Deflection Temperature and


Vicat Testers
Models 603 HDTM and 303 hdtm
Fig 1. Model 603 HDTM shown with
two test stations, one for deflection
temperature and one for Vicat test-
ing. The stations are raised for easy
and rapid specimen insertion prior
to a test and for specimen removal
at the end of the test.

Fig 2. Model 303 HDTM shown


with optional third test station, for
deflection temperature and / or
Vicat testing. Model here shown
with three stations, gantry lowered,
and optional cooling connections.

The Models 603 HDTM and 303 HDTM oil bath by pneumatic pistons and the test
are the latest generation of Tinius Olsen’s can start.
digitally controlled Automatic Deflection The start can be initiated by either the
Temperature/Vicat test equipment, with an handheld terminal or directly from a PC
automated testing sequence that running Tinius Olsen’s HDV software.
proceeds according to user defined control Once started, the test is completely auto-
and configuration parameters. matic and is performed according to the
The basic model 603 HDTM includes user defined program. Upon test comple-
a six station bath, two test stations, and tion, power to the heater is removed and
a handheld terminal. Up to four more the cooling cycle is initiated.
stations can be added and each test Test parameters are entered through the
frame can be configured with optional numeric keypad on the handheld terminal. Fig 3. Close-up
accessories for either Vicat or deflection Entries for each station include test type, of test station,
temperature testing, including both 100 deflection/penetration, span, specimen which is
mm edgewise and 64 mm flatwise deflec- dimensions, and stress or force. Starting configured
tion temperature, as well as the four inch temperature, soak time, maximum temper- for a heat
span test. Additionally, the 603 HDTM is ature and rate of temperature rise are also deflection test.
ready to be linked to Tinius Olsen’s HDV entered for control of the bath unit. These
software so that the PC can configure the configurations are stored and may be used
ples. In the event that a specimen is dis-
controller, collect the test data, generate a for future tests.
lodged from the station during the course
test report, and save the results. The 603 HDTM has a built-in heat
of the test, it will be safely caught in the
The 303 HDTM shares these same fea- exchanger so that once the test is com-
specimen basket, keeping the heat transfer
tures of the 603 HDTM, however it has a plete a rapid cooldown is
medium as clean as possible.
maximum of three stations. automatically initiated. Using regular mains
The handheld terminal also shows a
On both machines, all test stations are water, the temperature of the silicon oil can
continuous display of temperature and
pneumatically raised from the bath at the be reduced from a maximum of 300°C to
deflection/penetration for each station
touch of a button on the handheld terminal, 20° above the water inlet temperature in
throughout the test.
allowing easy placement of the test speci- approximately 20 minutes. The 303 HDTM
mens. Once the specimens are loaded has an optional cooling system.
into the respective test positions, all the Once the cooling cycle is complete, all
stations are lowered collectively into the the test stations can be collectively raised
allowing easy removal of the tested sam-

2
Key Features Technical Specifications
• Conforms to ISO 75, ISO 306, ASTM MODEL 603 303
D648, and ASTM D1525 MAXIMUM NUMBER
OF STATIONS 6 3
• Fully automatic control of entire
TEMPERATURE oC
test cycle 23o to 300o
RANGE
• Bath has port with an exhaust fan to
TEMPERATURE oC
remove interior oil fumes 50o per hour or 2o per minute
RAMP
• Air bearing-guided loading rods for TEMPERATURE
virtually friction-free load application DISPLAY oC 0.1
• Electronic transducers integrated into RESOLUTION
the loading rod assemblies for 0.001 TEMPERATURE Platinum RTD located adjacent to the load application point
mm (0.0001 in) resolution of deflec- SENSOR at each station
tion or penetration DEFLECTION /
• Loading nose and rod assemblies PENETRATION LVDT
MEASUREMENT
provide 76 grams nominal load
for ISO 75 ‘flatwise’ deflection DEFLECTION /
PENETRATION
temperature tests on 4 mm x 10 mm mm 0.001
DISPLAY
specimens at 0.45 Mpa stress RESOLUTION
• Built-in heat exchanger for rapid Maximum of 20o above cooling water temperature in 20
COOLDOWN RATE **
system cooldown (optional on Model minutes.
303 HDTM) TEMPERATURE Independent dual systems using thermostatic switch in bath
• Pneumatic station lift for easy SAFETY LIMIT and keypad selectable software limiting
specimen insertion and removal DIMENSIONS
mm 969 x 769 x 572 826 x 639 x 585
• Automatic correction for thermal (wxdxh)

expansion of test frames in 38.5 x 30.5 x 22.5 32.5 x 25.5 x 23


• Built-in specimen basket to catch WEIGHT kg 132 87
any dislodged specimens lb 290 190
• Accessories available include
Heat
additional test stations, deflection REQUIRED 16 liters (3.7 US gallons)
Transfer 11 liters (2.9 US gallons)
UTILITIES min
temperature loading noses, Vicat Medium
loading noses and needles, weights, Water Water supply for cooldown
64 mm span supports for ‘flatwise’
Clean Air Dry air filtered to 50 microns at 40psi (2.7 bar) min
deflection temperature testing (the
stations are predrilled to accept Power 220 +/- 10%, 50/60 Hz, 1 phase, 4.5 kW
these supports)
** Note - Cooling is an option on the Model 303, but standard on the Model 603

Operating Environment:
Environmental Temperature Range: 60 to 100°F (15 to 38°C)
Storage Temperature Range: 14 to 115°F (-10 to 45°C)
Humidity Range: 10% to 90% non-condensing, wet bulb method
Power: standard optional voltages 220/240 VAC, 50 – 60 Hz; power must
be free of spikes and surges exceeding 10% of the nominal voltage
Notes: All models conform to all relevant European CE Health and Safety Directives.
Specifications subject to change without notice.

Fig 4.
Close-up of
a test station
showing
the Vicat
configuration.

Fig 5. Both the Model 603 HDTM and 303 HDTM can be controlled by a PC.

3
HDV Software

• For easy testing control and data stor-


age using a PC.
• User selected reporting and
exporting formats including report
generation in HTML format.
• Built-in SPC programs for X bar, R, and
frequency distribution charts and his-
tograms.
• Produces temperature vs
penetration (or deflection) for
individual stations in real time.
• Test mode allows configuring,
running, and saving of tests and
results.
• Recall mode permits viewing of
previously saved results and performs
database maintenance.
• Demo mode simplifies operator train-
ing.
• Two levels of password protection.

1065 Easton Road


Horsham, PA 19044 USA
(215) 675-7100
Fax (215) 441-0899

6 Perrywood Business Park


Honeycrock Lane, Salfords
Redhill, Surrey RH1 5DZ England
Contact Your Local Representative: +44 (0) 1737 765001
Fax +44 (0) 1737 764768

www.TiniusOlsen.com
info@TiniusOlsen.com

© 2006, Tinius Olsen 09/06 03-286 Printed in U.K

Você também pode gostar