Você está na página 1de 12

Analytica Chimica Acta 488 (2003) 243–254

Analyser of chromium and/or cobalt


L.A. Tortajada-Genaro, P. Campı́ns-Falcó∗ , F. Bosch-Reig
Departament de Quı́mica Analı́tica, Facultad de Quı́mica, Universitat de València, C/Dr. Moliner 50, E-46100 Burjassot, Valencia, Spain
Received 9 January 2003; received in revised form 14 April 2003; accepted 27 May 2003

Abstract
Two stopped-flow manifolds have been proposed for individual or simultaneous determination of chromium and cobalt in
water samples. Automated procedures based on multicommutation systems have emphasised the differences of their catalytic
effect in luminol–hydrogen peroxide chemiluminescence reaction. A more rapid decay of signal was observed for Co for
both configurations (flow injection or continuous injection). The influence of chemical and hydrodynamic variables has been
studied in order to establish the robustness of method. The analysis rate was lower 1.5 min per replicate.
Chemometric tools have been employed for the resolution of their contributions. Partial least squares (PLS) and H-point
standard additions method (HPSAM) were used as multivariate calibration models. The percentages of explained variance were
97–99% (two factors). PLS and HPSAM obtained similar results. HPSAM provided a simple calibration model contributing
to develop an analyser for chromium and/or cobalt.
Standard mixtures, spiked samples and a certified reference material validated the proposed strategy. The applicability has
been demonstrated by the determination of Cr and Co concentration in different water samples. The best results have been
obtained for continuous injection providing more robust predictions. The achieved detection limit was 0.2 ␮g/l for both metals.
© 2003 Elsevier B.V. All rights reserved.
Keywords: Multicommutation; Chromium and cobalt; Simultaneous determination; Chemiluminescence; Water matrix; HPSAM; PLS

1. Introduction solenoid valve. The approach has been used to per-


form binary sampling, masking feasible the reduction
Computer-controlled switching systems are widely of reagent consumption, sequential determinations
used in some analytical areas. It supposes a flow switc- and sequential management of incompatible reagents
hing technique in order to achieve automation of steps [3–7].
of an analytical process. On-line sample washing, Some applications have been proposed for water
preconcentration, extraction, dilution or hyphenated analysis: majority metals [8], phenols [9,10] and an-
techniques have been proposed. Many of the strategies ions [11,12]. However, the solenoid-activated rotatory
described come from the field of chromatography [1]. valves have only been employed as sample introduc-
Nevertheless, multicommutation is a novel ap- tion system based on time. So, simple flow registers
proach in flow analysis [2]. It can be implemented by are obtained by using these manifolds.
using discrete commutation devices, such as three-way Stopped-flow techniques have also been employed
in order to emphasise kinetic differences between
∗ Corresponding author. Tel.: +34-96-35-43-002; some analytes in different detection systems. Some
fax: +34-96-35-44-322. compound mixtures have been resolved obtaining
E-mail address: pilar.campins@uv.es (P. Campı́ns-Falcó). time-intensity curves [13–19]. Kinetic methods based

0003-2670/03/$ – see front matter © 2003 Elsevier B.V. All rights reserved.
doi:10.1016/S0003-2670(03)00677-9
244 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

on flow assemblies and chemiluminescence detection additives, the signal of sample at time ti will be a con-
have been described using temporal difference in tribution of both species (SS,i = SX,i + SY,i ), where
activator effect [20,21] and temporal injection [22]. SS ,i , SX,i and SY ,i are the sample signal, signal of
In this paper, two automated assemblies are pro- species X and signal of species Y at ti , respectively.
posed for simultaneous determination of chromium
and cobalt. Computer controls a six-way valve and 2.1. HPSAM for flow injection
a rotatory valve in order to register time-intensity
curves. The chemiluminescence (CL) registers from In this paper, HPSAM [24] has been modified for
luminol–hydrogen peroxide reaction are obtained stopped-flow data. The time-intensity register is com-
according to stopped-flow theoretical background. posed from two regions, according to point where the
In the first manifold, sample introduction is discon- flow is stopped. Previous to maximum intensity, the
tinuous. The time-intensity register is composed from register is obtained from a flow injection system. The
two regions divided where the flow is stopped. Previ- response is result of dispersion of sample zone within
ous to maximum intensity, the register is obtained from the carrier and the chemical process of formation
a flow injection system. After maximum intensity, the of a chemical species. After maximum intensity, the
register is obtained from a stopped-flow system. In this register is obtained from a stopped-flow system. In
region, the response is result of decomposition of CL this region, the response is result of decomposition of
species. In the second manifold, sample introduction chemiluminescence species. In Fig. 1a, an example
is continuous. Previous to maximum intensity, the reg- of injected stopped-flow registers is shown.
ister is obtained from a system of continuous sample For the determination of species X, HPSAM is based
injection and after a decay of CL intensity is observed. on selection of two times ti and tj in such a way that
For the resolution of binary mixtures, partial least the interferent signals of species Y are equal to SY,i =
squares regression (PLS) and H-point standard addi- SY,j ( SY,i,j = 0). ti corresponds to FI region and tj
tion method (HPSAM) have used. HPSAM [23] is corresponds to stopped-flow region.
other useful multivariate calibration method in equi- The difference of signals at two times ti and tj will
librium and in kinetic studies. The method has been be for the sample:
developed for the determination of unbiased analyte
concentrations in the event that the presence of a direct SS,i,j = SS,j − SS,i = SX,j + SY,j − SX,i − SY,j
interferent and/or the total Youden blank is known. = SX,i,j + SY,i,j = SX,i,j (1)
HPSAM has been also applied for the correction of
measured signal in FIA systems [24]. Therefore, standard addition calibrations of species
The usefulness of the new proposed method is X provides values of slopes Mi and Mj at two previ-
demonstrated by applying it to simultaneous determi- ously selected times, ti and tj , In absence of matrix
nation of cobalt and chromium in some water samples. effect, the Mi and Mj values can be calculated from
These metals are widespread in water, existing in vari- standard calibration curves, directly. These calibration
ous chemical, physical and morphological forms. Very lines intersect at H-point, with (−cH , SH ) coordinates.
low levels are presented in underground and surface The H-point is a function of the analyte concentration
waters but industrial wastes often introduce them into (cX ) expressed in the following equation:
hydric cycle. Their release must be carefully moni- SS,i,j SX,i,j
tored because many of their forms are toxic. Then, −cH = = = cX (2)
Mj − M i Mj − M i
laws in the most of the countries have fixed the max-
imum concentration for Co and Cr in drinking water. The species Y concentration is calculated by inter-
polation of the SH value in a conventional calibration
graph. Other possibility is to develop another equation
2. Theorical background similar to Eq. (1) for species Y where SX,i,j = 0.
The method can generally be performed with more
Let us consider a sample S which is a binary mix- than one pair of variables (ti , tj ), in which case it can
ture of species X and species Y. If the responses are be considered a multivariate method. The selection of
L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254 245

position of chemiluminescence species. In Fig. 1b,


an example of stopped-flow registers for continuous
injection is shown.
The method is based on the well-known property
for most of analytical registers. For instance, in spec-
trophotometry, the ratio between absorbance values at
two wavelengths is a constant for a pure compound.
This fact is used for the detection of overlapped peaks
[27,28]. The study of chemiluminescence registers has
demonstrated that the ratio of signals at two times is
also a constant. For times values ti and tj , KX,i,j ratio
is defined for the X species as:
SX,i
KX,i,j = (3)
SX,j

It means the ratio between the signal at time ti (SX,i )


and the signal at time tj (SX,j ) for X species. So, the
difference SX,i − KX,i,j SX,j will be equal to zero.
Considering this weighted difference of signals at
both time values ti and tj for the sample, the following
equation is obtained:

SS,i − KX,i,j × SS,j


= SX,i + SY,i − KX,i,j × SX,j − KX,i,j × SY,j
= SY,i − KX,i,j × SY,j (4)

Therefore, the weighted increment of signals only


depends on Y species. A calibration procedure based
on standard addition of Y species will be provided
Fig. 1. Hypothetical analytical signal for stopped-flow manifold: the calibration slope at the selected time values ti
(a) flow injection and (b) continuous injection. and tj . In the absence of matrix effect, such values
can be calculated from standard solution of Y species
variables is based on the best ratio between noise and directly.
analyte increments. The following step is the estimation of Y species
concentration in samples. The concentration can
2.2. HPSAM for continuous injection flow be calculated by interpolation of the corresponding
SS,i − KX,i,j SS,j values in the calibration curves
In this paper, K-ratio [25,26], a variant of HPSAM, SY,i − KX,i,j SY,j versus cY .
has been modified for injected stopped-flow data.
The time-intensity register is composed from two re-
gions, according to point where the continuous flow 3. Experimental section
is stopped. Previous to maximum intensity, the regis-
ter is obtained from a system of continuous sample 3.1. Apparatus and reagents
injection. The response is only the chemical process
of formation of a chemical species. After maximum A Jasco CL-1525 (Tokyo, Japan) chemilumines-
intensity, the register is obtained from a stopped-flow cence detector, a LabPro Rheodyne six-way injection
system. In this region, the response is result of decom- valve and a LEA six-position valve were used for the
246 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

measuring. It was controlled by Borwin chromatog- 10.8). The concentration of hydrogen peroxide was
raphy software and the interface Hercule lite (JMBS, 0.01 mol/l.
France). A Gilson Miniplus peristaltic pump was used
to drive the reagents to the detection cell. Tygon tubing 3.2.1. Assembly I. FI procedure
(i.d. = 0.8, 2.0, 2.5 mm) was used for the peristaltic The carrier was composed of carbonate (0.05 mol/l,
pump. Other tubing was made of PTFE with i.d. = pH 10.8) and EDTA (2 × 10−3 mol/l). The scheme of
0.5 mm. The light emission intensity was recorded as FI manifold is shown in Fig. 2a. The luminol reagent
function of time. (L) was mixed with peroxide reagent (P). The stan-
The following reagents were used: chromium (III) dard solutions or samples (S) were injected in carrier
nitrate (Panreac, Spain), cobalt nitrate (Panreac), hy- (C) by action of valve 1. The reagent flow met with
drogen peroxide (Merck, Germany), luminol (Fluka, carrier flow in T-piece. Chemiluminescence emission
Switzerland), sodium carbonate (Merck), chlorhydric was registered in the detection system. Valve 2 was
acid (trace pur, Merck) and nitric acid (trace pur). The always operating in position V2A, then the solution
solutions were prepared in water (nanopure, Sybron, was driven to detector position.
Barnstead, Spain). The concentrations of stock stan-
dard solutions were 100 mg/l of chromium (III) or 3.2.2. Assembly II. Stop-flow procedure by flow
cobalt (II). injection
The manifold was the same than described above
3.2. Procedures and shown in Fig. 2a. The mixing of reagents and
injection of sample were the same than FI procedure.
Working standard solutions were prepared in HCl, However, switching of valve 2 stopped the flow in the
2 × 10−3 mol/l, daily. Luminol (1.2 × 10−4 mol/l) detection cell because solution was driven to waste
was prepared in carbonate solution (0.01 mol/l, pH outlet.

Fig. 2. (a) Scheme of FI system and flow injected stopped-flow system and (b) Scheme of continuous stopped-flow system. V1: valve 6-ways,
V2: valve 2-ways, PP: peristaltic pump, D: detector, W: waste, C: carrier stream, S: sample stream, P: peroxide stream, L: luminol stream.
L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254 247

3.2.3. Assembly III. Stop-flow procedure by 3.2.5. Sampling


continuous injection Samples were collected in different points of Va-
The carrier was composed of carbonate (0.125 mol/l, lencia area (Spain) and were filtered using a cellu-
pH 10.8) and EDTA (5 × 10−3 mol/l). The scheme of lose acetate filter (0.45 ␮m). Natural water samples
FI manifold is shown in Fig. 2b. The luminol reagent were acidified with HCl, 5 × 10−3 mol/l, but the acid-
(L) was mixed with peroxide reagent (P). The stan- ification of metallurgic industrial wastewater was not
dard solutions or samples (S) were mixed with carrier performed (presence of S2− ). Sampling and storage
(C) in T-piece when the valve 2 was in position V2A. containers and other glassware or plastic material was
The reagent flow met to carrier flow in other T-piece. acid-washed (nanopure water washing, bath HCl 10%
Chemiluminescence emission was registered in the for 12 h and nanopure water washing).
detection system. The flow stopped in the detection Recovery assays were performed in a river water
cell by switching of valve 1, because solution was sample. Spiked samples were prepared by addition of
driven to waste outlet. 1 + 0, 0 + 1 and 1 + 1 ␮g/l of Cr and Co, respectively.
In order to obtain the registers, the following con- The certificate reference material SRM 1640 (NIST,
trol sequences were performed. The sequences permit- USA), composed of natural fresh water, was also anal-
ted to control the sample injection and/or stop-flow. ysed.
Switching of valves was realised by methodology se- In order to calculate the total concentration of
quences in Borwin software. In Table 1, the general chromium, a reduction treatment procedure was per-
step in accordance with the number of replicates (r) is formed. To the samples 0.08% H2 O2 was added and
shown, being r = 1, 2, 3, . . . , n. In all studies, three heated (80 ◦ C, 30 min).
replicates were performed at least, r ≥ 3. For all calculations, Unscrambler (CAMO, Norway)
and Visual Basic routines for Excel were used.
3.2.4. Calibration set
Different calibration sets were assayed. For uni-
variate calibration, standard solutions of chromium or 4. Results and discussion
cobalt were measured 0–4 ␮g/l for instrumental gain
of 10 and 0–20 ␮g/l for instrumental gain of 1. Stan- 4.1. Optimisation of chemiluminescence signal
dard bicomponent mixtures were also prepared. The
composition of solutions followed a partial design Different chemical and physical parameters were
52 . optimised for the three configurations. According to

Table 1
Protocol sequences for controlling of valves for different procedures
Step Time (min) Valve Action Description

Flow injection
2(r − 1) + 1 1.1r + 0.1 V1 A→B Sample insertion (replicate r)
2r + 2 1.1r + 0.4 V1 B→A Filling of sample coil
Stopped-flow and flow injection
2r + 1 1.4(r − 1) + 0.1 V2 A → B Sample insertion (replicate r)
2r + 2 1.4(r − 1) + 0.3 V1 A → B Stop-flow
2r + 3 1.4(r − 1) + 0.45 V1 B → A Filling of sample coil
2r + 4 1.4(r − 1) + 1.0 V2 B → A Activation of flow
Stopped-flow and continuous injection
1 0.1 V2 A → B Sample stream insertion
2r + 2 1.1(r − 1) + 0.5 V1 B → A Stop-flow (replicate r)
2r + 3 1.1(r − 1) + 0.5 V1 A → B Activation of flow
n – V2 B → A Reinitiating
V2 in position for the entire sequence.
248 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

initial experiences, most critical variables were cho- for flow injection manifold and 0.4 min for continuous
sen. The studied range was selected based on the in- injection manifold. The injection frequency expressed
fluence of variable in chemiluminescence signal o its in function of time for each replicate were 1.1 min
evolution in time. The optimisation criterion was the for flow manifold, 1.4 min for stopped-flow and flow
maximum signal and maximum repeatability. So, CL injection manifold and 1.1 min for stopped-flow and
signal of standard solutions (0.5–2 ␮g/l of Cr or Co) continuous injection manifold. Therefore, the opti-
was registered as function of time. mised sequences is shown in Table 1.
The chemical variables of studied reaction were: In order to assess the robustness of method, some
luminol concentration, carbonate concentration, pH experiments were performed. The optimal conditions
(stream L and stream C), H2 O2 concentration and were varied in a factor of 10–20%. The objective was
EDTA concentration. Table 2 shows the optimal val- to study the influence of the effect of experimental
ues. EDTA concentration was chosen using selectivity variations or error sources in the response. The critical
criteria, it means, a selective signal of Cr or Co against variables were pH of carbonate buffer and EDTA con-
other metals and similar intensity of signal for both centrations. Specially, variations of cobalt signal were
metals. It is important to use high quality reagents, be- observed. For the rest of variables the signal changes
cause the presence of impurities involves bad results were acceptable according to precision and accuracy
(interferences or high base-line). criteria.
The hydrodynamic variables were optimised for
three configurations. Flow rate of sample stream, car- 4.2. Unicomponent and univariate determination
rier stream and reagent streams were studied. The
length of reaction coil and volume of sample coil Analytical parameters for the determination of
were also optimised, considering the distance between chromium and cobalt were calculated in conditions
consecutive peaks. The selected values are shown in where only one analyte was measured. However,
Table 2. the determination in water sample with both metals
The automation of procedure involved the optimi- in measurable forms was not possible because the
sation of control sequence. Then, the time values for interference effect is mutual.
switching of valves were studied, the objective was to Height peak was chosen as analytical signal for
stop the flow in the point maximal response. It made flow manifold and stopped-flow manifold using flow
possible to register some replicates avoiding over- injection (Fig. 3). The parameters of univariate lin-
lapped peaks and guaranteeing to fill the sample coil ear calibration are summarised in Table 3 (interval
between replicates. Stopped-flow times were 0.15 min 0–4 ␮g/l).

Table 2
Experimental conditions for chemical and hydrodynamic variables
Variables Unity Stream Assembly I Assembly II Assembly III

Luminol concentration mol/l L 1.2 × 10−4 1.2 × 10−4 1.2 × 10−4


Carbonate concentration mol/l L 0.01 0.01 0.01
C 0.05 0.05 0.125
Buffer pH – L 10.8 10.8 10.8
– C 10.8 10.8 10.8
H2 O2 concentration mol/l P 0.01 0.01 0.01
EDTA concentration mol/l C 2 × 10−3 2 × 10−3 5 × 10−3
Flow-rate ml/min L 1.2 1.2 1.2
P 1.2 1.2 1.2
C 3.5 3.5 1.2
S 1.2 1.2 2.8
Reactor length cm – 10 10 10
Sample coil volume ␮l – 200 200 –
L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254 249

Fig. 3. Normalised analytical signal for standard solutions of chromium (0–2 ␮g/l) and cobalt (0–2 ␮g/l) and for certificate reference
material (1:25) obtained stopped-flow configurations by flow injection (a) and continuous injection (b).

Detection limits were calculated from three-time cence signal is most significant for Co standards than
derivation of blank solution (LDN/S) and from cal- Cr standards, the sample has an intermediate decay.
ibration curve (LD 3syx /b). The repeatability (RSD) This behaviour was observed for both configurations
was calculated from the measurement of standard so- of stopped-flow. The signal curve is the same for dif-
lutions (1 ␮g/l, replicates = 5). ferent concentrations of each metal ion as it can be
It can be seen, the sensibility values, detection lim- seen in this figure.
its and repeatability values were similar for three man- The application of HPSAM needed the selection
ifolds. Additionally, the standard solutions for linear of pairs of variables according to some requirements
response interval were measured using a lower in- depending on the option studied.
strumental gain (factor 10). The regression parameters For the configuration of flow injection, the oper-
were calculated. Detection limits and sensibility val- ative criteria was based on a distance between vari-
ues were 10 times higher than values show in Table 3. ables higher than 4, a difference of interferent signals
lower than 5% and a determination coefficient (r2 ) for
4.3. Multivariate calibration models: PLS and (Si − Sj versus c) curve higher than 0.99.
HPSAM For the configuration of continuous injection, the
variables were chosen according to a suitable K-ratio
Fig. 3 shows the normalised registers obtained for for the corresponding interferent and analyte. It sup-
standard solutions of chromium and cobalt and for a posed a distance between variables higher than 4, a
sample. It can be seen the decay of chemilumines- weighted increment higher than 10% for the analyte

Table 3
Analytical parameters for unicomponent and univariate determination
Parameter Assembly I Assembly II Assembly III

Cr Co Cr Co Cr Co
a 32 ± 8 14 ± 14 8±9 8±4 19 ± 3 26 ± 8
b 149 ± 3 188 ± 6 132 ± 4 144 ± 4 187 ± 2 174 ± 4
r2 0.992 0.99 0.991 0.994 0.999 0.993
syx 20 29 19 9 9 22
LD N/S 0.2 0.2 0.1 0.1 0.1 0.1
LD 3syx /b 0.4 0.5 0.4 0.2 0.1 0.4
RSD 5.5 5.3 5.2 4.2 2.0 3.3
250 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

Fig. 4. Increment of signal for standard solutions of chromium and cobalt obtained stopped-flow configurations: (a) flow injection Cr,
47–115 increment, (b) flow injection Co, 51–80 increment, (c) continuous injection Cr, 27–80 increment and (d) continuous injection Co,
27–120 increment.

and a determination coefficient (r2 ) for (Si − K Sj ver-


sus c) curve higher than 0.99.
An example of increment selected for each config- Table 4
uration is shown in Fig. 4. Using a selected time pair PLS and HPSAM models for calibration set
with the same signal of Y species (Co or Cr), it is pos- Model Parameter Assembly Assembly
sible to verify than the increment is independent to II III
concentration of this species. Five concentrations of HPSAM Increments Assayed 12246 12246
Co or Cr between 0 and 2 ␮g/l were included in the Cr selection 25 32
Co selection 20 12
plots of Fig. 4. A tendency of the points in function SEP Cr 0.27 0.18
of the Y concentration was not observed. Then, it is calibration Co 0.69 0.18
demonstrated that the cancellation of the Y species is
PLS Factors Assayed 5 5
satisfactory. Selected 2 2
The predictions were calculated as an average of % Explained X-block 99.7 99.7
results provided for the pairs of variables with the Variance Y-block cal 97.9 99
highest slope of calibration curves. Table 4 shows the Y-block val 97.3 98.8
SEP Cr 0.16 0.1
assayed increments, selected increments and calcu-
calibration Co 0.14 0.16
lated errors of prediction for calibration set.
L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254 251

Fig. 5. Predictions of binary mixtures measured in stopped-flow assemblies with different injection configurations: (a) Cr assembly II, (b)
Co assembly II, (c) Cr assembly III and (d) Co assembly III.

In order to evaluate the prediction of previous mod- (SEP) are summarised in Table 4. SEP values were
els, binary standard solutions were determined. The calculated for calibration set.
composition of these binary solutions described a par- The results were similar to provided for both meth-
tial 52 design for both metals. The prediction results ods, then HPSAM can be considered as an alternative
are plotted in Fig. 5. It can be observed that the best re- to PLS calibration. Although, the advantage of HP-
sults were obtained by using the assembly II. Compar- SAM lies in the simplicity and the correction of matrix
ing both metals, the prediction of chromium is more effect when it is present.
accurate and precise. Any difference was detected be-
tween both calibration models. 4.4. Determination in water samples
PLS-2 models were obtained for both configura-
tions from unicomponent standard solutions of Cr or The proposed configurations were applied for the
Co. X-block was composed of CL registers versus simultaneous determination of chromium and cobalt
time (column centred) for 160 variables (0–0.08 min in different water samples. Previous to measure-
interval). Percentages of explained variance, number ment step, reduction treatment (Cr(VI) to Cr(III))
of factors and root means standard errors of prediction was performed in order to determine total chromium
252 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

Fig. 6. Recovery percentages for spiked water sample measured in stopped-flow assemblies by flow injection (a) and continuous injection
(b) for both calibration models.

concentration. This procedure was based in the reac- Samples collected in different locations in Valen-
tion with H2 O2 in acidic conditions and T = 80 ◦ C. cia area (Spain) were analysed. They are river water,
Fig. 6 shows the recovery values obtained for a wa- coastal water, harbour water and wastewater of met-
ter sample spiked with 1 ␮g/l of Cr, 1 ␮g/l of Co or allurgic industries. The water parameters and the con-
1 ␮g/l of Cr and 1 ␮g/l of Co. Both calibration mod- centrations of Cr and Co for both configurations and
els (PLS and HPSAM) and assemblies II and III were both calibration models are summarised in Table 5.
also compared. In order to compare both stopped-flow configura-
The percentages of recovery were better for assem- tions (assemblies II and III), a t-test for paired samples
bly III. Similar prediction values were observed for was performed. The following t values were obtained:
PLS and HPSAM models. 0.57 (Cr-PLS), −1.29 (Cr-HPSAM), −3.75 (Co-PLS)
A certificate reference material (SMR 1640) was and −0.07 (Co-HPSAM). Therefore, the predictions
analysed. It is a fresh water sample in which chromium were similar except for Co-PLS (ttab = 2.57, 95%
and cobalt contents are 38.6 ± 1.6 ␮g/l and 20.28 ± and d.f. = 5). However, the predictions were more
0.31 ␮g/l, respectively. In Fig. 3 are represented the robust and precise for continuous injection. The con-
CL registers obtained for both configurations. The pre- centrations obtained for certificate reference material
dicted concentrations are included in Table 5. indicated that the prediction errors for continuous

Table 5
Prediction of samples
Sample pH Sma Model Assembly II Assembly III

Cr Co Cr Co

CRM (1.25)b – – PLS-2 1.87 ± 0.10 0.59 ± 0.09 1.6 ± 0.3 0.69 ± 0.17
HPSAM 1.3 ± 0.2 0.9 ± 0.3 1.6 ± 0.4 0.67 ± 0.16
Harbour 8.52 2.8 PLS-2 0.61 ± 0.16 1.7 ± 0.6 0.75 ± 0.05 1.79 ± 0.10
HPSAM 0.6 ± 0.3 1.46 ± 0.11 0.52 ± 0.11 1.70 ± 0.09
Industrial (1:10) 8.22 17 PLS-2 1.1 ± 0.3 1.5 ± 0.3 1.46 ± 0.09 1.8 ± 0.4
HPSAM 1.51 ± 0.15 1.6 ± 0.3 1.43 ± 0.14 1.6 ± 0.2
Industrial (1:10) 7.80 10 PLS-2 1.86 ± 0.04 0.43 ± 0.12 1.6 ± 0.3 0.58 ± 0.05
HPSAM 1.10 ± 0.05 0.35 ± 0.06 1.4 ± 0.2 0.66 ± 0.07
Replicates n = 3.
a SM: suspended material (mg/l).
b Reference values: Cr, 1.54; Co, 0.81.
L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254 253

assembly were lower than 10% for Cr and Co, but strated by the determination of Cr and Co concentra-
higher for flow assembly. tion in different water samples.
The comparison of the calibration models (PLS and Comparing to other described methods, these strate-
HPSAM) involved a t-test for paired samples. The fol- gies have some advantages like low cost, rapid re-
lowing t values were obtained: 1.47 (Cr, discontinu- sponse and no limitations of specialised personnel.
ous), 1.85 (Cr, continuous), −0.40 (Co, discontinuous) These characteristics are very important for periodical
and 1.76 (Co, continuous). Therefore, any difference monitoring of water quality.
of prediction was found between both methods (ttab =
2.57, 95% and d.f. = 5). The standard deviation for
replicates of samples were also similar for PLS and Acknowledgements
HPSAM.
The programming of HPSAM calibration for the The authors are grateful to the DGICYT (Project no
procedure described is easy. Then, it is possible to PB 97-1387) for financial support. L.A.T.G. expresses
propose an analyser of chromium and/or cobalt. A their gratitude to Ministerio de Eduación y Cultura
sequence of calibration standards and samples is (Spain) for the predoctoral grant.
programmed. The concentrations of Cr and Co are
provided direct and automatically.
References

5. Conclusions [1] P. Campı́ns-Falcó, R. Herráez-Hernández, A. Sevillano-


Cabeza, J. Chromatogr. B 619 (1993) 177.
[2] B.F. Reis, M.F. Giné, E.A.G. Zagatto, J.L.F.C. Lima, R.A.
In this paper, the combination of two multicommu-
Lapa, Anal. Chim. Acta 293 (1994) 129.
tated assemblies, the stopped-flow technique and two [3] F. Albertús, A. Cladera, V. Cerdá, Analyst 125 (2000) 2364.
multivariate calibration models has been proposed. [4] F. Albertús, B. Horstkotte, A. Cladera, V. Cerdá, Analyst 124
The proposed methodology has achieved the simulta- (1999) 1373.
neous determination of Cr and Co in water samples [5] B.P. Bubnis, M.R. Straka, G.E. Pacey, Talanta 30 (1983) 841.
by a chemiluminescence detection system. [6] A.N. Araújo, J.L.F. Costa-Lima, B.F. Reis, E.A.G. Zagatto,
Anal. Chim. Acta 310 (1995) 447.
The proposed assemblies were based on differ-
[7] P.K. Dasgupta, K. Petersen, Anal. Chem. 62 (1990) 395.
ent injection configurations. The critical variables [8] F.R.P. Rocha, P.B. Martelli, R.M. Frizzarin, B.F. Reis, Anal.
(chemical or hydrodynamic) of the systems have been Chim. Acta 366 (1998) 45.
optimised. The robustness of method has been also [9] B.F. Reis, A. Morales-Rubio, M. de la Guardia, Anal. Chim.
studied. Although both configurations have presented Acta 392 (1999) 265.
similar analytical parameters for unicomponent de- [10] J. Michalowski, P. Halaburda, A. Kojlo, Anal. Lett. 33 (2000)
1373.
termination, some differences have been observed for [11] F.R.P. Rocha, B.F. Reis, Anal. Chim. Acta 409 (2000) 227.
simultaneous determination. The best results have [12] F.R.P. Rocha, P.B. Martelli, B.F. Reis, Anal. Chim. Acta 438
been obtained for continuous injection providing (2001) 11.
more robust predictions for standard solutions and [13] T. Pérez-Ruiz, C. Martı́nez-Lozano, V. Tomás, J. Fenoll,
samples. Analyst 125 (2000) 2372.
[14] W. Jianhua, H. Ronghuan, Analysis 21 (1993) 421.
Respect to the comparison of calibration models,
[15] J.H. Wang, R.H. He, Anal. Chim. Acta 303 (1995) 241.
PLS and HPSAM have provided similar results. Both [16] J.H. Wang, R.H. He, Mikrochim. Acta 117 (1994) 23.
models can be used without distinction between cal- [17] P.D. Bryan, A.C. Capomacchia, J. Pharm. Biomed. Anal. 9
culation times, predictive error of calibration or val- (1991) 855.
idation, recovery percentages or prediction in water [18] D. Pérez-Bendito, A. Gómez-Hens, M. Silva, J. Pharm.
samples. Therefore HPSAM can be an alternative to Biomed. Anal. 14 (1996) 917.
[19] Y.S. Hsieh, S.R. Crouch, Anal. Chim. Acta 309 (1995) 277.
PLS in the chemiluminescence analysis.
[20] A. Navas-Dı́az, J.A. González-Garcı́a, Anal. Chem. 66 (1994)
The proposed methods have been validated by us- 988.
ing standard mixtures, spiked samples and a certified [21] S. Panadero, A. Gómez-Hens, D. Pérez-Bendito, Anal. Chim.
reference material. The applicability has been demon- Acta 329 (1996) 135.
254 L.A. Tortajada-Genaro et al. / Analytica Chimica Acta 488 (2003) 243–254

[22] N.P. Evmiridis, N.K. Thanasoulias, A.G. Vlessidis, Anal. [25] F. Blasco-Gómez, P. Campı́ns-Falcó, F. Bosch-Reig, G.M.
Chim. Acta 398 (1999) 191. Liu, Analyst 123 (1998) 2857.
[23] F. Bosch-Reig, P. Campı́ns-Falcó, Analyst 113 (1988) [26] L. Guomin, M. Xueliang, L. Ping, Anal. Lett. 26 (1993) 801.
1011. [27] A.F. Poile, R.D. Coulon, J. Chromatogr. 204 (1981) 149.
[24] P. Campı́ns-Falcó, F. Blasco-Gómez, F. Bosch-Reig, Talanta [28] A.F. Fell, H.P. Scott, R. Gill, A.C. Moffat, J. Chromatogr.
47 (1998) 193. 282 (1983) 123.

Você também pode gostar