation values from all the projections, photons that are produced when the posi-
and using specific reconstruction and fil- trons released by the PET isotopes interact
tration methods. with electrons in the matter. The two pho-
tons produced by each annihilation trav-
The most important aspect of the CT el through matter in opposite directions
image in PET/CT, besides the anatomical and interact with the detector crystals.
information that it provides, is its utility in If these photons are recorded by the de-
correcting the attenuation within the pa- tector within the same time window, they
tient’s tissues of the gamma photons used are called coincident. If two photons are
to produce the PET image. This process, recorded by two crystals, the system will
termed CT attenuation correction (CTAC), assume that the annihilation process has
is necessary to minimise introduction of been produced somewhere on the line
bias into the PET image, which should between these crystals, the so-called line
precisely reflect the biodistribution of the of response (LOR). Faster detectors may
radiotracer. Based on the attenuation val- calculate the time interval between the ar-
ues calculated with the CT technique, also rivals of the two annihilation photons and
referred to as CT numbers or Hounsfield can more accurately estimate the origin of
units, the tissue densities are calculated; the annihilation process and the presence
once the density map of the investigated of the radiotracer; as a consequence, this
tissues is known, the attenuation of gam- so-called time-of-flight (TOF) technique
ma photons detected by PET is calculat- offers superior spatial resolution. LOR or
ed. An attenuation map is elaborated and TOF information is recorded on sinograms
applied to the PET non-corrected image and, using specific reconstruction meth-
based on the acquired data and the CTAC ods, the PET image is obtained[3]. In order
PET image is produced. This attenuation to gain a fuller understanding of the basic
correction process is performed after the principles of PET and PET/CT, the reader
PET data have been acquired. These PET can access the EANM website to consult

QUALIT Y CONTROL OF NUCLEAR MEDICINE 45


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 4

the technologist’s guide produced for this ing any relevant differences in relation to
Q UA L I T Y CO NTR O L
PE T SYS TE M

purpose: Principles and Practice of PET/CT. the various PET scanners on the market.
Part 1 – A technologist’s guide{4]. While we shall describe the principles of
Before any PET image is analysed, it the procedures, we shall not detail the
is mandatory to guarantee that quality practical aspects.
standards for the equipment are met. This
is the aim of the QC and QA procedures
which we shall describe in this chapter. ACCEPTANCE TESTING FOR PET
We shall try to define the concepts of QC Since PET and PET/CT techniques were
and QA, starting with acceptance test- first introduced into clinical medical im-
ing procedures and then describing the aging, the need for standardisation has
basic QC tests that are required for PET grown tremendously as the sensitivi-
equipment. There are various vendors on ty and the applicability of the method
the market who provide several types of have increased. In response to this trend,
PET/CT scanner, designed according to American and European associations
the same principles but with slight differ- and bodies, such as the Society of Nu-
ences in both the technologies used for clear Medicine, the National Electrical
detection or reconstruction and the QC Manufacturers Association (NEMA) and
and QA procedures. In modern PET scan- the International Electrotechnical Com-
ners, the QC and QA procedures are in- mission, started to establish and develop
creasingly automatic or semi-automatic, sets of standards for positron emission
which excludes potential human errors tomographs. The benefit is that now all
to the greatest possible extent. Never- the manufacturers can specify the per-
theless, it is important to understand the formance of their equipment using the
principles of the technique and to be same set of parameters and standards,
able to recognise the artefacts than can which can be measured and verified af-
be produced if QC or QA fails. It is also ter installation. NEMA standards have
necessary to know the measures to be been imposed over time and now all
taken in this circumstance in order to the vendors relate the performance of
eliminate errors from the diagnostic in- their equipment to these standards. Fol-
formation and thereby benefit patients. lowing the elaboration of the first NEMA
Here we aim to review the QC and QA NU2-1994 standard, the need for further
procedures that are of the greatest im- standardisation gradually grew, and sev-
portance in the practice of PET, identify- eral updates have since been produced:

46 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 4

NEMA NU2-2001, NEMA NU2-2007 and analysed by calculating the count rate for

Q UA L I T Y CO NTR O L
PE T SYS TE M
NEMA NU2-2012[5]. each slice and for each sleeve using a de-
After installation of the equipment, ac- cay correction formula and an attenuation
ceptance testing procedures need to be correction for each sleeve[6].
performed in order to ensure that the lev-
els of performance specified by the man- Spatial resolution
ufacturer are met. The acceptance tests The spatial resolution should be assessed
and the parameters verified usually refer during the acceptance testing, at the end
to the NEMA standards. If the acceptance of the warranty and whenever there is sus-
tests are failed, the system will need to be picion that the performance of the equip-
corrected. It is also important to perform ment has suffered importance changes.
a new set of acceptance tests prior to the The principle of the test consists in image
end of the warranty. acquisition using three point sources (with
Depending on the type of scanner, var- an activity of approx. 1 MBq) that have a di-
ious tests will form part of the acceptance ameter less than 1 mm and are suspended
testing procedures and these are dis- in the air to avoid any effect of scattered
cussed below. radiation; these point sources are placed 1
cm vertically from the centre of the FOV,
Sensitivity 10 cm vertically from the centre of the FOV
The sensitivity is determined during the and 10 cm horizontally from the centre of
acceptance tests, at the end of the war- the FOV. Two acquisitions are performed:
ranty and when the system suffers import- one in the centre of the axial FOV and one
ant changes in its performance. For this at a quarter of the axial FOV. The spatial
test, a 7-cm line source of 5 MBq is used, resolution is derived on the basis of the
placed in a phantom with five aluminium full-width at half-maximum (FWHM) of
sleeves with a wall thickness of 1.25 mm. the response function calculated for each
Acquisitions are performed starting with source[6].
the smallest sleeve and with the source in
the centre of the field of view (FOV) (or at Energy resolution
10 cm offset from the centre if variations A test relevant for scanners that use only
in sensitivity within the FOV are being singles-based attenuation correction and
checked); for subsequent acquisitions the calibration is performed by acquiring more
attenuation is increased by adding the than 10 kcts after placing a point source
other sleeves one by one. Sensitivity is of 18F with an activity of approx. 1 MBq in

QUALIT Y CONTROL OF NUCLEAR MEDICINE 47


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 4

the centre of the FOV. Usually, the energy of clinical imaging conditions. This can
Q UA L I T Y CO NTR O L
PE T SYS TE M

resolution is analysed using the vendor’s be achieved by acquiring the data using
procedure for energy testing[6]. two types of phantom. The first is a body
phantom with spheres, a so-called image
Scatter fraction, count losses and quality phantom, which consists in three
random measurements compartments: a body compartment, six
This test is very important since the scat- hollow spheres with diameters of 1, 1.3,
ter and random photons may affect im- 1.7, 2.2, 2.8 and 3.7 cm and a cylindrical
age quality. The test reflects the count rate central insert filled with material that sim-
performance of the scanner, expressed as ulates lung tissue’s attenuation properties.
noise equivalent count rate. To test this, This body compartment of this phantom
a polyethylene cylindrical phantom of 70 is filled with 18F solution. A sphere-to-
cm length and 200 mm diameter is used; background ratio of 4:1 is used for the
the cylinder is traversed from end to end first study for the four smallest spheres
by a hole that is parallel to the axis and at (the other two are filled with cold water)
a radial offset of 45 mm, to contain the line and a second study is performed using a
source. From the acquisition data, prompt ratio of 8:1. The second phantom is used
and random sinograms are obtained and, to mimic the radioactivity present outside
based on Poisson statistics, the scatter of the FOV in clinical situations. This is a cy-
fraction is calculated using the vendor’s lindrical phantom similar to the phantom
specifications and applying complex sta- described for the determination of scatter
tistical calculations[6]. fraction, with a concentration activity in
the line source equal to the concentration
Image quality and accuracy of of the background compartment of the
attenuation and scatter correction and image quality phantom.
quantitation One transverse slice centered on the
It is very difficult to simulate the distri- spheres is used for image analysis. ROIs are
bution of the radiopharmaceutical in drawn on each sphere and in addition 12
the patient’s body and to assess the im- ROIs are drawn in the background com-
age quality using a phantom. The image partment at a distance of 15 mm from the
quality, the attenuation and scatter cor- edge. The per cent contrast is calculated
rection and even the quantitation on the for each sphere (hot and cold) in relation
PET image can be evaluated reproducing to the background average counts mea-
as closely as possible the particularities sured on the previously drawn ROIs.

48 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 4

The accuracy of the attenuation and the tolerance values established by the

Q UA L I T Y CO NTR O L
PE T SYS TE M
scatter correction is estimated using ROIs manufacturer.
drawn on the lung compartment. The
counts measured in the lung insert of PET normalisation
the phantom are related to the counts Even when non-uniformities are within
measured in the ROIs placed in the back- the tolerance level, it is important that the
ground compartment as described above. reconstructed image is normalised to en-
Using the option provided by the soft- sure optimal uniformity. For this purpose,
ware, the concentration of the radioactivi- a rotating 68Ge or a uniform 68Ge cylindrical
ty is displayed and compared with the true phantom may be used. Calibration data
radioactivity concentration, known from are obtained which are used to normalise
the time of preparation of the 18F solution the acquired data in the clinical mode.
when filling the phantom[6].
Radioactivity concentration calibration
Coincidence timing resolution For this test, a cylindrical fillable 18F phan-
Coincidence timing resolution is estimat- tom is used, performing an acquisition
ed by histogramming the differences in with consistent data statistics. It is also pos-
the arrival times of annihilation photons sible to evaluate the SUV accuracy by scan-
after acquiring data using a point source ning a phantom with a known activity us-
placed in the centre of the FOV within a ing a multibed protocol. If the information
scattering material[6]. relating to the tracer and phantom (e.g.
activity, calibration time, phantom weight,
Uniformity of the reconstructed image phantom volume) is recorded as part of
This test is a measure of the system re- the patient data, then the measured SUV
sponse to a homogeneous radioactivity should be 1. Practically, this calibration test
distribution. A cylindrical hollow phan- evaluates the ability of the system to cor-
tom filled with 18F solution or, alterna- rectly measure the SUV.
tively, a cylindrical 68Ge/68Ga cylindrical
phantom is used. On the reconstructed The above represent a minimum set of
slices, the non-uniformities are assessed tests that need to be performed in relation
by determining the counts on square to the PET system during the acceptance
ROIs of 1×1 cm. The evaluation is per- testing for a PET/CT scanner. It is recom-
formed on each slice and the maximum mended that the tests are repeated before
value of non-uniformity should be within the end of the warranty to ensure that the

QUALIT Y CONTROL OF NUCLEAR MEDICINE 49


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 4

system will conform to the original accep- daily QC procedures for the main types of
Q UA L I T Y CO NTR O L
PE T SYS TE M

tance testing and the manufacturer’s per- PET/CT scanner.


formance specifications at the start of the For the PET/CT scanners produced by
subsequent cycle of use. Philips, the daily QC test is performed using a
point source placed in the centre of the FOV.
The performance of the system is evaluated
ROUTINE QC FOR PET on the basis of the sonograms obtained[5].
Even though acceptance testing remains For the PET/CT scanners produced by
the most important element of equip- Siemens, the daily QC is performed us-
ment testing, it is the responsibility of the ing a cylindrical 68Ge phantom placed in
operator (nuclear medicine technologist the centre of the FOV. The obtained sino-
and/or physicist) to ensure that the equip- grams are evaluated to identify eventual
ment continues to work optimally, thereby abnormalities[5].
ensuring that the first premise for acquisi- General Electric designed a calibration
tion of good diagnostic information is met. system based on a rod source of 68Ge placed
In order to respect the principles of behind the scanner, which is shielded when
radiation protection, the routine QC pro- not in use. It is extended in the FOV and con-
cedures must be successfully completed stantly rotated in front of the crystals to en-
before any radiopharmaceutical is admin- sure a constant exposure of all the crystals
istered to the patient[7]. (Fig. 1). The system evaluates the function-
The daily QC and QA process starts with ality of the detector but also estimates the
the simplest procedures like visual inspec- lifetime of the calibration source[5].
tion of the scanner, synchronisation of the The daily QC tests evaluate the constan-
clocks in the department and inspection cy of the detector’s functionality and may
of the handling systems for patients. The detect malfunctions such as failure of crys-
main daily QC steps, however, relate to tals or modules. The principle of the daily
the working parameters of the detector, QC is to correlate the data obtained from
namely coincidence detection, singles, the sinograms with the levels established
dead time, timing resolution and energy during the calibration of the system. If the
resolution. The procedure is automatic or differences are beyond the tolerance levels
semi-automatic, using a phantom, a point established by the manufacturer, recalibra-
source or a rod source, and depends on tion or maintenance should be performed.
the manufacturer’s recommendations. The frequency of QC tests should be
Therefore here we shall briefly present the in accordance with the manufacturer’s

50 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 4

Figure 1

Q UA L I T Y CO NTR O L
PE T SYS TE M
Calibration system based on a 68Ge rod source placed in the FOV of a GE PET/CT scanner
Discovery series, Optima 560 (courtesy of CT-Clinic Positron Emission Tomography)

recommendations and the national reg- for acquisition of accurate diagnostic in-
ulations established for nuclear medicine formation and also for reproducibility of
practice. These recommendations and scanning conditions, which in turn is nec-
regulations should be strictly respect- essary to allow correct comparison and
ed to ensure proper functionality of the evaluation of images during the diagnos-
equipment; the latter is a prerequisite tic sequences.

REFERENCES
1. International Atomic Energy Agency. Standard Op- tice of PET/CT. Part 1. A technologist’s guide. Vienna:
erating Procedures for PET/CT: A Practical Approach European Association of Nuclear Medicine; 2010.
for Use in Adult Oncology. Vienna: International 5. International Atomic Energy Agency. PET/CT atlas
Atomic Energy Agency; 2013. on quality control and image artefacts. Vienna: Inter-
2. Testanera G, van den Broek W, eds. Principles and national Atomic Energy Agency; 2014.
practice of PET/CT. Part 2. A technologist’s guide. 6. International Atomic Energy Agency. Quality assur-
European Association of Nuclear Medicine; 2011. ance for PET and PET/CT systems. Vienna: Interna-
3. Saha GB. Basics of PET imaging - physics, chemistry tional Atomic Energy Agency; 2009.
and regulations [internet]. New York: Springer Sci- 7. Busemann Sokole E, Płachcínska A, Britten A, Lyra
ence+Business Media Inc.; 2005. Available from: http:// Georgosopoulou M, Tindale W, Klett R. Routine quality
www.springerlink.com/index/10.1007/b138655 control recommendations for nuclear medicine instru-
4. Hogg P, Testanera G, eds. EANM principles and prac- mentation. Eur J Nucl Med Mol Imaging 2010;37:662–71.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 51


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
5
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
CON T RO L
Q UAL I T Y
SYSTE M
CT

by Urban Zdešar and


Dean Pekarovič
CHAPTER 5

INTRODUCTION
Q UA L I T Y CO NTR O L
C T SYS TE M

CT scanners produce cross-sectional images of high radiographic


contrast, which is especially important when diagnosing disease in
organs or tissues in which object contrast is intrinsically low. Modern CT
scanners are also very fast, which makes CT the technique of choice for a
growing number of examinations.

CT scanners are used not only in diag- by the dose to the patient – a higher dose
nostic radiology but also in radiotherapy usually means better image quality, but
and nuclear medicine. In radiotherapy sole attention to the latter can cause doses
departments, they are used for treatment to become inappropriately high. An effec-
planning and in nuclear medicine depart- tive quality assurance (QA) system should
ments mostly as a part of SPECT/CT or therefore be implemented to assure ad-
PET/CT systems. In spite of rapid techno- equate performance and optimised use
logical development, a drawback of CT is of the CT system. QA begins already with
still the relatively high patient radiation the specifications of the equipment being
exposure. Patient doses per procedure are purchased. When equipment is installed,
rarely below 1 mSv and can easily reach an acceptance test is performed; this is
several tens of mSv. In some procedures best done on a collaborative basis by the
(e.g. perfusion imaging), doses can even service engineer, medical physicist and
reach the threshold for deterministic ef- radiographer who will be responsible for
fects [1]. An exception to this can be nu- the scanner. At acceptance testing, the CT
clear medicine, if CT images are used for scanner and its operation are tested to es-
attenuation correction only and are there- tablish whether the scanner functions as
fore not of diagnostic quality. designed, whether it complies with regu-
latory requirements and whether it meets
The performance of the CT scanner and the requirements set in the purchase
its components has a significant influence specification. Acceptance testing is also
on the outcome (image) and the patient useful in establishing baselines for future
exposure. CT image quality is influenced QC testing.

54 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

Quality control (QC) is an important teria. However, one should not forget

Q UA L I T Y CO NTR O L
C T SYS TE M
element of the QA system. It is intended that proper functioning of equipment is
to verify that CT scanner performance is merely a necessary condition for the per-
consistent with the predetermined re- formance of optimised diagnostic pro-
quirements. The QC programme should cedures. Appropriate use (protocols) of
therefore include a list of tested parame- procedures is also of vast importance and
ters, a description of testing methods and should therefore also be monitored. This
the tolerances used to check that quality can be done using different techniques,
requirements are met. It should also in- for example clinical audits and dose mon-
clude anticipated actions to adjust or cor- itoring (tracking).
rect performance if the tolerance limits are
exceeded.
TESTS TO BE CONDUCTED BY
There are many books and guidelines MEDICAL PHYSICISTS
dealing with QA and QC in CT (e.g. [2–5]). Tests described in this section are divided
QC tests are usually divided into basic into three types: (a) geometry, positioning
and optional or according to the recom- and movement tests, (b) image quality
mended frequency. Some QC tests need tests and (c) CT dosimetry tests. The basic
to be performed frequently (daily, weekly equipment needed to perform these tests
or monthly), so it is recommended that is an image quality phantom and a CT
these tests be performed by radiographers dosimetry set (dosimetry phantom, ioni-
working on a CT scanner. More compre- sation chamber and electrometer). As test
hensive testing is done semi-annually or procedures are at least partially depen-
annually and is usually performed by med- dent on the available equipment, the user
ical physicists. should read the instructions provided with
the equipment. We have tried to describe
The focus of this guide is technical the test procedures listed below as gener-
QC, which refers to procedures designed ally as possible, so that the information is
to show that the CT equipment and not specific to particular equipment.
its performance conform to the set cri-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 55


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5

Geometry, positioning
and movement tests
Q UA L I T Y CO NTR O L
C T SYS TE M

Alignment of lasers with the scan plane


Purpose. Alignment of the scan localisation laser and the scan plane is import-
ant because basic patient positioning is done using alignment lasers.
Testing can be performed using narrow film strip (e.g. GAFCHROMIC®
CT strips) or a phantom with a suitable module.

Description. When using radiochromic film strip:


» Position radiochromic film strip on the table or on the phantom and
centre it vertically and horizontally using lasers.
» Mark laser position on the film or (if the film has its own markings)
centre it at a certain position (Fig. 1)
» Scan using narrow collimation.
» Observe/measure the difference of scan plane (blackening of film)
and marked laser indicator [cf. Fig. 1, where one can observe that the
beam (beam blackening) was slightly to the right of the centre for all
three positions at which CT beams were centred: -30 mm, 0 mm and
+30 mm].
CT test phantoms usually include a module with a narrow absorb-
er (wire) or two sets of opposing ramps, the centres of which pass
through the actual axes of the module:
» Centre the test device in the external light field parallel to the tomo-
graphic plane.
» Move the test device into the tomographic plane and check that the
test device is aligned with the internal field light.
» Scan the test object with a narrow axial scan over the range of ±3
mm about the centre of the light field.
» Observe images and evaluate alignment.

Tolerances. Laser indication of scan plane should be within ±2 mm.

56 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

Q UA L I T Y CO NTR O L
C T SYS TE M
Accuracy of indicated table movements
Purpose. The test is performed to check that the table movements are consis-
tent with the planned movements.

Description. The movement test should be performed with a weighted table (e.g.
with a person sitting on it):
» Attach a tape measure to the table in such a way that the central laser
crosses it at a certain number.
» Set the table position to 0.
» Move the table for a certain distance (e.g. 30 cm) and see whether the
laser crosses the tape measure at the expected distance.
» Return the table to the previous position and check it with the laser
line on the tape measure.

Tolerances. Movement accuracy should be within 2 mm.

Irradiated beam width


Purpose. To check that the irradiated slice width is within the specifications.

Description. The test can be performed together with the test of alignment of lasers
with the scan plane (see above). Proceed as follows:
» Position the film strip on the table or on the phantom and centre it
vertically and horizontally using lasers.
» Using the central laser, position the film so that the laser crosses the
film at 0 mm.
» Scan the film strip using the axial head protocol and a certain thickness.
» Move the table for a certain distance (e.g. 30 mm) and scan with a
different thickness.
» Observe the image (see Fig. 1 for example).
» If strip film is scanned, radiation profiles can be deduced and the slice
width accurately measured (Fig. 2).

Tolerances. Beam widths within manufacturer’s specifications.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 57


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5

Geometry, positioning
and movement tests
Q UA L I T Y CO NTR O L
C T SYS TE M

Image slice width


Purpose. To check that the image slice width is as selected within the protocol.

Equipment. Phantom with a suitable module (e.g. CATPHAN or PRO-CT phantom) –


usually there is an insert with wire ramps at a certain angle (α). When wire
length on the image is measured, slice width can be calculated (Fig. 3).

Description. Proceed as follows:


» Place the phantom with a suitable module on a stand on a CT table and
set the gantry to the vertical position.
» Centre it horizontally and vertically using lasers.
» Use routine head protocol.
» Set the scanning plane on the middle of the module and perform an
axial scan.
» Determine the CT number maximum (CTmax) for wire ramps. Restrict the
window to 1 or the lowest selectable value and move the window cen-
tre to the point at which the ramp image almost disappears. This level
value will be the maximum for the ramp.
» Determine the background CT number (CTb) using the average HU
value near the ramps.
» Calculate the CT number corresponding to half the maximum height (CThalf):
(CTmax – CTb)
CThalf = + CTb
2
» Set the window centre to the calculated CThalf and the window width
to 1.
» Measure the length of the ramp to obtain full width at half maximum
(FWHM).
» Calculate the image slice width T: T = FWHM · tan(α)

Tolerances. Deviation of image slice width from nominal value:


+ 0.5 mm for beams <2 mm;
± 1 mm for slices >2 mm.

58 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

Q UA L I T Y CO NTR O L
C T SYS TE M
Figure 1

Film strip irradiated at three different positions determined by central laser


(-30 mm, 0 mm and 30 mm) using three different beam widths

Figure 2

Profiles of beam widths 4×4 mm, 4×2 mm and 4×0.5 mm)

Figure 3

Calculation of image slice width (T):


T = FWHM • tan(α)

QUALIT Y CONTROL OF NUCLEAR MEDICINE 59


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5

Image quality tests


Q UA L I T Y CO NTR O L
C T SYS TE M

CT number uniformity and noise


Purpose. To check the uniformity of a CT image of a homogeneous object by
measuring the CT numbers (Hounsfield units, HU) across the image of
a homogeneous phantom [a CT dose index (CTDI) phantom can also
be used].

Description. Proceed as follows:


» Place the phantom on the table or head support and centre it vertically
and horizontally.
» Choose the protocol most often used clinically (head or abdomen,
depending on the phantom).
» Make a scout image and scan the phantom as would be done clinically.
» On reconstructed images, select regions of interest (ROIs) in the centre
and in four places at the periphery (usually 12 h, 3 h, 6 h and 9 h).
» Read the average CT number and standard deviation (noise).
» Calculate the difference in average CT number between peripheral
ROIs and the centrally placed ROI.
» Compare to reference values.

Tolerances. Tolerances for uniformity and noise are as follows:


Uniformity: Deviation of CT number from specified value ±20 HU
Noise: Deviation of noise from the specified values <25%

Artefact evaluation
Purpose. To evaluate the presence/absence of artefacts on images.

Equipment. Image quality phantom or other homogeneous phantom.

Description. Proceed as follows:


» Place the phantom on the table or head support and centre it vertically
and horizontally.
» Choose the most clinically used head protocol on the CT scanner.
» Set the scanning plane in the middle of the homogeneous module and
make a scan.
» Evaluate the image visually using a narrow window centred at the CT
number of the phantom material.

Tolerances. There should be no significant artefacts or distortions.

60 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

Q UA L I T Y CO NTR O L
C T SYS TE M
CT number accuracy
Purpose. To check that the CT numbers of different materials are within the
range of expected values.

Equipment. Image quality phantom with different materials with known CT numbers
(for example: CATPHAN, PRO-CT phantom, Mini CT QC Phantom, ACR CT
accreditation phantom).

Description. Proceed as follows:


» Place the phantom on the table or head support and centre it vertically
and horizontally.
» Choose the most clinically used head protocol on the CT scanner.
» Set the scanning plane in the middle of the phantom module contain-
ing different materials and make a scan.
» Read the average values of CT numbers in the regions of interest (ROIs)
placed in the centre of each material in the module.
» Compare the measured CT numbers with the reference values.

Tolerances. Deviation of CT numbers from specified values:


± 10 HU for water and air
± 20 HU for other materials

Spatial (high contrast) resolution


Purpose. To test the high contrast resolution of the CT scanner.

Equipment. Image quality phantom with high contrast resolution module.

Description. Proceed as follows:


» Place the phantom on the table or head support and centre it vertically
and horizontally.
» Choose the most clinically used head protocol.
» Set the scanning plane in the middle of the phantom high contrast module.
» Evaluate the number of resolved high contrast patterns.
» Compare the result with the limiting values.

Tolerances. Deviation of spatial resolution ≤10% from manufacturer’s specification or


0.5 lp/mm, whichever is greater.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 61


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5

Image quality tests


Q UA L I T Y CO NTR O L
C T SYS TE M

Low contrast resolution


Purpose. To test the low contrast resolution of the CT scanner.

Equipment. Image quality phantom with low contrast resolution module.

Description. Proceed as follows:


» Place the phantom on the table or head support and centre it vertically
and horizontally.
» Choose the most clinically used head protocol.
» Set the scanning plane in the middle of the phantom low contrast
module.
» Evaluate the number of resolved low contrast objects.
» Compare the result with the values at acceptance.

62 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

CT dosimetry tests

Q UA L I T Y CO NTR O L
C T SYS TE M
Accuracy of indicated dose parameters
Purpose. To check the accuracy of CT dose parameters (CTDIvol and/or DLP) displayed on the
CT scanner control console.
Equipment. Calibrated dosimeter with 100-mm pencil ionisation chamber (CTDI probe) and
PMMA CTDI phantom (16 cm and 32 cm PMMA).
Description. Proceed as follows:
» Place the CTDI head phantom on the head support and centre it vertically and
horizontally using lasers.
» Place the CTDI probe in the centre hole of the phantom.
» Prepare the CTDI probe for the measurements as described in the manufacturer’s
manual.
» Place the CTDI probe in the CTDI phantom in such a way that the centre of the
CTDI probe will be in the centre of the CTDI phantom.
» Centre the table at this position (put the position to zero or note the actual position).
» Make a scout image of the CTDI phantom. Set the start and end of the scan in
such a way that the centre of the CTDI phantom (CTDI probe) is in the middle of
the scanning area.
» Set the table movement to 0. During this test, the table should stay in a fixed position.
» Select the axial scan using exposure parameters close to the most frequently
used head scan.
»M ake a scan (1 rotation) and record the measured dose (PKL). Repeat to obtain at
least three measurements. The ionisation chamber measures the product of air-ker-
ma and length (PKL), which is usually called the dose–length product (DLP). To calcu-
late the CTDI, use the following formula:
PKL
CTDI = where N•T is the nominal beam collimation (N slices of thickness T).
N•T
» The dose parameter, which is usually displayed on the CT scanner control
console, is the weighted CT dose index CTDIw, which is a combination of mea-
surements in the centre of the phantom and at the periphery (CTDIc and CTDIp,
respectively):
1 2
CTDIW = CTDIC + CTDIP
3 3
» R epeat the measurements by placing the CTDI probe in all four periphery holes (po-
sitions: 12 h, 3 h, 6 h, 9 h). All unused holes of the phantom must be filled with plugs.
» Calculate the deviation of measured CTDI value to CTDI displayed on the control
console.
Tolerances. Deviation of measured dose from indicated dose <20%.
Notes. • Measurements should be done for all tube voltages that are used in practice.
• The same procedure is followed using a body phantom (32 cm PMMA) and a standard body protocol.
• Measurements of PKL in the centre can also be used to test some other properties of the CT scan-
ner, such as linearity with tube load (mAs), repeatability and dependence on beam width.
• Instead of measurements within the phantom, measurements free in air can be performed. In this
case, an ionisation chamber is attached to the centre of the CT scanner gantry.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 63


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5

RADIOGRAPHERS’ TESTS evaluated by the QC team and reported


Q UA L I T Y CO NTR O L
C T SYS TE M

The role of the radiographer is to per- to the service engineer if necessary.


form and understand more frequent CT The main objective of the CT number
scanner tests (daily, weekly, monthly). The test is to ensure that the relative calibra-
American College of Radiologist (ACR) tion of all CT numbers to water remains
and International Atomic Energy Agency within acceptable limits and that quan-
(IAEA) have recommended that radiogra- tum noise and electronic system noise do
phers should be competent to perform not increase. After scanning, the phantom
basic tests [2, 6]. More specifically, the ACR CT number and its standard deviation in
has recommended that the QC tests listed specified ROIs are recorded and compared
in Table 1 should be performed by the ra- with reference values set by the manufac-
diographer. turer or medical physicist. If any remarks
are documented or reported to the QC ra-
Radiographer QC Frequency diographer regarding artefacts on images,
Water CT number and all image series must be analysed. The QC
Daily radiographer should try to identify possi-
standard Deviation
ble causes of artefacts on images.
Artefact evaluation Daily
In addition a visual QC checklist for ra-
Wet laser printer QC Weekly diographers has been suggested by the
Visual checklist Weekly ACR[6] in order to ensure that other com-
Dry laser printer QC Monthly
ponents of the CT scanner are functioning
satisfactorily and to aid in avoiding repeat-
Display monitor QC Monthly
ed scans. According to this checklist, the
Table 1: CT QC tests for radiographers following items should be checked on a
recommended by the ACR [6] monthly basis:
»» Table height indicator functioning
As a first action, a daily calibration (air »» Table position indicator functioning
calibration) should be performed before »» Angulation indicator functioning
any other test. All tests must be performed »» Laser localisation light functioning
using the same exposure and reconstruc- »» High-voltage cable/other cables safely
tion parameters in order to ensure that attached (and not frayed)
reproducible results are obtained. All un- »» Acceptable smoothness of table motion
expected changes in the constancy of »» X-ray on indicator functioning
measurements should be documented, »» Exposure switch functioning

64 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 5

»» Display window width/level ucational programmes. It provides an op-

Q UA L I T Y CO NTR O L
C T SYS TE M
»» Panel switches/lights/meters working portunity to expand learning outcomes
»» Door interlocks functioning for educational institutions and improve
»» Warning labels present the competencies of radiographers who
»» Intercom system functioning already perform CT procedures. In order
»» Postings present to ensure the provision of an effective, safe
»» Service records present and efficient service, radiographers’ train-
ing should incorporate instruction on QA
Most modern CT scanners have their own and QC practices, to include: legislation,
QC procedure in which the test phantom regulations and guidelines, test equip-
provided with the scanner is used. In CT ment and methodologies, programme de-
software QC packages, tests are automat- sign and implementation and reporting[7].
ed and measurement data are stored in With effective analysis and collaboration
CT software and can be accessed and an- with medical physicists and radiologists,
alysed. Such tests are not time consuming QC tests can also become an important
and can therefore be performed relatively training tool regarding the ways in which
quickly. As such they can be easily incor- parameters monitored during these tests
porated into a QA programme. affect image quality and dose. Although
QC phantoms cannot fully simulate pa-
tients, they can still provide radiographers
CONCLUSION with useful knowledge on the perfor-
Radiographers must be a part of the QC mance of their CT scanner and can be
team in order to secure, maintain or im- used as a valuable tool in the optimisation
prove the health and well-being of the process.
patient. Radiographers have a specific re-
sponsibility to actively participate in QA
and therefore should have the compe-
tences required to perform QC tests for
CT scanners and to evaluate the results of
routine QA and QC tests.
The document “Guidelines on Radiation
Protection Education and Training of Med-
ical Professionals in the European Union”[7]
is useful for the evaluation of existing ed-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 65


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 5
Q UA L I T Y CO NTR O L
C T SYS TE M

REFERENCES

1. Bauhs JA, Vrieze TJ, Primak AN, Bruesewitz MR, Mc- 5. American College of Radiology. ACR–AAPM techni-
Collough CH. CT dosimetry: comparison of mea- cal standard for diagnostic medical physics perfor-
surement techniques and devices. Radiographics mance monitoring of computed tomography (CT)
2008;28:245–253. equipment. ACR Technical Standard 2012 Resolu-
2. IAEA HHS-19. Quality Assurance Programme for tion No. 34.
Computed Tomography: Diagnostic and Therapy 6. American College of Radiology. 2012 Computed
Applications. International Atomic Energy Agency, Tomography. Quality Control Manual. 2012. https://
Human Health Series No. 19, 2012. www.acr.org/~/media/ACR%20No%20Index/Docu-
3. International Atomic Energy Agency. IAEA Human ments/QC%20Manual/2012CTQCManual1a
Health Series No. 19. Quality assurance programme 7. European Commission. Radiation Protection No.
for computed tomography: diagnostic and ther- 175. Guidelines on radiation protection education
apy applications. 2012. http://www-pub.iaea.org/ and training of medical professionals in the Euro-
MTCD/Publications/PDF/Pub1557_web.pdf pean Union. 2014. https://ec.europa.eu/energy/en/
4. ImPACT CT Scanner Evaluation Group. ImPACT Infor- radiation-protection-publications
mation Leaflet 1. CT scanner acceptance testing. 2001.
http://www.impactscan.org/download/acceptan-
cetesting.pdf

66 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
SPE C T/C T –

6
O PTIM I SAT I O N
O F SP E C T A N D
ACQ UI SI T I O N A N D
RECO N ST RU C T ION

by Maximilien Vermandel
and Hélène Lahousse
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
CHAPTER 6

INTRODUCTION
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

Single-photon emission computed tomography (SPECT) is a medical


imaging technique that allows in vivo measurements of the 3D
distribution of gamma photon-emitting tracers. Technetium-99m (99mTc)
is the radioactive isotope most commonly and widely used with SPECT
imaging. 99mTc SPECT is mostly used in association with a vector in order
to study the metabolism of a given organ.

The range of medical indications for SPECT remains of interest. Indeed, for nephrology
is wide and includes bone resorption, indications, quantification is expected to
studied using a bisphosphonate labelled estimate renal function or cortical transit
with 99mTc, and renal function, investigated time. In cardiology, cardiac function can
using 99mTc-labelled mercaptoacetyltrigly- also be analysed using 99mTc-pertechne-
cine (MAG3). 99mTc is also used with sesta- tate-labelled red blood cells and calculat-
mibi to study myocardial perfusion. In neu- ing the portion of isotope ejected during
rology, brain perfusion analysis is achieved each heartbeat to determine the ejection
using 99mTc coupled with a neutral and li- fraction. In particular, radiotracer uptake is
pophilic complex whose molecular mass is measured at the end of systole and dias-
sufficiently low to allow passage across the tole to estimate the ejection fraction, fol-
blood-brain barrier. However, some other lowed by evaluation of the kinetics in 3D.
isotopes are also used, such as iodine-123 Moreover, prior to therapy, the avidity of
in thyroid scintigraphy for hyperthyroidism the thyroid for iodine-123 may be assessed
or congenital hypothyroidism, indium-111 through measurement of the uptake rate,
with a somatostatin analogue for neuro- which serves as the basis for estimation of
endocrine tumours, krypton-81m for lung the iodine-131 activity to be prescribed.
ventilation, thallium-201 for evaluation
of cardiac viability and iodine-131 for the Since the range of applications of SPECT
treatment of thyroid cancer. imaging is wide, acquisition parameters
must be chosen carefully, in accordance
For some of these applications, the vi- with the indication being investigated.
sual inspection of SPECT images provides Indeed, SPECT studies depend on the set-
sufficient information (e.g. bone fracture, tings and methods used during acquisi-
arthritis, metastasis). For other applications, tion, reconstruction and post-processing.
however, quantification on SPECT images These parameters may drastically affect

68 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

the image quality and have to be carefully SPECT AND SPECT/CT IMAGING
considered both when performing the ac- INSTRUMENTATION
quisition itself and when interpreting the
images. Gamma camera basics
The first section of this chapter presents A gamma camera is a set of planar detec-
important background information on the tors, usually two (dual-head), mounted on a
principles of SPECT instrumentation. The gantry that enables simultaneous detector
following section is devoted to the factors acquisition of parts of a patient’s body from
that affect image reconstruction quality different orientations. Detection of the gam-
and the parameters that must be consid- ma photons emitted by the decay of the ra-
ered in order to achieve optimal acquisi- dioactive isotopes relies on the scintillation
tion. Finally, some new advancements in principle[1]. A basic sketch of a detector is
SPECT are discussed. shown in Fig. 1.

Figure 1

Transverse view of a gamma camera detector

Photomultiplier tubes

Light guide

Scintillator

Collimators

Emitted gamma photon

QUALIT Y CONTROL OF NUCLEAR MEDICINE 69


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

A gamma camera acquires two-dimen- tion of multiple projections from multiple


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

sional projectional images of the distribu- positions enables reconstruction of the


tion of a radiotracer based on detection of three-dimensional distribution of gamma
the emitted photons. As illustrated in Fig. photon-emitting tracers (Fig. 2). SPECT is
1, an emitted photon (or event) passes usually acquired in addition to planar scin-
through the collimator, which aims to se- tigraphy in order to refine the diagnosis.
lect photons from certain directions. The SPECT imaging is basically achieved by
collimator consists of an array of holes, rotation of the detectors around the sub-
most often parallel holes, resembling an ject. The motion of the detectors can be
assembly of lead straws. On Fig. 1, non-par- either step by step (SS) or continuous, the
allel photons would be stopped by the latter option being preferred for unstable
hole walls, the so-called septa. Once they patients (e.g. those with dementia). During
have passed through the collimator, pho- continuous SPECT acquisition, continuous
tons interact with the scintillator, which motion cycles are employed to maximise
is a large-area NaI(Tl) scintillation crystal, the number of events measured. In the
where they are absorbed; the absorbed event of unexpected patient movements,
energy is then re-emitted, with release of one or several cycles may be rejected be-
photons in the visible light spectrum. Ba- fore the reconstruction phase. In contrast,
sically, the scintillator is a “wavelength con-
Figure 2
verter” that converts high-energy photons
to low-energy photons. The low-energy Illustration of the acquisition of multiple
photons are then guided through the light projections to achieve SPECT imaging
guide to photomultiplier tubes (PMTs) that
serve to convert the light signal into an
electrical signal. PMTs are arranged in a 2D
array in such a manner that the electrical
signal collected at each PMT enables esti-
mation of the 2D location of each event.

SPECT
When used in static or whole-body mode,
gamma camera detectors only allow for
planar imaging of the radiotracer distribu-
tion. However, in SPECT imaging, acquisi-

70 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

SS acquisition requires just a single cycle, pose of attenuation and scatter correction

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
which basically entails rotation of each de- during tomographic reconstruction.
tector through 180° around the subject in The use of hybrid technologies may
order to cover the full 360° range (for de- entail significantly increased radiation ex-
tector configuration in 180° or “H” mode). posure for the patient. In order to reduce
Step duration or speed and number of cy- the exposure while maintaining good im-
cles in continuous mode depend primarily age quality, manufacturers have released
on the intensity of the radiotracer uptake systems that automatically modulate the
and the patient’s condition. tube current (mAs) according to the den-
Finally, image reconstruction is achieved sity of tissues being explored. Basically, the
using specific algorithms to render the system automatically adapts, in real time,
spatial distribution of the radiotracer into the tube current to the thickness of tissue
matrices of 64×64, 128×128 or 256×256 crossed by the X-ray beam as deduced
pixels according to the location and size of from the topogram acquisition (Fig. 3). For
the organ studied. example, the X-ray beam is more strongly
attenuated by the shoulders than by the
SPECT/CT abdomen owing to the lower density of
Image reconstruction with conventional the latter; thus decreasing the tube current
SPECT acquisitions cannot be corrected for for the abdominal exploration allows for a
attenuation or scatter when considering substantial dose reduction. However, in
non-heterogeneous areas, an exception specific cases the dose reduction system
being the brain, for which Chang model- may not be activated. For instance, when
ling gives satisfactory results. For the pur- the CT scan is only acquired for the pur-
pose of such correction, the SPECT gantry pose of attenuation correction, minimal
can be coupled to a computed tomograph dose exposure is expected and thus high
(CT) so that a CT scan can be acquired in voltage and tube current are carefully bal-
conjunction with SPECT imaging. This hy- anced to achieve this goal.
brid imaging system has two major advan-
tages: First, the quasi-simultaneous acquisi- Collimator
tion of SPECT and CT allows for the fusion In the chain of acquisition, the collimator
of functional and anatomical data with type is the first element to be considered
minimal concerns regarding registration. for optimisation of SPECT image acquisition
Second, the CT images, converted to a map since it closely depends on (1) the energy of
of electronic densities, are used for the pur- the emitted photons to be measured and

QUALIT Y CONTROL OF NUCLEAR MEDICINE 71


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

Figure 3
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

Illustration of a dose reduction system.


a) Estimation of the mAs on the Z-axis according to the topogram.
b) Dose level for different locations when using the dose reduction system

(a)

(b)

72 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

(2) the location being investigated. As de- limit the study duration and thus restrict

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
scribed previously, the collimator is an array the patient’s movements as far as possible.
of holes or honeycomb structure. Thus, the
main factors influencing the acquisition are Depending on the energy, the collima-
hole length, septa width and hole diameter tor is classified as low energy (99mTc, 123I,
(Fig. 4). For instance, increasing the length 201
Tl), medium energy (111In, 67Ga) or high
of the hole improves the collimation and energy (131I).
spatial resolution but leads to a lower sen-
sitivity. Septa thickness is related to emitted Collimator types also differ according
photon energy and is thicker for higher en- to the geometry of the holes. Most colli-
ergy photons (e.g. 0.4 mm for 140-keV pho- mators used in SPECT imaging are of two
tons versus 1.9 mm for >300-keV photons). types: parallel-hole collimators and con-

Figure 4 Figure 5

Main characteristics of a collimator Parallel-hole (a) and


affecting the image acquisition fanbeam (b) collimators

3
1 – Hole diameter | 2 – Hole length | 3 – Septa width (a) (b)

Hole diameter (or hole size) influences the verging-hole (fanbeam) collimators (Fig. 5).
spatial resolution: a smaller hole diameter The collimator type to be used for a given
increases the spatial resolution but decreas- imaging application depends on the ratio
es the sensitivity. In general, the choice of between the size of the field of view (FOV),
collimator for SPECT imaging should pri- the size of the detector and the expected
oritise sensitivity over spatial resolution to spatial resolution and sensitivity[2].

QUALIT Y CONTROL OF NUCLEAR MEDICINE 73


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

The parallel-hole collimator is the histori- improve photon interactions and the sen-
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

cal design by Anger and this is still the stan- sitivity. Indeed, a thick crystal increases
dard collimator employed in clinical prac- the probability of total absorption of the
tice. For a parallel-hole collimator, the point incident photons when the energy is high.
source sensitivity is the same over the whole However, the resultant gain in sensitivity
FOV. When the organ or location under in- comes at the cost of a reduction in spatial
vestigation is smaller than the FOV, the use resolution.
of fanbeam collimators results in improved The thickness of the crystal/scintillator
performance in terms of both sensitivity is specific to the gamma camera and de-
and spatial resolution[3]. This latter configu- pends on the type of radiotracer used. For
ration may be preferred for brain imaging, a low-energy tracer, a thin crystal should
for instance. be preferred (3/8 inch) while a thicker
Finally, for cardiology imaging, specific crystal (5/8 inch) is to be preferred for
collimators allowing for faster cardiac SPECT higher energy tracers.
acquisition have recently been released[4]. Finally, the choice of crystal thickness
depends on the type of daily use. For in-
Crystal stance, SPECT with 99mTc will be acquired
Interaction of emitted photons with the using a thin crystal, whereas if most uses
scintillator crystal leads to the emission of involve 131I, a thick crystal will be pre-
fluorescence photons (light) in all direc- ferred.
tions. Only a limited fraction of light trav-
els to the PMT. Thus, the scintillator is sur- Photomultiplier tubes
rounded by a reflector material on all sur- Light emitted inside the crystal is collect-
faces, except the surface in contact with ed by the PMTs in order to convert the
the PMT, in order to maximise light col- light signal into an electrical signal (Fig.
lection at the PMT. Additionally, the crys- 7). PMTs comprise an evacuated glass
tal is coupled to the PMT array through a package with a photocathode, to convert
transparent medium of the same index of light to electrons, followed by a string of
refraction as the crystal so that reflection electron-multiplying dynodes. The ampli-
is minimised at the interface between the fication gain is high (>106). High voltage
PMT and the scintillator. for the whole PMT array and gain for each
The thickness of the crystal impacts PMT have to be carefully fine-tuned in or-
on both sensitivity and spatial resolution. der to achieve a similar response on all the
As illustrated in Fig. 6, a thick crystal will FOV for both the energy of incident pho-

74 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

Figure 6

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
Illustration of the effect of the crystal thickness on light collection at the PMT array.
In (a), the crystal is thicker, improving the sensitivity but limiting the spatial resolution (the cone
of light collected at the PMT array is wider). In (b), not all of the photons interact with the crystal
but the spatial resolution is higher (i.e. the cone of light collected at the PMT is narrower).

PMT array PMT array

(a) (b)

Figure 7

Illustration of a PMT converting light into an electrical signal

tons and the sensitivity. Furthermore, the the gantry. Variation in the PMT response
response of each detector has to be ho- according to angle may have an impact on
mogeneous independently of the angle of the reconstructed image.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 75


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

IMAGE RECONSTRUCTION frequently used being filtered back pro-


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

jection (FBP). FBP consists in filtering pro-


Algorithms jection data before propagating the pixel
Reconstruction of the images from projec- value on the slice to be reconstructed.
tions is a crucial stage in SPECT imaging. Because FBP is more computationally ef-
After the acquisition process, each pixel ficient (fast), it has long been preferred.
of the projection images at each angular However, FBP requires a huge amount of
position contains the number of events projection data to enable well-defined im-
detected. The principle of reconstruction ages to be obtained. In view of the high
consists in building a slice from the differ- number of projections acquired during
ent projections. Figure 8 shows an example X-Ray Computerised Tomography (CT)
of reconstruction from planar acquisition. examination, FBP leads to satisfactory
Basically, two types of reconstruction image reconstruction. However, the low
are used: analytic methods and iterative resolution of scintigraphic imaging and
methods[5]. Analytic methods are based the low number of projections available
on a back-projection algorithm, the most prevent proper rendering of the three-di-

Figure 8

10 16 8 10 16 8

9 X X X 9 4 2 3

14 X X X 14 2 10 2

11 X X X 11 4 4 3

(a) (b)

a) Two projections acquired at two angular positions.


b) Reconstruction of the slice from the events contain in each pixel of the projections

76 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

mensional distribution of the radiotracer the faster is the convergence. According-

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
when reconstructed using FBP methods. ly, the number of subsets and the number
Iterative methods address this issue and, of iterations should be carefully balanced.
thanks to the performance of the most Indeed, if the number of subsets and it-
recent generations of computers, they are erations is too small, the algorithm does
now commonly used for SPECT image re- not converge and the result is a poorly
construction. The general concept of itera- contrasted and blurred image, while if the
tive methods is to solve p=A×f where p is number of iterations is too large in respect
a vector of the projection data (i.e. projec- to the number of subsets, the reconstruct-
tions acquired), f is a vector representing ed image displays an increased noise level
the three-dimensional distribution to be (Figs. 10 and 11).
reconstructed (i.e. tomographic images)
and A is a function of projection (i.e. geom- Other benefits of iterative reconstruc-
etry of acquisition). The principle of itera- tion are that the projection function may
tive algorithms is to find a solution (i.e. f) include different acquisition parameters in
by successive estimates. Ordered subsets order to optimise image rendering. Indeed,
expectation maximisation (OSEM) is the spatial resolution (both intrinsic and ex-
most commonly applied algorithm and trinsic, depending on the collimator used),
is based on the maximum likelihood ex- attenuation by the patient, Compton scat-
pectation maximisation (MLEM). The aim ter (in the patient, collimator and/or crys-
of MLEM algorithms is to find a solution Figure 9
as the best estimate of f than can produce
projections p with the highest likelihood.
OSEM allows acceleration of computation
by dividing the set of projections into sub-
sets (Fig. 9). Thus, independently from the
acquisition duration and the numbers of
projections, the main factors influencing
the image quality are the number of sub-
sets and the numbers of iterations. (a) (b)
Finally, the higher the number of itera-
tions, the higher is the spatial resolution, Examples of ordered subsets:
but with an increase in the noise level, a) set of projections divided into four subsets;
and the higher the number of subsets, b) set of projections divided into two subsets

QUALIT Y CONTROL OF NUCLEAR MEDICINE 77


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

Figure 10

(a) 4 iterations, 4 subsets (b) 4 iterations, 16 subsets


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

Example of the poor contrast and blurring that occur if the number of subsets and iterations
is too small. Tomographic reconstruction of a bone study (99mTc), no attenuation reconstruction,
5-mm Gaussian post-filter, 64 angular positions (128 projections). Reconstruction was achieved
with OSEM, with 4 iterations and either 4 (a) or 16 subsets (b)

Figure 11

(a) 4i, 4s (b) 8i; 4s

(c) 16i, 4s (d) 64i, 4s

Example of the increase in noise level that occurs with a higher number of iterations. Tomographic
reconstruction of a bone study (99mTc), no attenuation reconstruction, 5-mm Gaussian post-filter,
64 angular positions (128 projections). Reconstruction was achieved with OSEM, 4 subsets and
an increasing number of iterations: 4 (a), 8 (b), 16 (c), and 64 (d)

78 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

tal) and collimator septal penetration can subject to higher attenuation on their

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
be modelled to optimise the reconstruc- paths. In other words, a photon emitted
tion. These reconstruction options with from the surface of the patient is less atten-
the possibility of applying post-processing uated than a photon emitted from a deep
filters must be implemented within acqui- location. As a consequence, the emission
sition and/or post-processing software to of photons from deeper locations are un-
facilitate the optimisation of image quality. derestimated on the projections (Fig. 12).

Attenuation correction For some locations, attenuation cor-


Emitted photons have different trajecto- rection may be achieved with the Chang
ries across the patient’s body, such that algorithm. However, this algorithm is only
those emitted from deeper locations are suitable for a homogeneous location, such

Figure 12

Example of photons emitted from different depths, with more attenuation occurring for photons
emitted from deeper locations

QUALIT Y CONTROL OF NUCLEAR MEDICINE 79


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

Figure 13
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

(a) (b)

(c) (d)

(e) (f )

Examples of images reconstructed without and with the use of attenuation correction based on
CT acquisition. a,b) A phantom filled with 99mTc solution and images reconstructed without (a)
and with (b) attenuation correction. c,d) A plain phantom with two sources of 177Lu in the centre
and images reconstructed without (c) and with (d) attenuation correction. e,f) Patient images
reconstructed from a 99mTc study without (e) and with (f) attenuation correction. The images clearly
show that without attenuation correction, events at depth are underestimated.

80 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

as the brain. For heterogeneous areas, an 256×256 up to 512×512. The size of the ac-

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
electronic density map of tissues acquired quisition matrix will define the size of the
from CT is preferred for attenuation cor- tomographic images to be reconstructed.
rection. Figure 13 shows examples of re- Thus, a higher matrix size will result in re-
construction without and with the use of cording of finer details within the limits of
attenuation correction based on CT acqui- the intrinsic spatial resolution. Matrix size
sition. selection directly affects the signal to noise
ratio (SNR) of acquired data. Indeed, larger
Matrix size and zoom matrix sizes lead to events being spread
Events detected on a projection are distrib- over a greater number of pixels; according-
uted on a grid, a so-called matrix. Each pic- ly, to maintain the same SNR with a larger
ture element (pixel) of the matrix basically matrix, the acquisition duration must be
covers the entire FOV and contains the longer so as to increase the number of
total number of events detected during events. Thus, acquisition duration and ma-
the acquisition. A matrix is described by trix size must be carefully balanced.
the number of rows (m) and columns (n) Zoom is used for the magnification of
defining its size: m × n. For SPECT imag- the object (e.g. specific body location,
ing, standard matrix sizes are 128×128 or Fig. 14) while keeping the matrix size un-

Figure 14

Illustration of a 4x4 acquisition matrix for the study of a kidney. a) The matrix covers the entire
FOV. b) A zoom of 2 is applied during the acquisition to achieve higher spatial resolution.

(a) (b)

QUALIT Y CONTROL OF NUCLEAR MEDICINE 81


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 6

changed. For example, an unzoomed im- CONCLUSION


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –

age with a 128x128 matrix and a pixel size Optimisation of SPECT/CT acquisition re-
of 5 mm will allow a pixel size of 2.5 mm lies on multiple factors. Obviously, the pro-
with the use of a zoom of 2. Finally, the tocol acquisition may be optimised taking
zoom allows for acquisition of the same into account both the technical aspects
matrix size on a limited FOV area but the detailed above and the ALARA principle
duration of frame acquisition may need (whereby the radiation exposure is kept
to be altered in order to maintain the SNR. “as low as reasonably achievable”). Howev-
er, patient status must be taken into con-
sideration. Indeed, the patient’s age and
NEW ADVANCEMENTS IN SPECT comorbidities may influence the design
Today, the newest CT scan reconstruction of the acquisition protocol, which thus
technologies, including iterative recon- may vary depending on the context. For
struction, are contributing in reducing instance, the exploration of pelvic pain by
patients’ radiation exposure. Additionally, means of bone scintigraphy in an elderly
new detection technologies will also help patient largely justifies a CT acquisition
to minimise this exposure. Indeed, the with SPECT because (1) risk of fracture of
new detectors based on semiconductor the pelvis is high and the bladder uptake
technology (CZT: cadmium-zinc-telluride) may prevent study of the sacrum and (2)
directly convert emitted photons into an the CT dose exposure remains acceptable
electrical pulse without the use of a scintil- since SPECT alone contributes the major-
lator or PMT. Thus, CZT detectors increase ity of the patient’s overall exposure[8]. Fur-
spatial resolution, energy resolution and thermore, the duration of the SPECT/CT
sensitivity. Finally, the improved sensitiv- acquisition may be reduced if uptake on
ity of this technology, initially developed planar scintigraphy is high. On the other
for cardiology applications[6], means that hand, in a child with gait abnormalities,
less radiotracer activity is required for an acquisition of a SPECT/CT after planar
equivalent image quality and allows re- study may be considered unnecessary if a
duction of the study duration, thereby complementary examination is scheduled
avoiding movement artefacts and reduc- (e.g. MRI), thereby avoiding needless dose
ing overall patient exposure[7]. exposure.

82 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 6

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F S PE C T A ND S PE C T/C T –
REFERENCES

1. Peterson TE, Furenlid LR. SPECT detectors: The Anger 6. Mouden M, Timmer JR, Ottervanger JP, Reiffers S,
camera and beyond. Phys Med Biol 2011;56:R145– Oostdijk AH, Knollema S, et al. Impact of a new ul-
R182. trafast CZT SPECT camera for myocardial perfusion
2. Van Audenhaege K, Van Holen R, Vandenberghe S, imaging: Fewer equivocal results and lower radiation
Vanhove C, Metzler SD, Moore SC. Review of SPECT dose. Eur J Nucl Med Mol Imaging 2012;39:1048–1055.
collimator selection, optimization, and fabrica- 7. Agostini D, Marie PY, Ben-Haim S, Rouzet F, Songy
tion for clinical and preclinical imaging. Med Phys B, Giordano A, et al. Performance of cardiac cadmi-
2015;42:4796–4813. um-zinc-telluride gamma camera imaging in coro-
3. Tsui BM, Gullberg GT, Edgerton ER, Gilland DR, Perry nary artery disease: A review from the cardiovascular
JR, McCartney WH. Design and clinical utility of a fan committee of the european association of nucle-
beam collimator for SPECT imaging of the head. J ar medicine (EANM). Eur J Nucl Med Mol Imaging
Nucl Med 1986;27:810–819. 2016;43:2423–2432.
4. Gremillet E, Agostini D. How to use cardiac IQ•SPECT 8. Rausch I, Fuchsel FG, Kuderer C, Hentschel M, Beyer T.
routinely? An overview of tips and tricks from prac- Radiation exposure levels of routine SPECT/CT imag-
tical experience to the literature. Eur J Nucl Med Mol ing protocols. Eur J Radiol 2016;85:1627–1636.
Imaging 2016;43:707–710.
5. Bruyant PP. Analytic and iterative reconstruction al-
gorithms in SPECT. J Nucl Med 2002;43:1343–1358.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 83


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
7
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
RECO N ST RU C T ION
O PTIM I SAT I O N
ACQ UI SI T I O N
O F PE T/C T —
AND

by Leesa Ross and Dusty York


CHAPTER 7

INTRODUCTION
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

Footnote: This chapter will mainly focus on whole-body FDG imaging.


For specific brain and cardiac PET/CT, please refer to the Tech Guides on
Brain Imaging and Myocardial Perfusion Imaging.

Positron emission tomography (PET) is diopharmaceutical is fluorine-18 fluorode-


able to evaluate the metabolic activity of oxyglucose (18F-FDG), which can be used
target tissues. The principle behind PET for total body imaging as well as for brain
is that two 511-keV photons are emitted and cardiac imaging (Fig. 1). Acquisition
from an annihilation reaction between a and reconstruction are important steps in
positron and an electron. When the 511- the scanning process. Without standards
keV photons are detected simultaneously and protocols, the patient outcome may
by the PET scanner, the process of image suffer. It is best to use the manufactur-
formation is begun. The isotopes car- er’s settings; however, with advances and
bon-11 and germanium-68 are frequently changes in the profession, due to research,
used; however, the most common PET ra- optimisation may require modifications.

Figure 1

A: A PET/CT image consisting of coronal whole-body CT image. B: PET image with CT attenuation
correction. C: Fused PET/CT image (C). Courtesy of the Journal of Nuclear Medicine Technology
(JNMT)[1]

86 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

Patient preparation for a whole-body and by the sides for head and neck im-

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
PET scan with 18F-FDG includes no con- aging. Elevating the arms when imaging
sumption of food, simple carbohydrates the torso will reduce beam-hardening ar-
or liquids other than plain (unflavoured) tefacts, as well as artefacts caused by trun-
water for 4 h prior to 18F-FDG injection for cation of the measured field of view (FOV).
the non-diabetic patient. It is also recom- If imaging the head and neck area, as well
mended that the patient should not con- as the torso, it is beneficial to acquire two
sume food after midnight the night before separate acquisitions — one with the arms
the study. A low carbohydrate diet is pre- elevated for torso acquisition and one
ferred for one day prior to the study[2]. with the arms by the sides for head and
The patient’s blood glucose level should neck imaging. In radiation therapy treat-
be evaluated prior to injection of 18F-FDG. ment planning, the position of the patient
For clinical studies involving the whole should be the same as for radiotherapy
body, the blood glucose level should be treatment set-up.
lower than 11 mmol/L (~200 mg/dL). Pa- Skull base to proximal thigh imaging is
tients with glucose levels out of this range recommended for most tumour imaging,
may be excluded from the study[3]. The typically from the external auditory me-
patient should be injected in a quiet area atus to the mid-thigh region. For tumours
and remain in that location with the lights with an increased potential for scalp, skull,
dimmed for 60–120 min post injection. or brain involvement or for lower extrem-
The patient should be kept warm and re- ity involvement, whole-body imaging is
quested to sit still during the waiting pe- performed. Limited-area imaging can be
riod. For the purposes of radiation safety performed when there is a specific area of
and compliance with ALARA, explanation interest.
of the procedure should be completed pri- PET acquisition and reconstruction
or to injection, thereby reducing radiation protocols should always begin with the
exposure to the technologist. factory-recommended settings. Any mod-
Before the patient is positioned on the ifications in protocol should be evaluated
table, he or she should be instructed to carefully to avoid affecting patient images
empty the bladder to enhance comfort in a negative manner. Over time, software
and decrease the opportunity for artefacts. upgrades and changes in professional rec-
The patient is then placed in the supine ommendations may occur that warrant
position, with arms elevated and support- modification of the factory-recommended
ed above the head for torso acquisition settings.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 87


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

Many brands of scanners are available EMISSION IMAGES


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

for use by the nuclear medicine commu- The acquisition time varies from 2 to 5
nity. Selection is based on patient popula- min or longer per bed position for body
tion, department needs, costs and facility imaging. Continuous bed movement
vendor preference. may also be utilised to image the pa-
Various professional organisations and tient, if the scanner has that capability.
societies have developed performance Acquisition time varies according to the
guidelines, including the American Col- administered activity, the patient body
lege of Radiology (ACR). The ACR-rec- weight, 2D versus 3D and the count rate
ommended scanner specifications are as capability and sensitivity of the PET scan-
follows: ner. The average imaging time for skull
to mid-thigh acquisition is 15–45 min. A
For the CT scanner: variation in time per bed position can be
a. Spiral scan time: <5 s (<2 s is preferable) used if the system is capable of that tech-
b. S lice thickness and collimation: nique. This enables a faster scan time by
<5 mm (<2 mm is preferable) permitting reduction in the time per bed
c. Limiting spatial resolution: >8 lp/cm position in areas outside of the torso be-
for >32-cm display field of view (DFOV) cause those areas have less attenuation.
and >10 lp/cm for <24-cm DFOV Administration of a higher activity can
also reduce the image acquisition time,
For the PET scanner: but compliance with ALARA principles
a. In-plane spatial resolution: <6.5 mm requires a lower dose and longer acqui-
b. A  xial resolution: <6.5 mm sition. Formulas are available to calculate
c. S ensitivity (3D): >4.0 cps/kBq dose taking into account bed overlap and
d. S ensitivity (2D): >1.0 cps/kBq patient weight, and whether maximum or
e. U niformity: <5% minimum activity is used.
When using systems with a high count
For the combined PET/CT scanner: rate capability, 18F-FDG activity and scan
a. M
 aximum co-scan range (CT and PET): duration for each bed position must
>160 cm be adjusted so that the product of the
b. M aximum patient weight: 18
F-FDG activity and scan duration ±10% is
>159 kg equal to or greater than the specifications
c. P
 atient port diameter: >59 cm set out below[3]:

88 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

The figures for systems with the scanner. Acquisition systems with sep-

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
bed overlap of <25% are: ta are 2D scanners. Recent scanners devel-
»P
 roduct of <MBq/kg × min/bed > oped in the early 2000s use 3D technology
27.5 for 2D scans and do not use septa[5]. Some scanners are
»P
 roduct of MBq/kg × min/bed > capable of 2D and 3D acquisition[6].
13.8 for 3D scans 3D acquisition increases sensitivity four
to ten times due to an increase in lines of
The dosage is then calculated as follows: response (LORs) resulting from the detec-
»» FDG activity in MBq for 2D scans = tion of cross-plane events[7]. Removal of
27.5 × weight/(min/bed) septa in 3D acquisitions provides a higher
»» FDG activity in MBq for 3D scans = detection probability. Increased sensitivity
13.8 × weight/(min/bed) also results in increased random and scat-
ter contributions. To compensate, these
And for systems with scanners are also equipped with new and
a bed overlap of 50%: fast detectors using new scintillation crys-
»» Product of MBq/kg × min/bed > tals. Faster crystals provide better count
6.9 (3D only) rate performance. A short coincidence
»» FDG activity in MBq = 6.9 × weight/ timing window can be applied, which re-
(min/bed) sults in a reduction of random and scatter
coincidences. Due to low sensitivity at the
No matter what technique is used to deter- edge of the detector ring in 3D acquisition,
mine the administered activity, the activity bed overlap from one bed position to the
within the FOV should not exceed the peak next may be warranted. In continuous
count rate capability of the system. motion acquisition there is no need for
overlap, as there is uniform detector sen-
sitivity over the patient[6].
2D VERSUS 3D ACQUISITION
PET images can be acquired in 2D or 3D
mode. Earlier scanners used septa made DYNAMIC IMAGES
of lead or tungsten that were positioned Dynamic images may also be obtained
within the FOV. The septa provided a way when a specific area is to be imaged as a
to limit photons emitted or scattered out- flow study. A bed position of 15–26 cm
side the transverse or transaxial plane to is capable of imaging smaller regions. Dy-
the detector. This limited the sensitivity of namic PET images are similar to dynamic

QUALIT Y CONTROL OF NUCLEAR MEDICINE 89


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

gamma camera images. Dynamic PET im- STANDARD UPTAKE VALUE


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

aging acquires complete volume images Reported data should include the type of
within a set time frame. Typical times for isotope utilised, the patient’s height and
evaluating arterial flow are 2 s per image body weight, FDG activity administered,
followed by a longer imaging time to assay time and time of injection. This
demonstrate further uptake of the tracer aids in standardised uptake value (SUV)
over time[7]. reporting. SUV is used to quantitate ac-
tivity in body organs, abnormalities or
the total body and supplements visual
LIST MODE interpretation of an area. It is derived as
If the system is capable of it, images may follows[8]:
be acquired in list mode. Events, timing
markers and physiological gating markers (Region of interest (kBq/mL)
are stored as lists in list mode; the data SUVBW =
Injected activity with
may be rebinned (recreated) during use of decay-corrected residual
cardiac or respiratory gating. These data subtracted (MBq/kg)/patient
can then be formatted into sinograms for weight (kg) where BW is
reconstruction. Time of flight scanners are body weight.
capable of acquiring in list mode; there-
fore, list mode may become the standard
method for acquiring data[7].
ACQUISITION PARAMETERS
Further instructions on obtaining the
DUAL TIME POINT emission image include that the online
For certain disease states, dual time point randoms correction should be based on
imaging may be beneficial. This involves the “delayed coincidence time window”
an initial scan, a waiting period and then technique or randoms correction using
a delayed scan. In the case of tumour im- a model based on (block) single count
aging there will be an increase in activity rates. This has an effect on image con-
over time, whereas in infections there will trast. Decay correction must be ‘on’, as
be a decrease over time[6]. Thus delayed well. Shorter image times result because
imaging can increase the specificity of of the increase in contrast that occurs
whether activity is due to either tumour or from the high-quality image obtained
inflammation. with time of flight[3].

90 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

TRANSMISSION IMAGE of 1.0 will result in no spacing. A pitch

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
The typical PET/CT scan involves the ac- of less than 1.0 results in overlap, while
quisition of a topogram followed by single a pitch of 2.0, or higher, results in spaces
or multiple helical CT scans. CT can be between slices, equal to the thickness of
performed for attenuation correction, ana- the slice. Pitch is inversely proportional to
tomical alignment or acquisition of an op- patient dose. Increasing the table speed
timised diagnostic scan. When obtaining or pitch reduces the radiation dose in pro-
attenuation-corrected/anatomical align- portion to the increase in pitch (unless the
ment images, a low milliampere-seconds scanner automatically increases mA with
setting is used to reduce the radiation increase in pitch). However, image quality
dose. In an optimised diagnostic CT scan, will be reduced[9].
standard CT milliampere-seconds settings Slice thickness and increment define
are used. This results in better spatial res- the positional relationship of image slices
olution of the CT scan. Modulation of the and the amount of separation between
tube current may also be used to reduce the slices or potential overlap. If the slice
radiation dose in patients without metal thickness is equal to the increment, there
implants[3]. This can be accomplished with is no separation. If the slice thickness is
dose reduction software developed by the less than the increment there is overlap.
manufacturer of the scanner, when avail- It is beneficial to have some overlap when
able, and enables optimal low mA and kV attempting to detect smaller lesions[9].
settings to be used. It is always the case Regarding rotation time, at slower times
that selection of CT acquisition param- a compensatory increase in mA is required
eters, such as pitch, rotation time, slice to add more X-rays to the image. The
thickness, increment, voltage and tube slower the rotation time, the more blurred
current, depends on the purpose of the the image. A faster rotation time is ben-
CT scan. If the CT image is obtained solely eficial when body motion is a concern.
for the purpose of attenuation correction, Increasing the rotation time will result in
voltage and current can be reduced to better resolution because each image is
near the lowest the scanner is capable of generated from more projections[9].
producing and still be acceptable. Other CT acquisition parameters are
Pitch is the ratio of table movement voltage and tube current. Denser areas re-
through the gantry during one 360° rota- quire a higher voltage. This voltage is mea-
tion, relative to beam collimation. The low- sured in kVp. The higher the mA setting, or
er the pitch, the more the overlap. A pitch tube current, the higher the quality of the

QUALIT Y CONTROL OF NUCLEAR MEDICINE 91


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

image acquired; however, radiation dose no longer available. Since 2006, scanners
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

will increase, as well[9]. have used CT to accomplish attenuation


Attenuation correction is critical for correction[5]. A topogram image is ob-
both visual and quantitative accuracy of tained prior to the PET scan, followed by
PET images. Attenuation will vary from single or multiple helical CT. When ob-
patient to patient due to body make-up taining a CT scan for attenuation correc-
and size. The only way to accurately eval- tion, a low milliampere-seconds setting is
uate the attenuation coefficient is to pass used to reduce radiation exposure. Some
a beam of radiation through the body vendors offer patient-specific software
and measure the attenuation. The first which calculates the lowest acceptable
commercial PET/CT scanners appeared dose to the patient (Tables 1 and 2).
in 2001. By 2006 PET-only systems were

Table 1: Example of a CT protocol based on Table 2: Example of a PET protocol based on


current clinical practicea current clinical practicea

Topogram 35 mA Scan duration Continuous bed


120 kVp motion 1–2 mm/s
1024 mm 2–5 min/bed position
Dose modulation Patient-specific dose Matrix 256
parameters reduction techniques
Zoom 1
available on most
systems Reconstruction Iterations 2, subsets 21

Slice 5 mm Filter Gaussian

Rotation time 0.5 s FWHM 5 mm

Pitch 0.8 s
a
Check vendor specifications for scanner utilised. For
acquisition and reconstruction protocols involving
Reconstruction for AC B30s, medium smooth
the brain and heart, please refer to the Tech Guide
FOV 780 mm
publications by the EANM on Brain Imaging and
Reconstruction for B30f medium smooth Myocardial Perfusion Imaging
imaging 5×5 mm, 500 mm FOV
a
Check vendor specifications for scanner utilised. For
acquisition and reconstruction protocols involving
the brain and heart, please refer to the Tech Guide
publications by the EANM on Brain Imaging and
Myocardial Perfusion Imaging

92 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

Figure 2

USE OF CONTRAST

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
Contrast may be used when performing the
CT portion of the scan to enhance demon-
stration of the anatomy. Contrast can be
administered orally or intravenously. Oral
contrast is administered when the area of in-
terest is in the gastrointestinal tract. Intrave-
nous (IV) contrast is administered when vas-
cular structures need to be differentiated[3].
The patient’s creatinine level and/or glo-
merular filtration rate should be tested pri-
or to administration of IV contrast in order
to reduce the possibility of damage to the
kidneys. Metformin, an oral hypoglycae-
mic agent, should be discontinued for 48
h post IV contrast injection[6].

BREATHING PROTOCOL
The CT for attenuation correction should be A curvilinear cold artefact (arrow) is commonly
acquired during the resting phase or shallow seen at the dome of the diaphragm/liver or at
breathing. If the CT scanner has six or fewer the lung base because of a respiration mismatch
detector rings, use of a breath-hold protocol on PET images with CT attenuation correction.
Courtesy of the Journal of Nuclear Medicine
in normal expiration should be considered
Technology (JNMT)[1]
for the duration of scanning of the thorax
and upper abdomen. Deep breath-hold for
chest CT acquisition can cause misregistra- facturer and age of the equipment. There is
tion artefacts on the emission image (Fig. 2). typically a proprietary file format, unique to
each vendor and model, which contains the
raw data. Raw data are first formatted into
IMAGE FORMATION AND sinograms. The sinograms represent the
RECONSTRUCTION activity across the detector at each projec-
The method of reconstruction of PET imag- tion angle. The sinograms are reconstructed
es can vary greatly according to the manu- into axial image slices and corrections are

QUALIT Y CONTROL OF NUCLEAR MEDICINE 93


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

applied. Other projection planes are then the process of back-projecting data across
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

created from the PET and the CT, to include the imaging matrix. It is a quick and sim-
coronal and sagittal. The CT scan data are ple method. However, the produced im-
used to create an attenuation correction ages suffer from extra noise and streaking
map for use in image reconstruction. artefacts which result in an overall reduc-
tion of contrast. FBP has consequently
been replaced by iterative methods.
IMAGE RECONSTRUCTION Iterative reconstruction is the process
METHODS of creating estimates of actual radiophar-
PET/CT image reconstruction is similar to maceutical distribution in an object. Each
SPECT. Image reconstruction is a process projection estimate is compared with the
of calculating the 3D activity distribution measured projection data, and the differ-
from the sinograms. During reconstruc- ence between them is then used to create
tion the data are normalised, with appli- a new estimate[10]. This process continues
cation of corrections for random coin- until the difference between the esti-
cidences, scatter radiation, attenuation, mate and the measured values reaches a
dead time and decay. Reconstruction specified value. A point is reached in the
algorithms and order of application vary number of iterations where there will be
between equipment manufacturers and no further improvement in image quality.
may be proprietary. Once this point has been reached, further
Reconstruction algorithms are general- iterations can actually begin to degrade
ly classified into two methods: analytical the image quality. It is important to deter-
and iterative. Image reconstruction differs mine the number of iterations which will
for 2D and 3D imaging. One sinogram is yield optimal image quality. The number
created for each 2D slice during 2D recon- of iterations is preset by the user. While
struction and one sinogram is created for iterative reconstruction produces images
each projection plane during 3D mode with less noise and artefacts than FBP, it is
image reconstruction[9]. The majority of computationally intense and takes signifi-
current scanners are 3D; therefore 3D im- cantly longer than FBP.
age reconstruction will predominantly be Maximum likelihood expectation max-
discussed in this chapter. imisation (MLEM) was one of the first
In the past, the most commonly used iterative algorithms. MLEM requires the
analytical image reconstruction method use of all the image data. The process
was filtered back-projection (FBP). FBP is proved to be time consuming and oth-

94 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

er techniques were developed which are converted into 2D sonograms, followed

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
faster and more efficient. Ordered subset by completion of the process of iterative
expectation maximisation (OSEM) is one reconstruction. Methods available to com-
of the most common iterative algorithms. plete this conversion of data from 3D to 2D
OSEM does not require the use of all the include single slice rebinning (SSRB) and
image data; instead only a subset of the Fourier rebinning (FORE). SSRB rebins 3D
projection data is used, resulting in a dras- data into 2D projection sinograms. SSRB
tic reduction in reconstruction time. Two results in a blurring of image data and is not
approaches are used: (a) a higher number commonly used today since there are bet-
of subsets (8–24) and small number of it- ter approaches. FORE incorporates a more
erations (2–8) and (b) fewer subsets (2–8) accurate process for converting data from
but more iterations (8–24). Both methods 3D to 2D projection data and is therefore
maintain the image quality despite the the most widely used technique. Rebin-
reduction in reconstruction time. In clin- ning results in reduced resolution as the
ical practice, use of fewer iterations and distance from the axial centre of the FOV
increased subsets appears to be the most increases. Full 3D reconstruction does not
widely used approach. Other iterative require the conversion of data and is the
methods have been developed by propri- optimal method to avoid the resolution
etary vendors as well[7]. loss associated with rebinning. Fully 3D
Use of 3D image reconstruction requires reconstruction is only available on mod-
additional steps beyond the requirements ern scanners but is the standard on new
of 2D image reconstruction. Reconstruction equipment. There is a significant increase
options for 3D image data are determined in data in the reconstruction process due
by the age of the scanner and software lim- to the increase in the number of LORs con-
its. The reconstruction of 3D data requires tained in each sinogram, which means that
that the raw data, 3D sinograms, be either additional computer components must be
rebinned into 2D information or recon- devoted to this step[11]. The process of full
structed completely as a full 3D volume[10]. 3D reconstruction is very complicated and
Significantly more data are acquired in 3D is beyond the scope of this chapter.
mode due to the increase in sensitivity. Along with reconstruction algorithms, the
Reconstructing 3D data in full 3D mode is data must also undergo filtering. The filter
computationally intense and currently this most commonly used in PET whole-body
option is only available on newer scanners. imaging, for the purpose of smoothing the
Traditional methods require 3D data to be image, is the Gaussian filter. This filter is not

QUALIT Y CONTROL OF NUCLEAR MEDICINE 95


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

defined by a parameter in frequency space TIME OF FLIGHT


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

like most filters used in nuclear medicine; Time of flight (ToF) PET systems are not a
rather, it is defined by its full-width half-max- new concept. ToF PET coincidence imag-
imum (FWHM) in pixel space, which is more ing was used during the 1980s on exper-
commonly known as the spatial domain. imental systems[12]. The first PET scanners
The width of the filter is measured in milli- used bismuth germinate (BGO) crystals,
metres. Increasing the FWHM on the Gauss- which have a low stopping power and
ian filter will result in a smoother image. A limited spatial resolution and cannot sup-
high-resolution image can be produced by port ToF PET. ToF systems were limited to
applying a FWHM value of 5 mm. research until the early 2000s. Current PET
systems provide scintillators with a higher
density, shorter decay time and modern
SCATTER CORRECTION electronics which allow for faster com-
Scatter correction is essential to achieve puting power. Lutetium oxyorthosilicate
accurate PET data. In 3D acquisition (LSO) and lutetium-yttrium oxyorthosili-
modes, scatter coincidences are even cate (LYSO) both have good timing reso-
greater than in 2D mode. Therefore, scat- lution and increased stopping power and
ter correction methods are even more energy resolution. ToF PET is common on
important. There are two scatter correc- today’s scanners. ToF PET allows for more
tion methods: energy window-based accurate determination of where an event
methods and calculation-based methods. has taken place. If an event occurs any-
Accurate scatter correction is difficult to where other than the midline of an LOR,
determine. Energy window-based meth- the crystal closer to the annihilation event
ods require acquisition of a second en- will detect its photon first. The time differ-
ergy window below the original energy ence between the arrival of the two pho-
window. Calculation-based methods are tons is used to determine where along
preferred. Such methods are modelled the LOR the event took place. As a result,
rather than measured, the amount of scat- ToF PET increases image resolution. The
ter correction depending on the model greatest benefit is found in larger patients,
applied. Single scatter simulation is a who suffer most from poor image quality.
currently used calculation-based method. ToF PET requires additional considerations
Scatter correction is applied during image during reconstruction. The addition of ToF
reconstruction and should take place prior data prompts a shift from sinogram to list
to attenuation correction. mode reconstruction[12].

96 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

POINT SPREAD FUNCTION to the partial volume effect, metallic im-

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
CORRECTION plants, respiratory motion, contrast medi-
Point spread function (PSF) correction is a um and truncation.
software option that corrects for the blur-
ring arising from mispositioning of events Partial volume effect
due to variation in depth of detection The partial volume effect is the apparent
within the patient. Some vendors refer to loss of density or concentration of a small
such correction as high-definition imag- object which occurs on PET and CT when
ing. PSF corrections are applied during re- lesions are small compared with the reso-
construction as part of the reconstruction lution of the scanner. The size of an object
algorithm. Detections along the edges of or lesion needs to be approximately three
the axial FOV can be slightly misplaced, times greater than the image resolution for
resulting in blurring of small outer lesions. the activity to be accurately represented[7].
PSF corrects for the blurring and enhances
visualisation of these lesions. PSF correc- Metallic implants
tions are available on new scanners. Streaking artefacts are generated when me-
tallic objects exist within the patient, such
as dental fillings, hip prosthetics or chemo-
PITFALLS AND therapy ports. The metallic object leads to
ARTEFACTS an overestimation of PET activity in the re-
Due to issues beyond the control of the gion of the object when the CT attenuation
technologist, common artefacts can be correction map is applied. This will lead to
demonstrated on the PET/CT scan due a false positive PET finding. However, not

Figure 3 (A) (B) (C)

A) Example of the way in which high-density metallic implants generate streaking artefacts and high CT
numbers (arrow) on CT images. B) High CT numbers will then be mapped to high PET attenuation
coefficients, leading to overestimation of activity concentration. C) PET image without attenuation
correction helps rule out metal-induced artefacts. Courtesy of the Journal of Nuclear Medicine
Technology (JNMT)[1]

QUALIT Y CONTROL OF NUCLEAR MEDICINE 97


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

all metallic objects produce false positives. use of the CT image to define the location
ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

For example, in the case of hip prosthetics, of the lesion will help in interpretation. To
the implants produce high CT numbers minimise respiratory motion artefacts, the
because of their high photon absorption. breathing procedure should be explained
These implants will also attenuate the PET to the patient thoroughly before begin-
511-keV photons. This results in no emis- ning the study (Fig. 4).
sion data and demonstration of a cold area.
When CT attenuation correction is per- Figure 4
formed, the PET images show diminished
uptake in that area. In these cases, the
non-attenuated PET images can be useful
for interpretation of the study (Fig. 3).

Respiratory motion
Due to the nature of the human body, re-
spiratory motion is the most common arte-
fact demonstrated on a PET/CT image. The
artefact is due to the difference between
the chest position on the PET image and
the chest position on the CT image. The A 58-year-old man with colon cancer.
image resulting from a PET scan is an aver- A) Lesion at the dome of the liver is mislocalised
age of many breathing cycles while the CT to the right lung (arrow) because of respiratory
image is obtained at a single stage of one motion. B) Image without attenuation
breathing cycle. To partially compensate correction shows that all lesions are confined to
for this possible artefact, the CT image can the liver. Courtesy of the Journal of Nuclear
Medicine Technology (JNMT)[1]
be obtained at mid-expiration, at mid-in-
spiration or during shallow breathing. The
image produced has a curvilinear cold
area at the lung–diaphragm interface. The Contrast media
most critical impact of this artefact is on liv- As previously mentioned, contrast agents
er lesions. Because of this motion, a liver can be administered in PET/CT studies.
lesion may appear to be in the lung, simu- Contrast agents such as barium sulphate
lating a lung nodule. This type of artefact is and iodine that are within the patient from
a misregistration of lesions. In these cases, a previous study can give rise to a false pos-

98 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

itive finding on the PET image because they mon in larger patients or patients imaged

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
mimic metallic object artefacts. The higher with their arms down by their sides. When
the concentration of the contrast, the high- a patient extends beyond the CT FOV, the
er the CT number, resulting in higher PET at- extended anatomy is truncated and not
tenuation coefficients. This causes an over- represented in the reconstructed CT im-
estimation of the PET tracer uptake, which age. This results in no attenuation correc-

Figure 5 (A) (B) (C)

A 61-year-old patient with lung cancer who ingested barium for an oesophagogram one day
before PET/CT scan. A) Concentration of contrast medium in colon (arrow) is increased because
of significant water reabsorption, shown on the CT image. B) High CT numbers of residual barium
overcorrect attenuation of PET emission data and mimic increased 18F-FDG uptake on the PET image
with CT attenuation correction. C) No increase in 18F-FDG uptake is seen on the image without
attenuation correction. Courtesy of the Journal of Nuclear Medicine Technology (JNMT)[1]

is the reason for the false positive study. A tion values for the corresponding region.
technologist should be aware of whether The consequence is a bias on the PET at-
the patient has had a previous study involv- tenuation-corrected images, which under-
ing use of contrast. In the case of prior con- estimate the SUV in the region in question.
trast use and presence of the resulting arte- Truncation also produces streaking arte-
fact, the non-attenuated PET images can be facts at the edge of the CT image, resulting
used for interpretation (Fig. 5). in an overestimation of the attenuation
coefficients to be used to correct the PET
Truncation data. To avoid this artefact, the patient
Truncation artefacts occur because of the should be positioned in the centre of the
difference in size of the FOV between the FOV and imaged with the arms above his
CT and PET images. They are most com- or her head, when possible (Fig. 6).

QUALIT Y CONTROL OF NUCLEAR MEDICINE 99


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 7

Figure 6 (A) (B)


ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —

A 54-year-old man with a history of metastatic melanoma (arrow). The CT image (A) appears truncated
at the sides and biases the PET attenuation-corrected image (B). Courtesy of the Journal of Nuclear
Medicine Technology (JNMT)[1]

100 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 7

ACQ UI S I TI O N A ND RE CO NS TRUC TI O N
O PTIM I S ATI O N O F PE T/C T —
REFERENCES
1. Sureshbabu W, Mawlawi O. PET/CT imaging artifacts. 7. Christian PE, Swanston NM. PET study guide. 2nd ed.
J Nucl Med Technol 2005;33:156–161. Reston, VA: SNMMI; 2015.
2. Boellaard R, O’Doherty MJ, Weber WA,   Mottaghy 8. 
Thie JA. Understanding the standardized uptake
FM, Lonsdale MN, Stroobants SG, et al. FDG PET and value, its methods, and implications for usage. J Nucl
PET/CT: EANM procedure guidelines for tumour PET Med 2004;45:1431–1434.
imaging: Version 1.0. Eur J Nucl Med Mol Imaging 9. Seeram E. Computed tomography: Physical princi-
2009;37:181–200. ples, clinical applications, and quality control. 3rd ed.
3. Boellaard R, Delgado-Bolton R, Oyen WJG, Giam- United States: Elsevier Health Sciences; 2008.
marile F, Tatsch K, Eschner W, et al. FDG PET/CT: EANM 10. Prekeges J. Nuclear medicine instrumentation. 2nd
procedure guidelines for tumour imaging: Version ed. Sudbury, MA: Jones & Bartlett; 2012.
2.0. Eur J Nucl Med Mol Imaging 2014;42:328–354. 11. Lartizien C, Kinahan P, Swensson R,  Comtat C, Lin
4. The American College of Radiology. ACR–SPR M,  Villemagne V,  Trébossen R. Evaluating image
practice parameter for performing FDG-PET/CT reconstruction methods for tumor detection in
in oncology. 2016. https://www.acr.org/~/me- 3-dimensional whole-body PET oncology imaging.
dia/71B746780F934F6D8A1BA5CCA5167EDB.pdf. J Nucl Med 2003;44:276–290.
Accessed January 12, 2017. 12. Vandenberghe S, Mikhaylova E, D’Hoe E, Mollet P,
5. Townsend DW. Combined PET/CT: the historical per- Karp JS. Recent developments in time-of-flight PET.
spective. Semin Ultrasound CT MR 2008;29:232–235. EJNMMI Phys 2016;3:3.
6. Hogg P, Testanera G, eds. Principles and practice
of PET/CT. Part 1. A technologist’s guide. EANM;
2012. http://www.eanm.org/content-eanm/up-
loads/2016/11/gl_Principles_and_Practice_of_PET-
CT_Part_1.pdf. Accessed January 11, 2017.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 101


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
8
ACCR E D I TAT I ON
FOR C L I N I C A L
T RIA L S: T H E
EAN M E A R L
PROJE C T

by Giorgio Testanera and Michel de Groot I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
CHAPTER 8

INTRODUCTION
Multiple research studies have shown the benefit of combined positron
THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

emission tomography (PET) and computed tomography (CT) in clinical


decision making using 2-[18F]fluoro-2-deoxy-D-glucose (FDG) as the
leading radiopharmaceutical for diagnosis, staging, prognosis and
response monitoring. PET/CT has the ability to provide quantitative
information through image analysis of the distribution of a wide variety
of PET radiopharmaceuticals, including FDG, using standardised uptake
values (SUVs).

Both quantification and the development greatest tools for establishing the position
of new and promising PET radiopharma- of PET/CT as a state of the art modality in
ceuticals have become more important oncology patient care within the compet-
in clinical decision making and improved itive environment of diagnostic imaging.
quality of care for the individual patient[1, 2]. Owing to its widespread use in tumour
The use of FDG-PET as a surrogate tool response assessment studies, the SUV, de-
for monitoring of therapy response offers fined as the activity concentration ratio in
better patient care by allowing individu- tissues, can be considered the measure to
alisation of treatment and avoidance of be used in order to standardise PET scans.
ineffective treatments: surgery, radiation Absolute SUV is used for definition of pa-
therapy or chemotherapy. Imaging bio- tient eligibility, patient stratification and
markers are of paramount importance, lesion selection, while relative SUV (per-
not only for patient evaluation but also for centage changes from pre-treatment val-
drug development. ues) can also be used to evaluate response
to therapy. However, SUV measurements
present uncertainties that may result in
QUANTITATIVE PET/CT resistance to their use in clinical trials of
Quantification in PET/CT can be defined cancer therapies. Possible causes of un-
as the use of a combination of methods certainties are as follows[3]:
to enhance image resolution, reduce im-
age noise and perform data analysis, and Technical factors
use of the maximum standardised uptake »» 
Relative calibration between the PET
value (SUVmax) can be considered as the scanner and dose calibrator (10%)
standard quantitative method in practice. »» Residual activity in the syringe (5%)
Assessment of cellular uptake is one of the »» Incorrect synchronisation of clocks (10%)

104 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

»» Injection vs. calibration time (10%) within the scope of EANM Research Lim-

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
»» Quality of administration (50%) ited (EARL) activities, was originally based
on FDG-PET and PET/CT: EANM procedure
Physics-related factors guidelines for tumour PET imaging: version
»» Scan acquisition parameters (15%) 1.0, published in the European Journal of
»» Image reconstruction parameters (30%) Nuclear Medicine and Molecular Imaging
»» Use of contrast agents (15%) (EJNMMI). This widely accepted guideline,
»» ROI (50%) revised to version 2.0 in 2014, aims at har-
monising quantification in multicentre
Biological factors studies and providing a minimum stan-
»» Uptake period (15%) dard for the acquisition and interpretation
»» Patient motion and breathing (30%) of PET and PET/CT scans obtained with
»» Blood glucose levels (15%) FDG. The guideline specifically addresses
patient preparation, dosage of FDG as a
Furthermore, the variability in methodolo- function of scanner model, patient weight
gy across centres hinders the exchange of and scan duration, data acquisition, image
SUV measurements. Therefore, standardi- reconstruction, data analysis and QC pro-
sation of FDG-PET whole-body procedures cedures[2, 4].
is essential in multicentre trials. Accord-
ingly, the nuclear medicine community is The FDG-PET/CT accreditation ensures
working to achieve standardisation of PET/ harmonised quantitative performance
CT imaging and to improve the quality of of PET/CT systems within a multicentre
the images generated, both in clinical rou- setting through the standardisation of ac-
tine and in clinical trials [1, 2]. quisition and processing of PET/CT scans.
This rigorous harmonisation of imaging
systems enables PET/CT sites to compare,
EARL exchange and combine FDG-PET/CT find-
In 2010, the European Association of Nu- ings as data are collected and processed.
clear Medicine (EANM) initiated a pro- SUVs can also be reliably used owing to
gramme for the accreditation of PET/CT the resultant reduction in inter- and in-
scanners using FDG in order to support tra-institute variability. The standardisation
compliance with requirements regarding achieved by the accreditation programme
quality control (QC) and quality assurance relates to imaging procedures and meth-
of PET/CT systems. The programme, run odology, including patient preparation,

QUALIT Y CONTROL OF NUCLEAR MEDICINE 105


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

scan acquisition and image processing gramme. The Italian Lymphoma Founda-
THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

and analysis, which is of the utmost im- tion (FIL), Lymphoma Study Association
portance for quality assurance in daily (LYSA) and GELA (Groupe d’Etude des
clinical practice as well as in multicentre Lymphomes de l’Adulte, i.e. Adult Lym-
trials[2]. phoma Study Group) are further groups
that have taken the EANM imaging guide-
line as the basis for PET/CT QC within their
COLLABORATION WITH OTHER studies. The Society of Nuclear Medicine,
INSTITUTIONS India (SNMI) and the Brazilian Society of
The EARL FDG-PET/CT accreditation pro- Nuclear Medicine (Sociedade Brasileira
gramme is strongly supported by the de Medicina Nuclear/SBMN) have also
European Organisation for Research and expressed great interest in the EARL FDG-
Treatment of Cancer Imaging Group (EO- PET/CT accreditation programme.
RTC-IG), which was founded in 2009 to en- The EANM imaging guideline is wide-
sure standardisation of image acquisition ly accepted and is also well represented
and quality assurance for EORTC clinical within the draft of the Uniform Protocols
trials with regard to CT, PET, magnetic res- for Imaging in Clinical Trials (UPICT), es-
onance imaging (MRI) and other imaging tablished by the Quantitative Imaging
modalities as they become available[5]. Biomarkers Alliance (QIBA).
EANM/EARL and the International Atom-
ic Energy Agency (IAEA) are also working
together with respect to quality stan- MULTICENTRE CLINICAL TRIALS
dards in nuclear medicine. EANM/EARL is The goal of the EARL FDG-PET/CT ac-
aware of the different harmonisation and creditation programme is to enhance the
standardisation efforts regarding PET/ quality standard of PET/CT investigations
CT worldwide and is eager to contribute for both daily use and multicentre studies.
to a possible alignment of accreditation FDG-PET/CT accreditation ensures similar
programmes. To this end, discussions are performance of PET/CT systems within a
being held with the Society of Nuclear multicentre setting through the harmoni-
Medicine and Molecular Imaging (SNMMI) sation of acquisition and processing of PET/
Clinical Trials Network (CTN). CT scans. Centres can compare, exchange
Furthermore, the Japanese Society of and combine FDG-PET/CT findings, includ-
Nuclear Medicine (JSNM) has adopted ing SUVs, since data are collected and pro-
the basis of the EARL accreditation pro- cessed in a standardised manner[2].

106 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

In order to ensure that data can be ex- sation. Centres seeking accreditation also

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
changed between centres participating in benefit from:
a multicentre trial organised in collabora- »» EARL’s knowledge, which is based on the
tion with EARL, it is essential that all PET/ contributions of worldwide-recognised
CT systems used in the trial are accredited imaging experts, with a global leading
by EARL. Furthermore, all centres should role in PET/CT research, who provide ad-
be required to apply the EARL-approved vice within the programme and monitor
parameter settings (used for processing its development.
of the QC phantom images) for the recon- »» Imaging results that can be compared,
struction of imaging during the course of exchanged and combined, since the
the trial. The sites can apply additional re- prerequisite for evaluation of imaging re-
constructions for other purposes, local use sults within preclinical and clinical (mul-
or diagnostic interpretations. The EANM ticentre) trials is comparable scanner
provides recommendations for acquisition performance across multiple sites (re-
times per bed position in combination duction of inter- and intra-institute vari-
with dosage per kg patient weight for var- ability in SUV results; provision of lower/
ious types of PET/CT system in the EANM upper limits of recovery coefficients; cal-
imaging guideline[2]. ibration factor within ±10%).
»» Accurate, reproducible and quantitative
An up-to-date list of accredited centres assessment enabled through standard-
participating in the EARL FDG-PET/CT ac- isation of methodology, including pa-
creditation programme, also known as the tient preparation, scan acquisition, im-
Centres of Excellence (CoE), can be found age processing and analysis.
via the EARL website. »» Reliable and quantitative imaging biomark-
er results generated within multicentre
clinical trials, leading to an enhanced
ACCREDITATION BENEFITS outcome (e.g. information on biological/
The current established and scientifical- pathological processes and response to
ly validated FDG-PET/CT accreditation therapeutic intervention) and thus ac-
programme provides independent qual- celerating compound development and
ity control/assurance within multicentre approval by the regulatory agencies and
clinical trials and specifically addresses lowering costs in the long run[4].
the needs of the pharmaceutical industry »» An exploratory further optimisation,
regarding harmonisation and standardi- presently being evaluated by EARL. This

QUALIT Y CONTROL OF NUCLEAR MEDICINE 107


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

procedure would allow lowering of After initial accreditation, centres are


THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

the administered FDG activity for PET/ requested at fixed quarterly intervals to
CT systems with higher sensitivity or meet the standard requirements, as de-
improved performance using new en- scribed in the most recent versions of the
hanced technology (e.g. better time- EANM imaging guideline and manual, in
of-flight performance, continuous bed order to retain their accreditation. Failure
motion or extended axial field of view, to adhere to the deadlines for submission
i.e. length of bed position). may lead to EARL putting a centre’s ac-
creditation on hold. The documentation in
the initial accreditation procedure needs
PROCESSES INVOLVED IN to be submitted only once, whereas the
ACCREDITATION QC measurements and documents need
Sites which are seeking EARL FDG-PET/CT to be regularly performed and submitted
accreditation for the first time need to pass to retain the accreditation. An update of
through the initial procedure. This proce- the online questionnaire in the first quar-
dure includes submission of the online ter of each year is a prerequisite for sub-
questionnaire and signing of the statement mission of the QC documents[2, 6].
and signet policy. Additionally, sites are
asked to perform and submit QC data with-
in the subsequent 3 weeks. For this pur- SPECIFIC QUALITY CONTROLS
pose, sites have to perform calibration QC
measurements using a cylindrical calibra- Calibration QC
tion phantom and image quality QC mea- Calibration QC measurements should be
surements using a NEMA NU2-2001/2007 performed for initial accreditation and
image quality phantom and to submit on- have to be repeated every 3 months. The
line the image data in DICOM format and aim is to verify that the average activi-
the results to EARL. A detailed step-by-step ty concentration and/or SUV within the
procedure is published in the FDG-PET/CT phantom is within 10% of the expected
accreditation manual to assist sites in ade- value. For this purpose a cylindrical cali-
quately performing these measurements. bration phantom of any dimension but
After submission and independent review with a precisely known volume is required
of the results by EARL, the site is granted (it is preferable to obtain the phantom via
EARL FDG-PET/CT accreditation, assuming the manufacturer, ensuring that it is spe-
the results meet EARL requirements [2, 6]. cific for the used PET/CT model; if this is

108 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

not possible, the phantom should be 20 ly 70 MBq (65–75 MBq) FDG. The FDG ac-

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
cm in diameter and 20–30 cm long) (Fig. tivity is dispensed into the phantom when
11). The process of cross-calibration de- it is completely full with water, ensuring
termines the correct and direct (cross- or that all the activity is in the phantom by
relative) calibration of the PET/CT system flushing the syringe a few times. The phan-
with the institution’s own dose calibrator tom must then be vigorously shaken in
or against another one which is used to order to homogenise the distribution of
determine patient-specific FDG activities[4]. the activity.

Figure 1

Calibration QC setup (© Giorgio Testanera and Fabio Provenzano, Humanitas Research Hospital)

All relevant information on the standard After positioning the phantom in the
operating procedure (SOP) needs to be centre of the gantry, a PET/CT scan consist-
recorded accurately, including the exact ing of at least two PET bed positions needs
phantom volume, the scanner hardware to be acquired. PET/CT scan acquisition
and software, the exact doses and the time and reconstruction should be performed
of performance of every step. identically to patient studies as prescribed
It is first necessary to prepare a 5- to 10- in the clinical protocol. However, for statis-
ml syringe with an activity of approximate- tical reasons somewhat longer emission

QUALIT Y CONTROL OF NUCLEAR MEDICINE 109


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

times (e.g. 5–10 min per bed position) » to determine/check the correctness of
THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

are recommended, with inclusion of a calibration and quantification using


a standard transmission scan or (low- a non-cylindrical (calibration) phantom
dose) CT for the purpose of attenuation containing a set of high-contrast spher-
correction. ical objects
Reconstructions should be per- » to measure standardised “activity
formed with corrections for attenuation, concentration or SUV recovery coef-
scatter, normalisation, decay and dead ficients” as a function of sphere (tu-
time, i.e. all corrections necessary for mour) size
quantification [6].

Figure 2

Typical calibration QC images

Image quality QC The main aim of this procedure is to


Although correct cross-calibration is guar- guarantee comparable quantitative PET/
anteed using the calibration procedure CT system performance with respect to
described above, differences in SUV quan- SUV recovery and quantification. Image
tification may still occur between centres quality QC measurements should be
as a result of differences in the reconstruc- performed for initial accreditation and
tion and data analysis methodology. Con- thereafter have to be performed and
sequently, an image quality QC procedure submitted every year. These measure-
has been developed: ments are based on calculation of SUV

110 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

recovery coefficients using the NEMA mixed with the FDG contained in one

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
NU2-2007 image quality phantom (Fig. of the syringes (20 MBq), flushing the
3). The recovery coefficient is determined syringe several times and homogenis-
as a function of the sphere size, using the ing the solution (Fig. 4). After this step,
maximum pixel value and the A50 vol- the different diameter spheres inside the
ume of interest (VOI). As for calibration phantom should be filled, leaving no air
QC, all relevant information on the SOP inside of the spheres.
must be recorded accurately, including
the exact volume of the phantom, the The next step is to remove 30 ml wa-
scanner hardware and software, the ex- ter from the background compartment

Figure 3 Figure 4

The NEMA NU2-2007 image The image quality phantom preparation


quality phantom procedure, step 1

act doses and the time of performance of of the phantom and to add 20 MBq of
every step [4, 6]. FDG, flushing several times to ensure that
The first step is to prepare two syring- all activity has been dispensed from the
es, each containing 20 MBq of FDG at syringe into the phantom. At this point,
the expected phantom acquisition time. the background compartment should be
To prepare the solution for use in the filled entirely with water and the solution
spheres, a bottle should be filled with homogenised by shaking the phantom
exactly 1,000 ml of water, which is then vigorously (Fig. 5).

QUALIT Y CONTROL OF NUCLEAR MEDICINE 111


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

Figure 5 Figure 6
THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

The image quality phantom preparation The image quality phantom


procedure, step 2 scanning procedure

It is now possible to acquire a routine » Equipment used in obtaining EARL ac-


quantitative whole-body FDG-PET scan of creditation (e.g. dose calibrator, PET
at least two PET bed positions (at least 5 and/or CT system, clocks) should be
min per bed position), covering the entire able to perform accurately and precisely
phantom, with use of standard acquisition within operational tolerances; typically
parameters as specified in the guidelines this means that equipment is calibrated
(Fig. 6). It is important to position the and periodically serviced.
phantom so that the spheres are located » All clocks involved in the procedure
at the centre of the axial field of view. (dose calibrator, PET or PET/CT system)
should be synchronised with the offi-
Reconstructions should be performed cial local time to within 1 min in order
with corrections for attenuation, scatter, to prevent unnecessary deviations in
normalisation, decay and dead time, i.e. activities.
all corrections necessary for quantification » Residual activity within the syringe will
(Fig. 7)[4]. result in incorrect verification of PET or
PET/CT system calibration; therefore
post-injection measurement of activity
GENERAL SUGGESTIONS in the syringe is advisable.
All procedures require great accuracy, and » Accurate quantification in PET or PET/CT
various pitfalls are possible: can be threatened by high dead time or

112 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

Figure 7

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
Typical image quality phantom images

random fractions. To prevent this, it is specific system (manufacturer and mod-


important to check whether the count el) and that EARL needs to be informed if
rates exceed the limits of the scanner a PET/CT system has been replaced and/
acquisition system. If count rates do ex- or upgraded (since this may have conse-
ceed these limits, it is suggested that a quences for the current accreditation) or if
QC with a lower FDG activity should be a centre wishes to obtain accreditation for
performed, perhaps with preparation of an additional PET/CT system [2, 3].
the phantom a couple of hours in ad-
vance to allow for radioactive decay.
» The use of FDG with high activity concen- DATA ANALYSIS
trations can result in inaccurate dosing; in The dedicated software that has been de-
these cases (a portion of ) the stock solu- veloped to ensure standardised analysis of
tion should be diluted to clinically used QC documents allows the automatic im-
activity concentrations before use. port of data from the scan report forms to
the analysis tool, which minimises sources
Additionally, sites should always be of error and further standardises and fa-
aware that accreditation is granted for a cilitates work procedures. The calibration

QUALIT Y CONTROL OF NUCLEAR MEDICINE 113


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

Figure 8
THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

Image quality results verification with dedicated software

114 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 8

QC phantom measurements reveals the in the accreditation procedure may help

THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:
cross-calibration factor between the PET/ technologists to develop this new mindset
CT system and the dose calibrator. This fac- and to raise operating standards. There is
tor needs to be within ±10% from 1. The a great ongoing debate in Europe regard-
image quality QC phantom measurements ing the basic competencies and advanced
are analysed in order to determine the practice of nuclear medicine technolo-
background calibration factor by using sev- gists. A consensus document[7] has distin-
eral VOIs placed in the uniform background guished two levels, as follows:
compartment. The maximum allowable
calibration deviation is again ±10% from 1. Entry level
Additionally, the SUV recovery coefficients »» A competence and skill set that is con-
are calculated for each sphere (Fig. 8). A sidered necessary to ensure that nuclear
cold spot recovery using a central insert to medicine procedures are conducted to
verify the accuracy of scatter correction is an appropriate level
also performed. The software automatically »» This competence and skill set would be
compares the results of both QC phantom acquired during basic training/formative
analyses with the EARL specifications [2]. professional education

Advanced practice
TECHNOLOGIST’S ROLE »» A competence and skill set that is ac-
AND DAILY PRACTICE quired after basic training
In PET centres, technologists have the ma- »» The competence and skill set would be
jor role of acquiring images with PET and at a higher cognitive and clinical level
PET/CT scanners and ensuring that the ac- than basic training/formative profession-
quired images meet the quality standards al education
established by departmental criteria. The »» The competence and skill set would
first step in assuring that quantitative im- seek to improve patient care and man-
aging is accurate is compliance with all QC agement
procedures stipulated for the PET scanner »» The competence and skill set would
by national regulations, the vendor and seek to offer clinical career progression
local regulations. Working at an accred- opportunities
ited site for clinical trials implies not only
greater accuracy in standard daily practice It is true that the technologist’s role in EARL
but also a different mindset. Involvement accreditation can be considered a basic

QUALIT Y CONTROL OF NUCLEAR MEDICINE 115


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 8

competency, since it mainly relates to QC, developing advanced practice in nuclear


THE E A NM E A RL PR OJ E C T
ACC RE D I TATI O N FO R C L I NI C A L TRI A L S:

but meeting the high-quality standards medicine departments.


required surely pushes technologists to
improve their working skills. This improve- Useful links
ment, often requiring a completely new http://earl.eanm.org/cms/website.php
skill set, can be considered at the higher (EARL website – home)
cognitive level and fosters improvement
of patient care and departmental efficien- Copyright Information
cy. Therefore, the role of the technologist All figures are the property of Giorgio Tes-
can be considered to be either basic or tanera and Fabio Provenzano, Humanitas
advanced, depending on whether he or Research Hospital
she is taking the lead in the process, and
can be a good starting point in further

REFERENCES
1. Boellaard R, Oyen WJG, Hoekstra CJ, Hoekstra OS, 4. Boellaard R, Delgado-Bolton R, Oyen WJG, Giam-
Visser EP, Willemsen AT, et al. The Netherlands pro- marile F, Tatsch K, Eschner W. FDG PET/CT: EANM
tocol for standardisation and quantification of FDG procedure guidelines for tumour imaging: version
whole body PET studies in multi-centre trials. Eur J 2.0. Eur J Nucl Med Mol Imaging 2015;42:328–354.
Nucl Med Mol Imaging 2008;35:2320–2333. 5. DeSouza N, Hoekstra OS, Nestle U, Stroobants S,
2. Ettinger S, Testanera G, Sera T, Boellaard R, Verzijlber- Boellaard R, Schaefer-Prokop C, et al. EORTC Imaging
gen F, Chiti A. Quality visits: The EANM/EARL FDG-PET/ Group. Eur J Cancer Suppl 2012;10:82–7.
CT accreditation programme. In: Glaudemans AWJM, 6. Manual for EARL FDG-PET/CT Accreditation, version
Medema J, van Zanten AK, Dierckx RAJO, Ahaus CTBK, 2.1 (April 2013). EARL, EANM Research Ltd.
eds. Quality in nuclear medicine. Heidelberg Berlin New 7. Waterstram-Rich K, Hogg P, Testanera G, Medvedec
York: Springer International Publishing; 2017:415–427. H, Dennan SE, Knapp W, et al. Euro-American discus-
3. Boellard R. Standards for PET image acquisition and sion document on entry-level and advance practice.
quantitative data analysis. J Nucl Med 2009;50 Suppl Nucl Med Technol 2011;39:240–248.
1:11S–20S.

FURTHER SUGGESTED READING

» Flyer. EARL FDG-PET/CT Accreditation (March 2013). » Tatsch K. Standardisation and harmonisation boost
EARL, EANM Research Ltd. the credibility of nuclear medicine procedures. Eur J
Nucl Med Mol Imaging 2012;39:186–7.

116 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
9QUA L I T Y CON TROL OFACC
I N S TRUME N TATION AND
NUCRELE
D IA
THEPROTO
R ME
TATI
E A NM ECO
O NDIC
FO RINE
L SPR
A RL TAOJ
C L I NI C A L TRI A L S:
NDA
E C TRDIS AT IO N

117
CHAPTER 8

QUALIT Y CONTROL OF NUCLEAR MEDICINE


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
R ADIO N UC L I D E
C AL IBR ATO R

by Aljaz Socan
CHAPTER 9

INTRODUCTION
When undertaking any practices involving the human administration of
RA DIO NUC L I D E C A L I BRATO R

a radiopharmaceutical, it is essential to use sensitive instrumentation for


the detection and measurement of radioactive emissions. This allows a
practical measure for dispensing the low amounts required for diagnostic
uses and an accurate measure of therapeutic levels of radioactivity. This is
also especially important for experimental work contributing to scientific
knowledge relating radiation doses to biological effects.

When radiations from a radioactive mate- (ionisation chambers). These detectors


rial pass through matter, they interact with respond to radiation by means of ionisa-
atoms and molecules and transfer energy tion-induced electrical currents.
to them. The transfer of energy has two A volume of gas is contained between
effects: ionisation and excitation. Ionisa- two electrodes having a voltage differ-
tion occurs when the energy transferred ence (and thus an electric field) between
is sufficient to cause an orbital electron to them. The negative electrode is called the
be stripped away from its parent atom or cathode, the positive electrode the anode
molecule, thus creating an ion pair (a neg- (Fig. 1). Under normal circumstances, the
atively charged electron and a positively gas is an insulator and no electrical current
charged atom or molecule). Excitation oc- flows between the electrodes. However,
curs when electrons are perturbed from radiation passing through the gas causes
their normal arrangement in an atom or ionisation, both direct ionisation from the
molecule, thus creating an atom or mol- incident radiation and secondary ionisa-
ecule in an excited state. Both of these tion from delta rays [1]. Gas-filled detectors
processes are involved in the detection include ionisation chambers, proportional
of radiation events; however, ionisation counters and Geiger-Müller (GM) count-
is the primary event, and hence the term ers. The use of these detectors in nuclear
ionising radiation is frequently used when medicine is somewhat limited because
referring to the emissions from radioactive their stopping power and detection ef-
material[1]. ficiency for X-rays and gamma rays are
quite low; however, they find some use for
applications in which detection efficiency
BASIC PRINCIPLES is not a major factor and for detection and
Most gas-filled detectors belong to a class measurement of non-penetrating, parti-
of detectors called ionisation detectors cle-type radiations [1].

118 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

Figure 5

RA DIO NUC L I D E C A L I BRATO R


Basic principles of a gas-filled detector. Electrical charge liberated by ionising radiation is collected by
positive (anode) and negative (cathode) electrodes. (Adopted from [1])

The electrons produced by ionisation cal current) increases. When the voltage
are attracted to the positive electrode becomes sufficient to cause complete
and the ionised atoms to the negative collection of all of the charges produced,
electrode, causing a momentary flow of a the saturation region begins. The voltage
small amount of electrical current. at which the saturation region begins is
For maximum efficiency of operation, called the saturation voltage (Vs, typically
the voltage between the electrodes must Vs ≈ 50–300 V). Ionisation chambers are
be sufficient to ensure complete collec- operated at voltages in the saturation
tion of ions and electrons produced by ra- region (Fig. 2). This ensures a maximum
diation within the chamber. If the voltage response to radiation and also that the
is too low, some of the ions and electrons response will be relatively insensitive
simply recombine with one another with- to instabilities in the voltage applied to
out contributing to electrical current flow the electrodes. The amount of electrical
(recombination region)[1]. charge current released in an ionisation
As the voltage increases there is less chamber by a single ionising radiation
recombination and the response (electri- event is very small[1].

QUALIT Y CONTROL OF NUCLEAR MEDICINE 119


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 9

Because of the small amount of electri- and measured. Small amounts of electrical
RA DIO NUC L I D E C A L I BRATO R

cal charge or current involved, ionisation current are measured using sensitive cur-
chambers generally are not used to record rent-measuring devices called electrome-
or count individual radiation events. In- ters[1].
stead, the total amount of current passing
through the chamber caused by a beam
of radiation is measured. Alternatively, the RADIONUCLIDE CALIBRATOR
electrical charge released in the chamber A radionuclide calibrator is a device con-
by the radiation beam may be collected sisting of an ionisation chamber and an

Figure 2

Voltage response curve (charge collected vs. voltage applied to the electrodes) for a typical ionisation
chamber. In usual operation, applied voltage exceeds saturation voltage Vs to ensure complete collection
of liberated charge. (Adopted from [1])

120 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

Figure 3

RA DIO NUC L I D E C A L I BRATO R


Schematic view of radionuclide calibrator. (Adopted from [6])

electrometer (Fig. 3). Unlike other types large for assay with NaI(Tl) detector sys-
of ionisation chamber, radionuclide cal- tems.
ibrators employ sealed and pressurised
chambers filled with argon gas. This elim- Because ionisation chambers have no
inates the effect of changing barometric inherent ability for energy discrimination,
pressure on output readings. Ionisation they cannot be used to select different
chamber radionuclide calibrators assay gamma ray energies for measurement, as
the total amount of activity present by is possible with detectors having pulse-
measuring the total amount of ionisation height analysis capabilities. An approach
produced by the sample. Radionuclide that is used to distinguish low-energy
calibrators typically are calibrated to read from high-energy gamma ray emitters
directly in units of activity (Bq or Ci), with (e.g. 99mTc vs. 99Mo) is to measure the sam-
switches to set the display for different ple with and without a few millimetres of
radionuclides. They are used for assaying lead shielding around the source (Fig. 4).
relatively large quantities (i.e. MBq range) Effectively, only the activity of the high-en-
of gamma ray-emitting radioactivity too ergy emitter is recorded with the shielding

QUALIT Y CONTROL OF NUCLEAR MEDICINE 121


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 9

in place, whereas the total activity of both counter depends on sample positioning[1].
RA DIO NUC L I D E C A L I BRATO R

emitters is recorded with the shielding ab- If a small volume of radioactive solution of
sent. This technique can be used to detect constant activity in a test tube is diluted
tens of kBq quantities of 99Mo in the pres- progressively by adding water to it, the
ence of tens or even hundreds of MBq of counting rate recorded from the sample
99m
Tc [1]. in a standard well detector progressively
decreases, even though the total activity
Figure 4 in the sample remains constant. In es-
sence, the geometric efficiency for the
sample decreases as portions of the ac-
tivity are displaced to the top of the well.
If the volume of a sample is increased by
adding radioactive solution at a constant
concentration, the counting rate first in-
creases linearly with sample volume (or
activity) but the proportionality is lost as
the volume approaches and then exceeds
the top of the well. Eventually there is little
change with increasing sample volume,
although the total activity is increasing.
For example, an increase in sample vol-
99
Mo breakthrough test lead shield
ume in a standard test tube from 7 to 8
(“radionuclide purity testing”) mL, i.e. a 14% increase in volume, increases
the counting rate by only about 1%[1]. Thus
the sample volume has significant effects
As with the NaI(Tl) well counter, radio- on the counting rate with well counters.
nuclide calibrators are subject to sample For the stated reasons, sample volumes
volume effects (the fraction of gamma should be the same when comparing two
rays escaping through the hole at the samples. A technique applicable when
end of the well depends on the position adequate sample volumes are available is
of the source in the well: it is about 7% if to use identical test tubes for all samples
the source is near the bottom of the well and to fill them such that the volume of
but increases to 50% if it is near the top) activity inside the well itself does not differ
and the geometric efficiency of a well between samples.

122 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

Absorption of gamma rays within the non-linearities requiring adjustment or

RA DIO NUC L I D E C A L I BRATO R


sample volume or by the walls of the test correction of readings. In applying this
tube is not a major factor except when technique, it is necessary to correct for
low-energy sources, such as 125I (27–35 99
Mo contamination using the shielding
keV), are counted. Identical test tubes technique, especially after several 99mTc
and carefully prepared samples of equal half-lives have elapsed[1].
volume should be used when comparing
samples of these radionuclides[1].
The discussed effects should be inves- QUALITY ASSURANCE OF
tigated experimentally when a new dose RADIONUCLIDE CALIBRATORS
calibrator is acquired (acceptance testing), For radionuclide calibrator installation, op-
so that correction factors can be applied eration and maintenance, the manufactur-
in its use, if necessary. For example, a er’s operating manual should be followed.
quantity of activity can be measured in a Only authorised personnel should operate
very small volume (e.g. 0.1 mL in a 1-mL the calibrator, and up-to-date instructions
syringe). Activity can be diluted progres- on the operation and maintenance of
sively afterwards to larger volumes in equipment should be readily available for
larger syringes and then in beakers, and reference and use. The radionuclide cali-
so forth, to determine the amount by brator should be placed on a solid, vibra-
which the instrument reading changes tion-free base and, as recommended by
with the sample volume[1]. Another pa- the manufacturer, operated at a relatively
rameter worth evaluating is linearity of constant temperature and humidity and
response versus sample activity; this may away from direct sunlight and any room
be determined conveniently by recording heater or air conditioner. The area should
the reading for a 99mTc source of moder- not be affected by high-activity sources
ately high activity (e.g. 1 GBq, or whatev- and additional shielding may be required
er is the approximate maximum amount for background reduction and/or person-
of activity that the dose calibrator will be nel exposure reduction [2].
used to assay) and then recording the Radionuclide calibrators may be locat-
readings during a 24- to 48-h period (four ed within laminar airflow workbenches or
to eight half-lives) to determine whether isolators to facilitate the compounding of
they follow the expected decay curve for sterile radiopharmaceutical preparations.
99m
Tc. Deviations from the expected decay Since all surfaces in the aseptic processing
curve may indicate instrument electronic area require frequent cleaning and disin-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 123


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 9

fection, the radionuclide calibrator must tine tests[3]. The most thorough assess-
RA DIO NUC L I D E C A L I BRATO R

be able to withstand frequent cleaning ment of calibrator performance occurs


and disinfection with a variety of agents. at acceptance testing. Routine perfor-
Extreme caution must be exercised to mance testing includes most of the same
avoid damaging the radionuclide calibra- measurements. At acceptance testing or
tor chamber and readout unit(s) when before first use, calibration settings for
using cleaning and disinfection agents. radionuclides in source geometries oth-
The surfaces of chamber plastic liners and er than those provided by the calibrator
plastic dippers in the aseptic processing manufacturers must be determined if the
area must be cleaned and disinfected by potential assay uncertainty is unaccept-
appropriate means, such as wiping with a able (greater than 5%) [2–4]. Radiopharma-
sterile 70% isopropyl alcohol-dampened ceutical manufacturers and commercial
cloth. Readout units may also be protect- nuclear pharmacies should be required
ed with plastic covers that can be disin- to provide dosages whose assays are ac-
fected frequently with sterile 70% isopro- curate to within ±5%. These dosages can
pyl alcohol wipes[2]. be used to calibrate facility calibrators for
the respective source geometries. At ac-
Acceptance testing ceptance testing or before first use, those
The International Commission on Ra- individuals who will use the radionuclide
diological Protection (ICRP) defines an calibrator should be instructed in calibra-
acceptance test as a “test carried out at tor operation, maintenance and quality
the request and with the participation of control as appropriate. Instruction should
the user or his representative to ascertain include reading and comprehending the
by determination of proper performance manufacturer’s operating manual[2].
parameters that the instrument meets
the specifications claimed by the vendor”. Routine QC tests
The ICRP recommends that an accep- Routine quality control (QC) tests are re-
tance test be carried out at the time of peated at specific intervals to establish
installation and when appropriate after a and document changes from the initial
major service. In addition to ascertaining performance of the radionuclide calibra-
that the radionuclide calibrator meets the tor established at acceptance testing. The
vendor specifications, test or reference overall objective of performance testing
data are obtained at acceptance testing is to assure the continued accuracy of the
and used for comparison with future rou- dosage assays[2].

124 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

The tests recommended by EANM Phys- Clock accuracy

RA DIO NUC L I D E C A L I BRATO R


ics Committee and the EANM Working For radionuclide calibrators that incorpo-
Group on Nuclear Medicine Instrumenta- rate a clock, the accuracy of the stored
tion Quality Control[5] are as follows and time should be checked. The time should
summarised in Table 1: be synchronised to a standard time. The
»» Physical inspection time should be accurate to within 1 min
»» High voltage [3]
. Accurate time measurements are essen-
»» Clock accuracy tial when working with radionuclides that
»» Zero adjustment have short half-lives and/or for quantita-
»» Background counts/contamination check tive or semiquantitative imaging. Clock ac-
»» Constancy (check source, relative response) curacy should be checked following pow-
»» Stability er outages or when investigating aberrant
»» Accuracy test readings. Other facility clocks referenced
»» Linearity during dosage administration should be
synchronised with the radionuclide cali-
Physical inspection brator clock. Calibrator clock adjustments
Check the calibrator and source holders should be performed in accordance with
for damage. Damaged source holders the instructions in the operator’s manual.
should be repaired or replaced. Check
the display screen for proper operation Zero adjustment
and the console for keypad damage or The zero setting should be tested and re-
damage to or malfunction of any push- corded on each day of use prior to first use,
buttons/switches/dials. Check to assure and compared with the manufacturer’s
that the chamber liner is in place and that tolerance in accordance with the instruc-
small items (e.g. needle caps) have not tions in the operator’s manual and before
fallen into the well. any adjustments are performed.

High voltage Background response/contamination


Test the high voltage and compare the check
result with the manufacturer’s tolerances Background may be caused by external
in accordance with the instructions in the radiation fields, chamber/dipper/liner
operator’s manual. Constant and correct contamination or electronic noise. The
operating voltage is essential for accurate test should be performed with the source
activity measurement. holder/liner in place in the chamber; near-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 125


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 9

Routine test Purpose Frequency Comments


Physical To check system and any Daily The chamber may be concealed, and not
RA DIO NUC L I D E C A L I BRATO R

inspection source holders and other accessible for physical inspection, but
accessories for damage the loose accessories should be checked
High voltage To check the constancy and Daily/as Essential for an accurate activity
correct operating voltage recommended measurement
by manufacturer
Clock To check that the calibrator Daily Essential for calibrating radioactivity
accuracy clock is the same as the time to a specific time of day; clock time
of day throughout the department must be
the same (i.e. all wall clocks and internal
computer clocks)
Zero To check that the display is at Daily Record the zero setting (before any
adjustment zero when no radioactivity is adjustment); a drift in the “zero” reading
present may indicate that the instrument needs
repair
Background To check background response Daily Perform the test with the source holder/
counts under operational conditions liner in place in the chamber; remove
appropriate for a particular nearby radioactive sources that might
radionuclide; to detect cause an incorrect background reading
contamination
Constancy To check the stability and Daily Measure a long half-life radionuclide, e.g.
reproducibility of the ionisation 137Cs, with its own calibration factor;
chamber, electrometer and also, obtain relative measurements for
calibrator nuclide settings each nuclide setting to be used that day
Stability To check the short-term Yearly Counting precision is a measure of the
counting precision stability of the whole system, and is
measured by repeated measurements
and application of the chi-square test
Accuracy To check the accuracy of the Yearly This requires readings of sources of
activity reading known activity; refer to the supplier and
national measurement standards for
guidance
Linearity To confirm that the calibration Six monthly/ The change in response when the
setting for a particular yearly measurement range is changed should
radionuclide indicates the be minimal; the range of use should be
correct activity over the entire chosen between the maximum activity
range of use to be measured (e.g. in the GBq range
for a 99mTc eluate) and the lowest
activity to be measured (e.g. 1 MBq) for a
particular radionuclide

Table 1: Routine QC tests for a radionuclide calibrator. Equipment type: gas ionisation
chamber. (Adapted from Eur J Nucl Med Mol Imaging 2010;37:662–671 [6])

126 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

by radioactive sources that might cause results kept for the life of the chamber. The

RA DIO NUC L I D E C A L I BRATO R


an incorrect background reading must be source should be measured on its own
removed. The test should be performed on setting (e.g. 137Cs on 137Cs). Using the same
each radionuclide setting to be used that procedure, the source is also assayed on all
day. The magnitude of the background commonly used settings (e.g. 137Cs on 99mTc,
should be established at acceptance test- 137
Cs on 131I, 137Cs on 18F, etc.). This is referred
ing and measured on each day of use prior to as a “relative response test”[3] and is a
to first use and checked at each use. The measure of the constancy of the calibrator
measurement should be taken with no response for commonly used settings. If a
radioactive source in the chamber and on standard source is used rather than a test
the most common radionuclide setting source, the measurement obtained on the
with the source holder and contamination setting for the source radionuclide can also
shield in place. Any increase in the back- serve as an accuracy test. Measurements
ground above the normal value should should be within ±5% of the decay-cor-
be investigated. For calibrators that have a rected initial values. For secondary stan-
background adjust function, background dard radionuclide calibrators and reference
should be within the allowed range of radionuclide calibrators, measurements
adjustment. Routine performance tests should be within ±2%. The measurements
should be corrected for significant back- should be recorded and available for regu-
ground contribution. latory review [2].

Constancy (check source and relative in- Accuracy test


strument response) Measurements are taken with the refer-
Routinely measuring a long half-life check ence source in the source holder in the
source (or standard source) allows the user measurement position following the rec-
to demonstrate the constancy of the cali- ommended daily tests. In practice, accu-
brator’s response (e.g. electrometer stabili- racy testing involves testing with one or
ty or gas pressure changes) over time. The more traceable standards. The standards
measurements are taken (following the are typically in a solid plastic matrix in a
above daily tests) with a long half-life sol- vial format and include 57Co, 133Ba, 137Cs
id check source in the source holder in the and 60Co. The geometry of these standards
measurement position. The measurements is typically not identical to that of the stan-
are compared to the initial measurements dard sources used by manufacturers to
performed at acceptance testing and the calibrate their systems. The accuracy test

QUALIT Y CONTROL OF NUCLEAR MEDICINE 127


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 9

is not a calibration; it is a test of system be used to determine linearity. The grad-


RA DIO NUC L I D E C A L I BRATO R

stability. Ideally, the test should use stan- ed source method involves manipula-
dards of radionuclides that are employed tion and accurate measurement of stock
by the radionuclide calibrator manufac- solution aliquots. The decaying source
turer to set the system when transporting method is recommended for secondary
calibration settings to production models standard radionuclide calibrators and ref-
(e.g. 57Co and 137Cs or 57Co and 60Co). As- erence radionuclide calibrators. For “field”
certaining which radionuclides were used instruments, the decaying source meth-
by the manufacturer should be part of the od should be used at acceptance testing
purchase process. The use of one source and following repair. The shield method
(e.g. 137Cs) in combination with the routine should be sufficient for annual testing;
“relative response tests” should be suffi- alternatively, the facility can employ the
cient for most medical facilities. For more decaying source method. Measurements
complex programmes, instrument stabili- are taken following the daily tests. The
ty should also be checked annually with decaying source measurements are tak-
at least two traceable reference sources en with the source in the source holder
and the radionuclides used should vary in the measurement position. The graded
from year to year[2]. Measurements of the shielding measurements are taken in ac-
long-lived standards and the two trace- cordance with the shield manufacturer’s
able reference sources should be within instructions. At acceptance testing and
±5% of the decay-corrected initial values. following repair, measurements should be
Secondary standard radionuclide calibra- taken using the decaying source method
tors and reference radionuclide calibrators from the highest activity (highest current)
should be within ±2%[3]. The measure- measured down to approximately 1 MBq.
ments should be recorded and available Annually, measurements should be taken
for regulatory review. between the maximum activity admin-
istered and 1 MBq over the range of use.
System linearity Measurements should be within ±5%
A calibrator is considered linear if the ra- of the expected values. For secondary
tio of the measured response to the pre- standard radionuclide calibrators and ref-
dicted response remains constant over erence radionuclide calibrators, linearity
the range of current inputs. The decaying testing should be performed quarterly us-
source method, the shield method (Fig. ing the decaying source method and the
5) and the graded source method may measurements should be within ±2%[3].

128 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 9

For the shield method, the shields should Documentation

RA DIO NUC L I D E C A L I BRATO R


be calibrated on a radionuclide calibrator Sufficient records need to be maintained
whose linearity using the decaying source to demonstrate proper calibrator oper-
method is within ±5%. The measurements ation, including personnel training and
should be recorded and available for regu- competence testing, and adherence to the
latory review[3]. quality assurance programme. The details
of any calibrator maintenance or repair
Figure 5 should also be recorded[3].
Radionuclide calibrators should be op-
erated in accordance with the manufac-
turer’s instructions. If the manufacturer
recommends additional tests, they should
be performed at the frequency recom-
mended by the manufacturer. Additional
tests are normally limited to tests of the
electronic circuitry and other tests that are
specific to the manufacturer’s systems. As
part of the purchase process, the manufac-
Linearity test kit (shield method): enables user
turer should be asked to document addi-
to simulate decay for a range from a few hours
up to 4 days. (Adopted from CapintecTM) tional routine tests[2].

REFERENCES
1. Cherry SR, Sorenson JA, Phelps ME. Physics in nu- ing 2010;37:672–681.
clear medicine. 4th ed. Philadelphia: Elsevier; 2012. 5. EANM Physics Committee,  Busemann Sokole
2. American Association of Physicists in Medicine. The E,  Płachcínska A,  Britten A;  EANM Working Group
selection, use, calibration, and quality assurance of on Nuclear Medicine Instrumentation Quality
radionuclide calibrators used in nuclear medicine. Control,  Lyra Georgosopoulou M,  et al. Routine
Report of AAPM Task Group 181, June 2012. quality control recommendations for nuclear med-
3. International Atomic Energy Agency. Quality as- icine instrumentation. Eur J Nucl Med Mol Imaging
surance for radioactivity measurements in nuclear 2010;37:662–671.
medicine. Technical Report Series No. 454. Vienna: 6. Measurement Good Practice Guide No.93; Protocol
IAEA; 2006. for Establishing and Maintaining the Calibration of
4. Busemann Sokole E, Płachcínska A, Britten A; EANM Medical Radionuclide Calibrators and their Quality
Physics Committee. Acceptance testing for nuclear Control; National Physical Laboratory, May 2006,
medicine instrumentation. Eur J Nucl Med Mol Imag- ISSN: 1368-6550.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 129


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
1
10
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
PROBE S A N D ORG A N
IN T R AO P E R AT I V E
GAMMA PRO B E S:

UPTA KE PRO B E S

by Kristof Baete
CHAPTER 10

EQUIPMENT AND and collimator, which makes it larger and


PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E

BASIC PRINCIPLES heavier. Therefore, the probe is usually


A gamma probe is a radiation detector mounted to a fixture or a holder, and al-
that allows for the detection and locali- lows for the measurement of a radionu-
sation of gamma ray-emitting radionu- clide in vivo using a standardised source–
clides[1, 2]. From this general concept, two detector setup. The most common appli-
device types have been derived for in vivo cation of this device is the determination
application in nuclear medicine, without of iodine uptake in the thyroid[5]. However,
the need for imaging: (a) the intraop- an organ uptake probe can also be used
erative probe and (b) the organ uptake for whole-body dosimetry in molecular ra-
probe. The clinical use of these devices diotherapy or as a measuring instrument
is different, but the principles of opera- for isolated limb perfusion.
tion are quite similar and require a largely
equivalent quality management (QM). The gamma probe, in general, needs to
be well shielded to protect the detector
The intraoperative gamma probe is a against unwanted radiation from its sur-
compact and hand-held device for conve- roundings. In front of the detector, a prima-
nient localisation of radionuclides in the ry collimator usually allows the radiation to
body during surgical interventions[3]. The enter the detector along a certain aperture
technique, for which a probe is used, is of- angle. Manufacturers tend to offer a variety
ten referred to as radio-guided surgery. The of intraoperative probe units, and addi-
main application is the sentinel lymph node tional or exchangeable collimators. Organ
(SLN) procedure, in which a gamma probe uptake probes, on the other hand, are typ-
can not only detect and localise an SLN but ically manufactured with a dedicated wide
also reveal occult disease and allows mini- or flat field collimator for reliable operation.
misation of surgical invasiveness[4]. For that The detector unit of a gamma probe is
purpose, specific probe shapes have been usually connected to a console or readout
designed, e.g. with a bent tip or laparoscop- unit. For better agility, some intraoperative
ic units. Intraoperative beta probes also ex- probes have detectors that are wirelessly
ist. However, the specific quality aspects are connected to the readout unit. The settings
beyond the scope of this chapter. on the console of an intraoperative probe
are usually limited, and primarily allow for
The organ uptake probe, on the other selection of a radionuclide, count rate in-
hand, uses a more optimised detector dicator and sound settings. The acquisition

132 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 10

console of an uptake probe is generally able electrical signal. Semiconductor de-

PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E
much more extensive. tectors are usually fabricated out of crys-
talline materials, such as CdTe, CdZnTe (or
Two types of radiation detector are used CZT) and HgI2.
for the manufacture of gamma probes: (a)
scintillation detectors, or scintillators, and Both detector types enable energy dis-
(b) semiconductors. crimination of the radiation and allow for
the selection of upper and lower energy
A scintillator is mostly a crystal that is level discriminators. Events that are re-
able to absorb incident gamma rays. The corded within the energy range can be ac-
energy of the charged particles, created cepted for further processing, while events
by the ionisation of the incident gamma outside of that range can be rejected or
rays, is re-emitted as scintillation light. used for the purpose of scatter correction.
This happens such that the light output In uptake probes, the entire energy spec-
of the crystal is proportional to the energy trum can be recorded.
of the incident gamma ray. Subsequently,
a light detector converts the scintillation The higher density and effective atom-
light to a measurable signal that is again ic number of inorganic scintillators allow
proportional to the energy of the incident for a better detector efficiency at higher
gamma ray. Scintillator crystals for gam- energies. However, scintillators in combi-
ma probes are usually inorganic materials, nation with a PMT are sensitive to environ-
such as NaI(Tl) or CsI(Tl). For higher ener- mental changes and therefore have a less
gy applications, BGO, LSO or GSO may be stable energy calibration. Scintillator-based
used owing to their better stopping pow- probes with a photodiode have shown an
er. Photomultiplier tubes (PMTs) or photo- improved ruggedness compared with PMT-
diodes are used for the detection of the based systems. Nevertheless, scintillators
scintillation photons. have a poorer energy resolution than semi-
conductors; this means that a semiconduc-
In a semiconductor, the incident radia- tor has a better scatter rejection capability.
tion is measured directly by the amount
of electron-hole pairs that are created in
proportion to the energy of the gamma QUALITY ASSURANCE
ray. The charge carriers are collected using The international Basic Safety Standards,
electrodes which then produce a measur- laid down in European Council Directive

QUALIT Y CONTROL OF NUCLEAR MEDICINE 133


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 10

2013/59/Euratom, recommend that for all For intraoperative probes, it is very


PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E

medical radiological equipment a quality important to follow the manufacturer’s


assurance (QA) programme needs to be recommendations regarding instrument
in place in order to uncover defects and cleaning, disinfection and sterile operation.
to prevent procedural mistakes during the Ignoring the advice can cause serious sys-
execution of tasks. For the verification of tem damage. Manufacturers may also pro-
good equipment manufacturing, such a vide specific directions for wireless detec-
QA programme should include at least: tor probe units, such as recharging of bat-
(a) the execution of formal acceptance teries and how to safely store probe units.
testing of the instrumentation and (b) the
installation of a QC programme. The QA QA and QC recommendations for gamma
should also involve good registration and probes in general are available in the litera-
documentation of all information related ture[7–10]. More specific guidelines for intra-
to the use of the system. A good foun- operative probes exist[11–13], as do minimum
dation for the QM is described in the IEC system performance requirements[14, 15].
61223-1 technical report [6].

The issue of protocol standardisation ACCEPTANCE AND REFERENCE


for the clinical use of gamma probes is TESTS
a further aspect of QA. The sensitivity of
an uptake probe, for example, relies par- Before a gamma probe can be used in
ticularly on the source–detector distance clinical practice, its performance should
and it is important to know the exact data be assessed by a thorough verification of
acquisition requirements for acquisition equipment specifications. This assessment
of reliable results. Therefore, standard op- is done by a medical physics expert. The
erational procedures should be available acceptance test results are the reference
to all end-users. Procedural mistakes or values for the QC tests that will be execut-
pitfalls may include the misuse of colli- ed periodically from the moment of com-
mators, isotope or energy settings, and missioning. Therefore, the acceptance
display or audible settings. It is important tests should be documented well and ap-
that all aspects related to the routine use pended to the equipment log.
of the device are clarified in written pro-
cedures that have been explained well to Acceptance tests are usually more com-
the end-user. prehensive than QC tests. A performance

134 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 10

analysis encompasses more than those The power supply of the device is crit-

PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E
aspects that need to be verified once the ical and should be verified. Also, the bat-
equipment is in routine use. Performance tery-supplied intraoperative probe unit
tests for gamma probes have been pro- should be checked. The energy response
posed in the literature[7, 11, 13]. The most of an uptake probe depends on the stabil-
common acceptance and reference tests, ity of the high voltage and detector gain,
in general, are: but also on the environmental conditions,
such as temperature. Therefore, it is ex-
»» Physical inspection, shielding tremely important to assess the detector
»» Power supply, high voltage, energy calibration or peaking. After recep-
detector gain, battery tion, the energy settings should be verified
»» Energy peaking or calibration, for all used radionuclides, and if necessary
energy settings the instrument should be (re)calibrated.
»» Background count rate For an uptake probe, direct spectral analy-
»» Energy resolution sis of the energy peak of a reference source
»» Sensitivity (e.g. caesium-137) is usually obvious. How-
»» Counting precision ever, visualisation of the energy spectrum
»» Linearity of the response of an intraoperative probe by the end-user
is not always possible.
Some device-specific acceptance tests
should be added, such as the source ge- The background count rate of a gamma
ometry influence for uptake probes and probe is an essential property. Each mea-
spatial resolution for intraoperative probes. surement should in fact be preceded or
followed by a reading of the background
First of all, physical inspection of the count rate. This system parameter de-
probe should be carried out immediately af- pends on the energy settings. For any of
ter its reception. Any problem or deficiency these settings, a typical background count
observed during the warranty period should rate should be analysed and reference val-
be reported and repaired at once. Physical ues should be determined. It is important
impact to the detector, or its shielding, can to perform a follow-up of background
lead to malfunctioning or changes in effi- measurements for various detector loca-
ciency. For intraoperative probes, the signal tions and positions. In addition to the am-
from a nearby injection site, caused by a leak bient conditions, the electronics and con-
in the shielding, may conceal an SLN. necting cables can have an impact.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 135


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 10

The energy resolution of a gamma sion. The decay of a radioactive source is a


PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E

probe indicates how well it is able to dis- random process that obeys Poisson statis-
tinguish between closely spaced energy tics. A chi-square test can be used to verify
peaks of detected gamma rays in the en- that the short-term detector sensitivity is
ergy spectrum. A good energy resolution in accordance with the Poisson distribu-
enables the gamma probe to reject scat- tion [1]. With a series of measurements, a
tered radiation. Hence, measuring energy chi-square value can be computed and
resolution can identify the potential for a converted to a probability. A high prob-
good system performance. The energy ability (>0.99) means that the variations
resolution of the probe is quantified by in the measurements are smaller than
the full-width at half-maximum (FWHM) statistically expected. This may indicate
of the gamma peak at a certain energy. In that, for example, periodic noise is being
uptake probes, FWHM can be reported by measured instead of a radioactive source.
the console as part of the energy calibra- A low probability (<0.01) indicates that
tion or peaking procedure. the measurements are too irregular to
follow the Poisson nature of the source,
An important parameter is the sensitiv- and there may be something wrong with
ity of the gamma probe. In a standardised the device. A probability of 0.5 indicates a
setup, with a reproducible source-to-de- perfect detector behaviour according to
tector distance and a reference source Poisson statistics.
activity, the count rate response can be
determined for the available energy set- The linearity of the detector response,
tings. The results of these sensitivity mea- or the count rate capability, verifies the
surements are the reference values for the sensitivity at different levels of source ac-
subsequent QC tests. These QC measure- tivity. Radiation detectors are influenced
ments are therefore usually referred to as by a number of effects at increased count
a test of the constancy or the long-term rates. Some of these effects can lead to a
stability of the sensitivity. For a more re- shift of the measured energy spectrum.
alistic analysis, sensitivity may also be as- Even though this effect is small for most
sessed taking into account the influence applications, it should be recognised and
of scatter [11]. measured at least during acceptance test-
ing. The most important effect, however,
The short-term stability of the sensitivity is dead time or the loss of counts. With a
is often referred to as the count rate preci- decaying source, for example, one can

136 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 10

compare the observed detector count rate such that the QC results of the device are

PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E
with the true count rate of the source. brought back in line with the reference
values. Moreover, trend analysis of the QC
For the uptake probe, a specific accep- results is a powerful tool for observation of
tance test should be added. Since the in- the performance of the instrument over
strument is used to quantify an amount of time.
activity, it is important to know the influ-
ence of different source geometries on the QC recommendations and test fre-
response of the detector. The result of this quency proposals have been described
assessment reveals the robustness of the in a number of publications [8–10]. The most
measurement setup and source (i.e. organ) important routine QC tests for gamma
positioning. probes, in general, are:

For the intraoperative gamma probe, »» Physical inspection


the directional or angular detector sensi- »» Power supply, high voltage, gain, battery
tivity is very important. A narrow measure- »» Energy peak verification
ment field, using an appropriate collima- »» Background count rate
tor, allows for a high signal to background »» Sensitivity or constancy
ratio and SLN detectability. In addition, the »» Counting precision
lateral detector sensitivity defines the spa-
tial resolution of the probe and the ability Physical inspection of the gamma probe
to separate neighbouring lesions[12]. should be performed at regular inter-
vals, preferably before each use. The user
should be aware that any observation or
QUALITY CONTROL knowledge of imprudent use, e.g. a drop
The acceptance and reference tests are or collision of the detector unit, can result
complemented by the subsequent QC in system malfunctioning or changes in
programme. This part of the overall QM the response. Such events should be re-
focusses on the fulfilment of the quality re- ported immediately and registered in the
quirements or acceptability criteria by the instrument log, and their impact analysed
performance of verification tests in clinical by a quality inspection of the shielding
practice, at regular intervals. If the criteria and detector sensitivity and verification of
are not met, corrective actions should be the energy settings. Also, the power sup-
imposed by the medical physics expert, ply should be checked before each use,

QUALIT Y CONTROL OF NUCLEAR MEDICINE 137


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 10

especially for those mobile units that use property that should be observed prior
PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E

batteries. to any use. There is the risk of contami-


nation of the instrument, or its surround-
Regular verification of the energy set- ings, or the influence of an unexpected
tings of a gamma detector is preferred. local radioactive source. In addition, there
For intraoperative probes, the energy may be a sudden change in the detector
spectrum analysis is unfortunately not background reading due to, for example,
always available to the end-user. On the electronic noise. The observation of an
other hand, daily verification of the energy elevated background should always be
peak for quantitative measurements with investigated before proceeding.
uptake probes is easy and highly recom-
mended. Moreover, a trend analysis of the The follow-up of detector sensitivity, as
energy settings has the power to reveal determined during acceptance testing,
subtle performance changes or drift. The is known as the constancy test. This QC
same holds true for the energy resolution. test verifies whether the sensitivity has
A sudden increase in the energy resolu- changed, and it is an easy test to assess
tion may be a sign of crystal damage (e.g. influences on the detector response, for
a crack) or decoupling of the light detec- example, due to incorrect detector energy
tor (PMT, photodiode). A more gradual settings or the condition of the hardware.
broadening of the energy resolution may As explained above, counting precision
indicate crystal yellowing due to humidity analysis may reveal subtle detector issues.
or a deterioration in the light detector.
In conclusion, carrying out sustainable
As mentioned above, the background QM leads to the correct and safe use of
count rate of a probe is an important gamma probes in clinical routine.

138 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 10

PR O BE S A ND O R GA N UPTA KE PR O BE S
GAMMA PR O BE S: I NTRAO PE RATI V E
REFERENCES

1. Cherry SR, Sorenson JA, Phelps ME, eds. Physics in quality control recommendations for nuclear med-
nuclear medicine. 4th ed. Philadelphia: Saunders; icine instrumentation. Eur J Nucl Med Mol Imaging
2012. 2010;37:662–671.
2. Bailey DL, Humm JL, Todd-Pokropek A, van Aswegen 9. 
Quality control of nuclear medicine instruments.
A. Nuclear medicine physics. Vienna: International IAEA TECDOC 602. Vienna: International Atomic En-
Atomic Energy Agency; 2015. Available from: http:// ergy Agency; 1991.
www-pub.iaea.org/books/IAEABooks/10368/Nucle- 10. International Electrotechnical Commission. Nuclear
ar-Medicine-Physics medicine instrumentation – Routine tests – Part 1:
3. Heller S, Zanzonico P. Nuclear probes and intra- Radiation counting systems. 2001. IEC/TR 61948-1.
operative gamma cameras. Semin Nucl Med 2011; http://www.iec.ch.
41:166–181. 11. NEMA NU-3. Performance measurements and qual-
4. Povoski SP,  Neff RL,  Mojzisik CM,  O’Malley DM,  Hin- ity control guidelines for non-imaging intraopera-
kle GH, Hall NC, et al. A comprehensive overview of tive gamma probes. Rosslyn, VA: National Electrical
radioguided surgery using gamma detection probe Manufacturers Association; 2004.
technology. World J Surg Oncol 2009;7:11. 12. Wengenmair W, Kopp J, Sciuk J. Quality criteria of
5. Hänscheid H, Canzi C,  Eschner W,  Flux G,  Luster gamma probes: Requirements and future develop-
M,  Strigari L,  Lassmann M. EANM Dosimetry Com- ments. In: Shauer AJ, Becker W, Reiser MF, Possinger
mittee Series on Standard Operational Procedures K, eds. The sentinel lymph node concept. Berlin Hei-
for Pre-Therapeutic Dosimetry – II. Dosimetry prior to delberg New York: Springer; 2005.
radioiodine therapy of benign thyroid diseases. Eur J 13. Chiesa C, Toscano F, Mariani M, Bomnardieri E.
Nucl Med Mol Imaging 2013;40:1126–1134. Physical performance parameters of intraoperative
6. International Electrotechnical Commission. Evalua- probes. In: Mariani G, Giuliano AE, Strauss HW, eds.
tion and routine testing in medical imaging depart- Radioguided surgery: A comprehensive team ap-
ments – Part 1: General aspects. 1993; IEC/TR 61223- proach. New York: Springer; 2008.
1. http://www.iec.ch. 14. Matheoud R, Giorgione R,  Valzano S,  Sacchetti
7. Busemann Sokole E,  Płachcínska A, Britten A; EANM G,  Colombo E,  Brambilla M. Minimum acceptable
Physics Committee. Acceptance testing for nuclear sensitivity of intraoperative gamma probes used
medicine instrumentation. Eur J Nucl Med Mol Imag- for sentinel lymph node detection in melanoma
ing 2010;37:672–681. patients. Phys Med 2014;30:822–826.
8. EANM Physics Committee,  Busemann Sokole 15. European Commission. Radiation Protection 162.
E,  Płachcínska A,  Britten A;  EANM Working Group Criteria for acceptability of medical radiological
on Nuclear Medicine Instrumentation Quality equipment used in diagnostic radiology, nuclear
Control,  Lyra Georgosopoulou M,  et al. Routine medicine and radiotherapy. 2012.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 139


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
1
11
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
COUNT E R S
WEL L

by Petra Kolenc Peitl


and Marko Krošelj
CHAPTER 11

PRINCIPLES OF RADIATION energy of radiation, detector volume,


W E L L CO UNTE RS

DETECTION source geometry and detector dead time.


Radiation is detected through its inter-
action with matter. When ionising radi- Gas-filled detectors
ation interacts with matter, ions can be Gas-filled detectors include ionisation
produced directly or indirectly. Alpha or chambers, proportional counters and Gei-
beta particles are highly energised parti- ger-Müller (GM) counters. The detector
cles that interact directly with atoms and consists of gas contained between two
produce charged ion pairs. On the other electrodes with a suitable high voltage
hand, X-rays and gamma photons interact difference between them. Gas is normally
primarily with orbiting electrons, which an insulator and no electrical current flows
are ejected and then interact with other between the electrodes. When ionising
atoms, producing charged ions[1]. radiation passes through the chamber, it
There are two basic principles of radia- causes ionisation of the gas molecules,
tion detection: producing negatively charged electrons
»» Ionisation is caused when the radiation and positively charged gas ions. Electrons
is passing through. The ions produced are attracted to the positively charged
are detected and measured. electrode (anode) and ionised atoms to
»» Electronic excitation in atoms or mol- the negative electrode (cathode), causing
ecules is caused by the radiation. The a small amount of electrical current which
excess of energy is then dissipated and is proportional to the number of ionising
detected and measured. photons or particles in the chamber. To
achieve maximum efficiency of gas-filled
detectors, the voltage between the elec-
GAMMA AND BETA RADIATION trodes must be optimal so as to ensure
DETECTORS complete collection of electrons and ions
Three types of detector are used for the (Fig. 1). If the voltage is too low, recombi-
quantitative determination of radioactiv- nation of some of the ions and electrons
ity: gas-filled ionisation detectors, scintil- occurs, with no contribution to electrical
lation detectors and semiconductors. No flow. As the voltage is increased (to 50–
detector is suitable for universal detection 300 V), all of the primary charges arising
of all types of radiation. The factors influ- from the initial ionising event are collect-
encing the choice of detector are sensitiv- ed (ionisation chamber region). Moreover,
ity, response to different radiation types, increasing the voltage further (to approx.

142 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 11

Figure 1

W E L L CO UNTE RS
Voltage-response curve for gas-filled detectors

300–1000 V, depending on the detector in some materials a portion of the energy


gas) will allow the moving electrons to is released as a weak flash of light (scintilla-
become ionising particles themselves, tion). When such materials (scintillators or
and the signal is amplified (proportional scintillants) are used in the detection of ra-
counter). With a voltage of 1000–2000 V diation, the detectors are called scintillation
(depending on geometry) an avalanche is detectors. Scintillators can be divided into
created and the resulting pulse becomes two general types: inorganic materials (sol-
independent of the size of the original ion- id crystals) and organic materials dissolved
ising event that triggered the signal (Gei- in liquids. Organic liquids and plastic scintil-
ger-Müller counter)[1, 2]. lators are usually used for the detection of
beta particles and fast neutrons. The higher
Scintillation detectors density and atomic number of inorganic
After the radiation causes excitation of at- crystals usually lead to a better detection
oms and molecules, the energy is released efficiency, so these scintillators are used for
during the de-excitation. Most of the ener- the detection of X-rays and gamma rays[3].
gy is released as thermal energy; however, The most common crystal scintillators

QUALIT Y CONTROL OF NUCLEAR MEDICINE 143


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 11

used in nuclear medicine instrumentation radiation detection: linear response with


W E L L CO UNTE RS

are summarised in Table 1. The amount of deposited energy, negligible absorption


light produced when gamma rays or beta of energy in the entrance window of the
particles interact with a scintillator is pro- detector, excellent energy resolution, for-
portional to the energy deposited by the mation of pulses with fast rise times and
radiation in the scintillation material. These small detector sizes. The signal formation
scintillations are detected by the light cell in a semiconductor detector is very similar
or photomultiplier (PM), which converts to that in an ionisation chamber. Ionisa-
light energy into weak electric currents tion is produced within the sensitive vol-
that are amplified and converted into volt- ume of the detector and converted to a
age pulses. The heights of these pulses are voltage pulse that is amplified and count-
related to the radiation energy and the ed[4]. The most commonly used semicon-
number of pulses is related to the activity ductor materials are elemental silicon (Si)
of the radioisotope[1, 3]. and germanium (Ge). Both Si and Ge have
a limited use in the detection of X-rays and
Semiconductor detectors gamma rays: (a) they need ancillary cool-
In comparison to other materials, semi- ing systems due to a significant amount
conductors have unique characteristics of thermally induced electrical current at
that make them very suitable for ionising room temperature; (b) they have mod-

Table 1: Basic characteristics of common scintillator crystals used in


nuclear medicine applications [1, 3]
Crystal Atomic Density Hygroscopic Rugged Decay Emission
number, Z (g/cm3) constant wavelength
(ns) (nm)
NaI(Tl) 51 3.67 Yes No 200 410
BaF2 53 4.88 No Yes 0.8 220 and 310
Csl(Tl) 54 4.51 Slightly Yes 1000 565
No (cleaves
Gd2SiO5(Ce) (GSO) 59 6.71 No 60 430
easily)
Lu2SiO5(Ce) (LSO) 65 7.40 No Yes 40 420
Bi4Ge3O12 (BGO) 75 7.13 No Yes 300 480

144 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 11

Figure 2

W E L L CO UNTE RS
Left panel: Schematic diagram of well counter.
Right panel: Most automated multiple sample systems use a so-called through-hole detector

est stopping powers and (c) impurities studies of glomerular filtration rate, cere-
are present even in relatively pure crystal brospinal fluid leak, red cell mass, plasma
and capture electrons released in ionisa- volume and radioimmunoassay tests with
tion, leading to lower detection efficiency. 125
I-radiolabelled probes[7,8]. Preclinically,
Compound semiconductors, such as cad- well counters are used in in vitro cell stud-
mium zinc telluride (CZT), can overcome ies, biodistribution studies, determination
these drawbacks[2,5]. CZT detectors have of radionuclidic impurities (e.g. determi-
been applied to various medical imaging nation of 68Ge in the 68Ge/68Ga generator
modalities in CT, PET and SPECT[6]. eluate) etc. Well counters usually comprise
a well-type detector, which surrounds the
sample and therefore provides high detec-
NaI(Tl) WELL COUNTERS tion efficiency (Fig. 2). A photomultiplier
Well counters are an important instrument tube (PMT) with associated electronics
in nuclear medicine since they play a key backs the detector. Very high detection ef-
role in preclinical and clinical practice. The ficiency of the NaI(Tl) well counters limits
most important clinical applications are the amount of activity that can be count-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 145


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 11

ed (~37 kBq). With higher levels of activity, by self-absorption, even with large crys-
W E L L CO UNTE RS

problems regarding dead time can be en- tal sizes.


countered. Dead time is the time required »» It is relatively inexpensive.
to process individual detected events. The »» The scintillation light is well matched in
shorter the dead time, the smaller are the wavelength to the peak response of the
dead time losses. NaI(Tl) and semiconduc- PM photocathode.
tor detector systems usually have a dead
time in the range of 0.5–5 μs[2]. However, NaI(Tl) detectors also have
some disadvantages: the crystal is quite
The majority of well counters employed fragile and easily damaged; NaI is hygro-
in nuclear medicine use thallium-activat- scopic, which can impair light transmis-
ed sodium iodide [NaI(Tl)] crystals as scin- sion to the PM upon exposure to moisture;
tillators. NaI(Tl) detectors enable detection and at energies >250 keV a large volume
of the majority of the gamma-emitting of NaI(Tl) is required for adequate detec-
radionuclides in nuclear medicine (e.g. tion efficiency.
99m
Tc, 123I, 111In, 201Tl). Since pure sodium io- The efficiency of detection depends on
dide crystals are scintillators only at liquid the following factors:
nitrogen temperatures, thallium is added »» The energy of gamma photons:
in small amounts as a so-called activator high-energy photons can pass com-
centre in the crystal matrix and is respon- pletely through the crystal matrix with-
sible for the scintillation effect at room out energy dissipation.
temperature. There are many advantag- »» Crystal size: larger crystals absorb
es of NaI(Tl) which make these detectors high-energy photons more efficiently,
suitable for almost all routine applications but the size of the crystal is limited by
in nuclear medicine: the optical transmission of scintillation.
»» Due to its relatively high density, NaI(Tl) »» Geometry: it is essential that samples
is a good absorber with high detection are counted using the same volumes,
efficiency for penetrating radiation in the same size tubes and always in the
(X-rays and gamma rays in the energy same position within the counter.
range from 50 to 250 keV). »» Photomultiplier high voltage: PMTs have
»» It is a relatively efficient scintillator, since to be supplied with an optimal and sta-
it yields one visible light photon per 30 ble source of high voltage. At too low a
eV radiation energy absorbed. voltage, no electrical pulse is produced
»» Little loss of scintillation light is caused in PMTs. At increased voltage, sponta-

146 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 11

Figure 3

W E L L CO UNTE RS
Left panel: Example of an automated multisample well type counter.
Right panel: Example of computerised routine QC tests

neous electron emission and thermal pendently from the sample volume. Sys-
noise contribute to an increased count tems with an automated sample arm can
rate[1, 9]. save a lot of time, since the sample vials
The overall advantages of NaI(Tl) make can be loaded into a rack and counted
these detectors the first choice for almost automatically. One disadvantage of auto-
all routine applications in nuclear med- mated systems is that they are not as well
icine involving the detection of gamma shielded as manual well counters, since
rays in the 50–250 keV energy range. there is no lead shielding directly on the
In daily practice, rather than manual top of the sample being counted. This
well counters, automated multiple sam- can increase background counting rates,
ple systems are used (Fig. 3). Most such particularly from the other samples in the
systems use a so-called through-hole rack, which can be problematic, especial-
detector, where the sample hole pass- ly when low-activity samples are counted
es throughout the NaI(Tl) crystal and together with high-activity samples in the
the PMT is connected to the side of the rack.
scintillator (Fig. 2). Samples in such a de- For very high throughput one can use
tector can be automatically positioned multidetector systems which can contain
at the centre of the NaI(Tl) crystal, inde- multiple NaI(Tl) scintillation detectors. This

QUALIT Y CONTROL OF NUCLEAR MEDICINE 147


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 11

allows multiple samples to be counted strict acceptance testing after installation


W E L L CO UNTE RS

simultaneously. The individual detectors to verify that the instrument meets spec-
are separated and shielded from each ifications and performs according to its
other to prevent crosstalk, which can give clinical purpose. Acceptance testing re-
rise to substantial counting error when sults are the reference data for future QC
counting high-energy gamma rays. Back- tests, and some of them may be repeated
ground measurements in one detector periodically (every 6 or 12 months) or after
while counting a sample in an adjacent a major service or component change[14].
one can be used to estimate the extent of At the time of installation of a well count-
the crosstalk[2]. er, physical inspection, energy window
Because of the high-energy photons, calibration and checking of the energy
in the case of PET applications the NaI(Tl) resolution, sensitivity and linearity are per-
crystal thickness needs to be optimised formed as acceptance testing[15].
and the shielding around the detector Since the electronic components and
has to be sufficient to prevent crosstalk detectors of well counters can fail or de-
from either the samples in adjacent rack teriorate over time, a good quality assur-
positions or the adjacent detector in the ance programme should be employed
case of multidetector systems[10,11]. Re- to ensure consistently accurate results.
cently, a high-sensitivity well counter with The routine QC tests for well counters in-
a bismuth germanate (BGO) detector clude energy window calibration (energy
was developed for preclinical animal PET peaking), measurement of background,
investigations in which quantitative mea- energy resolution, constancy, efficiency,
surements of small blood volumes and reproducibility (chi-square test), linearity
activities are needed[12]. and minimal detectable activity[13,14]. Table
2 provides an overview of basic QC tests,
their purpose and their frequency. Some
QUALITY CONTROL of the tests are described in detail below:
Quality control (QC) is defined as an es- »» When a well counter is equipped with
tablished set of ongoing measurements a multi-channel analyser, the energy
and analyses designed to ensure that the spectrum should be checked to verify
performance of a procedure or instrument that the photopeak of the radionuclide
is within a predefined acceptable range[13]. coincides with the preset photopeak
Every nuclear medicine instrument, in- energy window. Usually, energy peaking
cluding well counters, must undergo is done with a 20% photopeak energy

148 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 11

Table 2: Summary of routine QC tests performed for well counters

QC test Frequencya Purpose

W E L L CO UNTE RS
Background Daily To detect contamination and to determine background
radiation
Constancy Daily Day-to-day counting of long-lived reference sources
Energy window Daily / quarterly To verify that the photopeak of the radionuclide coincides
calibration with the preset photopeak energy window
Energy resolution Quarterly To evaluate the sharpness of a photopeak of reference
source radionuclide
Reproducibility Quarterly / To check variations in the set of measurements of reference
half-yearly source (chi-square test)
Efficiency Annually To determine the sensitivity (ε: efficiency) in cpm/Bq for each
radionuclide
a
The type of test and the frequency depend on the manufacturer’s instructions, national legislation and
different guidelines.

window, equivalent to 90–110% of the »» Energy resolution evaluates the sharp-


gamma ray energy of the radionuclide. ness of a photopeak produced when
»» The background counting rate should be a well detector is exposed to a single
checked daily before use for each photo- energy radionuclide. It is expressed as
peak energy window. It is of utmost im- the percentage full-width at half-maxi-
portance to determine the level of back- mum (FWHM) and is checked with a ref-
ground radiation, which can be relatively erence source, such as 137Cs. Widening
high and variable in nuclear medicine of the peak indicates a malfunction of a
departments. Measuring a blank sample certain system component (e.g. break-
(empty tube) will also show any contami- down of the crystal to photomultiplier
nation of the counting well. tube seal).
»» Constancy: at least one reference source »» Calibration for each radionuclide should
traceable to a certified metrology insti- be performed at installation, annually
tute should be counted daily. Day-to- and after any major repair. Efficiency in
day counting rates should be ±10%. cpm/Bq can be determined using a pre-
Long-lived radionuclides, such as 57Co, cisely calibrated sample, prepared by ap-
68
Ge, 129I and 137Cs, can be used as refer- propriate dilution of a sufficiently large
ence sources for well counter constancy activity of radionuclide to allow accurate
determination. measurement in a radionuclide cali-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 149


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 11

brator. On the other hand, a surrogate or by aliquoting the initial solution with
W E L L CO UNTE RS

radionuclide (reference standards, e.g. known radioactivity.


57
Co instead of 99mTc or 68Ge instead of »» Minimum detectable activity (MDA) is
18
F) can be used to routinely determine the smallest amount of activity detect-
the efficiency. In this case, the factor for able from background which can be
conversion of the efficiency of the surro- quantified at a given confidence level
gate to that of a particular radionuclide of 95%.
has to be determined at installation of
the well counter. Whichever tests are performed in the
»» Reproducibility of a well counter is quality assurance programme for a well
checked with the chi-square test, which counter, it is essential that the procedures
shows whether random variations in a are documented and the results are re-
set of measurements are consistent with corded (Fig. 3). Test methods and results
the expected Poisson distribution. Any in- have to be maintained in electronic or
consistency with a Poisson distribution is hard-copy form and archived in order to
indicative of a well counter malfunction. accurately compare all QC tests with the
»» Linearity of activity response is checked acceptance test results[14–16].
either with the decaying source method

150 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 11

W E L L CO UNTE RS
REFERENCES

1. Theobald T. Sampson’s textbook of radiopharmacy. http://hidex.com/products/gamma-counters/hi-


London: Pharmaceutical Press; 2010. dex-automatic-gamma-counter/.
2. Cherry SR, Sorenson JA, Phelps ME. Physics in nuclear 11. Lodge MA, Holt DP, Kinahan PE, Wong DF, Wahl
medicine. 3rd ed. Philadelphia: Saunders; 2003. RL. Performance assessment of a NaI (Tl) gamma
3. Melcher CL. Scintillation crystals for PET. J Nucl Med counter for PET applications with methods for im-
2000;41:1051. proved quantitative accuracy and greater standard-
4. Seco J, Clasie B, Partridge M. Review on the charac- ization. EJNMMI Phys 2015;2:11.
teristics of radiation detectors for dosimetry and im- 12. Yamamoto S, Watabe H, Kanai Y, Watabe T, Imai-
aging. Phys Med Biol 2014;59:R303. zumi M, Shimosegawa E, et al. Development of a
5. Owens A, Peacock A. Compound semiconduc- high-sensitivity BGO well counter for small animal
tor radiation detectors. Nuclear Instruments and PET studies. Radiol Phys Technol 2012;5:59–62.
Methods in Physics Research. Section A: Accelerators, 13. Zanzonico P. Routine quality control of clinical nu-
Spectrometers, Detectors and Associated Equipment clear medicine instrumentation: a brief review. J
2004;531:18–37. Nucl Med 2008;49:1114–1131.
6. Iniewski K. CZT detector technology for medical im- 14. Busemann Sokole E, Płachcínska A, Britten A, Lyra
aging. J Instrum 2014;9:C11001. Georgosopoulou M, Tindale W, Klett R. Routine
7. Goldsmith SJ. Radioimmunoassay: Review of basic quality control recommendations for nuclear med-
principles. Semin Nucl Med 1975;5:125–152. icine instrumentation. Eur J Nucl Med Mol Imaging
8. Fairbanks VF, Klee GG, Wiseman GA, Hoyer JD, Tefferi 2010;37:662–671.
A, Petitt RM, et al. Measurement of blood volume and 15. Busemann Sokole E, Płachcínska A, Britten A. Ac-
red cell mass: re-examination of 51Cr and 125I meth- ceptance testing for nuclear medicine instrumen-
ods. Blood Cells Mol Dis 1996;22:169–186. tation. Eur J Nucl Med Mol Imaging 2010;37:672–681.
9. Lubran MM. Quality control of gamma counters. I. 16. Alessi AM. Quality assurance programs for up-
Experimental evaluation of sources of error. Ann Clin take probes and well counters. 2014 [23.2.2017].
Lab Sci 1977;7:57–67. Available from: http://www.biodex.com/nucle-
10. Hidex. Automatic gamma counter for nuclear med- ar-medicine/blog/quality-assurance-programs-up-
icine and PET. 2015 [23.2.2017]. Available from: take-probes-and-well-counters.

QUALIT Y CONTROL OF NUCLEAR MEDICINE 151


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
1
12
QUA L I T Y CON TROL OF NUC LE A R ME DIC INE
I N S TRUME N TATION AND PROTO CO L S TA NDA RDIS AT IO N
MEAS U R E M E N T
PROT E C T I O N
EQUI PM E N T
R ADIAT I O N

by Søren Holm
CHAPTER 12

OVERVIEW OF PHYSICAL er most applications in nuclear medicine.


ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

ENTITIES AND UNITS IN For alpha particles a value of 20 is applied


DOSIMETRY FOR RADIATION in radiation protection. To distinguish the
PROTECTION equivalent dose from the absorbed dose,
The fundamental entity in radiation do- the unit of HT is changed to sievert (Sv).
simetry is the absorbed dose D (and dose To establish a single “risk parameter”, the
rate dD/dt). This is a purely physical entity, values of HT are finally weighted togeth-
defined as the energy E (unit joule, J) ab- er using tissue factors wT that reflect in-
sorbed in a small (tissue) volume divided dividual organs’ radiation sensitivity. The
by the mass m (unit kg) of that volume. resulting entity is the effective dose H (unit
The unit of D therefore is J/kg, but is given Sv), which is used when setting limits for
the special name gray (Gy). The biological whole-body exposure. The precise defini-
effects of radiation depend not only on tions and the tables of wR and wT can be
the amount of energy imparted, but also found in ICRP 103[1] and the IAEA Hand-
on its distribution at the microscopic lev- book on Nuclear Medicine Physics[2]. The
el, which in turn depends on the charac- dosimetry concepts are summarised in
teristics of the radiation. Therefore the ab- Table 1.
sorbed dose is averaged over an organ or Effective dose is not directly measur-
a tissue (T) and weighted by a factor, the able even in a uniform external field. For
radiation weighting factor wR, to give the practical use in radiation protection the
equivalent dose HT. The factor wR equals following “operational” entities have been
one for photons (X-ray and gamma) and defined (for details, see IAEA): Ambient
for electrons (beta- or beta+), which cov- dose equivalent H*(10) (unit sievert) corre-

Table 1: Entities and units in radiation measurement and protection

Name Activity Absorbed Radiation Equivalent Tissue Effective


dose weighting dose weighting dose
factor factor
Symbol A D wR HT wT E
Unit becquerel, gray, Gy Sv/Gy sievert, Sv No unit sievert, Sv
Bq 1 Gy = = 1 for beta 1 Sv = (sum of all 1 Sv =
1 J/kg and gamma 1 J/kg wT equals 1) 1 J/kg
= 20 for alpha
Operational dose entities:
Ambient dose equivalent H*(10) (unit sievert)
Personal dose equivalent Hp(d) (unit sievert) – with d (mm) = 0.07 (skin), 0.3 (eye lens), 10 (body)

154 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

sponding to equivalent dose at a depth Two important issues of interest that

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
of 10 mm in tissue (equivalent) material, were already present in previous directives,
and personal dose equivalent Hp(d) (unit but may be interpreted differently in the
sievert) defined as the equivalent dose at new implementation, are: categorisation
a depth d in soft tissue below a specified of (exposed) workers and classification of
point in the body. Relevant values of d are workplaces. Workplaces must be divided
d=10 mm (used for estimating effective into controlled areas with more restricted
dose), d=0.07 mm (used for skin dose) access and surveillance and supervised
and d=0.3 mm (for dose to the eye lens). areas with a lower risk and therefore more
These entities are listed here because ready access. Workers are divided into two
radiation protection equipment will nor- groups: group A, who are “liable to receive
mally be specified (calibrated) in terms of an effective dose greater than 6 mSv per
H* or Hp(d). year or an equivalent dose greater than
15 mSv per year for the lens of the eye or
greater than 150 mSv per year for skin and
RULES AND REGULATIONS IN extremities”, and group B, who are not. It
RADIATION PROTECTION is possible to design nuclear medicine de-
The fundamental concepts are laid out by partments and to plan the work so that it
ICRP in their recommendations, the most is highly unlikely that anyone (technolo-
recent general one being from 2007 (ICRP gists, physicians or medical physicists) will
103[1]). Based on this, international (IAEA) exceed the dose limits listed above. How-
and regional (EU) Basic Safety Standards ever, even then it will still be necessary to
(BSS) have been issued[3,4]. The EU BSS-di- monitor (most of ) the staff to document
rective (Euratom 2013/59[4]) must be im- this. The rest of this chapter shows the
plemented in the national legislations types of equipment that can be used for
of all the EU member states by February monitoring of persons and workplaces.
2018. Despite the European cooperation
between authorities and attempts to har-
monise, there will be some differences DOSIMETERS FOR PERSONAL
between countries. Further, local institu- MONITORING
tions may have their own (stronger) de- Monitoring of effective dose and finger
mands, and therefore it is recommended dose has been routine for decades. The
always to look (also) at what local rules recently lowered dose limit to the eye
may apply. lens[5] has created some uncertainty as to

QUALIT Y CONTROL OF NUCLEAR MEDICINE 155


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 12

whether special measures are necessary (rates) and also provide an alarm function
ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

for documenting compliance with this for “active” protection. These two kinds of
limit. Currently there is no consensus on dosimeter form a useful supplement to
the issue, but in nuclear medicine (unlike each other: the passive one for legal use
interventional radiology) it seems likely and the active one as a reminder in daily
that the measure of effective dose will be processes. Dosimeters for personal moni-
sufficient to cover also the dose to the eye toring are normally calibrated to provide
lens. Radiochemistry may be an exception Hp(d) and results are reported in mSv.
to this. The legal, integrating dosimeters
Two kinds of dosimeter are used: pas- should be worn at all times in the depart-
sive, integrating devices that require a ment by all staff for whom this is relevant

Figure 1

a) Personal film dosimeter. b) Open (empty) film holder showing the metal filters and openings
allowing the distinction of different radiation qualities

special process to provide a reading and (as defined by the medical physics expert
active instruments (battery included!) or a radiation protection officer). They
that can immediately show actual dose should normally be worn at the front of

156 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

the body and at belt height. When work- terial can be stored and will retain the in-

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
ing in front of a shielded workplace, the formation for potential later re-evaluation.
dosimeter may be placed at chest level in Against the film material speaks the rather
order to avoid underestimation of the ex- high uncertainty at low exposure, which
posure here. Pregnant staff members (to over the years has become more import-
the extent that they are allowed to work at ant as the dose limits have been reduced
all in the department) should always wear and the price of the silver in the emulsion
the dosimeter at or near belt level to give has increased.
a better representation of the dose to the An alternative that is becoming increas-
foetus. ingly common is the use of thermolumi-
Electronic dosimeters can be considered nescence dosimetry (TLD). Many materials
optional. They are most important for new have the property, when irradiated, that
staff or when introducing new procedures. some of the absorbed energy will get
They may also be given to short-term visi- “trapped”. Electrons are excited out of their
tors for whom the monthly (or 3-monthly) stable ground state and end up in a high-
dosimeters do not make sense. er energy state that is also quite stable at
The classical personal dose meter is film normal (room) temperatures. By heating
based (Fig. 1a); the interaction of ionising (thermo = T) the material to 180–260°C,
radiation with a photographic film was the electrons are “lifted” out of the traps
what in 1896 led Becquerel to the detec- and will fall back emitting light (lumines-
tion of radioactivity. As with other film ma- cence = L). The amount of light can be
terials, processing (developing, fixing) is re- collected and will be proportional to the
quired to show the exposure, and a special energy absorbed in the first place (dose
device is needed to read and convert the = D). For dosimetry, the most common
gray scale into a dose value. The placement basis material is lithium-fluoride (LiF) with
of the film in a specially designed film certain impurities added to determine its
holder (Fig. 1b) with different thickness- properties in detail. The material is “tissue
es of plastic material and metal filters (Al, equivalent” and can be formed into tab-
Sn, In, Cd, Pb) to some extent allows de- lets (Fig. 2); in addition to being used for
termination of the radiation quality (soft/ whole-body monitoring (Fig. 3), these tab-
hard gamma, soft/hard beta, neutron). lets are small enough that they not only
Further, contamination by small drops of can be mounted in finger rings or wrist do-
radioactivity can easily be identified. Film simeters (Fig. 4) but can even be attached
has the advantage that the developed ma- to the fingertips (and covered by gloves).

QUALIT Y CONTROL OF NUCLEAR MEDICINE 157


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 12

As soon as the reading (heating) has been (unexpected) higher dose levels. The set-
ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

performed, the information vanishes by ting of the alarm level requires some con-
“annealing” and the tablet can be reused. sideration when working with patients
The minimum detection limit for TLD is who may be unnecessarily scared if they
lower than for film, but TLD provides little observe that they are triggering the signal.
information on radiation quality, contam- Film and TLD require no quality control
ination is more difficult to observe and from the user except for a simple check
re-examination after the first reading is that the film holder or TLD package is
not possible. not damaged. Most often the handling
Electronic dosimeters (Fig. 5) are usual- of film or the TLD is performed by cen-
ly based on silicon solid-state detectors, tralised institutions that must be accred-

Figure 2 Figure 3

The size of a TLD tablet is only a few The author’s current TLD
millimetres

where ionising radiation produces pairs ited to perform that task and adhere to
of electrons/holes that can be collected certain standards, e.g. ISO/IEC 17025, ISO
and converted to a dose reading in µSv/h. 14146. Electronic dosimeters will typically
The advantage of such an instrument is loudly signify that they are running low
straightforward. The user can keep an eye on battery; if they do not do so, then it
on the dose rate during different work pro- is the responsibility of the user to check
cesses and can even be warned against and detect this. The function of alarm

158 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

levels can be checked (again, by the user) tering material in such a measurement

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
simply by moving the device gradually should be minimised.
towards any (sufficiently strong) gamma Most personal electronic dosimeters
source, e.g. a 57Co flood source, a 137Cs or have an energy response (efficiency) that
68
Ge calibration source or a vial contain- rapidly falls off towards zero for photon
ing99mTc or 18F (in this last case, the device energies below 30–50 keV, dependent on
must be brought just above the vial as the type. This may result in underestima-
it is shielded at the sides!). The absolute tion of doses, in particular if a major part of
calibration should be certified at delivery, the radiation consists of scattered radiation
and constancy of the reading can then be at low energy. However, a comparison in a
checked using, for example, a long-lived PET department has shown that results ob-

Figure 4 Figure 5

TLD tablets can be used for finger and wrist Two examples of personal
dosimetry and even attached to the fingertips electronic dosimeters

calibration source in a fixed geometry. tained using dosimeter film, a TLD and an
Proving the correct absolute calibration electronic dosimeter are reasonably consis-
from first principles is not easy even with tent[6]. In such a comparison, one key issue
a known, certified calibration source. Re- is the handling of (natural) background
member (if trying this) that the distance radiation, which is corrected for in the in-
square law is valid only for the direct radi- tegrating film or TLD systems by subtrac-
ation and therefore the presence of scat- tion of the measured value from an “unex-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 159


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 12

posed” sample, whereas for the electronic ment. There are many manufacturers and
ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

device it must typically be corrected by varieties, and individual specifications may


subtraction of the product of an assumed vary from the ones mentioned here.
background dose rate and the number of
hours that the device has been turned on. Portable devices
A portable survey meter like the one in Fig.
6 is useful for estimating the dose rate in
INSTRUMENTS FOR DOSE the work place or the emission from a pa-
RATE AND CONTAMINATION tient. The illustrated example is based on
MEASUREMENTS two Geiger-Müller (GM) tubes. Using spe-
The ability to detect and monitor ambient cial (metal) filters around the tubes yields
dose rates and control for contaminations an energy response that is reasonably
is mandatory in a nuclear medicine de- independent of the photon energy. The
partment to maintain a safe working en- instrument is calibrated to approximately
yield the ambient dose equivalent H*, and
Figure 6 the display shows mSv/h. It can handle
gamma radiation from a fraction of nat-
ural background (<0.01 µSv/h) up to dose
rates (10 Sv/h) infrequently encountered
in a nuclear medicine department. By it-
self, the device is not sensitive to beta-ra-
diation. It can be equipped with an ex-
ternal probe (end-window GM tube) that
allows beta particles to enter and provide
a signal which is normally displayed as
counts per second. The quality control
of such an instrument consists in battery
A portable survey meter for measurement of check, check for potential contamination
dose rate of the instrument (in a low background
environment) and constancy control with
vironment. The basic detection principles a known (preferably long-lived) source.
have been described in Chapter 11. The in- Deviations of up to ±10% from the ex-
struments shown in this chapter must be pected values are usually considered ac-
seen only as examples of classes of equip- ceptable.

160 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

Contamination monitors exist in many detector for particles (alpha, beta) must

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
forms, but typically they are either large-ar- have a thin entry window to facilitate
ea (100–200 cm2), rectangular detectors their access, and a detector for photons
(Fig. 7a,b) or circular (cylindrical) detectors (X-ray, gamma) must have a high densi-
that detect at an “end-window” of diame- ty and high atomic number to provide
ter 2–6 cm (Fig. 7c). Obviously the large-ar- stopping power. Figure 8 shows a gam-
ea detectors in general have a higher ef- ma source (137Cs) in the upper half and

Figure 7

(a) (b) (c)

a) Contamination monitor with a large detection area. b) For particle measurements, the bottom plate
must be slid off. Note the mesh protecting the thin cover foil. c) Small circular contamination monitor

ficiency which is useful when surveying a a beta source (36Cl) in the lower half. The
large surface. The smaller ones, on the oth- instrument to the right has an end-win-
er hand, are easier to apply in the search dow GM tube which is sensitive to beta
for localized spills (single drops), where the particles and will even detect some alpha
large-area monitors yield a rather constant particles. It has a rather low sensitivity
signal when moving over the spot. for gamma photons because it is a gas
Some instruments look quite similar, detector. To the left is a scintillation (NaI)
and yet their probes have very different detector which is highly sensitive to gam-
properties. It is easily understandable that ma rays (and X-rays). The alumina cover
a detector can function only if the radi- of the crystal, however, does not allow
ation of concern has a high probability beta particles (at least below 1 MeV) to
of entering the active detector volume enter. It can therefore be very important
and a high probability of being stopped to choose the right instrument in a given
(interacting) in that volume. Therefore a situation. Some instruments (like the one

QUALIT Y CONTROL OF NUCLEAR MEDICINE 161


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 12

in Fig. 7a) use a plastic scintillator covered stancy with a relevant source, preferably
ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

with a thin foil (lightproof ) and allow for long lived, in a fixed geometry. For the
detection of both particles and photons end-window detectors (and cylindrical
with reasonable efficiency. NaI detectors) a point source will be a nat-
Quality control of contamination ural choice. For large-area monitors, a foil
monitors should include battery check, of, for example, 100 cm2 with a weak gam-
background check and check for con- ma or beta activity is convenient (Fig. 9).

Figure 8

(a)

(b) (c)

A demonstration of the difference in detection properties between a scintillation counter (left) and
a GM tube (right). In (a), with the detectors pointing to a gamma source, the scintillation counter
yields 100 cps while the GM tube only shows 5–6. With the beta source, the scintillation counter (b)
is down to 5 cps (mostly background) while the GM tube in (c) shows almost 50 cps

162 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

Figure 9 should still have enough energy to pene-

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
trate the cover foil and trigger the detec-
tor, which may use energy discrimination
to distinguish alpha from beta particles
(and noise). A ~3-μm cover foil may cor-
respond to 0.5 cm of air, so to be sure of
detection the distance should not exceed
2 cm (Fig. 10). Then it is obvious that there
is a high probability of contaminating the
foil by touching potential activity directly.
It is possible to cover the instrument by a
single layer of standard household plastic
foil (8–10 µm) and still have a signal. In this
way a potential contamination can easily

be removed by replacing the plastic foil.
Alpha monitors should be checked for
constancy with an appropriate source,
which could be a point source of 241Am,
Example of calibration/test sources which has a half-life of 433 years.
for large-area detectors

Stationary instruments
Not all contamination monitors can A hand-foot monitor (Fig. 11) is a conve-
detect alpha particles. A particularly thin nient piece of equipment for a fast check
window is required for the particles to en- of personnel contamination. One of the
ter. When measuring alpha radiation on a hand detectors can be removed from its
surface, it is also important to remember position to scan clothes. In older instru-
that the range even in air is quite limited ments, the panels used to be proportion-
and that the energy is gradually lost along al counters (gas-based) but today plastic
the path. For protection, it is a reasonably scintillator systems are common. Quality
easy rule of thumb that the range in air is control of this kind of monitor is the same
always below 10 cm. For a 5-MeV alpha as for portable contamination monitors,
particle, however, the range is really only i.e. checking at regular intervals for back-
4.5 cm, and after passing the air gap from ground and constancy, e.g. with a long-
the source to the detector, the particle lived beta-active sheet (14C or 90Sr).

QUALIT Y CONTROL OF NUCLEAR MEDICINE 163


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
CHAPTER 12

Figure 10
ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N

When measuring for alpha-particles, a close distance to the (potential) source is mandatory

Figure 11

At certain places in the department it


may be relevant to install a survey instru-
ment that can warn against (temporary)
high dose rates. This could be outside a
cyclotron room, next to a radiochemistry
lab or in other places where a (temporary)
warning is appropriate. Readings can be
checked with a 137Cs source in a reproduc-
ible position (assuming low or corrected
background).

Acknowledgements. The author wish-


es to thank technologists Marianne Fed-
erspiel, Elin Lindell, Eunice Saxtoft and
Susanne Svalling for valuable input to and
A hand-foot monitor is useful for a fast discussion and criticism of the contents of
check of personal contamination this chapter.

164 Q UA L IT Y CON T ROL OF NUC LEAR M EDIC INE


I N S T RU M E N TAT I ON AND PROTOCOL STANDARD ISATIO N
CHAPTER 12

ME A S URE M E NT E Q UI PM E NT
RA DIATI O N PR OTE C TI O N
REFERENCES

1. ICRP. The 2007 Recommendations of the Interna- gers arising from exposure to ionising radiation and
tional Commission on Radiological Protection. ICRP repealing Directives 89/618/Euratom, 90/641/Eura-
publication 103. Ann ICRP 2007;37(2–4):1–332. tom, 96/29/Euratom, 97/43/Euratom and 2003/122/
2. International Atomic Energy Agency. Nuclear Euratom. OJ of the EU. L13; 57: 1–73 (2014).
medicine physics. Vienna: IAEA, Vienna; 2015. Web: 5. ICRP statement on tissue reactions and early and
http://www-pub.iaea.org/books/IAEABooks/10368/ late effects of radiation in normal tissues and organs
Nuclear-Medicine-Physics-A-Handbook-for-Teach- – threshold doses for tissue reactions in a radiation
ers-and-Students. Accessed 1 March, 2017. protection context. ICRP publication 118. Ann ICRP
3. International Basic Safety Standards for Protection 2012;41(1–2):1–322.
Against Ionizing Radiation and for the Safety of 6. Dall B, Svalling S, Holm S. A comparison of three dif-
Radiation Sources. IAEA Safety Series No. 115. ISBN: ferent methods for measuring whole body radiation
92-0-104295-7. exposure. Eur J Nucl Med Mol Imaging 2008;35 (Suppl
4. European Council Directive 2013/59/Euratom on ba- 2):S214.
sic safety standards for protection against the dan-

QUALIT Y CONTROL OF NUCLEAR MEDICINE 165


INSTRUMENTATION AND PROTOCOL S TANDARDIS ATION
IMPRINT

ISBN: 978-3-902785-13-8

Publisher:
European Association of Nuclear Medicine
Schmalzhofgasse 26, 1060 Vienna, Austria
Phone: +43-(0)1-890 27 44 | Fax: +43-(0)1-890 44 27-9
Email: office@eanm.org | URL: www.eanm.org

Main-Editor:
Sebastijan Rep

Co-Editors:
Marie Claire Attard
Luca Camoni
Andrea Santos

English Language Editing:


Rick Mills

Project Management:
Magdalena Schwartz

Content:
No responsibility is taken for the correctness of this information.
Information as per date of printing September 2017.

Layout & Design


Robert Punz
European Association of Nuclear Medicine
Schmalzhofgasse 26, 1060 Vienna, Austria
Phone: +43-(0)1-890 27 44 | Fax: +43-(0)1-890 44 27-9
Email: office@eanm.org | URL: www.eanm.org

Printing
“agensketterl” Druckerei GmbH
Druckhausstraße 1, A-2540 Bad Vöslau, Austria
Email: info@diedrucker.biz | URL: www.agensketterl.at