1.SPECIFICATIONS
CONSTANT
50Kw 50Kw 80Kw
OUTPUT(Kw)
DUAL SPEED
OPTIONAL OPTIONAL
ROTOR STARTER
LOW / DUAL LOW / DUAL
(HIGH / LOW)
1.
POWER LINE REQUIREMENTS
50Kw Three-Phase
Line Voltage: 380VAC±5%
Line Frequency: 50/60Hz
Momentary Current: 100Amps
Stanby Current: 5Amps
80Kw Three-Phase
Line Voltage: 380VAC±5%
Line Frequency: 50/60Hz
Momentary Current: 150Amps
Stanby Current: 5Amps
PAGE 1 / 27
GEMSS Medical NHFG Service Manual
Minimum
Minimum Minimum
Recommend HFG
Recommend Recommend Groun Apparent
Main ed Momenta
Mode ed ed d Mains
Voltag Main ry
l Generator Distribution Wire Resistan
e Disconnect Line
Service Transformer Size ce
to Current
Rating Rating
HFG
RMX-
#4
550R 380VA #6
100A 100A 65kVA (25mm 0.17Ω
(50K C (15mm2)
2)
w)
RMX-
#4
750R 380VA #6
100A 100A 85kVA (25mm 0.13Ω
(50K C (15mm2)
2
w)
RMX-
#4
850R 380VA #6
150A 100A 105kVA (25mm 0.11Ω
(80K C (15mm2)
2
w)
2..INSTALL
INSTALL POWER LINE
PAGE 2 / 27
GEMSS Medical NHFG Service Manual
TRANSFORMER CONNECTION
RMX-850R
Transformer Connection
PAGE 3 / 27
GEMSS Medical NHFG Service Manual
RMX-550/750R(Low Rotor)
Transformer Connection
ROTOR CONNECTION
1 5.W(Com)
2 4.B(Main)
3 3.G(Phase)
5 1.Tube Temp
6 2.Tube Temp
PAGE 4 / 27
GEMSS Medical NHFG Service Manual
3. SETTING
DIP SWITCH Check(Product set up check)
CPU BOARD
PAGE 5 / 27
GEMSS Medical NHFG Service Manual
PAGE 6 / 27
GEMSS Medical NHFG Service Manual
LD2
LD1
VDD
GND
PAGE 7 / 27
GEMSS Medical NHFG Service Manual
PAGE 8 / 27
GEMSS Medical NHFG Service Manual
PAGE 9 / 27
GEMSS Medical NHFG Service Manual
CPU BOARD
PAGE 10 / 27
GEMSS Medical NHFG Service Manual
FILAMENT BOARD
PAGE 11 / 27
GEMSS Medical NHFG Service Manual
Adjust to VR1
TP3-GND (Factory Setting)
PAGE 12 / 27
GEMSS Medical NHFG Service Manual
J4
J4
J5
High Rotor : 360VAC
Rotor Power
Low Rotor : 200VAC
J6
200~220VAC Fan Power
PAGE 13 / 27
GEMSS Medical NHFG Service Manual
J7
PAGE 14 / 27
GEMSS Medical NHFG Service Manual
LED DESCRIPTION
START
RUN STOP
180Hz 60Hz
ON/OFF
START LED ON
STOP LED ON It is Standby during LOW ROTOR driving
60Hz LED ON condition.
START LED ON
STOP LED ON
“READY” BUTTON is pressed during
60Hz LED ON
LOW ROTOR driving condition.
RUN LED ON
ON/OFF LED ON
START LED ON
STOP LED ON
During HIGH ROTOR driving condition.
180Hz LED ON
PAGE 15 / 27
GEMSS Medical NHFG Service Manual
START LED ON
STOP LED ON
“READY” BUTTON is pressed during
180Hz LED ON
HIGH ROTOR driving condition.
RUN LED ON
ON/OFF LED ON
J10
High Rotor : 360~380VAC
200~380VAC
Low Rotor : 200~220VAC
PAGE 16 / 27
GEMSS Medical NHFG Service Manual
5V 10usec
PAGE 17 / 27
GEMSS Medical NHFG Service Manual
2V 10msec
Error 08 Fault_ mA
☞Check New Control mA BD or Control Main BD
Replacement of NEW CONTROL MAIN BD MODULE is recommended.
Error 10 Fault_Fil_Power
①Check if the voltage of J4 on N-Filament BD is measured as the order of
+12VDC,-12VDC, and GND
②Check if DC48~52V is measured on TP7 and TP8.
Error 11 Fault_Fil_Over
Replacement of N-FILAMENT BD is required.
Error 12 Fault_Fil_Low
①Check if FILAMENT of X-RAY TUBE is Pre-heating
②HIGHTENSION VOLTAGE CHECK
PAGE 19 / 27
GEMSS Medical NHFG Service Manual
The voltage shall be measured as above picture and more than DC520V shall be
measured. If the voltage is less than 520VDC, an ERROR can be occurred when
X-ray is exposure on high condition.
PAGE 20 / 27
GEMSS Medical NHFG Service Manual
Error 25 Fault_PW_CPU3
☞Check CPU Power Board
Same action is required as ERROR 23..
Error 26 Fault_PW_CPU4
☞Check CPU Power Board
Same action is required as ERROR 23.
Error 27 Fault_PW_CTL1
☞Check New Control Main BD
Check if the voltage of +12V -12V GND is supplied from J13 on NEW CONTROL
BD.
Error 28 Fault_PW_CTL2
☞Check New Control Main BD
Check if the voltage of +12V -12V GND is supplied from J13 on NEW CONTROL
BD.
Error 29 Fault_AD_WORK
☞Check CPU Board
Replacement of CPU BD is required.
Error 30 Fault_KV_EEPROM
☞Check CPU Board EEPROM
It occurs when the KV data at EEPROM of CPU BD is over than its limited value.
Replacement of EEPROM is recommended.
Error 31 Fault_MA_EEPROM
☞Check CPU Board EEPROM
It occurs when the mA data at EEPROM of CPU BD is over than its limited value.
Replacement of EEPROM is recommended.
Error 32 Fault_S_PREHEAT_EEPROM
☞Check CPU Board EEPROM
It occurs when the SMALL FOCUS PREHEATING data at EEPROM of CPU BD is
over than its limited value.
Replacement of EEPROM is recommended.
Error 33 Fault_L_PREHEAT_EEPROM
☞Check CPU Board EEPROM
It occurs when the LARGE FOCUS PREHEATING data at EEPROM of CPU BD is
over than its limited value.
Replacement of EEPROM is recommended.
Error 35 Fault_Feedback_High
☞Check CPU Board & New Control BD
It occurs when the value of FEEDBACK is too high.
Replacement of CPU BOARD and NEW CONTROL MAIN BD is recommended.
PAGE 21 / 27
GEMSS Medical NHFG Service Manual
6.CALIBRATION
mA CALIBRATION
PAGE 22 / 27
GEMSS Medical NHFG Service Manual
Large Focus
mA adjust
Small Focus
mA adjust
N-FILAMENT BOARD
PAGE 23 / 27
GEMSS Medical NHFG Service Manual
Kv CALIBRATION
FEEDBACK WAVEFORM
20Kv/1V
SMALL FOCUS TEST 20Ma/1V
TIME: 0.05S
VALUE WAVEFORM
80Kv
100mA
120Kv
100mA
140Kv
100mA
PAGE 25 / 27
GEMSS Medical NHFG Service Manual
20Kv/1V
LARGE FOCUS TEST 100mA/1V
TIME: 0.05S
VALUE WAVEFORM
80Kv
200mA
110Kv
320mA
80Kv
400mA
PAGE 26 / 27
GEMSS Medical NHFG Service Manual
7.BLOCK DIAGRAM