Specifications
2
times. Unlike conventional simultaneous systems
Radial Mode Axial Mode VDV Mode SVDV Mode which sequence these read steps, AIT can conduct
SVDV these readings at the same time providing true
configuration simultaneous measurement
VDV
configuration
• The VistaChip II features the fastest available read-
RV
configuration out speed of any spectroscopic CCD detector – with
a 1 MHz clocking speed to process pixels. Total
Easy access to pre-optics windows for user read-out time for a full illumination of all pixels on
serviceability/maintenance. Computer-optimized the detector is approximately 0.8 seconds. Duplex
echelle optical design uses a single entrance slit read-out circuitry is provided on both sides of the
and focuses the echelle image onto a single CCD detector, halving read-out processing time
detector. No moving optical parts to ensure lowest
detection limits and maximum stability. 400 mm • CCD detector features anti-blooming protection on
focal length polychromator is thermostatted to 35 ºC each individual pixel, enabling the simultaneous
for excellent stability. Features a CaF2 prism cross measurement of trace level analytes in the presence
disperser and echelle grating (94.74 lines/mm) creating of nearby intense signals
an echellogram of 70 orders projected onto the CCD
detector, which is custom-designed to exactly match • The VistaChip II is hermetically-sealed which means
the image produced by the echelle optics. Mass flow it requires no argon purge to achieve excellent
controlled polychromator purge (argon or nitrogen), with sensitivity in the UV range of the spectrum. This
easy access user replaceable filter. feature also reduces time from plasma ignition to
analysis as there are no delays from waiting to
CCD detector purge air from the detector
3
• Fast Automated Curve-fitting Technique (FACT) for • Windows 7, 64-bit compatible
online spectral deconvolution of complex spectra.
Inter Element Correction (IEC) technique also • Software interface available in English, Japanese,
included Simplified Chinese, French, German, Italian,
Spanish, Portuguese, Polish and Russian
• Calibration routines for multi-element external
calibration and method of standard addition • Simplify diagnosis of instrument and utility
supply faults with enhanced diagnostic software.
• Calibration reslopes eliminate the need for full Actionable error messages, a simple graphic display
recalibration dashboard and automated instrument self checks
on all sub modules assist to maximize instrument
• User-customizable Quality Control Protocols (QCP) uptime
designed to meet US EPA and other international
compliance standards • Optional software available to assist in achieving
compliance to the US FDA’s 21 CFR Part 11
• Perform rapid qualitative screening and semi requirements for audit trails, electronic signatures
quantitate all elements in each sample using and access privileges
IntelliQuant
Optional Pro software pack:
• Unique IntelliQuant 'heat map' results display
paradigm provides rapid and intuitive visualization • Select 3rd party autosampler support
of results, at a glance, for all elements in each
sample • Tracking of nebulizer back pressure and argon
emission intensity for monitoring and trouble
• Fully editable sample label list with optional shooting sample introduction system
customer and batch label fields
• Control of the AVS 6 and 7 accessories
• Full control of all fully integrated Advanced Valve
systems (AVS 4, 6 and 7). These switching valves • Rate generated QCP
will increase sample throughput, save gas costs and
reduce your consumeable costs by extending the • Live export of data to spreadsheet
lifetime of torches, spraychambers, nebulisers and
pump tubes • Oxygen addition
• Autosampler rack and tube positions can be edited Warm up time from standby mode of <20 minutes from
for true random access sampling plasma ignition.
4
Signal stability Application-specific sample introduction options
Typically stable to less than 1% RSD over 8 hours A range of optimized torches and sample introduction
without internal standardization or any form of drift kits is available for:
correction.
• aqueous samples
Typical resolution
• organic solvents
Element Wavelength (nm) Resolution (pm)
As 188.980 <7
• high salt/matrix samples
Mo 202.032 <7
• samples containing hydrofluoric acid (HF)
Zn 213.857 <7.5
Pb 220.353 <8 You can minimize costs with demountable torches,
Cr 267.716 <9.5 designed for easy maintenance, fast changeover, and
Cu 327.396 <13
economical operation.
Ba 614.172 <34
Installation requirements
Accessories and peripherals System installation
Agilent offers a full range of configurable accessories For details of ICP-OES installation requirements refer to
and peripherals for the 5110 ICP-OES, including: the Agilent 5110 ICP-OES Site Preparation Guide.
AVS Advanced Valve System Dimensions
The fully integrated AVS (optional) simplifies setup Width Depth Height Weight
and use while delivering ultra-high sample throughput
800 mm 740 mm 940 mm 106 kg
without compromise to analytical performance.
31.5 in 29 in 37 in 233 lb
Available in 4 port, 6 port and 7 port options
Exhaust requirements
SPS 4 Autosampler
The 5110 ICP-OES incorporates corrosion-resistant
Designed to meet the needs of high-throughput materials, and an on-board fan maintains internal
laboratories requiring a fast, compact, high capacity positive pressure to keep acid vapors out. Exhaust flow
(360 samples) reliable autosampler. minimum requirements are 2.5 m3/min (88 ft3/min).
User serviceable course dust/particulate filter included
Multimode Sample Introduction System (MSIS) on air supply inlet.
5
Access and serviceability Customer support policy
All connections for power, gas, water and Warranty
communications are accessed from the side rather than
the rear of the instrument. Self-diagnosing electronics Twelve (12) months, though this may vary according to
constantly monitor instrument status, allowing rapid location.
identification of component health issues.
Agilent service guarantee
Power requirement
If your Agilent instrument requires service while
2.9 kVA, single phase mains input voltage of between covered by an Agilent service agreement, we guarantee
200-240 VAC (50-60 Hz), drawing a maximum of repair or we will replace your instrument for free. No
15 amps. other manufacturer or service provider offers this level
of commitment to keeping your lab running at maximum
Instrument qualification services productivity.
www.agilent.com/chem
Agilent shall not be liable for errors contained herein or for
incidental or consequential damages in connection with
the furnishing, performance or use of this material.