PD534-012

Rotating
Anode CIRCLEX
X-ray tube
Assembly 0.3/0.8P324&0.6/1.2P324DK-85

GENERAL
The Shimadzu CIRCLEX Focal spot combination of
0.3/0.8P324DK-85 & 0.3mm/0.8mm (for over-table) or
0.6/1.2P324DK-85, Rotating Anode 0.6mm/1.2mm (for C-arm table
X-ray tube assemblies are rated to Cvision & over-table) is available.
150kV and feature a 100mm (4inch)
target of Rhenium-tungsten faced CONFIGURATION
Molybdenum with a target angle 12˚. The CIRCLEX 0.3/0.8P324DK-85&
Anode heat content is 280kJ {400kHU} 0.6/1.2P324DK-85, Rotating Anode
and the maximum anode heat X-ray tube assemblies are composed
dissipation rate is 1600W {2250HU/s}. as follows.
The anticrystalization agent is mixed
in tungsten, protecting the surface of
the anode from becoming rough and CIRCLEX CIRCLEX
cracked, so that reduction of effective Model 0.3/0.8P324DK-85 0.6/1.2P324DK-85
X-ray dose is minimized over a long (Nominal focal spot; 0.3/0.8mm) (Nominal focal spot; 0.6/1.2mm)
time of use. The target disk coated Supporter B-80
with high heat emission material Fan F-3
enables stable X-ray output.
The CIRCLEX 0.3/0.8P324DK-85 &
0.6/1.2P324DK-85 are designed for
gastro-intestinal examination,
high-speed serialography.
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
SPECIFICATIONS
CIRCLEX CIRCLEX
0.3/0.8P324DK-85 0.6/1.2P324DK-85

Nominal focal spot 0.3mm 0.8mm 0.6mm 1.2mm

Nominal X-ray tube Fluorography 125kV

voltage Radiography 150kV

X-ray tube Max. heat content 1100kJ {1600kHU}

assembly Max. continuous heat dissipation 470W {660HU/s}

Max. anode heat content 280kJ {400kHU}

X-ray tube Max. anode heat dissipation rate 1600W {2250HU/s}

Max. continuous heat dissipation 300W {420HU/s}

X-ray tube Max. Anode stationary 200W {280HU/s} 355W {500HU/s} 355W {500HU/s} 425W {600HU/s}

Fluoroscopic load (10min.) Anode rotating 355W {500HU/s} 425W {600HU/s} 425W {600HU/s} 425W {600HU/s}

Nominal anode input power (0.1sec, 180Hz)


11kW 54kW 38kW 92kW
(0.1sec.)

Max. filament voltage 7.4V 14.3V 12.5V 18.4V

Max. filament current*1 5.0A 5.6A 5.6A 5.6A

Material Rhenium-tungsten faced molybdenum

Anode target Angle 12°

Diameter 100mm (4inch)

Direction of anode rotation is counterclockwise as viewed from the


cathode side, and R.P.M are as follows.
Anode rotation*2 2700min.-1 at 50Hz
3200min.-1 at 60Hz
9700min.-1 at 180Hz

Equivalent filtration Min. 1.5mm Al equivalent

Leakage radiation in hour from the X-ray tube assembly and collimator
Is less than 0.87mGy at a distance of 1 meter from the focal spot.
Leakage radiation
However, leakage radiation in an hour from the collimator is less than
0.304mGy

X-ray radiation field 350×350mm (13.8×13.8inch) at 1000mm (39.4inch) SID

Min. ambient temperature -10℃

The thermal switch actuates to break the high voltage generating circuit
Safety circuit
when the temperature inside the tube housing reaches 80℃.

Weight 22.5kg (49.5) (inclusive of supporter and fan)

*1 Refers to the maximum value usable in tube current regulation, and its limit in constant use is less than 4.6A for
focus of 0.3mm, less than 5.2A for focus of 0.8mm, and less than 5.2A for both of 0.6mm focus and 1.2mm focus.
*2 Combination with the optional SA-40, SA-41, SA-42, SA-50 and SA-60, starter.

2
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
HEATING AND COOLING CHARTS

X-ray Tube anode

X-ray Tube
X-RAY TUBE ASSEMBLY assembly
HEATING AND COOLING CURVE

1600

470W
X-ray tube assembly heat content (kHU)

1400 600W

1200

300W
1000

800

600

Cooling
冷却
400

200

0
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
Time (min)

3
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
RADIOGRAPHIC RATING CHARTS

Three Phase - Full Wave

0.3mm Focal Spot, 180Hz 0.8mm Focal Spot, 180Hz

0.6mm Focal Spot, 180Hz 1.2mm Focal Spot, 180Hz

4
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
EMISSION CHARTS
The charts show the emission characteristic under three phase full wave operation.
In the case of single phase full wave operation, mA values are decreased by approximately 30% for same filament current.

0.3mm Focal Spot 0.8mm Focal Spot

0.6mm Focal Spot 1.2mm Focal Spot

Emission & Filament Characteristics Emission & Filament Characteristics


0.6P324DK 1.2P324DK
600 14 1,000 18
100kV
900 16
500 12 80kV
800 14
125kV
X-ray Tube Current (mA)

X-ray Tube Current (mA)

100kV 80kV 700 12


Filament Voltage (V)

Filament Voltage (V)


400 10 150kV 60kV
e
600
ta g 10
125kV 60kV
t Vol
300 8 500 m en 8
150kV Fila 40kV
40kV 400 6
ge
200 Volta 6
nt
me 300 4
Fila
200 2
100 4
100 0

0 2 0
3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4
Filament Current (A) Filament Current (A)

5
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
HIGH SPEED SERIALOGRAPHIC RATING CHARTS
N.B. The load per exposure must be less than 70% or 80% of the maximum rating
given by the radiographic rating charts.

Load per exposures (HU)


= kV×mA×s ( )
or kV×mA×s×1.35 ( 6 pulse )
or kV×mA×s×1.41 ( 12 pulse or )

Load per exposures (J)


= kV×mA×s×0.71 ( )
or kV×mA×s×0.96 ( 6 pulse )
or kV×mA×s ( 12 pulse or )

DIMENSIONS
245 (9.6) Max. 20 (0.8)

145.2 (5.7)
φ46 (1.8)

φ107 (4.2)
Focal Track
φ84 (3.3)

φ136 (5.4)

1.5 (0.06) 103 (4.1) 69 (2.7)


21 18
(0.8) (0.7)

X-ray Tube UNIT : mm (inch)

6
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
DIMENSIONS (cont.)

131 (5.2) 84.5 (3.3)


229 (9.0)

92 (3.6)
(2.0)
52
focal

100 (3.9)
spot

57 ± 1
(2.2)
464 (18.3) Reference axis
240 (9.4)
161 (6.3) 140 (5.5)
4-M6 12 deep U63.5
P.C.D. 92 20 Mount
φ184 (7.2)

φ158 (6.2)

15 (0.6)
148 (5.8)
UNIT : mm (inch)

X-ray Tube Assembly

UNIT : mm (inch)

Supporter and Fan

7
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85

Remarks
* Every value in this Product Date Sheet is a standard value, and it may vary a little from the actual at each site.
* The appearances and specifications are subject to change for reasons of improvement without notice.
* Certain configurations may not be available pending regulatory clearance. Contact your Shimadzu
representative for information on specific configurations.

SHIMADZU CORPORATION. International Marketing Division


Shimadzu Corporation Medical Systems Group
3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax: 81(3)3219-5710 has been certified by TUV Rheinland as a
Cable Add.: SHIMADZU TOKYO manufacturer of medical equipment and systems in
compliance with ISO9001: 2000 Quality
URL http://www.shimadzu.com Management Systems and EN ISO13485: 2000
Medical Equipment Quality Management Systems.