Escolar Documentos
Profissional Documentos
Cultura Documentos
2415170-100
Rev 0
© 2005
GE HUALUN Medical Systems
All Rights Reserved
Revision History
Rev Date Change Description
A April 2005 Draft for ME
0 Aug 2005 Initial Release
IMPORTANT SAVE THESE INSTRUCTIONS. . PLEASE READ THIS MANUAL BEFORE USING EverView 7500 C-Arm X-Ray Imaging System.
This manual may not be reproduced, in whole or in part, without the written permission of GE Healthcare.
Other product and company names mentioned herein are the property of their respective owners.
The contents of this document are accurate at the time of publication. However, changes in design and additional features can, at
any time, be incorporated in the hardware and software and may not be reflected in this version of the document. Contact GE
Healthcare Technical Support for clarification, if discrepancies arise.
The text of this manual was originally written, approved and published by the manufacturer in English.
GE HUALUN Medical Systems, a General Electric company, going to market as GE Healthcare.
Page II
Introduction and Safety
REGULATORY REQUIREMENTS
This product complies with the regulatory requirements of the following:
Council Directive 93/42/EEC concerning medical devices: the CE 0459
label affixed to the product testifies compliance to the Directive.
The location of the CE 0459 label is described in page 6-8 of Operator manual.
European registered place of business:
GE Medical System Europe
Quality Assurance and safety Regulatory Manager
BP34
Address: 283 Rue de la Miniere 78533 Buc Cedex France
Tel:+33 (0) 1 30 70 40 40
International Electrotechnical Commission (IEC), international standards organization, when
applicable.
Page III
EverView 7500 C-Arm X-Ray Imaging System
Page IV
Introduction and Safety
本维修手册仅存有英文版本。
警告 非 GEMS 公司的维修人员要求非英文版本的维修手册,客户需自行
负 责翻译。
未详细阅读和了解本维修手册之前,不得进行维修。
忽略本注意事项会对维修人员,操作人员或病人造成触电、机械伤
害或其他伤害。
Page V
EverView 7500 C-Arm X-Ray Imaging System
Contents
Chapter 1 Introduction and Safety.........................................................................................1-1
1.1 Introduction .......................................................................................................................................1-2
1.1.1 Objective................................................................................................................................................... 1-2
1.1.2 Scope.......................................................................................................................................................... 1-2
1.1.3 Service Philosophy .............................................................................................................................. 1-2
1.1.4 Qualifications ......................................................................................................................................... 1-2
1.1.5 Unauthorized Modifications............................................................................................................ 1-2
1.1.6 Certified Components........................................................................................................................ 1-2
1.1.7 Manual Organization and Use ....................................................................................................... 1-3
1.2 Safety ...................................................................................................................................................1-3
1.2.1 Safety Hazard Alerts........................................................................................................................... 1-3
1.2.2 Ingress of Fluids .................................................................................................................................... 1-4
1.2.3 Electrical Shock ..................................................................................................................................... 1-4
1.2.4 Implosion.................................................................................................................................................. 1-5
1.2.5 Explosion .................................................................................................................................................. 1-5
1.2.6 Thermal Harm........................................................................................................................................ 1-5
1.2.7 Equipment Stability ............................................................................................................................. 1-5
1.2.8 X-Ray Radiation .................................................................................................................................... 1-6
1.2.9 Motorized Mechanical Movement................................................................................................ 1-6
1.2.10 Safety Interlock .................................................................................................................................. 1-6
1.2.11 Labels ...................................................................................................................................................... 1-7
1.2.12 Symbols................................................................................................................................................1-14
1.3 Electromagnetic Compatibility Statement ....................................................................... 1-15
Chapter 2 Systems OverView ...................................................................................................2-1
2.1 System Descriptions......................................................................................................................2-2
2.2 Motions of the C-Arm ....................................................................................................................2-4
2.3 EverView 7500 C-Arm X-Ray Imaging System Block .....................................................2-5
2.4 EverView 7500 C-Arm X-Ray Imaging System Schematics ........................................2-6
Chapter 3 Installation....................................................................................................................3-1
3.1 Pre-Installation .................................................................................................................................3-2
3.1.1 Dimensions and Weights of System........................................................................................... 3-2
3.1.2 Site Requirement .................................................................................................................................. 3-2
3.1.3 Power Supply Requirements........................................................................................................... 3-2
3.1.4 Environments Requirements........................................................................................................... 3-3
3.2 Unpacking ..........................................................................................................................................3-3
3.3 Power Supply Adaptation............................................................................................................3-9
3.3.1 OverView .................................................................................................................................................. 3-9
3.3.2 Change the System Frequency..................................................................................................... 3-9
3.3.3 Change the Power Supply Cables and plugs .......................................................................3-10
3.3.4 Voltage Adaption ................................................................................................................................3-12
3.4 Software Configuration............................................................................................................. 3-13
3.5 Options Installation ..................................................................................................................... 3-14
3.5.1 Sony Video Printer Installation.....................................................................................................3-14
3.5.2 DAP Installation ...................................................................................................................................3-15
3.6 Installation Check......................................................................................................................... 3-17
3.6.1 Label and Symbol Inspection.......................................................................................................3-17
3.6.2 Visual inspection.................................................................................................................................3-18
3.6.3 Mechanical Inspection.....................................................................................................................3-25
3.6.4 System on check ................................................................................................................................3-26
Page VI
Introduction and Safety
Page VII
EverView 7500 C-Arm X-Ray Imaging System
Page VIII
Introduction and Safety
Page IX
Chapter 1
Introduction and
Safety
EverView 7500 C-Arm X-Ray Imaging System
1.1 Introduction
1.1.1 Objective
Provide EverView 7500 C-Arm X-Ray Imaging System service documentation that is consistently
organized, easy to access and is designed to aggressively expedite fault isolation and decrease
product down-time.
1.1.2 Scope
This manual’s contents document the necessary information for service.
1.1.4 Qualifications
Reading this manual does not qualify an untrained person to service the equipment. This
publication assumes the reader is a skilled and fully qualified electronic technician who has
experience servicing medical imaging systems. Factory trained service personnel will benefit the
most from this material.
Page 1-2
Introduction and Safety
1.2 Safety
Do not attempt to service or maintain this equipment before reading this entire section.
CAUTION Caution indicates a potentially hazardous situation which, if not avoided, may
result in moderate to minor injury, equipment damage or loss of data.
Page 1-3
EverView 7500 C-Arm X-Ray Imaging System
The equipment is not rated for water-tight operation. Do not allow any liquids to
WARNING
leak into the equipment where they can cause electric shock, fires, and short
circuits. If leakage does occur, ensure that all internal circuitry is completely dry
before attempting to operate the equipment.
Never store or operate the system in the presence of conductive fluids like water
WARNING or saline solution unless the system is adequately protected by an approved
bagging or draping system.
The equipment must be serviced only by personnel familiar with proper safety
WARNING
procedures for working near these components.
Have someone watch while you work near high voltages. This person must
WARNING remain clear of all circuitry and be prepared to turn off the power to the system
and render aid in an emergency.
High voltage cables, electrolytic capacitors, and CRTs can retain a dangerous
WARNING static charge for long periods after power has been removed. Some devices can
acquire a charge spontaneously without direct contact with other circuitry. Do
not touch these components unless power has been completely removed and
they have been discharged.
Never operate the system unless all potential equalization conductors (green
CAUTION wire with a yellow stripe) and related components are fastened in place.
Dangerous electrical shock and improper equipment operation can otherwise
result.
Remove all metal rings and watchbands before working on system circuitry.
CAUTION Skin burns and damage from involuntary muscle contractions can result if metal
jewelry shorts electrical circuits.
Page 1-4
Introduction and Safety
1.2.4 Implosion
EverView 7500 C-Arm X-Ray Imaging System has a Cathode Ray Tube (CRT) , do not locate
objects so that they might fall and strike the tube causing it to implode. Use caution when
working around the tube. The coating on the glass can also produce toxic dust and fumes. If a
CRT implodes:
1. Remove power immediately and evacuate the area.
2. Contact site administration regarding hazardous material disposal.
1.2.5 Explosion
If an abnormal condition occurs, such as the room fills with flammable gas, steps must be taken
to prevent the gas from contacting components within the equipment. Follow these guidelines:
1. Do not turn the system on or off.
2. Do not un-plug or plug-in system or peripheral equipment power cords.
3. Do not operate light switches or other electrical equipment.
4. Evacuate all personnel immediately.
5. Notify the hospital administration and/or fire department.
6. Ventilate the room to clear the air of all flammable vapor or gas.
7. Identify and remove the source of the flammable liquid, vapor, or gas.
Page 1-5
EverView 7500 C-Arm X-Ray Imaging System
WARNING Never bypass, jumper, or otherwise disable any equipment safety feature.
Page 1-6
Introduction and Safety
1.2.11 Labels
Even though this manual is supplied only in English, a foreign language equivalent
NOTE
of the following labels may appear on your system.
1. Quantity: 1
This label indicates the position of the C-Arm
Wig-Wag movement handle. When the handle is
rotated to vertical position, the Wig-Wag
brake is released. When the handle is in
horizontal position, the Wig-Wag brake is
locked.
2. Quantity: 1
This label indicates the position of the C-Arm
Wig-Wag movement handle. When the handle is
rotated to vertical position, the Wig-Wag
brake is released. When the handle is in
horizontal position, the Wig-Wag brake is
locked.
3. Quantity: 2
This label indicates the position of the C-Arm
orbital movement and rotation handle. When the
handle is rotated to the dashed position, the
brake is released. When the handle is in
horizontal position, the brake is locked.
4. Quantity: 2
This symbol indicates the position of the C-Arm
orbital movement and rotation handle. When the
handle is rotated to the dashed position, the
brake is released. When the handle is in
horizontal position, the brake is locked.
5. Quantity: 1
This symbol indicates the position of the C-Arm
horizontal movement handle. When the handle
is rotated to the horizontal position, the brake
is released. When the handle is in vertical
position, the brake is locked.
Page 1-7
EverView 7500 C-Arm X-Ray Imaging System
6. Quantity: 1
This is the mark of image intensifier.
IMAGING INTENSIFIER
P.N. 2414497
7. Quantity: 1
This label indicates that the Mobile C-Arm’s
mechanical subassemblies position should be
placed in their most compact position before
transporting or moving the system. Refer to OM
chapter 4 Mechanical Movement for details.
① ② ③
8. Quantity: 2
10. Quantity: 1
This label on the control panel of the C-Arm
means that you can refer to Section 2.4 in this
11. Quantity: 1
This label near the orbital brake handle means
that there is accessible moving parts and keep
hands/fingers away.
12. Quantity: 1
This label near the safety switch means that you
must take care to rotate the C-Arm larger than
! 155°.
Page 1-8
Introduction and Safety
13. Quantity: 1
This label on the Workstation means that you can
refer to the WorkStation keyboard section in this
14. Quantity: 1
This label on the USB interface means you should
open the USB cover then insert the USB stick.
!
15. Quantity: 1
This label on the footswitch means that the
degree of protection of the footswitch is IP X8.
IPX8 The electrical mechanism within the footswitch is
protected from the effects of continuous
immersion in water.
16. Quantity: 1
This symbol indicates the location of Emergency
Stop Switch on the EverView 7500 equipment.
17. Quantity: 1
This mark means that there is ionization radiation.
18. Quantity: 1
System nameplate.
C-Arm X-ray Imaging System
TYPE: EverView 7500
Indicates manufacture information and input
S.N. :
MANUFACTURED : CLASS 1
power requirements.
INPUT: ~ 110/120 | 220/230/240 V, 50Hz /60Hz
Maximum Line Impedance: 0.3Ω/0.6Ω
POWER:( momentary) 4.5kVA | 5.0kVA
POWER:(continuous) 1.4kVA
GE HUALUN MEDICAL SYSTEMS CO. Ltd.
No. 1 ,Yong Chang North Road, Economic-Technological Development Zone,
BEIJING,P.R. CHINA 100176
This label is prefabricated on the system in Chinese, and the other version of the
NOTE labels are delivered with the system. It can be installed by GE qualified service
engineer according to the requirements of the local users.
Page 1-9
EverView 7500 C-Arm X-Ray Imaging System
19. Quantity: 1
X-ray Generator This symbol Indicates manufacture information
P.N.: 2397125 and input power requirements of X-Ray generator.
S.N.:
DATE:
RATING VOLTAGE: 110kV
TUBE MODEL: 110/3 DF
INH. FILTERATION: 3.5mm AL
FOCAL SPOTS: 1.4mm 0.6mm
GE HUALUN MEDICAL SYSTEMS CO. Ltd.
No 1 Yong Chang North Road Beijing Economic-Technological Development
Area 100176 BEIJING CHINA.
20. Quantity: 1
This symbol indicates the system was tested by
a Notified Body and was found to be in
0459 compliance with the requirements of all
relevant directives and standards in effect
within the European Union at the time of
manufacture.
21. Quantity: 1
This symbol indicates the generator was tested
by a Notified Body and was found to be in
0459 compliance with the requirements of all
relevant directives and standards in effect
within the European Union at the time of
manufacture.
22. Quantity: 1
2339039,CASSETTE HOLDER
This is the mark of the cassette holder.
23. Quantity: 1
WARNING: This X-ray unit may be dangerous to PATIENT and OPERATOR unless
safe exposure factors and operating instructions are observed
Page 1-10
Introduction and Safety
24. Quantity: 2
This label indicates that waste electronic and
electrical equipment must not be disposed of as
unsorted municipal waste and must be collected
separately. Please contact an authorized
representative of the manufacturer for
information concerning the decommissioning of
your equipment.
25. Quantity: 1
This symbol indicates the system was tested by
CCC authorized body and was found to be in
compliance with the requirements of all
S
relevant directives and standards at the time
of manufacture.
26. Quantity: 1
This Label indicates the system was tested by
ITS authorized body and was found to be in
T ERT EK
I N
CM ETL CLASSIFIED
C US
Medical Equipment compliance with the requirements of all
relevant directives and standards at the time
L I
S T E D
9801745
of manufacture.
27. Quantity: 1
This label is prefabricated on the system in Chinese, and the other version of the
NOTE label are delivered with the system. It can be installed by GE qualified service
engineer according to the requirements of the local users.
28. Quantity: 1
This is the mark of hard disk.
HARD DISK
P.N. 2414947
29. Quantity: 1
Page 1-11
EverView 7500 C-Arm X-Ray Imaging System
30. Quantity: 1
This label is near the capacitor to warn that you
!
shall discharge the capacitor before touching it.
31. Quantity: 1
33. Quantity: 1
Collimator Assembly
This is the mark of the collimator.
P.N.: 2415086
34. Quantity: 1
This is the mark of J1 connector.
N L1 N1 L
1 2 3 4 5 6 7 8 9 10 11 12
J1
2
3
6
7
8
1
35. Quantity: 1
9
13
440v
14
10
11 15
Fusel 1
J2 220v 16
12
Fusel 1
17
13 Fusel 2
24v 18
Fusel 2
14
This is the mark of J2 connector.
Page 1-12
Introduction and Safety
36. Quantity: 1
This is the mark of monitor connector.
L N
24V Indicator
220V N
220V L
Page 1-13
EverView 7500 C-Arm X-Ray Imaging System
1.2.12 Symbols
Be familiar with the following symbols that may appear on equipment and schematics, so you
can safely maintain and operate the system:
Dangerous Voltage
This symbol identifies areas where hazardous voltages may be present. Use
appropriate safety precautions.
X-Ray Source
This X-Ray Source symbol can be found on the X-Ray generator. It indicates the
location of the X-Ray source and the controls used to produce ionizing X-Radiation.
Use appropriate precautions and protective equipment when using the system
and at all times when X-Rays are present.
This symbol is used to identify controls that will produce ionizing X-Radiation when
activated. Always use appropriate precautions when generating X-Rays.
This symbol identifies system terminals which, when connected together, are at the
same potential (not necessarily earth ground). Green/yellow wires normally connect
potential equalization terminals together.
Attention
Type B
This symbol indicates Type B applied parts. User is protected against electric shock by
protective earth grounding of all accessible conductive parts or the use of
non-conductive parts.
Page 1-14
Introduction and Safety
This equipment generates, uses, and can radiate radio frequency energy. The
NOTE
equipment may cause or be subject to radio frequency interference with other
medical and non–medical devices and radio communications. To provide
reasonable protection against such interference, the EverView 7500 system
complies with emissions limits for a Group 1, Class A Medical Devices and has
applicable immunity level as stated in EN 60601-1-2:2001.
Page 1-15
EverView 7500 C-Arm X-Ray Imaging System
Page 1-16
Introduction and Safety
The EverView 7500 system is suitable for use in the specified electromagnetic environment.
The purchaser or user of the EverView 7500 system should assure that it is used in an
electromagnetic environment as described below:
Emissions Test Compliance Electromagnetic Environment
The EverView 7500 system uses RF energy only
for its internal function. Therefore, its RF emissions
RF Emissions Group1
are very low and are not likely to cause any
CISPR11 interference in nearby electronic equipment.
RF Emissions Class A The EverView 7500 system is suitable for use in all
CISPR11 establishments other than domestic and those directly
connected to the public low-voltage power supply
Harmonic emissions Not
network that supplies buildings used for domestic
applicable purposes.
IEC 61000-3-2
Voltage fluctuations/ Not
flicker emissions applicable
IEC 61000-3-3
Page 1-17
EverView 7500 C-Arm X-Ray Imaging System
Page 1-18
Introduction and Safety
3,5
Radiated RF 3 V/m [E1=] 3 V/m d =[ ] P
3
IEC 61000-4-3 80 MHz to
3,5
2.5 GHz d =[ ] P 80 MHz to 800 MHz
3
7
d =[ ] P 800 MHz to 2,5 GHz
3
Note: P is the power rating of the transmitter in
watts (W) according to the transmitter manufac-
turer and d is the recommended separation dis-
tance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey,* are less
than the compliance level in each frequency
range.**
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
*Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field
strength exceeds the RF compliance level above, observe the EverView 7500 system to verify normal operation in each
use location. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the [EQUIPMENT and/or SYSTEM].
Page 1-19
EverView 7500 C-Arm X-Ray Imaging System
**Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
The Recommended Separation Distances are listed in the next table.
Note: These are guidelines. Actual conditions may vary.
Frequency of
Transmitter
150KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Equation
3,5 3,5 7
d =[ ] P d =[ ] P d =[ ] P
3 3 3
Rated Power of
Transmitter DISTANCE DISTANCE DISTANCE
(W) (meters) (meters) (meters)
100 12 12 23
For transmitters rated at a power not listed above, the DISTANCE can be estimated using the equation
in the corresponding column, where P is the power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Page 1-20
Chapter 2
Systems OverView
EverView 7500 C-Arm X-Ray Imaging System
The EverView 7500 C-Arm X-Ray Imaging System is intended to provide fluoroscopic and film
imaging of the patient in diagnostic and surgical applications. Clinical applications may
include, but are not limited to, surgery, critical care and emergency room procedures. This
product will be used on a daily basis by such users. The product may be used for other
imaging applications at the physician’s discretion.
EverView 7500 C-Arm X-Ray imaging system includes:
• Mobil C-Arm
Is composed of generator, monoblock, collimator, image intersifier, CCD camera head
and control panel.
• Mobil WorkStation
Is composed of 2 monitors, 1 computer hard disk, keyboard, etc.
• Options
Sony Thermal Printer
DAP
Cassette holder, 10×12” (24cm×30 cm)
Spacer (30 cm)
Page 2-2
System Overview
The C-Arms derive their name from the rigid, C-shaped weldment that supports the Image
Intensifier, collimator, CCD Camera, and X-ray tube. These components together are commonly
called the “C”, and the entire assembly from the casters up is known as the “C-Arm”.
The X-Ray tube generates an X-ray beam that is shaped and oriented by the collimator. After
passing through patient anatomy, the beam is directed onto a film cassette to produce a
radiographic film image, or through the image intensifier and CCD camera, to convert the X-ray
image to a video signal that is transmitted to the WorkStation for display, processing, and
storage.
Page 2-3
EverView 7500 C-Arm X-Ray Imaging System
Orbital Movement
Page 2-4
System Overview
Collimator
Filament PCB KV/MA PCB Exposure
Current Control KV & MA Control
Brightness Self-regulation
Management
of Filament
Tabletop MP4 MP3 Controller
MP2
Grid Driver PCB
MP1
II/CCD IBS
Operator Desk
CCU Display PCB Exp Foot Switch
Optics
50Hz CCIR X-Ray ON Warning Light
100Hz
Rotation
Sensor
Imaging Computer
75Hz VGA
CCD Recordt
17"
Monitor 2
752X582 17" Display
CCD Database
Monitor 1
Camera
Keyboard
Video
PAL
Workstation Out
Page 2-5
2.4 EverView 7500 C-Arm X-Ray Imaging System Schematics
1 2 3 4 5 6
Revision Content of design change Designer Date
PJ7-1 PJ7-2
Note! 1
5 6 The low voltage wires (24V) should be enwrapped with additional
PJ7-4 PJ7-5
3
1
4
2
insulate shell if bundled together with high voltage wires. 2
PJ8-1
3
PJ8-2
PJ3-1
J1
1 Brn
CL02A310TJ 24VAC
1 BT1
9 Brn 13 MPJ1-3 Brn 2390714 A
A PJ3-3
2 Blk
2 440V 14 MPJ1-1 Blu 2339458
3 Wht
3 10 Gry
4 Red J3B J5 J2 J3
2 4 15 97 4 J5
J3A
4 11 Org 15
Fuse1 J0
PJ3-4 Yel
2339471
5
5 1 5 2414335
12220V Blk
6 Gry
13 Yel 1 6
Fuse1
2 4 15 97 3 J2
J4
CCU
PJ3-6
2390715
7 Blu 14 24V Blu 1 7 1 CJ4
CJ3
6 Display Board:
15 Fuse2 MX8-7
J1
8 7 CJ8 CJ7
3 2414090
CJ6
Brn 16 22V MX8-6 2414334
9 Grn
PJ1-1
8
17 18
Fuse2 CCD
24 15 9 71 10 Red2 4 15 95 7 Connect Board :
PJ1-3
11
PJ1-5 Blu2 4 15 95 0
2414092
Blu 12
PJ1-9 2414369
2 41 59 6 4
For detail connection 24 15 9 55
I.I
J2-12
please refer to page 3
J2-13
AL3
AD1
AL4
PJ4-1 1 BT2
24 1 59 69
5 A:
MX3-3/4
MX2 1234567
MX13
PJ4-2 MX1
MX3-5
2
PJ4-3 3 MX14
MX8 AC AR
6
PJ4-4 4 AL3 AL4 AL5 AL6
7 2 41 5 98 2 MP1 MP2 MP3 MP4 MP5 MP6 MP7
24 15 97 5
PJ4-7
PJ4-8
1
2 41 59 7 7 MPJ1-7
MP2-J1
MX3
Blk
MX15-1
MX15-3
MPJ1-5
B Red
AD2
B
Red FJ1-1 Red AD1
MPJ1-8 2 1
Blu FJ1-2 Blu Blu
CJ8-4 2 41 6 03 9 I:
IJ
2 415 96 8
AL2-2
Blk AL2
CJ8-1 2 41 59 7 6 AL1-2
MPJ1
Emergency AL2-1
220VL
220VN
J2-13
220VL
220VN
COM
+24V
GND
COM
COM
-15V
+15V
Switch
GND
+5V
S:
V1
V1
J2-14
SJ
MPJ1-6 1
MX4
Brn
Mother Board ASM: 2
2 41 59 6 3 AL1
2 4 15 94 8 2 41 59 7 8 2415166 M X3 - 1/ 2
6 1
1 J2-1 H.S
AL2-2
AL2-1
GND
AL1-2
2 J2-2 GND Brn Blu
Blu
Brn
PJ7
3 J2-3 Blk
PJ6
4 J2-4 2339469
5 J2-5
24 159 58
2 4 16 05 2 H:
Blu
Red
Brn
MX5-5Red
MX5-4Grn
PJ5-1 2339470
Wht
MX5-6
2414086
PJ5
PJ5-4
2414482 HJ2
2 41 5 98 0
CJ7-1
L
P1
24 15 97 9
CJ7-3
PJ1
MD
PJ1-6 OX1 2 41 59 5 3
C I1 1 2 3 4 5
N C
PJ1-8 A
2372499
B
2372500 1 4
F:
C Motor
J3 100Hz
Video
FJ3
+ Driving
Blu
Brn
J1-1
GND O: OX2 1 MX5-1 J1-2
Male 2 MX5-2 J1-3
J2 50Hz 5 2PIN
Video 3 MX5-3 J5-F 3 1
Female
Filter GND 4 Blue&Green
Male
FJ1 FJ2
24 15 94 7
16AWG
2 41 59 7 0 1 J5-M 2 1
2 41 5 94 9 2 41 59 5 1
Note!
OX3
Do the right wire 序号 零部件号 零部件名称 备 注 数 量
24 15 98 1
24 1 59 67
24 15 96 5
connection between
the coxial cabel and J4-M Male
3PIN
Brn Blu BNC connector J4-F Femal FT 比例 材料 表面处理 补充编号
2 4 15 9 7 2 1 Proximit
LO +u
L-
2 4 15 96 6 y Sensor
a
a
a
K1 K2
Wiring Diagram D
A
A
D
A
A
24 1 59 45
F.S
FT
FT-EN 文件类别 图号 版本号 页号
Brn For detail connection
L N Blu please refer to page 3 2 1
3 4
5
北京通用电气华伦医疗设备有限公司 SCH S10002PE 1
0
2339441(2415147) Double grounding connection
IX2 with tie-in protected by
GE HUALUN MEDICAL SYSTEMS
Female heat-shrinking pipe
3
1 2 3 4 5 6
Page 2-6
System Overview
VGA
Key board
video in
power
241 47 25
printer
YJ-2 YJ-1 24 14 72 2
VGA
24 1472 4
power
24 1472 5
24 1472 5
female male
key board&mouse
4 3 2 1 4 3 2 1 video in
2
AWG18 Note: a grounding label should 24 14 72 6 3
Length =10cm be attached around here
4
GND
5
7
Shafler
24 1472 3
24 1472 4
Filter
J3 J2
OMRON
G2R-2-S-12VDC
1
5 4
6 3
7 2
8
220VN
220VL
GND
1 2 3 4 5
A
IX1 B
C
video
GND
GND
video
Page 2-7
EverView 7500 C-Arm X-Ray Imaging System
Page 2-8
Chapter 3
Installation
EverView 7500 C-Arm X-Ray Imaging system
3.1 Pre-Installation
3.1.1 Dimensions and Weights of System
Dimensions:
C-Arm: 1730mm x 760mm x 1700mm (Length x Width x Height )
WorkStation: 840mm x 650mm x 1630mm (Width x Depth x Height )
Weight:s:
C-Arm: 480 Kg (with packing) 250Kg (net weight)
WorkStation: 225 Kg (with packing) 150Kg (net weight)
The floor should be hard enough to support the system. It should also be
WARNING
smooth and horizontal .
Page 3-2
Installation
3.2 Unpacking
There are 2 packages for EverView 7500 C-Arm X-Ray Imaging system. One is for
NOTE C-Arm (L X W X H=1900mmX900mmX1970mm), the other is WorkStation (L X W X
H=800mmX960mmX1820mm). It is easy for one FE to do unpackage in 2 hours.
The WorkStation and Mobile C-Arm ramp is located in rear panel of the Mobile
NOTE
C-Arm’s enclosure. You should open the Mobile C-Arm’s enclosure to get the ramp
first.
Before unpacking, inspect the shipping enclosure /boxes for damage and report to
NOTE the shipper if available.
The unpacking of the components of the system before it is moved into the operation room will
be made for checking if there is damage or non-apparent damage occurring during
transportation.
The packages will be minutely examined during installation, in order not to overlook parts of
small dimensions or envelopes containing screws, washers, etc.
If damage is observed, immediately notify the transport service and the transport insurance
service of the factory.
The same will apply for non–apparent damage due to transport discovered when unpacking the
material or during installation, UP TO 15 DAYS MAXIMUM AFTER DELIVERY.
A transport service will generally not pay a bill for non-apparent damage unless an inspection
has been demanded within 15 days after delivery.
If damage is discovered, immediately contact the services concerned, specifying the type of
apparatus, the serial number and the order number if possible, and describing the nature of the
damage.
Transport the equipment to the operating site, check the parts according to packing list (refer to
the packing list with the system for details).
Page 3-3
EverView 7500 C-Arm X-Ray Imaging system
Please follow the steps below to unpack the EverView 7500 C-Arm X-Ray Imaging System:
1. First , place the C-Arm package as pictures below and leave enough space for backward
moving of the C-Arm.
Front (Image
Intensifier) side of
the C-Arm, with a
block outside
Right panel
Rear panel
(ramp attached)
2. Unscrew the screws of the fastening angle iron. Remove the angle iron. Remove the top
cover first.
3. Remove the wood block on right side ( from rear side to front side view) of the package of
the C-Arm. Remove the right side panel. Remove the 2 crossbeams on the left side panel
as crowbars. Then remove the left panel and rear panel, front panel of the package.
Remove the ramp from rear panel and put it on the rear side of the C-Arm. Remove the
plastic on the C-Arm.
Page 3-4
Installation
crossbeams
Wood block
4. Unscrew the 2 bolts with a 8 mm sleeve wrench and remove the support woods.
2 bolts
5. Loosen the fastening nut of the support assembly, then unscrew the screw bolt.
Page 3-5
EverView 7500 C-Arm X-Ray Imaging system
6. Unscrew the 4 screws of the fastening assembly at the front of the C-Arm. Remove the
fastening assemblies with a crowbar.
7. Unscrew the 8 screws of the 2 fastening assemblies on each side of the C-Arm and
remove the fastening assemblies with a crowbar ( only right side for illustration).
8. Check and Verify that all the mechanical movements brakes are locked. Release the
wheel brake. Hold the right steering handle in a vertical position to keep the rear wheels
in a straight direction. Then carefully move the C-Arm backward along the ramp to the
site you desire.
WARNING The C-Arm weights about 250 Kg. Two persons should move it slowly to
prevent lose of control.
Page 3-6
Installation
9. Remove the top cover and side covers of the WorkStation Package. Remove the front and
back fastening wood blocks.
10. Lift the WorkStation with a crowbar, and draw out the stow-woods under the
WorkStation. Put the wood ramp in position, and push the WorkStation along the ramp.
11. Move the WorkStation to the site near the C-Arm and lock the wheels.
12. Insert the male socket of the WorkStation’s integrated cable into the female socket
located on the right side of the C-Arm’s cover. Push the socket in and lock it in position.
13. Plug the C-Arm power plug into a properly rated AC receptacle. Power on the system.
Before power on the system, please check and make sure that the power
WARNING
frequency and voltage of the system are compatible with the power supply. Or
you must change the power supply and frequency settings of the system.
Refer to Section 3.3 for details.
14. Press the Raise C-Arm Command Key to lift the C-Arm for about 20 cm height.
Page 3-7
EverView 7500 C-Arm X-Ray Imaging system
15. Use a pliers or 17mm wrench to loosen the fasten nut then unscrew the screw rod.
Remove the backward-forward fastening assembly.
16. Loosen the horizontal movement brake handle and push the C-Arm forward. Lock the
brake handle. Lower the C-Arm to it’s lowest position and then power off the system.
17. Move it to the site you desire.
Page 3-8
Installation
If your power is 60 Hz, please change the jumper on MP5 from 50 Hz to 60 Hz. The location of
the jumper is as the picture below.
Page 3-9
EverView 7500 C-Arm X-Ray Imaging system
1. Open the cable kits: 110 V power cable kit (2415136).Remove the back cover of the
C-Arm control panel. Loosen the fasten screws of the incoming circuit breaker and
remove the 220 V power supply cable.
2. There are 2 types of plugs provided by GE Healthcare: European type and American
type. Select or buy the relative plugs according to your local regulatory requirements
and sockets style. They must be reliably grounded.
3. If the system is used in Canada and Australian and USA, please select and buy the
relative plugs according to the list below. The selected plugs shall be listed or recognized
type plugs. The manufacture doesn’t provide these types of plugs.
Countries Rated Current Comments Voltage Frequency
Canada 20 A Hospital Grade 120 V 60 Hz
Plug
Australian 16 A transparent 230 V 50 Hz
USA 20 A "Hospital grade" 120 V 60 Hz
or "Hospital
only" type plug
4. Connect the plug to the power supply cable and connect the power supply cable to the
incoming terminals of the K2 circuit breaker. Then change the bridge connectors of J1
terminal blocks. ( The follows are the schematics for 110 V and 220 V system FYR).
You should change the 2 spacers’ position too.
NOTE
Page 3-10
Installation
Brn 9
PJ1-1
a
a
a
10
Breaker Breaker
Unlock
Unlock
11
PJ1-3
K1 K2
Blu PJ1-5
A
A
A
A
12
PJ1-9
Brn
L N Blu
Brn
a
a
Unlock
Unlock
10
K1 K2
11
PJ1-3 A
A
A
A
PJ1-5
Blu 12
PJ1-9
Brn
L N Blu
5. Plug the power supply plug into the socket and check if the system can power on
normally.
Page 3-11
EverView 7500 C-Arm X-Ray Imaging system
N L1 N1 L
J1
11
10
12
9
3
8
7
2
5
4
1
6
1
8
2
7
6
5
Page 3-12
Installation
Page 3-13
EverView 7500 C-Arm X-Ray Imaging system
The power and video signal of the View Printer- UP-895MD is provided by
WARNING EverView system. Please do not connect the power supply and video to other
printer. Please do not connect other external power supply and video signal to
this printer. Otherwise, it may cause electrical shock to user.
CAUTION The Video Printer is not fixed in the cabinet. Please pay attention when moving
the WorkStation.
Connect one end of the video cable to the video input port of the printer. Connect the other end
to the Video output socket on the WorkStation.
Video Output
Connect the socket plug of the power cable to the video printer, cut off the other end, strip off 3
wire ends, press the cable lugs and then connect the 3 wires to the terminal blocks at the bottom
of the workstation (brown line to the brown power supply line, and blue to the blue, green-yellow
to the green-yellow ).
Put the power supply cable along the wiring harness to get a good layout.
After installation finished, power on and test the printer to see whether it can work normally.
Page 3-14
Installation
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING serious injury or death from electrical shock, and/or damage to the equipment.
Use appropriate precautions.
To prevent ESD damage, ensure that you are properly grounded with a grounded
CAUTION
wrist strap before handling any electronic components sensitive to ESD.
NOTE
The system being upgraded may differ slightly from what is illustrated.
1. Power off the EverView 7500 system and unplug the power supply plug of the system from
the socket.
2. Disonnect the intergrated cable.
3. Unscrew the 4 screws of tube cover, remove the tube cover.
Side View
4. Insert the RJ45 plug of the power supply wire in the socket of the DAP kit.
Socket
RJ45 plug
Page 3-15
EverView 7500 C-Arm X-Ray Imaging system
6. Check the C-Arm and confirm the updation. Then power on the system, check if the display
and operation of the DAP work properly.
Page 3-16
Installation
Page 3-17
EverView 7500 C-Arm X-Ray Imaging system
Control
Panel
Steering
Handles
Base &
Frame
Front and
back
wheels
Monoblock
Comments:
Indicate the location of damage of defects, if any, on the figure and add comments to describe
the damage if necessary.(Reference Mobile C-Arm Damage Inspection on previous page)
Page 3-18
Installation
Page 3-19
EverView 7500 C-Arm X-Ray Imaging system
Comments:
Indicate the location of damage or defects, if any, on the figure and add comments to describe
the damage if necessary. (Refer to WorkStation Damage Insepection on previous page.)
Page 3-20
Installation
Page 3-21
EverView 7500 C-Arm X-Ray Imaging system
Page 3-22
Installation
3. Remove the 8 screws of the WorkStation cover , remove the cover. Inspect the internal
cables. Check if there are loosen parts.
Page 3-23
EverView 7500 C-Arm X-Ray Imaging system
2. Remove the C-Arm’s covers and inspect the internal cables and PCB seating. Look for loose
parts. See below figure.
Page 3-24
Installation
• Brakes
1. Check the Wig-Wag brake operation. Check Wig-Wag for ease of movement (side to
side swing ) without excessive play when the brake handle is turned to release
position. When the brake handle is in lock position, check that it is hard to enable
Wig-Wag movement.
2. Check the Horizontal Movement brake operation. Extend and retract the Horizontal
Cross-Arm. Check for ease of movement without excessive play. When the brake
handle is in lock position, check that it is hard to enable Horizontal movement.
3. Check the rotation brake operation. Check the rotation for ease of movement without
excessive play. When the brake handle is in lock position, check that it is hard to
enable rotation movement.
4. Check the C-Arm orbital rotation brake operation. Check orbital rotation of the C-Arm
for ease of movement without excessive play. When the brake handle is in lock
position, check that it is hard to enable orbital rotation.
5. Check the operation of the rear wheel pedal brakes.
6. Check for proper operation of the steering handle and control of the rear wheels.
Check for ease of movement without excessive play.
7. Check the operation of the brakes of workstation.
• Steering
With the Mobile C-Arm’s rear wheels brake released, turn the right steering handle
90° to the left and right. Verify that the rear wheels turn 90° to the left and right.
Place the Mobile C-Arm’s rear wheel brake in the neutral (flat) position. Push the
C-Arm slowly and verify that when the steering handle is turned the Mobile C-Arm
turns.
• Wheels
Push the C-Arm and verify that the wheels roll smoothly, the wheels turn freely.
Push the WorkStation and verify that the wheels roll smoothly without vibration
and that they are securely attached to the wheels.
Page 3-25
EverView 7500 C-Arm X-Ray Imaging system
5. Press (Raise C-arm Command) key on the panel to raise the C-am, the
lower the C-arm. When the movement stops, press two keys at the same
time and the C-arm will be lowered to it’s lowest position.
Page 3-26
Installation
WARNING
The system will produce X-Rays during the following procedures. Take
appropriate measures and precautions.
1. Manual Fluoroscopy
Place spacial resolution card in front of the image intensifier, the angle between the horizontal
axis of the image and the test card must be 45°。Select Manual Fluoroscopy mode
and appropriate kV value,take an exposure ,and stop when image is stable. The image is frozen
on the main monitor, verify the resolution in lp/mm is equal to, or greater than, the values listed
below. Repeat and select any of three field sizes (NORM, MAG1 and MAG2) to verify the resolution.
Field Size NORM MAG1 MAG2
Resolution (lp/mm) 1.4 1.8 2.2
2. Auto Fluoroscopy
Test Equipment : 1 mm Copper filter ( 3 for each system) No: P/N 00-887682-01.
(This equipment is not provided with the system by the manufacturer).
Page 3-27
EverView 7500 C-Arm X-Ray Imaging system
Select Manual HLF mode + Pulse mode via the C-arm control panel, the
X-Ray Exposure indicator flashes when taking an exposure.
4. Manual HLF
Select Manual HLF mode + , verify that a short-term HLF exposure is taken
and automatically stopped; Audible tone beeps by the buzzer shows it’s in correct function.
5. Auto HLF
Place an object in front of the intensifier, select Auto HLF mode + to take
exposure. It will stop automatically.
6. I.I inspection
Place something on the I.I., take exposure under 3 fileds (9”, 6”,4.5”). Verify that the image
changes with the field.
7. Recursive Factor
rotation key (left then right), the image on the monitor will rotate
relatively.
Enable the Motion Detection key, select Fluoroscopy mode , take an continuous exposure and
press the CCD rotation key , the image lag on the monitor will be smaller
than before.
Page 3-28
Installation
9. Semi-transparent Shutters
Select manual fluoroscopy mode to take exposure intermittently. Open, close and rotate the
shutters to verify the functions. When the shutters are opened to the largest, the shutter will not
display on the monitor.
10. IRIS
Select manual fluoroscopy mode to take exposure intermittently, open and close the IRIS to
verify the functions. When the IRIS are opened to the largest, the IRIS will not display on the
monitor.
Page 3-29
EverView 7500 C-Arm X-Ray Imaging system
11. Radiography
Fasten the cassette holder on the intensifier; insert a cassette (10”×12”,with a film). Select
Radiography mode and take a short exposure under 50KV, 4mAs, verify that beeps are heard
and no errors are displayed, the developed film is clear.
Page 3-30
Chapter 4
Calibration
EverView 7500 C-Arm X-Ray Imaging system
4.1 Overview
The Calibration described herein should be performed each time you replace a major system
component. When possible, run these tests in sequence that appear in this section.
Tools
In addition to common tools, the following tools are recommended:
• Extended board 2391302
• Mp6 Dummy board 2415249
• Digital Volt-Ohm-Meter--------------------- Type C31/1 V
• kVp meter--------------------------------------Type PMX-1/R
• mAs meter-------------------------------------Type MA9021
• Oscilloscope-----------------------------------Type FLUKE 196
• DC current meter------------------------------C31/1-mA
The system will produce X-Rays when implementing the following procedures.
WARNING Take appropriate measures and precautions to protect yourself and others from
X-Ray exposure.
You must obey the following steps to guarantee the safe operation of the C-Arm.
CAUTION Improper adjustment may cause premature failure of major system components.
Page 4-2
Calibration
MP3
Board
W3 jumper
Remove W Jumper,
Connect both probes
to W test point.
Page 4-3
EverView 7500 C-Arm X-Ray Imaging system
4.2.1 Measurement
1. Normal Fluoroscopy current measurement
Insert the MP3 PCB on one extended PCB and connect a DC current meter into the line of
MAMEAS2. (Remove the W jumper and connect the DC current meter probes to the two
pins).
2. High Level Fluoroscopy current measurement
Connect jumper HIDO on MP2 to disable automatic stop exposure function, please note
that exposure time should be as short as possible.
MP2 board
HIDO Jumper
Page 4-4
Calibration
4.2.2 Adjustment
Fluoroscopy current adjustment
1. Pre-adjust
a) Select manual fluoroscopy mode.
b) Power on the system.
c) Set the KV value to 110KV and measure the test point TP5 on MP3. The measured
value should be 9.3V (adjust POT3 on MP3 board to reach this value).
d) Set the KV value to 40KV and measure the test point TP5 on MP3. The measured
value should be 0.3V (adjust POT3 on MP3 board to reach this value).
e) Pre-heating: Adjust POT7 on MP3 to make the measurement value of TP2 is about
3.7V.
f) Power off the system.
2. Adjust
a. Put MP3 board on the extended board.
b. Set DC current meter to mA range.
c. Connect both probes to ‘W’ test point on extended board.
d. Remove ‘W’ jumper.
e. Power on the system.
f. Set mA as the values below:
mA settings 0.1 0.5 1.0 2.0 4.0
Measurement
3. Take exposures. Check the current (0.5~4.0 mA) measured by the meter according to
the above values. The error should be within the ±10% ranges of the settings.
Otherwise adjust POT4 on MP4 board. Rotating it clockwise can increase the current ,
rotating it counter-clockwise can decrease the current.
4. To adjust 0.1mA accuracy, adjust POT8 on MP3.
Raise kV step by step, otherwise instant high current may cause damage to
WARNING the equipment. 1/8 circle of knob in the meter will be a proper stroke to
prevent an instant high current.
Do not adjust the POT4 on MP4 board when unnecessary. Once do it, you
WARNING
must calibrate the HLF current and Radiography current (mAs).
5. Power off, disconnect extended board, put the jumper W back to extended board.
Page 4-5
EverView 7500 C-Arm X-Ray Imaging system
2. Take exposure. Check the measured values and verify that the errors are within the ±10%
range of the values above. Otherwise, adjust POT 6 on MP3.
3. Power off the system.
4. Disconnect extended board, remove probes from W test point on extender board.
5. Disconnect jumper HIDO on MP2 board.
Page 4-6
Calibration
Page 4-7
EverView 7500 C-Arm X-Ray Imaging system
k) Set 40kV, take exposure, the Oscilloscope wave peak should be 5V±0.2, if not, adjust
POT3 on MP4 and take exposure again. Don’t care the current meter and mAs meter.
l) Set 40kV 40mAs, take exposure and ensure the current meter reading is 20mA ±10%. If
not, adjust POT3 a little.
m) Set 110kV 40mAs, take exposure, make sure the current meter reading is 20mA ±10%.
If not , adjust POT2 a little.
2. Timer adjust.
a) Set the following kV and mAs values:
1mAs(110kV) 80mAs(40kV) 4mAs(80KV) 20mAs (100KV)
b) Take exposure, check and verify the errors of the mAs value is within ±10% range.
c) If not, adjust POT9 on MP3.
d) If the mAs still beyond the 10% range, for 80kV and 110kV, re-adjust POT2 on MP4 a
little, for 40kV re-adjust POT3 on MP4 a little.
3. Power off, take off extended board, connect jumper W.
Page 4-8
Calibration
The mAs meter must be connected tightly, otherwise, it may cause injury or
WARNING damage. This test is only performed in the factory when it is required by
authorized organizations according to the regulations.
4.4.1 Measurement
The X-ray generator is composed of high voltage transformer, high voltage rectifier and X-ray
tube in one pressurized oil tank. The X-ray tube voltage can be measured with indirect method
by the kVp meter.
Load factor Fluoroscopy: 110 kV/4.0 mA, 99 kV/3.4 mA; 66 kV/1.6mA
Radiography: 110 kV/2 mAs,99 kV/2 mAs, 66 kV/2 mAs
4.4.2 Adjustment
1. Pre-adjustment
a) Set Digital Volt-Ohm-Meter meter to DC.
b) Connect 1 probe to TP1 on MP3 PCB, the other to the ground.
c) Power on the system.
d) Set the kV value to 40KV, the reading on the meter should be 4V±0.2V, otherwise
adjust POT2 on MP3 board.
e) Set the kV value to 110KV, the reading on the meter should be 11V±0.2V, otherwise
adjust POT1 on MP3 board.
2. Put kVP meter 50cm or closer (depend on the kV settings) from FS( Focal Spot) of the
tube.
3. Fluoroscopy kV adjustment
Set the values as below:
KV Display 66kV 99kV 110kV
Measurement
The errors of the KVP should be within ±7% of the values above, Otherwise adjust POT116 on
MP5. (Counter-clockwise to increase kV, and clockwise to decrease kV).
Also monitor DC Voltage on MP6 (+HRPV and –HRPV), the trend is as same as kV trend,
it should be no more than 370V.
Page 4-9
EverView 7500 C-Arm X-Ray Imaging system
4. Radiography KV adjustment
Select the Radiography mode and set 2mAs, press the HandSwitch to verify the test
value according to the table below.
KV Display 66kV 99kV 110kV
Measurement
The errors of the KVP should be ±7% of the KV values above, Otherwise adjust POT2 on
MP5. (Counter-clockwise to increase kV, and clockwise to decrease kV)
Also monitor DC Voltage on MP6 (+HRPV and –HRPV), the trend is the same as kV trend,
it should be no more than 395V.
Before perform the Radiography KV adjustment procedure, you must adjust
NOTE
the Fluoroscopy KV. Step 3 and Step 4 must not be reversed.
Page 4-10
Calibration
Brake Cover
2. Loosen 2 fastening screws, adjust the brake by 4 adjustment screws to adjust the
brake force. After the brake force is moderate (no less than 200 N) , fasten the 2
fastening screws.
Fastening Adjustment
Screw Screw
Page 4-11
EverView 7500 C-Arm X-Ray Imaging system
3. The 2 limited screws are used to adjust the range of the brake movement. You can
adjust it when necessary.
Limited
Screws
Page 4-12
Calibration
Fixed
Screw
Page 4-13
EverView 7500 C-Arm X-Ray Imaging system
3. Then you can find a hole used to lock the C-Arm rotation, there is a pin extended into
the hole. To repair this equipment, you must use a stick to push the pin back and
install the push screw. See the follow Fig.
Page 4-14
Calibration
2. Take off these two limited screw and then you can loose all fixed screws, turn the nut to
adjust the brake force.
Page 4-15
EverView 7500 C-Arm X-Ray Imaging system
4.8 Others
For Camera iris adjustment (dose adjustment), Blanking adjustment, -ABS window adjustment,
Refer to Appendix chapter 11 for details.
Page 4-16
Chapter 5
Software
EverView 7500 C-Arm X-Ray Imaging system
5.1.1 Overview
Page 5-2
Software Maintenance
Click the combo box (PIC. 2) and select the proper user to logon. Different users are assigned
with different operating rights:
• ENG for development engineer
• MFG for manufactory engineer
• SVC for service engineer
Page 5-3
EverView 7500 C-Arm X-Ray Imaging system
For software version higher than 1.02, if the password is correct, 7500 software will be closed
automatically and main window of maintenance feature will pop up. After that the service
engineer can access Operating System.
NOTE
The logon password for user SVC is “zhq”, PLEASE KEEP it CONFIDENTIAL.
Page 5-4
Software Maintenance
Page 5-5
EverView 7500 C-Arm X-Ray Imaging system
After press the Repair User Data button, a message box as shown in PIC. 5 will pop up if the
database has crashed. Press Yes button to repair it.
Page 5-6
Software Maintenance
This operation will delete all users data, so it is only for test purpose in manufacture.
NOTE Press OK button in the pop-up message box if you are sure to do so.
Page 5-7
EverView 7500 C-Arm X-Ray Imaging system
USER: ENG
BUTTON: Remove Software
DESCRIPTION: You can remove 7500 software, including
Click Remove Software button to prompt Remove Software dialog. See PIC. 7.
Page 5-8
Software Maintenance
Patient Information
Page 5-9
EverView 7500 C-Arm X-Ray Imaging system
5.2.1 Overview
There are three CD disks for every 7500 system, they are named as Recover CD, Application CD
and Operating System CD.
Page 5-10
Software Maintenance
6. The system will ask you if you want to keep user data. The default value is Yes. If you do not
want to save the data, you can select No. Click on Next button, the installation will start.
If you select No, the images and patient information saved in the system will be
NOTE
lost.
Page 5-11
EverView 7500 C-Arm X-Ray Imaging system
7. After the installation is finished, the system will give a Success! prompt. You can click on
Finish button. The software is installed successfully.
8. Wait for a few minutes, the DaoJet database installation Wizard will start automatically.
Page 5-12
Software Maintenance
9. Select the default settings and click Next button to go all the way down the install process
until the screen below pop up. Click on OK button to continue.
10. After the installation is finished, the following screen will pop up.
This application is compatible to previous versions, but you still need a special
NOTE service operation manual to update it in order to avoid damage of the system
and the database.
Page 5-13
EverView 7500 C-Arm X-Ray Imaging system
5.2.3 Recover CD
DESCRIPTION: The recover CD can only be used to repair the operating system when
necessary.
Steps:
2. Power on the system. Press "Delete" key quickly and hold it on to launch the BIOS SETUP
UTILITY.
3. Press the "Right" key to open the "Boot" page. Press "Down" key to select "Boot Device
Priority" and press "Enter" key.
4. Use "Enter" and "Down" key to set the "1st Boot Device" as the CDRW drive, and "2nd Boot
Device" as the hard disk.
5. Press "F10" and "Enter" key to save the change and restart the system.
6. When it is prompted "Press Any Key To Recover Operating System...", press "Space" key to
recover operating system, which may take about 15 minutes.
7. When recovery is done, the system will be rebooted automatically. Please do not press any
key until the Window XP is boot up.
8. When Window XP is up, insert the software installation CD into the CD drive and install
EverView 7500 software according to section 5.2.2.
9. After the software is installed, repeat step 2 and step 3 to set the 1st boot device as the hard
disk and the 2nd boot device as the CDRW drive.
10. Press "F10" and "Enter" key to save the change and exit the BIOS SETUP UTILITY.
Page 5-14
Chapter 6
Trouble Shooting
EverView 7500 C-Arm X-Ray Imaging system
b) Power on the equipment, and check the connection of the power indicator.
4. Fault indicator
A). If Fault Indicator is not lit, go to the next step;
B). If Fault Indicator is lit:
a) Press the Emergency Stop Switch, and open the main control cabinet;
There is the risk of electric shock when implementing the following steps.
CAUTION
Take appropriate precautions.
Page 6-2
Trouble shooting
In this board, the signals of alarm are as follows( They may cause the light of fault indicator):
• ALTH-HC tube overheat alarm > 55 oC H2 on the Motherboard will light
• FILAL-HC filament over current alarm H3 on the Motherboard will light
• Over HV inverter sample voltage PWRSMDV is over pre-setting value
• Low HV inverter sample voltage HVTDV is less than preset H7 on the
Motherboard will light
d) The operator needs power off the system and wait for 1minute, then power on
the system to reset the Alarm.
6.workstation check
A). If workstation works normally, go to the next step;
B). Check and verify the monitor is powered on.
CAUTION The following step has the risk of electric shock. Take appropriate precautions.
Page 6-3
EverView 7500 C-Arm X-Ray Imaging system
Begin
Yes
Yes
No
No
Other indicators
correct? Check MP2, MP3 and display PCBs
Yes
No
Display of the Check monitor power cable, filter,
monitors correct? computer power supply
Yes
End
Page 6-4
Trouble shooting
Page 6-5
EverView 7500 C-Arm X-Ray Imaging system
POT 2 POT 1
TP1
Page 6-6
Trouble shooting
Before follow the following step, the maximum DC voltage might exceed 380V,
CAUTION
take appropriate precautions.
Page 6-7
EverView 7500 C-Arm X-Ray Imaging system
System Check
Page 6-8
Trouble shooting
N Check
kV/mA is
Display PCB
OK?
and MP2、MP3
Y
Measure the MP3 KVC
signal
1V=10kV
N Check
KVC: 4--11V?
MP3
Y
Select Radiography
Check
N
Display
kV is ok?
MP2 and
MP3
Y
Page 6-9
EverView 7500 C-Arm X-Ray Imaging system
N Check display
mAs Range is
board
ok??
Check MP2
Y
Power down
Page 6-10
Trouble shooting
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING
serious injury or death from electrical shock. Take appropriate precautions.
Page 6-11
EverView 7500 C-Arm X-Ray Imaging system
b). The value of wave is about 0.6V, or readjustment shall be made according
to the current calibration procedure in Chapter 5. The wave should be as
the following.
Page 6-12
Trouble shooting
Page 6-13
EverView 7500 C-Arm X-Ray Imaging system
Page 6-14
Trouble shooting
Fluoroscopy Check
Begin Fluoroscopy
check
Attach working
card
Power on
use Manul mode
The monitor
brightness is N Check the IBS signal
suitably and the in MP3
kV is stably?
Y
5
End
Page 6-15
EverView 7500 C-Arm X-Ray Imaging system
N Check
It is correct?
MP5,MP7
Y
Use a load instead of the
tube HV primary,Check
the wave of the HV
N Check MP5,
The high voltage
MP6, MP7 and
wave is correct?
Motherboard
Y
Check the
tube
assembly
Page 6-16
Trouble shooting
Check the
image chain
Check
the II
IBS:0--12V,is
fluctuate when the Check the
brightness is cable of II
changing?
Check
MP3
Page 6-17
EverView 7500 C-Arm X-Ray Imaging system
Electrical circuits inside the equipment use voltages that are capable of causing
WARNING serious injury or death from electrical shock. Take appropriate precautions.
1. In normal condition
• Insert the MP3 board to mother board via extended board;
• Use an oscilloscope to measure the voltage of W test point on extended board.
• Power on and select radiography mode.
• Adjust KV value to 70 and mAs to 20.
• Press handswitch to have an exposure, the exposure time is about 1 second and
the test result is approximate –5V.
• Power off the system.
2. In abnormal condition
A). An alarm is given from the fault indicator during an exposure.
a) Shut off the system and check the tube’s temperature. If the temperature is too
high (the temperature of tube cannot exceed 55℃, or an alarm will be given),
re-start the machine 20 minutes later and make the exposure inspection again.
b) MP2 board is connected to motherboard via connection board.
c). Power on and take an exposure, if an alarm is given again, check the signal
“FILAL-HC, AL, ALTH-HC, HVTDV” on extended board.
d). FILAL-HC is filament over-current alarm signal, AL is Hi-voltage primary
over-voltage alarm signal, ALTH-HC is overheating alarm signal of tube, and
HVTDV is Hi-voltage primary low-voltage alarm & sampling signal.
B). The voltage tested on the oscilloscope is far away from 5V;
Readjust it in line with description in Chapter 4.
C). During the exposure, the buzzer beeps continuously but the alarming indicator is off.
a). Shut off the system for 1 minutes. Then restarts the system and conduct
exposure. If the same problem happens again, open the cover of console under
power-up condition, and adjust the voltage of TP9 on MP7 board to 13.8V, but no
more than 14V.
b). Conduct exposure, if the same problem happens again, check if the fluoroscopy
current is correct according to measuring methods described in Chapter 4.
c). Replace the tube.
Page 6-18
Trouble shooting
D). Others
During the exposure, the exposure lamp shines and the buzzer beeps continuously
at the same time. Use a DC-voltage-meter to check the HDSW-EN signal on MP5.
Value of the signal is about 15V or check the signals (HDSW-EN、HDSW-SECURITY
and H6)on mother board. When H6 is on, check the HANDSWITCH signal.
Page 6-19
Trouble shooting
Fault 2:
Image looks gloomy
Reason
Contrast/Brightness parameter is improper.
Repair:
Click right button of mouse when software runs in MAIN screen, then select
Contrast/Brightness item to adjust contrast/ brightness parameters.
Fault 3:
Archive function can’t work.
Reason 1
Archive disks damaged.
Repair:
Repair or replace those devices including USB disk, CDRW disk.
Page 6-21
EverView 7500 C-Arm X-Ray Imaging system
Page 6-22
Chapter 7
Parts Replacement
EverView 7500 C-Arm X-Ray Imaging system
Electrical circuits inside the equipment use voltages that are capable of Causing
WARNING
serious injury or death from electrical shock. Use appropriate precautions.
Please use specified fuse.
If the system has power applied, power off the machine and unplug the AC
WARNING power cord from the wall outlet.
Page 7-2
Parts Replacement
Page 7-3
EverView 7500 C-Arm X-Ray Imaging system
Page 7-4
Parts Replacement
Page 7-5
EverView 7500 C-Arm X-Ray Imaging system
1 2 3 4 5 6 7 8
Page 7-6
Parts Replacement
10 11 34
17
Page 7-7
EverView 7500 C-Arm X-Ray Imaging system
43 44 45 35 36
41 39 42
Page 7-8
Parts Replacement
40 13 46 37 38
Page 7-9
EverView 7500 C-Arm X-Ray Imaging system
37 38 33
14 15
Page 7-10
Parts Replacement
31 32
18 19 20 22 21
Page 7-11
EverView 7500 C-Arm X-Ray Imaging system
47 26
24
Page 7-12
Parts Replacement
16
Page 7-13
EverView 7500 C-Arm X-Ray Imaging system
63
48
48
56
56
23
62
64
27
61
Page 7-14
Parts Replacement
54
58
57
60
59
12
65
61
49
Page 7-15
EverView 7500 C-Arm X-Ray Imaging system
51
52
50
55
66
67
Page 7-16
Parts Replacement
53
Page 7-17
EverView 7500 C-Arm X-Ray Imaging system
NOTE 2415142 kit includes 2 Power Supply fuses, 2 AC24V fuses and 2 fuse sockets.
Power supply
Fuse(6.3 A)
AC 24V Fuse(1 A)
Page 7-18
Parts Replacement
MP1
4. Power on the system, press the semi-transparent shutter close for more than 10s.
5. Check and verify the movement of the leaf is smooth.
6. Select fluoroscopy mode to take exposure, verify the shutters rotation is at the
center of the blanking circle, if not, loosen the screws, which is used to fasten the
shutter, re-position the shutters, until the shutters opening is at the center of the
blanking circle, see above Fig for reference.
7. Fasten locating screws.
8. Select the radioscopy mode, adjust the resistor POT303 on the MP1 board, let the
collimator leafs open to maximum position and the voltage measured on the two
poles of the drive motor less than 3V.
9. Power off.
Page 7-19
EverView 7500 C-Arm X-Ray Imaging system
10. Connect Iris (see below Fig for reference), and install it on its position but don’t screw
down.
11. Take off the bridge gear, close the IRIS manually, then power on the system.
12. Press the IRIS close Icon for more than 10s, and then adjust the resistor on the IRIS
until the motor don’t rotate. Keep the motor resting and install the bridge well.
13. Select the manual fluoroscope mode and take exposure, verify the IRIS field is at the
center of the blanking circle. If not move the position to adjust. Screw down the
fastening nuts.
14. Open the IRIS to maximum, take exposure and adjust the POT302 on the MP1 board
to make the X-ray field in tangent with blanking circle.
15. Place the Beam Alignment Fixture Tool (9 inch) in front of the Image Intensifier, close
the IRIS to minimum, take an exposure to measure the minimum field size is less
than 45mm, if not repeat the step14 and step 16.
16. Select MAG1 field mode, take exposure and adjust the POT2, to make the X-ray field
in tangent with blanking field.
17. Select MAG2 field mode, take exposure and adjust the POT1, to make the X-ray field
in tangent with blanking field.
18. Select the radioscopy mode, adjust the resistor POT304 on the MP1 board let the IRIS
open to maximum position but the voltage measured on the two poles of the drive
motor less than 3V.
19. Power off, take off extended board, fix pressing strip, and recover the console.
Page 7-20
Parts Replacement
Page 7-21
EverView 7500 C-Arm X-Ray Imaging system
Page 7-22
Parts Replacement
4. Connect a DC voltage meter (500V) to +PRHV and -PRHV on MP6 board, power on, the
value is about 140V.
5. Press kV+, the voltage rises smoothly. When 110kV is displayed, adjust POT116 to make
the voltage about 360V.
6. Switch to radiography mode, and then adjust POT2 to make the voltage be 370V.
7. Adjust kV according to chapter 4.4.
8. Power off. Install the strip pressing the PCB and recover the console.
Page 7-23
EverView 7500 C-Arm X-Ray Imaging system
Page 7-24
Parts Replacement
3. Rotate the brake handle to vertical position and remove the 2 screws of the fixing block.
4. Push the rotation brake handle backward with your left hand, remove the block with your
right hand.
Page 7-25
EverView 7500 C-Arm X-Ray Imaging system
5. Rotate the brake handle to make the fixing screw upward, unscrew the fixing screw and
remove the handle.
6. Grasp the left end of the brake shaft, knock on the right end of the shaft, remove the bearing
bushing on the left, then remove the shaft from the shaft from the right side. Refer to the
pictures below.
7. Get the new brake assembly, remove the handles, and put the shaft in position as the
pictures below.
Page 7-26
Parts Replacement
9. Rotate the handles to the upward position ( as pictures below), put the block in position and
fasten the screws of the block.
NOTE
While putting the block, shake the shaft with your left hand to make it in position.
10. Rotate the brake handle to lock position, then release the brake. Verify if the brake can be
operated smoothly. Release the brake, rotate the C-Arm, then lock the brake. And check if the
brake force is enough. If the brake force is not enough, follow the step 11 to adjust the brake
force; Otherwise please jump to step 12.
Page 7-27
Parts Replacement
2. Unscrew the 2 fastening screws of the brake plate, then you can lift the brake assembly.
3. Then you can replace the old assembly with a new one. After that, check and verify that if
the brake force of the orbital brake is appropriate. If not, please adjust the brake force as step
4. Otherwise you can jump to step 5.
Page 7-29
EverView 7500 C-Arm X-Ray Imaging system
4. Loosen the 2 fastening screws, fasten or loosen the 4 adjustment screws to adjust the brake
force. Rotate the screws clockwise to decrease the brake force, rotate them
counter-clockwise to increase the brake force. Check and verify that the brake force is
appropriate.
Fastening
Screw(2)
Adjustment
Screws( 4)
Page 7-30
Parts Replacement
2. Remove the limited screws on the shaft. Loosen all of the fixing screws of the brake nut.
Remove the brake handles.
Fixing Limited
screws screws
Page 7-31
EverView 7500 C-Arm X-Ray Imaging system
4. Adjust the nut on the shaft to get the appropriate brake force. Rotate it clockwise to increase
the brake force, rotate it counter-clockwise to decrease the brake force.
Nut
5. Reassemble the brake assembly and cover and fasten the screws.
Page 7-32
Parts Replacement
2. You can adjust the brake force by adjusting the 4 screws of the brake plate.
3. You can remove the old brake plate and replace it with a new one.
4. If the brake handle is not vertical when in lock state, you can loosen the nut below to adjust
the handle to vertical position.
Page 7-33
EverView 7500 C-Arm X-Ray Imaging system
Page 7-34
Parts Replacement
1 GND GND GND GND GND GND GND GND GND GND HVTN HVTN HVTGND HVTGND A
2 15V 15V 15V 15V 15V 15V 15V 15V 15V 15V HVTN HVTN HVTGND HVTGND B
3 -15V -15V -15V -15V -15V -15V -15V -15V -15V -15V C
6 IRD IRI ALTH-LC HVCOM-HC MAC KVC MAC KVC KVC HVCOM-HC HVTP1 HVTP1 F
7 OPEN CLOSE MNSYNC GENRES-HC IBS GENRES-HC GENRES-HC GENRESHC MNSYNC HVTP HVTP H
15 SWMT1 SWMT2 MAS2 MAS1 DATA6 DATA5 FILLA1 FILLA1 GND GND S
17 SRMT1 SRMT2 MAS6 MAS5 FILSM2 FILSM2 FFOND FFOD PWRSMDV HVTDV CL13N CL13 U
18 FTALR MAS7 FILC0M FILC0M GND GND DISCOM-HC RECTTR FFOD FFO V
HEATALM
19 FSMT1 FSMT2 HDSW-LC FTSW-LC GND MAMEAS2 CL13N CL13 _HC FFOND FFON W
Emergency_s
21 KEY KEY wt 0VREF KEY KEY KEY KEY KEY KEY KEY KEY KEY KEY Y
22 GND GND GND GND GND GND GND GND GND GND GND GND GND GND Z
Page 7-35
EverView 7500 C-Arm X-Ray Imaging system
Page 7-36
Chapter 8
Periodic Maintenance
EverView 7500 C-Arm X-Ray Imaging system
Page 8-2
Periodic Maintenance
After taking X-rays, the high voltage cables can retain a lethal charge, even if
WARNING you power down the system. Even though the contacts are recessed, high
voltages can arc several inches from their terminals if the cables are removed
from the connector wells. Use caution when disconnecting high voltage
cables even when the power is turned off. If you remove these connections,
immediately discharge their contacts against the high voltage tank case
(ground).
Many of the PCB's in this system contain components which are sensitive to
CAUTION
Electro-Static Discharge (ESD). Observe ESD safety procedures.
Page 8-3
EverView 7500 C-Arm X-Ray Imaging system
8.5.1 Tools
Common hand tools
Page 8-4
Periodic Maintenance
8.6 PM Checklist
Serial No: GE HuaLun Medical Systems, Inc
Start Date: EverView 7500 C-Arm X-Ray Imaging
System
GE HuaLun Elapsed Time Meter:
Periodic Maintenance Inspection
PM Schedule: Semi-Annual
Report
SEMI-ANNUAL PM
1.1Mechanical Inspections 1.2Electrical Inspections
□Brakes □Horizontal Movement □Electrical Plug/Power Cord
1.SAFETY
Page 8-5
EverView 7500 C-Arm X-Ray Imaging system
Page 8-6
Periodic Maintenance
COMMENTS
Page 8-7
EverView 7500 C-Arm X-Ray Imaging system
Page 8-8
Periodic Maintenance
Electrical circuits inside the equipment use voltages that are capable of
WARNING
causing serious injury or death from electrical shock. Use appropriate
precautions.
1. If the system has power applied, place the C-Arm power switch in the OFF position and
unplug the C-Arm's AC power cord from the wall outlet.
2. Remove C-Arm and Workstation covers as needed.
Do not perform this continuity test with the AC plug plugged into a power
WARNING
receptacle.
Page 8-9
EverView 7500 C-Arm X-Ray Imaging system
From To Spec
Pin 4 of the FootSwitch Connector Screw on the bottom of the < 0.5 Ω
FootSwitch
2. Connect the Integrated Cable to the C-Arm and verify continuity between the
following:
From To Spec
Screw on the top left corner of the WorkStation AC Power Plug Ground Pin < 0.5 Ω
Screw on the tube cover AC Power Plug Ground Pin < 0.5 Ω
Screw on the bottom of the FootSwitch AC Power Plug Ground Pin < 0.5 Ω
3. Record only the highest resistance on the PM form.
8.9.1 Power ON
1. Connect the integrated cable in the C-Arm connector and lock it. Connect the
FootSwitch cable to the C-Arm.
2. Plug the AC power cord into the wall receptacle.
3. Press the system on key located on the C-Arm control panel.
1. Take X-Ray exposure and verify the following Image Orientation functions:
A. Image rotation key rotates the image clockwise and counter-clockwise.
B. Image Invert key inverts and reinverts the image.
2. Press the field size key to change the field size: NORM, MAG1 and MAG2.
3. Using the Converging Line pairs/millimeter (lp/mm) tool on the Image Intensifier, verify the
resolution in lp/mm is equal to, or greater than, the values listed below: Indicate your
measurements on the Checklist.
Page 8-10
Periodic Maintenance
4.5″ II 6″ II 9″ II
MAG2 ≥ 2.2 MAG1 ≥ 1.8 NORM ≥ 1.2
4. Press the F3 key to verify the image on the left monitor shifts to the right monitor. Verify
the resolution in lp/mm is equal to, or greater than, the values listed above.
5. Press each of the following collimator keys and verify the correct collimator function:
A. Press the COLLIMATOR IRIS OPEN/CLOSE key.Press the Footswitch or HandSwitch and
verify that the size of the circle varies as the iris is opened and closed.
B. Press the COLLIMATOR LEAF OPEN/CLOSE key and press the Footswitch or
HandSwitch. Verify that the distance between the two lines increases and decreases
as the leaves are opened and closed.
C. Press the COLLIMATOR LEAF ROTATE key.Press the Footswitch or HandSwitch and
verify that the image displays on the Workstation monitor and that the icon rotates
as the collimator leaves are rotated.
6. Press the kVp key and verify that the technique can be adjusted manually.
7. Select Auto Fluoro mode and the NORM field size. Verify that the kVp tracks accurately as
copper filters are inserted into the field per the following ranges (9-inch). Indicate your
measurements on the Checklist.
8. Select High Level Fluoro (HLF) mode via the HLF key on the C-Arm control panel.
A. Cover the II with the lead blocker and press the Foot/handswitch to take 1s exposure.
B. Verify that the audible tone beeps “Di…..” .
C. Verify that the mA is driven higher than normal Fluoro (KV is the same).
9. Select PULSE mode via the C-Arm Control Panel.
A. Press the foot/handswitch.
B. Verify that the X-ray indicator flashes 2 times/3 seconds .
10 . Select Radiography mode and:
Take a 50 kVp 80 mAs shot and verify that Audible “Di……” beeps are heard after the
exposure and no errors occur.
Page 8-11
EverView 7500 C-Arm X-Ray Imaging system
Page 8-12
Periodic Maintenance
WARNING The following steps produce X-rays. Use appropriate safety precautions.
Page 8-13
EverView 7500 C-Arm X-Ray Imaging system
5. Calibrate the DAP: Shorten the pin 3 with pin 7 of the crystal terminal for no less than
0.5 s, the display value of the DAP will decrease 0.5%. Shorten the pin 2 with pin 3 of
the crystal terminal for no less than 0.5 s, the display value of the DAP will increase
0.5%.
Page 8-14
Periodic Maintenance
Page 8-15
EverView 7500 C-Arm X-Ray Imaging system
3. Lock the wheel brakes completely. Push the Workstation and verify that the casters lock in
place and the wheels roll in a straight line.
4. Release the Workstation's wheel brake. Push the Workstation and verify that the wheels
and casters turn freely.
8.13.2 Power ON
Complete the Power On test.
1. Apply Power to the system and verify that the Workstation and C-Arm complete the boot
sequence and are functional.
2. While pressing the Emergency Stop switch, verify that vertical column operation and
X-ray generation are disabled.
4. Reboot the system by pulling the Emergency Stop Switch out, pressing the Power Off key.
Wait for a few minutes and then back to Power On key on the C-Arm.
5. Verify that the system boots-up error free.
Page 8-16
Chapter 9
Circuit Schematics
EverView 7500 C-Arm X-Ray Imaging system
List
Page 9-2
Circuit Schematics
Page 9-3
EverView 7500 C-Arm X-Ray Imaging system
Page 9-4
Circuit Schematics
Page 9-5
EverView 7500 C-Arm X-Ray Imaging system
Page 9-6
Circuit Schematics
Page 9-7
EverView 7500 C-Arm X-Ray Imaging system
Page 9-8
Circuit Schematics
Page 9-9
EverView 7500 C-Arm X-Ray Imaging system
Page 9-10
Circuit Schematics
Page 9-11
EverView 7500 C-Arm X-Ray Imaging system
Page 9-12
Circuit Schematics
Page 9-13
EverView 7500 C-Arm X-Ray Imaging system
Page 9-14
Circuit Schematics
Page 9-15
EverView 7500 C-Arm X-Ray Imaging system
Page 9-16
Circuit Schematics
Page 9-17
EverView 7500 C-Arm X-Ray Imaging system
Page 9-18
Circuit Schematics
Page 9-19
EverView 7500 C-Arm X-Ray Imaging system
Page 9-20
Circuit Schematics
Page 9-21
EverView 7500 C-Arm X-Ray Imaging system
Page 9-22
Circuit Schematics
Page 9-23
EverView 7500 C-Arm X-Ray Imaging system
Page 9-24
Circuit Schematics
Page 9-25
EverView 7500 C-Arm X-Ray Imaging system
Page 9-26
Chapter 10
Technical Reference
EverView 7500 C-Arm X-Ray Imaging system
Page 10-2
Technical Reference
10.13 Tolerance:
The tolerance of the tube voltage ±7 %
The tolerance of the tube current ±10 %
The tolerance of the current time product ±10 %
Page 10-4
Technical Reference
Patients must be kept as far as possible from the focus, with minimum
WARNING distance of 20cm. Make sure not to operate when a plastic cover in x-ray
assembly is disassembled, which properly make the distance meet the above
requirements.
The operating table matched this equipment should be made of small
NOTE
attenuation equivalent material, otherwise, X-ray dose may be increased.
EverView 7500 C-Arm X-Ray Imaging system
Page 10-6
Chapter 11
Appendix
EverView 7500 C-Arm X-Ray Imaging system
11.1.1 Components
The X-Ray Imaging Unit consists of the following items :
1 2
5
3
4
Page 11-2
Technical Reference
AIR FLOW
CCU
Page 11-3
EverView 7500 C-Arm X-Ray Imaging system
The cable between the camera head and the CCU must be shielded to be in compliance with
the electromagnetic standard IEC 60601-1-2
The shielding of the cable must be connected to the housing of the Image Head and to the
housing of the CCU.
The cable must include a 75 Ω coaxial wire for the video.
Page 11-4
Technical Reference
1
9
2
10
3
11
4
12
5
13
6 control input 3 (UC3)
control input 2 (UC2) 14
7 control input 1
15
8 + 24 V DC power in (UE)
power 0 volt (0VP)
(1) Open the circular top cover of the Image head protecting the HVPS (2 screws).
(2) Connect the High-Voltage Power Supply interface as follows (a connection kit is
provided).
UE (15) 24 V DC
HV3-3 0VP (8)
N
UC1 (14) M1
UC2 (6)
UC3(13) M2
Page 11-5
EverView 7500 C-Arm X-Ray Imaging system
Page 11-6
Technical Reference
1
+ 24 V DC 14
2 0 V - ground
AGC off 15
3 ABC -
ABC + 16
4 Analog ABC voltage
Image rotation 17
5 Image rotation
Last Image Hold 18
6 DRF on / off
Recursive filter 0 19
7 Recursive filter 4
Recursive filter 8 20
8 Recursive filter 16
Horizontal inversion 21
9 Vertical inversion
Freeze A 22
10 Freeze B
23
11
24
12
25
13
Note : some of these contacts are unused, depending upon camera configuration.
Page 11-7
EverView 7500 C-Arm X-Ray Imaging system
4. Protective earth
Three connection points for Protective Earth are available on the RIU:
• One is located near the camera head,
• An other one is located on the HVPS,
• An other one is located on the CCU.
Page 11-8
Technical Reference
LIH (5)
Analog
ABC (16) DRF (18)
V inv (21)
ABC- (15)
RF 0 (6) Video 3
(100 or 120 Hz)
RF 4 (19)
RF 8 (7)
ABC
RF 16 (20) 10-bit D/A
converter
Image Image
processing : 50 (60) buffer Freeze A (9)
CCD Video 10-bit A/D to
sensor amplifier converter DRF 100 (120)
LIH Hz
Image
H&V inv converter Freeze B (22)
buffer
6. Video 2 output
(1) The Video 2 output displays the image present on the CCU bus. It could be
either the live image coming from the CCD (Last image Hold =1) or a stored
image coming from the image memory (Last image Hold =0).
(2) The live image displayed through the Video 2 output includes noise reduction
feature with the motion detection.
7. Video 3 output
(1) While Freeze A = 1, the Video 3 output displays the image present on the CCU
bus. This image can be horizontally or/and vertically inverted using the H&V
image inversion control inputs.
(2) The image is frozen on the Video 3 output when Freeze A = 0.
8. Video 4 output
(1) The Video 4 output works like the Video 3 output with the Freeze B control input
instead of Freeze A.
(2) The H&V image inversion control inputs act simultaneously on both channels A
and B. They are not active on frozen images.
Page 11-9
EverView 7500 C-Arm X-Ray Imaging system
Interface to
the equipment
The Freeze A and the Last Image Hold inputs used simultaneously allow to freeze the live
image (LIH).
Remark: replace Freeze A by Freeze B in case of the monitor is connected on the Video 4
output.
Page 11-10
Technical Reference
Interface to the
equipment Monitor A : Monitor B :
live image stored images
+
Last Image Hold
The Freeze A and the Last Image Hold inputs used simultaneously allow to freeze the live
image (LIH).
The Freeze B input is used to maintain the stored image on the monitor B after reading.
The Last Image Hold input is used to maintain the last image on the monitor A when used in
combination with the Freeze A input or to read a stored image.
Page 11-11
EverView 7500 C-Arm X-Ray Imaging system
11.1.4 ADJUSTMENTS
1. System adjustment location
The RIU includes 3 kinds of adjustments:
- The mechanical adjustment of the camera head (centering, focus, iris) ;
- The electrical adjustment of the Camera Control Unit (video gain, white level, video
mask ...) ;
- The electrical adjustment of the HVPS (U1, U2 and U3 voltages).
camera HVPS
head adjustment
adjustment
Camera
Control Unit
adjustment
Page 11-12
Technical Reference
Without its top cover, the radiological imaging unit is not in compliance
with x-ray DHHS regulation during the adjustments of the camera.
Page 11-13
EverView 7500 C-Arm X-Ray Imaging system
output image
of the XRII
CCD
active area
NOTE
The required torque is 4 N.m in case of this camera with image rotation (M5
screws).
Mechanical focusing
Normally, the factory adjustment should not be modified, except in case of camera
replacement. The focusing procedure is as follows:
1- Open the camera head top cover (remove the 2 M2.5 screws for cameras with
image rotation).
2 x M 2.5 screws to
3 x M5 screws for remove the camera
camera head head cover
centering
Page 11-14
Technical Reference
The camera must imperatively be turned off before removing the top cover of
NOTE
the camera head (risk of short circuit).
Two mechanical adjustments are available on the top of the camera head as shown on the
following picture:
Mechanical focusing
Iris adjustment
adjustment
close open
focus
VCCD
Test point 1
Page 11-15
EverView 7500 C-Arm X-Ray Imaging system
Iris adjustment
As it is done in the factory, for an input dose of 30 µR/s (75 kVp - 2 mm Cu - uniform X-ray
flow) in the entrance plane of the Radiological Imaging Unit, in normal mode, the iris has to be
adjusted to obtain a CCD signal of 1.3V on the test point TP1.
Use a X-Ray dosemeter to know the X-Ray dose.
In addition, the Automatic Brightness Control (ABC) was adjusted, to maintain the average level
in the center of the CCD at 1.3 volts measured on test point TP1 - see 6.3.4
1- See the previous picture for the iris adjustment location.
2- Use a small screwdriver to adjust the iris aperture.
To recover the factory iris adjustment, see "iris adjustment" on the Thales test
NOTE report. This value indicated the position of the factory adjustment in quarter of
turn with respect to the totally open position.
3. Camera Control Unit adjustment
Page 11-16
Technical Reference
1
R94 J5
2
P1 P2 P3
J2
J6
1 2 3 4
1
R157
SW2 P4
SW4
2
J4 P11
R156 R163
1
SW3
P5 P6 P8 P9 P10 P7
2 N23 ANB1
The videomask has to be adjusted internally to the image itself ; no edge must be visible inside.
Page 11-17
EverView 7500 C-Arm X-Ray Imaging system
Page 11-18
Technical Reference
WHEN USING THIS FUNCTION, SOME PRECAUTIONS HAVE TO BE TAKEN TO LIMIT THE
MAXIMUM DOSE IN ORDER TO PREVENT PATIENT INJURY.
Measuring window
The measuring window is used for the Automatic Gain Control and for the Automatic
Brightness Control. The average signal in the window is taken into account to control the
video gain in AGC mode and to control the dose (ABC).
The measuring window can be displayed on the monitor using the jumper J3.
The diameter of the measuring window can be adjusted using P5.
The measuring window is centered in the circular blanking. The H & V shift
NOTE
of the circular blanking also move the measuring window.
Analog ABC output
The signal in the measuring window (Vlev) is compared to a reference (Vset) in order to
generate the ABC signals.
The analog ABC voltage is available on the pin 16 of J0 interface connector. It varies between
0 V and + 12 V. If the video level in the measuring window (Vlev) is lower than the preset
voltage reference (Vset), the analog ABC voltage is close to 0 V. It means that the entrance
dose is too low: the X-ray generator will increase the dose and the analog ABC voltage will
goes up. If the video level in the measuring window (Vlev) is higher than the preset voltage
reference (Vset), the analog ABC voltage is close to + 12 V. It means that the entrance dose is
too high: the X-ray generator will decrease the dose and the analog ABC voltage will goes
down. The entrance dose will be stabilized at the level which maintain the analog ABC voltage
at about + 6 V.
Page 11-19
EverView 7500 C-Arm X-Ray Imaging system
set-point adjustment
Vset is adjusted using P4, ABC set-point. It has been adjusted in our factory to get + 6 V on
the analog ABC voltage output when the video level after the CCD is at 1.3V.
The following figure shows analog ABC voltage versus the entrance dose and the effect of the
6V
Entrance dose
0V
The black level of the video signal in AGC mode is adjusted using R157, black level, on the
N23 ANB1 board.
The amplitude of the video signal in AGC mode is adjusted using R163, white auto level, on the
N23 ANB1 board.
Page 11-20
Technical Reference
The selection of the J1 output signal is done using the jumper SW4 on the N23 ANB1 board.
4. High-Voltage Power Supply adjustment
HV3-3 HVPS allows the adjustment of U1, U2 and U3 voltages in their respective ranges, to
obtain the 3 different input fields on the XRII.
The HVPS voltages have been adjusted in the factory and, except in special cases, they have
not to be adjusted again.
This adjustment can be done using the adjustment slots on the side of the power supply ; to
have a better access to the potentiometers, it's necessary to dismount the HVPS without
disconnecting the cables.
Don't change
Mag2
U1 Mag1
Normal
Mag2
U2 Mag1
Normal mode :
- select this mode using the C-arm
- if necessary, adjust gently the U3 potentiometer to optimize the field of view
- adjust gently the U2 potentiometer to optimize the resolution
Page 11-21
EverView 7500 C-Arm X-Ray Imaging system
Page 11-22
Technical Reference
All the FRUs are delivered with preadjusted parameters : after change, they will be able to
provide an image which will only have to be optimized according to the field constraints.
The parameters which will be optimized will be :
- HVPS replacement : G3 voltages for fields of view, G2 voltage for resolution
- CCU replacement : video mask size and position, possibly shading correction
- camera head : focus, blemishes, centering, iris ; add CCU adjustment (see above)
- XRII replacement : all the above adjustment
G2 (black)
G1 (green)
G3 (red)
G4 (black)
PC (red)
-gauge (blue)
G5 (30 kV)
GND (green-yellow)
Page 11-23
EverView 7500 C-Arm X-Ray Imaging system
Page 11-24
Technical Reference
Page 11-25
EverView 7500 C-Arm X-Ray Imaging system
G5 (30kV) : push the cable completely in the connection hole, screw the cap carefully.
G1 to G4 cables : push the cable completely in the connection hole, screw the cap.
PC (RED)
-GAUGE (BLUE)
G3 (RED)
GND (GREEN-YELLOW)
G2 (BLACK)
G1 (GREEN)
INPUTS (24V, MODES)
G5 (30KV)
Page 11-26
Technical Reference
Page 11-27
EverView 7500 C-Arm X-Ray Imaging system
TH 7195 adjustment
After the HVPS replacement, when switching on the RIU, the pre-adjustment of the TH 7195
HVPS for 28HP2 XRII or RIU allows to obtain directly a useful image. But, since every tube and
HVPS are a bit different, it's necessary to adjust gently :
- U2 voltage to optimize the image resolution.
- U3 voltage to optimize the field of view
Use the 3 encoders at the top of the TH 7195
U1 U2 U3
Normal mode :
- select this mode using the C-arm
- if necessary, adjust gently the G3 encoders to optimize the field of view
- adjust gently the G2 encoder to optimize the resolution
it's not necessary to adjust G1
Mag 1 mode :
- select this mode using the C-arm
- if necessary, adjust gently the G3 encoders to optimize the field of view
- adjust gently the G2 encoder to optimize the resolution
it's not necessary to adjust G1
Mag 2 mode :
- select this mode using the C-arm
- if necessary, adjust gently the G3 encoders to optimize the field of view
- adjust gently the G2 encoder to optimize the resolution
it's not necessary to adjust G1
Page 11-28
Technical Reference
Adjustments :
- video mask / circular blanking : adjust its size and position according to section Video Mask/
Circular Blanking Adjustment in 11.1.4, part 3.
The video mask has to be adjusted internally to the image itself ; no image edge must be visible
inside.
- Shading adjustment : after image vizualisation, if the uniformity is good enough, don't adjust
anything ; if the edge is globally too dark, a shading correction could be done according to
section Shading Correction Adjustment in 11.1.4, part 3.
use the horizontal and vertical dome corrections gently but check the result for all the image
modes (normal mode, mag1, mag2) ; avoid to use the tilt corrections due to possible rotation
effects.
3. Camera head replacement
Disconnect the camera head cable
Dismount the camera head using the 3 screws on its basis :
Lead sheet
3 screws
Check orientation
between cable,
camera head hole
and HVPS
Dismount the camera head from the orange basis (4 screws) and mount the new one.
Page 11-29
EverView 7500 C-Arm X-Ray Imaging system
4 screws
! The camera head FRU should include a lead sheet on its top
Adjustments :
- Check the image blemishes ; if the image quality is not good enough, dismount the camera
head, clean and mount again.
- centering : using the 3 basis screws, adjust horizontally and vertically the camera head
position according to Camera Head Centering in 11.1.4, part 2.
- iris aperture : it was adjusted in factory using a typical XRII ; if iris has to be adjusted, refer to
IRIS Adjustment in 11.1.4, part 2.
Page 11-30
Technical Reference
- video mask / circular blanking : adjust its size and position according to Video Mask/ Circular
Blanking Adjustment in 11.1.4, part 3.
The videomask has to be adjusted internally to the image itself ; no edge must be visible inside.
- Shading adjustment : after image vizualisation, if the uniformity is good enough, don't adjust
anything ; if the edge is globally too dark, a shading correction could be done according to
Shading Correction Adjustment in 11.1.4, part 3.
use the horizontal and vertical dome corrections gently but check the result for all the image
modes (normal mode, mag1, mag2) ; avoid to use the tilt corrections due to possible rotation
effects.
4. XRII replacement
- Disconnect and dismount HVPS and its brackets according to section 11.1.5, part 1 HVPS
Replacement.
- Mount the same camera head according to section 11.1.5, part 1 HVPS Replacement.
- Mount the same HVPS according to section 11.1.5, part 1 HVPS Replacement.
Adjustments :
- adjust camera head and CCU according to section 11.1.5, part 1 HVPS Replacement.
Page 11-31
EverView 7500 C-Arm X-Ray Imaging system
Please read the following instructions as they are designed to prevent damage to property and
harm to the user.
Failure to follow the notes below could result in bodily harm or damage to
WARNING equipment.
• Use only a properly grounded power plug and receptacle. An improper ground may
cause electric shock or equipment damage.
• Do not remove cover (or back). No user serviceable parts inside.
• Maintenance and servicing are limited to service personnel only. High voltage parts and
heated parts inside may cause a risk of electrical shock and injury.
• If your monitor does not operate normally - in particular, if there are any unusual
sounds or smells coming from it - unplug it immediately and contact an authorized
dealer or service.
• Never insert anything into the monitor.
• Do not replace fuse.
Failure to follow the notes below could result in bodily harm or damage to
CAUTION equipment.
• Do not cover the vents on the monitor cabinet. Bad ventilation may cause a breakdown
or fire.
• Put your monitor in a location with low humidity and minimum of dust.
• Place the monitor on a flat and stable surface. The monitor can cause injury by falling.
• When cleaning the monitor case or the surface of the CDT, wipe with a slightly
moistened, soft fabric. Do not clean monitor with water. Use a water-diluted mild
detergent.
• Do not spray detergent directly on the monitor.
• Do not place any heavy objects on the monitor.
• For each hour of looking at the monitor, you should let your eyes rest for 5 minutes.
• Do not use or store inflammable substances near the monitor.
• Keep the monitor away from any magnetic substances.
• Please transmit signals according to the broadcast standard shown in the specification.
Page 11-32
Technical Reference
11.2.1 Introduction
• This monitor is a very high brightness, high-resolution digitally controlled autoscan
monochrome monitor.
• This monitor, specifically designed for medical applications, gives unique performances
in term of brightness, resolution and spot size.
• A rich set of functions allows quick installation, flexible user preference customization
and easy maintenance.
• Almost all the internal functions of the monitor are digitally controlled, and they can be
stored in a resident memory that may contain up to 18 different formats, 2 of which are
factory preset.
• The monitor complies with the Energy Star Computers Program initiated by the EPA.
• The monitor will automatically synchronize within a wide range of scanning frequencies
and determine the specific operating mode.
• Monitor behaviors, image quality and screen geometry adjustments can be changed or
selected via the hardware OSD by means of the front panel push buttons.
• The monitor provides special features as:
Full compatibility with all mono and color video cards, progressive and interlaced
timing.
Keyboard lock facility.
Page 11-33
EverView 7500 C-Arm X-Ray Imaging system
(5) If signal format of system is SOG (sync on green), Signal Selection (switch① in figure
1) switches to Video.
(6) If signal format of computer is Separate Sync, Signal Selection (switch① in figure 1)
switches to VGA.
4. Setup Daisy Chain Selection
(7) If you use the sole monitor, Daisy Chain Selection (switch④ in figure 1) switches to 75
Ω.
(8) If you use some monitors in Daisy Chain configuration, the units shall be connected as
shown in figure 2. Daisy Chain Selection of the end monitor switches to 75Ω, and the
other monitors switch to High.
Figure1
Connectors on the back of the monitor
Ω Ω Ω
Figure 2
Daisy Chain Configuration
Page 11-34
Technical Reference
Power On/Off
• Press “ ” to activate OSD menu, then continued press “ ” button more than 2s to
lock keyboard.
Page 11-35
EverView 7500 C-Arm X-Ray Imaging system
Page 11-36
Technical Reference
Page 11-37
EverView 7500 C-Arm X-Ray Imaging system
Page 11-38
Technical Reference
• Technical data
Nominal high voltage 110 kV
Nominal anobe input power 2.2 kW
Nominal input power (10 minutes ON) 0.6 kW
Maximum anode heat content 53 kJ
Target material Tungsten
Target angle 10°
Small focal spot 0.6 x 1.4 mm
Large focal spot 1.4 mm
Filtration equiv. 0.55 mm Al
Anode Stationary
Typical filament specification 4.2 V, 3.5 A
Maximum filament current 4.6 A
Weight 0.65 kg
Page 11-39
EverView 7500 C-Arm X-Ray Imaging system
Page 11-40
Technical Reference
Page 11-41
EverView 7500 C-Arm X-Ray Imaging system
Page 11-42
Technical Reference
Page 11-43
EverView 7500 C-Arm X-Ray Imaging system
Page 11-44
Technical Reference
The X-ray generator assembly consists of high voltage transformer, high voltage rectifier
and X-ray tube in one pressurized tank. It can be used in EverView 7500 C-arm X-ray
imaging system.
Page 11-45
EverView 7500 C-Arm X-Ray Imaging system
1. Technical specification
Power:
Standard Fluroscopy(small focal spot)
110kV×4.0mA=440W (maximum)
HLF(small focal spot, pulse)
110kV×8.0mA=880W (maximum)
Radiography(Large focal spot)
110kV×20mA=2.2KW (maximum)
Nominal high voltage 110kV
Permanent filtration : 2.5mmAl
Isolation: oil
Heat protection: 55°C temperature switch
Focus mark: dot
Maximum thermal capacity: 800kJ (1040kHU)
Weight: 23kg
Maximum heat disappation 177W
Maximum X-ray field(on II input screen) φ210mm
Heating/cooling charcteristics:
Page 11-46
Technical Reference
2.Configuration
Time (min)
Reference Axis : The axis from focal spot to the center point of the Image Intensifier.
Note:The focal spot marked in the diagram is the midpoint between large focal spot and
small focal spot, which the windage is no more than 2mm from the benchmark
axes.
3.X-ray tube
Refer to section 11.3.
Page 11-47
EverView 7500 C-Arm X-Ray Imaging system
Fig 1
X-ray tube
Fig 2
Page 11-48
Technical Reference
b) Connect 9-core connector to glower cable and connect 4-core connector to high voltage
cable.
9-core connector
4core connector
D-sub9 male
D-sub9 female
c) Connect D-sub9 male to Iris and connect D-sub9 female to parallel shutter, please see
picture below for details.
Iris
D-sub9 male
Parallel shutter
D-sub9 female
Page 11-49
EverView 7500 C-Arm X-Ray Imaging system
Page 11-50
Address
SERVICE OFFICES
For service issues, contact the Service Office for your country.
AMERICAS
WAUKESHA, WI 53188
MILWAUKEE, WI 53201
LATIN AMERICA
GE Medical Systems Tel: +1–305–497–1200
USA
ASIA
NIPPON (Japan)
GEMSA Headquarters Tel: 81–426–48–2940
REST OF ASIA
GE Pacific Pte. Ltd (GEMS Asia) Tel: +65–291–8528
SINGAPORE 0316
EverView 7500 C-Arm X-Ray Imaging system
EUROPE
BELGIE/BELGIQUE (Belgium)
GE Medical Systems Benelux NV/SA (NL) Tel:+32 2 644 09 38
DANMARK (Denmark)
GE Medical Systems Danmark Tel: +45 45 51 00 55
DK–2840 HOLTE
DEUTSCHLAND (Germany)
GE Medical Systems Deutschland GmbH & Tel: 49–6102–36–0
Martin–Behaim–Strasse 10
Beethovenstrasse 239
42655 Solingen
ESPAׁA (Spain)
GE Medical Systems Espaסa Tel: +34 1 663 25 00
E–28100 ALCOBENDAS
FRANCE (France)
GE Medical Systems Tel: +33 (0)1 30 70 40 40
BP34
HELLAS (Greece)
GE Medical Systems Hellas Tel: +30 1 93 24 582
ITALIA (Italy)
GE Medical Systems Italia Tel: +39 02 64 22 01
20126 MILANO
NEDERLAND (Netherlands)
GE Medical Systems Nederland B.V. Tel: +31 73 6 457 457
NL–5231 DD S HERTOGENBOSCH
ײSTERREICH (Austria)
GE GesmbH Medical Systems Austria Tel: 0660 8651 (geb hrenfrei)
POLSKA (Poland)
GE Medical Systems Polska
PORTUGAL (Portugal)
GE Medical Systems Portuguesa S.A.
ROSSIYA (Russia)
GE Medical Systems Tel:+7 095 935 72 41
Kosmodamianskaya nab. 52, Bldg 1, 6th Floor Fax:+7 095 935 73 46 and 48
Tlx:613020 GEMED SU
SVERIGE (Sweden)
GE Medical Systems Tel: +46 8 457 95 20
S–113 85 STOCKHOLM
SCHWEIZ/SUISSE (Switzerland)
EverView 7500 C-Arm X-Ray Imaging system
CH–6010 KRIENS
T RKIYE (Turkey)
GE Medical Systems Turkiye A.S. Tel: +90 212 75 5552
Gayretteppe
ISTANBUL
UNITED KINGDOM
IGE Medical Systems
SLOUGH
Imagination at work