SPESIFIKASI PRODUK

PROFESSIONAL BLOOD PRESSURE MONITOR HBP-1100


Manufactured by OMRON HEALTHCARE Co.Ltd
53. Kunotsubo, Terado-cho
Muko, Kyoto,617-002 JAPAN
Name of Medical device Professional Blood Pressure Monitor HBP-1100
Production Facility OMRON (DALIAN) CO,LTD
Measurement Technology Pulse rate accuracy
Measurement Method Oscillometric ( full automatic ) & Auscultation
method ( semi automatic by stateschope )
Measurement Process Dynamic Linear Deflation Method
Pressure Display range 0 to 300 mmHG
Pressure display accuracy within ± 3mmHg or 2 %
NIBP Measurement range SYS 60 to 250 mmHg, DIA 40 to 200 mmHg
PULSE 40 to 200/min
NIBP Accuracy Maximum mean error within ± 5mmHg
Maximum standard deviation within 8 mmHg
Pulse rate Accuracy within ± 5 % of reading
Dry cell battery Input Voltage : AC 100 V to 240 V
Weight Main unit : approx. 510 g ( not including accessories
and optional part)
Safety Standards IEC60601-1:1988+A1:1991+A2:1995
Medical electrical equipment-part1 : General
requirements for safety
Dimension Main unit :
5.12 x 6.89 x 4.72 (inch ) 130 x 176 x 120 ( mm) (W x
H x D)AC Adapter : 2.17 x 0.98 x 2.76 (inch) 56 x 26 x
70 (mm) (W x H x D )
Environemnt conditions Temperature range : 6 to 40° C (41 to 104°F)
humidity range : 15 to 85% RH (not condensed)
Storage and transportation Temperature range : -20 to 60 C, (-4 to 140°F)
humidity : 10 to 95 % RH (not condensed)
Display 7 segment LCD
Reference Standard ANSI/AAMI SP-10:2002+A1:2003+A2:2006/(R)
2008 ISO81060-2:2009
AC Adapter Input voltage range : AC 100 V to 240 V
Rated current : 0.5 A
Frequency : 50/60 Hz
Output voltage range : DC 6 V ±5%
Degree of protection Type BF
Package Storage / Paket pembelian Main unit HBP-1100, Manset ukuran M, AC adaptor,
instruction manual, warranty card

1
INFORMASI TAMBAHAN :
Sertifikasi :
EN ISO 9001 : 2008
EN ISO 14000
EN ISO 13485:2003 +AC :2009

IZIN EDAR : KEMENKES RI AKL 20501511489