Você está na página 1de 2

SAMPLING PROCESS

One mechanical arm performs all sampling operations with:


G liquid level sensing (capacitive)
G reagent pre-warming at 37° C
G automatic probe washing
G sample volume range: 2.0 µl ÷ 99 µl (0.25 µl incr.)

G reagent 1 volume range: 2.0 µl ÷ 348 µl (1 µl incr.)


LIST OF DEDICATED APPLICATIONS SPECIFICATIONS
G reagent 2 volume range: 2.0 µl ÷ 330 µl (1 µl incr.)
G reagent 3 volume range: 2.0 µl ÷ 330 µl (1 µl incr.)
The SAT 450 has all parameters set based on SYSTEM DESCRIPTION
G sample pre-dilution
AMS reagent kits given below. Benchtop, fully automatic, random access, open analyzer for G sample post-dilution
clinical and immunoturbidimetric assays G sample post-concentration

ASSAY TYPES
CLINICAL CHEMISTRY AND IMMUNOTURBIDIMETRY REACTION PLATE
G End point, Kinetic, Differential, Bichromatic, holds 80 cuvettes divide into 4 racks with 20 cuvettes each
Fixed Time and Immunoturbidimetric and performs the following operations:
Code Description Method
G up to 999 tests, all user programmable
G automatic cuvette washing

SA4175 00 AMYLASE L Enzymatic CNPG3 REAGENTS ON BOARD G continuous monitoring of cuvette quality
SA4201 00 ALBUMIN L Colorimetric (BCG) G incubation temperature control at 37° C ± 0.3
Up to 72 reagents can be located into the refrigerated
SA4231 00 BILIRUBIN Total Colorimetric (Jendrassik-Groff)
compartment and distribuited on: COMPUTER (minimum requirements)
SA4256 00 BILIRUBIN Direct Colorimetric (Jendrassik-Groff)
G 4 removable racks
G CPU: Intel or AMD
SA4323 00 CALCIUM ARSENAZO Colorimetric
G 50/20/5 ml containers can be loaded
SA4325 00 CALCIUM OCP L Colorimetric G MEMORY: Ram 2 Gb
G 8 extra positions for 5 ml containers
SA4330 00 CHLORIDE L Colorimetric G HARD DISK: 250 Gb

SA4340 00 TOTAL CHOLESTEROL L Enzymatic colorimetric SAMPLE LOADING G MONITOR: 17“


G COMMUNICATION: USB and/or RS232 port
SA4375 00 HDL DIRECT L Direct enzymatic G 4 universal, separate racks for continuos loading of samples,
SA4380 00 LDL DIRECT L Direct enzymatic G HOST COMMUNICATION: RS232 or Ethernet
calibrators and controls
HOMOCYSTEINE L Enzymatic G SOFTWARE: WINDOWS XP - VISTA - 7 (32 bit only)
G each rack contains up to 15 cups and/or tubes 10-16 mm
SA4410 00
SA4420 00 ENZYMATIC CREATININE L Enzymatic colorimetric
diam., 40-100 mm height OPTIONS
SA4450 00 CREATININE L Jaffè-mod.
G 8 extra position for urgent samples, calibrators and controls
SA4544 00 GAMMA-GT L Enzymatic G Direct potentiometry I.S.E. module for Na+, K+, CI- (140
G STAT sample execution is allowed at any time
SA4546 00 LDH - L Enzymatic DGKC (P->L) tests/hour)
SA4565 00 PHOSPHOROUS UV L Enzymatic THROUGHPUT G Positive barcode readers for reagents and samples (code

SA4575 00 GLUCOSE L Enzymatic colorimetric Up to 420 tests/hour including I.S.E. 128, codebar, code 2 of 5 interleaved, code 39)
SA4645 00 MAGNESIUM XB L Colorimetric (277 tests/hour photometric) INSTALLATION REQUIREMENTS
SA4710 00 TOTAL PROTEIN L Colorimetric (Biuret)
SA4740 00 PYROGALLOL PROTEIN L Colorimetric READING SYSTEM Power Supply: 90 ÷ 250 VAC, 47 ÷ 63 Hz
SA4815 00 TRIGLYCERIDES L Enzymatic colorimetric G direct reading, dual channel photometer Electric consumption: 386 Watt
SA4865 00 URIC ACID L Enzymatic colorimetric G 200 µl minimum reading volume Water Consumption: 1.5 l/h
SA4886 00 IRON FERENE L Colorimetric G 9 narrow band, automatically selected interferential filters DIMENSIONS, WEIGHT & ENVIRONMENT
SA4910 00 GOT/AST - L Enzymatic IFCC-SCE 340, 380, 405, 492, 510, 546, 577, 620, 690 Analytical Module:
SA4920 00 GPT/ALT - L Enzymatic IFCC-SCE G halogen lamp 6V/10W
G Height: 50 cm
G linear from 0.0001 to 3.200 Abs.
SA4933 00 ALP - L Enzymatic DGKC
G Width: 68 cm
SA4960 00 UREA UV L Enzymatic G high resolution (0.0001 Abs.)
G Lenght: 105 cm
SA4975 00 CHOLINESTERASE Enzymatic DGKC
G Weight: 49 Kg
SA5418 00 ENZYMATIC HbA1c Direct Enzymatic
AMS SpA Room Temperature: 18° C ÷ 30° C
SA5424 00 GLICOTEST HbA1c Direct Latex immunoturbidimetric
Instruments Factory, Sales & Customer Service
SA5432 00 CK-NAC L Enzymatic IFCC Via E. Barsanti 17/A
Relative Humidity: 20% ÷ 85%
SA8422 00 ASO Latex L Latex immunoturbidimetric 00012 Guidonia - Rome (Italy) All the above features can be changed without prior notice.
SA8427 00 CRP Latex L Latex immunoturbidimetric Tel. +39 0774 354441
SA8432 00 RF Latex L Latex immunoturbidimetric Fax +39 0774 578035 IVD

Zesi Tipografia srl - Tel. +39 06 4513002


www.ams-analyzers.com
SA8437 00 FERRITIN Latex L Latex immunoturbidimetric
info@ams-analyzers.com
SA8439 00 MICROALBUMIN Latex L Latex immunoturbidimetric
SA8472 00 IgA L Immunoturbidimetric Reagents Factory, Sales & Customer Service
SA8482 00 IgG L Immunoturbidimetric Via G. Galilei 38
SA8485 00 C3 L Immunoturbidimetric 20096 Seggiano di Pioltello - Milan (Italy)
Tel. +39 02 9291891
SA8487 00 IgM L Immunoturbidimetric
Fax.+39 02 92918939
SA8490 00 C4 L Immunoturbidimetric www.ams-analyzers.com AMS is a Company of Group
SA8494 00 TRANSFERRIN L Immunoturbidimetric infodiagnostics@ams-analyzers.com
SAT 450 is a fully automatic, random access analyzer designed to support the modern laboratory in achieving
improved cost efficiency, reduced rerun and manual handling. It provides quality results while following rigid quality
system procedures and in strict compliance with regulatory agency requirements.

1
I.S.E. MODULE (OPTIONAL) SOFTWARE OVERVIEW
The SAT 450 operating software runs under multitasking
SIMPLICITY, PRODUCTIVITY, WINDOWS XP, VISTA and 7 (32 bit only).
QUALITY RESULTS Its bright graphical user interface simplifies routine work and
· direct measurement
offers the operator a concise guide through the operating
of Sodium, Potassium
and Chloride functions.
· easy access for MONITOR SCREEN (1)
reagent replacement
G real time presentation of samples and reagents status
· maintenance-free electrodes
G dynamic monitoring of test execution
a single, heated level
sensing arm minimizes G analyzer detailed status with error messages, incubation 2
mechanical complexity temperature control, START, STOP, and PAUSE functions
and rapidly moves to all
METHODOLOGY PROGRAMMING (2)
position for:
G
G optimized sample all parameters are completely user programmable
and reagent mixing G up to 999 methods can be memorized
COST EFFICIENCY G all settings are user programmable
G up to 3 different reagents
addition to each method G user selectable pre-dilution and re-dilution ratio
G user selectable samples re-run option
G 80 long life, 6 mm optical path,semi-disposable cuvettes G test calculation models are user selectable: Factor, Stan-
G sample pre-dilution (4 separate racks with 20 cuvettes each) dard, Linear Fit, Point to Point, Quadratic, Cubic, Quartic,
G sample rerun with or without dilution G selective cuvette wash with 3 different
Cubic Inv., Log-Logit, Log-Log Cubic
G probe double washing between each sampling cycle liquids: acid, alkaline and neutral
G inner and outer probe washing after every cycle
3
G low water consumption: 1.5 l/h AUTOMATIC DETECTION OF:
G 200 µl minimum reaction volume G end point limit
G cuvette quality automatic check G substrate depletion
ON BOARD REAGENTS G fixed time and kinetic reaction linearity

GRAPH SCREEN (3)


G display of test reaction curves
G display of test calibration curves
G Levy-Jennings QC statistic
PRACTICAL DESIGN
DIAGNOSTIC SOFTWARE (4)
G each instrument part can be singularly tested
G remote diagnostic capability 4
FLEXIBILITY
MORE FEATURES
advanced technology and
G multilanguage files (user programmable)
sophisticated system G 4 universal racks for cups and primary tubes G bi-directional communication compliant with ASTM and CLSI
enginering allow easy
G continuous loading of up to 15 samples on standard
access for maintenance
each rack (serum, plasma, urine) G automatic transmission of patients data and results through
G up to 64 reagents on 4 separate racks purposes by simply
G additional 8 positions
lifting up the entire serial port or Ethernet
G additional 8 position G STAT samples loading allowed at any time G reagents data traceability and management
G on board storage at 8 °C ÷ 12 °C
light-weight analytical chassis
G standards and controls can be placed on the
G barcode tracking of reagents (optional) same racks with no limitations
G possibility to close the system with dedicated reagents (optional) G positive identification of samples (optional)