Escolar Documentos
Profissional Documentos
Cultura Documentos
Submitted by:
GROUP 1
Cusipag, Jonelou A.
Decena, Clarissa H.
Donato, Redfield M.
Submitted to:
October 2018
Abstract
I. Introduction ...................................................................................................................... 1
A. Objectives .................................................................................................................... 2
A. Recommendations...................................................................................................... 30
B. Conclusions ............................................................................................................... 30
List of Figures
Figure 1.Activated carbon, a) milled carbon which is powder, b) granulated active carbon .... 3
Figure 2. Fixed bed adsorber ..................................................................................................... 4
Figure 3.Preparation of Reference Solutions setup.................................................................... 9
Figure 4. Adsorption Isotherm Determination setup ............................................................... 10
Figure 5. Design of Fixed-bed Adsorption schematic diagram ............................................... 11
Figure 6. Schematic diagram for preparation of reference solutions ....................................... 12
Figure 7.Schematic diagram for adsorption isotherm determination....................................... 13
Figure 8. Schematic diagram for design of fixed-bed adsorption ............................................ 14
Figure 9.Density versus Concentration .................................................................................... 16
Figure 10. Subdivided Equations for Density-Concentration .................................................. 17
Figure 11. Breakthrough curve for Height =10 cm at 1% (w/w) dye solution ........................ 25
Figure 12. Breakthrough curve for Height =10 cm at 0.5% (w/w) dye solution ..................... 26
Figure 13. Breakthrough curve for Height =5 cm at 1% (w/w) dye solution .......................... 28
Figure 14. Breakthrough curve for Height =5 cm at 0.5% (w/w) dye solution ....................... 29
List of Tables
Separation occurs because differences in molecular weight, shape, or polarity cause some
molecules to be held more strongly on the surface than others because the pores are too small
to admit the larger molecules. In many cases, the adsorbing components, or adsorbate, is held
strongly enough to permit complete removal of that component from the fluid with very little
adsorption of other components. (McCabe, Smith, & Harriott, 1993)
Adsorption can also be defined based on the strength of interaction between the adsorbent
and adsorbed molecules. The first definition is through a process that includes intermolecular
forces specifically, Van der Waals, interaction between the substrate and the adsorbate. It
would be easy to recognize this when the phenomenon is a general one and it occurs in any
fluid/solid system and when the evidence for the perturbation of the electronic states of
adsorbent and adsorbate is minimal. Also, when the adsorbed species are chemically identical
with those in the fluid phase, so that the chemical nature of the fluid is not altered by adsorption
and desorption. This is called physisorption. The other definition of adsorption lies as
chemisorption which is defined as chemical bonds involved in sticking the adsorbate to the
adsorbent. When the phenomenon is characterized by chemical specificity and when the
changes in the electronic state may be detectable by suitable physical means, it is known to be
chemisorption.
1
The process opposite of adsorption is known as desorption or stripping. This is
characterized by the release of a substance from another from the surface or through the surface.
When an equilibrium situation between a bulk phase and an adsorbing surface is altered,
desorption occurs. After adsorption, the adsorbed chemical will stay on the substrate
indefinitely, if the temperature remains low. However, as temperature rises, desorption is meant
to occur. (Geankoplis, Hersel, & Lepek, 2018)
A. Objectives
Generally, the objective of this experiment is to evaluate the performance of a packed bed
of activated carbon adsorbent to purify or cleanse varying concentrations of dye solution.
Specifically, it aims to:
Adsorption isotherms show the dependence of the amount of adsorbed substance from the
solution concentration at constant temperature. The adsorption capacity of adsorbent is
determined by using the adsorption isotherm models. The most commonly used iso-therms for
2
modelling adsorption processes in liquid phase are the Freundlich and Langmuir isotherm
models. Adsorption isotherms help to design experimental system and to evaluate it, based on
isotherm models, such as Freundlich and Langmuir isotherm models, and to find out a deviation
between experimental data and isotherm models. Adsorption isotherms are helping to describe
surface processes of adsorbent, type of adsorbate monolayer or mul-tilayer and capacity of
adsorbent, which were used for this work. (Willis, 2009), (Repo, 2011)
The Langmuir isotherm derived from assuming a uniform surface applies well for gases
that are weakly adsorbed given by the relation W=Wmax (Kc/(1+Kc)); whereas Freundlich
equation W=bcm, where b and m are constants and m<1, fits well for liquids.
Activated carbon (Fig. 1) is the well-known adsorbent which is widely used in many
industries, such as water treatment and medical production. It has exceptionally strong
adsorption properties and complicated structure. Activated carbon can be produced by physical
or chemical activation from material such as coal, lignite, nutshells, peat, wood and petroleum
pitches.
Figure 1.Activated carbon, a) milled carbon which is powder, b) granulated active carbon
A fixed-bed adsorption system is a widely used method for the adsorption of solutes
from fluids. Granular particles are packed inside a fixed-bed. The liquid to be treated is passed
down through the packed bed at a constant flow rate. But in cases of gases, it works as the
opposite, it starts from the bottom and with the use of a blower, it would go up. Resistance to
mass transfer are important in the process and this would be in an unsteady state. The efficiency
of the process is determined by the overall dynamics of the system rather than just the
equilibrium considerations.
3
Figure 2. Fixed bed adsorber
The major part of adsorption takes place in a relatively narrow adsorption or mass-
transfer zone at any time. As the solution continues to flow, the S-shaped mass-transfer zone
moves down the column. The concentration in the outlet becomes zero until the mass-transfer
zone starts to reach the tower outlet at a given time t4. Then, an outlet concentration starts to
rise and at time t5, the outlet concentration has risen to a point called the break-point.
After the break-point time is reached, the concentration rises rapidly up to the end of
the breakthrough curve where the bed is considered ineffective. The break-point concentration
represents the maximum that can be discarded and is often taken as 0.05 to 0.1.
4
III. Materials and Methodology
A. Materials and Chemicals
The materials used for this experiment are shown in Table 1 as follows:
5
Without a funnel, spillage may occur. a tube
or pipe that is wide at the top and narrow at
the bottom, used for guiding liquid or powder
into a small opening.
Other materials not mentioned in the table are the following: filter paper, funnel, stop watch
and ruler. The chemical used in this experiment are activated carbon which is the adsorbent in
granular form of about 200 grams, a dyeing agent, the adsorbate, in powder form (of about 15-
20 packs, each pack consisting 2 grams) and distilled water as the solvent (we used at most 6L
for the whole procedure) which is used for the preparation for the different dye concentrations.
6
B. Experimental Procedure
i. Preparation of Reference Solutions
For the preparation of reference solution, prepare 250 ± 1 mL with 0, 0.2, 0.4, 0.6, 0.8 and
1% w/w dye solution using distilled water as the solvent, to do this 0 g of dye must be added
to 250 ± 1 mL distilled water to produce 0% w/w dye solution, 0.510 ± 0.01 g of dye must be
added to 250 ± 1 mL distilled water to produce 0.2% w/w dye solution, 1.01 ± 0.01 g of dye
must be added to 250 ± 1 mL distilled water to produce 0.4% w/w dye solution, 1.51 ± 0.01 g
of dye must be added to 250 ± 1 mL distilled water to produce 0.6% w/w dye solution, 2.02 ±
0.01 g of dye must be added to 250 ± 1 mL distilled water to produce 0.8% w/w dye solution,
2.53 ± 0.01 g of dye must be added to 250 ± 1 mL distilled water to produce 1.0% w/w dye
solution. After that, determine the density of the individual solution at room temperature using
the hydrometer, this is done by placing the solution in a 250 ± 1 mL graduated cylinder then
let the hydrometer float in it and then read the measurement. Next, make a plot with the
concentration as a function of density using Microsoft excel. Then, determine the best equation
that represents the experimental data by getting the equation with R value nearest to 1.
7
iii. Design of Fixed-bed Adsorption
For the design of a fixed-bed adsorption, it was represented onto a burette loaded with a
predetermined amount of activated carbon until it reaches a compacted height of 10 cm. Then
the weight of burette and the weight of burette filled with 10 cm amount of activated carbon
was measured. The weight of activated carbon was obtained by subtracting the weight of the
loaded burette of 10 cm activated carbon minus that to the weight of burette. After the 10 cm
load of activated carbon in a burette, it was set-up in the iron stand.
A new 1% w/w dye solution was prepared by using 250 mL of distilled water and 2.54 ±
0.01 g of dye. The solution of 1 % w/w dye solution was filled in a graduated cylinder. After
that, the burette’s stopcock was set quarterly open. The 1% w/w solution was poured into the
burette containing the 10 cm load of activated carbon. The timer was started to about 10
minutes and the effluent volume was recorded using a 25 mL graduated cylinder. The mass of
the 25 mL graduated cylinder containing the effluent volume was measured using a balance.
The mass also of the 25 mL graduated cylinder was measured. The final mass of the effluent
was determined by subtracting the mass of the 25 mL graduated cylinder to the mass of the
empty 25 mL graduated cylinder. This is necessary to determine the average volumetric flow
rate and the density of the effluent. For the average volumetric flow rate expressed in mL/min,
it can be determined by dividing the effluent volume to that of the time which is 10 min. And
for the density expressed in g/mL, it is determined by dividing the final mass of the effluent to
that of the final volume of the effluent. The determination of density of the effluent was done
for number of trials of set of 10 minutes until the density of the effluent is constant.
The same set of procedures were done also for 0.5% w/w solution and using 10 cm of
burette-loaded activated carbon referring to the packed bed height , packed bed height of 5 cm
and 0.5% w/w dye solution and lastly packed bed height of 5 cm and 1% w/w dye solution.
The preparation of 0.5 % w/w solution was prepared by using 1.26 ± 0.01 g of dye and 250 mL
of distilled water. The density determined is correlated by plotting the density versus the
concentration of the effluent. The breakthrough curve is then generated by plotting C/Co versus
time. Lastly, height of used bed (HB) and height of unused bed (HUB) is determined by using
the formula indicated in the literature.
8
I. Preparation of Reference Solutions
Hydrometer
Decanted
solution
9
II. Determination of Adsorption Isotherm
Decanting
Solution
Activated
carbon
10
III. Design of Fixed-bed adsorption
Effluent
11
1. Preparation of Reference Solutions
Prepare
0
0.2% w/w (0.501 g activated carbon + 250 ml water);
0.4% w/w (1.004 g activated carbon + 250 ml water);
0.6% w/w (1.509 g activated carbon + 250 ml water);
0.8% w/w (2.016 g activated carbon + 250 ml water);
1% w/w (2.525 g activated carbon + 250 ml water)
dye solution. Use distilled water as the solvent.
12
2. Adsorption Isotherm Determination
13
3. Design of Fixed-bed adsorption
14
IV. Results and Discussion
Gathered Data
Take note that the standard density of water at 1 atmospheric pressure and 25 oC is
approximately 1000 kg/m3 (1 g/ml). The volume of prepared water at each beaker is 250 ml
therefore, the mass of water being calculated for each beaker is 250 g. The mathematical
representation for mwater calculation is;
mass water
* volume
1g
mass water
* 250ml
ml
mass water
250 g
This explains why the uncertainty for the mass of water is in terms of volume instead of
mass.
15
ii. Concentration VS. Density Plot
Density Concentration
0.985 0
0.99 0.00204
0.993 0.00402
0.996 0.006
0.997 0.00802
0.999 0.01
Table 3. Concentration VS. Density
Since the R2 does not approaches 1 close enough, the data representing the graph will be
cut into three parts.
16
iii. Determination of best equation that represents the experimental data using
Microsoft Excel
0.01
0.01
y = 0.99x - 0.979
0.00802 R² = 1
0.008
0.006 R² = 1
0.00402
0.004 y = 0.66x - 0.6514
R² = 1
0.00204
0.002
0 y = 0.408x - 0.4019
0 R² = 1
0.98 0.99 1 1.01 1.02 1.03 1.04 1.05 1.06
-0.002
Density (g/ml)
The corresponding equation of every part will be identified with respect to its interval. This
method will be used for the reason that the graphing by parts is more accurate than graphing
by whole. We take into consideration that for every equations hold, a corresponding density
limit must be applied.
Therefore
At 0.985<density<0.99, Eqn 1 will be used
At 0.99<density<0.996, Eqn 2 will be used
At 0.996<density<0.997, Eqn 3 will be used
At 0.997<density<0.999, Equation 4 will be used
17
Where y-concentration and x-density
Density
Beaker Number
3 min 6 min 9 min
2 0.975 0.951
3 0.976 0.965
4 0.985 0.966
5 0.999 0.972
18
m10cm activated carbon 0.005 kg (± 0.5 g)
0.5% (w/w) dye solution:
mdye 1.26 ± 0.01 g
vwater 250 ml
Bed Height: 10 cm
Loading Concentration: 0.5% (w/w) dye solution
Stopcock Opening: Quarterly open
c. For bed height of 5 cm and loading concentration of 1% (w/w)
Bed Height: 5 cm
19
Bed Height: 5 cm
Table 4. Average volumetric flow rate for height 10 cm and 1% (w/w) dye
solution
Comparing Table 2 and Table 3, of the same bed height of 10 cm but of different initial
loading concentration, a higher volumetric flow rate is achieved for a loading concentration of
0.5% (w/w) than 1.0% (w/w) dye solution. For every loading concentration introduced into the
burette, it adsorbs as it flows down as it has contacted the adsorbent’s surface. When the flow
rate of the effluent’s solution for 0.5% (w/w) loading concentration is higher than 1% (w/w)
20
dye solution, less is adsorbed hence more flow rate flows. The same is also achieved for Table
6 and 7.
Volume of solution
Time(min) Volumetric Flowrate (mL/min)
(±0.1 ml)
10 83.9 8.394763343
20 70.0 7
30 60.4 6.03707
40 43.0 4.3025
Average Volumetric Flow Rate: 6.433583336 mL/min
Table 6. Average volumetric flow rate for height of 5 cm and 1% (w/w) dye solution
Volume of solution
Time(min) Volumetric Flowrate (mL/min)
(±0.1 ml)
10 115.0 11.5
20 144.0 14.4
30 122.0 12.2
40 116.5 11.65
50 119.8 11.98
Average Volumetric Flow Rate: 12.346 mL/min
Table 7. Average volumetric flow rate for height of 5 cm and 0.5% (w/w) dye solution
Comparing Table 2 and Table 3, of the same bed height of 10 cm but of different initial
loading concentration, a higher volumetric flow rate is achieved for a loading concentration
of % (w/w) dye solution.
21
(±0.01 g) (±0.01 g) (±0.1 ml) (±0.1 g/ml)
Mass of 25 mL
Mass of Volume of Density of
Graduated
Time(min) Solution solution Solution
Cylinder + Solution
(±0.01 g) (±0.01 g) (±0.1 ml) (±0.1 g/ml)
10 374.88 83.36 83.9 0.993
20 288.24 69.6 70.0 0.994
30 278.89 60.25 60.4 0.998
40 188.79 43.03 43.0 1.000
Table 10. Density of effluent for height 5 cm and 1% (w/w) dye solution
22
Mass of 25 mL Graduated Mass of Volume of Density of
Time(min) Cylinder + Solution Solution solution Solution
(±0.01 g) (±0.01 g) (±0.1 ml) (±0.1 g/ml)
10 477.43 113.85 115.0 0.990
20 577.33 143.136 144.0 0.994
30 483.91 121.268 122.0 0.994
40 477.62 116.1505 116.5 0.997
50 480.23 119.67 119.8 0.999
Table 11. Density of effluent for height 5 cm and 0.5% (w/w) dye solution
To correlate the density in terms of concentration, we use the 4 equations generated in part
A for which density limits are hold to determine the concentration for the analysis of
breakthrough curve which are the following:
Where:
Equation 1 …… y = 0.408x - 0.4019 y-concentration
Equation 2 …… y = 0.66x - 0.6514 x-density
Equation 3 …… y = 2.02x - 2.0059
Equation 4 …… y = 0.99x - 0.979
Therefore
At 0.985<density<0.99, Eqn 1 will be used
At 0.99<density<0.996, Eqn 2 will be used
At 0.996<density<0.997, Eqn 3 will be used
At 0.997<density<0.999, Equation 4 will be used
Density of Holds
Concentration
Time(min) Solution for Co C/co
(y)
(±0.1 g/ml) Equation
10 0.990 1 0.002124615 0.01 0.212462
30 0.992 2 0.00332 0.01 0.332
23
40 0.993 2 0.003980818 0.01 0.39808
50 0.996 3 0.006693333 0.01 0.66933
60 0.999 4 0.012094348 0.01 1.20943
70 1.000 4 0.014024308 0.01 1.40243
0.01
0.008
0.006
0.004
0.002
0
0.988 0.990 0.992 0.994 0.996 0.998 1.000 1.002
Density
Time(min) C/co
10 0.212462
20 0.332
30 0.39808
40 0.66933
50 1.20943
60 1.40243
Table 12. Breakthrough table for H=10 cm, Co= 1% (w/w) dye solution
24
Breakthrough Curve @H=10 cm, 1% (w/w) dye solution
1.6
1.4
1.2
1
C/Co
0.8
0.6 2.491 mL/min
0.4
0.2
0
0 10 20 30 40 50 60 70
Time
Figure 11. Breakthrough curve for Height =10 cm at 1% (w/w) dye solution
Figure 4 shows the breakthrough curve for height of 10 cm at 1% (w/w) dye solution. Based
from the graph, of an average volumetric flow rate of 2.491 mL/min, there is a steady increasing
of breakthrough concentration at a time from 10 to 30 min which indicates a slow uptake of
dye. As time progresses, the breakthrough curve becomes steep and breakthrough is reached
quickly because of the residence time of the dye solution in the column, which is long enough
for adsorption equilibrium to be reached at high flow rate.
Density of
Holds
Solution
Time(min) for C Co C/Co
(±0.1
Equation
g/ml)
10 0.986 1 0.00038833 0.005 0.077666
20 0.987 1 0.000835857 0.005 0.167171
30 0.995 2 0.0053 0.005 1.06
40 0.998 4 0.00902 0.005 1.804
25
Density VS. Concentration
0.01
0.009
0.008 R² = 0.9699
0.007
Concentration C
0.006
0.005
0.004
0.003
0.002
0.001
0
0.984 0.986 0.988 0.990 0.992 0.994 0.996 0.998 1.000
Density
Time(min) C/Co
10 0.077666
20 0.167171
30 1.06
40 1.804
Table 13. Breakthrough table for H=10 cm, Co=0.5% (w/w) dye solution
1
0.8 6.19652 mL/min
0.6
0.4
0.2
0
0 10 20 30 40 50
Time
Figure 12. Breakthrough curve for Height =10 cm at 0.5% (w/w) dye solution
26
Figure 5 shows the generated breakthrough curve for a bed height of 10 cm at 0.5% (w/w)
loading concentration. As can be seen from the figure, breakthrough curve becomes steeper as
time progresses which means that the contact time between the activated carbon and the dye
solution is minimized, leading to early breakthrough.
Density of Holds
Time(min) Solution for concentration, C Co C/Co
(±0.1 g/ml) Equation
10 0.993 2 0.00398 0.01 0.398
20 0.994 2 0.004828571 0.01 0.482857
30 0.998 4 0.009020679 0.01 0.902068
40 1.000 4 0.011115049 0.01 1.111505
0.008
0.006
0.004
0.002
0
0.992 0.993 0.994 0.995 0.996 0.997 0.998 0.999 1.000 1.001
Density
Time(min)
C/Co
10 0.398000003
20 0.482857143
30 0.902067891
40 1.111504939
27
Table 14. Breakthrough table for H=5 cm, Co=1% (w/w) dye solution
0.8
C/Co
0.6
6.43358 mL/min
0.4
0.2
0
0 10 20 30 40 50
Time
Figure 6 shows the measure of C/Co over time for height equal to 5 cm at 1% dye solution.
Slow uptake of dye is noticed in the initial stage and rate increased thereafter and finally
reached saturation. As flow rate increased, the breakthrough curve becomes steep and
breakthrough is reached quickly. This is because of the residence time of the dye solution in
the column, which is long enough for adsorption equilibrium to be reached at high flow rate.
Steeper breakthrough curve means that the contact time between the activated carbon and the
dye solution is minimized, leading to early breakthrough. Increasing the flow rate gave rise to
a shorter time for saturation.
Density of Solution
Time(min) Holds for Equation C Co C/Co
(±0.1 g/ml)
10 0.990 1 0.00202 0.005 0.404
20 0.994 2 0.00464 0.005 0.928
30 0.994 2 0.00464 0.005 0.928
40 0.997 3 0.00804 0.005 1.608
50 0.999 4 0.00992571 0.005 1.985142
28
Density VS. Concentration
0.012
0.01
Concentration C R² = 0.977
0.008
0.006
0.004
0.002
0
0.988 0.990 0.992 0.994 0.996 0.998 1.000
Density
Time(min) C/Co
10 0.404
20 0.928
30 0.928
40 1.608
50 1.985141903
Table 15. Breakthrough table for H=5 cm, Co=0.5% (w/w) dye solution
1.5
C/Co
1 12.346 mL/min
0.5
0
0 10 20 30 40 50 60
Time
Figure 14. Breakthrough curve for Height =5 cm at 0.5% (w/w) dye solution
29
Based from Figure 7, it can be concluded that there is rapid uptake of the dye solution in
the initial stage, after 10 seconds the rate decreases and the rate increased thereafter and finally
saturation is reached. The results indicate that at smaller bed height the effluent concentration
ratio increased more rapidly than for a higher bed height. An increase in the flow rate decreases
the volume of dye solution treated; consequently, this decreases the service time of the
activated carbon bed. The data satisfies from the experiment proves the theoretical assumption
of a break through curve (when the slope at the center part of the S curve stays constant, this
means the column saturation is achieved).
B. Conclusions
VI. References
Geankoplis, C., Hersel, A. H., & Lepek, D. H. (2018). Transport Processes and Separation
Process Principles (5th ed.). Prentice-Hall International, Inc.
Mall, I., Srivastava, V., & Agarwal, N. (2006). Removal of Orange-G and methyl violet dyes
by adsorption onto bagasse fly ash kinetic study and equilibrium isotherm analyses.
Dyes and Pigments, 210-223.
McCabe, W., Smith, J., & Harriott, P. (1993). Unit Operations of Chemical Engineering, 5th
edition. McGraw-Hill, Inc.
Perry, R., & Green, D. (2008). Perry's Chemical Engineers' Handbook. McGraw-Hill.
Repo, E. (2011). EDTA- and DTPA- Functionalized Silica Gel and Chitosan Adsorbents
bents for the Removal of Heavy Metals From Aqueous Solutions. Thesis,
Lappeenranta Uni-versity of Technology.
Wankat, P. C. (2016). Separation Process Engineering: Mass Transfer Analysis, 4th edition.
Prentice-Hall International, Inc.
Willis, T. (2009). Sorbents – Properties, Materials and Applications. New York: Nova
Science Publishers, Inc.
30
VII. Appendix
Raw Data for Bed height = 10 cm, C=1% (w/w) dye solution
Density Error
Concentration Error
Raw Data for Bed height = 5 cm, C=1% (w/w) dye solution
31
Density error
Concentration
Raw Data for Bed height = 10 cm, C=0.5% (w/w) dye solution
32
Density error
Concentration
Raw Data for Bed height = 5 cm, C=0.5% (w/w) dye solution
Density error
33
Concentration
34
ii. Determination of the average volumetric flow rate
35
Construction of breakthrough curve
36